Professional Documents
Culture Documents
Sontec Instruments
A guide to ensure you, that every time you need an instrument to work properly, it Does.
Needle Holders
For manual preparations, instruments have to be For cleaning, it is recommended to use either
immersed into a combined disinfecting and cleaning solu- lintfree soft cloths, plastic brushes or cleaning pistols.
tion with proven disinfecting effect. Drying with a pneumatic-air pistol is particularly safe and
The instructions of the manufacturer have to be effective and should therefore be given preference over
strictly followed regarding concentration, temperature any other drying method.
and induction time. Special attention has to be paid to the In general, dental instruments can be prepared
manufacturer's instructions with regard to material com- like surgical instruments. Instructions for preparation of
patibility of instruments not made of high-grade steel.
Use fresh disinfecting and cleaning solutions ev-
10
endoscope in the container filled with instrument disin-
dental instruments, recommended for separate treatment,
fecting and cleaning agent and make sure that the surface
are given as follows.
is thoroughly wet.
Hand pieces and angled hand pieces as well as
Clean the channels with the brush belonging to
turbines must not be immersed. The outside is either
the system followed by flushing with the cleaning solution.
cleaned with a cloth or sprayed with disinfectant. Only the
For this purpose some manufacturers offer a hand pump.
methods specified by the manufacturers have to be applied
Take particular care when cleaning the distal end (optical
for internal cleaning and care.
system, Albarran lever etc.).
As a result of the material used, rotating dental
Immediately after this pretreatment abundantly
instruments have to be placed in special disinfecting and
rinse the instrument inside and outside with water. Then
cleaning solutions. In order to avoid corrosion, the instru-
place the flexible endoscope into the container with
ments, after quick rinsing, have to be dried at once and
instrument disinfecting solution so that all channels are
treated with corrosion protecting agent suitable for steril-
filled. The hand, 1.c. pump can now be used. Take care to
ization. In the case of ceramic-bonded or plastic-bonded
disinfect the extraction nozzle. Soaking time and solution
grinding tools, one must first check whether the disin-
concentration have to be strictly observed according to the
fecting and cleaning agents are suitable agents for these
instructions of the manufacturer.
instruments. Unsuitable agents may destroy the bonding
After chemical disinfection, thoroughly rinse all
materials.
surfaces and channels leaving no residues. To avoid water
Root canal instruments are sensitive to me-
spots use demineralized water. Additional sterile filtration
chanical damage and, therefore , have to be prepared
of the water prevents undesired recontamination.
separately. Root canal instruments with color-anodized
Dry the outside of the flexible endoscope with
handles will loose their color codes when attacked by
a lint-free cloth. Drying the channels should be done ac-
alkaline solutions.
cording to the manufacturer's instructions by using a hand
The various components of the surgical mo-
pump, light source and suction pump or by means of com-
tor line have to be externally cleaned with lint-free cloth
pressed-air of max. 0.5 bar. Undesired recontamination is
impregnated with disinfecting and cleaning agents. Soft
prevented when sterile and filtrated compressed-air is used.
brushes can also be used. When spraying the surfaces with
Elastic instruments with lockable cavities, such as
disinfection spray, wipe the parts afterward with a cloth.
bellows and breathing masks, have to be cleaned and dis-
Under no circumstances should these components be im-
infected in closed conditions thus avoiding the penetration
mersed. Any penetrated liquid has to be removed immedi-
of liquid into the cavities.
ately by turning the aperture upside down.
To avoid damage on diaphragms and functional
Simple tools can be prepared like surgical instru-
parts of the breathing system, no compressed air should be
ments. Rigid endoscopes have cavities and channels which
used for cleaning. Machine preparation allows disinfec-
are difficult to clean. Careful preparation of these instru-
tion, cleaning, after rinsing and drying without having the
ments requires:
instrument handled.
•Immediately after drying, treat all rotating •Insufficient mechanical or manual cleaning.
dental instruments (drills, burrs) with a •Unsuitable cleaning, disinfecting and care
corrosion protection agent, suitable for steriliza- agents.
tion in hot air or steam. •Failure to observe the dosage instructions for
cleaning, disinfecting or care agents.
Motor lines have to be lubricated and maintained
•Remnants of cleaning and disinfecting agents-
with agents recommended by the manufacturer. Hand
insufficient rinsing.
pieces which are not watertight have to be sprayed with
special care agents for inside cleaning and lubrication. •Poor water quality.
Apply a few drops of special oil into the supply •Residue in the sterilizing steam, when steam
channel of pneumatic surgical motor. To distribute the oil quality is not accordance with recommenda-
inside , run the motor for a few seconds with compressed- tions.
air. Proper lubrication and maintenance of the motor •Remnants of medications, marking pens or
line is of great importance, therefore, the manufacturers chemo-indicators.
instructions must be followed. •Procedural faults e.g. not cleaning brand-new
Rigid endoscopes must be lubricated. surgical instruments prior to sterilization.
However, joints and non maintenance-free stop-
cocks may have to be treated with special oil or special These and other causes for spots on surgical
grease according to the manufacturer's instructions. instruments show the complexity and difficulty of the
The only necessary maintenance on flexible endo- problems dealt with here. To facilitate tracing and finding
scopes is to treat the valves with silicon oil before inserting the cause for such stains, it is recommended to cooper-
them into the valve housing. Do not spray them with car ate with competent manufacturers. By making use of the
agents as the propellent gases will damage the instruments. company's service, you will not only take advantage of
Only silicon oils and grease-free gels should be their practical experience but their well-equipped laborato-
used as lubricants. Agents containing vaseline or paraffin ries as well.
cause swelling or softening of rubber parts. To avoid permanent damage, instruments with
Refrain from treating elastic instruments and remnants on the surface have to undergo a special treat-
breathing systems with lubricants prior to sterilization. ment. The method of treatment is adapted to the cause of
Special care and maintenance measures are prescribed by the stain. In order to avoid damage and subsequent corro-
the manufacturer, should the need arise. sion due to metal friction, under no circumstance use metal
Elastic instruments of silicon rubber may not be brushes or metal sponges to remove stains. More details
about causes of stains and corrosion are given in chapter
14
12. end of the fibre optic cable, hold it against the light and
Each surgical instrument is designed for a specific then look into the other end. Little black spots indicate
purpose. Inspection has to be carried out to ensure that breaks in the fibres. A large number of breaks reduces the
they function as they should. If any doubt, a reliable man- light output. Such fibre optics as well as endoscopes with
ufacturer can advise you on suitable inspection methods. surface damage and surface deformation should be sent for
Especially fine and delicate instruments are repair.
inspected under the magnifying glass. In order to avoid Normally, clouding of the optical system on flex-
damage during transportation, place the instruments in ible endoscopes can be removed by using a wooden appli-
specially designed racks or use special holding devices to cator (not metal) soaked in alcohol. If this does not help,
prevent them from slipping. return this instrument to the manufacturer for inspection.
Faultless surgical instruments should not be Optical fibre breaks in the optical wave guide can
packed together with instruments having damaged sur- be seen by holding the proximal end against the light and
faces. Older instruments with chipped chromium and/or looking into the distal end. Quite a number of breaks con-
nickel coating may cause discoloration or corrosion on siderably reduce the light output. Such damaged instru-
high-grade surgical instruments. It is, therefore, recom- ments have to be returned to the manufacturer.
mended to discard such instruments or pack them sepa- Endoscopes with visible external damage (e.g.
rately. deformation in the insertion part or on the supply tube,
Handles, cables and cables for neutral electrodes overstretched angled rubber) have to be returned to the
for HF-surgery have to be checked for faultless function. manufacturer.
(Caution: defective contact). It is compulsory to Flexible endoscopes are best dried while sus-
sort out defective parts. pended and should be stored unfolded. Locks and feed-
Prior to sterilization the surgical motor line with reels should be loosened. The transportation case should
accessories should undergo a functional inspection accord- be used for storage.
ing to the manufacturer's instructions. Prior to every use, check all functions of the
Units operating with compressed air should be endoscope. Use only silicon oils or greaseless gels as
checked for leaks and be given a functional inspection. lubricants. Under no circumstances should vaseline or
Any leakage will be audible or can be found by immersing lubricants containing paraffin be used as these cause soft-
the hose into water. ening or swelling.
For checking the exhaust channel additionally Breathing systems have to be inspected according
connect the pneumatic surgical motor to the compressed- to the manufacturer's instructions.
air hose. With the motor running, leaks can be best Elastic instruments have to be inspected ac-
detected in the water. cording to their function and range of use. The most
Simple tools are inspected as general surgical in- important inspections are:
struments. In order to avoid damage during transportation,
store the tools in special racks or place them in suitable •Bellows have to be undamaged and airtight
holding devices to prevent them from slipping.
The leak test for tubing sets for cooling liquid •Filling system of the bellows must not show any
can be carried out by means of a clamp and a large syringe leakage.
filled with water. Fill the tubing with water; close one end
with the clamp, and insert and empty the filled syringe in
to the other end. •Lumina of catheters and probes have to be free.
In case rigid endoscopes are used, check the ad-
ditional instruments for HF-surgery for absolutely perfect •Connections have to meet functional safety.
installation. Exchange damaged parts.
Surfaces of electrodes/loops must be free from •There should be no changes of design, e.g.
encrustations which can be removed with fine abrasive radius of curvature of tracheal tubes.
powder. Take care that neither insulation nor clips get
damaged or deformed. Elastic instruments with faults or damage have to
To avoid damage to optical systems, clean them be replaced. Frequent problems are
carefully with a swab, moistened with alcohol. If this does
not remove the clouding return the part to the manu- •Dissolvation (blister forming)
facturer for inspection. Damage can be avoided by using •Cracked surface
wooden or plastic applicators, metal is not suitable.
Optical wave guides and fibre-optic cables have to be •Sticky surface
checked for optical fibre breaks. To find this out, take one •Hardenings
•Porous surface
•Discoloration
15
•Rotating dental instruments (e.g. burrs or
To prevent premature failure, take care that elastic milling cutters) can be autoclaved.
instruments are stored in a dry place without being kinked
or overstretched. •Hand pieces should, wherever possible, be
o
autoclaved at 134 C, due to the shorter
Sterilization induction time.
General
Sterilization conditions as well as units have to be
•Check whether turbines are suitable for auto
claving with the manufacturer.
in conformity with valid quality standards.
Follow the sterilization instructions of the
manufacturer. Sterilizing accessories as well as sterilizing •For mouth mirrors, refer to manufacturer's
packings have to meet the requirements of the both the instructions.
instruments as well as the sterilizing method used.
Autoclaving All components of the surgical motor line, meant
o
Normally, autoclaving is performed with satu- for sterile application, can be autoclaved at 134 C. Refer
o
rated water steam at 134 C. In special cases a temperature to the manufacturer's instructions. Special instructions of
o
of 121 C can be used for a longer time. the manufacturer have to be observed for storage during
Sterilization procedure has to be standard suitably sterilization.
for the goods to be sterilized. Sterilizing packings have Hoses for compressed-air have to be protected
to meet the valid standards with regard to quality and ap- against pressing during sterilization.
plicable to the procedures selected. Apart from the optical system with rigid endo-
Steam used for sterilization has to be free from scopes, all endoscope parts (insert, trocar, trocar sheath,
any contamination and should neither impede the process shaft, mandrel) have to be separated from each other and
nor do damage to the sterilizer of the goods to be ster- can then be sterilized like other surgical instruments.
ilized. In order to guarantee this, meet the recommend- Sterilizing instructions from the manufacturer have to be
ations of pr EN 285 regarding the quality of the water in observed for optical systems. Optical systems suitable
o
the tanks as well as the condensate, otherwise rust particles for autoclaving should be processed at 134 C instead of at
o
from the conducting system may cause corrosion or a 121 C due to the shorter thermal stressing.
high content of silicic acid may lead to discoloration of the Flexible endoscopes are not autoclavable due
instruments. to restricted temperature compatibility. If necessary, use
Due to heating and cooling down during the gas sterilization. Instruments used for endoscopy, such as
sterilization process, a surgical instrument with a closed clamps, catheters etc. may be autoclaved.
ratchet may suffer from tension stress which causes stress Elastic instruments with and without bellows
cracking in joints or deterioration of the clamping force. made of silicon and natural latex are suitable for auto-
Therefore, such instruments have to be sterilized either in claving. Due to the shorter thermal stress, preference is
o
open condition or closed on the first ratchet only. given to a processing at 134 C. Items of thermo-materials
The loading weight of perforated trays filled with (plastic) may only be autoclaved if recommended by the
instruments should not exceed 10kg. manufacturer.
After sterilization, instruments have to be stored When elastic instruments are autoclaved, take
dry until used again. Instruments as well as the inner care that the cavities, e.g. edge of mask, bellows are open
covering of sterilized goods have to be absolutely dry after in order to avoid damage due to changes in pressure.
having cooled down to room temperature. Prior to sterilization, cavities closed with valve
Excessive condensate production during ster- (e.g. bellow catheters) have to be sucked free of air and
ilization is avoided by observation of the recommended water by means of a syringe.
maximum weight for loaded perforated trays. Drying is Functional parts of breathing systems can be
o
facilitated by wrapping the perforated trays into a cloth autoclaved at max. 134 C. Cavities must not be closed in
within the container or external paper packing. order to avoid damage to the valves.
If heavy sets are unavoidable, the instruments Hot-Air Sterilization
should be distributed among several perforated trays. In When surgical instruments are hot-air steril-
addition, special measures may be necessary for drying. ized, please take care to load and operate the sterilizers
In general, dental instruments can be autoclaved properly. To ensure safe sterilization, the temperature
o o
like surgical instruments. When separate treatment for should not be below 180 C but should not exceed 200 C as
dental instruments is necessary, obey to the following this may cause structural changes leading to irreversable
instructions. damage, especially as far as microsurgical instrument are
concerned. Instruments with parts of rubber, plastic, or
textile as well as plastic-coated instruments and handles
16
for electrodes are not suitable for hot-air sterilization. can be given only by the manufacturer of endoscopes.
The general use of lubricating agents should be After airing, flexible endoscopes should be stored in an
omitted prior to hot air sterilization. Only oil the joints and extended position.
ratchets of surgical instruments. Elastic instruments of thermolabile plastic are not
Dental instruments can generally be sterilized like autoclavable but can be gas sterilized if the manufacturer
surgical instruments. When separate treatments for dental gives instructions about a suitable procedure.
instruments is necessary, obey to the following instructions: Elastic instruments of rubber and functional parts
of breathing systems do not have to be gas sterilized; use
•Temperature should not exceed 180oC when autoclaving.
sterilizing hand pieces.
Treatment of brand new instruments
•Turbines are not suitable for hot-air sterilization.
Packings of brand-new instruments have to be
removed and instruments have to be stored in dry rooms,
•Burrs and milling cutters must not be sterilized
o open to air. Temperature fluctuations may otherwise lead
at a temperature exceeding 180 C in order to
to condensation within the plastic packing and thus cor-
avoid softening.
rosion.
Under no circumstances store instruments in
•The sterilizing of mouth mirrors has to be cupboards or rooms where chemicals are kept which can
carried out in accordance with the manufacturers produce corrosion vapors.
instructions. Prior to first use, brand-new instruments have to
be prepared. First remove any protective caps or foils.
Components of the motor line are only partly suit- Cleaning, rinsing, lubrication, inspection and sterilization
able for hot-air sterilization, due to the various materials have to be carried out according to the procedures previ-
used. ously described.
Rigid endoscopes are not suitable for hot-air ster- Prior to the first preparation, microsurgical instru-
ilization. ments have to be placed in racks or holding devices to
Flexible endoscopes are not suitable for hot-air avoid damage.
sterilization. Elastic instruments have to be kept in their
Elastic instruments are not suitable for hot-air original packing and stored in a dry, cool and dark place.
sterilization. Breathing systems are not suitable for hot-air When ordering supply, please keep in mind that in addition
sterilization. to wear through use, elastic instruments are prone to aging
Gas Sterilization even when in storage.
Gas sterilization should only be used when no Functional parts of the breathing system fre-
other method is suitable. quently contains valves or membranes which can get
Components of motor line should only be gas ster- sticky when stored for a longer period. Such valves or
ilized when explicitly recommended by the manufacturer. membranes have to be tested and operated before being
Optical systems of rigid endoscopes may be gas put to use.
sterilized, however, follow the instructions of the manu-
facturer. Special information
Flexible endoscopes can be gas sterilized at a tem-
o
perature limit of 60 C. Use higher temperatures only when
indicated by the manufacturer. •By following these instructions properly, there
For gas sterilization, the flexible endoscope is is no difference in the preparation of instru-
packed in a transparent sterilizing hose allowing a curva- ments with a mirror finish or matte surface.
ture diameter of at least 30cm. It is very important to make •These instructions do not refer to disposable
sure that the venting cap is adapted to the supply connector items
as otherwise irreversible damage will occur. •Instruments and cables with optical waveguides
To ensure protection against mechanical dam- can generally be prepared like surgical
age, the sealed flexible endoscope is put into a wire basket instruments, if the manufacturers have not given
belonging to the gas sterilizer. Again, pay attention that the other instructions. Only hot-air sterilization and
curvature diameter is not less than 30cm. ultrasonic baths cannot be used.
Goods sterilized with ethylene oxide require suf- •Fibre-optic cables should not be bent, nor coiled
ficient airing times before being used again. Depending on too tightly.
the goods sterilized and on available airing conditions, such
airing times can differ considerably. Reliable airing times
•Cables and handles for HF-surgery can be
17
machine prepared and are autoclavable. silicic acid. The remedy is in-time regeneration of the
exchanger-consult an expert.
For all other preparation processes, refer to the
instructions of the manufacturer. Materials
Individual instruments
Cold Sterilization
Cleaning
Sontec Instruments
7248 South Tucson WayEnglewood, CO 80112
(303)790-9411(800)821-7496Fax(303)792-2606
A TRADITION OF FINE SURGICAL INSTRUMENTS