Owlstone Inc.

Detecting H2S Scavengers


and their Byproducts with
Lonestar
Owlstone Whitepaper
Alasdair Edge
alasdair.edge@owlstone.co.uk
006012 v 1.0

Owlstones Lonestar Analyzer provides an accurate, portable and easy-to-use means for nonspecialist personnel to carry out real-time, at-line monitoring of H2S scavengers and their reaction
by-products.

Detecting H2S Scavengers and their Byproducts


with Lonestar

Hydrogen Sulfide in Crude: A Problematic Contaminant


Hydrogen sulfide (H2S) is a poisonous, corrosive contaminant found in unrefined crude oil and natural gas.
It may also be produced by heating sulfur-containing compounds in oil during the refining process. H2S is
highly toxic, with 1000 ppm most likely fatal within a few inhalations, and an acceptable ceiling
concentration of just 20 ppm given by the Occupational Safety and Health Administration (OSHA Report).
Although it has a strong and characteristic odor (of rotten eggs) at low concentrations, at levels above
about 100 ppm it can paralyze the olfactory nerve, dulling the sense of smell and meaning operators may
be unaware of the presence of potentially dangerous amounts of the gas. It is heavier than air, so will tend
to collect at the bottom of confined spaces.
H2S in crude oil and natural gas can also cause damage to pipework, either by reacting directly with steel to
create an iron sulfide corrosion film, or by increasing the acidity of the liquid/gas mixture in the pipes (Sun
and Nesic 2007). Furthermore, when dissolved in water, H2S may be oxidized to form elemental sulfur. This
can also produce an iron sulfide corrosion film when in direct contact with the metal surface (Fang, Young
and Nesic 2008).
The average sulfur content (and by implication the H2S content) of crude oil input to US refineries has been
on an upward trend over the last 20 years (EIA report), while permissible emissions limits have been
tending downwards (EPA Standards of Performance for Petroleum Refineries).
For all of these reasons, it is essential to remove H2S from crude oil as quickly and efficiently as possible. To
do so, many companies utilize chemical compounds known as scavengers, which remove H2S in a process
sometimes known as stripping.

What Kinds of Scavengers Exist?


There are two broad classes: regenerative and non-regenerative, which undergo reversible and nonreversible reactions with H2S respectively. Regenerative include alkanolamines (MEA, DEA, MDEA and
DGA). May also include oxidizers such as peroxide and caustic (NaOH or NaOH/KOH). Non-regenerative
include metal oxides (e.g. Fe2O3 or ZnO), oxidizing chemicals such as KMnO4 or K2Cr2O7, aldehydes, metal
carboxylates and chelates, and the most widely-used, triazines (Madsen 2011). Of these, it is 1,3,5-tri-(2hydroxyethyl)-hexahydro-s-triazine (HHTT, CAS number 4719-04-4) that is primarily used in oil refining,
where it is usually simply referred to as triazine.

How Does the HHTT Scavenging Process Work?


The HHTT molecule consists of a central s-triazine ring, with a 2-hydroxyethyl group attached to each
nitrogen. In the scavenging reaction, the sulfur replaces the nitrogen and 2-hydroxyethyl, which then
combine with the hydrogen from the H2S to form monoethanolamine (MEA). In principle, all three
nitrogens could be replaced, to leave trithiane (C3H6S3) and three MEA molecules. Under normal refinery
conditions, though, only two sulfurs displace nitrogen from the HHTT, forming dithiazine (5-(2-

All rights reserved 2013. Owlstone Inc.

Detecting H2S Scavengers and their Byproducts


with Lonestar

hydroxyethyl)hexahydro-1,3,5-dithiazine) and two MEA molecules. Pictorially, the reaction proceeds as


follows:

Figure 1 - Triazine Scavenging Reaction

What Are the Effects of Excess Scavengers?


HHTT scavengers are effective in removing H2S from crude oil, but are not themselves free of problems.
Refineries have reported fouling of systems and solid deposits caused by crude oil treated with HHTT
scavengers. It is believed (Taylor and Matherly 2011, Taylor et al 2012) that this is due to the formation of a
polymer of dithiazine linked by long carbon and sulfur chains. This material is also known as amorphous
dithiazine, and appears to begin forming when HHTT is around 60% spent. In order to prevent the
formation of amorphous dithiazine, levels of HHTT, MEA and dithiazine must be closely monitored.
Owlstones Lonestar Analyzer provides a rapid, portable and easy-to-use way of doing so.

How Are Scavengers Detected and Scavenger Levels Monitored?


The Lonestar system provides chemical identification and quantification using field asymmetric ion mobility
spectrometry (FAIMS). The FAIMS process comprises three phases: ionization, separation and
measurement.
In the first phase, vapor from a volatile sample is ionized using a 63Ni ionizer. The ions then pass into a
channel, where they are exposed to an electric field perpendicular to their direction of travel, created by an
alternating voltage of the form shown in Figure 2.

All rights reserved 2013. Owlstone Inc.

Detecting H2S Scavengers and their Byproducts


with Lonestar

Figure 2 Alternating Square Wave


This voltage causes the ions to travel in a saw-tooth trajectory, as in Figure 3. The product of the voltage
and the time for the short-duration, high positive voltage and long-duration, low negative voltage sections
are the same. This means that if the ions mobility, k (where v=kE, v is the velocity of ion travel and E is the
electric field strength) is the same in the high and low electric fields, then it will travel through the channel
along the path marked 1 in Figure 3. In general, though, ions mobility changes with field strength. Ions
whose mobility increases with field strength will travel along a path like that marked 2, and hit one side of
the filter, while those whose mobility decreases will travel along a path like that marked 3, and hit the
other side of the filter. Measuring the ion current leaving the filter allows us to tell many ions travelled
along path 1, allowing us to work out the concentration of those ions in the original mixture.

Figure 3 Ion trajectories in FAIMS

All rights reserved 2013. Owlstone Inc.

Detecting H2S Scavengers and their Byproducts


with Lonestar

If we wish to change which ions pass through the filter, we can apply an additional DC compensation
voltage (CV) across the channel, which will result in an additional drift velocity towards one of the sides.
The strength of the compensation voltage necessary to allow a particular ion to pass through the filter
depends on the difference between its high- and low-field mobilities, and this is how FAIMS can be used to
test for particular substances. By applying the compensation voltage for the substance we are interested in,
and then measuring whether an ion current is produced at the end of the channel, we can test whether
that substance is present.
HHTT, MEA and dithiazine all pass through the system at different compensation voltages, so the amounts
of each present in a particular crude oil sample can be determined by measuring the ion currents at their
respective compensation voltages. Calibration for each compound is carried out by measuring the ion
current produced by a range of samples of known concentrations, and then interpolating the results.

Experimental Setup and Sample Preparation


FAIMS analysis is very simple to carry out using the Lonestar. Crude oil is placed into a glass vial, which is
then inserted into a sample holder and attached to the system. The air flow through the system is opened,
to ensure the sample vapor enters the ion channels, and analysis is carried out at the push of a button. The
whole process can be seen in action here.

All rights reserved 2013. Owlstone Inc.

Detecting H2S Scavengers and their Byproducts


with Lonestar

Figure 4 Lonestar photograph and experimental schematic

Owlstone Results: Detection of Scavengers in Crude


Figure 5 shows the ion current recorded as the compensation voltage is swept from -6 to +6V. The peak due
to triazine (HHTT), seen at around -0.4V, is clearly separated from the peak due to the crude oil
background, at around 0.8V. This shows that detection of HHTT in crude can be easily achieved.

Figure 5 - Triazine peak for desalter crude

All rights reserved 2013. Owlstone Inc.

Detecting H2S Scavengers and their Byproducts


with Lonestar

Figure 6 shows the calibration curve for triazine (HHTT) in crude oil. The limit of detection for the system is
taken to be 3x the standard deviation of the background noise, which in this case is equivalent to
approximately 0.2 A.U. Thus, triazine can be detected down to 3ppm with this setup.

Figure 6 - Calibration curve for Eagle Ford crude, showing detection of HHTT down to 3ppm.

Owlstone Results: Detection of Scavengers and By-Products in Water


Figure 7 shows the results obtained from a commercially available HHTT scavenger sample diluted 1000
times with water, which was then spiked with known concentrations (25ppm) of MEA and dithiazine. The
clear separation between the labeled triazine (HHTT), dithiazine and MEA peaks shows that the three
compounds can be simultaneously detected and quantified.

Figure 7 - Scavenger with 25 ppm of dithiazine and MEA added, showing capacity for simultaneous
measurement.

All rights reserved 2013. Owlstone Inc.

Detecting H2S Scavengers and their Byproducts


with Lonestar

Figure 8 is a calibration curve for dithiazine, showing detection and quantification down to 25ppm.

Dithiazine Ion Count (A.U.)

3
y = 0.012x + 0.032
R = 0.991

2.5
2
1.5
1
0.5
0
0

50

100

150

200

250

Dithiazine Concentration (ppm)

Figure 8 Calibration curve of dithiazine in water

Conclusion
By detecting and quantifying (HHTT) triazine and MEA in crude oil, and triazine, MEA and dithiazine in
water, Lonestar provides the means to monitor the entire scavenging process and prevent potentially
expensive fouling by amorphous dithiazine.

All rights reserved 2013. Owlstone Inc.

Detecting H2S Scavengers and their Byproducts


with Lonestar

Lonestar Specifications
Technology

Field Asymmetric Ion Mobility Spectrometry (FAIMS)

Analytes

Triazine (CAS: 4719-04-4)


Monoethanolamine (CAS: 141-43-5)
Dithiazine

Dynamic range

3-600ppm

Accuracy

+/- 5% of maximum concentration (+/-30ppm)

Analysis time

<10 min

Calibration frequency

6 months

Features of Lonestar H2S Scavenger Analyzer


Selective detection of H2S scavengers and reaction products in range of crude oils
Quantitation of scavenger between 3 and 600ppm
Ease of operation - No special sample preparation or extraction required, operated by
non-specialist.
Analysis time under 10 minutes

Benefits of Deploying Lonestar H2S Scavenger Analyzer


Petroleum processors can identify impurities on the spot, before custody transfer
Avoid corrosion and fouling in downstream equipment

All rights reserved 2013. Owlstone Inc.