Nortec 500 Familiarization Scanner Hole Inspection

Presented by: James Bittner Senior Sales Engineer, ECT Portables & Sensors

Scanner Bolt Holt Testing


Bolt hole inspection has been one of the earliest eddy current inspection since the advent of eddy current as a testing method.

Scanner Bolt Holt Testing


General instrument set-up
Bolt Hole Probes: Used for detecting
flaws inside fastener holes, which are commonly found in aircraft Fuselage and pylons

Rotary Hole Probes: Used with rotating scanner guns and turned at high speeds.

Scanner Bolt Holt Testing


General instrument set-up
LAMINATED TEST SPECIMEN COIL MOTORIZED BOLTHOLE PROBE

CRACK

IMPEDANCE DISPLAY

Reflection-Differential

Three coils are making contact with the surface, one driver and two pick-ups. More sensitive to small surface cracks and less sensitive to lift-off. Most of the rotating scanner bolt hole probes uses this type of coil arrangement.

Nortec 500 General Instrument set-up

N500 with scanner capability


Nortec 500: Four models at various price points and specific inspection features.

N500D
Scanner Support capable

N500S
Scanner Support capable

Basic Operation

ALARM (Alarm Menu)


ALARM Menu Functions include: Controls functions for Box 1 through 3 alarms. Controls functions of Polar alarm. Controls functions of Sweep alarm.
8

Basic Operation

DISP (Screen Display Menu)


Display Menu Functions include: Control functions of SCREEN SETUP Controls functions of D ERASE SETUP. Controls functions of MAIN DSP F1.
9

Basic Operation

SCAN (Scanner Menu)


SCAN Menu Functions include: Controls functions for Sweep. Controls functions of Sweep Time. Controls RPM of scanner.
10

Controls Sync Angle function. Controls Waterfall function.

Nortec 500 Series I/O Ports


16-Pin LEMO scanner cable will plug

11

Nortec Rotating Bolt Hole Scanners


Fischer 4pin (Universal) or Lemo 4pin versions available. 600 3000 RPM Compatible with NDT-19eII and NDT-24 but at fixed 1200 RPM. Alarm LED

Lemo Stepped 4pin connector. 600 2400 RPM Compatible with NDT-19eII and NDT-24 but at fixed 1200 RPM. Alarm LED

Fischer 4pin (Universal) connector. 600 3000 RPM Compatible with NDT-19eII and NDT-24 but at fixed 1200 RPM. Alarm LED Staveley Tri-axial (accommodates Gulton style probes). 40 240 RPM Compatible with NDT-19eII but at variable range of 40 - 200 RPM.

12

Nortec 500 Demonstration


Scanner Bolt Hole Testing

Scanner Bolt Hole Testing

Instrument Set-up; Step-1 Note All Nortec Scanner guns are PowerLinked.

Plug scanner cable 16-pin to 16-pin scanner cable into the instrument. Instrument will default to PowerLink mode. Press LOG/ENTER soft key.

14

Scanner Bolt Holt Testing

Instrument Set-up; Step-2: Install scanner probe into scanner gun and locate bolt Hole standard.

Press MAN function key, then press SCAN function key to adjust scanner screen to.

15

Scanner Bolt Holt Testing

Instrument Set-up; Set-3 Split Screen Display Note this follows general practices for Airbus, Boeing & USAF for bolt hole inspection.

Press SWP MODE soft key.

16

Scanner Bolt Holt Testing

Instrument Set-up; Set-3a Split Screen Display

Rotate Smart knob or press soft key until sweep mode is set for SWP MODE EXT/IMP. This will give you spit screen display for both Sweep and Impedance display for instrument step for scanner hole inspection.
17

Scanner Bolt Holt Testing

Instrument Set-up; Set-4, Screen Errors

Now press DISP function key, then press PERSIST soft key and adjust persist to read 0.4-0.6s seconds.

18

Scanner Bolt Holt Testing

Instrument Set-up; Set-5 Angle/Phase Adjustment. Note Angle/Phase will change from probe to probe. So if you exchange probe repeat this process

Press Angle soft key. Turn on scanner gun and put the hole probe on the surface of the standard. Adjust angle until you get a horizon display.

19

Scanner Bolt Holt Testing

Instrument Set-up; Set-6 Filter Adjustment for Scanner Inspection. It will be necessary to have scanner hole probe in the hole standard to complete this process.

Press MAIN function key then FILTER SETUP soft key, this will take you to filter set-up for Hi & Low pass filter.
20

Scanner Bolt Holt Testing

Instrument Set-up; Set-6a Adjustment for HI & LOW PASS FILTER.

To adjust HIPASS & LOWPASS soft key and rotate Smart knob as required per procedure requirements.
21

Scanner Bolt Holt Testing

Now you will be ready to do scanner bolt hole inspection.

22

Scanner Bolt Holt Testing

23

Scanner Bolt Holt Testing

24

Thank you and safe travels with Olympus NDT