Professional Documents
Culture Documents
Solution:
6
Page 1 of 27
Q2. What is the output voltage (0) of the following amplifier circuit?
Solution:
12 V.
As we know that no current flows into the OP-AMP or the positive and the negative terminal
are at the same voltage (1). The diagram below shows the flow of currents.
1 =
5
= 1
5
2 =
1 1 1 2
=
= 1
1
1
1 = 2 1 = 2
0 = 1 10 2 = 2 + 10 1 = 12
Page 2 of 27
Q3. Given a resistor with 5% accuracy, what will be the percent accuracy in ?
i) 2.5%
ii) 5%
iii) -5%
1
iv) %.
Solution: ii
= 1/
Page 3 of 27
Q4. A DC voltage across a resistor is measured with an accuracy of 3%. If the resistor is
valued with a 5% accuracy, what is the percent accuracy in determining the power dissipated
by the resistor from values of the resistor and measured voltage?
Solution:
2
1
=
=
= + + = 3% + 3% + 5% = 11%
Page 4 of 27
Q5. A Wheatstone bridge for measuring resistance is shown below, where 1 , 2 , and 3 are
resistors of the bridge circuit and is the resistor to be measured. The bridge circuit is
balanced by adjusting 3 such that the voltmeter indicates zero reading. Then the value of
can be represented in terms of the bridge resistors.
(i) Show that the balanced condition for the
constant.
1 +2
3 +
1 3
2
=
2 3 = 1
=
= ,
1 + 2 3 +
2
(ii)
=
=
+
2
1 + 2 3 + +
1 + /
1 + /
1
1
1 + 1 + + /
1 /2 + 1 3 / + 1 + /
(1 + + / ) (1 + )(1 + / )
=
(1 + )(1 + + / )
=
(1 + )(1 + + / )
(iii) =
2 3
1
,
(1 + )2
= 1 + 2 + 3 = 3%
1
Page 5 of 27
Q6. You are measuring temperature of a solution using a mercury thermometer. You have
done measurements 10 times and obtained the following readings. 30.1, 30.2, 29.8, 32.1,
30.5, 29.9, 30.1, 29.5, 31.1, and 30.6. Calculate the mean and the standard deviation of the
measured temperatures.
Solution:
Mean:
30.1+30.2+29.8+32.1+30.5+29.9+30.1+29.5+31.1+30.6
= =
= 30.39
Standard Deviation:
( )2
1
10
= 0.7505
Page 6 of 27
Q7. Below are some hypothetical results from mammogram for breast cancer screening.
What is the specificity of the mammogram?
Mammogram positive
Mammogram negative
Total
Breast cancer
70
30
100
No breast cancer
180
720
900
Total
250
750
1000
What are the sensitivity and specificity and prevalence of the disease?
Solution:
Sensitivity = 70/100 = 70%
Specificity = 720/900 = 80%
Prevalence = 100/1000 = 10%
Page 7 of 27
//
+
=
=
( ) + //
( ) +
+
=
Page 8 of 27
Q9. A capacitive pressure measurement scheme is shown below where the movable and fixed
metal diaphragms form a capacitor . Assume that the common area of the diaphragm, the
dielectric constant of the capacitor are known, and the distance between the diaphragm is
inversely proportional to the gas pressure.
(i) Show that the gas pressure can be
measured by measuring the capacitance
value .
that = , =
(ii) =
, =
, = +
Page 9 of 27
Q10. If blood flowing in a blood vessel with a radius of 2 mm cuts a magnetic field with flux
density = 0.1 tesla, and the blood volume flow rate is 10 cm3/s, what is the electric
potential that can be detected across the blood vessel?
Solution:
2
=
=
4
4
4 4 10 0.013 0.1
=
= 0.318
=
3.14 2 2 103
Page 10 of 27
Page 11 of 27
Page 12 of 27
1 1
It is known that = 3150 and (300) = 1600 . Find the sensitivity of the thermistor at
= 300 = 310 , respectively.
Solution:
() = (0 )
()
= (0 )
1 1
0
1 1
0
= (0 )
1 1
0
1 1
1 1
1
1 1
1 1
()
0
0
0
= (0 )
= (0 )
1 1
()
1
1
= (0 ) 0 2 = () 2
()
= 2 ()
3150
(300) =
(300) = 56 /
3002
3150
(310) =
(310) = 37.4 /
3102
Page 13 of 27
Page 14 of 27
Page 15 of 27
Q16. Why cannot Ground-fault interrupter (GFI) be used on life support device circuitry?
Solution:
GFI stops the power supply when there is a fault. This may lead to death of the patient on life
support device.
Page 16 of 27
Q17. Explain how a line isolation monitor (LIM) can be used to detect a neutral line shorted
to the ground.
Solution
To detect fault of the neutral line the Line isolation monitor (LIM) needs to be connected
between the hot and ground lines. When there is no fault the current flowing through the
LIM is very small due to the very small stray capacitance. Once the neutral get shorted to the
ground, the current flowing through the LIM increases. The ammeter connected in the LIM
shows the fault.
Equivalent circuit
2
2 + 1
When the neutral gets shorted to the ground the equivalent circuit becomes,
So the LIM is set in such a way that whenever there is a current flow higher than the set limit,
it will set an alarm, which indicates that there is a fault in the neutral wire.
Page 17 of 27
Q18. Calculate the maximal safe capacitance between a liquid-filled catheter and dc-isolated
pressure sensor leads for a 230 volts, 50 Hz fault in the sensor leads if its equivalent circuit is
as shown in Figure below.
Solution:
As this is a microshock hazard, so the maximal safe current is 10 A.
1
Current = || =
230
||
= 10
|| =
230
= 2.3 107
105
1
This shows that the capacitance shouldnt be more than 138 pF for safety.
Page 18 of 27
Q19. A receptacle tester has the circuit shown below. Indicate which LEDs will be ON or
OFF for fault conditions given in Table.
Conditions
Hot open
Neutral open
H and N reversed, N open
Ground open
Hot and ground reversed
Conditions
Hot open
Neutral open
H and N reversed, N open
Ground open
Hot and ground reversed
1
OFF
ON
OFF
OFF
ON
2
OFF
OFF
ON
OFF
ON
3
OFF
OFF
OFF
ON
OFF
Solution:
Page 19 of 27
Page 20 of 27
Q21. Below are three different volumes related to human lung. What are i) the definitions
and ii) typical values for an adult lung? Tidal volume, Functional Residual Capacity (FRC),
Dead Space.
Solution:
Tidal Volume: volume of gas inspired or expired during each normal respiratory cycle. 500
ml.
FRC: Amount of gas remaining in the lungs at the resting expiration. 2400 ml.
Dead Space: Volume of air that is not available for gas exchange. 150 ml.
Page 21 of 27
Page 22 of 27
Q23. a) Figure below shows a spirometer. In its simplest form, the bellow is mechanically
articulated to a biased potentiometer such that the wiper arm voltage is proportional to
volume. Derive the relationship of the voltage and volume.
b) If the spirometer has a radius of 10 cm, a maximum height of 10 cm, and = 10 ,
plot the air volume versus the output voltage.
Solution:
(a) is the maximal distance change that the bellows can change. So the maximum volume
change is = . Since the spirometer volume ( ) is proportional to the output
voltage ( ), =
If we adjust linkage such that at the potentiometer output has = .
=
As the changes in distance of bellows and output of potentiometer are proportional. We have
(b) As = , = , = , we have
( ) ( )
=
=
= . /
( )
So, = .
Page 23 of 27
Q24. Briefly explain nitrogen washout method for estimation of lung residual volume (RV)?
Solution
Nitrogen washout method for estimating lung residual volume is using a modified spirometer
setup and a nitrogen analyzer. Here, the patient starts inhaling pure oxygen through a one
way tube. Then he exhales air into the spirometer through another one way tube. When
exhaling he expels a mixture of oxygen, nitrogen and carbon-dioxide into the spirometer. So
after many breathing cycle the nitrogen content in the lung and in the spirometer will come to
equilibrium.
By doing a mass balance of nitrogen before and after the experiment, one can calculate the
RV. Provided the person started the breathing cycle at the moment when the lung volume was
at RV.
Lets say at the beginning (time 1 , and the lung volume is ), the total number of 2 moles
in the lung is given by,
2 (1 )
At the beginning the total number of 2 moles in the spirometer is NIL (0).
At the end (time 2 , and the lung volume is ), the total number of 2 moles in the lung is,
2 (2 )
At the end (time 2 , and the spirometer volume is (2 ), the total number of 2 moles in the
spirometer is,
(2 )
2 (2 )
By doing a mass balance (since no 2 has gone out of the lungs other than spirometer),
(2 )
2 (1 )
2 (2 )
= 2 (2 )
=
2 (2 ) (2 )
2 (1 ) 2 (2 )
Page 24 of 27
() = 1010
0
6.02 = 1010
0
10 = 6.02/10
0
0 = 3.99 4
Page 25 of 27
Page 26 of 27
Page 27 of 27