FIGARO

PRODUCT INFORMATION

Features:

GS Oxygen Sensors
Applications:
* Medical - Anesthetic instruments, respirators, oxygen-enrichers * Biotechnology - Oxygen incubators * Food industry - Refrigeration, greenhouses * Safety - Air conditioners, oxygen detectors, re detectors

* Long life KE-25 - 5 years / KE-50 - 10 years in ambient air * Virtually no inuence from CO2, CO, H2S, NOx, H2 * Low cost * Operates in normal ambient temperatures * Stable output signal * No external power supply required for sensor operation * No warmup time is required

The GS Oxygen Sensor KE series (KE-25 and KE-50) is a unique galvanic cell type oxygen sensor which was developed in Japan in 1985. Its most notable features are long life expectency, excellent chemical durability, and it is not inuenced by CO2. The KE series oxygen sensor is ideal to meet the ever-increasing demand for oxygen monitoring in various elds such as combustion gas monitoring, the biochemical eld, medical applications, domestic combustion appliances, etc.

Sensitivity characteristics (typical values under std.


test conditions)
325 300 250
Output Voltage (mV)

Response time (typical)


120 100 80 60 40 20
KE-25

150 100 50
KE-50
KE-25

Output Ratio (%)

200

KE-50

10

20

30

40

50

60

70

80 90 100

30

60

90
Time (sec.)

120

150

180

Oxygen Concentration (%)

Dimensions
KE-25/KE-50 standard version KE-25F1 (w/o flange)
22.70.5

KE-25F3 (threaded top) 22.70.5

KE-25F4 (O-ring top) 22.70.5

280.5 22.50.3

230.5

230.5

805 42
1.5

47.30.5

JAPAN STORAGE

KENO.94

23.20.5

13

13

5 9

5 9

9
Unit = mm

15

15

Specications
Item

Model

K E- 2 5

KE-50 0~100% O2

Measurement range Accuracy (Note 1) Atmospheric pressure Operating conditions Temperature Relative humidity Response time (90%) (Note 4) Initial output voltage under standard test conditions Test gas Standard test conditions Atmospheric pressure Temperature Relative humidity Linearity Offset voltage Temperature characteristics (Note 3) (Va-V0)/(V100-V0) (Note 2) V0 VH/Va VL/Va

Notes: 1) When calibrated at both 0% and 100% of O2, accuracy in the range from 0-100% O2 shall be within 1% of full scale for KE-25 and 2% of full scale for KE-50. 2) Va = output voltage at 21% O2 V0 = output voltage at 0% O2 V100 = output voltage at 100% O2 3) Va = output voltage at 25C VH = output voltage at 40C VL = output voltage at 5C 4) Sensors should be used under conditions where the air exchange is greater than 200~300ml per minute in order to obtain the response speed as specied in Table 1.

1% (full scale)

2% (full scale)

811hPa ~1216hPa 5~40C 10 ~ 90%R.H. (no condensation) 142 seconds 10.0~15.5mV 21% O 2 10135hPa 25C1C 605% 0.210.02 605 seconds 47~65mV

0.5mV
0.91~1.09 0.91~1.09

6.0mV

FIGARO ENGINEERING, INC. 1-5-11 Senba-nishi Mino. Osaka 562-8505 Phone: (81)-72-728-2561 Fax: (81)-72-728-0467 email: garo@garo.co.jp

REV: 08/03

4.5

23.20.5

23.20.5

16 16.6

15

M16 x P1.0

2.5

570.5

Lead wire

500.5

OXYGEN

47.30.5

230.5