You are on page 1of 27

61

Two-photon Microscopy and


Imaging

Patrick Theer, Bernd Kuhn, Dorine Keusters, and Winfried Denk


Max Planck Institute for Medical Research, Heidelberg, Germany

1 Introduction 63
1.1 Optical Sectioning 64
1.2 Increased Penetration Depth 66
1.3 Reduced Photodamage and Photobleaching 67

2 The Multiphoton Microscope 67


2.1 Light Sources 68
2.2 Excitation Pathway 69
2.3 Final Focusing and Resolution 69
2.4 Detection 70

3 Limitations 71
3.1 Temporal Resolution 71
3.2 Spatial Resolution 72
3.3 Penetration Depth 73
3.4 Photodamage and Photobleaching 74

4 Fluorophores 74
4.1 Comparison of 1PA and 2PA 74
4.2 Intrinsic Chromophores 75
4.3 Synthetic Dyes 76
4.3.1 Ion Indicators 76
4.3.2 Voltage-sensitive Dyes 76
4.4 Genetically Encoded Fluorophores and Indicators 77

5 Applications 77
5.1 Neurobiology 77
5.2 Calcium Imaging 79
5.3 Imaging of Metabolic Activity 80

Encyclopedia of Molecular Cell Biology and Molecular Medicine, 2nd Edition. Volume 15
Edited by Robert A. Meyers.
Copyright  2005 Wiley-VCH Verlag GmbH & Co. KGaA, Weinheim
ISBN: 3-527-30652-8
62 Two-photon Microscopy and Imaging

5.4 Photoactivation (Uncaging) 80


5.5 Human Diseases and Therapy 81

6 Future Directions 82

Bibliography 83
Books and Reviews 83
Primary Literature 83

Keywords
Confocal Microscopy (CM)
Microscopy technique where optical sectioning is achieved by using a detector pinhole
that is confocal to the excitation focus to reject the out-of-focus signal.

Fluorescence
Light-emitting process of special molecules, called fluorophores, after excitation with
photons of higher energy.

Laser-scanning Microscopy
Microscopy technique, in which the light source is focused to a diffraction-limited spot
and the image is formed by scanning this spot across the sample while sequentially
recording the signal generated at each position.

Multiphoton Absorption (MPA)


Process where two or more photons are absorbed simultaneously to excite a transition
with an energy equal to the sum of the energy of the photons.

Multiphoton Microscopy (MPM)


Microscopy technique where the contrast mechanism is based on a multiphoton
absorption process. Usually, but not always, the detected signal is fluorescence.

Two-photon Microscopy (2PM)


Most widely used variant of multiphoton microscopy.

Whole-field Detection
Detection scheme for laser-scanning microscopy where the light from the whole field
of view is detected.

Wide-field Imaging
Imaging technique where the whole field of view is illuminated simultaneously and
imaged onto a spatially resolving detector, for example, a charge-coupled device
(CCD) camera.
Two-photon Microscopy and Imaging 63

 Multiphoton microscopy (MPM) is an imaging technique that employs signals,


such as fluorescence, generated by a multiphoton absorption (MPA) process as its
contrast mechanism. Compared to other fluorescence microscopy techniques, MPM
often has a higher sensitivity, superior penetration into scattering tissue, and causes
less photodamage to the sample and less photobleaching of the fluorophore. These
advantages make MPM the method of choice for optophysiological experiments and
also make MPM the only technique available for high-resolution in vivo imaging
deep within scattering tissue, such as the brain. The two-photon microscope (2PM)
is the kind of MPM used in the overwhelming majority of applications since it
has virtually all of the specific advantages of MPM. Recent advances in staining
techniques including the booming development of genetically encoded fluorophores
keep increasing the field of actual and possible MPM applications.

1 a signal – such as fluorescence – that is


Introduction generated by absorption of light from
the beam.
Over the past 15 years, multiphoton mi- Molecules that can easily absorb light
croscopy (MPM) has developed into an generally have a structure where sev-
indispensable imaging tool in the life eral atoms of the molecule are connected
sciences. MPM has higher sensitivity, su- by conjugated single–double bonds. As
perior penetration into scattering tissue, a consequence, the most weakly bound
and often causes less photodamage in sam- electrons are delocalized and thus inter-
ples than alternative imaging techniques act easily with the electromagnetic field.
such as confocal microscopy (CM). This Near-ultraviolet or visible light is usually
means that MPM – combined, in partic- sufficient to promote the outermost elec-
ular, with one of many new genetically tron to an excited level (Fig. 1). In general,
encoded and functional dyes – has become the molecule is excited electronically and
the method of choice for in vivo imaging vibrationally, with the latter excitation de-
in strongly scattering tissues such as skin, caying very fast (∼10−12 s). From the state
muscle, and tumors, and even deep within reached after vibrational relaxation, which
intact organs such as kidney, pancreas, is still electronically excited, the molecule
lymph nodes, and brain. then relaxes within nanoseconds (10−9 s)
This article provides an overview of func- to a state that is only vibrationally excited.
tional principles, instrumentation, and The energy difference can be carried away
some of the applications of MPM. Fur- by fluorescence emission of a photon.
ther details can be found in the books and The difference between one photon
reviews and primary literature listed at the (1P)-excited and multiphoton (MP)-excited
end of this article. fluorescence microscopy is that molecular
In laser-scanning microscopy, a three- absorption is of a single photon in the first
dimensional image is generated by scan- case and a photon multiplet in the second
ning the position of a tight laser focus (Fig. 1). In a 1P process, the photon energy
around in the sample while recording, equals the energy difference between the
as a function of the focus position, ground state and an excited state. Light
64 Two-photon Microscopy and Imaging

Fig. 1Principle of one-photon and


Excited two-photon excited fluorescence. In 1PE
state
fluorescence, a molecule is brought to
the excited state by absorbing a single
One-photon Two-photon (blue) photon. Vibrational relaxation
excitation excitation brings the molecule to the lowest
excited state, from which it returns to
the ground state by emitting a
Virtual
longer-wavelength (green) photon. In
state
2PE fluorescence, the molecule is
brought to the excited state by
simultaneously absorbing two NIR (red)
photons. The molecule again
vibrationally relaxes, and emits a
fluorescence photon from the same
Fluorescence Ground level as in 1PE fluorescence (see color
state
plate p. xxv).

with a quantum energy less than the gap 1.1


between the ground state and the lowest Optical Sectioning
lying electronically excited state cannot ex-
cite the molecule by a 1P process, but if the In 1P microscopy, the signal generated
is proportional to the light intensity (the
light intensity is sufficiently high, multi-
power divided by the beam area: I = P/A).
ple photons of lower energy can cooperate
As a result, the signal generated in each
and combine their energies to promote the
slice of a sample depends on the slice
molecule into the excited state. For exam-
thickness, but not the distance from the
ple, two 800 nm photons can excite a tran-
focus, since variations in the beam area
sition that normally requires 400 nm light.
are, as long as the power stays constant,
One may picture this process as the first
exactly balanced by inverse variations in
photon exciting the molecule to a virtual
the excitation rate per molecule. When
state from where the second photon then
imaging thick samples, the light from
brings the molecule into its (real) excited the focus is thus always contaminated, in
state. For this process to occur, the two fact, dominated by fluorescence generated
photons have to arrive at the fluorophore above and below the focus. This is the
within the lifetime of the virtual state, bane of wide-field fluorescence imaging
which is less than 1 fs (1 fs = 10−15 s) of 3-dimensional (3D) samples. In CM,
as defined by the Heisenberg uncertainty this problem is solved by placing a
principle (τ ≤ h̄/E). The likelihood that pinhole in front of the detector. The
two photons arrive at a fluorophore almost pinhole is confocal with the excitation
simultaneously is proportional to the in- focus and rejects light that did not
tensity squared and is all but negligible for originate from that focus (Fig. 2). This
light sources other than a focused laser. property of the confocal microscope, called
Therefore, two-photon absorption (2PA) optical sectioning, vastly improves the
was observed only after the invention of the ability to obtain 3-D fluorescence images,
laser. It is the strongly superlinear depen- particularly of thick biological samples.
dence on the light intensity that gives 2PM The first nonlinear optical microscope
(and MPM in general) its main advantages. used the second-harmonic signal and
Two-photon Microscopy and Imaging 65

Focal
plane

(a) (b)

PMT PMT
Pinhole

Lens Lens

Dichroic Dichroic
mirror mirror
Objective Objective
Sample Sample

Focal
plane

(c) (d)
Fig. 2 Optical sectioning in one-photon confocal and in two-photon
microscopy. (a) With 1PE, fluorescence is generated throughout the
illumination cone. For 2PE (b), because of the nonlinear intensity
dependence, fluorescence generation is limited to a small volume
around the focus. In CM, optical sectioning is achieved by placing a
confocal pinhole in front of the detector. This pinhole rejects
(out-of-focus) fluorescence that is generated above and below the focus
(c), fluorescence generated by scattered excitation light (e), and in-focus
fluorescence that is scattered on its way to the detector (e). Because no
out-of-focus fluorescence is generated in a two-photon microscope, no
pinhole is necessary (d) and even fluorescence that is scattered on its
way to the surface can be detected (f ).
66 Two-photon Microscopy and Imaging

PMT PMT
Pinhole

Lens Lens

Dichroic Dichroic
mirror mirror
Objective Objective
Sample Sample

Focal
plane

(e) (f)
Fig. 2 (Continued)

wide-field imaging. Scanning the focus, Common to scanned-focus nonlin-


which was introduced later, strongly en- ear microscopies is that they provide
hances the focal but not the out-of-focus excitation-based optical sectioning. This
signal, and thus provides optical sec- is because, different from the linear (1P)
tioning. The hopes placed on harmonic case, the total excitation in a slice in-
imaging for biological specimens were, tersecting the beam strongly drops with
however, not fulfilled, at least initially. distance to the focus (Fig. 2). In fluores-
Two photon–excited fluorescence mi- cence microscopy, excitation-based rather
croscopy proved to be much more imme- than confocal optical sectioning is par-
diately applicable to biological problems, ticularly helpful since photodamage and
due, first, to the central importance of flu- photobleaching are enormously reduced
orescent labeling in biological microscopy (see below).
(see Sect. 4) and second, to the differences
in the physical characters of fluorescence 1.2
and harmonic generation. The main dif- Increased Penetration Depth
ference being that harmonic generation is
a coherent process, where field strengths MPM can generate high-resolution im-
rather than intensities add, and the gener- ages from focal planes deep inside the
ated signal depends on the square of the tissue, much deeper than CM, because
chromophore concentration, and is liable signal is generated only at the focus allow-
to interferences effects; in fluorescence, ing scattered fluorescence to be collected
which is incoherent, light from different without loss of resolution or optical sec-
fluorophores does not interfere with each tioning. In CM, only light passing through
other and the signal increases linearly with the confocal pinhole is detected. However,
the fluorophore concentration. due to scattering, the fraction of the signal
Two-photon Microscopy and Imaging 67

originating from the focus that does pass light-induced damage to the sample
the pinhole in the CM decreases exponen- (photodamage) or the fluorophore
tially with imaging depth. Compensating (photobleaching), which in 1PM occur
for the loss in signal by increasing the ex- throughout the sample even though
citation power in CM in most cases would information is only gained from the focal
lead to an unacceptable increase in photo- slice. In MPM, damage is confined to
damage (see below). A further reason for those fluorophores (and their immediate
the deeper penetration of MPM is that, vicinity) that do provide information.
for the same fluorescent label it uses ex- A reduced relative MP absorbance
citation light with a wavelength that is of intrinsic fluorophores may further
longer (up to two times for 2PM) and contribute to the vastly improved tissue
hence scattered much less. Occasionally viability sometimes seen even in tissue
helpful is the larger separation between that scatters only little.
the excitation and emission wavelengths
in MP excited fluorescence, which allows a
2
more spectrally complete detection of the
The Multiphoton Microscope
fluorescence.
1.3 The layout (Fig. 3) of a generic MPM is
Reduced Photodamage and very similar to that of a laser-scanning
Photobleaching CM, from which an MPM can often
be constructed by modification. In this
Protracted and high-resolution CM section, the components of an MPM will
imaging is often limited by excitation be discussed individually.

NIR Detector
fs-laser
Barrier filter

Collection
lens
Intensity
control

x -y scan Dichroic
mirrors mirror
Scan Tube
lens lens
Objective

Sample on x -y -z stage

Fig. 3 Setup of a generic two-photon microscope. Light from an ultrafast laser or laser
amplifier is focused to a diffraction-limited spot by an objective. The beam is
raster-scanned across the sample and the fluorescence generated at each position is
collected by the objective, separated from the excitation light by a dichroic mirror and
additional filters either in a transmission or reflectance configuration (see inset), and
detected by a whole-field detector.
68 Two-photon Microscopy and Imaging

2.1 where   indicates an average over time, δ


Light Sources is the two-photon absorption cross section
(often expressed in units of Göppert-Mayer
The development of reliable ultrafast laser (GM); 1 GM = 10−50 cm4 s/photon), and
sources operating in the near infrared C is the fluorophore concentration. For a
(NIR) coincided with the invention of spot with area A, power and intensity are
MPM and has been an important factor related by:
in the broad adoption of MPM, since with
continuous wave (cw) lasers, even 2PM P(r)
I(r, t) = . (2)
is only marginally possible for brightly A
labeled samples. Today the vast majority of For a square pulse with pulse length τ ,
MPMs use mode-locked titanium sapphire the ‘‘two-photon advantage’’ is
lasers (Ti:sapphire, Ti:Al2 O3 ), which can

be tuned from about 680 to 1100 nm, 1 T
I(r, t)2 dt
allowing 2PE of almost all relevant labels, I(r, t)2  T 0 1
=  2 = .
and 3PE into the deeper UV. The pulse I(r, t)2  T f τ
1
length of this laser type is around ∼100 fs I(r, t)dt
with a repetition rate of around 100 MHz, T 0
i.e. every 10 ns a short burst of light arrives (3)
at the sample while for the remainder Equation 1 then becomes:
of this period no light is incident upon
P(r)2 1
the sample. Alternative laser sources that nF (r) ∝ δC . (4)
A2 f τ
produce similar pulse lengths are based
on Cr:LiSrAlFl, Cr:Forsterite, or Nd:YLiF Thus, for example, to obtain the same
crystals, or on Yb-doped optical fibers, all of signal with 1 ps pulses rather than with
which have much smaller tuning ranges, 100 fs pulses, the average power must be

but with the exception of the Cr:LiSrAlFl higher by a factor of 10.
laser, operate at longer wavelengths than Implicit in this derivation is the
can be easily provided by the Ti:Sapphire assumption that the 2P absorption cross
laser. A very large tuning range but section δ is constant across the spectrum of
less total power is provided by optical the laser pulse. For very short pulses this is
parametric amplifiers. not true because – due to the uncertainty
An ultrafast laser increases the sig- principle – the shorter the pulse duration
nal levels in the 2PM by 5 orders (τ ), the broader its spectral width
of magnitude as compared to a cw- (ν). For a transform-limited (unchirped)
laser of the same average power. This Gaussian-shaped pulse τ ν = 0.44. A
is because the average rate at which 100-fs laser pulse centered around 800 nm
fluorescence photons are generated by then ranges (full width at half maximum;
2PA is proportional to the average FWHM) over 10 nm in wavelength and a
square of the instantaneous light inten- 20-fs pulse over 50 nm, which is already
sity I: as wide as the main features in a
typical fluorophore absorption spectrum.
nF (r) ∝ δCI(r, t)2  However, in practice the main limitation
I(r, t)2  for using shorter pulses is group delay
= δCI(r, t)2 , (1) dispersion (GDD), which occurs because
I(r, t)2
Two-photon Microscopy and Imaging 69

different wavelength components of the 2.2


laser pulse travel at different speeds inside Excitation Pathway
all materials, including of course, optical
glass. After passing the microscope optics, The pathway for the excitation light in
the long-wavelength (‘‘red’’) components a MPM is essentially that of a laser-
of the pulse arrive at the focus ahead of the scanning CM: shutter, beam conditioning
short-wavelength (‘‘blue’’) components, and intensity control, scan mirrors, scan
leading to a so-called chirped pulse. lens, tube lens, objective, and, finally, the
Because the wider spectrum of a shorter sample (see Fig. 3). As in every laser-
pulse accentuates this effect, an initially scanning microscope, the scan mirrors (or
shorter pulse can actually become longer a point in between for proximity coupled
once it reaches the focus than an initially scanners) is conjugate to (i.e. imaged into)
longer pulse. Owing to the GDD in the objective exit pupil. Coatings on lenses
a typical microscope, a 20-fs pulse, for and mirrors need, of course, to be IR
example, is stretched to about 700 fs, while transmissive or reflective, respectively. The
a 100-fs pulse is stretched to only 200 fs. intensity control needs to be compatible
Without compensation (prechirping, see with ultrashort pulses and should allow
below), the shortest pulses that can the automatic compensation of depth
be attained at the sample are about dependent losses (see Sect. 3.3) and should
150 fs. Therefore, lasers producing 100 to be fast enough to shut off the beam during
200 fs pulses are typically used in 2PM. retrace, i.e. while the focus is moved from
the end of a scan line to the beginning of
It is possible (but rarely necessary) to
the next.
compensate for the microscope’s GDD
by the use of an optical arrangement
2.3
(consisting of prisms, gratings, or special Final Focusing and Resolution
dielectric coatings) that has negative
GDD. Negative GDD prechirps the pulse, Many standard microscope objectives can
giving the blue components a head be used for MPM, albeit with compromises
start over the red components. This in correction and transmissivity. While
head start compensates all or part of this was an issue in the early times of
the speed disadvantage that the blue MPM, most objectives now are either cor-
components have when traveling through rected and transparent throughout the IR
the microscope optics. Prechirping allows or come as special ‘‘IR’’ versions. To mini-
the arrival of transform-limited pulses as mize tissue damage (see below), detection
short as 15 fs at the focus. For such short efficiency has to be maximized by keep-
pulses, compensation of higher-order ing the transmission at the fluorescence
dispersion and chromatic aberrations wavelength(s) as high as possible. Optical
needs to be considered as well. For higher correction at the fluorescence wavelength
than 2PA, the effects of using shorter is not necessary. The numerical aper-
pulses are even more dramatic (the 3PA ture (NA) of the objective determines
rate, for example, increases by 10 orders of the resolution (laterally ρ ∝ 1/NA, axi-
magnitude compared with cw-excitation ally z ∝ 1/(NA)2 ) and the peak intensity
when using the typical Ti:sapphire (∝ (NA)2 ), and thus the peak excitation
laser). rate per fluorophore (∝ (NA)4 ), but (and
70 Two-photon Microscopy and Imaging

this is a special property of 2PA, that will usually be water or a physiological


neither 1P nor higher-order multiphoton salt solution. Refractive-index mismatch
absorption, MPA possess) when the con- between immersion medium and sample
centration of fluorophores is uniform, the that was not taken into account during the
total amount of fluorescence generated lens design results in a loss of resolution
throughout the focal volume V = ρ 2 z and signal strength.
is independent of the NA:
4
2.4
2 NA 2 21 Detection
nF tot ∝ δCP ρ z ∝ δCP .
fτ fτ
(5) To preserve the main strength of MPM, i.e.
A more extensive treatment that takes the reduction of photodamage, the detec-
into account the exact spatial and temporal tion efficiency must not be compromised.
intensity distribution of the light in the This is because molecular excitation in-
sample yields: evitably causes photodamage to the sample
4ηgnλ 1 (see Sect. 3.4) and thus the number of
nF tot = δCP2 , (6) molecular excitations necessary to gener-
πh2 c2 fτ
ate a sufficient number of detected photons
where η is the quantum-efficiency of need to be minimized. In nonscattering
the fluorophore (i.e. the percentage of samples, the main determining factor for
excited molecules that emit their excess detection efficiency is the solid angle of
energy in the form of a photon), g = collection (assuming, of course, the avoid-
f τ I(r, t)2 /I(r, t)2 and depends on the ance of absorption and reflection losses as
exact temporal intensity distribution of much as possible). Thus, a high NA objec-
the pulse. For the square pulse used tive is crucial if epi detection (through
in equation (3), g = 1. For more realistic the objective lens, the usual configura-
shapes such as a Gaussian or a hyperbolic tion) is used. Better still is a combination
secant intensity distribution, g = 0.66 and of epi detection (where, for working dis-
0.59, respectively; n, λ, h, and c are the tance reasons, a compromise with respect
index of refraction, wavelength of the to the NA may be necessary) with trans-
excitation light, Planck’s constant, and the detection through a high NA condenser.
vacuum speed of light, respectively. The separation of the fluorescence from
While scattering per se need not reduce the excitation light by a dichroic mir-
the resolution even if only a minor part ror is standard practice in fluorescence
of the beam energy reaches the focus microscopy, as is the further separation
unscattered (ballistically), eventually the into multiple spectral channels. Suppres-
stronger attenuation of the lateral rays sion of the excitation light often is easier
due to their longer path length inside in MPM because the wavelength sep-
the tissue does lead to a reduction in aration is larger and the sensitivity of
the effective NA, and thus a larger focus some detectors declines sharply toward
size. Further, practical, considerations in longer wavelengths. In contrast to a laser-
selecting an objective are the detection scanning CM, there is no need in the
efficiency (see below), working distance, MPM to ‘‘de-scan’’ the fluorescence in
which tends to decrease with increasing order to thread it through the confocal
NA, and the immersion medium, which pinhole. In fact, the detection optics in
Two-photon Microscopy and Imaging 71

a MPM can be rather crude but need to 3.1


accept large opening angles and have a Temporal Resolution
large field of view for deep imaging in
scattering samples. The temporal resolution of optical tech-
Deep imaging in scattering samples niques can be very high. In fact, some
is where no other known fluorescence of the fastest measurements are based on
microscopy technique can compete with pulsed lasers. When fluorescence is used,
MPM because MPM can image deep in the excited-state lifetime (in the nanosec-
scattering samples virtually without loss of ond range) is the ultimate limit. In practice,
resolution and sensitivity as long as the temporal resolution is often limited by the
detection pathway collects a large fraction time required to obtain a sufficient signal-
of the diffuse fluorescence light emerg- to-noise ratio. When a whole image is to
be acquired, this time has to be multi-
ing from the sample. Unlike in other
plied with the number of pixels in a frame.
techniques, the diffuse fluorescence con-
Another limitation of all laser-scanning
tains high-resolution information about
microscopes is that the focus has to be
the focus because the excitation is highly
raster-scanned across the field of view. In
localized to a volume of less than one
most CMs and MPMs, this is achieved
femtoliter. The improvement possible by
by mechanically scanning the beam using
using whole-field detection has been di-
galvanometer mounted mirrors, which are
rectly demonstrated by switching quickly
limited in frequency response to the low
between whole-field and descanned detec-
kilohertz regime, setting the rate at which
tion. The main drawback of whole-field scan lines can be acquired to about 1 kHz
detection is its incompatibility with de- or so, depending somewhat on the ex-
tectors that only accept a small phase tent of the scan. A high-resolution (1000
space volume (the product of angular lines) image thus takes at least one sec-
and area acceptances). This precludes ond. By sacrificing one spatial dimension
the use of photon-counting avalanche and scanning a single line only, a time
photodiodes (APDs), which are the high- resolution of 1 ms is common, and when
est quantum-efficiency photon detectors measuring only from one spot, a time
available, and of spectrometer detectors. resolution in the microsecond range can
Fortunately, the characteristics of photo easily be achieved. On the other hand,
multiplier tubes (PMTs), which have large to acquire a high-resolution volume im-
sensitive areas and acceptance angles age (a stack of optical sections) can take
and are the standard detectors for CM, many minutes. When trying to measure
have also improved, closing the quantum- dynamic physiological signals from many
efficiency gap. locations simultaneously, as it is desirable,
for example, in neuroscience, image acqui-
sition time is a serious limitation. Attempts
3 have, therefore, been made to speed up the
Limitations scanning process by either using resonant
scanners, or rotating polygonal mirrors,
In this section, the limitation of spatial and or by scanning the beam nonmechanically
temporal resolution, penetration depth using an acousto-optical deflector (AOD).
and photodamage will be discussed. Owing to the spread wavelength spectrum
72 Two-photon Microscopy and Imaging

of the excitation pulses, the AOD, which point-spread function) is completely de-
is based on diffraction by a sound wave, is termined by the intensity distribution (in
plagued by angular dispersion, which can fact, the point-spread function equals the
only be sufficiently compensated for a lim- intensity squared) at the focus. Thus, the
ited field of view. The main attraction of resolution depends on the wavelength of
AOD scanning is the capability of random- the excitation light and on the NA of the
access scanning: the focus can be moved objective. With 700-nm light and an NA of
between distant location within microsec- 1.3, a resolution (FWHM) of ∼200 nm in
onds without illuminating (and potentially the lateral and 600 nm in the axial direction
damaging) the sample area in between. can be achieved.
Acquisition speed is limited ultimately, Even though in CM the resolution is
as mentioned above, by the fluorescence determined by the combined (multiplied)
decay time, which has to be shorter than excitation and detection point-spread func-
the pixel dwell time. But the limited tions, the mathematical expression for the
rate at which a fluorophore can produce resolution of CM and 2PM are identi-
photons, even if the excitation light is cal. This holds, however, only under the
arbitrarily intense, usually enforces a somewhat unrealistic assumption that the
longer pixel dwell time just to collect pinhole has zero diameter and the 1PE
enough photons. Beyond those limits, and 2PE wavelengths and the emission
an increase is only possible by scanning wavelength are all the same. Realistically,
multiple points, arranged either in a 2D the pinhole has to be opened up to gain
pattern or in a line. This approach is sufficient signal, the emission wavelength
rather successful in CM, but it requires is longer (both reducing the resolution of
confocal or at least spatially resolved CM) and the excitation wavelength is close
detection, since at each point in time, to twice as long for 2P as for1P (reducing
fluorescence can come from any of the the resolution in 2PM).
foci. Thus, using it in MPM means As a result, compared with CM, the
giving up one of the main advantages resolution in 2PM (and MPM, in general)
of MPM, namely the ability to detect is typically somewhat worse, but can be
fluorescence irrespective of how it reaches improved by reintroducing a confocal
the detector (see above) and accepting pinhole, albeit at the expense (particularly
either image degradation or signal loss. in scattering specimens) of detection
Other drawbacks are anisotropic lateral efficiency. 2PM can also be combined
and reduced axial resolution (for line with the 4Pi microscope, which greatly
illumination) and the increased laser- improves the resolution in the axial
power requirement (for any multipoint direction, and where using 2PE rather than
method). 1PE considerably reduces side lobes in the
point-spread function and thus fills in the
3.2 holes in the modulation transfer function.
Spatial Resolution In general, at a given wavelength, res-
olution improvements beyond the Abbe
In the generic configuration (with whole- limit require optical nonlinearity, which
field detection), there is no spatial se- is exploited in the Stimulated Emission
lectivity in the detection and hence the Depletion (STED) microscope. This mi-
resolution of MPM (i.e. the size of the croscope is based on the saturation of
Two-photon Microscopy and Imaging 73

excited-state depletion, which has poten- To increase the imaging depth further,
tially unlimited resolution, because this the efficiency in generating 2P fluores-
saturation contains arbitrarily high or- cence needs to be increased, which is possi-
ders of nonlinearity. 2PE, by contrast, ble by reducing the laser pulse–repetition
contains only a second order (quadratic) rate while increasing the pulse energy.
nonlinearity, and consequently, in 2PM The use of a so-called regenerative amplifier,
the resolution improvement beyond the which generates pulses that are about 250
Abbe limit is moderate. times as intense as those directly from the
laser ‘‘oscillator’’ but are also by the same
3.3 factor less frequent, resulting in an un-
Penetration Depth changed average power, permits imaging
to about 1000 µm. A somewhat techni-
In scattering samples where absorption cal but important issue is that each pixel
of the light is negligible, the intensity of needs to contain at least one pulse, limiting
the ballistic light, i.e. light that has not the pixel rate to several hundred kilohertz
been scattered, decreases with depth in (the repetition rate of the amplifier). At
the sample as: such pulse rates, the average 2PA rate is
  increased by a factor of about 250.
d
Iball = I0 exp − , (7) Now the imaging depth is, however,
ls
no longer limited by the available laser
where ls is the scattering length, i.e. power but instead by the fluorescence
the average distance a photon travels background generated near the surface of
before it is scattered. For near-infrared the sample. The reason for this is that
wavelengths, ls in brain tissue is around in order to keep the fluorescence gener-
200 µm. To estimate the imaging depth, ation at the focus constant, the amount
one can assume that only the ballistic of ballistic light that reaches the focus
light that reaches the focus contributes to must also be kept the same, and thus the
the generation of two-photon fluorescence, laser power that needs to enter the sample
which seems to hold for up to at least must increase exponentially (equation 7).
several scattering mean-free-path lengths. At large depths, this exponential increase
Then the number of fluorescence photons overcomes the quadratic decrease in light
generated when the focus is at a depth d intensity, which results from the increased
below the surface can be written as: beam cross section with distance from the
1 g 8nλ (−2d/ls ) focus. Eventually, the intensity at the sur-
nF tot = ηδCP2 e . (8) face becomes comparable to the intensity
2 f τ πh2 c2
at the focus and surface fluorescence starts
As long as the detection system is de- to dominate. For samples with staining
signed to capture most of the fluorescence, throughout the sample, this limit may well
which consists almost entirely of scattered be impossible to overcome, thus defining
light, the maximum imaging depth de- the ultimate depth limit of 2PM (a similar
pends only on the laser power that is logic applies to higher-order MPM). This
available. With the power available from limit increases somewhat with the NA, but
standard mode-locked Ti:sapphire lasers, that may not help much since at higher
imaging depths of 2 to 3 scattering lengths NAs, focus blurring due to wavefront aber-
(400–600 µm) are achieved routinely. rations becomes a more serious issue, and
74 Two-photon Microscopy and Imaging

may, in turn, require the use of adaptive nonlinearities such as locally overwhelm-
optics as a remedy. ing cellular repair mechanisms should
be avoidable by faster scanning. Often,
3.4 as mostly anecdotal evidence suggests,
Photodamage and Photobleaching damage decreases with increasing wave-
length, which may be due to reduced
Cell viability is obviously important for 2PA by endogenous chromophores. While
imaging living tissue. Excitation of in- photobleaching can be quantitatively mea-
trinsic or introduced chromophores often sured, this is more difficult with damage
leads to photochemical effects, such as to cells, since photostress may be present
destruction of the chromophore (bleach- but may not be evident below a certain
ing) or damage and subsequent cell-death, instantaneous or cumulative threshold.
which are mostly mediated by reactive This is consistent with the observation
oxygen species. This problem is exacer- in several studies that at low intensities
bated in laser-scanning CM because of the (up to 1014 − 1015 W m−2 which corre-
rather poor utilization of the generated sponds to 1–5 mW at the focus) no
fluorescence light and the appearance of obvious damage occurs. As the intensity
superlinear photochemical effects at high increases, severe damage, optical break-
intensities, probably due to excited-state down, and strong luminescence eventually
absorption. Utilization of the generated do occur.
fluorescence is generally much better in
MPM, with the improvement increasing
as the specimen becomes more scattering 4
and the imaging depth increases. Fluorophores
However, in clear (i.e. nonscattering)
specimens, where the detection efficiency 4.1
in the CM for focal photons can be rather Comparison of 1PA and 2PA
high, bleaching and photodamage in the
2PM can actually be exacerbated, but is While the fluorescence process itself is
also clearly reduced in some cases com- largely independent of the mode of excita-
pared with CM. Strategies for avoiding or tion (and hence the fluorescence emission
reducing damage depend on the under- spectra are the same for 1PE and 2PE),
lying photophysics and photochemistry. the absorption spectra of 1PE and 2PE are
If damage is due to single-photon ab- expected and found to be quite different in
sorption, increasing the 2PE efficiency by some cases. The reasons for this are: (1) To
reducing pulse length or repetition rate excite a molecule from its ground state to
is helpful. Damage is, however, rarely the excited state, quantum-mechanical se-
dominated by 1PA for typical tissue, but lection rules have to be fulfilled in addition
can become so in pigmented cells or to energy conservation. For example, the
if much longer wavelengths are used angular momentum (usually only the pho-
and water absorption becomes an issue. ton spin is relevant since the absorption
If higher-order instantaneous nonlinear- of a photon with nonzero orbital angular
ities, such as 3P absorption or optical momentum is unlikely) that is carried by a
breakdown are the culprits, increasing photon must be taken up by the molecule
the pulse length can help. Chemical upon absorption. The selection rules for
Two-photon Microscopy and Imaging 75

2PA are different from those for 1PA, since attempts to predict 2P cross sections and
in the first case, two photons, each carrying to develop, on the basis of theoretical
a spin of 1 h̄, are absorbed. This means that calculations, chromophores with large 2P
the angular momentum of the molecule cross sections.
has to be unchanged (if two photons of In the following, we will give a short
opposite spin are absorbed) or changed by overview of chromophores that are widely
2 h̄. A change by 1 h̄ (also corresponding to used with 2PM, with particular focus
a change in parity) is not allowed. How- on dyes with biological relevance. Under
ever, for complex asymmetric molecules, this premise, the chromophores can be
these selection rules are not necessarily categorized into intrinsic, synthetic, and
as strict, due to interaction with molecu- genetically encoded dyes, each group
lar vibrations and rotations. (2) Since the containing dyes with a variety of different
transition to the excited state occurs in a indicator features.
sense via all molecular states, which serve
as a combined ‘‘virtual’’ intermediate state, 4.2
excitation is possible even if there is no di- Intrinsic Chromophores
rect orbital overlap between ground and
target state. Biological tissue is naturally fluorescent
For the use of fluorescent dyes in to a varying degree even without adding
MPM, it is very important to know chemically synthesized molecules or in-
their MP-absorption spectra in order to troducing the gene of a fluorescent protein
choose the best dye and then excite (see below). The molecular origin of this
it optimally. While it is not easy to intrinsic ‘‘autofluorescence’’ depends on
measure MP-absorption spectra, at least the excitation wavelength. For UV excita-
the 2P-absorption spectra of many com-
tion, nucleotides and the aromatic amino
monly used chromophores are now avail-
acids are dominant, for excitation with visi-
able (http://www.drbio.cornell.edu/) and
ble wavelengths, other compounds such as
it turns out that all dyes used with 1PE can
nicotinamide adenine dinucleotide (phos-
also be used with 2PE, albeit with a larger
phate) NAD(P)H and flavin adenine dinu-
spread in their cross sections because, in
cleotide FAD are the main contributors.
a way, transition matrix elements enter
twice in the 2PA process. Tryptophane fluorescence is rarely used
In some cases, the 2PA spectra plotted for 2PM, but 2PE fluorescence correlation
on a λ/2 scale are rather similar to spectroscopy (FCS) is possible for proteins
the 1PA spectra plotted on a λ scale. containing a large number of tryptophane
In most cases, however, the excitation residues. NAD(P)H is of special interest
maxima are shifted to the blue and the as its fluorescence intensity depends
spectra are broader. Additional spectral on its oxidation state. With excitation
features at wavelengths longer than the at 800 nm, NAD(P)H and flavoprotein
‘‘red’’ 1P absorption edge are not seen, fluorescence can be used together for a
nor would they be expected. The reason quantitative ratiometric measurement of
for this is that fluorescence emission only cellular metabolism.
occurs for molecules where the transition One of the advantages of MPA for
to the lowest lying excited state 1P is exciting cellular autofluorescence is that
allowed. There have been a number of it permits access to transitions that
76 Two-photon Microscopy and Imaging

require for 1PE, ultraviolet light of wave- moderate 2P cross section. Much larger
lengths that do not pass most micro- cross sections are found in xanthene-
scope objectives. derived indicators, which also use the
On one hand, endogenous fluorophores BAPTA Ca2+ -binding group and are now
can be useful because tissues can be im- available with a wide range of Ca2+
aged without dyes having to be applied affinities, covering the entire physiological
to the tissue. On the other hand, endoge- range. A selection of different peak emis-
nous fluorophores contribute most of the sion wavelengths (from 530 to 670 nm)
unwanted background when exogenous ensures discrimination ability against GFP
dyes, which provide enormous specificity or other cellular labels if needed.
and tailored indicator properties, are to be In tissue imaging, labeling without
imaged. A special case is the green fluo- damage to cells is crucial and several
rescent protein (GFP, see below), which methods have been used in connection
occurs naturally in the jellyfish and has with MPM. (1) Loading individual cells
a biological function there, but in other by diffusion or iontophoresis from dye-
organisms GFP is an exogenous stain in filled micropipettes in brain slices or
the sense that it needs to be introduced by in vivo; (2) applying cell-permeable ace-
molecular genetic means. toxymethoxyl (AM) esters variants, which
become trapped in cells by intracellular es-
4.3 terases. AM loading is straightforward in
Synthetic Dyes isolated cells, but is also possible in brain
slices and even in vivo; and (3) particle
Because of the specific advantage of (‘‘gene’’)-gun delivery.
MPM for the observation of living tissue, For an overview of fluorescent indicators
the most frequently used synthetic dyes see http://www.probes.com/. All of these
are those that provide functional signals, indicators can be efficiently 2P excited
called fluorescent indicators, in particular, somewhere between 700 and 1000 nm,
ion-sensitive, and (more recently) voltage- i.e. they are accessible when using a
sensitive probes. Ti:sapphire laser.

4.3.1 Ion Indicators 4.3.2 Voltage-sensitive Dyes


In many cellular responses, changes in ion Voltage-sensitive dyes (VSDs) measure
concentrations are involved. Most impor- electrical activity directly rather than in-
tant is Ca2+ , with its key position in cellular directly via Ca2+ influx. Most VSD-based
signal transduction; also important are measurements have, until recently, not
Na+ , K+ , Mg2+ , H+ , and Cl− . Ion in- used laser-scanning microscopy, since the
dicators go back to Arnaldus de Villanova, fractional fluorescence changes that oc-
a doctor and alchemist (∼1300 AD), who cur with physiological voltage swings are
invented litmus, an indicator that changes small and thus need for detection and
its color (absorption spectrum) depend- quantification of a large number of fluores-
ing on the concentration of H+ ions. The cent photons. Recently, it was shown that
first 2PM calcium measurements where by using band-edge excitation, fractional
done using Indo-1, a BAPTA-based Ca2+ changes are substantially increased, and
indicator, which is, however, excited by are for equivalent excitation wavelengths,
rather short wavelength and has a rather larger with 2PE than with 1PE.
Two-photon Microscopy and Imaging 77

4.4 only method that allows morphological


Genetically Encoded Fluorophores and and functional imaging in thick living
Indicators tissue and in live animals. While many
of the essential uses of MPM are in
Labeling with synthetic dyes is in most the field of neuroscience, the range of
cases not cell-type specific. For example, biological structures and processes in-
when injecting AM ester-Ca2+ indicator vestigated using MPM have continuously
into brain tissue, many cell types take up grown. Applications of MPM now include
the dye, and it is impossible to stain, for ex- photodynamic therapy, noninvasive opti-
ample, only neurons. Genetically encoded cal biopsy of human skin, the study of
protein ‘‘dyes,’’ often called XFPs by gen- cellular metabolism, embryogenesis, im-
eralizing from GFP, the green fluorescent munology, tumor pathophysiology, and
protein, with ‘‘X’’ standing for the color neurodegenerative disease. In the fol-
such as for yellow in YFP and for cyan lowing, we will discuss some of these
in CFP, overcome this problem because applications in more detail.
their expression can be put under the con-
trol of specific promoters that are only 5.1
active in certain cell types. XFPs can also Neurobiology
be fused to other proteins to then map
their dynamic distribution in the cell. Fu- After it became clear that 2PM is par-
sion protein with XFPs have also been ticularly well suited to high-resolution
designed to act as indicators for Ca2+ , imaging in scattering tissue, it was ap-
Cl− , pH, and other cellular signals and plied to neural tissue, which is strongly
physiological parameters such as cAMP, scattering and requires the use of in-
phosphorylation, redox potential, protein tact pieces of tissue for functional studies
kinase B and C, as well as membrane volt- owing to the high degree of cellular in-
age. The molecular genetic methods used terconnectivity. Early studies focused on
to label intact animals are either the cre- Ca2+ dynamics in dendritic spines. Then
ation of transgenic animals, infection with it became clear that it is possible to im-
properly engineered viral vectors, or in vivo age Ca2+ dynamics with high spatial and
electroporation. temporal resolution in the wholly intact
For imaging in intact animals XFPs brain as well, where 2PM can now be
and MPM are rather synergistic, since used to guide recording pipettes to se-
together they solve the deep labeling and lected cells. It has also been possible to
the deep imaging problem, with 2PM even study the structural stability and plastic-
allowing imaging through the thinned but ity of neuronal morphology in explanted
intact skull. tissue and, over periods of months, in
intact animals.
2PM has helped to reveal neural struc-
5 ture and function at various levels ranging
Applications from individual synapses to entire neural
networks (see also Fig. 4). The retina, a
On the basis of the unique imaging spatially separate but functionally integral
properties described earlier, MPM has part of the brain, has been a particularly re-
become the preferred and virtually the warding specimen for 2PM. Because of the
78 Two-photon Microscopy and Imaging

0 0

200 200

621 µm 621 µm
400 400

600 860 µm 600 860 µm

800 800
1028 µm 1028 µm

1000 1000

(a) 50 µm (b) 50 µm

ds
dt
∆t /∆s
∆t
∆s

900 µm
300 ms

25 µm
(c)
Fig. 4 (a) x-z projection and single planar scans of GFP labeled neurons and
(b) stained vasculature obtained throughout almost the entire gray matter of the
mouse neocortex. (c) Blood flow measurement in the mouse neocortex at 900 µm
below the brain surface. Blood cells appear as shadows in surrounding blood plasma.
Their motion traces out shaded bands in an image consisting of line scans repeatedly
taken along the capillaries (dashed line in the planar scan). Blood flow parameters as
velocity, linear density, average spacing, and flux of red blood cells can be inferred from
the slope (dt/ds) of, and the distance (ds) and time (dt) between shaded bands.

retina’s high sensitivity for UV and visible in fact, intensities used commonly for flu-
light, 1PE of common fluorescent probes orescence microscopy completely bleach
inevitably perturbs the observed specimen; the photopigments within seconds. In
Two-photon Microscopy and Imaging 79

contrast, photoreceptors are blind to light muscular contraction and neurotransmit-


in the near-infrared wavelength region, ter secretion, and because [Ca2+ ] tran-
allowing the use of IR-excited 2PM to mea- sients can often be used as a proxy for
sure light stimulus-evoked calcium signal electrical excitation. The particular advan-
in the functionally intact retina, which has tages of MPM are, again, excitation-based
been used to study the mechanism of mo- optical sectioning and highly efficient flu-
tion detection in retinal cells (see Fig. 5). orescence collection in scattering tissue.
A disadvantage of MPM is that ratiomet-
5.2 ric indicators that require dual-wavelength
Calcium Imaging excitation are difficult to use because a
second laser is required for the second
‘‘Calcium imaging’’ constitutes the bulk excitation wavelength (at this point in
of ion imaging, both with 1PM and with time, changing the wavelength even in
MPM, because of the important role of a fully automated Ti:Sapphire lasers still
Ca2+ in coupling electrical processes to takes seconds). Therefore, ratiometric cal-
biochemical processes, for example, in cium imaging with 2PM has to resort

50 µm

(a)
Fig. 5 Two-photon optophysiology in the retina. Dye-filled ‘‘starburst’’
amacrine cell (a) in flat-mounted rabbit retina. Like many amacrine cells,
this neuron bears no axon; it receives inputs and makes output synapses
with its dendrites. Starburst cells are involved in the detection of image
motion. Using (b) two-photon microscopy, light stimulus-evoked Ca2+
signals (green trace) were recorded in the dendritic tips of a starburst
amacrine cell. Simultaneously, membrane voltage (black trace) was
measured at the soma using a patch electrode (see schematic drawing).
The light stimulus, a concentric sinusoidal wave, induced stronger
responses when it was expanding (left) than when it was contracting
(right) (see color plate p. xxv).
80 Two-photon Microscopy and Imaging

‘Expansion’ ‘Contraction’

40%
[Ca2+] ∆F/F

Vm 1 mV

0.5 s
(b)
Fig. 5 (Continued)

to Indo-1, which has drawbacks, such as the activity of respiratory enzymes and
a short absorption wavelength and fast thus is an intrinsic probe of cellular
bleaching, but has been used to mea- metabolism. 1P-CM of two-dimensional
sure calcium oscillations in tumor mast NADH/NADPH-fluorescence maps is pos-
cells and calcium transients in cardiac my- sible, as has been shown in rabbit cornea,
ocytes. More common nowadays is the use but is hampered by photobleaching and
of dye mixtures that provide virtually all photodamage due to the UV illumina-
the advantages of proper ratiometric in- tion (∼360 nm) needed for the excitation
dicators. Furthermore, single-wavelength of NADH/NADPH. Those problems are
indicators are often sufficient, particularly substantially reduced with 2PE so that
if one is interested mainly in the temporal now nearly continuous NADH/NADPH
[Ca2+ ] dynamics (see Figs. 5 and 6). Satis- imaging is possible throughout the entire
factory calibration can often be achieved by thickness (400 µm) of the rabbit cornea.
using the intensity prior to stimulation as Because metabolic events can be fast but
a reference, such as was done in the early may need to be monitored for extended
studies of [Ca2+ ] dynamics in dendrites, time periods, low-damage imaging is of
and dendritic spines in brain slices, and in paramount importance. With MPM, it has
whole animals. become possible to follow the response to
glucose application in β-cells inside intact
5.3 pancreatic islets and in muscle cells.
Imaging of Metabolic Activity
5.4
The fluorescence from the reduced pyri- Photoactivation (Uncaging)
dine nucleotides NADH and NADPH,
which are metabolic intermediates, can The excellent localization of MPE can be
be used to localize and characterize used to confine photochemical generation
Two-photon Microscopy and Imaging 81

(a) 50 µm (b) 5 µm
Fig. 6 Measurement of calcium dynamics in dendritic spines of a hippocampal
neuron (a) using simultaneous excitation, at 910 nm, of two different
fluorophores – a red fluorescent Ca2+ -insensitive dye (Alexa-594) helping in
visualization of small structures and a green fluorescent Ca2+ indicator
(Fluo-5F), which is very dim at rest, but shows large fluorescence changes in
activated spines. (b) Following extracellular electric stimulation, presynaptic
fibers release transmitter in the synaptic cleft whose binding to corresponding
receptors on the postsynaptic membrane causes a Ca2+ rise in a single dendritic
spine leading to an increase in green fluorescence observed by successive planar
scans of a dendritic branchlet (see color plate p. xxvi).

of chemical compounds, such as flu- combination of 2P uncaging and bleaching


orescent tracers or biological signaling of fluorescence has been used to measure
molecules, to volumes smaller than a fem- diffusional resistance of spine necks.
toliter. Typically, this is achieved by the
illumination of a photolabile precursor 5.5
that then absorbs two or more photons Human Diseases and Therapy
and decays via a photochemical pathway
into the desired product. Scanning 2P pho- In conjunction with animal models, MPM
tochemical microscopy, first demonstrated has provided some insights into human
by mapping the distribution of neurotrans- diseases such as Alzheimer’s demen-
mitter receptors on cultured muscle-like tia. There, 2PM in vivo imaging allows
cells, has, however, been hampered by the the observation over time of how senile
very low 2PE cross sections of the available plaques (amyloid-beta peptide aggregates)
‘‘caged’’ compounds (an example of the develop. This is extremely helpful when
rule that weak 1P absorbers tend to be very evaluating antiplaque therapies, since the
weak 2P absorbers) even though there has fate of individual plaques can be fol-
been recent progress in the synthesis of lowed over weeks and months by repeated
better caged compounds, having allowed imaging through a thinned skull. These
the study of calcium ‘‘sparks’’ in cardiac experiments have shown that, surpris-
muscle cells and the mapping of gluta- ingly, plaques do not change size af-
mate sensitivities on dendritic spines. A ter formation.
82 Two-photon Microscopy and Imaging

2PM has also been used to measure MPM, the range of applications that bene-
blood flow at the level of single capillar- fit from MPM continues to expand. Much
ies and to observe leucocytes-endothelial of the current improvement efforts are
interactions in tumors in vivo at tissue directed at excitation and detection effi-
depths that are not accessible otherwise. ciencies, which, as was discussed at length
Such studies have revealed, for example, earlier, is the main factor determining
the depth dependent dynamics of tumor imaging depth and specific (i.e. per in-
angiogenesis. In the area of immunology, formation gained) photodamage. Novel
2PM allows the observation of lymphocytes chemical fluorophores with large two pho-
in their native environment such as mon- ton–absorption cross sections still await
itoring the motility of T cells that invade significant biological applications. Closer
the brain during neuroinflammation. to being really useful appear to be semicon-
Potential clinical applications on MPM ductor quantum dots (QDs), which have
are optical skin biopsy and photodynamic the largest 2PA cross sections measured,
therapy. Traditional biopsy is an invasive are very photostable, and are available with
procedure, which requires the removal and widely varying but narrow emission spec-
tra. Recent advances have all but solved
fixation of tissue before imaging. During
early problems with solubility, quenching,
this process, biochemical information is
and toxicity. Different coats allow covalent
often poorly preserved. Optical biopsy does
linking to biorecognition molecules, such
not require the removal of any tissue
as antibodies or biotin/avidin, so that QDs
sample, but it remains to be seen whether
can be used as specific fluorescent probes.
the quality of 2PM images will become
In conjunction with their broad excitation
good enough to allow a pathological
spectra, QDs are very well suited for mul-
analysis with accuracy comparable to that
tilabel imaging.
of traditional biopsies. An entirely different route to improving
Photodynamic therapy is based on the excitation efficiency is the use of coherent
preferential accumulation of certain pho- control, which involves tailoring the phase
tosensitizing agents in tumorous or other- and amplitude of a laser pulse to optimize
wise abnormal tissues and the subsequent the optical response of a molecule, to
selective tissue destruction by illumina- achieve selective excitation, or to reduce
tion. 2PE photodynamic therapy has the photobleaching.
potential of providing more spatially spe- Inhomogeneities in the index of refrac-
cific destruction of, for example, cancer tion across the tissue lead to degradation
tissue than 1PE, which can be confined of the focus and thus a reduction in 2PE
not as easily or not at all. efficiency. In particular at large depths,
therefore, adaptive aberration correction
can significantly improve excitation.
6 In order to image structures that are
Future Directions beyond the penetration depth of the MPM,
overlying tissue has to be removed. In
Among the ‘‘exponentially’’ growing num- order to keep the lateral extent of such
ber of papers on MPM, about half are removal, limited ‘‘endoscopic’’ approaches
still on methods development. With these are being developed that use a small
ongoing improvements and extensions of diameter gradient-index (GRIN) lens as
Two-photon Microscopy and Imaging 83

the objective. This limits the NA and the Denk, W., Piston, D.W., Webb, W.W. (1995)
field of view, but may be the only way to Two-photon molecular excitation in laser
scanning microscopy, in: Pawley, J. (Ed.) The
reach deeper brain structures such as the
Handbook of Confocal Microscopy, Plenum,
hippocampus in vivo. New York, pp. 445–458.
Regular MP microscopes are bulky and Diaspro, A., 2002 Confocal and Two-photon
thus most in vivo imaging is done in Microscopy: Foundations, Applications, and
anesthetized and immobilized animals. To Advances, Wiley-Liss, New York, pp. xix,
achieve the goal of high-resolution imag- 567.
Helmchen, F., Denk, W. (2002) New develop-
ing in freely moving animals, scanners
ments in multiphoton microscopy, Curr. Opin.
and optics have to be miniaturized. This Neurobiol. 12(5), 593–601.
requires a different scanning mechanism König, K. (2000) Multiphoton microscopy in life
because the weight of a galvo scanner rules sciences, J. Microsc. (Oxford) 200, 83–104.
out its use in a headpiece that can weigh at Masters, B.R., Thompson, B.D. (Eds.) (2003)
most a few tens of grams without encum- Selected Papers on Multi-photon Excitation
Microscopy, SPIE Milestone Series, Vol.
bering an animal such as a rat too much. MS175. SPIE Press, Bellingham, D.C.
An alternative is the ‘‘2P-fiberscope’’ in Mertz, J. (2004) Nonlinear microscopy: new
which scanning is achieved by deflecting techniques and applications, Curr. Opin.
the free end of the fiber that is used to Neurobiol. 14(5), 610–616.
deliver the excitation light to the head- So, P.T.C., et al. (2000) Two-photon excitation
fluorescence microscopy, Annu. Rev. Biomed.
piece. The deflection can, for example,
Eng. 2, 399–429.
be achieved by using resonant oscilla- Zipfel, W.R., Williams, R.M., Webb, W.W.
tions. One challenge peculiar to nonlinear (2003) Nonlinear magic: multiphoton
fiber microscopy is that special measures, microscopy in the biosciences, Nat. Biotechnol.
including the use of large core or mi- 21(11), 1368–1376.
crostructured fibers, have to be taken to
limit the degradation of the pulse shape Primary Literature
by nonlinear optical effects in the beam-
Abbe, E. (1873) Beiträge zur Theorie des
delivery fiber.
Mikroskops und der mikroskopischen
Wahrnehmung, Schultzes Arch. Mikrosk. Anat.
9, 413–468.
See also Alzheimer’s Disease; Cal- Albota, M., et al. (1998) Design of organic
molecules with large two-photon absorption
cium Biochemistry; Metabolic Ba- cross sections, Science 281(5383), 1653–1656.
sis of Cellular Energy; Molecular Barad, Y., et al. (1997) Nonlinear scanning laser
microscopy by third harmonic generation,
Neurobiology, Single-Cell; Neuron Appl. Phys. Lett. 70(8), 922–924.
Chemistry. Bardeen, C.J., et al. (1999) Effect of pulse shape
on the efficiency of multiphoton processes:
implications for biological microscopy, J.
Bibliography Biomed. Opt. 4(3), 362–367.
Beaurepaire, E., Mertz, J. (2002) Epifluorescence
collection in two-photon microscopy, Appl.
Books and Reviews Opt. 41(25), 5376–5382.
Beaurepaire, E., Oheim, M., Mertz, J. (2001)
Denk, W., Svoboda, K. (1997) Photon upman- Ultra-deep two-photon fluorescence excitation
ship: why multiphoton imaging is more than in turbid media, Opt. Commun. 188(1–4),
a gimmick, Neuron 18(3), 351–357. 25–29.
84 Two-photon Microscopy and Imaging

Bennett, B.D., et al. (1996) Quantitative sub- de Grauw, G.J., et al. (1999) Imaging properties
cellular imaging of glucose metabolism within in two-photon excitation microscopy and
intact pancreatic islets, J. Biol. Chem. 271(7), effects of refractive-index mismatch in thick
3647–3651. specimens, Appl. Opt. 38(28), 5995–6003.
Bewersdorf, J., Pick, R., Hell, S.W. (1998) Denk, W. (1994) 2-photon scanning photo-
Multifocal multiphoton microscopy, Opt. Lett. chemical microscopy-mapping ligand-gated
23(9), 655–657. ion-channel distributions, Proc. Natl. Acad. Sci.
Bhawalkar, J.D., et al. (1997) Two-photon U.S.A. 91(14), 6629–6633.
photodynamic therapy, J. Clin. Laser Med. Surg. Denk, W., et al. (1994) Anatomical and
15, 201–204. functional imaging of neurons using 2-photon
Bird, D., Gu, M. (2003) Two-photon fluorescence laser-scanning microscopy, J. Neurosci.
endoscopy with a micro-optic scanning head, Methods 54(2), 151–162.
Opt. Lett. 28(17), 1552–1554. Denk, W., Detwiler, P.B. (1999) Optical
Birge, R.R. (1986) 2-photon spectroscopy of recording of light-evoked calcium signals in
protein-bound chromophores, Acc. Chem. Res. the functionally intact retina, Proc. Natl. Acad.
19(5), 138–146. Sci. U.S.A. 96(12), 7035–7040.
Brakenhoff, G.J., et al. (1996) Real-time two- Denk, W., Strickler, J.H., Webb, W.W. (1990)
photon confocal microscopy using a Two-photon laser scanning fluorescence
femtosecond, amplified Ti:sapphire system, J. microscopy, Science 248, 73–76.
Microsc. (Oxford) 181, 253–259. Denk, W., Sugimori, M., Llinas, R. (1995) Two
Brown, E.B., et al. (2001) In vivo measurement types of calcium response limited to single
of gene expression, angiogenesis and
spines in cerebellar Purkinje cells, Proc. Natl.
physiological function in tumors using
Acad. Sci. U.S.A. 92(18), 8279–8282.
multiphoton laser scanning microscopy, Nat.
Eilers, J., et al. (2001) GABA-mediated Ca2+
Med. 7(7), 866–870.
signaling in developing rat cerebellar Purkinje
Centonze, V.E., White, J.G. (1998) Multiphoton
neurones, J. Physiol. (London) 536(2),
excitation provides optical sections from
429–437.
deeper within scattering specimens than
Engert, F., Bonhoeffer, T. (1999) Dendritic spine
confocal imaging, Biophys. J. 75(4),
changes associated with hippocampal long-
2015–2024.
term synaptic plasticity, Nature 399(6731),
Chalfie, M., et al. (1994) Green fluorescent
66–70.
protein as a marker for gene expression,
Science 263(5148), 802–805. Euler, T., Detwiler, P.B., Denk, W. (2002)
Chan, W.C.W., Nie, S.M. (1998) Quantum dot Directionally selective calcium signals in
bioconjugates for ultrasensitive nonisotopic dendrites of starburst amacrine cells, Nature
detection, Science 281(5385), 2016–2018. 418(6900), 845–852.
Chance, B., Thorell, B. (1959) Localization and Fan, G.Y., et al. (1999) Video-rate scanning two-
kinetics of reduced pyridine nucleotide in photon excitation fluorescence microscopy
living cells by microfluorometry, J. Biol. Chem. and ratio imaging with cameleons, Biophys.
234(11), 3044–3050. J. 76(5), 2412–2420.
Christie, R.H., et al. (2001) Growth arrest of Feierabend, M., Ruckel, M., Denk, W. (2004)
individual senile plaques in a model of Coherence-gated wave-front sensing in
Alzheimer’s disease observed by in vivo strongly scattering samples, Opt. Lett. 29(19),
multiphoton microscopy, J. Neurosci. 21(3), 2255–2257.
858–864. Fork, R.L., Martinez, O.E., Gordon, J.P. (1984)
Chung, M.A., Lee, K.S., Jung, S.D. Two- Negative dispersion using pairs of prisms, Opt.
photon absorption cross sections of Lett. 9(5), 150–152.
dithienothiophene-based molecules, ETRI J. Friedrich, D.M. (1982) 2-photon molecular-
2002. 24(3), 221–225. spectroscopy, J. Chem. Educ. 59(6), 472–481.
Cohen, L.B., et al. (1974) Changes in axon Gannaway, J.N., Sheppard, C.J.R. (1978) Second-
fluorescence during activity: molecular probes harmonic imaging in the scanning optical
of membrane potential, J. Membr. Biol. 19(1), microscope, Opt. Quantum Electron. 10(5),
1–36. 435–439.
Two-photon Microscopy and Imaging 85

Goeppert-Mayer, M. (1931) Ueber Elemen- mammalian cells, Proc. Natl. Acad. Sci. U.S.A.
tarakte mit zwei Quantenspruengen, Ann. 96(11), 6255–6260.
Phys. 9, 273. Jovin, T.M. 2003 Quantum dots finally come of
Griesbeck, O. (2004) Fluorescent proteins as age, Nat. Biotechnol. 21(1), 32–33.
sensors for cellular functions, Curr. Opin. Jung, J.C., Schnitzer, M.J. (2003) Multiphoton
Neurobiol. 14(5), 636–641. endoscopy, Opt. Lett. 28(11), 902–904.
Grutzendler, J., Kasthuri, N., Gan, W.B. (2002) Kaiser, W., Garrett, C.B.G. (1961) Two-photon
Long-term dendritic spine stability in the adult excitation in CaF2 :Eu2+ , Phys. Rev. Lett. 7(6),
cortex, Nature 420(6917), 812–816. 229–231.
Gu, M. (1996) Resolution in 3-photon Kawano, H., et al. (2003) Attenuation of
fluorescence scanning microscopy, Opt. Lett. photobleaching in two-photon excitation
21(13), 988–990. fluorescence from green fluorescent protein
Gu, M., Sheppard, C.J.R. (1995) Comparison of with shaped excitation pulses, Biochem.
3-dimensional imaging properties between Biophys. Res. Commun. 311(3), 592–596.
2-photon and single-photon fluorescence Kettunen, P., et al. (2002) Imaging calcium
microscopy, J. Microsc. (Oxford) 177, 128–137. dynamics in the nervous system by means
Hanninen, P., Soini, E., Hell, S. (1994) Contin- of ballistic delivery of indicators, J. Neurosci.
uous wave excitation two-photon fluorescence Methods 119(1), 37–43.
microscopy, J. Microsc. (Oxford) 176, 222–225. Kim, K.H., Buehler, C., So, P.T.C. (1999) High-
Hell, S., Stelzer, E.H.K. (1992) Fundamental speed, two-photon scanning microscope, Appl.
improvement of resolution with a 4pi- Opt. 38(28), 6004–6009.
confocal fluorescence microscope using 2- Kleinfeld, D., et al. (1998) Fluctuations and
photon excitation, Opt. Commun. 93(5–6), stimulus-induced changes in blood flow
277–282.
observed in individual capillaries in layers 2
Hell, S., Wichmann, J. (1994) Breaking the
through 4 of rat neocortex, Proc. Natl. Acad.
diffraction resolution limit by stimulated
Sci. U.S.A. 95(26), 15741–15746.
emission: stimulated emission depletion
Koester, H.J., et al. (1999) Ca2+ fluorescence
fluorescence microscopy, Opt. Lett. 19(11),
imaging with pico- and femtosecond two-
780–782.
photon excitation: signal and photodamage,
Hell, S.W., Dyba, M., Jakobs, S. (2004) Concepts
Biophys. J. 77(4), 2226–2236.
for nanoscale resolution in fluorescence
König, K., Tirlapur, U.K. (2002) Cellular and
microscopy, Curr. Opin. Neurobiol. 14(5),
599–609. Subcellular Perturbations during Multiphoton
Hellwarth, R., Christiansen, P. (1974) Nonlinear Microscopy, in: Diaspro, A. (Ed.) Confocal
optical microscopy examination of structure and Two-photon Microscopy. Foundations,
in polycrystalline ZnSe, Opt. Commun. 12(3), Applications and Advances, Wiley-Liss, New
318–322. York, pp. 191–205.
Helmchen, F., et al. (1999) In vivo dendritic Kuhn, B., Fromherz, P., Denk, W. (2004) High
calcium dynamics in deep-layer cortical sensitivity of stark-shift voltage-sensing dyes
pyramidal neurons, Nat. Neurosci. 2(11), by one- or two-photon excitation near the red
989–996. spectral edge, Biophys. J. 87(1), 631–639.
Helmchen, F., et al. (2001) A miniature Lakowicz, J.R., et al. (1999) Advances in
head-mounted two-photon microscope high- fluorescence spectroscopy: multi-photon
resolution brain imaging in freely moving excitation, engineered proteins, modulation
animals, Neuron 31(6), 903–912. sensing and microsecond rhenium metal-
Helmchen, F., Tank, D.W., Denk, W. (2002) ligand complexes, Acta. Phys. Pol. A 95(1),
Enhanced two-photon excitation through 179–196.
optical fiber by single-mode propagation in Larson, D.R., et al. (2003) Water-soluble
a large core, Appl. Opt. 41(15), 2930–2934. quantum dots for multiphoton fluorescence
Hentschel, M., et al. (2001) Attosecond metrol- imaging in vivo, Science 300(5624), 1434–1436.
ogy, Nature 414(6863), 509–513. Lechleiter, J.D., Lin, D.T., Sieneart, I. (2002)
Hockberger, P.E., et al. (1999) Activation Multi-photon laser scanning microscopy using
of flavin-containing oxidases underlies an acoustic optical deflector, Biophys. J. 83(4),
light-induced production of H2O2 in 2292–2299.
86 Two-photon Microscopy and Imaging

Lipp, P., Niggli, E. (1998) Fundamental calcium Muller, M., et al. (1998) Dispersion pre-
release events revealed by two-photon compensation of 15 femtosecond optical
excitation photolysis of caged calcium in pulses for high-numerical-aperture objectives,
guinea-pig cardiac myocytes, J. Physiol. J. Microsc. (Oxford) 191, 141–150.
(London) 508(3), 801–809. Neil, M.A.A., et al. (2000) Adaptive aberration
Lippitz, M., et al. (2002) Two-photon excitation correction in a two-photon microscope, J.
microscopy of tryptophan-containing proteins, Microsc. (Oxford) 200, 105–108.
Proc. Natl. Acad. Sci. U.S.A. 99(5), 2772–2777. Nielsen, T., et al. 2001 High efficiency
Lombardo, J.A., et al. (2003) Amyloid-beta anti- beam splitter for multifocal multiphoton
body treatment leads to rapid normalization microscopy, J. Microsc. 201, 368–376.
of plaque-induced neuritic alterations, J. Neu- Nitsch, R., et al. (2004) Direct impact of T cells on
rosci. 23(34), 10879–10883. neurons revealed by two-photon microscopy
Maiti, S., et al. (1997) Measuring serotonin in living brain tissue, J. Neurosci. 24(10),
distribution in live cells with three-photon 2458–2464.
excitation, Science 275(5299), 530–532. Norris, T.B. (1992) Femtosecond pulse amplifica-
Maletic-Savatic, M., Malinow, R., Svoboda, K. tion at 250 kHz with a Ti-sapphire regenerative
(1999) Rapid dendritic morphogenesis in CA1 amplifier and application to continuum gen-
hippocampal dendrites induced by synaptic eration, Opt. Lett. 17(14), 1009–1011.
activity, Science 283(5409), 1923–1927. Ouzounov, D.G., et al. (2002) Delivery of
Margrie, T.W., et al. (2003) Targeted whole-cell nanojoule femtosecond pulses through large-
recordings in the mammalian brain in vivo, core microstructured fibers, Opt. Lett. 27(17),
Neuron 39(6), 911–918. 1513–1515.
Masters, B.R., Kriete, A., Kukulies, J. (1993) Pastirk, I., et al. (2003) Selective two-photon
Ultraviolet confocalfluoresence microscopy in
microscopy with shaped femtosecond pulses,
the in vitro cornea: redox metabolic imaging,
Opt. Express 11(14), 1695–1701.
Appl. Opt. 32(4), 592–596.
Patterson, G.H., Piston, D.W. (2000) Photo-
Masters, B.R., So, P.T.C., Gratton, E. (1997) Mul-
bleaching in two-photon excitation mi-
tiphoton excitation fluorescence microscopy
croscopy, Biophys. J. 78(4), 2159–2162.
and spectroscopy of in vivo human skin, Bio-
Pawley, J., (Ed.) 1995 Handbook of Biological
phys. J. 72(6), 2405–2412.
Confocal Microscopy, 2nd edition, Kluwer
Matsuzaki, M., et al. (2001) Dendritic spine
geometry is critical for AMPA receptor Academic Publishers, Norwell, MA.
expression in hippocampal CA1 pyramidal Pelliccioli, A.P., Wirz, J 2002 Photoremovable
neurons, Nat. Neurosci. 4(11), 1086–1092. protecting groups: reaction mechanisms and
McClain, W.M. Excited state symmetry applications, Photochem. Photobiol. Sci. 1(7),
assignment through polarized two-photon 441–458.
absorption studies of fluids, J. Chem. Phys. Piston, D.W., et al. (1994) 2-photon-excitation
1971. 55(6), 2789–2796. fluorescence imaging of 3-dimensional
Mempel, T.R., et al. (2004) In vivo imaging of calcium-ion activity, Appl. Opt. 33(4), 662–669.
leukocyte trafficking in blood vessels and Piston, D.W., Knobel, S.M. (1999) Real-time
tissues, Curr. Opin. Immunol. 16(4), 406–417. analysis of glucose metabolism by microscopy,
Miller, M.J., et al. (2002) Two-photon imaging Trends Endocrinol. Metab. 10(10), 413–417.
of lymphocyte motility and antigen response Piston, D.W., Masters, B.R., Webb, W.W. (1995)
in intact lymph node, Science 296(5574), 3-dimensionally resolved Nad(P)H cellular
1869–1873. metabolic redox imaging of the in-situ
Minsky, M. (1961) Microscopy Apparatus, in US cornea with 2-photon excitation laser-scanning
Patent, 3013467. microscopy, J. Microsc. (Oxford) 178, 20–27.
Minta, A., Kao, J.P., Tsien, R.Y. (1989) Fluores- Potter, S.M., Pine, J., Fraser, S.E. (1996) Neural
cent indicators for cytosolic calcium based on transplant staining with DiI and vital imaging
rhodamine and fluorescein chromophores, J. by 2-photon laser-scanning microscopy,
Biol. Chem. 264(14), 8171–8178. Scanning Microsc. Suppl. 10, 189–199.
Miyawaki, A., et al. (1997) Fluorescent indicators Prasher, D.C., et al. (1992) Primary structure
for Ca2+ based on green fluorescent proteins of the Aequorea victoria green-fluorescent
and calmodulin, Nature 388(6645), 882–887. protein, Gene 111(2), 229–233.
Two-photon Microscopy and Imaging 87

Shen, Y.R. (1984) The Principles of Nonlinear Theer, P., Hasan, M.T., Denk, W. (2003) Two-
Optics, Wiley, New York. photon imaging to a depth of 1000 mu m in
Sheppard, C.J.R. (1980) Scanning optical living brains by use of a Ti:Al2O3 regenerative
microscope, Electron. Power 26(2), 166–172. amplifier, Opt. Lett. 28(12), 1022–1024.
Shirakawa, A., Sakane, I., Kobayashi, T. (1998) Trachtenberg, J.T., et al. (2002) Long-term in vivo
Pulse-front-matched optical parametric ampli- imaging of experience-dependent synaptic
fication for sub-10-fs pulse generation tunable plasticity in adult cortex, Nature 420(6917),
in the visible and near infrared, Opt. Lett. 788–794.
23(16), 1292–1294. Treacy, E. (1969) Optical Pulse Compression
Spence, D.E., Kean, P.N., Sibbett, W. (1991) 60- With Diffraction Gratings, IEEE J. Quantum
fsec pulse generation from a self-mode-locked Electron. QE-5(9), 454–458.
Ti:sapphire laser, Opt. Lett. 16(1), 42–44. Tsien, R.Y. (1981) A non-disruptive technique
Squier, J., Muller, M. (2001) High resolution for loading calcium buffers and indicators into
nonlinear microscopy: a review of sources and cells, Nature 290(5806), 527–528.
methods for achieving optimal imaging, Rev. Tsien, R.Y. (1989) Fluorescent-Probes of Cell
Sci. Instrum. 72(7), 2855–2867. Signaling, Annu. Rev. Neurosci. 12, 227–253.
Squirrell, J.M., et al. (1999) Long-term two- White, J., Amos, W., Fordham, M. (1987) An
photon fluorescence imaging of mammalian evaluation of confocal versus conventional
embryos without compromising viability, Nat. imaging of biological structures by
Biotechnol. 17(8), 763–767. fluorescence light microscopy, J. Cell Biol. 105,
Stelzer, E.H.K., et al. (1994) Nonlinear absorp- 41–48.
tion extends confocal fluorescence microscopy Wilson, T., Sheppard, C. (1984) Theory and
into the ultra-violet regime and confines the Practice of Scanning Optical Microscopy,
illumination volume, Opt. Commun. 104(4–6), Academic Press, London.
223–228. Xu, C., Webb, W.W. (1996) Measurement of two-
Stosiek, C., et al. (2003) In vivo two-photon photon excitation cross sections of molecular
calcium imaging of neuronal networks, Proc. fluorophores with data from 690 to 1050 nm,
Natl. Acad. Sci. U.S.A. 100(12), 7319–7324. J. Opt. Soc. Am. B 13(3), 481–491.
Svaasand, L.O., Ellingsen, R. (1983) Optical Yaroslavsky, A.N., et al. (2002) Optical properties
properties of human brain, Photochem. of selected native and coagulated human
Photobiol. 38(3), 293–299. brain tissues in vitro in the visible and near
Svoboda, K., et al. (1997) In vivo dendritic infrared spectral range, Phys. Med. Biol. 47(12),
calcium dynamics in neocortical pyramidal 2059–2073.
neurons, Nature 385(6612), 161–165. Yasuda, R., et al. (2004) Imaging calcium
Svoboda, K., Tank, D.W., Denk, W. (1996) concentration dynamics in small neuronal
Direct measurement of coupling between compartments, Sci. STKE 2004(219), pl5.
dendritic spines and shafts, Science 272(5262), Yuste, R., Denk, W. (1995) Dendritic spines as
716–719. basic functional units of neuronal integration,
Szmacinski, H., Gryczynski, I., Lakowicz, J.R. Nature 375(6533), 682–684.
(1998) Spatially localized ballistic two-photon Zhang, J., et al. (2002) Creating new fluorescent
excitation in scattering media, Biospectroscopy probes for cell biology, Nat. Rev. Mol. Cell Biol.
4(5), 303–310. 3(12), 906–918.
Tan, Y.P., et al. (1999) Fast scanning and Zumbusch, A., Holtom, G.R., Xie, X.S. (1999)
efficient photodetection in a simple two- Three-dimensional vibrational imaging by
photon microscope, J. Neurosci. Methods coherent anti-stokes Raman scattering, Phys.
92(1–2), 123–135. Rev. Lett. 82(20), 4142–4145.

Ubiquitin Mediation Complex:


see Proteasomes

You might also like