You are on page 1of 52

SYSTEM CLINIC / DELUX

EN User manual
SYSTEM CLINIC / DELUX

User manual

Komorów 2006
PART A

Prior to use of Viofor JPS Clinic/Delux apparatus, please read the present manual.
The User Manual for Viofor JPS apparatus is not meant as the main source of medical
information. Any information necessary for optimal use of the apparatus can be found in
professional publications and at Symposiums organized by Med & Life. It is also advis-
able to seek medical advice available on Med & Life Medical Infoline, share information
and experience with other MD’s and to visit the site: www.medandlife.com

Please read the User’s Manual!

NOTICE:
Please read the enclosed documentation!

Do not use during pregnancy


(see: contraindications)!

Do not use if wearing electronic implants


(see: contraindications)!

Beam of intensive red or/and infrared radiation! Protect your eyes!


Do not look at the light!
Direct eye contact with the applicator can cause risk
as in case of class 3A laser.

Fragments marked in italics present viofortherapy theory elements and are mainly
for doctor’s use. You do not need to know these parts of the manual in order to be able
to operate the apparatus.
Application of Viofor JPS System Clinic/Delux

Application of Viofor JPS System


Clinic/Viofor JPS System Delux with light energy.
INTRODUCTION
User Manual for Viofor JPS System Viofor JPS System Clinic with light energy
Clinic and Delux with light energy has been or Viofor JPS System Delux (further referred
designed for a wide range of users dealing to as Viofor JPS Clinic/Delux) is a unit con-
professionally with health care, i.e. doctors sisting of: a controller with basic and/or ad-
of various specializations (including dentists ditional accessories. Unique design solutions
and veterinarians) as well as physiotherapists enable a wide application of therapeutic prop-
and rehabilitation specialists, who use or are erties of variable magnetic fields. Viofor JPS
interested in application of variable magnetic Clinic/Delux provides a selection of physical
fields in medicine. parameters of variable magnetic fields used
In recent years, medical world has been in:
seeking new solutions in diagnostics and 1. magnetostimulation in JPS system
therapy, especially noninvasive ones, as well 2. magnetostimulation in JPS system paired
as trying to develop treatment methods that with light energy
have not been used sufficiently. 3. magnetotherapy* (only for Viofor JPS Sys-
Dynamic development observed in mathe- tem Clinic).
matics, physics and technology (especially
electronics) has affected a new quality physi-
cal medicine. Thanks to modern technologies * NOTICE. With asterisk are marked all
applied in medical devices the application and the information referring only to Viofor JPS
effectiveness of variable magnetic fields in System Clinic.
medicine has also been expanded.

1 CHARACTERISTICS OF VIOFOR JPS CLINIC/DELUX


– APPARATUS FOR MAGNETIC FIELD THERAPY
1.1 Usage Viofor JPS System Clinic/Delux can be
Viofor JPS Clinic/Delux is used for the fol- used:
lowing types of therapy: - in medical and physiotherapy clinics
- magnetostimulation of the whole body - in stomatology
or its parts - in rehabilitation
- magnetostimulation paired with light ener- - in sport medicine
gy of selected body parts - in veterinary.
- magnetotherapy* of selected body parts.

Notice: Viofor JPS System Clinic must be operated by qualified medical personnel!

2 TYPES OF THERAPY
2.1 Magnetostimulation
2.1.1 Characteristics pulse fields of complex impulse shapes and
of magnetostimulation signal structure of multi-peak frequency
Magnetic fields in magnetostimulation spectrum. Basic frequencies are within 180
generated by Viofor JPS System are im- – 195 Hz, frequencies of impulse packets

3
Application of Viofor JPS System Clinic/Delux
are within 12,5 to 29 Hz, packet groups due to penetration through the cell mem-
within 2,8-7,6 Hz, and series 0,08 – 0,3 Hz, brane to the cell interior of positive ions,
whereas median induction of the generated e.g. Na+ . In order to reverse this process,
fields are significantly lower of those used the cell needs energy, which can be ob-
in magnetotherapy. tained from ATP hydrolysis.
Viofor JPS Clinic/Delux generates variable 2. The biochemical effect is based on an
magnetic fields, which cause in the organism increased enzymatic activity as well as
changes similar to those caused by physical reduced processes related with ATP.
activity (physical exercise, physical activity), 3. The bioenergetic effect is a factor that
therefore support biological regeneration. stimulates the nutrition and growth of cells
Both mental and physical regeneration is and regulates intercellular processes lead-
tightly connected with exercise. Variable ing to regeneration of the organism.
magnetic field plays a similar beneficial role. The main biological effects of magneto-
Magnetostimulation stimulates in the orga- stimulation mechanisms are actions:
nism electromotor forces of such properties, - regenerative
which enable homeostasis, and in case of - vasodilatatory
its disturbance stimulate return to a normal - angiogenetic
state. - stabilizing cell membranes
Biological effect of variable magnetic field - anti-inflammatory
generated by Viofor JPS Clinic/Delux is the - relaxative
result of: - anti-spastic
1. electrodynamic effect of such field on ion - analgesic.
currents in the organism
2. magnetomechanic effect of the magnetic Analgesic action takes place mainly
field on particles with uncompensated due to an increase in the secretion of endog-
magnetic spin enous opiates from the ß-endorphin group,
3. ionic cyclotron resonance of cations and a substance responsible for the increase in
anions in the body fluids. pain threshold. The effect of analgesic action
Body fluids i.e. blood, the lymph and inter- occurs not only during exposure to magnetic
cellular fluid constitute so-called second mode fields, but is also observed after exposure is
conductors of the ionic character of electrical discontinued. This determines the biological
conductivity. hysteresis of magnetic field effect. Precisely
The essence of the Viofor JPS System is such a mechanism of analgesic action on the
the optimal choice of two different electrical internal opioid system confirms an antagonist
(and thus magnetic) signals as well as their of opiates blocking this effect (Naloxone).
mutual combination, in order to achieve the The basis for regeneration effect is main-
aforementioned three fundamental mecha- ly the intensification of the processes of oxy-
nisms of variable magnetic field absorption gen utilization and tissue respiration as a re-
by the living matter. Electromagnetic radiation sult of intensified diffusion as well as oxygen
may affect an organism in various ways. De- capture by hemoglobin and cytochrome. The
pending on the amount of absorbed energy, increased capture of oxygen is associated
a thermal effect and extra thermal effects with a stimulation of the processes of tissue
occur. Extra thermal effects: bioelectric, bio- respiration, DNA synthesis and acceleration
chemical and bioenergetic – due to the very of the mitotic cycle. Thanks to this, magnetic
low value of magnetic induction – are the fields treat trophic lesions of the shank, acce-
most essential in magnetostimulation. lerate the healing of ulcerous niches, improve
1. The bioelectric effect causes normaliza- the cellular metabolism of the central nervous
tion in the potential of the cell membrane. system and accelerate regeneration of da-
In pathological cases, the potential drops maged peripheral nerves.

4
Application of Viofor JPS System Clinic/Delux
Very weak currents generated due to inter- 2. soft tissue:
action on piezoelectric substances e.g. colla- - extrajoint rheumatism - so-called
gen, stimulate the activity of osteoblasts. It is fibromyalgia
aided by intensification of vasodilatatory ef- - traumas of soft tissue
fects. It has also been observed that a mag- - conditions after injury(cutting) of tissue
netic field inhibits demineralization processes continuity (including post-surgical)
of the bones and, at the same time, increases 3. migraines
their mineralization – which is evaluated by
densitometry. II. Regeneration action
Anti-inflammatory effect is connected 1. osseous system:
with the stimulation of formation of c-AMP - osteoporosis, particularly post-menopausal
and E prostaglandin. Prostaglandin influences (concerns prophylaxis as well as therapy)
the accumulation of c-AMP, which decreases - Sudeck’s syndrome
the secretion of inflammation mediators from - conditions after bone fracture
neutrophiles, basophiles, mastocytes and and breakage
lymphocytes. - conditions after operations
In case of the nervous system, magnetic on the osseous system
fields also improve the interneuron conductivity - conditions after operations
and modulation of neuron activity as well as in which anastomosis has been carried
the pineal gland’s twenty-four hour rhythms of out with screws or nails
melatonin secretion, which is called “the scav- - pseudoarthrosis
enger of free radicals” responsible, among 2. soft tissues:
others, for the aging process. - conditions after injury with
The variable magnetic fields generated by or without damage to skin continuity
the Viofor JPS clearly modify the amplitude - burns
value of alpha and theta wave rhythms in proper - bedsores
directions, characteristic for the physiological
state of relaxation - objectively measurable and III. Action improving peripheral circulation
indicated by value of the demobilization of the 1. improper local blood flow in the limbs:
autonomous system. The results of psycho- - in diabetic angiopathy
physical testing, as a part of psychotherapy - in atherosclerotic angiopathy
of emotionally immature people point to mag- - in crural ulceration
netostimulation as a method of choice for these 2. accelerated absorption of haematomas
people. What is more, it is a non-invasive meth- 3. decreased lymphatic oedema/swelling
od applied equally in medicine as in psychology.
IV. Anti-spastic action
2.1.2 Indications for magnetostimulation - conditions after apoplexy
Indications of magnetostimulation as a - degenerative diseases of the nervous
supplementary therapy (or by choice) include: system, accompanied by spasticity

I. Analgesic action V. Relaxative action


1. osseous-joint system: - neuroses
- degeneration of the spine - sleep disorders
and osseous-joint system - prophylactic calming action,
of the upper and lower limbs to counteract everyday stress
- rheumatoid inflammation of the joints - stress in the common meaning
- stiffening inflammation (rigor) of the word, and particularly
of the spinal joints after long-term psychic/mental tension
- joint injuries concentration disorders

5
Application of Viofor JPS System Clinic/Delux
VI. Action decreasing oxygen deficit through magnetotherapy, the contraindications should
effect on the processes of: be the same as in case of magnetotherapy:
- aerobic tissue respiration - pregnancy
- anaerobic tissue respiration - active neoplastic disease
- active tuberculosis of the lungs
VII.In dentistry: - bleeding from the alimentary tract
- in conservative teeth treatment, e.g. - serious infections of a viral, bacterial
pulp biological treatment, periodontal and myframe bedic origin
membrane irritation, in complications - the presence of electronic implants
after endodentic treatment - conditions after organ transplants
- acute and chronic inflammation Pregnancy is a state in which the dynam-
- damage of soft tissue ics of forming new tissues is the greatest. As a
- post traumatic states, e.g. impaired rule, apart from necessary actions, everything
healing of osseous system else should be avoided during pregnancy. Until
- after surgery (24 h), now, the influence of magnetic fields on the fe-
e.g. after tooth extraction tus is unknown. Due to the lack of information
- after surgical treatment on the effect of magnetic fields on the fetus,
of parodontal diseases the application of variable magnetic fields in
- complications after surgery pregnancy should be considered inadvisable.
(e.g. alveolar osteitis), after anesthesia, Neoplastic disease in the majority of
removal of scars or damaged nerves cases has been of a vague etiology, and the
- diseases of mucous membrane natural history of neoplastic diseases is also
in oral cavity, gum diseases, diseases known only in some cases. Experimental data
of small salivary glands regarding the effect of magnetic fields on neo-
- treatment of parodontal diseases plastic cell cultures and animals with an active
(supplementary anti-inflammatory neoplastic disease are contradictory. In ac-
treatment, fighting pain, tissue regen- cordance with the principle primum non nocere
eration, surgical scar removal) the application of variable magnetic fields in
- supplementary orthodontic treatment a neoplastic disease should not be used al-
(during braces treatment) though much data indicates a positive effect of
- supplementary prosthetic treatment magnetic fields on neoplasms.
(after application of prosthetics). Active tuberculosis of the lungs is
a contraindication to the application of mag-
2.1.3 Contraindications netic fields, which results from their interac-
to magnetostimulation tive mechanisms with tissue. The influence
Contraindications concern the application of magnetic fields on the utilization of oxygen,
of magnetostimulation as a method itself, and as well as on the blood flow, is beneficial.
not the Viofor JPS Clinic/Delux apparatus. Koch’s bacilli are aerobes, therefore, based on
Emitted fields are a contraindication to ap- knowledge of the biological effects of variable
plication in an existing disease – whereas magnetic fields, in case of TBC, they should
they are not a disease-developing factor. The not be applied.
diseases mentioned in contraindications have Bleeding from the alimentary tract
a source in other factors. At present, there is should be divided into massive hemorrhaging
no information in literature describing the con- – from the varices of the gullet and peptic ul-
traindications to magnetostimulation. How- cers – and bleeding accompanying diseases
ever, when taking into consideration the com- of the gastritis type. In the case of intensive
paratively short application time of magneto- bleeding, the therapy of choice is an endo-
stimulation and similar physical character of scopic or surgical intervention. In the case
of bleeding observed with gastritis erosiva

6
Application of Viofor JPS System Clinic/Delux
and gastritis haemorrhagica, caution is also pause as well as people with an increased
recommended in the application of magnetic tension of the neurovegetative system. The
fields. In case of gastric ulcer the relaxative above symptoms are observed particularly in
and anti-spastic action of variable magnetic the first days after exposure. The frequency of
fields could be considered, but the decision their occurrence may decrease through the ap-
of their application in this disease should be plication of an increasing value of intensity as
consulted with a doctor, due to the fact that it well as informing the patients of the possibility
is frequently accompanied by the gastritis. of their occurrence. These symptoms affect
Serious infections of a viral, bacterial approx. 4% of women and 1% of men.
and myframe bedic origin are a contrain- A temporary intensification of pain usu-
dication to the application of magnetic fields. ally occurs during the first period of exposure
Each type of these diseases should be and affects people with significant degenera-
treated by antiviral, antibiotic and antimy- tive changes of the osseous-joint system.
frame bedic pharmaceuticals. Summing up, after taking into consideration
Electronic implants applied in modern the above-presented observations, it should be
medicine, in most cases have good screening stated that the application of variable magnetic
from external interference. From a physical fields, is safe and the progress of experimental
point of view, the fields applied in magnetic and Clinic/Deluxal knowledge will in all likelihood
field therapy, particularly in magnetostimula- result in verification of the given information.
tion, have a small possibility of influencing the
operation of electronic implants. The person 2.2 Magnetostimulation paired
ordering the therapy rarely has knowledge of with light energy
the electronic implant’s screening quality. In
the case of an absence of this knowledge, it
is safer to refrain from exposure of the patient 2.2.1 Characteristics of magnetostimulation
to variable magnetic fields. paired with light energy
The method of magnetostimulation paired
2.1.4 Side effects, which may occur with light energy utilizes the magnetic field ef-
due to magnetostimulation fect described in paragraph 2.1 of the Manual.
Until now, no serious side effects of mag- Light energy has a mainly topical effect on
netostimulation have been reported. The fol- tissues. Its penetration inside the body de-
lowing effects might occur temporarily: pends on the light wavelength. Tissue reaction
- tingling sensation depends on energy absorption in each tissue
- a numbing sensation layer. The absorption effectiveness is mainly
- a warmth sensation affected by: tissue thickness, blood supply
- sleep disorders, particularly in elderly people and flow, water content and presence of pig-
- irritation ment. Infrared reactions start at the level of
- concentration disorders cell membrane, and reactions to red light – in
- a temporary intensification of pain. mitochondria.
A tingling, cramped and warmth sensa- At the tissue level, mechanism of biologi-
tion may occur particularly in the first three or cal effect of variable magnetic fields and light
four days of exposure. This concerns particularly energy are of similar nature. When applying
post-traumatic areas of the osseous system. magnetostimulation together with light, a sy-
Sleep disorders is observed only in ap- nergy effect could be expected, depending on
prox. 3% of older patients. They could be eas- individual characteristics.
ily eliminated by an exposure of increasing Main biological effects of magnetostimula-
level of induction. tion paired with light energy should be:
Irritation and/or concentration disorders - improved oxygen assimilation by its car-
concerns particularly women during meno- riers, which improves ATP synthesis in

7
Application of Viofor JPS System Clinic/Delux
oxidation-reduction systems aerobic and - joint strain and injuries
anaerobic respiration path; increased - rheumatoid inflammation of the joints
ATP concentration is directly related to - stiffening inflammation (rigor)
the increased activity of ATP-dependent of the spinal joints
enzymes, responsible, among others, for 2. soft tissue:
protein and collagen synthesis, which better - extrajoint rheumatism – so-called
stimulates tissue regeneration process fibromyalgia
- strong stimulation of DNA synthesis and - traumas of soft tissue
cell proliferation - zoster
- vasodilatation effect due to direct relaxation - neuralgia
of vessel smooth muscular coat, related
with accelerated processes of angiogenesis II. Regenerative effect
and tissue perfusion, which improves re- - skin and soft tissue inflammation
generation - skin and soft tissue post-inflammation
- direct effect on liquid crystal membrane condition
structure and on modification of membrane - psoriasis
enzyme activity related with increased ATP- - bedsores
dependent ion pumps through increased - chronic and subacute arthritis
ATP synthesis in mitochondria; each of the - initial treatment prior to massage,
above processes leads to changed mem- kinesitherapy and iontophoresis
brane permeability, thus to change in elec-
trolyte and water distribution between cells III. Improvement of peripheral circulation
or their organelle and surrounding areas; 1. improper local blood flow in the limbs:
both processes have a strong anti-inflam- - in diabetic angiopathy
matory and anti-swelling effect - in atherosclerotic angiopathy
- restored ATP-dependent membrane pump - in crural ulceration
activity, which has been lost due to dis- 2. accelerated absorption of haematomas
ease-related changes – sodium-potassium 3. decreased lymphatic oedema
and calcium pumps in neurons, which leads
to membrane hyperpolarization and stimu- IV. Anti-spastic effect
lus blocking, even those of high amplitude; - conditions after apoplectic stroke
as a result, pain stimuli conduction in affer- - degenerative diseases of the nervous
ent fibers is decreased, therefore a clear system, accompanied by spasticity
analgesic effect can be observed
- changed activity of serotoninergic synapses V. Decreased oxygen deficit and debt
and stimulation of beta-endorphin release through affecting the processes of:
and stimulation of pituitary gland-adrenal - aerobic tissue respiration
gland axis accompanied by increased re- - anaerobic tissue respiration.
lease of glycocorticoids adds to a strong
analgesic effect 2.2.3 Contraindications
- regulation of immune system and rheologi- to magnetostimulation and light energy
cal blood properties. treatment
Although there is no proof for negative
2.2.2 Indications for magnetostimulation effect of magnetostimulation on each body
with light energy function, contraindications should be the same
I. Analgesic effect as in case of variable magnetic field (see
1. osseous-joint system: paragraph 2.1.3.). Special attention should be
- degeneration of the spine and osseous- paid to local neoplastic changes and changes
joint system of the upper and lower limbs that are difficult to identify, predisposition to

8
Application of Viofor JPS System Clinic/Delux
bleeding as well as acute inflammation of skin - effect of structures with piezoelectric and
and soft tissue in the area exposed to therapy. magnetostrictive properties.
Patients who are administered drugs, which The above-listed biophysical effects of
cause photosensitization or show symptoms variable magnetic fields influence biological
of photosensitization should not be treated – it effects of these fields and their therapeutic
is easy to verify by checking patient’s history. application:
1. intensified process of oxygen utilization
2.2.4 Side effects which may occur due and tissue respiration
to magnetostimulation with light energy 2. vasodilatatory and angiogenetic effect
treatment 3. intensified regeneration of soft tissue
No significant side effects should be ex- 4. accelerated process of bone union
pected. The following effects might occur 5. anti-inflammatory and anti-swelling effect
temporarily: 6. analgesic effect
- tingling sensation, 7. modification of membrane transport and
- numbing sensation, ion distribution.
- warmth sensation,
- skin erythema in the area of treatment. 2.3.2 Indications for magnetotherapy
with Viofor JPS Clinic
2.3 Magnetotherapy*
I. Motor system diseases
2.3.1 The essence of magnetotherapy - delayed union of fractured bone
In Viofor JPS Clinic apparatus there have - complications of endoprosthesoplasty
been applied the most practically used param- of hip joint related with lack of stability
eters of magnetotherapy. The generated signals of endoprosthesis
are sinusoidal, triangular or rectangular, with - pseudoarthrosis
frequencies from 5 to 40 Hz regulated incre- - complications related to surgical
mentally every 5 Hz. Time set can be regulated bone extension
incrementally from 10 to 30 minutes every 5 - osteoporosis
minutes. There is one binary flat applicator. - Sudeck’s syndrome
The application of variable magnetic fields in - degenerative joint disease and arthritis
medicine is based on biophysical effects of such - states after bone fractures
fields on the living matter: - states after joint injury
- effect on uncompensated magnetic spins
of paramagnetic elements and free radi- II. Diseases of nervous system
cals as well as on diamagnetic molecules - diabetic angiopathy
- effect on liquid crystals in the organism, in - atherosclerotic angiopathy
particular on the components of cell mem- III. Diseases of skin and soft tissue
branes of liquid crystal properties - skin and soft tissue wounds
- effect on the structure of protein compo- - bacterial infection of skin and soft tissue
nents of membrane channels and ion dis- - burns
tribution - keloid
- shift of moving electric charges as the re- - crural ulceration and trophic lesions.
sult of Lorentz’ force and Hall’s effect
- changes of certain physical-chemical prop- 2.3.3 Contraindications
erties of water to magnetotherapy
- induced potential difference in the areas Despite the lack of proof for negative effect
filled with electrolyte of variable magnetic field of therapeutic pa-
- effect on depolarization of cells indicating rameters on body functions, there are however
automatism certain contraindications to magnetotherapy.
* Only for Viofor JPS Clinic

9
Application of Viofor JPS System Clinic/Delux
Presently the most common contraindica- sensation in the area exposed to the field,
tions are as follows: sometimes accompanied by general ir-
Pregnancy, neoplastic disease, electronic ritability and sleep disorders, lack of con-
implants, active tuberculosis of lungs, insulin- centration felt by some patients usually
dependent diabetes, thyrotoxicosis, bleeding only within first few days of therapy. These
from alimentary tract, states after organ trans- effects usually occur in a small group of
plants, serious viral, bacterial and myframe patients, are of transient nature and usually
bedic infections. occur in people with unstable vegetative
system. The effects can usually be pre-
2.3.4. Side effects that might occur vented by starting the treatment with low
due to magnetotherapy field induction values and short time of ex-
The only side effects occurring dur- posure, and then gradually increasing both
ing therapy with variable magnetic field is parameters to therapeutic values over the
subjective tingling, numbing and warmth next few days of treatment.

3 THERAPY SELECTION

Viofor JPS Clinic/Delux apparatus is a 3. Magnetostimulation is most of all an ad-


unique and original design that provides the dition to primary treatment. Changes in
user with the selection of one of three options treatment (particularly pharmacological)
of variable magnetic field therapy: may be introduced only after consultation
1. magnetostimulation in JPS System with a doctor.
2. magnetostimulation in JPS System 4. Introduction of a program according to set
paired with light energy D1 or D2 (see point.12) several days prior
3. magnetotherapy*. to therapy enables to determine individual
In order to establish the application reaction to magnetic field therapy.
parameters and therapy options 5. When lying down on a large applicator
for a given patient, the following factors – mat, the head should always be posi-
should be considered: tioned near the cable outlet.
- age 6. Due to the possibility of sleep disorders
- general health occurrences, particularly in older people,
- course of the main illness treatments should be avoided late in the
- co-existing illnesses evening. It particularly concerns treat-
- pharmaceuticals administered ments of the head and neck area with the
- the number of treatments use of the pad, elliptic and point applicator
and their duration as well as ring applicators.
- duration of magnetostimulation 7. Magnetostimulation with the use of the
- other types of physical therapy. large applicator (frame bed, mat, short
mat) can be applied in intervals of mini-
3.1 Magnetostimulation mum six hours, two or three times within
24 hours.
1. Therapy should be conducted upon 8. In case of long-term therapy
a diagnosis and doctor’s consultation. in chronic illnesses, a week interval
2. Magnetostimulation, like other forms of is recommended after several weeks
physical therapy, usually does not bring im- of application.
mediate effects. They depend on the nature 9. During therapy it is not possible to inter-
of the illness, its length and coexisting ill- fere with the application’s parameters
nesses. The individual reaction to this type – all keys, apart from START/STOP and
of treatment is also important. SOUND, are inactive at this time.

10
Application of Viofor JPS System Clinic/Delux
Applicator combinations

2)

2)

2)
1)

1)

2)

2)

2)

1)
2)

Applicator for magnetotherapy is not combinable with other applicators!

10. Possible subjective side effects tors of the same color cannot be used on
are usually of temporary character. one patient in respect to the same body
11. Usage of an adapter enables to use two part. It is, however, possible to use any
applicators at the same time without pos- two applicators on two patients. Other
sibility of applying different application combinations require consultations with
parameters. Below-listed applicator com- the distributor. The proper position of the
binations, when used simultaneously on applicators is described in paragraph 11
a patient, intensify the effect of variable – apparatus usage, magnetostimulation.
magnetic field, and, therefore, accelerate 12. Prophylaxis – according to default pa-
the therapeutic effect. Two elliptic applica- rameters set by the manufacturer.

* Only for Viofor JPS Clinic

11
Application of Viofor JPS System Clinic/Delux
3.1.1 Application parameter components 3. M3 – application with an increasing-de-
3.1.1.1 Programs (P) creasing intensity; the degree of field in-
1. P1 – JPS system with one impulse type tensity increases in 2 min from 0,5 to the
of simplified structure selected value, during the next 8 min re-
2. P2 – JPS system with two impulse types mains at the selected level, during the last
3. P3 – JPS system with two impulse types 2 min decreases to a level of 0,5 – applies
of different order and impulse structure as to the large and small applicator; apply
compared with P2 from an intensity of 3 or higher.

3.1.1.2 Application modes (M)


1. M1 – application with a constant intensity;
the selected field intensity is constant
throughout the whole time of application;
recommended in persons of good health
condition

3.1.1.3 Magnetic field intensity ( I )


Intensity from 0,5 to 12 corresponds with
proportional values of magnetic field emitted
by the applicators. It is recommended to start
or therapy from magnetic field intensity set at
1 and successively set to higher levels. The
speed of setting to higher levels depends on
2. M2 – application with increasing intensity; the age and general health of the user. For
the degree of field intensity increases persons of high sensitivity to magnetic fields
every 10 or 12 seconds from 0,5 to the se- it is recommended to start application from
lected intensity, cyclically during the time a level of 0,5. For older people and children,
of application; recommended in persons of changes may be introduced every 5-7 days,
lowered health condition for others every 1-2 days.
Proposed magnetic field intensity for adults:
- level 0,5 to 12, depending on age and indi-
vidual sensitivity to magnetic fields.
Proposed intensity for children:
- 0-2 years level 0,5-1
- 2-5 years level 1-2
- 5-11 years level 1-4
- 11-15 years level 1-6

3.1.1.4 Selection of application


parameters
Application parameters are selected
through a combination of: program (P), mode
(M) and intensity (I):

12
Application of Viofor JPS System Clinic/Delux
1. P1 + M1 – application with constant - the simultaneous application of several
intensity (8 min) large applicators on several athletes
2. P1 + M2 – application with increasing e.g. after training.
intensity (12 min) The two-coloured marking of the appli-
3. P1 + M3 – application with increasing- cators serves to define the direction of the
-decreasing intensity (12 min) magnetic field poles generated by the ap-
4. P2 + M1 and P3 + M1 – JPS system plicator. With the same direction of impulses
with constant intensity (10 min) generated by the control unit, each applica-
5. P2 + M2 and P3 + M2 – JPS system tor placed e.g. with the light side upwards
with increasing intensity (12 min) will have this same direction of magnetic
6. P2 + M3 and P3 + M3 – JPS system pole. This is essential in the simultaneous
with increasing-decreasing intensity application of two applicators on the same
(12 min). person.
The recommended modes of application
are based on the wide clinical experience CAUTION!
of a team of practicing physicians as well Do not apply applicators with sides of the
as multicenter research that evaluate same colour turned towards each other.
the effectiveness of magnetostimulation
with the Viofor JPS Clinic/Delux apparatus. The arrangement of the whole body or
Each of the 9 types of application its parts between two applicators generating
can be combined with chosen field intensity inhomogeneous fields allows one to achieve
according to the above-mentioned fields approximate to homogeneous.
instructions. One may apply two or three large applica-
In order to establish the application param- tors simultaneously (connected with one or
eters for a given patient, the following factors two adapters) as well as various combinations
should be considered: of applicators (e.g. large and point, large and
- age small, small and point). Point applicators are
- general fitness adapted to work with a large or small applica-
- course of the main illness tor with the light end turned upwards. In the
- co-existing illnesses event of using the application with two or three
- pharmaceuticals administered people simultaneously, the therapy parame-
- time of the day, when therapy will be ters for all persons are identical. It is not possi-
carried out ble to separately set field parameters for each
- duration of magnetostimulation of the simultaneously applied applicators.
- other types of physical therapy.
3.2 Magnetostimulation
3.1.2 The use of two or more applicators with light energy
Accessories (adapter, additional applica-
tors) enable the optional use of several ap- CAUTION!
plicators simultaneously with one or more Strong red and/or infrared
persons. Thanks to this one may among other radiation!
things increase the efficiency of treatments Protect your eyes!
or take advantage of the appliance more eco- Do not look at the light!
nomically. For example:
- a patient with rheumatoid arthritis and 1. Magnetostimulation paired
large degenerative lesions of the hip-joints with light energy is used locally during
– simultaneous application of the mat the series of repeated treatments
on the whole body and pad in the vicinity with the use of one of three magnetic-light
of the hip-joints applicators.

13
Application of Viofor JPS System Clinic/Delux
2. Each applicator generates a heteroge-
neous variable magnetic field in the JPS
system. The light area of 48 diodes is lo-
cated in within the coil and equals 19,6 cm2.
3. Light energy penetrating the tissue de-
pends on the radiation properties (wave-
lengths, angle of incidence), and on tissue
properties (rate of hydration, perfusion,
pigment content).
4. Tissue penetration will be the highest in
case of infrared magnetic-light applicator
(47 IR diodes, wavelength 855±5 nm,
1 R diode, wavelength 630±5 nm ),
intermediate in case of combined
magnetic-light applicator (24 IR diodes,
wavelength 855±5 nm, 24 R diodes,
wavelength 630±5 nm), and the lowest
in case of red magnetic-light applicator
(48 R diodes, wavelength 630±5 nm).
5. Energy power of each light type
will increase to intensities set
on the controller between 0.5 – 4.
It will be identical within intensity range
of 4 – 12.
6. Treatment with magnetic-light applicators
should be started from low induction
values of variable magnetic field, due to
the expected synergy of magnetostimula-
tion and light energy.
7. If the magnetostimulation has started with
the use of elliptic applicator, then it is
recommended to lower down the induction
of magnetic-light applicator by 50%
as compared to the previous one in re-
spect to the same body area.
8. Due to the applicator contact with the skin,
it is recommended to wash and dry the
skin area to be exposed.
9. When using magnetic-light applicators,
never direct the light in the eye area!

3.2.1 Application parameter components

3.2.1.1 Programs ( P )
For use of magnetic-light applicators all
three programs are recommended P1, P2
i P3. Parameters of energy of the light are
contained in the part C (Technical Data),
table 10.5

14
Application of Viofor JPS System Clinic/Delux
3.2.1.2 Application modes ( M ) 1. Red magnetic-light applicator R (wave-
Viofor JPS Clinic/Delux apparatus ena- length 630±5 nm) is used mainly for skin
bles three different application modes (M) disease treatment, such as unhealing
for variable magnetic field described in para- wounds and ulcerations (including bed-
graph 3.1.1.2. sores), common acne and zoster. It is rec-
When using magnetic-light applicators, ommended in dermatology and cosmetic
M1 mode is recommended. It is the applica- surgery
tion with constant intensity, where selected 2. Infrared magnetic-light applicator IR
intensity of variable magnetic field does not (wavelength 855±5 nm)in motor system
change during the application. diseases (arthrosis; pain syndromes in
discopathy; syndromes due to strain of
3.2.1.3 Magnetic field intensity ( I ) muscles and periarticular soft tissue;
Intensity from 0,5 to 12 corresponds with inflammation of tendons, fascia, tendon
proportionate values of variable magnetic sheaths and joint bursa; periarticular in-
field emitted by elliptic applicators. The high- flammation; bone fractures and delayed
est mean impulse value of variable heteroge- union of fractured bones) as well as in
neous magnetic field (in microteslas) on the neurological diseases (neuralgia of pe-
surface of magnetic-light applicators is 8.9 x 1 ripheral nerves – intercostal neuralgia,
at P2M1 (intensity level selected on the Viofor trigeminal neuralgia, post-zoster neuralgia,
JPS Clinic/Delux controller). sciatica; diabetic neuropathy).
It is recommended to start the therapy 3. Combined magnetic-light applicator R+IR
from the lowest intensity and increase it (wavelength 630±5 nm / 855±5 nm)
gradually. in chronic skin inflammations; keloid,
psoriasis, psoriatic joint inflammation,
3.2.1.4 Selection of application parameters ankylosing spondylitis, rheumatoid in-
Application parameters are selected flammation of the joints, post-thrombotic
through a combination of one of three pro- syndrome.
grams (P1, P2 or P3), application mode M1 Energy doses as well as type of light and
(recommended) and magnetic field intensity (I) the number of treatments should be selected
Treatment duration: individually; their parameters should be
- for P1: 8 minutes x time multiplier in the similar to those used in laser biostimulation,
controller (1, 2 or 3), i.e. 8, 16 or 24 min- depending on experience and preferences of
utes respectively the therapist.
- for P2 and P3 : 10 m x minutes x time mul-
tiplier in the controller (1, 2 or 3), 3.3 Magnetotherapy*
i.e. 10, 20 or 30 minutes respectively
When the induction in Viofor JPS Clinic/ 1. Magnetotherapy is applied locally during
Delux controller increases from 0.5 to 4, light the series of repeated treatments.
energy density increases as well, whereas 2. Viofor JPS Clinic apparatus is equipped
above these values, i.e. from 4 to 12, only with a flat, binary applicator
variable magnetic field induction will increase. generating variable heterogeneous
Due to the expected synergy of both types of magnetic field.
radiation, the therapy should be started from 3. When a given body part is placed between
low induction values of variable magnetic two elements of the flat applicator, the
field, and, therefore, from low light energy generated magnetic field is close the ho-
density level. mogenous one.
Selection of one of three magnetic-light 4. 1 m connection between two elements of
applicators should depend on the type of light the flat applicator enables simultaneous
required for a given form of therapy. exposure of two body parts.
* Only for Viofor JPS Clinic

15
Application of Viofor JPS System Clinic/Delux
5. It is forbidden to use two elements of the from 5 to 40 Hz, regulated incrementally
applicator with the same color sides facing every 5 Hz.
each other, except for therapy of different
parts of the body separately with each part 3.3.1.3 Shape of magnetic signals (S)
of applicator. Signals of variable magnetic field gener-
6. After the first few treatments, side effects ated by Viofor JPS Clinic apparatus can
of transient nature might occur (see 2.3.4) be sinusoidal, triangular or rectangular in
7. Due to possible sleep disorders, especially shape.
in older patients, treatment in the evening
or at nights is not recommended. 3.3.1.4 Treatment duration
Viofor JPS Clinic apparatus is equipped
3.3.1 Application parameter components with incremental time set from 10 to 30 mi-
nutes, every 5 minutes.
3.3.1.1 Magnetic field intensity
Intensity from 0,5 to 12 corresponds to 3.3.1.5 Selection of application parameters
proportionate values of variable magnetic field Application parameters are selected
emitted by a flat applicator. It is 6 x I for each by setting combination of: frequency, signal
applicator element, when used separately shape (S), intensity of magnetic field and
or 10 x I when they are used one on top of treatment duration on Viofor JPS Clinic con-
another. It is recommended to start treatment troller. The therapist can select application
from low intensity and increase it gradually. parameters and the number of treatments
individually, depending on own experience
3.3.1.2 Magnetic field frequency and preferences as well as latest findings
Viofor JPS Clinic apparatus enables se- in magnetotherapy.
lection of variable magnetic field frequency

16
Usage of Viofor JPS System Clinic/Delux

PART B

17
Usage of Viofor JPS System Clinic/Delux

Usage of Viofor JPS System Clinic/Delux


4 GENERAL INFORMATION

4.1 Apparatus back panel


1 – power switch 5 – magnetotherapy* / magnetostimulation
2 – voltage switch with ledtherapy (socket G3)
3 – fuse base 6 – magnetostimulation (socket G2)
4 – power cable 7 – magnetostimulation (socket G1)

Drawing 1: Viofor JPS System Clinic – apparatus back panel

Drawing 2: Viofor JPS System Delux – apparatus back panel

4.2 Apparatus front panel

1 – display 8 – IR receiver window


2 – STAND-BY diode 9 – key UP
3 – key ON – switch to active mode 10 – key RETURN
from STAND-BY mode 11 – key LEFT
4 – key ESCAPE 12 – key OK
5 – START/STOP diode 13 – key DOWN
6 – START/STOP key 14 – key RIGHT
7 – SOUND 15 – key NEXT
* Only for Viofor JPS Clinic

18
Usage of Viofor JPS System Clinic/Delux

Drawing 3: Viofor JPS System Clinic – apparatus front panel

Drawing 4: Viofor JPS System Delux – apparatus front panel

19
Usage of Viofor JPS System Clinic/Delux

5 VIOFOR JPS CLINIC/DELUX USER INSTRUCTIONS


5.1 General information Viofor JPS Clinic/Delux is operated by
a system of graphic windows. There is one
On the front panel of Viofor JPS Clinic/ main window, where the mode is selected be-
Delux there are: membrane keyboard for ap- tween MAGNETOSTIMULATION - MAGNE-
paratus operation control, graphic display of TOTHERAPY* - CONFIGURATION and three
the current state and LED diodes signalling additional windows for modes MAGNETO-
STAND-BY mode and electric flow in the ap- STIMULATION and MAGNETOTHERAPY:
plicator. These elements constitute the main SETTINGS, CONFIRMATION, WORK.
control panel of the apparatus. Modes are selected with the keys UP (9),
DOWN (13), and key OK (12) is used to ap-
Available keys and their functions: prove the settings and to display window SET-
TINGS. In any graphic window, current cursor
position is indicated by blinking highlighted
key ON (3) field. Current set value is indicated by high-
switch from STAND-BY mode lighted field. The type of applicator connected
to a given output is indicated as follows:
starting or stopping of application (6)
large applicator, mat;
large applicator, short mat,
key OK (12) approval applicator - frame bed
of a given selection
pad
key CANCEL (4) present operation,
move from submenu to main menu, point applicator S
close current window and return to
previous one point applicator Z

SOUND (7) signal turn on/off key; elliptic applicator

ring applicator Φ 65 cm
move the cursor DOWN (13)
or edit value
ring applicator Φ 30 cm
move the cursor UP (9)
or edit value magnetic-light applicator R

move the cursor RIGHT (14) magnetic-light applicator IR


or edit value
magnetic-light applicator RIR
move the cursor LEFT (11)
or edit value
applicator for magnetotherapy
move to the NEXT window (15)
applicator not coded
or edit value
or connected by adapter
RETURN to previous window (10) no applicator.
or edit value
* Only for Viofor JPS Clinic

20
Usage of Viofor JPS System Clinic/Delux
The device switches to “STAND-BY” mode netotherapy. The meters can be displayed
after two minutes of idleness. LCD display is in the main window by pressing two keys
blank. Pressing of key ON (3) or any other key together: ON (3) and ESCAPE (4).
turns the STAND-BY off. Pressing of START/
STOP key (6) prompts CONFIRMATION win- PLACEBO function is activated by press-
dow for magnetostimulation or magnetotherapy, ing the key START/STOP key (6) when the
depending on current cursor position. device is switched on with the power switch
The device is equipped with two independ- on the back panel.
ent meters for magnetostimulation and mag-

5.2 Configuration
5.2.1 Main window

Viofor JPS®
Delux

Drawing 5: Viofor JPS System Delux – main window, choice MAGNETOSTIMULATION

Viofor JPS®
Clinic

Drawing 6: Viofor JPS System Clinic – main window, choice MAGNETOSTIMULATION

21
Usage of Viofor JPS System Clinic/Delux
5.2.2 Window CONFIGURATION

Drawing 7: Window CONFIGURATION (Viofor JPS Clinic).

In the window CONFIGURATION user can Editing of a selected setting is done after a
choose the parameters of device: language of given position is confirmed by the key OK
interface (Polish, English, German, French), vol- (12), and then with the keys LEFT (11),
ume of sound (0-100%) and also the IR chan- RIGHT (14) where the value of a given set-
nel (the device Viofor JPS Clinic in future will ting changes immediately. Changes are
have the wireless connection with computer). confirmed by the key OK (12), and the key
Keys UP (9), DOWN (13) are used to ESCAPE (4) is used to undo the editing.
move from field to field in the main menu.

5.3 Mode MAGNETOSTIMULATION


5.3.1 Window SETTINGS

Drawing 8: Window SETTINGS. Example 1: output G1(mat), P2, M1, level 4, time 1x10 min

22
Usage of Viofor JPS System Clinic/Delux
In this window, the user can select the fol- plicator type in the programmable lists L1,
lowing settings: L2, L3 and L4, after placing the cursor on the
applicator icon. Saving of the set list is done
Output – enables socket identification of the by pressing OK (12), after placing the cursor
applicator socket (G1, G2, G3) used for a on SAVE.
given application
Key RETURN (10) is used to switch
Program – for selection of therapy P1, P2, P3; to main window and key NEXT (15) to switch
to window CONFIRMATION.
Application – for selection of application
mode M1, M2, M3; 5.3.1.1 Application parameters
programming
Level – therapy intensity from 0.5 to 12; To program the list, go to the menu
SETTINGS, set demanded param-
Cycle – sets the number of application repeti- eters: P, M, level of intensity and amount
tions (from 1 to 3) of cycles,and move the cursor with a key
DOWN (13) to the position LIST. With the key
List – enables opening of preset settings. OK (12) or RIGHT(14) place the cursor on
For magnetostimulation, there are four sets the chosen field of list. With the key RETURN
of settings programmable by the user, L1, (10) or NEXT (15) choose the list (L1-L4),
L2, L3, L4 and two sets of default settings and pressing the keys DOWN (13) and UP
D1, D2. Each set contains four independent (9) find position of the list, where you want
settings: to save the parameters. Move the cursor
on the icon of applicator, and with the keys
L1-1 to L1-4 DOWN (13) and UP (9) choose applicator
L2-1 to L2-4 which should be used with the parameters
L3-1 to L3-4 set, then move the cursor to the position
L4-1 to L4-4 SAVE and confirm with a OK (12) key.
D1-1 to D1-4
D2-1 to D2-4 5.3.1.2 Application parameters recalling
To recall chosen list, using the key
The above options are contained in the DOWN (13) move the cursor to the position
main menu of the window SETTINGS. Keys LIST, than with a key DOWN (13) and UP
UP (9), DOWN (13) are used to move from (9) or RIGHT(14) place the cursor on
field to field in the main menu. the number of list. With the keys RETURN
Editing of a selected setting is done after (10) or NEXT (15) choose the number
a given position is confirmed by the key of the list (L1-L4) and using the keys DOWN
OK (12), and then with the keys LEFT (11), (13) and UP(9) position number. Next
RIGHT (14), where the value of a given set- to the chosen number will appear the appli-
ting changes immediately. Changes are con- cator icon, programmed for this list. If such
firmed by the key OK (12), and the key ES- an applicator is connected to the controller,
CAPE (4) is used to undo the editing. In the after pressing the key OK (12), controller will
main menu, keys LEFT (11), RIGHT (14) choose this applicator. In case there is no
are used to select main settings from the list such applicator, controller will choose ap-
in the main menu, and keys UP (9), DOWN plicator “?” (unidentified, if such is connect-
(13) are used for selecting of a given setting. ed) or with the five short beeps will signal
In order to display a set list, when the cursor the lack of applicator.
is in the selected field, press OK (12). Keys
UP (9), DOWN (13) are used to set the ap-

23
Usage of Viofor JPS System Clinic/Delux
5.3.2 Window CONFIRMATION

Drawing 9: Window CONFIRMATION, settings according to example 1

The window displays the settings set switch to windows SETTINGS, and keys OK
in the window SETTINGS. It is possible (12), NEXT (15) or START/STOP (6) are
to switch on the sound signal with the key used to start magnetostimulation and switch
SOUND (7). Key RETURN (10) is used to to window WORK

5.2.4 Window WORK

Drawing 10: Window WORK, settings according to example 1

The main element of the window is a grammed cycles. The windows also displays
timer measuring the time left until comple- a programmed maximal level of the appli-
tion of stimulation. Timer shows total time of cator signal and its current value. Current
a stimulation cycle and the number of pro- impulses fed to the applicator are displayed

24
Usage of Viofor JPS System Clinic/Delux
graphically. The current flow is signaled by a is signaled by an intermittent sound, and
LED diode and by sound, if the key SOUND then the window switches to window CON-
(7) is pressed. Completion of the stimulation FIRMATION.

5.4 Mode MAGNETOTHERAPY (only Viofor JPS Clinic)


5.4.1 Window SETTINGS

Drawing 11: Window SETTINGS. Example 2: S1 (sinusoid), level 2, frequency 30 Hz, time 20 min

The only possible to use applicator in List – displays preset settings program-
the mode of work MAGNETOTHERAPY is a mable by the user - L1, L2, L3, L4. Each set
special two-part magnetotherapy applicator, contains four independent settings:
however the controller will recognize mag- L1-1 to L1-4
netic-light applicators (the icons will show on L2-1 to L2-4
the display). When there is incorrect applica- L3-1 to L3-4
tor connected to the controller, the applica- L4-1 to L4-4
tion will not start.
Editing in window SETTINGS is the same
In this window the user can edit the fol- as for the mode MAGNETOSTIMULATION.
lowing settings:
Key RETURN (10) is used to switch to
Shape – selects the type of signal used in selection window MAGNETOSTIMULA-
therapy: sinusoidal, triangular or rectangular; TION-MAGNETOTHERAPY, and key NEXT
(15) is used to switch to window CONFIR-
Level – sets therapy intensity on the scale MATION.
from 0.5 to 12;

Frequency – sets therapy signal frequency NOTICE


from 5 to 40 Hz every 5 Hz; In MAGNETOTHERAPY working mode it is
not possible to start application with connected
Time – sets therapy time from 10 to 30 min in magnetic-light applicator, although it will be
every 5 min recognized by the controller.

25
Usage of Viofor JPS System Clinic/Delux
5.4.2 Window CONFIRMATION

Drawing 12: Window CONFIRMATION, settings according to example 2

Works the same as for the mode MAGNETOSTIMULATION.

5.4.3 Window WORK

Drawing 13: Window WORK , settings according to example 2

Works the same as for the mode MAGNETOSTIMULATION

26
Usage of Viofor JPS System Clinic/Delux
5.5 Failure signal

Drawing 14: Failure window – no applicator

For modes MAGNETOSTIMULATION, - incorrect applicator;


MAGNETOTHERAPY, the device signals the - system error
following problems: - internal error

- no applicator Key OK (12) or ESCAPE (4) is used to


- unknown applicator; close the failure information window.

5.6 Remote control


The remote control device enables of application parameters. Since the device
remote controller operation as well as fast uses infrared signal, it should be directed at
programming and execution the grey circle (8) in the control desk.

5.6.1 Keys description


All keys marked yellow and red are used for programming
and for retrieving programmed application parameters.

27
Usage of Viofor JPS System Clinic/Delux

Drawing 15: Remote control

ON (1) – Activation from the STAND- G (4) – Output socket is selected with
BY mode the key LEFT or RIGHT and approved
with OK (15).
MS (2) – Work mode selection Mag-
netostimulation opens the Settings P (5) – Program is selected with the
window for magnetostimulation key LEFT or RIGHT and approved
with OK (15).
MT (3) – Work mode selection for
MAGNETOTHERAPY (not active for M/S (6) – Selection of the application
Viofor JPS Delux) method (for magnetostimulation) or
signal shape (for magnetotherapy*)
with the key LEFT or RIGHT and ap-
proved with OK (15).

28
Usage of Viofor JPS System Clinic/Delux
TIME (7) – Time setting. In case of OK (15) – Selection approval.
magnetostimulation, the key is used
for cycle function, whereas in case of MRC (16) – Retrieves the list. The key
magnetotherapy, for time function. Pa- retrieves the settings recorded
rameters are set with the key LEFT or in memory (list number and item
RIGHT and approved with OK (15). number).

HZ (8) – Frequency selection (only for MEM (17) – Save in memory. After the
Magnetotherapy ) – not active for Vio- parameters and applicator have been
for JPS Delux. Frequency is changed set, two keys pressed together save
with the key LEFT or RIGHT and ap- the settings on a selected list.
proved with OK (15).
D1, D2 (18) – Retrieve prepro-
grammed lists D1 and D2 for magne-
CODE (9)– inactive. tostimulation.

LEVEL (10) – Intensity level selection RETURN, NEXT (19) – Same as on


with the key LEFT or RIGHT and ap- keyboard – prompt to previous or next
proved with OK (15). window or select the item number
from the list.
SOUND (11) – Sound on/off.
1, 2, 3, 4 (20) – Item number on the
ESCAPE (12) – Go back to the previ- list.
ous window.

APPL (13) – Applicator (only for


magnetostimulation programming). L1-, L2-, L3-, L4- (21) – Retrieves
Sets the applicator function to be programmable lists L1 to L4 for both
programmed on the list. Applicator magnetostimulation and magneto-
changed with the LEFT or RIGHT . therapy.

LEFT, RIGHT, UP, DOWN . Cursor START/STOP (22) – Same as key


keys. Move the cursor horizontally START/STOP on the keyboard.
and vertically.

5.7 PLACEBO mode


To start PLACEBO mode, switch off the on this way will be a PLACEBO application,
controller with a switch on the back panel of next one will be already in normal mode.
the device. Than, holding the key START/
STOP (6) on the front panel of apparatus, NOTICE!
switch the device on and wait until the main
window appears.After choosing MAGNETO- PLACEBO mode will not start,
STIMULATION mode, set the parameters ap- if there is no applicator connected
plication and press the START/STOP (6) key. to the controller.
First application after switching the controller

29
Usage of Viofor JPS System Clinic/Delux

5.8 Application counter


Viofor JPS Delux controller has one which means, that until that moment,
and Viofor JPS Clinic two build-in application 12 magnetostimulation application and
counters - one for magnetostimulation and 2 magnetotherapy application has been per-
one for magnetotherapy. formed. In case of Viofor JPS Delux only one
reading will display on the screen, e.g.:
To display the counters go to the main win-
MS 12
dow (see drawing 3) and press simultaneous-
ly keys ON (3) and ESCAPE (4). In top left The application is counted if it lasts longer
corner of the display the counters will appear, that 30 sec. After moving to the next win-
for example (on Viofor JPS Clinic display): dow and returning to the main window, the
counters are not visible any more. If user
MS 12 wants to display them again it is neccessary
MT 2, to repeat the procedure.

6 INSTALLATION
Place the controller on a stable surface
(e.g. table) and connect selected applicators.
Plug the cable in.
A large applicator – mat should be
placed on flat and hard surface. It is rec-
ommended that applicators be located not
closer than 50 cm from large metal objects
(e.g. radiators).
In order to get the best results, the place
for therapy should be located in the area of
lowest possible external electromagnetic
fields.

7 SAFE USAGE OF VIOFOR JPS CLINIC

1. The apparatus transported or stored in tem- 3. Device can be installed and connected ac-
peratures below (negative) the temperature cording to the instruction contained in the
in the room, where it will be operated, it can present manual.
be plugged in after at least 4 hours from 4. Do not plug the device if the cable or plug
delivery. The equipment cannot be operated are damaged. Have the device serviced.
in humid places. 5. Protect the device out of reach of small
2. When installing Viofor JPS Clinic/Delux, children and mentally disabled persons.
check if the voltage is set correctly. 6. Keep away from pets.
7. If the device is flooded or gets wet, do not
NOTICE! Devices distributed in Europe are plug it in – have it serviced immediately.
set by default to 230 V. When switching 8. Before plugging the device in, put away
the voltage to 115 V (when using in other magnetic cards and electronic watches,
countries), fuse should be exchanged since they might be damaged. It applies
to T 1,25 A. both to patients and the service.

30
Usage of Viofor JPS System Clinic/Delux
9. Turn off all other electronic devices 12.After work:
(e.g. hearing aids, pain stimulators, etc.) - do not roll the mat, but fold it twice
10.For hygienic purposes, it is recommended - do not bend the pad
to cover the applicators with disposable - before cleaning the device, read appro-
foil or paper covers. priate section of the instruction
11. When the device is working: (see bellow)
- personnel must keep at least 1 meter - do not open the zipper around the mat.
away from an active applicator. 13.It is recommended to have the device
checked every 1 to 2 years (service is free
CAUTION! When using magnetic-light ap- of charge, transportation costs are cov-
plicators, do not direct the light in the eye ered by the client).
area! 14.Use only AAA alkaline batteries
- plug the applicator in and out only for the remote control.
when the power is off.

8 CLEANING, MAINTENANCE, REPAIRS


8.1 General information 8.2 Fuse replacing
- Clean the controller 1) Unplug the device from the wall socket.
with the plastic cleaner, do not use water 2) Remove the fuse handle and the fuse
- Applicators can be cleaned with a proper screwdriver. The fuse
with a cloth with a mild detergent and, handle is removed by one-quarter turn
if necessary, with disinfectant. counter-clockwise.
- Device does not require 3) Remove the fuse from the handle.
any maintenance. Insert a new fuse in accordance
- It is recommended to have the device with the instruction on the back panel.
checked at least every 2 years. 4) Place the handle in the socket
- Repairs can be done only by authorized with a proper screwdriver
service of Med & Life, as described in the and make one-quarter turn
warranty certificate. clockwise.
- When changing the fuse, 5) Plug the device in the wall
plug the controller off. socket and turn on the pow-
- If disposal is necessary, have it done er. If the device is not operat-
by a specialized company. ing, it should be serviced.

NOTICE: do not use fuses with parameters other than indicated.

9 WARRANTY TERMS
Warranty cover free repair or exchange - tears
of the device or its elements in case of manu- - discoloration due to usage of improper
facture and material defects. Exchange will detergents and cleaning agents
apply to the damaged element only. - detachment of magnetic-light
The warranty does not cover: applicator glass due to improper
• mechanical damages cleaning method
- due to device being dropped or hit - torn or broken power supply cable, ap-
- abrasions plicator cable or plug
- scratches - improper disinfection method
- dents (e.g. in autoclave)

31
Usage of Viofor JPS System Clinic/Delux
• regular wear and tear of applicators In case of repair or structural changes
• applicator deformation by unauthorized persons, Med & Life shall
• effects of atmospheric discharge have the right to revoke warranty obliga-
• pollution tions.
• flooding The manufacturer shall not be held re-
• damages due to sponsible for any body damages or injuries
- incorrect usage of the device due to incorrect usage of the device, its us-
- incorrect parameters of power supply age for purposes other than those indicated
- incorrect setting of the voltage switch in the instruction manual, or any changes in-
- incorrect fuse troduced to the device without prior approval
• broken warranty seals of the manufacturer.

32
Technical data and accessories

PART C

33
Technical data and accessories

Technical data and accessories


10 TECHNICAL DATA

10.1 Power supply

Power voltage rating 230/115 V ±10%

Frequency 50 - 60 Hz

Power input when ready for application (230 V) < 5 VA

Power input during application


Viofor JPS Clinic (230 V magnetotherapy,
level 12,rectangular signal, load 3 Ohm) < 40 VA

Viofor JPS Delux (230 V magnetostimulation, level 12) < 26 VA

Working mode Continuous running

Cut-out at the primary circuit T 0,63 A (for 110/115 V


– T 1,25 A) – time delay

Working temperature/humidity +10OC – +40OC/max. 80%

Storage and transport temperature/humidity -20OC – +50OC/max. 90%

Safety class II

Protection B (from electric shock)

10.2 Output circuits


10.2.1 Magnetostimulation

magnetostimulation output G1, G2 2 sockets “large stereo jack”

maximal voltage value on the output terminal <±15 V

maximal application current at the level 12 2A

application time 8/10/12 min

time multiplier (cycle): 1, 2, 3

impulse types 2

34
Technical data and accessories

10.2.2 Magnetotherapy (only Viofor JPS Clinic)


magnetotherapy and magnetostimulation with light energy 6-pin socket
output (G3)
maximal voltage value on the output terminal <±15 V
maximal application current at the level 12 2A
application time 10/15/20/25/30 min
frequency of the applied signal 5/10/15/20/25/30/35/40 Hz
signal type sinusoidal, triangular,
rectangular

10.2.3 Magnetic-light application

magnetotherapy* and magnetostimulation 6-pin socket


with light energy output (G3)
maximal voltage value on the output terminal <+25 V
maximal voltage (mean value) 0,5 A
output signal synchronized with magneto-
stimulation signal
safety class for light energy (incoherent light) 3A

10.2.4 Applicator coding

coding output applicators sockets


maximal voltage value on the output terminal <+5,5 V
maximal output voltage <+1,2 mA
coding element in the plug

10.2.5 Compatibility
Viofor JPS Clinic/Delux is compatible with all Viofor JPS applicators. It is designed to work with
IR remote control.

10.2.6 Casing
A special ABS casing has been used, of maximal dimensions:
width 330 mm
depth 280 mm
height 280 mm

10.2.7 Mass
Mass of the device without applicators – < 4kg

35
Technical data and accessories

11 APPLICATORS

11.1 Applicators for magnetostimulation

11.1.1 Large applicator - mat 11.1.4 Small applicator - pad


Used for the whole body treatment. Used for local treatment , can be
Contains 3 pairs of coils placed placed over or under selected body
along the mat and generating heterogene- part. Contains one pair of coils generating
ous variable magnetic field of different in- heterogeneous variable magnetic field.
tensity: coil located in the feet area are the The strongest effect of magnetic field is lo-
strongest, and in the shoulder area – the cated near each coil. Do not bend the pad.
weakest. Head should always be located Two color. Size 50 x 30 cm.
near the output cable. Mat is a portable ap- Installation: plug the applicator into any
plicator, in two colors, can be folded in 3. controller socket. Place the applicator on or
Size 180 x 65, folded 60 x 65 cm. under the treated body part. Set the desired
Installation: place the applicator on a hard application parameters and press START/
surface, away from any household appliances STOP button.
and TV or radio sets. Plug the applicator to
any controller socket. Lay down on the mat, 11.1.5 Point applicator (S and Z )
with your head near the output cable. Set the For local application. Generates het-
desired application parameters and press erogeneous variable magnetic field
START/STOP button. of point-like characteristics. Used
in stomatology, on selected acu-
11.1.2 Large applicator – short mat puncture points and in cases when
Used for the whole body treatment. Gen- a narrow beam of variable magnetic field
erates heterogeneous variable magnetic of significant induction is required. Can be
field. Used in cases, when it is recommended to used together with two-color applicator fac-
use two mats on one patient or if a regular mat ing light side up.
is not handy (e.g. after surgery). Two colors. Installation: plug the applicator into
Size 157 x 65 cm, side cable output. any controller socket. Point the plastic
Installation: plug the applicator into any control- applicator end to the place exposed
ler socket. Cover the patient with the mat. Set to treatment. Set the desired application
the desired application parameters and press parameters and press START/STOP button.
START/STOP button.
11.1.6 Elliptic applicator
11.1.3 Large applicator (light and green)
– frame bed For local application, fastened with
Metal frame bed, generates heteroge- the belt on limbs or with the tape in other
neous variable magnetic field. Used instead places of application. Generates heteroge-
of large applicator – mat in doctor’s office. neous variable magnetic field, whose area
Equipped with applicator roll guides for the of activity is within approx. 6 cm circle. Used
large ring applicator. Size: length 197 x width mostly for treatment of small and medium
60 x height 55 cm. joints, in stomatology and laryngology. Light
Installation: place the applicator away from and green applicators, when used separate-
household appliances and TV or radio sets. ly, work identically. If two elliptic applicators
Plug the applicator into any controller socket. are used on the same spot, e.g. wrist in the
Set the desired application parameters and same patient, they only applicators of differ-
press START/STOP button. ent color can be connected.

36
Technical data and accessories
Installation: plug the applicator into any be place exactly above the treated area)
controller socket. Place the applicator on the on a flat base. Set the desired parameters
treated area, fasten with the strap, with an and press START/STOP button.
adhesive tape or hold it in one position. Set
the desired application parameters and press 11.1.8 Ring applicator – large,
START/STOP button. 65 cm diameter
Used together with the large applica-
11.1.7 Ring applicator – small, tor – frame bed. Equipped with rolls that can
30 cm diameter be moved in the frame bed guides. In the
For local application on limbs. area of application, generates homogenous
Generates homogenous variable magnetic variable magnetic field, of line of force parallel
field of line of force parallel to the coil axis, to the coil axis, therefore parallel to patient’s
therefore parallel to patient’s body, which in- body, which increases application effective-
creases application effectiveness. Equipped ness. Used as local applicator for selected
with two cable sockets, marked H and L. body parts.
When applicator is connected to socket H, Installation: plug properly (see assembly
the induction of magnetic field generated by instruction) ring applicator installed on the
the applicator is three times stronger than in guide rails of the frame-bed applicator to any
socket L. controller socket. Lie down on the applicator
Installation: select therapy strength: H or L and set the ring applicator. Set the desired
and plug in the cable in the selected application parameters and press START/
applicator socket, in accordance STOP button. If frame-bed applicator and
with symbols on the cable. Plug the other large ring applicators have to be used
end to any controller socket. Place the limb simultaneously, they should be connected
inside the applicator (applicator should through an adapter.

11.2 Magnetic-light applicators


Three magnetic-light applicators for local use color of signal type of applicator
(red R, infrared IR and combined red/infra- diode
red R+IR. Generate heterogeneous variable
magnetic field of physical parameters of mag- a) orange red light applicator R
netostimulation in JPS System, whose area
of activity is within approx. 6 cm circle. Inside b) green infrared light applicator IR
the applicator coil there are:
a) red applicator R - 48 red LED diodes, c) yellow mixed light applicator RIR

a) b) c)
Drawing 16: Location of diodes in applicator: a) red, b) infrared and c) mixed
(in infrared applicator one red LED informs about emission of the invisible infrared radiation)

37
Technical data and accessories
b) infrared applicator IR - 47 infrared LED 11.2.3 Combined magnetic-light
diodes, 1 red LED diode, applicator RIR
c) combined applicator R+IR - 24 infrared (wavelength 630/855±5 nm; medium
and 24 red LED diodes action) in chronic skin inflammations; keloid,
Light area of magnetic-light applicators psoriasis, psoriatic joint inflammation, anky-
is 19,6 cm2. The applicator casing contains losing spondylitis, rheumatoid inflammation
a coil for magnetic field therapy and LED of the joints, post-thrombotic syndrome.
diodes set for light therapy. LED diode termi-
nals touch polycarbonate plate 0,75 mm thick. 11.3 Applicator
On the top of the casing there is a triangular for magnetotherapy
warning label, signal diode in color depending
on the applicator type (see table on the previ- Flat applicator used for magnetotherapy of
ous page) and company logo. selected body areas. Consists of two identi-
On the bottom there is a transparent glass, cal, connected two-color elements. Each
through which the light is emitted. generates heterogeneous variable magnetic
Plate is placed directly on patient’s body. field. Two-color coding is used to identify the
Applicators are fastened with the belt on limbs direction of magnetic field poles generated
and with tape on other parts of the body. Syn- by each one. In case of the same impulse di-
ergy of both types of electromagnetic radia- rection, each element of the applicator when,
tion is expected. for example, facing light side up, will have
Each applicator is equipped with a cable the same magnetic field pole direction. It
with a 6-pin plug. On the cable there are: should be remembered when using both ele-
- applicator serial number (from the plug’s ments place one on top of another, or at both
side), sides of the body part, so that the stronger
- applicator type identifier (from the casing local effect can be obtained. Connection of
side – R, IR, RIR) elements enables simultaneous therapy of
two body parts, which are within the applica-
11.2.1 Red magnetic-light tor range. When a given body part is placed
applicator R between two elements of the flat applicator,
(wavelength 630±5 nm; shallow ac- the generated magnetic field is close the
tion) is used mainly for skin disease treat- homogenous one.
ment, such as unhealing wounds and ulcera-
tions (including bedsores), common acne and
zoster. It is recommended in dermatology and
cosmetic surgery.

11.2.2 Infrared magnetic-light


applicator IR
(wavelength 855±5 nm; deep action)
in motor system diseases (arthrosis; pain syn-
dromes in discopathy; syndromes due to strain
of muscles and periarticular soft tissue; inflam-
mation of tendons, fascia, tendon sheaths
and joint bursa; periarticular inflammation;
bone fractures and delayed union of fractured
bones) as well as in neurological diseases
(neuralgia of peripheral nerves – intercostal
neuralgia, trigeminal neuralgia, post-zoster
neuralgia, sciatica; diabetic neuropathy).

38
Technical data and accessories
11.4 Maximal induction of magnetic field generated
by Viofor JPS System coils
B[µT]
B[µT]
Type of applicator Mean value Comments
Peak value impulse
(at setting P2 M1)
Large applicator
2.5 x I 28 x I Heterogeneous field1)
Mat/frame bed
Small applicator - pad 5xI 56 x I Heterogeneous field
Point applicator S 10 x I 112 x I Heterogeneous field
Point applicator Z 100 x I 1120 x I Heterogeneous field
Elliptic applicator (light
8.9 x I 100 x I Heterogeneous field
and green)
Small ring applicator 0.9 x I 10 x I for L
Homogenous field2)
Ø 30 cm 2.7 x I 30 x I for H
Large ring applicator
0.9 x I 10 x I Homogenous field
Ø 65 cm
Magnetic-light applica-
8.9 x I 100 x I Heterogeneous field
tor

Heterogeneous field;
for one part
Binary applicator 380 x I
for magnetotherapy 500 x I
Heterogeneous field;
for two parts piled up

Accuracy ±10%.
I – is the intensity set on Viofor JPS System controller.
1)
Heterogeneous field measured on the applicator surface in the area of maximal device readings.
2)
Homogenous field measured in the geometric center of the applicator.
Homogeneity with accuracy ±5% obtained in approx. 70% applicator internal volume.

11.5 Radiation value parameters of magnetic-light applicators connected


to Viofor JPS Clinic/Delux controller

Light area 19,6 cm2.

In R red applicators: 48 red LED diodes, ø5 mm, 2Θ1/2 = 15°

In IR infrared 47 infrared LED diodes, ø5 mm, 2Θ1/2 = 30°;


applicators: 1 red LED diode, ø5 mm, 2Θ1/2 = 15°

In RIR combined 24 red LED diodes, ø5 mm, 2Θ1/2 = 15° and


applicators 24 infrared LED diodes, ø5 mm, 2Θ1/2 = 30°

39
Technical data and accessories
Red and infrared light parameters constitute 50% of below listed parameter values
R RED APPLICATOR

Qty Unit 0,5 1 2 3 4 ÷ 12

Impulse power mW 210 210 210 210 210

Mean power for P1 mW 4,2 8,4 16,8 25,2 33,6

Mean power for P2 mW 4,9 9,7 19,5 29,2 39,0

Mean power for P3 mW 5,0 10,1 20,2 30,2 40,3

Impulse power density mW/cm2 10,7 10,7 10,7 10,7 10,7

Impulse energy mJ 0,10 0,21 0,42 0,63 0,84

Impulse energy density µJ/ cm2 5,4 10,7 21,4 32,1 42,9

Energy available during treatment for P1 J 2,0 4,0 8,1 12,1 16,1

Energy available during treatment for P2 J 2,9 5,8 11,7 17,5 23,4

Energy available during treatment for P3 J 3,0 6,0 12,1 18,1 24,2

Energy density for P1 J/cm2 0,10 0,20 0,41 0,62 0,82

Energy density for P2 J/cm2 0,14 0,25 0,60 0,90 1,19

Energy density for P3 J/cm2 0,15 0,30 0,62 0,93 1,23

Table shows light parameters of the red light applicator depending


on the intensity set on the controller.

40
Technical data and accessories
IR INFRARED APPLICATOR

Qty Unit 0,5 1 2 3 4 ÷ 12

Impulse power mW 1440 1440 1440 1440 1440

Mean power for P1 mW 28.8 57.6 115.2 172.8 230.4

Mean power for P2 mW 33.4 66.8 133.6 200.4 267.3

Mean power for P3 mW 34.6 69.1 138.2 207.4 276.5

Impulse power density mW/cm2 73.5 73.5 73.5 73.5 73.5

Impulse energy mJ 0.72 1.44 2.88 4.32 5.76

Impulse energy density µJ/ cm2 36.7 73.5 146.9 220.4 293.9

Energy available during treatment for P1 J 13.8 27.6 55.3 82.9 110.6

Energy available during treatment for P2 J 20.0 40.1 80.2 120.3 160.4

Energy available during treatment for P3 J 20.7 41.5 82.9 124.4 165.9

Energy density for P1 J/cm2 0.71 1.41 2.82 4.23 5.64

Energy density for P2 J/cm2 1.02 2.05 4.09 6.14 8.18

Energy density for P3 J/cm2 1.06 2.12 4.23 6.35 8.46

Table shows light parameters of the red light applicator depending


on the intensity set on the controller.
Wavelengths of emitted light (λP) are as follows:
• 630±5 nm for red light,
• 855±5 nm for infrared light.

0,5 1 2 3 4 ÷ 12
Intensity

ms 0,5 1 2 3 4
Width of impulse

Table shows widthdrawn of light impulses depending from intensity level set
on the Viofor JPS Clinic/Delux controller.

CAUTION!
STRONG RED AND/OR INFRARED RADIATION!
PROTECT YOUR EYES!
DO NOT LOOK DIRECTLY AT LIGHT!

41
Additional information

EC CONFORMITY DECLARATION

Med & Life herewith declares, that:


Systems for therapy and rehabilitation

• Viofor JPS System Clinic with magnetic field, with or without


light source (medical product class IIa - Annex IX, Rule 9)
• Viofor JPS System Delux with magnetic field, with or without
light source (medical product class IIa - Annex IX, Rule 9)

conform the essential requirements and the guidelines of the Council Directive
93/42/EEC - Medical Devices and with harmonized standards:
– EN IEC 60601-1 “Medical electrical equipment” with the supplements
and linked standards.
– PN-EN 60825:2000/A2:2002 (U) “Safety of laser products. Equipment classification,
requirements and user`s guide.”
is a subject of the procedure mentioned in the Annex II
of the Directive 93/42/EEC

Med & Life has implemented the Quality Management System according
to ISO 9001:2000/ISO 13485:2003 under the surveillance of the Notified
Body nr 0120, SGS United Kingdom Ltd

– in medical publications

– in specialist publications

and at www.madandlife.com

42
Awards and honors
Viofor JPS System Clinic is laureate of:
- main prize in 2002 in the competition Polish Product of the Future
- gold medal with merit on International Exhibition of Inventions “Innowacje 2003” in Gdańsk
- silver medal on International Exhibition of Inventions in Geneva 2004
- “Golden Aesculapius” of Poznań International Fairs SALUS’ 2004

Viofor JPS has won many prestigious national and international awards, such as:
- award of “Polish Product of the Future 2000”
- gold medal at the Health Spa and Rehabilitation Fair “Zdrój 2001”
- awards at 50th World Exhibition of Development, Research and New Technologies
“Brussels Eureka” in Brussels:
• gold medal and distinction
• special award of the Italian Delegation
• special award of Croatian Society of Inventors INOVA
- gold medal at “Concours Lepine” Fairs 2002 Paris
- gold medal at IENA 2002 Fairs in Nurnberg
- gold medal at the International Fair “Invest Technology 2002” in Warsaw
- medal at Intarg Fair 2002 in Katowice
and many others.

43
Awards and honors
Med & Life, the manufacturer of Viofor JPS System has a certified Quality System according to
ISO norms: ISO 9001:2000 and ISO 13485:2003.

Med & Life is also a recipient of many prestigious awards:


- Leader of Entrepreneurship 2000
- Cup and Diploma of the Minister of Education
and President of KBN for innovative activity
- Leader of Innovation 2002
- Award at “Huzar Europy 2003” competition.
For more information on the manufacturer see: www.medandlife.com

44
List of contents

LIST OF CONTENTS
1 CHARACTERISTICS OF VIOFOR JPS CLINIC/DELUX
– APPARATUS FOR MAGNETIC FIELD THERAPY............................................................. 3
1.1 Usage .......................................................................................................................... 3
2 TYPES OF THERAPY ........................................................................................................... 3
2.1 Magnetostimulation ........................................................................................................ 3
2.1.1 Characteristics of magnetostimulation ................................................................... 3
2.1.2 Indications for magnetostimulation ......................................................................... 5
2.1.3 Contraindications to magnetostimulation ............................................................... 6
2.1.4 Side effects, which may occur due to magnetostimulation..................................... 7
2.2 Magnetostimulation paired with light energy .................................................................. 7
2.2.1 Characteristics of magnetostimulation paired with light energy ............................. 7
2.2.2 Indications for magnetostimulation with light energy .............................................. 8
2.2.3 Contraindications to magnetostimulation and light energy treatment .................... 8
2.2.4 Side effects, which may occur due to magnetostimulation with light energy treatment ........9
2.3 Magnetotherapy ............................................................................................................. 9
2.3.1 The essence of magnetotherapy ............................................................................ 9
2.3.2 Indications for magnetotherapy with Viofor JPS Clinic ........................................... 9
2.3.3 Contraindications to magnetotherapy .................................................................... 9
2.3.4 Side effects, which may occur due to magnetotherapy ........................................... 10
3 THERAPY SELECTION ...................................................................................................... 10
3.1 Magnetostimulation ...................................................................................................... 10
3.1.1 Application parameter components ...................................................................... 12
3.1.1.1 Programs (P) ................................................................................................ 12
3.1.1.2 Application modes (M).................................................................................. 12
3.1.1.3 Magnetic field intensity (I) ............................................................................ 12
3.1.1.4 Selection of application parameters ............................................................. 12
3.1.2 The use of two or more applicators ...................................................................... 13
3.2 Magnetostimulation with light energy ........................................................................... 13
3.2.1 Application parameter components ...................................................................... 14
3.2.1.1 Programs (P) ................................................................................................ 14
3.2.1.2 Application modes (M).................................................................................. 15
3.2.1.3 Magnetic field intensity (I) ............................................................................ 15
3.2.1.4 Selection of application parameters ............................................................. 15
3.3 Magnetotherapy ........................................................................................................... 15
3.3.1 Application parameter components ...................................................................... 16
3.3.1.1 Magnetic field intensity................................................................................. 16
3.3.1.2 Magnetic field frequency .............................................................................. 16
3.3.1.3 Shape of magnetic signals (S) ..................................................................... 16
3.3.1.4 Treatment duration ....................................................................................... 16
3.3.1.5 Selection of application parameters ............................................................. 16
4 GENERAL INFORMATION ................................................................................................. 18
4.1 Apparatus back panel ................................................................................................... 18
4.2 Apparatus front panel ................................................................................................... 18
5 VIOFOR JPS CLINIC/DELUX USER INSTRUCTIONS ...................................................... 20
5.1 General information ...................................................................................................... 20
5.2 Configuration ............................................................................................................... 21
5.2.1 Main window ....................................................................................................... 21
5.2.2 Window CONFIGURATION.................................................................................. 22

45
List of contents
5.3 Mode MAGNETOSTIMULATION ................................................................................. 22
5.3.1 Window SETTINGS ............................................................................................. 22
5.3.1.1 Application parameters programming .......................................................... 23
5.3.1.2 Application parameters recalling .................................................................. 23
5.3.2 Window CONFIRMATION .................................................................................... 24
5.3.3 Window WORK .................................................................................................... 24
5.4 Mode MAGNETOTHERAPY* ....................................................................................... 25
5.4.1 Window SETTINGS ............................................................................................. 25
5.4.2 Window CONFIRMATION .................................................................................... 26
5.4.3 Window WORK .................................................................................................... 26
5.5 Failure signal ................................................................................................................ 27
5.6 Remote control ............................................................................................................. 27
5.6.1 Keys description ................................................................................................... 27
5.7 PLACEBO mode ....................................................................................................... 29
5.8 Application counter .................................................................................................... 30
6 INSTALLATION ................................................................................................................... 30
7 SAFE USAGE OF VIOFOR JPS CLINIC/DELUX ............................................................... 30
8 CLEANING, MAINTENANCE, REPAIRS. ........................................................................... 31
8.1 General information ...................................................................................................... 31
8.2 Fuse exchange ............................................................................................................. 31
9 WARRANTY TERMS .......................................................................................................... 31
10 TECHNICAL DATA .............................................................................................................. 34
10.1 Power supply .............................................................................................................. 34
10.2 Output circuits ............................................................................................................ 34
10.2.1 Magnetostimulation ............................................................................................ 34
10.2.2 Magnetotherapy* ................................................................................................ 35
10.2.3 Magnetic-light application ................................................................................... 35
10.2.4 Applicator coding ................................................................................................ 35
10.2.5 Compatibility ....................................................................................................... 35
10.2.6 Casing ................................................................................................................ 35
10.2.7 Mass ................................................................................................................... 35
11 APPLICATORS.................................................................................................................... 36
11.1 Applicators for magnetostimulation ........................................................................... 36
11.1.1 Large applicator - mat......................................................................................... 36
11.1.2 Large applicator - short mat................................................................................ 36
11.1.3 Large applicator - frame bed .............................................................................. 36
11.1.4 Small applicator - pad ......................................................................................... 36
11.1.5 Point applicator (S and Z) ................................................................................... 36
11.1.6 Elliptic applicator (light and green) ..................................................................... 36
11.1.7 Ring applicator – small, 30 cm diameter ............................................................ 37
11.1.8 Ring applicator – large, 65 cm diameter ............................................................. 37
11.2 Magnetic-light applicators ......................................................................................... 37
11.2.1 Red magnetic-light applicator R ......................................................................... 38
11.2.2 Infrared magnetic-light applicator IR................................................................... 38
11.2.3 Combined magnetic-light applicator RIR ............................................................ 38
11.3 Applicator for magnetotherapy .................................................................................. 38
11.4 Maximal induction of magnetic field generated by Viofor JPS System coils ............. 39
11.5 Radiation value parameters of magnetic-light applicators connected
to Viofor JPS Clinic/Delux controller .................................................................................... 39

* Only for Viofor JPS Clinic

46
NOTICE!
Disposal of old electrical & electronic equipment (applicable
throughout European Union and other European countries
with separate collection programs).

This symbol, found on your product or on its packaging, indicates


that this product should not be treated as household waste when
you wish to dispose of it. Instead, it should be handed over
to an applicable collection point for the recycling of electrical
and electronic equipment. By ensuring this product is disposed
of correctly, you will help prevent potential negative consequences
to the environment and human health which could otherwise be
caused by inappropriate disposal of this product. The recycling
of materials will help to conserve natural resources. For more
detailed information about the recycling of this product, please
contact local city office, household waste disposal service
or the retail store where you purchased this product.

AACD-EN-I-07/2006

47
NOTES

48
CO-OPERANT: