Tagging photons with gold nanoparticles as

localized absorbers in optical measurements in


turbid media
Serge Grabtchak,1,2,* Kristen B. Callaghan,1 William M. Whelan1,3
1
Department of Physics, University of Prince Edward Island, Charlottetown, PEI C1A4P3, Canada
2
Departments of Electrical and Computer Engineering, and Physics, Dalhousie University, Halifax, NS B3H3J5,
Canada
3
Atlantic Veterinary College, Charlottetown, PEI C1A4P3, Canada
*
sgrabtchak@upei.ca

Abstract: We analyze a role of a localized inclusion as a probe for spatial


distributions of migrating photons in turbid media. We present new
experimental data and two-dimensional analysis of radiance detection of a
localized absorptive inclusion formed by gold nanoparticles in Intralipid-
1% when the target is translated along the line connecting the light source
and detector. Data are analyzed using the novel analytical expression for the
relative angular photon distribution function for radiance developed by
extending the perturbation approach for fluence. Obtained photon maps
allow predicting conditions for detectability of inclusions for which
proximity to the detector is essential.
©2013 Optical Society of America
OCIS codes: (170.3660) Light propagation in tissues; (170.6510) Spectroscopy, tissue
diagnostics; (170.7050) Turbid media; (290.4210) Multiple scattering.

References and links


1. T. Kitai, T. Inomoto, M. Miwa, and T. Shikayama, “Fluorescence navigation with indocyanine green for
detecting sentinel lymph nodes in breast cancer,” Breast Cancer 12(3), 211–215 (2005).
2. B. E. Schaafsma, J. S. D. Mieog, M. Hutteman, J. R. van der Vorst, P. J. Kuppen, C. W. Löwik, J. V. Frangioni,
C. J. van de Velde, and A. L. Vahrmeijer, “The clinical use of indocyanine green as a near-infrared fluorescent
contrast agent for image-guided oncologic surgery,” J. Surg. Oncol. 104(3), 323–332 (2011).
3. M. Kacprzak, A. Liebert, P. Sawosz, N. Zolek, D. Milej, and R. Maniewski, “Time-resolved imaging of
fluorescent inclusions in optically turbid medium - phantom study,” Opto-Electron. Rev. 18(1), 37–47 (2010).
4. E. Boisselier and D. Astruc, “Gold nanoparticles in nanomedicine: preparations, imaging, diagnostics, therapies
and toxicity,” Chem. Soc. Rev. 38(6), 1759–1782 (2009).
5. E. C. Dreaden, A. M. Alkilany, X. Huang, C. J. Murphy, and M. A. El-Sayed, “The golden age: gold
nanoparticles for biomedicine,” Chem. Soc. Rev. 41(7), 2740–2779 (2012).
6. R. Wilson, “The use of gold nanoparticles in diagnostics and detection,” Chem. Soc. Rev. 37(9), 2028–2045
(2008).
7. X. H. Gao, Y. Y. Cui, R. M. Levenson, L. W. K. Chung, and S. M. Nie, “In vivo cancer targeting and imaging
with semiconductor quantum dots,” Nat. Biotechnol. 22(8), 969–976 (2004).
8. C. J. Wen, L. W. Zhang, S. A. Al-Suwayeh, T. C. Yen, and J. Y. Fang, “Theranostic liposomes loaded with
quantum dots and apomorphine for brain targeting and bioimaging,” Int. J. Nanomedicine 7, 1599–1611 (2012).
9. E. Hutter and D. Maysinger, “Gold Nanoparticles and Quantum Dots for Bioimaging,” Microsc. Res. Tech.
74(7), 592–604 (2011).
10. S. Fantini, M. A. Franceschinifantini, J. S. Maier, S. A. Walker, B. Barbieri, and E. Gratton, “Frequency-domain
mutlichannel optical detector for noninvasive tissue spectroscopy and oximetry,” Opt. Eng. 34(1), 32–42 (1995).
11. A. Pellicer and M. C. Bravo, “Near-infrared spectroscopy: A methodology-focused review,” Semin. Fetal
Neonatal Med. 16(1), 42–49 (2011).
12. M. Ono, Y. Kashio, M. Schweiger, H. Dehghani, S. R. Arridge, M. Firbank, and E. Okada, “Topographic
distribution of photon measurement density functions on the brain surface by hybrid radiosity-diffusion method,”
Opt. Rev. 7(5), 426–431 (2000).
13. H. Dehghani, S. R. Arridge, M. Schweiger, and D. T. Delpy, “Optical tomography in the presence of void
regions,” J. Opt.Soc. Am. A. 17, 1659–1670 (2000).
14. P. R. Bargo, S. A. Prahl, T. T. Goodell, R. A. Sleven, G. Koval, G. Blair, and S. L. Jacques, “In vivo
determination of optical properties of normal and tumor tissue with white light reflectance and an empirical light
transport model during endoscopy,” J. Biomed. Opt. 10(3), 034018 (2005).

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 2989
15. K. Bensalah, A. Tuncel, D. Peshwani, I. Zeltser, H. Liu, and J. Cadeddu, “Optical reflectance spectroscopy to
differentiate renal tumor from normal parenchyma,” J. Urol. 179(5), 2010–2013 (2008).
16. D. A. Boas, M. A. O’Leary, B. Chance, and A. G. Yodh, “Detection and characterization of optical
inhomogeneities with diffuse photon density waves: A signal-to-noise analysis,” Appl. Opt. 36(1), 75–92 (1997).
17. V. Ntziachristos, A. G. Yodh, M. Schnall, and B. Chance, “Concurrent MRI and diffuse optical tomography of
breast after indocyanine green enhancement,” Proc. Natl. Acad. Sci. U.S.A. 97(6), 2767–2772 (2000).
18. M. G. Giacomelli and A. Wax, “Imaging beyond the ballistic limit in coherence imaging using multiply scattered
light,” Opt. Express 19(5), 4268–4279 (2011).
19. A. Laidevant, L. Hervé, M. Debourdeau, J. Boutet, N. Grenier, and J. M. Dinten, “Fluorescence time-resolved
imaging system embedded in an ultrasound prostate probe,” Biomed. Opt. Express 2(1), 194–206 (2011).
20. X. Montet, V. Ntziachristos, J. Grimm, and R. Weissleder, “Tomographic fluorescence mapping of tumor
targets,” Cancer Res. 65(14), 6330–6336 (2005).
21. A. P. Mosk, A. Lagendijk, G. Lerosey, and M. Fink, “Controlling waves in space and time for imaging and
focusing in complex media,” Nat. Photonics 6(5), 283–292 (2012).
22. D. Piao, H. Xie, W. L. Zhang, J. S. Krasinski, G. L. Zhang, H. Dehghani, and B. W. Pogue, “Endoscopic, rapid
near-infrared optical tomography,” Opt. Lett. 31(19), 2876–2878 (2006).
23. S. A. Walker, D. A. Boas, and E. Gratton, “Photon density waves scattered from cylindrical inhomogeneities:
theory and experiments,” Appl. Opt. 37(10), 1935–1944 (1998).
24. M. Xu, M. Alrubaiee, S. K. Gayen, and R. R. Alfano, “Three-dimensional localization and optical imaging of
objects in turbid media with independent component analysis,” Appl. Opt. 44(10), 1889–1897 (2005).
25. M. E. Zevallos, S. K. Gayen, M. Alrubaiee, and R. R. Alfano, “Time-gated backscattered ballistic light imaging
of objects in turbid water,” Appl. Phys. Lett. 86(1), 011115 (2005).
26. S. R. Arridge, “Methods in diffuse optical imaging,” Philos Trans A Math Phys Eng Sci 369(1955), 4558–4576
(2011).
27. S. C. Feng, F. A. Zeng, and B. Chance, “Photon migration in the presence of a single defect: a perturbation
analysis,” Appl. Opt. 34(19), 3826–3837 (1995).
28. S. B. Colak, D. G. Papaioannou, G. W. ’t Hooft, M. B. van der Mark, H. Schomberg, J. C. Paasschens, J. B.
Melissen, and N. A. van Asten, “Tomographic image reconstruction from optical projections in light-diffusing
media,” Appl. Opt. 36(1), 180–213 (1997).
29. O. Barajas, A. M. Ballangrud, G. G. Miller, R. B. Moore, and J. Tulip, “Monte Carlo modelling of angular
radiance in tissue phantoms and human prostate: PDT light dosimetry,” Phys. Med. Biol. 42(9), 1675–1687
(1997).
30. S. Grabtchak, T. J. Palmer, and W. M. Whelan, “Detection of localized inclusions of gold nanoparticles in
Intralipid-1% by point-radiance spectroscopy,” J. Biomed. Opt. 16(7), 077003 (2011).
31. A. M. Ballangrud, P. J. Wilson, K. Brown, G. G. Miller, R. B. Moore, M. S. McPhee, and J. Tulip, “Anisotropy
of radiance in tissue phantoms and Dunning R3327 rat tumors: Radiance measurements with flat cleaved fiber
probes,” Lasers Surg. Med. 19(4), 471–479 (1996).
32. D. Dickey, O. Barajas, K. Brown, J. Tulip, and R. B. Moore, “Radiance modelling using the P3 approximation,”
Phys. Med. Biol. 43(12), 3559–3570 (1998).
33. D. J. Dickey, R. B. Moore, D. C. Rayner, and J. Tulip, “Light dosimetry using the P3 approximation,” Phys.
Med. Biol. 46(9), 2359–2370 (2001).
34. L. C. L. Chin, A. E. Worthington, W. M. Whelan, and I. A. Vitkin, “Determination of the optical properties of
turbid media using relative interstitial radiance measurements: Monte Carlo study, experimental validation, and
sensitivity analysis,” J. Biomed. Opt. 12(6), 064027 (2007).
35. L. C. L. Chin, B. Lloyd, W. M. Whelan, and I. A. Vitkin, “Interstitial point radiance spectroscopy of turbid
media,” J. Appl. Phys. 105(10), 102025 (2009).
36. S. Grabtchak, T. J. Palmer, and W. Whelan, “Radiance spectroscopy tool box for characterizing Au
nanoparticles in tissue mimicking phantoms as applied to prostate,” J. Cancer Sci. Ther. S1–008 (2011).
37. S. Grabtchak, T. J. Palmer, I. A. Vitkin, and W. M. Whelan, “Radiance detection of non-scattering inclusions in
turbid media,” Biomed. Opt. Express 3(11), 3001–3011 (2012).
38. S. Grabtchak and W. M. Whelan, “Separation of absorption and scattering properties of turbid media using
relative spectrally resolved cw radiance measurements,” Biomed. Opt. Express 3(10), 2371–2380 (2012).
39. S. Grabtchak, E. Tonkopi, and W. M. Whelan, “Optical detection of gold nanoparticles in a prostate-shaped
porcine phantom,” J. Biomed. Opt. 18(7), 077005 (2013).
40. A. Liemert and A. Kienle, “Analytical Green’s function of the radiative transfer radiance for the infinite
medium,” Phys. Rev. E Stat. Nonlin. Soft Matter Phys. 83(3), 036605 (2011).
41. L. C. L. Chin, W. M. Whelan, and I. A. Vitkin, “Information content of point radiance measurements in turbid
media: implications for interstitial optical property quantification,” Appl. Opt. 45(9), 2101–2114 (2006).
42. M. S. Patterson, B. C. Wilson, and D. R. Wyman, “The propagation of optical radiation in tissue I. Models of
radiation transport and their application,” Lasers Med. Sci. 6(2), 155–168 (1991).
43. M. S. Patterson, B. C. Wilson, and D. R. Wyman, “The propagation of optical radiation in tissue II. Optical
properties of tissues and resulting fluence distributions,” Lasers Med. Sci. 6(4), 379–390 (1991).
44. W. M. Star, “Light dosimetry in vivo,” Phys. Med. Biol. 42(5), 763–787 (1997).
45. A. M. Siegel, J. J. A. Marota, and D. A. Boas, “Design and evaluation of a continuous-wave diffuse optical
tomography system,” Opt. Express 4(8), 287–298 (1999).
46. A. Dimofte, J. C. Finlay, and T. C. Zhu, “A method for determination of the absorption and scattering properties
interstitially in turbid media,” Phys. Med. Biol. 50(10), 2291–2311 (2005).

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 2990
47. T. C. Zhu, J. C. Finlay, and S. M. Hahn, “Determination of the distribution of light, optical properties, drug
concentration, and tissue oxygenation in-vivo in human prostate during motexafin lutetium-mediated
photodynamic therapy,” J. Photochem. Photobiol. B 79(3), 231–241 (2005).
48. H. P. Xu and M. S. Patterson, “Determination of the optical properties of tissue-simulating phantoms from
interstitial frequency domain measurements of relative fluence and phase difference,” Opt. Express 14(14),
6485–6501 (2006).
49. K. K. H. Wang and T. C. Zhu, “Reconstruction of in-vivo optical properties for human prostate using interstitial
diffuse optical tomography,” Opt. Express 17(14), 11665–11672 (2009).
50. H. J. van Staveren, C. J. M. Moes, J. van Marie, S. A. Prahl, and M. J. C. van Gemert, “Light scattering in
Intralipid-10% in the wavelength range of 400-1100 nm,” Appl. Opt. 30(31), 4507–4514 (1991).
51. B. Chance, K. Kang, L. He, J. Weng, and E. Sevick, “Highly sensitive object location in tissue models with
linear in-phase and anti-phase multi-element optical arrays in one and two dimensions,” Proc. Natl. Acad. Sci.
U.S.A. 90(8), 3423–3427 (1993).
52. B. J. Tromberg, L. O. Svaasand, T. T. Tsay, and R. C. Haskell, “Properties of photon density waves in multiple-
scattering media,” Appl. Opt. 32(4), 607–616 (1993).
53. F. Bevilacqua, A. J. Berger, A. E. Cerussi, D. Jakubowski, and B. J. Tromberg, “Broadband absorption
spectroscopy in turbid media by combined frequency-domain and steady-state methods,” Appl. Opt. 39(34),
6498–6507 (2000).
54. Y. Yamada, “Fundamental studies of photon migration in biological tissues and their application to optical
tomography,” Opt. Rev. 7(5), 366–374 (2000).
55. R. B. Schulz, J. Ripoll, and V. Ntziachristos, “Noncontact optical tomography of turbid media,” Opt. Lett.
28(18), 1701–1703 (2003).
56. B. W. Pogue, S. C. Davis, X. M. Song, B. A. Brooksby, H. Dehghani, and K. D. Paulsen, “Image analysis
methods for diffuse optical tomography,” J. Biomed. Opt. 11(3), 033001 (2006).
57. T. Durduran, R. Choe, W. B. Baker, and A. G. Yodh, “Diffuse optics for tissue monitoring and tomography,”
Rep. Prog. Phys. 73(7), 076701 (2010).
58. S. Grabtchak, T. J. Palmer, F. Foschum, A. Liemert, A. Kienle, and W. M. Whelan, “Experimental spectro-
angular mapping of light distribution in turbid media,” J. Biomed. Opt. 17(6), 067007 (2012).
59. F. Martelli, M. Bassani, L. Alianelli, L. Zangheri, and G. Zaccanti, “Accuracy of the diffusion equation to
describe photon migration through an infinite medium: numerical and experimental investigation,” Phys. Med.
Biol. 45(5), 1359–1373 (2000).
60. A. Ishimaru, Wave Propagation and Scattering in Random Media (Academic Press New York, 1978).
61. S. K. Ghosh and T. Pal, “Interparticle coupling effect on the surface plasmon resonance of gold nanoparticles:
From theory to applications,” Chem. Rev. 107(11), 4797–4862 (2007).
62. S. Carraresi, T. S. Shatir, F. Martelli, and G. Zaccanti, “Accuracy of a perturbation model to predict the effect of
scattering and absorbing inhomogeneities on photon migration,” Appl. Opt. 40(25), 4622–4632 (2001).
63. A. Sassaroli, F. Martelli, and S. Fantini, “Higher-order perturbation theory for the diffusion equation in
heterogeneous media: application to layered and slab geometries,” Appl. Opt. 48(10), D62–D73 (2009).
64. F. Martelli, A. Pifferi, D. Contini, L. Spinelli, A. Torricelli, H. Wabnitz, R. Macdonald, A. Sassaroli, and G.
Zaccanti, “Phantoms for diffuse optical imaging based on totally absorbing objects, part 1: Basic concepts,” J.
Biomed. Opt. 18(6), 066014 (2013).

1. Introduction
Advances in detection of localized inclusions in multiple-scattering biological tissues (i.e.,
turbid or diffusive media) are mostly driven by needs of practical applications of biomedical
optics. In general, the localized inclusion can be considered as a region in turbid media with
optical properties different from those of the background. Such regions may contain
exogenous (i.e., indocyanine green [1–3], gold nanoparticles [4–6], quantum dots [7–9] etc.)
or endogenous (i.e., hemoglobin [10, 11]) chromophores, nonscattering liquids [12, 13] (i.e.,
cerebrospinal fluid, water) or malignant tissues [14, 15]. Various approaches have been
developed to detect objects of certain geometries in turbid media [16–26].
However, rather than being of a primary interest by themselves, localized inclusions can
play a different role by acting as tools for probing spatial distributions of migrating photons in
turbid media which in turn may provide researchers with a better understanding of conditions
for detectability of inclusions.
Introducing a localized inclusion with optical properties different from the surrounding
medium produces a perturbation that changes the original photon density distribution. In the
formalism of the perturbation approach [27, 28] developed for fluence with the diffusion
approximation, the distribution of photons can be reconstructed using a point localized
absorber as a probe that perturbs photon density in the selected location. When a small
localized absorber is placed at some position between the light source and fluence-based
detector, a reduction in the resulting fluence is measured. The measurements are repeated

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 2991
after moving the absorber to a new position eventually covering most of the plane (or even
the volume) of interest, and finally the photon-path-distribution function can be reconstructed
from multiple point measurements. In essence, this process is equivalent to tagging photons
with the localized absorber because only photons that pass through the region of the physical
location of the absorber will be removed from a subsequent migration and detected as a
reduction in the signal. Analytical expressions for the photon-path-distribution function
derived in [27, 28] allowed predicting of so-called banana-shape regions where the flux is
concentrated inside the turbid medium (with photon densities higher near the source and
detector and lower at the midway point). This is important because it can provide clear
information about optimal detector sensitivity to tumors or another localized inhomogeneity
in the selected point in a turbid medium [27].
Two important aspects of the perturbation approach from Ref [27]. should be emphasized:
1) the distribution function describes the fraction of the total number of photons diffusing
from the source to the detector that must travel through the selected point in space; 2) the
distribution function assumes an isotropic 4π collection window for the detector that
corresponds to fluence detection. If the detector has an anisotropic detection window as in the
case of radiance detection [29, 30], then the photon density distribution function derived for
fluence can’t be applied.
Detecting optical radiance is less understood compared to fluence for studying
propagation of light in turbid media with only a few groups of researchers exploring its utility
[29–39]. This is likely due to the relative complexity of radiance-based techniques with an
angular-sensitive detector and rotation along its axis which imposes certain difficulties for
practical use. On the other hand, the lack of a practical demand for radiance applications has
produced only few theoretical studies [38, 40, 41] that would enrich experiments and
stimulate further developments as was done for fluence based techniques. Indeed, the
simplicity of basic fluence measurements, the equivalency between the detector and the light
source (for example, the isotropic spherical diffuser can act in both capacities) and a relative
straightforwardness of results interpretation have triggered a spectacular growth in
applications of fluence-based techniques [12, 20, 22, 42–57]. Hence, the lack of the
perturbation formalism that can be easily applied to analysis of radiance data for localized
inclusions in turbid media seriously impedes further progress of the radiance-based methods.
The current work presents new experimental data and novel two-dimensional (2D)
analysis of radiance detection of the localized absorptive inclusion formed by gold
nanoparticles (Au NPs) in the infinite turbid medium (Intralipid-1%) geometry when the
absorptive target is translated along the line connecting the light source and detector. This is
the configuration that produces “banana” patterns in fluence measurements, and it was used
earlier by us to study effects of voids in turbid media [37]. Radiance data were presented as
spectro-angular maps of the radiance extinction ratio (RER) that were introduced previously
[37, 39]. The maps were analyzed using the novel analytical expression for the relative
angular photon distribution function developed in the current work within the framework of
the perturbation approach as in [27] but extended to radiance.
2. Methods and materials
2.1. Set-up for radiance measurements in Intralipid-1%
Optical radiance corresponds to a variation in the angular distribution of photons measured in
the selected point in turbid media. To obtain radiance data, an optical probe with a well-
defined angular detection window must be rotated along its axis. A schematic of the
experimental set-up is shown in Fig. 1(a). The set-up has been used in number of publications
[30, 39, 58] so only a brief description is given here. Tissue illumination was provided by a
tungsten halogen white light source (20 W) connected to a fiber with a 2-mm spherical
diffuser at the end with a total output power ~18 mW. A 600-micron side firing fiber (Pioneer
Optics) acted as a radiance detector. To prevent any secondary rays that do not undergo the
total internal reflection to couple back to the fiber from other directions, a cap from a heat-

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 2992
shrink white polymer tube with a hole aligned to the entrance window was placed at the end
of the side-firing fiber [39]. Both illuminating and detecting fibers were threaded through 15-
gauge needles for mechanical stability. The side firing fiber was mounted on a computer-
controlled rotation stage (Thorlabs). Radiance data were obtained by rotating the side firing
fiber over a 360° range with a 2° step. The side firing fiber was connected to a computer-
controlled USB 4000 spectrometer (Ocean Optics) that collected spectra at every angular
step. Depending on a number of averages, it took between 4 and 8 minutes to acquire a
complete single angular profile.

Fig. 1. a) Experimental set-up for radiance measurements of gold target in a Intralipid-1%, b)


conceptual top-view diagram illustrating the principle of distance dependent experiments.

The sample containing the capillary filled with the colloidal gold was controlled in the
X,Y-plane by two translation stages. Au NPs based inclusion was prepared by filling two
types of capillaries with the colloidal gold: ~3.5-mm (~0.8 mL, ~6 cm immersion length) and
2-mm diameter (~0.1 mL, ~6 cm immersion length) thin-wall quartz capillary tubes. The
colloids (BBI Solutions) were formed by spherical 5-nm Au NPs with the plasmon resonance
at 520 nm. Two different stock concentrations were used in various experiments: 3.95x1013
and 7x1014 particles/mL (referred as 20x in the text).
The Lucite box with blackened 18-cm walls was filled with Intralipid-1% suspension and
accommodated both detecting and illuminating fibers as well as the capillary tube with the
colloidal gold. Intralipid-1% suspension was prepared by volume dilution of Intralipid-20%
stock (Sigma Aldrich Canada). The size of the box ensured an approximation of infinite
medium geometry.
For distance dependent experiments, the source-detector separation was kept constant at
30 mm, and the target was translated with 5 mm (2 mm in some experiments) increments
from 5 mm (3 mm in some experiments) to 25 mm away from the detector keeping it in line
between the source and detector (i.e., at 0° angle) as shown in Fig. 1(b). In all experiments,
for every target position a set of radiance profiles was acquired by rotating the detecting fiber
over a 360° range with a 2° step. The reference measurement was taken in the Intralipid
solution for the corresponding source-detector separation prior to inserting the capillary with
Au NPs.
Optical properties of Intralipid-1% were extensively studied by us earlier [58], and
published data for the effective attenuation coefficient (μeff(λ)) and the diffusion constant
(D(λ)) were used in simulations in the current work. Of a particular importance for the current
work are the spectral dependencies of μeff(λ) for Intralipid-1% and the absorption coefficient
(μa(λ)) for 5-nm Au colloidal solutions that are shown in Fig. 2(a), 2(b).

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 2993
Fig. 2. a) Effective attenuation coefficient of Intralipid-1% and absorption coefficient of 5-nm
Au NPs colloidal solutions with a) 3.95x1013 particles/mL and b) 7x1014 particles/mL (marked
as 20x) concentrations.

Radiance data were presented in a form of spectro-angular maps that define relative
photon distribution in the angular domain. Spectro-angular mapping approach relies on
obtaining two data sets, one for the phantom without the inclusion and the other for the
phantom with the inclusion. Individual spectra (i.e., radiance profiles) collected at various
radiance detector angles in 0°-360° range in the phantom without inclusion were assembled
into “the Intralipid matrix”, IIntralipid with columns corresponding to wavelengths and rows to
angles. Placing the Au target at a certain distance from the detector along a selected angle and
repeating the measurements yielded “the Intralipid+Au matrix”, IIntralipid + Au. An element by
element ratio of two matrices, IIntralipid/IIntralipid + Au minimized contributions from Intralipid,
white light source and illuminating/detecting fibers, highlighting the target signal in a
particular background. The ratio is defined as the Radiance Extinction Ratio (RER) and can
be presented via surface or contour plots demonstrating intensity and spectral variations vs.
angle [37, 39].
2.2. Perturbation modeling for radiance
Diffusion approximation (DA) is a simplification which allows one to obtain an analytical
solution of the radiative transfer equation (RTE) and thus, analytical expressions for fluence
and radiance. This simplification can be achieved when absorption is much less than
scattering (which is true for most biological tissues) and when measurements are performed at
distances from the light source that are much bigger than the reciprocal reduced scattering
coefficient. A clear physical meaning and the ease of manipulation with fluence or radiance
under DA makes it attractive for testing theoretical predictions against experimentally
observed effects.
Consider the isotropic dc light source placed at rS and the radiance detector located at rD
in the infinite homogenous medium that exhibits absorption and multiple scattering effects
and can be treated within the framework of the diffusion theory. Since radiance measurements
are always performed interstitially inside tissues of interest, infinite medium geometry
corresponds to practical applications of radiance. Radiance measured by the detector can be
described by a familiar expression [59]:

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 2994
S0   D (λ )  
I 0 ( rS − rD , θ , λ ) = 1 + 3 + μ eff ( λ ) D ( λ )  cos θ 
π D(λ )   rS − rD  
2 2
4
(1)
exp ( − μ eff ( λ ) rS − rD )

rS − rD

where S0 is the source power, rS − rD is the source-detector separation, θ is the angle between
the direction of propagation and scattering, λ is the wavelength, D ( λ ) = 1 3( μ ′ ( λ ) + μ ( λ )) is the s a

diffusion coefficient, μeff(λ) is the effective attenuation coefficient of the background turbid
medium.
Once the inclusion is formed at position r by an absorbing sphere of radius a, the new
value of radiance affected by the perturbation, I ( r − r , θ , λ ) will be measured. The radiance
pert S D

extinction ratio (RER) introduced earlier can be constructed from two values as RER =
I ( r − r , θ , λ ) / I ( r − r , θ , λ ) . This ratio singles out properties of the inclusion making the
0 S D pert S D

absolute radiance measurements unnecessary [37]. The ratio also emphasizes attenuation of
light in the phantom with the inclusion such that values of RER>1correspond to the increased
extinction relative to the bare phantom while values of RER<1indicate that addition of the
inclusion reduces the overall extinction and results in a higher photon density. When the
inclusion is added to the phantom it replaces the equivalent volume of the background
medium, and the differential optical properties of two volumes will play an important role in
determining the directional changes in RER values. In principle, different perturbation
metrics can be introduced from RER as { I ( r − r ,θ , λ) - 0 S D

I ( r − r , θ , λ ) }/ I ( r − r , θ , λ ) ≡ΔI/ I 0 but as will be shown later, it bears essentially the


pert S D 0 S D

same information as RER.


Following the diffusion approximation, radiance can be approximated by first two terms
(isotropic fluence and anisotropic flux) of a series expansion in spherical harmonics [60]:
1 3
I pert ( rS − rD , θ , λ ) = Φ pert ( rS − rD , λ ) + D ( λ )∇Φ pert ( rS − rD , λ ) ⋅ sˆ (2)
4π 4π

where Φ pert ( rS − rD , λ ) is the fluence measured in the presence of the perturbation, ŝ is the unit
vector pointing to the direction of the perturbed flux propagation. One can see that knowledge
of the perturbed fluence allows determining perturbed radiance.
In order to obtain the expression for fluence in the presence of the inclusion, it is
instructive to invoke the perturbation approach developed for fluence for the infinite medium
geometry in the steady-state condition for a perfectly absorbing sphere [27, 28]. Following
this approach, the perturbed fluence can be presented as a sum of the fluence measured in the
absence of the perturbation and the fluence that is due to photons that come from the source rS
and pass through the region occupied by the inclusion at r, because only these photons will be
removed by absorption of the inclusion:
Φ pert ( rS − rD , λ ) = Φ 0 ( rS − rD , λ ) + Φ 1 ( rS − rD , r , λ ) (3)

The corresponding expressions for both terms from Eq. (3) are written below (the first is
known from the diffusion theory and the second was derived in [27]):
S0 exp ( − μ eff ( λ ) rS − rD )
Φ 0 ( rS − rD , λ ) = (4)
4π D ( λ ) rS − rD

aS 0 exp ( − μeff ( λ ) r − rS ) exp ( − μ ( λ ) ( rD − r − a ) )


Φ 1 ( rS − rD , r , λ ) = − k
eff
(5)
4π D ( λ ) r − rS rD − r

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 2995
where a is the radius of the absorbing sphere and k is the coefficient introduced by us when
defining the boundary condition on the surface of the absorbing sphere and is set to 1 for the
perfect absorber such that Eq. (5) matches the corresponding expression from Ref [27].
Note, that the positions of the source and detector are fixed while the inclusion can be
moved which makes r a spatial variable. Substituting Eq. (3) into Eq. (2) results in the
following expression:
1 1
I pert ( rS − rD , θ , λ ) = Φ 0 ( rS − rD , λ ) + Φ 1 ( rS − rD , r , λ )
4π 4π
(6)
3 3
+ D ( λ ) ∇Φ 0 ( rS − rD , λ ) ⋅ sˆ + D ( λ ) ∇Φ 1 ( rS − rD , r , λ ) ⋅ sˆ
4π 4π

Inspection of Eq. (6) indicates that since the third term does not depend on the variable r, it
will be zero. The next step is performing necessary differentiation in Eq. (6) and setting the
coordinate system. In the coordinate system the detector lies at the origin, the source is placed
at the distance d from the detector along X-axis and the inclusion is moved from the detector
to the source along X-axis. Then, r transforms to x and the following expressions can be
written for the perturbed and unperturbed radiance:
S0  exp ( − μ ( λ ) d ) exp ( − μ ( λ ) ( d − a ) )
I pert ( d , θ , λ ) = − ak
eff eff

2 2 
4 π D (λ )  d x(d − x )
(7)
( d − 2 x ) exp ( − μ eff ( λ ) ( d − a ) ) 
− 3akD ( λ ) cos θ

(d − x ) x 
2 2

I 0 (d , θ , λ ) =
S0 1 + 3  D ( λ ) + μ ( λ ) D ( λ )  cos θ  exp ( − μ ( λ ) d ) (8)
 
eff

4 π D ( λ )   d  
2 2 eff
d

It should be noted that the selected geometry has 2D symmetry such that the absorbing sphere
(or cylinder) is reduced to a circle and rotation of the detector and translation of the target are
performed in the same plane. The expression for RER will take the final form as following:
I 0 (d , θ , λ )
RER =
I pert ( d , θ , λ )

 D ( λ ) + μ ( λ ) D ( λ )  cos θ
1 + 3 (9)

 d 
eff


adk 3aD ( λ )( d − 2 x ) dk
1− exp ( − μ eff ( λ ) ( d − a ) ) − exp ( − μ eff ( λ ) ( d − a ) ) cos θ
x(d − x ) (d − x ) x
2 2

As follows from Eq. (9), modeling RER requires the knowledge of the measurement
geometry, size of the inclusion and optical properties of the background medium.
While the perfect absorber can be a useful concept, realistic absorbers exhibit finite
absorption values. For realistic absorbers, the fluence on the surface of the absorber does not
vanish completely but is reduced to some non-zero value so that k≠1. To account for the finite
absorption of the inclusion, we used a simple approximation based on the Beer-Lambert law
of optical attenuation. The absorbing circle of radius a (for 2D case) was assigned an
absorption coefficient μa(λ). The circle was subject to uniform illumination from all
directions. The fluence at the surface is determined by the transmission through the
inclusion’s medium with a 2a optical path that has replaced the similar shape of the turbid
medium and is also immersed into the turbid medium characterized by μeff(λ). Then, fluence
on the surface of the absorber was approximated to be reduced
by k = (1 − exp[ −( μ ( λ ) − μ ( λ ))2a ]) . This new boundary condition was incorporated via k into
a eff

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 2996
the derivation (i.e., Eqs. (5), (7), (9)) so that Eq. (9) can be used for both perfect and finite
absorbers given by k values.
3. Results and discussion
Presenting data via RER is extremely useful for isolating signatures of the localized inclusion
in spectral and angular domains in radiance measurements. To illustrate this, Fig. 3 shows
spectro-angular contour plots of raw radiance data for the Intralipid phantom (Fig. 3(a)) and
for the phantom with the 3.5 mm Au NPs inclusion (3.95x1013 particles/mL) (Fig. 3(b)) that
was placed between the source and detector (i.e., at 0° angle) at a distance of 5 mm from the
detector (with source-detector separation at 30 mm). In the color coding used in the contour
plots, red corresponds to a high signal with a maximal optical transmission while blue
highlights areas with lowered transmission. For both plots, the maximal transmission is
observed along 0° angle that is consistent with preferential forward scattering in Intralipid
solutions [58]. Spectral responses of both plots include contributions from wavelength-
dependent white light source, illuminating and detecting fibers. (Optical signatures of the
Intralipid solution can be isolated by normalizing to measurements in water [58].) Close
examination of two plots suggests that other than a red-shift of the maximal transmission for
Fig. 3(b), nothing indicates the presence at 0° of the Au NPs inclusion with certain spectral
characteristics.

Fig. 3. a) Spectro-angular contour plot of raw radiance data for Intralipid-1% ; b) spectro-
angular contour plot of raw radiance for Intralipid-1% with 3.5 mm inclusion with Au NPs
located at 0° and 5mm from the detector. Color bars have units of counts per second. Source-
detector separation 30 mm.

However, the ratio of the two plots expressed as RER in Fig. 4(a) clearly isolates the
signature of the inclusion. Red color corresponds to the regions of highest attenuation that
occurs in the vicinity of 520 nm along 0° angle. Figure 4(b) presents the same data as the
surface plot that emphasizes the spectral response which is characteristic for 5-nm Au NPs in
water (compare with Fig. 2(a)). Figure 4(a) also shows that RER undergoes changes with
wavelength such that for λ>680 nm RER<1. The contour line corresponding to RER = 1 is
presented in white. Reduction in RER values indicates that photon density has increased for
certain wavelengths relative to Intralipid solution after inserting the Au NPs inclusion. (Some
asymmetry seen in RER = 1 contour line was caused by a small angular misalignment in the
position of the inclusion). Thus, away from the plasmon resonance, the Au NPs inclusion not
only can lose its strong absorptive properties but, in fact, may even become more transparent
than the background turbid medium it has replaced.

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 2997
Fig. 4. a) Spectro-angular contour plot of the radiance extinction ratio (RER) the Au NPs
inclusion. b) Spectro-angular surface plot of the radiance extinction ratio (RER) of the Au NPs
inclusion. Inclusion: 3.5 mm capillary with 5nm Au NPs at 0° at 5 mm from the detector.
Source-detector separation 30 mm.

Continuing to move the Au NPs inclusion with 5 mm step toward the light source yields
four more spectro-angular contour plots (target at 10 mm, 15mm, 20 mm and 25 mm away
from the detector) presented in Figs. 5(a), 5(b), 5(c), 5(d)). A characteristic absorption peak
around 520 nm is observed in all plots. A noticeable angular broadening manifested in the
increased horizontal ellipticity of contour lines while moving the inclusion away from the
detector is another characteristic feature. From the intensity scale, one can see that RER tends
to achieve higher values when the inclusion is closer to the source or to the detector while in
the middle of the distance RER is minimal. These observations are quantified in Fig. 6 which
shows the variation of maximal values of RER (observed along 0°) and Full-Width-Half-
Maximum (FWHM) of the angular profile at 520 nm with distance.
Because it is extremely difficult to present the variation of RER in the spectral, spatial and
angular domains at the same time, we concentrated on two most interesting and important
representations. Since Au NPs are used mostly because of their optical properties at the
plasmon resonance, the first plot in Fig. 7(a) represents the distance variation in the relative
angular photon density distribution probed by the Au NPs inclusion at the wavelength of the
plasmon resonance, 520 nm. To create this plot, five individual angular profiles of RER at
520 nm measured at 5, 10, 15, 20, and 25 mm (from Figs. 4(a), 5(a)-5(d)) were combined into
the spatio-angular contour plot as in Fig. 7(a). Due to the nature of interaction between the
inclusion and photons, highest values of RER correspond to highest photon densities such that
the plot can be considered as a negative image of the photon density in the turbid medium.
Since the radius of contour lines corresponds to angular distribution of photons, one can see
that photons are better localized in the vicinity of the detector rather than of the light source.
Indeed, the radius of contour lines gradually increases while moving the target from the
detector to light source. It becomes more obvious when comparing the profiles measured at 5
and 25 mm because these two positions are symmetric relative to the midpoint (i.e., both have
5 mm separation from the detector or from the source). Variation of RER values along 0°
angle exhibits a minimum midway between the source and the detector while showing rising
toward the source and detector. Applying this knowledge to detection of inclusions means
that the closer the inclusion is to the detector, the better the angular detectability (or angular
localization). When data from Fig. 7(a) were represented (not shown) via the spatio-angular
contour plot using different metrics, ΔI/I0 (that was described in the modeling section) the
resulting plot had the appearance very similar to the one in Fig. 7(a) with the maximal value

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 2998
at ~0.63. Since both plots contained essentially the same information and demonstrated
similar features, we proceeded with RER-based analysis.

Fig. 5. Spectro-angular contour plots of the radiance extinction ratio (RER) the Au NPs
inclusion located at a) 10 mm, b) 15 mm, c) 20 mm, d) 25 mm from the detector at 0°.
Inclusion: 3.5 mm capillary with 5nm Au NPs. Source-detector separation 30 mm.

The second way of data representation is shown in Fig. 7(b) where the variation of RER
along 0° angle in the 450-900 nm spectral range with distance is displayed as the spectro-
spatial contour plot. One can notice that almost all wavelengths exhibit the pattern observed
for 520 nm in Fig. 6 along 0° i.e., rising toward the detector and source with the minimum
half-way between the detector and source. The shape of contour lines in 450-500 nm spectral
range somewhat resembles the absorption spectrum of Au NPs with the peak at 520 nm. The
contour line with RER = 1 separates the spectral range of the increased extinction (<680 nm)
from the one of the decreased extinction (>680 nm). In accordance with RER definition,
RER<1indicates that Intralipid solution with the added inclusion becomes more transparent
than the Intralipid alone.

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 2999
Fig. 6. Variation of maximal RER values observed along 0° and full-width-half-maximum
(FWHM) measured at 520 nm with distance extracted from Figs. 4(a), 5(a)-5(d).

Fig. 7. Experiment for 3.5 mm target with 3.95x1013 Au NPs/mL: a) spatio-angular contour
plot of RER measured at 520 nm, b) spectro-spatial contour plot of RER measured along 0°.
Plots created from data in Figs. 4(a), 5(a)-5(d).

Results of modeling using Eq. (9) for the perfect absorber of 3.5 mm diameter in the
Intralipid-1% solution are presented in Fig. 8 which shows the spatio-angular contour plot of
RER at 520 nm. Simulations were always performed with the same incremental step for the
detector-target separation as in the experiment. This figure can be compared directly with Fig.
7(a). Figure 8 shows qualitatively the same behavior as the Fig. 7(a): better angular
localization of photons near the detector, much broader angular distributions of photons
closer to the light source and well-defined minimum in RER at the midpoint between the
source and the detector. Quantitative comparison indicates that simulated values are about
30% lower than the experimental ones. Also, the simulated plot shows that for the inclusion
located closer to the light source the contour lines start bending upward which is opposite to
what was measured in the experiment. This plot demonstrates the relative angular distribution
of photons at the selected wavelength in a very characteristic asymmetric “hourglass” pattern
and can be considered analogous to known banana-type patterns for fluence.

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 3000
By definition, the absorption coefficient of the perfect absorber is much larger than the
extinction of the surrounding medium. The concept of the perfect absorber might be useful
when comparing optical properties at the selected wavelength, however extending the concept
to a wider spectral range would result in the uniform absorption spectrum of the black body.
Since the absorption coefficient of the perfect absorber is not present explicitly in Eq. (9),
simulations can’t generate the plot with spectrally resolved features similar to the one in Fig.
7(b).

Fig. 8. Simulated spatio-angular contour plot of RER at 520 nm using Eq. (9) for 3.5 mm
diameter perfect absorber.

Performing simulations using Eq. (9) for the realistic absorber resulted in two plots shown
in Figs. 9(a), 9(b): spatio-angular and spectro-spatial contour plots. Comparing Fig. 9(a) with
Fig. 7(a) one can see that in spite of the increased difference in absolute values of RER
between the two plots, there is still a satisfactory qualitative correspondence. The increased
gap between the experimental and simulated values should not come as a surprise. The
perfect absorber has already introduced values that were lower by ~30% so that any further
decrease in the absorption will only increase the discrepancy.
Detailed comparison of Fig. 9(b) with Fig. 7(b) demonstrates a very good degree of
correspondence between the two plots. (Note that the plot in Fig. 9(b) appears more discrete
because fewer wavelengths were used in the simulation than in the experiment.) With the
difference in absolute values still present, the contour line patterns remain similar. Both plots
demonstrate contour lines clearly delineating the plasmon peak around 520 nm at both ends of
the detector-target separation and a very distinct evolution of contour lines up to ~700 nm.
The contour line with RER = 1 is marked in white in both contour plots. The origin of this
line becomes clear after inspecting Fig. 2(a) that contains the wavelength-dependent
absorption coefficient of Au NPs in water and the effective attenuation coefficient of
Intralipid-1%. Below ~650 nm absorption of Au NPs dominates over the attenuation of
Intralipid-1%, however for wavelength >650 nm this reverses. Once this relation was
incorporated into the boundary condition, the crossover also appears in the simulated contour
plot. It is interesting that while the crossover in Fig. 2(a) occurs at ~650 nm, for the Au NPs
inclusion placed in the turbid medium the wavelength at which the crossover is observed

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 3001
changes with distance. Not only modeling and experiment produce almost the same shape for
the corresponding contour line but the values of crossover wavelengths are close (~690 nm
and ~700 nm for 5mm detector-target separation for the experiment and theory,
correspondingly). Some deviations between the model and the experiment occur for λ>700nm
where simulation shows a slightly different spectral dependence which is also manifested in
different distance dependent patterns.

Fig. 9. Simulation of a realistic absorber (3.5 mm diameter with 3.95x1013 Au NPs/mL): a)


spatio-angular contour plot of RER at 520 nm using Eq. (9), b) spectro-spatial contour plot of
RER at 0° using Eq. (9)

In order to get closer to perfect absorber conditions, colloidal solution with a 20x
increased concentration of Au NPs (7x1014 particles/mL) was used in subsequent
experiments. To improve the spatial resolution in mapping of the relative angular photon
distribution function, the size of the capillary tube was reduced from 3.5 mm to 2 mm, and the
incremental step in the translation of the target was set to 2 mm covering the range from 3 to
25 mm away from the detector. The corresponding optical properties of the 20x Au NPs
solution and Intralipid-1% are shown in Fig. 2(b). Figure 10 presents experimental data in two
adopted formats: spatio-angular plot for 530 nm (Fig. 10(a)) and spectro-spatial plot for 0°
angle (Fig. 10(b)). (The offset seen in Fig. 10(a) in 0° position is due to the target’s
misalignment.) Increased extinction values (up to ~4) indicate that this combination of the
concentration and tube size produced much stronger perturbation than the set of parameters
used previously.
The main tendencies however remain the same: a) the extinction is maximal near the
detector, goes through the minimum midway between the source and detector and rises again
near the source to values lower than those near the detector; b) angular localization of photons
is better when the target is closer to the detector such that the relative angular photon
distribution function resembles the shape of an asymmetric hourglass. Figure 10(b) shows
values of RER>1 for the entire spectral range of interest (450-900 nm). This behavior is
consistent with Fig. 2(b) that indicates that the absorption coefficient of Au NPs is larger than
the effective attenuation coefficient of Intralipid-1% in the entire spectral range and no
crossing occurs. Another feature that is seen in both in Fig. 10(b) and Fig. 2(b) is some red
shift and spectral broadening of the plasmon resonance. Even though at low Au NP
concentrations the resonance is well defined at 520 nm (as can be seen in Fig. 2(a)), higher
concentrations result in the reduced interparticle spacing that is manifested in spectral
broadening of the plasmon resonance and red-shifting to ~530 nm. Similar effects occur
during cluster formation and aggregation of Au NPs [61]. These phenomena highlight the
importance of the actual concentration of Au NPs in studied turbid media when the local

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 3002
concentration may significantly exceed the initial stock concentration of Au NPs solution thus
affecting spectral signatures of the detected Au NPs inclusions.

Fig. 10. Experiment for 2.0 mm target with 7x1014 Au NPs/mL: a) spatio-angular contour plot
of RER measured at 530 nm, b) spectro-spatial contour plot of RER measured along 0°.

Fig. 11. Simulation of the realistic absorber (2.0 mm diameter with 7x1014 Au NPs/mL
(i.e.,20x)): a) spatio-angular contour plot of RER at 530 nm using Eq. (9), b) spectro-spatial
contour plot of RER at 0° using Eq. (9)

Results of simulations corresponding to these measurements are shown in Fig. 11(a),


11(b). Figure 11(a) displays the spatio-angular plot for 530 nm for the realistic absorber.
Simulations for the perfect absorber produced a virtually indistinguishable plot indicating that
optical properties of Au NPs near the plasmon resonance with 20x concentration approximate
well those of the perfect absorber. Comparing Fig. 11(a) with Fig. 10(a) shows a reasonable
degree of qualitative agreement between the experiment and the theory: patterns of contour
lines are very similar as well as the degree of angular localization of photons near the detector
and source with the increased degree of deviation closer to the light source. Absolute values
of theoretical RER are about twice as low as the measured values. As in the case of the
previous sample, the flatness of contour lines in the vicinity of the light source observed
experimentally indicates a stronger degree photon delocalization than is seen in simulations.

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 3003
Comparing Fig. 11(b) and Fig. 10(b) indicates that modeling correctly predicts RER>1 in
the entire spectral range, shows strong localization of photons near the detector, minimum
photon density midway between the source and the detector and lower values of RER in the
vicinity of the source. It is interesting to note a remarkable degree of similarity of contour
lines in 450-650 nm spectral range. It appears that very strong absorption and relatively broad
plasmon resonance of 20x Au NPs solution smear the plasmon signature in contour lines by
producing almost uniform band in 450-600 nm range both in the experiment and modeling.
Other than the expected difference in absolute values of RER, modeling produces a different
spectral behavior in 700-900 nm range which is also reflected in the distance dependence. A
close examination of all modeling results for both samples indicates that this is common
feature most likely brought by the boundary condition used in modeling.
Translating the target in-line between the source and fluence detector and performing the
perturbation analysis [27, 28], produces a curve that is completely symmetric relative the
midpoint and shows the minimal photon density midway between the source and detector.
Extending target translation to the entire X,Y-plane in the perturbation approach for fluence
produces the so called banana pattern. The banana pattern is also symmetric relative the
midpoint so that it has the identical widths measured at the same distance away from the
source and detector [28]. It is brought by the fact that the expression for photon density
distribution function for fluence is completely symmetric [27] and replacing the source with
the detector results in the same pattern. Inspection of Eq. (9) that can be considered analogous
to the photon density distribution function but for radiance indicates the absence of such
symmetry.
Qualitatively the difference between fluence and radiance in the perturbation analysis can
be explained as follows. When the inclusion is closer to the fluence detector, almost all
photons that pass through the region occupied by the inclusion are detected by the detector
but this number is relatively low compare with the total number of photons emitted by the
light source. For the inclusion closer to the light source, it acts like a filter such that almost all
photons that originate from the source also pass though the inclusion’s location. However,
due to properties of the turbid medium only a small fraction of these photons reach the
detector. Hence, the fractions in both cases are the same. Introducing angular sensitivity for
the detector (as in the case of radiance) changes the picture described for fluence. Multiple
scattering reduces the number of photons reaching the finite angular detection window from
the specific direction for radiance (as opposed to unlimited 4π for fluence) after travelling
long distances when the inclusion is closer to the light source. Thus, it produces lower values
for RER as oppose to the case when the inclusion is closer to the detector. The same multiple
scattering also continuously broadens photon angular distribution with distance which results
in the loss of directionality. Hence, angular profiles measured closer to the light source will
exhibit such broadening effects as seen in the experiment.
The observed discrepancies between the experiment and simulations can be attributed to
several factors. Even though radiance detection was performed in-plane in our experiments,
the source of the perturbation was a non-uniformly illuminated cylinder containing Au NPs.
Thus, it is expected that absolute RER values obtained for the cylinder will deviate from those
for the thin disk as in the 2D model. To account for the actual shape of the inclusion the
model needs to be extended to a 3D geometry. Another factor is related to applicability of
diffusion approximation for describing radiance in the angular domain [40, 58, 59]. It has
been shown theoretically that the deviations for radiance, obtained under diffusion
approximation from more accurate Monte-Carlo modeling, can be observed for distances
from the source equal up to five reciprocal reduced scattering coefficients [59]. In fact, the
deviation between radiance under diffusion approximation and more accurate analytical
solution of RTE was shown to exist for Intralipid-1% up to 10 mm away from the source [38].
Since diffusion approximation was used twice in calculating RER, it is responsible for
reported differences for contour lines in vicinity of the light source between simulated and
experimental spatio-angular plots. Validity of diffusion approximation can be also

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 3004
undermined by very high absorption values of the localized inclusion which may exceed
locally the scattering coefficient of Intralipid.
It is remarkable that the simple model used in the current work was able to confirm most
essential features observed in the experiments. It not only validates the experimental approach
but provides important details about the relative angular photon distribution function which
plays a critical role in interpreting radiance data for localized inclusions. In order to obtain a
complete quantitative agreement between theory and experiment, perturbation approach
should be incorporated into more sophisticated and accurate treatments of radiance [40, 62–
64]. By improving accuracy, such models would provide the basis for extracting the
concentration of nanoparticles in the localized inclusion and the distance from the inclusion to
the detector from experimental data. The universality of the model also allows extending it to
analysis of the inclusion positioned at any distance and angle relative to the detector-source
direction thus generalizing model predictions and parameter extractions.
Radiance is an interstitial spectroscopic technique with some spatial resolution
capabilities. Once fully developed, it can identify an inclusion spectroscopically, localize it in
the angular domain by specifying the angular and radial coordinates and quantify it by
providing the absorption coefficient or concentration. It all can be achieved with a single
detector-source pair by rotating the detector over 360° range. A single fluence-based detector-
source pair lacks spatial sensitivity and can’t detect an inclusion. As a result, configurations
with arrays of sources and fluence detectors on the boundaries of regions of interest are
employed for this purpose. Such technique, diffuse optical tomography (DOT) is a truly
imaging technique that uses sophisticated back-projection algorithms to reconstruct the
image. To constrain the inverse problem and improve the resolution of the image, DOT is
usually paired with ultrasound or MRI techniques. In addition, highly non-linear
reconstruction algorithms are required when there is a large discrepancy between the
refractive index of the inclusion and the surrounding medium (as in the case of Au NPs in
biological tissues). Thus, all techniques have their merits and limitations. Radiance-based
approach can’t substitute DOT in applications where true imaging is needed. We consider
radiance as a complementary technique that can be used interstitially for diagnostic purposes
(for example, in prostate) when its capabilities (identify, localize and quantify) and simple
inexpensive hardware satisfy medical needs.
4. Conclusion
We present new experimental results on detecting signatures of Au NPs based localized
inclusions in Intralipid-1% phantom when the inclusion was translated along the line
connecting the detector and light source. The inclusion acts as a perturbation and allows
tagging photons that passed through the inclusion’s location recreating their distribution in the
turbid medium. It permits drawing conclusions about detectability of the inclusion anywhere
within 30-mm source-detector separation.
Using diffusion approximation and extending the perturbation approach to radiance, we
develop the novel analytical expression for the radiance extinction ratio (RER) that has a
meaning of the relative photon angular distribution function and is equivalent to the photon
path density distribution function for fluence. The derived expression for radiance is capable
of capturing qualitatively most essential features and trends observed in the experiments:
strongly asymmetric distribution of photons in the angular domain between the source and
detector with highest density of photons near the detector and the minimum midway between
the source and the detector, better angular confinement of photons in vicinity of the detector
and continuous deterioration of the angular resolution once the inclusion is moved away from
the detector. This is distinctly different from the photon distribution seen by the fluence
detector indicating that the anisotropic detector paired with a single light source is capable of
providing spatial sensitivity as opposed to the isotropic detector. Angular detectability is
always better for the inclusion located closer to the anisotropic detector but not the light
source.

#199147 - $15.00 USD Received 11 Oct 2013; revised 19 Nov 2013; accepted 21 Nov 2013; published 25 Nov 2013
(C) 2013 OSA 1 December 2013 | Vol. 4, No. 12 | DOI:10.1364/BOE.4.002989 | BIOMEDICAL OPTICS EXPRESS 3005
We show that differential optical properties of the inclusion and the background medium
are manifested in measured spectra leading to very distinct transitions at RER = 1 when
optical attenuations match.
The observed discrepancies between the model and experiment call for more accurate
modeling approaches that can further aid in extracting such important parameters as the
concentration of nanoparticles in the localized inclusion and the distance between the
inclusion and detector.
Acknowledgments
The authors acknowledge financial support from Natural Sciences and Engineering Research
Council, the Canadian Institutes of Health Research, Atlantic Canada Opportunity Agency,
and Undergraduate Student Research Award from Natural Sciences and Engineering
Research Council (to K.B.C). The work was also partially funded by the Canadian Research
Chair Program to W.M.W.