Overview of Contents:
_____________________________________________________________
1 About CALYPSO 2014
1.1 Important Information
1.2 Installation and System Related Notes
1.3 Notes on the Application
1.4 Notes from previous CALYPSO versions

2 Contents and latest changes

3 New Features in CALYPSO


3.1 Handling, Operation and Performance
3.1.1 Enhanced handling concept and touch screen operation
3.1.2 AutoRun - extended functionality
3.1.3 Measurement plan search function for features containing directory paths
3.1.4 Open dialog windows in fixed location
3.1.5 Flexible measurement strategy for 2D-Line
3.1.6 ID-Chip with CALYPSO
3.1.7 Optimized Scanning function for passive Sensors
3.1.8 Definable shortcuts / direct calls
3.1.9 Circular path: circular travel path for single-point measurements
3.1.10 Pattern function, exclude indices by multi-selection
3.2 Display and Evaluation
3.2.1 Fourier analysis for 2D lines
3.2.2 Form & location: Show actual points of a characteristic
3.3 Standards and Mathematics
3.3.1 ISO 14405-1 Symbols
3.4 Options
3.4.1 Curve: Performance enhancement
3.4.2 PCM: Activate and deactivate measuring volume illumination with O-INSPECT

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.4.3 PCM: paths for stylus system change


3.4.4 Universal post processor for CALYPSO Preset
3.5 Other
3.5.1 Stylus system change visible in Planner
3.5.2 Stylus System Creator Standalone installation
3.5.3 Optimized Error Reports

4 Installation Instructions
4.1 Installing the CALYPSO Basic Package
4.2 Installing CALYPSO ServicePacks and Patches
4.3 Data Backup
4.4 Installing the ViScan Drivers
4.5 Installing Software for Metrotom
5 Compatibility
5.1 CMM Validity
5.2 Controllers & Sensors
5.3 PC System
5.4 Software compatibility
5.5 CAD Interfaces
5.6 Compatibility of Measurement Plans and Programs
6 Contact

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

1 About CALYPSO 2014


Please read the following notes before installing this software!

1.1 Important information


Special Geometries - Circle on Cone
5.8.08 07.10.2014
As from CALYPSO version 5.8 onwards the average height in Z was calculated incorrectly if the following
conditions applied: Usage of Special Geometry Circle on Cone in conjunction with a constrained radius
combined with the cone axis orientated in Z. This problem was resolved with version 5.8.08.
True Position Rotate by an Angle
5.8.08 24.09.2014
The calculation of Position characteristics for which the corresponding feature is not parallel to the reference
system and has a cylindrical tolerance zone while utilizing the function "Rotate by an Angle" was corrected
as from CALYPSO 5.6.12 onwards. Comparison of results with evaluations created with previous versions
may show differences.
CALYPSO Basics Operating Instructions changed
5.8.08 24.09.2014
The addition "DIN ISO 1101" has been removed from the coaxiality characteristic description in the "Details
on defining characteristics" chapter of the CALYPSO Basics operating instructions.
Single Point in Freeform Surface Exporting List of Points and Banner Display
5.8.00 14.05.2014
The output of single points from a freeform surface feature with bestfit has been modified. The values for
single point position and single point deviation contained in the export file are affected by this modification. In
the banner display, the values for single point position are affected. These values can deviate from values
displayed in earlier CALYPSO versions. These changes have no effect on the overall bestfit result as the
respective calculation is made independently.
Filter Function and Outlier Elimination
5.8.00 11.02.2014
As from CALYPSO 2014 Version 5.8 the default measurement plan settings for filter and outlier elimination
for Form are no longer effective for Line Profile, Curve Form and Free Form Surface Profile. The filter
function and outlier elimination for these characteristics can now only be activated via the corresponding
evaluation dialog of the respective feature.
When running measurement plans created in previous CALYPSO versions the following circumstances can
cause results to differ in certain cases: No filter was set for the characteristic, additionally a filter was
selected in the default measurement plan settings for filter and outlier elimination for Form as well as
selecting a value below 1/7 mm (0,1429 mm or 0,0056 inch) for the segment step width.
The filter settings must be carried out in the feature dialog directly, should identical results to previous
versions be required.
ViScan - Hardlock Dongle use only with USB 2.0 Port
5.6 01.07.2013
The Hardlock dongle required for ViScan application must not be connected to USB 3.0 ports (usually
marked by blue colour code). Please ensure that the dongle is connected to a conventional USB 2.0 port.
FACS Automation - Individual modification required
5.6 01.07.2013
The FACS Automation Interface is customised individually to suit the unique requirements, therefore
individual modification is required for every single installation. Please contact our Support Team before
running your FACS application with CALYPSO 2014.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

C90 and C98 Controllers - End of Software Support


Software Support has been cancelled for CMMs using the C90 or the C98 controller.

5.6 01.07.2013

1.2 Installation and System Related Notes


CALYPSO 2014 Parallel Installation
5.8 27.02.2014
Before installing CALYPSO 2014 alongside CALYPSO 5.2 or CALYPSO 5.4 please ensure that the following
Service Packs are already installed: CALYPSO 5.2.20 or CALYPSO 5.4.16 respectively. If this is not the
case, the installation of CALYPSO 2014 can cause a licensing conflict and consequently prevent CALYPSO
from starting.
Improved Setup Procedure
5.6 01.07.2013
The CALYPSO installer (for the Basic Package, ServicePacks and Patches) has been revised for Version
2014. Please refer to the installations instructions in chapter 4 for details.
Metrotom Measuring Module
5.6 01.07.2013
Before installing CALYPSO 2014 with the Metrotom Measuring Module you must uninstall any existing
version of CALYPSO! Parallel installation of more than one version of CALYPSO in combination with the
Metrotom Measuring Module is not permissible.
'Preset/Erode' Option with multiple Installation
5.6 01.07.2013
If CALYPSO 2014 is installed parallel to CALYPSO 5.4 the 'Preset/Erode' Option must be updated for 5.4.
This update (5.4.0200) is available in conjunction with the CALYPSO ServicePack 5.4.20 or via download.

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

1.3 Notes on the Application


Blade XML Output
5.8.12 04.12.2014
With previous versions of CALYPSO, the XML output files for Blade were created without taking the full
decimal places into account. This was corrected for CALYPSO 2014, the number of decimal places used is
now based upon the value set in CALYPSO.
O-Inspect: Control Panel Potentiometer
5.8.12 11.11.2014
Caution! The potentiometer on the O-Inspect control panel has no influence on rotary table movement.
RC List Qualification with Linescan
5.8.08 13.11.2014
The RC List Qualification for the sensors Linescan and Linescan II has been enabled with this version.
As a non-standard method of CNC qualification, the preset limits for the Sigma value are monitored during
the List Qualification. Exceeding one of these limits will cause the stylus sytem to be marked as invalid. With
a standard CALYPSO installation the limit for Sigma is set to 0,01 mm. This setting is too low for a LineScan
Sensor and should either be corrected to a higher value or deactivated completely.
PCM Inapplicable commands
The following PCM commands are no longer available and cannot be used as from 5.8.08:

5.8.08 27.08.2014

getTemperatureCorrection().temperatureTableFrontBottom,
getTemperatureCorrection().temperatureTableFrontTop
getTemperatureCorrection().temperatureTableRearBottom
getTemperatureCorrection().temperatureTableRearTop
O-Inspect: Tactile-Optical Adjustment
5.8.04 18.06.2014
The measurement plan for the tactile-optical adjustment has been revised and improved. The adjustment is
now less sensitive to varying lighting conditions. Due to the enhanced stability of the measurement plan the
adjustment procedure has also become more user friendly.
Please use the following measurement plans for the tactile-optical adjustment:
Use the measurement plan tactile_optical_adjustment_and_check_V1.3_C5.4_ViScan, if your camera is
controlled via a Framegrabber board. This applies to machine types O-INSPECT 442 up to August 2013,
that are equipped with a ViScan camera.
In all other cases use the measurement plan tactile_optical_adjustment_and_check_V1.4_C5.4. This
applies to machine types O-INSPECT 322 and machine types O-INSPECT 442 from September 2013
onwards, that are equipped with a Manta camera.
Please note: the tactile-optical adjustment must be carried out regularly in order to guarantee the specified
accuracy
Yellow Stoplight
5.8 16.04.2014
By switching the stoplight to yellow a CNC-run can be paused and subsequently continued by switching the
light back to green. However, this functionality is not always applicable. Switching the stoplight to yellow
should be avoided as soon as the sensor is in probing search mode or already taking points (single points or
scanning).
METROTOM: Neue Funktion "Check gray value"
5.6.04 22.08.2013
The function "Check gray value" has been added to the CT functionality. This function checks the calculated
gray value along with the "automatic parameter definition". Please observe that the function will only operate
correctly if the CT data set is already visualized and that the part consists of single component material only.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Scanning optimization for passive sensors (Retry measurement after "airscanning")


5.6 01.07.2013
If the actual position of a feature differs largely from its nominal position, high speed scanning with passive
sensors (XXT) can lead to the so-called "airscanning" effect due to the loss of contact between the sensor
and the surface of the workpiece. In such a case, for newly created measurement plans, CALYPSO 2014 will
automatically repeat the measurement at half speed, taking the actual position of the feature into
consideration. This behaviour can also be activated for existing measurement plans created with earlier
CALYPSO versions by changing the compatibility settings (pulldown menu -> Resources -> Compatibility
Settings )
RT-AB Rotary Table Application
5.6 01.07.2013
When using the lifting and lowering function for loading the RT-AB rotary table please make sure that
subsequent measurements must be carried out in the same state as used for the definition of the base
alignment. Otherwise the subsequent measurement will not correspond to the base alignment!
Back to overview

1.4 Notes from previous CALYPSO versions


These notes also apply to the current version.

1.4.1 Operation
Device Administration
Please ensure that all CMM connections are closed (by closing traffic light windows) before removing a
device from the 'Measuring-Systems' page under 'Extras', 'Workroom', 'CMM'.
Calculating the alignment
When calculating an alignment, the evaluation setting ISO 5459 is not taken into consideration if the
compatibility mode 'Calculate Alignment as Base Alignment' is set to 'On'.
Selecting CMM System Setup for search path to nominal position
This setting is used for using the respective CMM default value for the probing search path. Previously the
value 5 (with **) was displayed as suggestion when selecting the default setting in the editor. However this
was not effective because of a limiting value prompt; the current CMM default value applied. Without
changing the measurement plan, 'CMM System Setup is now displayed as default setting (suggestion), the
performance does not change. If a numerical value is now given as search path, this value is effective even if
it is smaller than the previous limiting value. Too small a value should be avoided.
Probing Point Distribution in Slot Features
The distribution of points in circular sections of slot features changed from CALYPSO 4.6 onwards. Now the
same logic as for circles is used. Should an evaluation of a slot feature not be possible using an odd number
of probings (e.g. 5 points), an additional probing point should be added in order to have an even number of
probing points (e.g. increase no. of points from 5 to 6 ).
Measurement Plan Compatibility using Multiple-Strategy
Application of the multiple-strategy function in new measurement plans created with CALYPSO 4.4.04 or
newer or opening and saving existing measurement plans with CALYPSO 4.4.04 or newer will prevent the
usability of these plans with older CALYPSO versions!

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Measurement Plan Compatibility 'Sphere' with Circular Sections


Measurement plans created with versions older than CALYPSO 4.4 which contain spheres with circular
sections should be checked for the correct start angle in these features!

1.4.2 Rotary Table Operation


RT-AB Rotary Table Application
5.4.08 06.09.2012
When using the RT-AB, the lowered position must only be used for loading the rotary table and is not
suitable for measuring! Please make sure that the definition of the base alignment and subsequent
measurements are carried out with the RT-AB raised.
Moving an RT-AB Rotary Table on the CMM
5.4.04 04.07.2012
After moving a rotary table of the RT-AB type on a CMM, you are required to redefine the reference point of
the rotary table. Select 'Probe Reference Point of the Rotary Table' from the 'More' menu in the stoplight
window or from the rotary table menu .

RT Position for RT-AB Rotary Table


5.4.04 04.07.2012
When defining the rough position of the RT axis for an RT-AB rotary table in CALYPSO "RT position" the
user is prompted to probe the qualification point. This must be carried out with the rotary table set to 0
degrees. Please take notice of the information regarding the 0 degree position.
"Missing borehole" Function
5.4.04 13.06.2012
The "Missing borehole" function can only be used in conjunction with MT type sensors (e.g. VAST etc.) if the
designated stylus is qualified with a probing force of 100mN or higher. Otherwise the functionality cannot be
ensured. This also applies to the function "Search distance after nominal position".

1.4.3 Stylus Systems, Probes and Qualification


Reset during List Qualification
5.2.12 19.10.2011
A reset during List Qualification can cause the rotating carrier to stop in an undefined postion and the stylus
system name be changed to "not_defined". In this case the prompt to "Probe in stylus direction" must be
cancelled and you are required to select the stylus system manually before restarting the List Qualification.
Otherwise a collison can occur.
Air scanning function with Sensor type XXT TL2
The air scanning function is no longer supported for application with XXT TL2 sensors.

5.2 01.03.2011

Stylus data no longer valid when updating from CALYPSO 4.0 or older
5.0 20.05.2010
Existing stylus and configuration data can no longer be used when updating from older CALYPSO versions.
This concerns CALYPSO versions 4.0 as well as all older versions. If specific data has to be retained, please
contact our Software Hotline.
Stylus System Data Import
5.0 20.05.2010
Styli data exported with version 4.4.04.xx from systems with Zeiss or Renishaw index head (e.g. RDS, DSE,
MIH, PH9 and PH10) may not be imported into newer versions! The useability of imported styli when used on
a CMM is different than usage in simulation mode (without CMM). For use on a CMM re-qualification is
required, in simulation the styli can be used without re-qualifying.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Shaft Radius of the Reference Sphere


5.0 20.05.2010
The Shaft Radius of the Reference Sphere on the 'Reference sphere management' page must be set to a
value large enough to cover all of the stylus diameters in use:

In order to prevent collision during probe qualification the 'Shaft Radius' setting must be
increased to a value that sufficiently takes into consideration the largest stylus radius in
use (see D2 for the red sphere illustrated).
However, increasing the radius will also reduce the amount of possible angular positions
for qualification (RDS,DSE etc.). A workaround to this situation is achieved by saving
multiple copies of one and the same reference sphere with a varying shaft radius and
selecting the sphere according to the requirements for probe qualification.

RCCAA
5.0 20.05.2010
RCCAA qualified stylus systems must not and need not be re-qualified using the list qualification method.
Sphere Coverage Angle for qualification of passive sensors
5.0 20.05.2010
The Sphere Coverage angle for a full qualification of passive sensors (XXT, SP25) is limited to a minimum of
180, values lower than 180 are ignored. In order to qualify smaller styli a reference sphere with a smaller
diameter (e.g. a stylus of which the radius is known) is required. Following a full qualification it is possible to
re-qualify a stylus using the regular reference sphere. For this a lower sphere coverage angle may be
selected and a 6-point qualification is performed.
Interruption of VAST-XXT CNC-Qualification
4.10 17.11.2009
Interrupting a CNC-qualification of a VAST-XXT, will result in undefined probe data! In this case the stylus
must be re-qualified manually completely before using it for measurement or re-running the CNCqualification. Please use 'Qualify passive stylus, the selection 'Geometry Re-qualification is not sufficient.
Special characters in Stylus System Names
4.10 17.11.2009
Special characters must not be used in stylus names and stylus system names! Under no circumstances
may apostrophes (' ') or quotation marks (" ") be used in names for styli or stylus systems nor may these
names begin with a special character (e.g. ) or end with a dot! Underscores ('_') may be used as
separators.
Automatic configuration of temperature sensors
You have to reconnect CALYPSO with the controller after adding or removing a temperature sensor, if
automatic configuration is activated in the C99 controller.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Distance to Proberack
The preset default value for the distance to the proberack during stylus system changing of XXT and SP25
(for empty probe) was increased from 0 mm to 70 mm with CALYPSO 4.4. All holders which are created new
for XXT and SP25 are now preassigned with 70 mm. This can be adapted subsequently for each holder by
the Master user.
Side Styli and List Qualification
With the RDSCAA or DSECAA option, side styli can now also be qualified semi-automatically (except for
XXT sensors!). Please note: List qualification cannot be applied for stylus systems with side styli! However
as a result, on CMMs with articulating probes after a software update to CALYPSO 4.4.04 or newer, all the
styli have to be subsequently completely re-qualified! If you are using the RDSCAA or DSECAA option, it is
imperative that the MasterProbe can be used for the qualification of the fitting position. Passive sensors such
as XXT and SP25 cannot be used for this. To facilitate this, the List Qualification can be used (not for stylus
systems with side styli! ). The procedure is described on the following pages.

1. Open the Probing system qualification window

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

2. By clicking on the
icon, the dialog for setting
the RDS axis is opened.

3. Select the Styli list tab:

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

4. By clicking on the
icon, the message displayed
on the right will appear.
Select Overwrite and click
OK.
5. Close the window for setting the RDS axis with 'Close' and in the ''Probe
system qualification' window start the List Qualification using the caliper icon.
6. In the following dialog, please select List Qualification:

You are prompted to set the index holder to the angle A 0 B 0. You are then prompted to make a probing at
this angle position. This process also takes place if you have not defined a stylus in the angle position 0.
Using this position, the lengths of all the other styli in this stylus system are calculated and automatically
qualified.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

1.4.4 Results and Presentation


Overall Result in Printout
The characteristics In CALYPSO were re-catagorized in version 5.4. Text elements, coordinate systems and
deviations are listed separately. The Overall Result in the printout has changed accordingly:
New from CALYPSO 5.4:
Overall Result
All characteristics:
in tolerance
Out of tolerance
Over Warning Limit
Not Calculated
Number of alignments:
not calculated
Number of text elements:

Previous display :

170
148
21
1
0
3
0
2

Overall Result
All characteristics:
Out of tolerance
Over Warning Limit
Not Calculated

175
20
1
0

1.4.5 Optical Systems


GigE Camera Installation
5.4.12 07.07.2012
The application of a GigE camera (OI 322) requires changes to be made to the firewall on the CALYPSO
computer. Please refer to the enclosed information sheet or the details listed in the document 6000614107.010_Konfiguration_Firewall_Kurzversion.pdf in the Tols\GigE Sensor Tools folder of the DVD.
ViScan Mirror function
The mirror function in CALYPSO cannot be applied to optical plans.

5.4 26.04.2012

ViScan Slot holes


5.4 26.04.2012
Rounded or rectangular slot holes can only be measured in CNC-mode by using several circle or line
segments.
O-Inspect with Rotary Table
5.2 09.05.2011
While creating a measurement plan with an O-Inspect equipped with a rotary table (RT), the RT should be
set to 0, even if the RT functionality is not utilized. This can be achieved by calling RT-Reference (the RT
axis must be valid) or temporarily deactivating the RT altogether:
Pulldown menu > Extras > Settings > CMM > Rotary Table > Activate Rotary Table Functionality to OFF
As a default the RT is deactivated when a new plan is opened (by 'RT activated for this plan in the rotary
table settings page), however, by switching between optical and tactile probing, the internal incorporation of
the RT angle is activated and can lead to incorrect rotated values for features measured.
ViScan Scanning unknown Contour
5.0 20.05.2010
With version 5.2 the scanning direction of the optical scanning of unknown contours has been changed. If
necessary existing measurement plans must be adjusted.
ViScan Important Note for all users of F25, O-Inspect and CMMs equipped with ViScan 5.0 20.05.2010
Please take note of the following information regarding existing optical probe data (F25, O-Inspect or ViScan
camera data) created with CALYPSO 4.7.02 or older.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Any existing optical probe data originating from CALYPSO 4.7.02 or older must be deleted before using
Release 5.8. The required camera data must be re-qualified with Release 2014. Trying to switch to older
camera data with CALYPSO 2014 will be denied and can result in output of various error messages.
ViScan Upgrade
4.10 17.11.2009
After upgrading from older versions to CALYPSO 4.10 or higher the ViScan sensor must be re-qualified
using the chrome template and the reference sphere after installing! LN_653.
ViScan Manual CNC Mode
4.10
After focussing during manual CNC-runs the optical probing method must be manually re-selected.

17.11.2009

ViScan - Grafics Adapters


4.10 17.11.2009
The Grafics adapters ATI FireGL 8800 and NVIDIA Quadro FX1100 are not compatible with CALYPSO in
combination with the ViScan sensor as from version 5.0. The camera image is not displayed correctly in the
graphics window.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

1.4.6 Windows System and Software


Ghostscript 64Bit incompatible
5.4 19.03.2012
The 64Bit Version of GhostScript does not work with CALYPSO. If GhostScript is needed, please use the
32Bit Version.
ViScan Driver Installation and Firewall
5.4 25.01.2012
An activated firewall can prevent the ViScan dongle driver from being installed correctly. This results in the
error message 'Security Device not found' while trying to load a ViScan image in CALYPSO. We
recommend deactivating the firewall before starting the 'ViScanDrivers' installation.
Microsoft SQL Server required
5.2 05.05.2011
For running the application the Microsoft SQL Server is needed. The setup routine checks the prerequisites
of your operating system.

Change to
German
language

1. Follow the instructions step


by step.

2. SQL Server check dialog


will be shown.
Press Button for
detailed
information

3. Press OK.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Press Button for


detailed
information

4. Press OK.

In case of failure, please contact your IT administrator in order to enable the required
prerequisites.
Locally installed safety systems (firewall and/or intrusion prevention systems) can prevent access to the SQL
server which is installed automatically as a prerequisite of CALYPSO. In such a case the setup will be
cancelled. To adjust the configuration of your locally used firewall system please contact your IT department.
The communication between the Microsoft Installer (msiexec.exe) and the SQL server (sqlserver.exe)
processes must be assured.
The setup of the SQL server contains the package 'MSXML 6. On some systems the package 'MSXML 6
Service Pack 2' is already installed. This package can be incompatible with the package 'MSXML 6'. If the
SQL server setup is aborted during the installation of the 'MSXML 6' package 'MSXML 6 Service Pack 2'
must be uninstalled first.
.pdf Output Merge Function
GhostScript is required for merging .pdf results, version 8.56 or newer is required for error free application.
ViScan - Windows 7 Compatibility
5.0 20.05.2010
The ViScan Sensor can only be used with Windows 7 in combination with a suitable Framegrabber card.
Please refer to chapter 4.4.
64bit Windows System restrictions
Operation via HP-IB interface is only permissible with 32bit Windows systems!

5.0 20.05.2010

Windows 7 - Aero Mode


5.0 20.05.2010
Use of the Aero Mode with installations on Windows 7 systems may cause malfunction of the landscape view
of the graphics window in the custom printout. This can be corrected temporarily by switching off the Aero
Mode. Teleservice/Netviewer application can also be impaired by having Aero Mode active.
Windows 7 - Setting Properties for Network Cards and USB Hubs
In Windows 7, the NIC Power Management must be set so that the operating system cannot switch off
network cards or USB Hubs. To do this the following setting must be made in the Device Manager for the
network cards and USB-Hubs in use:

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

The 'Allow the computer to turn off this device to save power' box must not be checked.
Recommendation: In addition, set the power mode to 'Never' and the performance to 'High Performance' in
the system power management.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Windows 7 - Warning Message during Installation


After installing CALYPSO (successfully) on a Windows 7 system the following warning message might pop
up:

The message is confusing and does not relate to an unsuccessful install. It is caused by the outdated install
assistant tool. A newer version of the install assistant tool is not available.
The warning can be suppressed by executing the install in compatibility mode:

By right clicking the .exe select 'Properties' and set the compatibility mode for Windows XP (ServicePack 2).
Activate by clicking OK:

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Windows will 'remember' this setting and further installs will be carried out without the warning message.
Windows 7/64-Bit - Graphics drivers for NVIDIA FX1800
A modification must be made to the driver 266.45 when in use with the graphics adapter NV FX1800 on
Windows 7/64-Bit systems. The 3D setting must be changed to 'Autodesk Motion Builder compatible' to
prevent slow reaction when rotating or moving CAD models.
Windows 7 - Regional Settings
Should your CAD view display incomplete or inaccurate circles:

Open Windows 'Regions and Languages' in 'System control' and select 'Further settings', then replace the
decimal comma by a decimal point or vice versa.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Windows System Control Show window contents while dragging


In order to optimize the screen settings for CALYPSO the setting 'Show window contents while dragging'
should be deactivated. Adjust the setting in Advanced system settings/Visual effects:

Manipulation of Installed Files and Folders


Installed CALYPSO files must not be manipulated! By renaming or removing files that are installed in the
CALYPSO folder (including installed user data!), you can cause malfunction of your application. By doing so
the Windows Installer autorepair function might be launched unsuccessfully. Inserting program files manually
(via Windows Explorer) can also cause faulty operation.

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

2 Contents and latest changes

Contents:

CALYPSO 2014 - ServicePack 3 - Version 5.8.1200

Optional Updates:

Calypso2014CT32_580802Setup.exe - METROTOM module 32Bit


Calypso2014CT64_580802Setup.exe - METROTOM module 64Bit
StylusSystemCreatorSetup.exe - Version 1.7.5218
ZEISSBasicReportingSetup5027.exe
LineScan 1.20.7
CFS 1.19.3

Default Languages:

German, English

Additional Languages:

Instruction manuals:

German, English

Latest changes:
5.8.1200

4274
4462
6252
6914
10062
13380
13852
13922
14916
14985
15476
16350
16790
16812
16999
17368
17737

07.01.2015

Release Servicepack 3

SPR 6327: "Execute now" function default changed to automatic. Retry uses setting from previous
run.
Actual points now removed from point sets when deleted manually
Output of reference names in custom printout completed
Paste contents with Distance characteristics corrected
PDF output corrected
Exported probe data includes manual path
Coordinate system display corrected for the 3d cylinder plot
Bugfix: Error SDO 1015 when qualifying an XXT in SENSITIVE mode
Nominal Radius value for stylus system qualification corrected
SPR 6552: Inch/metric ssue with old formplot corrected
CNC output of Graphic Element containing more than 17 results enabled
Cut off stylus system dialog corrected
Left hand spiral ring evaluation corrected
Cylindricity plot display corrected
Bugfix: Simulated CNC stylus system change for XXT
Bugfix F25: "Not Defined" message corrected for MasterProbe qualification
Bugfix: Nominals corrected when creating Features

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

17777
18023
18168
18257
18319
18353
18355
18369
18496
18518
18688
18743
18769
18784
18789
18822
18846
18920
19037
19055
19082
19085
19223
19275
19446
19476
19497
19500

SPR 6590: Report of K2005 characteristic class from userattributes.ini in the custom printout
Generated Freeform Surface corrected
Editing the name of a base system enabled
CAD model of Rotary Table for Duramax added
"Angle between Features" used in a loop corrected
Improved CAD Filter "Point"
Improved CAD Illumination for colour display of deviations
Measurement plan start is now independent from logged in user
2-Point diameter mesasured optically corrected
SPR 6602: Circle Feature "Projection" Button, improved performance
Bugfix: Message not understood" upon CNC start
Creating minimum feature from points corrected
Path for ACR1 installation corrected
Point export in CNC for feature without characteristic corrected
Plot output in CNC mode corrected
Re-opening a measurement plan with a specific printer assigned enabled
Evaluation corrected for curve form measurement
Temporary file scratch.scr renamed to avoid .scr extension
Single point measurement of circle segment in CNC mode corrected
Incompatibilty issue with Microsoft Lync 2013 solved
Nominal value for inclination of cylinder from CAD model corrected
CAD graphics presentation enabled with 5m distance to base alignment
Black spots removed from CAD presentation as .pdf
Numeric decimal comma activated for Dutch user interface
Black points in Custom Printout with "Output to PDF..." corrected
Display Parasolid CAD model enabled
Bugfix: Measurement plan start in Autorun
Imprved CAD model handling in combination with Autorun

19516
19517
19521
19527
19583
19626
19902
17220
17759

RDS XXT Qualification corrected


Bugfix: CAD Evaluation in new result presentation
Bugfix: CAD evaluation in new result presentation
Bugfix PCM: wrong parameter assignment in case of para2 =-para1
Symmetry: Arrow direction corrected in projected tolerance zone
Bugfix: Macro update disrupted start alignment
SPR 6624: faulty 2D-line best fit corrected
Optimized collision examination in simulation
Optimized collision examination in simulation

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

18714
18789
19164
19975
19980

Bugfix: Output of cam stroke characteristic to .txt file


Plot output in CNC mode corrected
Stylus rotation corrected for CNC run in Simulation mode
Bugfix Curve: Selection of line-form actuals in CAD
DMIS-Import V2: Expand all nodes corrected

5.8.0800

2633
5276
6232
6835
8376
8643
9005
9625
9959
10059
10069
10214
10631
10893
11127
11162
11497
11517
12246
12280
12455
12558
12650
12781
12908
12912
12924
12974
12994
13039
13371

10.10.2014 - Release Servicepack 2

Farbklassen lassen sich nicht umstellen


Kegel mit Zwischenpositionen - KMG fhrt im Eilgang die Zw.-Pos. an
PH10T has the CAD model of the PH10M in the Probe Simulation
Messung mit Teilung und SG
The setting of scale limits on the Color Classes is not saved
CAD Ansicht in diesem Prfplan stimmt nicht
More than 16 straightness plots in a graphics element --> in cnc only the last page will be printed.
Color gradient display is unworking with 'recall feature points'.
Simulation: Keine Modelldaten fr MMZ-G 30 60 20 verfgbar
Missing CMM-CAD-Models
Werden exportierte Punkte einer Ebene in eine neue eingelesen, so ist die Orientierung verdreht.
'Export DMIS' calculates abnormal coordinate values.
Scan limit and default setting for strategy are not working correkt
Stylus tip is not positioned onto workpiece when probing
SPR 6501: XTR in Simulation does not work properly
RT rotates with "Base Alignment axis", the coordinate values in traffic light is not correct
Dateipfad von EXCEL-Report mu nach Update auf 5.6 in allen PP manuell angepasst werden
Kurvenmessug groe Abweichung
Problem whit a Rotary Table in Simulation.
Exception beim Messen der Freiformflche "Surface Boss Br4 palp z-"
Beim Exportieren eines Prfplans im DMIS-Format werden SAVE Anweisungen erzeugt, die scheinbar nicht
ganz ideal sind.
A MICURA can not be shown in simulation. There is no sab-file for MICURA in the cad-components directory.
Individuelle Einstellungen der Farbklassen werden nicht bernommen
Nach Ablauf des PP strtzt Acis ab.
Optik: Lichteinstellungen mit Dezimalzahl: Eingabe u. Messung man. OK; wird aber im CNC-Abl gerundet
unter Vorbereiter /Speichern / Laden - >> Strategie bei Kreis u Zyl. Einstellungen konfiguren
Sporadischer Fehler bei Lochbildmessung: Zu viele Iterationen bei Elementberechnung
Alignment doesn't appear in the selection list of the coordinate system on the Stop light window
Kunde mchte CAD-Modell fr seine Maschine haben: Accura II 12/24/08
Simulation: CAD-Modell von MMZ-G 20/50/20 ist nicht verfgbar
XTR in der Simulation geht immer noch nicht

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

13571 nderungen im Farbverlaufsdiagramm werden nicht abgespeichert.


Wenn in einem Rechteck in der Strategie eine Rechteckbahn verwendet wurde, so kommt im MAN-Modus
13715 eine Fehlermeldung
13774 The language setting of Printout don't apply to Blade-interface.
13830 Profile Element_System Error occurs after selecting "Evaluation" tab.
13837 Die Messelemente der vorhergehenden CAD-Prsentationen werden nicht gelscht.
13922 Many customers have an error SDO 1015 when qualifying an XXT in SENSITIVE mode
14034 Beim Exportieren eines Prfplans in DMIS kommt eine Fehlermeldung
14550 2d-Line with a point set in the strategy does not work in simulation
14617 Nachtrgliche Auswertung
14691 Probing speed becomes very slow , once it is stopped by ESC or Red light button
14754 Measurement pla with macros from Calypso 4.4.06 does not work in Calypso 5.2.24
CALYPSO-Ampel whrend des CNC-Ablaufes: der Knopf "Ablauf beenden mit Datensicherung" funktioniert
14767 nicht
14895 SPR 6550: Scan near the boundary of the CAD face: custom printout displayed an error in the profile.
wird eine Kurve in einem Koordinatensystem gescannt, wird je nach Scannrichtung die Tasterradiuskorrektur
14980 in die falsche Richtung gemacht.
14991 Bei einer Ebene mit 4 Polylines fhrt Calypso Kollision seit 5.6.08
wird eine Kurve mit mehreren Segmenten erzeugt, und dies ber Punkterckruf nochmal erstellt, lassen sich
15015 die Verbindungslinien beim Grafikelement nicht ausblenden
15068 in Calypso 5.4.20.06 oder .08 funktioniert im angegebenen Prfplan das Einlesen von Sollpunkten im CNC nicht.
15130 Bei einem Programm werden die Calypsocurven nicht mehr berechnet.
SPR 6553: If program is in inch, go to Feature Settings Editor->Geometry [Nominal Data] - changing X, Y, or Z
15196 will immediately multiply the typed number by 25.4
15198 Probleme bei Form und Lage Toleranzen einlesen mit UG
Wenn man bei CAD Ladeeinstellungen "Kein Modell" anwhlt, einen Prfplan ffnet und speichert, ist das
15273 "model.sab" zerstrt
15328 CAD Kommentar
15360 Basic Reporter: True Position with additional position result gives a tolerance for the coordinates.
SPR 6559: Helical path with a negative helix shows a wrong direction, but only in CAD representation, the
15366 move on the machine is OK
15393 Bei der Funktion "Prfplan spiegeln" ist die Ikone des Fahrzeugkoordinatensystems falsch dargestellt.
wird die Schrittweite eines Elements parametrisiert, und der Paramater gendert, wird dies im CNC Ablauf
15428 nicht aktualisiert
15476 More than 17 straightnesses in a graphic element: in CNC-mode only the second page is printed.
15576 The number of meas datas is different when points exported manual or in CNC
after the "traffic light" changes to red and after it is switched back to "green" the probing is became extremely
15613 slow
OI luft in Serienmessung mehrere Stunden (3-4 Stunden) ohne Probleme und pltzlich kommt folgende
15636 Fehlermeldung (viscan.jpg). Fokuspunkt konnte nicht fokusiert werden.
15640 Defect of circle path for sphere when there are single points and 3 or more circle paths.
15682 CAD view from Calypso 4.8 does not work correct in Calypso 5.6.08
15697 Fehler bei nachtrglicher Auswertung: werden vorher Istwerte gelscht, so wird beim Ablauf das Tastersystem

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

15739
15740
15789
15792
15798
15912
15915
16010
16025
16035
16085
16167
16172
16203

16264
16266
16284

16304
16319
16337
16381
16414
16419
16475
16500
16502
16504
16508
16524

berprft.
if he set a default strategy of 400 degrees for the measurement of a circle and then he creates a new circle, the
strategy is still fixed to 360 degrees.
Texteingabe wird Position nicht beibehalten
Zeichnungsnummer im Protokolkopf wird ins Balde static bertragen es sollte dynamic sein
Im Messelement Kurve werden beim ffnen der Parameterdateien evtl. vorhandene Punkterckrufe
abgewhlt
In a mirrored measurement plan we get a collision from safety +X to -X
Manuelle Ausrichtung kann in der CNC-Startmaske nicht mehr angewhlt werden, wenn vorher das
Startsystem deaktiviert wurde.
Exception bei Initialisierung
beim einstellen des Farbverlaufs fr die CAD-Auswertung werden die nderungen in den Farbklassen Farbverlauf nicht bernommen
Wird das Datum "kurz" verwendet und im Autorun das Protokoll vom letzten Ablauf geffnet, so steht das
falsche (aktuelle) Datum drin.
AutoRun: Inspection Plan Looping Problem When Activating Mini-Plan
Funktion Elementerkennung geht nicht
Error code -102
SPR 6569: Default setting for Strategy: angle range and and start angle for circle does not take effect
Prfmerkmal EINFACHE DISTANZ: Eingestellte Warngrenze wird falsch ins Tabellenfile bernommen (steht
immer auf 100%).
Punkte exportieren von Kurve - beim exportieren der Kurve ber rechte Maustaste wird eine leere DXF-Datei
erzeugt und die IST-Werte der Kurve gelscht, wenn die Kurve aber vorher einmalig geffnet wird funktioniert
es.
Sicherheitsgruppen
Calypso Duplex - Wird ein automatischer Tasterwechsel ausgefhrt, dann kommt die Meldung "die
anzufahrende Sicherheitsebene kann nicht angefahren werden"
Problem with Cone Addition. After Loading Measured Points, Calypso characteristic list and reports do not
update any characteristic that uses the radius or diameter of Cone Addition, including Radius, Diameter, and
True Position with MMC
Catia V R23
Frageliste, ob Elemente in das neue Basissystem transformiert werden sollen, ist ohne Scrollbalken.
Cad evaluation banner issue with curve form, specifically with 5 decimal places whether inch or metric.
Exception error occurs while moving stylus with a mouse in CAD view
Eintrge in forbiddenChars.txt ohne Wirkung
wird bei CAD Darstellung ein Kommentar eingegeben stimmt er eingegebene Text mit der Acis Ansicht nicht
berein
Measurement volume looks shifted in CAD view when "Drag Base Alignment" is activated
CAD model is not displayed in a plot window.
SPR 6580: Circle for an outside diameter switch to a cylinder will change to an inside diameter
F25: nach Update von Calypso 5.4 auf Calypso 5.8 kommt eine Meldung, dass der Messkopf nicht erkannt
wird: ZEISSdirekt
Acisabsturz

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

16569
16591
16604
16605
16666
16716
16763
16812
16813
16848
16850
16878
16897
16914
16957
16959
16974
17133
17195
17222
17382
17388
17415
17449
17454
17464
17475
17513
17569
17589
17603
17627
17694
17700
17716

Erweiterte Positionsausgabe ( ein ): bei der Achsabweichung wird im FlexReporting eine Toleranz ausgegeben
ein ber PCM erzeugter Offsetwert wird im CNC Ablauf nicht aktualisiert
Dongle-Lizenz wird im License Utility angezeigt, aber bei Calypso-Start kommt Fehlermeldung
Vorbelegung fr neue Elemente
Nach Installation von 5.6.12 oder 5.8 auf manuellen KMGs werden im Prfplan bei Antastungen keine
Messelemente angelegt
In CALYPSO 2014 kommt eine EXCEPTION wenn im Prfplan das Rohrmessprogramm enthalten ist und der
Prfplaneditor geffnet wird.
No CAD-model after update from 4.6 to 5.6
Wird beim neuen Zylinderplot die Mantellinien dargestellt, liegen diese nicht auf dem Zylindermantel. Beim
Innenzylinder liegen diese zu weit innen, beim Auenzylinder liegen diese zu weit auen.
Der neue Zylinderplot kann gedreht werden, auch um die eigene Achse. Dieser Drehpunkt liegt aber nicht
wirklich auf der Achse
CALYPSO 2014: CT - Computertomography: If you go into the Base Alignment and select teh "Approx.
Alignment" than ACIS crashed down.
No CAD-model after update from 4.6 to 5.6.1203
Error after opening plan only with metrotom dynamic device, no problems with STL
Show Silhouettes dosnt work coorect
Problem with a symmetry point in secundary alignment
Probing points in the start alignment are not on the right position.
Japanese license name IMT_CMM-OS_J is not valid in calypso.im
Nothing is displayed on banners in Japanese mode.
XTR funktionniert nicht in der Simulation
Measurement plan from Calypso 5.4 does make a collision at a slot in Calypso 5.6.12
Simulation with DuraMax in Simulation-collision
Sporadic errormessage "too much iterations, feature cannnot be calculated"
Exception after gear measurement gear pro hob protocol
The probing of the features for the basesystem is OK, but the calculation of the features seems to be wrong.
Fehler bei getActual von Torus radius2 und diameter2 wenn nach Inch umgestellt ist
Since updating to V5.6 the 3D Polar characteristic is reporting an invalid result, the same program measuring
in V5.4 reported correctly.
A measurment plan from Calypso 5.4 does not calculate the base system in Calypso 5.6.12
In the inspections many circles didnt display picture in the CAD window
True Position issue
Measurement plan in INCH mode: Feature Settings Editor -> if you typ in Retract Distance it is wrong
SPR 6588: Open UG file with form and location gives an exception
VAST XT gold probe head is not located at the bottom of the ram of MMZ-T in Simulation
Error on circle on cone. problem occured when we have upgraded to Caypso 5.6.1200
Grafics Element will not be printed if graphics element contains a group.
When I'm loged in not as Master but as normal User in Calypso and I click on Planner -> Save Work Area there
will be an Exception
Differences in Best Fit Curves when using the same points.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

SPR 6590 Ford would like to report the K2005 characteristic class from their userattributes.ini in the custom
17777 printout.
17879 Exception when trying to save an error-report
17887 Accura II 16-24-15 model
CAD, Create feature, Freeform Surface, generated from file: points in freeform surface are not identical with
18023 imported point list.
18031 Save load defaults for strategy doesnt work
18044 lineform does not calculate correctly when the datum is a cylinder
18265 new cmm-connection with temperature probe in probe head lead to "not_defined"
18319 Angle between Features gives sometimes a wrong result, if used in a loop
5.8.0400

5844
7440
7886
10483
11698
11848
11899
12924
13071
13100
13186
13378
13929
14084
14101
14229
14429
14497
14764
14777
14931
15319
15325
15464
15472
15479
15483
15496

14.07.2014 - Release Servicepack 1

Curve: Nominal display corrected


FACS: output to QDB-System enabled
VDA CAD model removed after delete
Bugfix : Contents of curve selection column
Bugfix : CMM connection
Simulation: 4 point circle with RT scanning probe path corrected
Bugfix: Saving multiple printout
Iteration for bore pattern measurement optimized
F9 Start corrected
Measurement of outer circles and outer cylinders corrected
Vector was displayed after feature was deleted
Bugfix: CNC Run in Autorun corrected
ViScan: Autofocus point optimized
Points export corrected for freeform
Bugfix : CMM-OS connection to CMM
O-Inspect: PCM control for illumination of measuring volume
Bugfix : DMIS Export
Collision during final feature resolved
Corrected Scanning velocity following update
Icon colour corrected in curve result
Bugfix parallel curve
Stylus name characters corrected
RT Axis
Bugfix: Creation of 3D-curve from one section
ViScan: Optical point in CNC corrected
RPS alignment improved
Interruption of CNC run corrected
Bugfix: CNC run

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

15500
15505
15518
15534
15545
15555
15558
15661
15693
15743
15981
16001
16014
16015
16031
16035
16069
16167
16218
16281

Results display corrected


Calculation of consecutive MP commands corrected
Corrected display of pallet position in Autorun
Green check button in characteristic corrected
Icons disolay of large buttons corrected
Multiple selection of deviation banners corrected
SPR 6564: Bugfix, changed nominals in RPS alignment
Simulation: Curve with rotary table corrected
Subsequent evaluation corrected
Enhanced Dynamic Tensor qualification
Bugfix: workroom CMM
Bugfix: CNC Run
Bugfix: craetion of plane feature
Text output corrected
Defective point in 3d curve corrected
AutoRun: loop enabled
Simulation: optimized single point probing path display
Bugfix: Error code -102
SPR 6571: DME stylus qualification corrected
Bugfix : PC-DMIS Import

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3 CALYPSO New Features


3.1 Handling, Operation and Performance
3.1.1 Enhanced handling concept and touch screen operation
Completely adaptable user interface and support for numerous touch commands to work without a mouse.

Configurable menu functions Individual menu entries or entire menu trees can be hidden.
These special configurations can be saved in a file and imported from there again. To return to the original
CALYPSO settings, the Suggest button has to be pushed.

User interface switchable between drop-down interface or ribbon interface.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

New one-click log. The CAD view is generated with a single click as a portrait or landscape log depending
on the page ratio.
.

The element tree and the presentation log can be shown or hidden by sliding to the left or right.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Users can navigate through the element tree and the presentation log by sliding up or down.

Interactive presentation log. Characteristics in the log can be selected directly. The results are immediately
shown in the CAD display. Double-clicking in the log opens the characteristic for editing. The presentation
log updates immediately during the measurement.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

New integrated power feature


The power feature concept allows you to
integrate your
own or user-defined
characteristics and analyses into CALYPSO.
The toolboxes can be docked.

Manual measurement wizards


When measuring manually, only the selected
geometry is taken into consideration by feature
recognition, i.e. only circle, cylinder or planes.
Element assignment is used when the measuring
points in a generated measurement plan are
assigned to the respective features. If no
assignment is found, a new feture is generated.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.1.2 AutoRun - extended functionality


New functions for the automation interface simplify operation and enhance visualization.
Mandatory output of measuring run information when CNC starts in AutoRun. The system configurator
of AutoRun is able to configure measurement plan or pallet icons. It can specify that a measurement plan, a
document, PDF, image, etc. opens when CNC starts. The CNC run is delayed until the user has seen the
document and continues or discontinues the run.
The document is compiled by the customer and usually contains information about clamping, strategies on
how the base system will be calibrated, required styli, layout for stylus racks, etc. Machine operators
therefore control if the workpiece is correctly clamped, if the right stylus is in the rack, etc.

Retroactively open compact log. In the heat of the battle, it is possible for a user to close a log before
looking at it. This usually results in a loss of actual values in AutoRun. In the context menu of an icon, the
user can retroactively open the compact log again (but only from the last run).
Save CALYPSO session file in CNC (and also in AutoRun) Until now, it was only possible to manually
save session files of special characteristic point quantities. Session files are special lists with nominal and
actual points that are particularly used by CT or line scan applications. The session files can be imported via
CALYPSO or ACIS Viewer.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.1.3 Measurement plan search function for features containing directory paths
Additional search function enables simple search for all features containing directory paths.
You can search for features throughout the measurement plan, which contain the directory path
Search for Paths in the global PCM parameters (under CNC, measurement plan function, parameters)
Paths, PCM file commands and PCM system commands in the input/output settings of the measurement
plan, features, characteristics and travel paths. Paths in names for output files. Paths in features such as
point recall from file. Paths in characteristics such as tolerance curve from the file for waviness. Special PCM
files such as inspection_start_pcm_txt. The icons in the results window appear as they would for a search for
parameters/formulas. Double-clicking a characteristic or feature marks it in the tree view. There is also a
replace function.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.1.4 Open dialog windows in fixed location


The user-friendliness of CALYPSO will improve by opening windows at their last known position and support
by adapting the windows to the individual needs of the user and opening them at their last position.
If the Open at last position setting (previously Open in defined position) is activated, all windows that
previously opened with the standard setting now open at their last position. This does not include standard
windows such as warning, info and inquiry windows.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.1.5 Flexible measurement strategy for 2D-Line


New line measurement strategy for 2D line and measurement strategy for surface line for cones.
Interrupted lines and surfaces can now be measured with a single 2D line feature.
Line measurement strategy for 2D lines
Operators have the option of moving, extending or
shortening a line via two parameters that they can enter in
the user interface. Furthermore, they have the possibility of
inverting the measuring direction by entering a negative
value for the line. These two parameters (entry fields) can
also be parameterized via PCM.
Note: It is not possible to freely define the start and end
points. The points must be located on the geometry (2D
line).

Line measurement strategy for surface lines


The functionality of Surface line for cones corresponds to
that of Surface line for cylinders. It is possible to scan the
surface line or measure with single points. Furthermore, it is
also possible to enter additional parameters such as
scanning speed, step width, number of points, etc.
Note: Unlike Surface line for cylinders, it is not possible to
define an analysis range (analysis) for Surface line for
cones.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.1.6 ID-Chip with CALYPSO


Users of all CMMs with ID wiring and programmed sensors (VAST, XXT...) will benefit from the automatic
recognition of styli via ID chip. Better collision protection through recognition of mix-ups and easier operation
are the results.

As CALYPSO learns the ID chips of the stylus systems used, it enables the clear recognition of the
stylus system

Recognition of the stylus system when (re)starting and when manually changing the stylus

Simplified work following an error or collision during stylus change-out

Fast recognition of mistakes with adapter plates

A continuation of a CNC run may also be possible when a rack is improperly loaded as a result of
software-guided user actions.

The behavior of the software can be modified through a few mode to the intended purpose and the
existing hardware configuration (all sensors detectable/ mixed operation)

Wizard for (fully automatic) learning or review of a full rack with a single click

Display of the IDs in the stylus data management system, the individual IDs of these chain(s) are
printed on the sensors/ plates (exception: XXT plates)

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

If a wrongly used sensor can be identified with absolute certainty, the name of the recognized stylus system
is shown, otherwise only the ID chain of the plate is shown.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

If a component or stylus system is relearned for multi-level sensors (e.g. XXT, TP200)all previously learned
stylus systems that use these components are automatically updated. This simplifies change-out, e.g. for a
single XXT

Learning, as well as any errors and warnings are shown in the results window.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.1.7 Optimized Scanning function for passive Sensors


Automatically adapted repeat measurements with air scanning

When scanning with passive sensors (XXT), the stylus can partially lift from the material (air
scanning) if the location error of the feature exceeds the nominal value by too much and at high
travel
speeds,
meaning
that
not
all
areas
are
measured
as
intended.
Until now, this was only shown in the Run notes window and in the log.
With air scanning (at least 1 point without contact to the workpiece), a repeat measurement of the
affected
feature
occurs
automatically.
(Scanning with specification of the measured actual location and at half scanning speed).

Activation:
To avoid unintentionally changing the behavior of existing measurement plans through a new version, the
improved behavior must be intentionally activated by the user via the Scanning optimization for passive
sensors compatibility mode (drop-down menu -> setup -> Compatibility settings...). For new measurement
plans (from CALYPSO 5.6), the enhanced behavior is automatically activated.
Sphere of action:
Activation is based on the measurement plan. A repeat measurement occurs when the feature contains at
least one circular or helical scanning path and at least 1 point without contact to the workpiece was
recognized. The entire feature is then re-measured.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.1.8 Definable shortcuts / direct calls


Fast access to functions through assignable function keys lead to vastly improved operation.

Access

Drop-down menu Setup Access functions


Access functions with numbers (F12) and shortcuts
The functions of the drop-down window can be accessed with assignable shortcuts (Alt+ or Shift-Alt+) and
function numbers, (particularly for access to control panels without alphanumerical input).

List of functions (F7)

The list of functions is an easily opened window that shows the last 20 (max.) functions used.

Through a simple selection (various possibilities), these functions can be directly accessed without
going back into the drop-down menu.

Every function accessed via the drop-down menu is added to the end of the list; the others move
upwards.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

A selection of frequently needed functions can be specified (mark with _ as prefix). These are not
displaced when a new function is accessed.

Access functions by name (F5)


The functions of the drop-down menu can be accessed by entering their names.

Auto-complete enables users to access the functions more quickly be entering just a few letters,
which generates a list of hits.

This eliminates the need to look for a function in the drop-down menu.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Circle with 4-point macro at the push of a button on the control panel
Pressing the F11 key (control panel or keyboard) inserts a new circle in the features list and also opens the
4-point circle macro in the strategy (alternatively via the drop-down menu Measure Circle with 4-point
macro). Following an information window about the stylus positioning in the in center of the circle, the circle is
measured semi-automatically with 4 points as before.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.1.9

Circular path: circular travel path for single-point measurements

Circular path measurement strategy (circular path on plane) in the Plane feature. It is now possible to
configure circular travel paths if Single points has been activated. CALYPSO users can now also use
Circular path on plane for problematic assemblies (see description).
If the Single points function is activated in the Circular path on plane measurement strategy, the travel path
is comprised of the discrete probing points from straight travel paths. This happens although the name
indicates a circular path. With Circular path on plane, only the probing points lie on a circular path. For the
travel path, the shortest possible distance, i.e. the direct path, is selected. This behavior can lead to
collisions in the CNC run if the Circular path on plane is located on a groove on the workpiece or within an
assembly (e.g. bolt). Therefore, the Circular path on plane is limited in use. As a result, CALYPSO users now
have the option of selecting the form of the travel paths themselves: either direct travel paths as before, or
new circular paths.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

The new behavior of the travel path is integrated into the existing system context. The new, circular travel
path in no way replaces the existing travel path. CALYPSO users have the opportunity to choose from two
types of travel paths: direct or circular. The standard behavior is the direct travel path as before.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.1.10 Pattern function, exclude indices by multi-selection


This simplification allows for faster programming when using the "Exclude indices" functionality with patterns.

The list of indices can be more easily operated by the user

Instead of each index that you want to exclude, click individual indices. Hold the shift key to mark
index segments.

Simple checkbox to mark or unmark all indices

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.2 Display and Evaluation


3.2.1 Fourier analysis for 2D lines
Extended functionality for the evaluation of 2D lines

2D line can now be selected as a feature for the waviness characteristic

Requirement for the analysis of the Fourier analysis: 2D line must be measured with a sufficient
number of points

Fourier plot output: as before

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.2.2 Form & location: Show actual points of a characteristic


Optimized grafic display of the evaluation calculation in the CAD window.

Points:
The actual points of a characteristic can be displayed in the CALYPSO CAD window. The error of
the points is calculated against the CAD model and disproportionately displayed. However, this error
does not belong to the selected feature.

Results:
The display of a characteristic in the CAD window takes into account the references and all settings
in the characteristic such as limitation of the degrees of freedom, filter, etc.
It is possible export the min/max points

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Enhancement of the results display input page:

Automatic alignment: display is adjusted to monitor and rotated to provide a proper perspective
Nominal geometry + color selection
Line type of the boundary curves for the tolerance surfaces (except freeform surfaces)
Axis display + color selection: applies only for rotationally symmetric features
Min/max points expanded + color selection of the info boxes
Polyline supplemental to the other displays
The point display can be activated/deactivated for size

All settings can be changed via the characteristic measurement plan editor:

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.3 Standards and Mathematics


3.3.1 ISO 14405-1 Symbols
Symbols added in the new ISO 14405-1 standard have been implemented in CALYPSO

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

With the evaluation settings for


diameter,
the
existing
methods
(Gaussian,
minimum,
minimum
circumscribed, maximum inscribed, ...)
are identified with icons from the
standard

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.4 Options
3.4.1 Curve: Performance enhancement
Shorter analysis times, faster CAD display, improved measurement strategies

Improved analysis times 5% (2D curve) 50% (3D curve)

Considerably faster CAD display (>Factor 10)

Improved measurement strategy different speeds and point densities within a measuring segment

Optimized rotary table scanning


Number of points now defined the same as in the strategy
Point spacing based on the surface, not the stylus center point (=>fewer loops with stylus correction)
Support for measuring point transfer after change to rotary table angle

Analysis and optimization suggestions of the used curve settings for each curve
(Best fit search distance nominal point spacing number of measured points)

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.4.2 PCM: Activate and deactivate measuring volume illumination with O-INSPECT

Activate and deactivate measuring volume illumination on the O-INSPECT.


Automatic activation with critical features, resulting in more comfortable control by the operator, primarily
during setup operations

For visual inspection, illumination can be activated in the measurement plan via PCM with
contact measurements and subsequently deactivated via PCM for optical measurements. As
a result, the optical measurements are not impaired by the measuring lab illumination

Command names:

measVolumeIlluminationOn() : Activation of measuring lab illumination


measVolumeIlluminationOff() : Deactivation of measuring lab illumination

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.4.3 PCM: Paths for stylus system change


Option to define travel paths for travel from and to the workpiece to the handover position of the stylus
system changer. Use for single and pallet measurements

Syntax for conditions and loops must no longer be manually entered, but can be simply imported
from the list.

Option to define travel paths for travel from and to the workpiece to the handover position of the
stylus system changer.

Safe path through specification of intermediate positions for critical conditions

Different definitions are possible depending on the measurement plan and rack assignments.

Use for single and pallet measurements.

The paths correspond to the existing travel paths (park position, CNC end, ...)

Activation:
Drop-down menu CNC travel paths Travel behavior for stylus system change

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Dynamic path selection with PCM (PCM option required)

Enables path selection depending on various conditions (pallet location, rack assignment, ...) for the
path before and after stylus system change.

Therefore, only special paths for certain styli for example

Differentiation for manual stylus system change possible

Simplified PCM code generation: To simplify the generation of PCM code, the context menu (right mouse
button) in the PCM editor has been enhanced for path determination.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

PCM functions: Insert the associated PCM function at the current cursor position.
Feature selection: Show all feature names of the type selected in the pop-up menu. Insert the feature name
selected from the list at the current cursor position. Characteristics, features and log header data was added
to all PCM editors in the context menu and in the icon bar (see above). Use via getActual or similar PCM
function for which the name of features is needed.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.4.4 Universal post processor for CALYPSO Preset

Change of sign, commutation of axis and the sequence of transformations (first rotation, then
translation or vice versa) can be easily adjusted via the user interface

The number of decimal places can be specified individually for coordinates and angles

The export file can consist of straight value lines, but can also contain fixed header and / or - footer
up to a table structure (eg Heidenhain controller)

Additional information such as index (electrode number, pallet ID, storage space number), part
name, file name, coordinate system name and tolerance status (at the inspection) may also be
included in the export file.

The Index is automatically incremented, but can also be fixed to a value.

For numeric values, the decimal separator (point or comma) can be preset for the target system regardless of the regional settings.
Editor for post processors

Parameters which will be exported via


the post processor at the end of the
measurement

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.5 Other
3.5.1 Stylus system change visible in Planner
Stylus change movements completely simulated. The entire work environment can be simulated in advance
offline.

Stylus system changer, as well as the required stylus systems can be defined in Planner. The entire setup is
visible in the simulation.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.5.2 Stylus System Creator Standalone Installation


Stylus system creator as own option. Generate and use stylus systems without Planner. Generate full stylus
documentation on any computer with a single click.

Download available in German, English and Chinese at www.probes.zeiss.com

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

3.5.3 Optimized Error Reports


CALYPSO users receive a suggestion in the error report about how to complete the report. This will simplify
the work for customer support and development and lead to faster troubleshooting.

The more clearly an error is described, the more time a support employee or developer saves. Unfortunately,
not all error reports contain the required information and this guide is designed to improve handling.

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

4 Installation Instructions
4.1 Installing the CALYPSO 2014 Basic DVD
Please refer to all notes regarding the installation in chapter 1.2.
CALYPSO is a single-user application. It cannot be installed on a network and used as a client-server
application. In order to install CALYPSO you must have full local Administrator rights.
If you are performing an update of an existing CALYPSO installation, the existing installation should
be uninstalled first.
Existing data such as measurement plans, probe data or other CMM-specific data will not be deleted by
uninstalling CALYPSO. However we still recommend creating backup copies of your data at regular intervals.
Uninstall is performed using the Windows Uninstall function. In Windows, click on 'Start' and select
'CALYPSO' via 'Settings\Control Panel\Add or Remove Programs' and then click on 'Remove'.
In order to comply with the Microsoft recommendations for separating programs and corresponding data on
Windows 7 systems the installation of CALYPSO resides in the following folders:
Programs:
Program data:
User data:

C:\Program Files (x86)\Zeiss\CALYPSO 5.8


C:\ProgramData\Zeiss\CALYPSO 5.8
C:\Users\Public\Documents\Zeiss\CALYPSO

If the existing CALYPSO version is not uninstalled first, CALYPSO 2014 will be installed alongside the
existing version, thereby enabling alternative usage of both versions. Before installing CALYPSO 2014
alongside CALYPSO 5.2 or CALYPSO 5.4 please ensure that the Service Packs listed below are already
installed. If this is not the case, the installation of CALYPSO 2014 can cause a licensing conflict and
consequently prevent CALYPSO from starting.

Pay close attention when using multiple versions of CALYPSO alternately!


- ServicePack 5.2.20 or higher is required for parallel installation with version 5.2.
- ServicePack 5.4.16 or higher is required for parallel installation with version 5.4.
- When using the automatic stylus system changer please consider the increased danger of getting
stylus systems mixed up! After switching versions make sure that the currently selected stylus
system is actually mounted in the probe before calling the automatic changer.
- Measurement plans saved with a new CALYPSO version cannot subsequently be opened with an
older version. Pay attention to warning messages.
- Check for correct path settings when using external evaluation programs which access CALYPSO
data.
- Parallel installation of CALYPSO combined with the Metrotom Module is not permissible.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Installation Procedure:
When the DVD is inserted, the installation routine usually starts automatically. If this is not the case, use the
Windows Explorer to select 'CALYPSO.exe' on the DVD and start the installation by double-clicking.

Select 'Install'

Read the notes and select 'Continue'

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Select 'CALYPSO'

CALYPSO 2014 will automatically install Windows Installer 4.5, dotNET Framework 3.5 SP1, dotNET
Framework 4.0, SQL Server 2008, a PDF printer set, various C++ Redistributables, ZEISS Basic Reporting,
ZEISS Stylus System Creator and the ZEISS License Activation Utility as prerequisites should these not
already be available on the target system. This can require one or more system reboots and can take a long
time (especially the dotNET Framework).
Following the prerequisites the CALYPSO 2014 setup is unpacked and the installation is initialized and
started, the terms of the license agreement are displayed and must be accepted.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

The CALYPSO setup automatically recognizes whether an older version is already installed and will proceed
to show the following dialog:

If you wish to install the new CALYPSO version in addition to an existing verson you will have to confirm
taking note of the safety precautions mentioned above in order to continue with the setup. Installing an
older version alongside an existing newer version is not permissible!
If several versions of CALYPSO are installed simultaneously the respective application data is stored in
separate folders, in this case the application data folder for CALYPSO 2014 is:
C:\Users\Public\Documents\Zeiss\CALYPSO 5.8.
Note: If CALYPSO 2014 is installed alongside CALYPSO 5.4 or CALYPSO 5.2 the CALYPSO
ServicePacks 5.4.16 or newer or 5.2.20 or newer must be installed first.

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

4.2 Installing CALYPSO 2014 ServicePacks and Patches


Required program modifications and additions are provided as ServicePacks or Patches. These can arrive
by eMail, on data medium or via download. The procedure is basically the same for all media types.
If you have received a CALYPSO ServicePack or a Patch together with the CALYPSO Basic DVD, the
ServicePack or Patch must be installed after the installation of the Basic DVD.
Prerequisite for the installation of a ServicePack or a Patch is always an existing installation of the
corresponding basic version from a DVD. The version currently installed can be seen in CALYPSO in the
Miscellaneous menu. The version currently installed is also automatically output in the 'Error Report. You
reach this via the 'Extras menu.
Note on Patch versions: Patch versions contain urgent corrections to recently reported bugs and
subsequently deployed after undergoing minimum test coverage. Should your system be affected in
manner we recommend installing the Patch. Alternatively, the next ServicePack will also contain
appropriate bugfixes. ServicePacks are subjected to the full test coverage before being released.
detailed information regarding current Patch versions please contact our support team.

are
this
the
For

For current ServicePacks and Patches concerning CALYPSO visit our website:
http://extranet.zeiss.de/abtimt/download.nsf
Please contact our Software Support in order to receive the required access data.

4.2.1 ServicePack Installation


1. Exit CALYPSO first if it is still running.
2. Use Windows Explorer to navigate to the folder where you have stored the ServicePack file on your
computer.
3. Start the ServicePack installation by double-clicking the desired ServicePack file.
Example for ServicePack 5.4.04:

The ServicePack installation will then run through completely without requiring any further input until you
reach the final dialog.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

4. Click on 'Finish' to finalize the installation.


As from Version 5.4, CALYPSO ServicePacks can also downgrade your installation. That means you can
return to an older version of CALYPSO (e.g. from 5.4.08 to 5.4.04) if required. Simply install '5.4.04' over the
installed '5.4.08' version, which will then automatically be uninstalled, and CALYPSO version 5.4.04 will be
restored.

4.2.2 Patch Installation


1. Exit CALYPSO first if it is still running.
2. Use Windows Explorer to navigate to the folder where you have stored the Patch file on your computer.
3. Start the Patch installation by double-clicking the desired Setup.msp file.
Example for Patch 5.4.0001:

The Patch installation will then run through completely without requiring any further input until you reach the
final dialog.
4. Click on 'Finish' to finalize the installation.
Patches for CALYPSO can also be removed if required. To do so, use the Windows Control Panel and select
'Uninstall program'. Click on 'Show updates' to display the Patches. Select the Patch you would like to
remove and click on 'Uninstall'. After the Patch removal CALYPSO will return to its previous state.

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

4.3 CMM Data Backup

The 'Save CMM data' and 'Install CMM data' backup functions are not available for CALYPSO 2014.

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

4.4 Installing the ViScan Drivers


Hardware drivers are required for ViScan operation, the setup routine described in this chapter installs all of
the required drivers.
Older Framegrabber cards, as shown here, cannot be used with Windows 7!

Newer Framegrabber cards, as shown below, can be used with Windows 7.

The setup is run from the CALYPSO DVD by selecting Drivers and Components and Optical Sensors'.
The ViScan Dongle must be inserted (Type 1 on the parallel port or Type 2 on a USB 2.0 port. Caution: Take
care not to connect the dongle to a USB 3.0 port (blue colour code))!

1. Select Drivers and Components and select Optical Sensors.

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

2. Select 'ViScan' and then select the appropriate Version for your windows system (32-Bit or 64-Bit):

3. After Hardlock and IFC the Sapera framework is installed. Confirm with Next:

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

4. Confirm with Yes:

5. Confirm with Next:

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

6. Confirm with Next:

7. Select No, I will restart my computer later and confirm with Finish:

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

8. The driver for the Framegrabber will now be installed. Confirm with Next:

9. Confirm with Yes:

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

10. Confirm with Next:

11. Select Run-time and confirm with Next:

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

12. Confirm with Next:

13. Finally the current Framegrabber Firmware will be loaded. Confirm with Automatic:

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

14. Close the installation with Yes ...

and 'OK':

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

4.5 Installing Software for Metrotom

Depending upon the operating system in use (32-Bit or 64-Bit) the Metrotom Measuring Module needs to be
installed in addition to the CALYPSO basic software.
Note: Parallel installation of more than one version of CALYPSO in combination with the Metrotom
Measuring Module is not permissible.

Metrotom-CT install sequence:


1. Install CALYPSO basic version.
2. Select 'Drivers and Components' and 'Metrotom Measuring Module' from the menu on the basic DVD and
then decide between the 32-Bit and 64-Bit versions.

3. Finally install CALYPSO ServicePack or Patch if applicable.

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

5 Compatibility

Please read the Notes and Familiar Problems in Chapter 1 before you use this software.

5.1 CMM Validity


CALYPSO 2014 is intended for application on the following coordinate measuring machines:
Bridge measuring machines:

Vista
MICURA
CONTURA
CONTURA G2
CONTURA G3
SPECTRUM
ACCURA
ACCURA II
PRISMO
UPMC
CenterMax
GageMax

Large measuring machines:

MMZ-T
MMZ-G
MMZ-E
MMZ-B

Special measuring machines:

F-25
O-Inspect
METROTOM
DuraMax

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

5.2 Controllers & Sensors


Permissible combinations of CMM type, controller and firmware version:
Control

Firmware

Interface

Probe

C 99

>=10.30 Vista
rec om m end ed 11.04
>=10.47 Prismo,
Contura
rec om m end ed 11.04
>=11.04 Centermax
>=14.07 Access
>=15.00 Gagemax
UPMC
>=16.05 , Accura,
Spectrum(11*)
>= 18.08 Contura G2
>= 20.03,PRISMO MASS
ACCURA MASS
>=21.00 DuraMax
>=20.15 F25
>=20.15 O-Inspect
>=24.0 ACCURA 2
>=26.0 MICURA
>=29.10 CONTURA G3

LAN

Direct: ST 3, ST, VAST, VAST Gold, XT, DT(3*),


HSS(19*), XXT 18 pin (18*), CFS (29*)*), VAST
XTR(30*)
Holder RDS: RST, TP6, TP20 (14*,17*), TP200(14*,
17*; 20*),XXT 5 pin (20*, 21*), SP25(14*, 20*), XXT
18 pin (18*, 20*), Linescan(25*), XDT (27*, 28*),
Linescan2(32*)
Holder PH9/PH10 (16*): TP2, TP6, TP20(16*),
TP200(3*, 9*,16*, 20*)
Holder MIH/MH8: TP2, TP6

Notes:

This matrix assumes that the controls have been configured correctly.
On column 2 - The software must not be used with firmware releases older than the minimum release stated. Some problems
also could arise with newer interm ediate firmware releases which are however older than the current release, see remark at
the bottom of this section. The software was tested with up to date CMMs and current firmware releases. This configuration
will guarantee maximum performance and full functionality.
Not all stated firmware releases and CMM types (column 2) are compatible with all probe types (column 4)! They are each
referenced independently to this software release.

3*) Valid for firmware revision 11.04 or higher


9*) Valid for firmware revision 16.03 or higher
11*) Eclipse and Spectrum New deliveries only with firmware >= 18.06. For software updates on CMMs with ST3 without MPS the
firmware need not be changed.
12*) For multi-stage probe change system with RDS valid from firmware revision 18.06 or higher
14*) Valid for firmware revision 18.06 or higher (Please also refer to the Firmware conditions).
16*) The multi-stage probe change system is not implem ented for this holder.
17*) The firmware release 18.06 requires CALYPSO version 4.2 for the probe change system
18*) The XXT 18 pin requires firmware >= 19.05. (Please also refer to the Firmware conditions).
19*} The HSS probe is supported for firmware 15.04 or higher.
20*) not for horizontalarm CMMs with DSE,CSC.
21*) The 5 pin XXT requires firmware 19.06.
25*) Linescan is supported for firmware 20.03, WBScan >= 1.11 and MASS wiring
27*) The XDT sensor requires firmware >= 23.04. (Please also refer to the Firmware conditions).
28*) The XDT sensor requires CALYPSO version 5.0
29*) For CFS (Chromatic Focus Sensor) application Firmware >= 24 is mandatory.
30*) For VAST XTR application Firmware >= 26.10 is mandatory.
32*) Linescan 2 from Firmware >=30.04, WBScan >= 1.20 and MASS wiring

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Notes on Functionality:
Use of the "Cylindrical or Circular Travel Path" function requires Firmware >= 19.06.
The table shown above comprises released configurations that have been tested with up-to-date
coordinate measuring machines only. Such a configuration will guarantee full performance and
complete functionality.
Further combinations as well as the application of other components are permissible, e.g. via
Retrofit. Application of modified systems needs to be confirmed separately by Carl Zeiss Industrielle
Messtechnik GmbH.
The C98 controller is no longer supported with CALYPSO 2014. Application of this software version
in combination with a C98 controller is solely on the users own responsibility.
Operation of older combinations of hardware and firmware may lead to restrictions (see above *).
Restrictions may result regarding e.g.:
-

Measurement
Scanning functions
Support of probe changes
System performance

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

5.3 PC System

Recommended data systems:


ZEISS Entry Workstation:

Z220 HP Workstation G1
-

ZEISS High End Workstation:

HP Single Unit (CMT) Packaging


HP Z220 CMT 400W 90% Efficient Chassis
Windows 7 Ultimate Edition 64bit OS
Intel Core i3-2120 3.3 3M 2C GT1 CPU
8GB DDR3-1600 ECC (2x4GB) RAM
128GB SATA 1st SSD
500GB 7200 RPM SATA 6G 2nd HDD
NVIDIA NVS 310 512MB GFX
2 Netzwerkanschlsse
HP No Keyboard
HP USB 1000dpi Laser Maus
16X SuperMulti DVDRW SATA 1st ODD
HP TPM Disabled
China Regulatory CCC Compliance Mark
HP 3-3-3 CMT Warranty
Windows 7 64 Bit Factory Image Recovery
HP Z220 Workstation Country Kit
CMT SATA HDD2 Kabel

Z820 HP Highend Workstation G1


-

HP Single Unit Packaging


HP Z820 1125W 90 Efficient Chassis
Windows7 Ultimate Edition64BIT
Windows 7 64 Bit Factory Image Recovery
Xeon E5-2667 6C 2.90 15MB 1600 CPU-1
Xeon E5-2667 6C 2.90 15MB 1600 CPU-2
Integrierter Intel 82579LM PCIe Dual GbE
HP 64GB (8x8GB) DDR3-1600 2-CPU Reg RAM
NVIDIA Quadro 4000 2.0GB Graphics
HP 256GB SATA 1st SSD
HP 2TB SATA 7200 2nd HDD
3 Netzwerkanschlsse
HP No Keyboard
HP USB Laser Scroll Mouse
16X SuperMulti DVDRW SATA 1st ODD
HP TPM Disabled
China Regulatory CCC Compliance Mark
HP 3-3-3 Warranty
HP Z820 Localization Kit
HP Dual Processor Air Cooling Kit

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

Minimum system requirements:


The following minimum requirements must be fulfilled for installing and operating CALYPSO 2014:
PC with 1,6GHz and 2GByte RAM, mouse, printer and data backup facility.
The minimum screen resolution is 1280x1024 pixels. This is the default size for the ' CALYPSO User Desk'
window.
2 network connections must be possible (incl. installed driver):1x for control connection, 1x for customer
network. The TCP/IP network protocol must be installed. The installed graphics card must be installed with
OPEN-GL drivers. At least one of the ZEISS approved graphics cards should be used (NVIDIA Quadro FX
3800, NVIDIA Quadro2000 or ATI FirePro V3800).

Windows Operating Systems:


The following Windows versions are suitable for running CALYPSO 2014:
Windows 7
Windows 7

32bit
64bit

CALYPSO 2014 has been tested successfully on Windows 7 systems. Please note the following restrictions:
- The ViScan Sensor can only be used with Windows 7 in combination with suitable hardware. Please refer
to chapter 4.4.

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

5.4 Software compatibility

CALYPSO is a single-user application. It is cannot be installed on a network and used as a client-server


application.
GEAR PRO
Due to the changed directory structure, operation of GEAR PRO 4.0 and older GEAR PRO versions in
combination with CALYPSO 2014 is not possible. Operation of GEAR PRO and CALYPSO 2014 is possible
as from service pack GEAR PRO 4.0.2 or the version GEAR PRO 5.0 onward.
MCC
Due to the changed directory structure, operation of MCC 3.1 and older MCC versions in combination with
CALYPSO 2014 is not possible. Operation of MCC and CALYPSO 2014 will be possible from the future
version MCC 3.2 onward.
FACS
The FACS automation interface is customized to each customers needs and may be affected by the
changed directory structure. Before operating CALYPSO 2014 with your FACS application, you should
contact our support team.
PiWeb
If CALYPSO 2014 is installed in combination with PiWeb, version 3.8 (or newer) of PiWeb is required. Older
versions of PiWeb can cause problems creating PDF-files.

5.5 CAD Interfaces


CALYPSO supports the following CAD Interfaces:
CAD Software
Catia 5
Catia 4
Pro/ENGINEER
Unigraphics
SolidWorks
Inventor
JT Open*
Parasolid
IGES
VDAFS
STEP

Version
24
4.2.4
Wildfire 5.0 & Creo 2.0
NX 9.0
2014
2015
9.0
26
5.3
2.0
AP214

* JT conversions are made with the converter from Holometrics up to version 9.5

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

5.6 Compatibility of Measurement Plans and Programs


Compatibility Measurement Plans and Programs Warranty/Guarantee Exclusion

ZEISS will neither warrant nor guarantee functionality for measurement plans or programs created by third
party suppliers or by licensee/customer or that such measurement plans or programs will run error free on
ZEISS software or systems respectively. ZEISS excludes any form of warranty or guarantee regarding
measurement plans and programs supplied by a third party especially upon implementation of software
upgrades or new program versions.
In particular, ZEISS emphasizes the resulting possibility of lack or alteration of performance relating to
measurement plans should one or more of the following points apply:
- Exchanged computer operating system between software revisions
- Essential alterations to calculation algorithms
- Elimination of software errors and malfunction
- Changed dependencies between software options
- Improper programming of measurement plans

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

6 Contact
If you have questions, ideas or problems regarding CALYPSO, please always use the integrated CALYPSO
Error Report to contact us.
You can call the form for this by selecting 'Error Report' from the 'Extras' menu.
In the 'Comment' field you can add any text to explain your problem.
With the pull down menu in the Error Report, you can save the report (to forward this per email) or print it (for
faxing). Our email address and the fax number can be round in the report together with the respective phone
numbers.
For Germany:

For USA:

Carl Zeiss Industrielle Messtechnik GmbH


SES-AH Software Support
73446 Oberkochen

Carl Zeiss IMT Corp.


Software Support
Novi MI 48377

Tel.: 07364-20-6336 oder 6337


Fax: 07364-20-4304

Tel.: 1-800-327-9735
Fax: 248-624-1258 or
763-535-9792
eMail: calypso@zeiss.com

eMail: calypso-support.metrology.de@zeiss.com
www.zeiss.de/imt

www.zeiss.com

ServicePack Downloads
For current ServicePacks concerning CALYPSO visit our website:
http://extranet.zeiss.de/abtimt/download.nsf
Please contact our Software Support in order to receive the required access data.

Metrology Academy Login Up to Date Info on Metrology Software under Windows


Get the latest information regarding Zeiss metrology software on Windows systems here:
http://academy.imt.zeiss.de
With a login and a password you can access 'Software Knowledge' und 'Windows Tips' for the latest tips and
tricks.

Back to overview

CALYPSO Copyright 1996-2014 Carl Zeiss Industrielle Messtechnik GmbH. All rights reserved.

19.12.2014

You might also like