Instrument Application
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Uric Acid
Chemistry
Standard
0
37 C
2.5 7.7
<= 20, Use Reflex Test, Dilution Ratio = 2
mg/dl
1
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Reagent [Uric Acid Reagent], Volume [200.0 ul]
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High
Incubate 00:01:00 [0h 1min 0sec]
Read with Primary Filter [505], Differential Filter [630]
Add Sample, Volume [5.0 ul]
Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High
Incubate 00:10:00 [0h 10min 0sec]
Read with Primary Filter [505], Differential Filter [630]
Document number: APP-U7581-CW2910 Rev. 6-12
Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com
Test: Albumin
Catalog # : A7502
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Albumin
Chemistry
Standard
0
37 C
3.5 5.3
0.5 8.0, Use Reflex Test, Dilution Ratio = 2
g/dl
1
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Reagent [Albumin Reagent], Volume [200.0 ul]
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High
Incubate 00:01:00 [0h 1 min 0 sec]
Read with Primary Filter [630], Differential Filter [700]
Add Sample, Volume [2.0 ul]
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low
Incubate 00:01:00 [0h 1 min 0 sec]
Read with Primary Filter [630], Differential Filter [700]
Document number: APP-A7502-CW2910 Rev. 6-12
Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Alk Phos
Chemistry
Factor
Factor = 2187 (Adjusted 3988.1)
0
37 C
35 123
<= 800, Use Reflex Test, Dilution Ratio = 2
U/L
0
********************************************************************************************************************
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [5.0 ul]
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
ALT (SGPT)
Chemistry
Factor
Factor = -1768 (Adjusted -3004.2)
0
37 C
4 36
<= 500, Use Reflex Test, Dilution Ratio = 2
U/L
0
********************************************************************************************************************
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [20.0 ul]
Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low
Test: Amylase
Catalog # : A7564
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Amylase
Chemistry
Factor
Factor = 3178 (Adjusted 5795.3)
0
37 C
25 125
<= 2000, Use Reflex Test, Dilution Ratio = 2
U/L
0
********************************************************************************************************************
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [5.0 ul]
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
AST (SGOT)
Chemistry
Factor
Factor = -1768 (Adjusted -3004.2)
0
37 C
5 34
<= 500, Use Reflex Test, Dilution Ratio = 2
U/L
0
********************************************************************************************************************
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [20.0 ul]
Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
HDL
Chemistry
Standard
0
37 C
30 85
2 150, Use Reflex Test, Dilution Ratio = 2
mg/dl
0
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
autoHDL/LDL Calibrator, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [3.0 ul]
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Direct Bilirubin
Chemistry
Standard
0
37 C
0.0 0.5
<= 20.0, Use Reflex Test, Dilution Ratio = 2
mg/dl
1
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [25.0 ul]
Aspiration Speed=4, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Total Bilirubin
Chemistry
Standard
0
37 C
0.2 1.2
<= 20.0, Use Reflex Test, Dilution Ratio = 2
mg/dl
1
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Reagent [Total Bilirubin working Reagent], Volume [200.0 ul]
Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High
Incubate 00:01:00 [0h 1min 0sec]
Read with Primary Filter [545], Differential Filter [630]
Add Sample, Volume [10.0 ul]
Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High
Incubate 00:05:00 [0h 5min 0sec]
Read with Primary Filter [545], Differential Filter [630]
Test: BUN
Catalog # : B7552
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
BUN
Chemistry
Standard
0
37 C
7 18
<= 150, Use Reflex Test, Dilution Ratio = 2
mg/dl
0
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [2.0 ul]
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low
Test: Calcium AZ II
Catalog # : C7529
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Ca AZ II
Chemistry
Standard
0
37 C
8.5 10.4
<= 15.0, Use Reflex Test, Dilution Ratio = 2
mg/dl
1
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Reagent [Calcium AZ II Reagent], Volume [200.0 ul]
Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High
Incubate 00:01:00 [0h 1min 0sec]
Read with Primary Filter [600], Differential Filter [700]
Add Sample, Volume [2.0 ul]
Aspiration Speed=1, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low
Incubate 00:03:00 [0h 3min 0sec]
Read with Primary Filter [600], Differential Filter [700]
Document number: APP-C7529-CW2910 Rev. 6-12
Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Calcium CPC
Chemistry
Standard
0
37 C
8.5 10.5
<= 20.0, Use Reflex Test, Dilution Ratio = 2
mg/dl
1
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Reagent [Calcium Color Reagent], Volume [100.0 ul]
Aspiration Speed=2, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High
Add Reagent [Calcium Buffer Reagent], Volume [100.0 ul]
Aspiration Speed=2, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High
Incubate 00:01:00 [0h 1min 0sec]
Read with Primary Filter [545], Differential Filter [700]
Add Sample, Volume [4.0 ul]
Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High
Incubate 00:01:00 [0h 1min 0sec]
Read with Primary Filter [545], Differential Filter [700]
Document number: APP-C7503-CW2910 Rev. 6-12
Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com
Test: CK
Catalog # : C7522
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
CK
Chemistry
Factor
Factor = 6592 (Adjusted 12021.0)
0
37 C
<= 160
<= 1500, Use Reflex Test, Dilution Ratio = 2
U/L
0
********************************************************************************************************************
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [5.0 ul]
Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low
Test: Chloride
Catalog # : C7501
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Chloride
Chemistry
Standard
0
37 C
98 106
80 120, Use Reflex Test, Dilution Ratio = 2
mEq/L
0
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Reagent [Chloride Reagent], Volume [200.0 ul]
Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High
Incubate 00:01:00 [0h 1min 0sec]
Read with Primary Filter [505], Differential Filter [630]
Add Sample, Volume [2.0 ul]
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=1, Dispense Height=High
Incubate 00:05:00 [0h 5min 0sec]
Read with Primary Filter [505], Differential Filter [630]
Test: CO2
Catalog # : C7502
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
CO2
Chemistry
Standard
0
37 C
23 34
2 - 40, Use Reflex Test, Dilution Ratio = 2
mEq/L
0
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [2.0 ul]
Aspiration Speed=1, Air Gap=2ul, Dispense Speed=1, Dispense Height=Low
Add Reagent [CO2 Reagent], Volume [200.0 ul]
Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High
Incubate 00:01:30 [0h 1min 30sec]
Kinetic Read, Filters [405/505], Fixed Time [5min 0sec]
Document number: APP-C7502-CW2910 Rev. 6-12
Test: Cholesterol
Catalog # : C7510
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Cholesterol
Chemistry
Standard
0
37 C
< 200
<=700, Use Reflex Test, Dilution Ratio = 2
mg/dl
0
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Reagent [Cholesterol Reagent], Volume [240.0 ul]
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High
Incubate 00:01:00 [0h 1min 0sec]
Read with Primary Filter [505], Differential Filter [630]
Add Sample, Volume [2.4 ul]
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low
Incubate 00:10:00 [0h 10min 0sec]
Read with Primary Filter [505], Differential Filter [630]
Document number: APP-C7510-CW2910 Rev. 6-12
Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com
Test: Creatinine
Catalog # : C7539
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Creatinine
Chemistry
Standard
0
37 C
0.4 1.4
<= 25.0, Use Reflex Test, Dilution Ratio = 2
mg/dl
1
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [10.0 ul]
Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low
Add Reagent [Creatinine working Reagent], Volume [200.0 ul]
Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High
Incubate 00:00:50 [0h 0min 50sec]
Kinetic Read, Filters [505/630], Fixed Time [2min 30sec]
Test: Phosphorus
Catalog # : P7516
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Phosphorus
Chemistry
Standard
0
37 C
2.5 4.8
<= 12.0, Use Reflex Test, Dilution Ratio = 2
mg/dl
1
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Reagent [Phosphorus Reagent], Volume [200.0 ul]
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High
Incubate 00:01:00 [0h 1min 0sec]
Read with Primary Filter [340], Differential Filter [630]
Add Sample, Volume [4.0 ul]
Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=Low
Incubate 00:05:00 [0h 5min 0sec]
Read with Primary Filter [340], Differential Filter [405]
Document number: APP-P7516-CW2910 Rev. 6-12
Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com
Test: Gamma GT
Catalog # : G7571
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
GGT
Chemistry
Factor
Factor = 2211 (Adjusted 3935.9)
0
37 C
9 54
<= 800, Use Reflex Test, Dilution Ratio = 2
U/L
0
********************************************************************************************************************
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [10.0 ul]
Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Glucose (Hex)
Chemistry
Standard
0
37 C
70 110
<= 500, Use Reflex Test, Dilution Ratio = 2
mg/dl
0
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Reagent [Glucose (Hex) Reagent], Volume [240.0 ul]
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High
Incubate 00:01:00 [0h 1min 0sec]
Read with Primary Filter [340], Differential Filter [405]
Add Sample, Volume [2.4 ul]
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low
Incubate 00:05:00 [0h 5min 0sec]
Read with Primary Filter [340], Differential Filter [405]
Document number: APP-G7517-CW2910 Rev. 6-12
Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Glucose (Ox)
Chemistry
Standard
0
37 C
70 105
<= 500, Use Reflex Test, Dilution Ratio = 2
mg/dl
0
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Reagent [Glucose (Ox) Reagent], Volume [240.0 ul]
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High
Incubate 00:01:00 [0h 1min 0sec]
Read with Primary Filter [505], Differential Filter [630]
Add Sample, Volume [2.4 ul]
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low
Incubate 00:10:00 [0h 10min 0sec]
Read with Primary Filter [505], Differential Filter [630]
Document number: APP-G7521-CW2910 Rev. 6-12
Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com
Test: HbA1c
Catalog # : H7541
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
HbA1c
Chemistry
Cubic Spline
1
37 C
< 6%
2 16
%
1
********************************************************************************************************************
Blank used:
Valid interval: 14 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
autoHDL/LDL Calibrator, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [5.0 ul]
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Iron, Total
Chemistry
Standard
0
37 C
60 150
<= 500, Use Reflex Test, Dilution Ratio = 2
ug/dl
0
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [30.0 ul]
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=1, Dispense Height=Low
Test: LDH
Catalog # : L7572
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
LDH
Chemistry
Factor
Factor = 3376 (Adjusted 6010.0)
0
37 C
80 285
<= 1000, Use Reflex Test, Dilution Ratio = 2
U/L
0
********************************************************************************************************************
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [5.0 ul]
Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low
Test: Lipase
Catalog # : L7503
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Lipase
Chemistry
Standard
0
37 C
0 62
<= 600, Use Reflex Test, Dilution Ratio = 2
mg/dl
0
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
Lipase Standard, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [2.5 ul]
Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low
Test: Magnesium
Catalog # : HM929
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Magnesium
Chemistry
Standard
0
37 C
1.5 2.6
<= 4.86, Use Reflex Test, Dilution Ratio = 2
mg/dl
2
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Reagent [Magnesium R1 Reagent], Volume [150.0 ul]
Aspiration Speed=2, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High
Add Reagent [Magnesium R2 Reagent], Volume [150.0 ul]
Aspiration Speed=2, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High
Incubate 00:01:00 [0h 1min 0sec]
Read with Primary Filter [545], Differential Filter [700]
Add Sample, Volume [2.0 ul]
Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High
Incubate 00:01:00 [0h 1min 0sec]
Read with Primary Filter [545], Differential Filter [700]
Document number: APP-HM929-CW2910 Rev. 6-12
Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com
Test: Microprotein
Catalog # : P7582
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Microprotein
Chemistry
Standard
0
37 C
28 141
2 250, Use Reflex Test, Dilution Ratio = 2
mg/24 hrs
0
for CSF and Urine
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
Microprotein Standard, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Reagent [Microprotein Reagent], Volume [200.0 ul]
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High
Incubate 00:01:00 [0h 1min 0sec]
Read with Primary Filter [600], Differential Filter [700]
Add Sample, Volume [4.0 ul]
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low
Incubate 00:05:00 [0h 5min 0sec]
Read with Primary Filter [600], Differential Filter [700]
Document number: APP-P7582-CW2910 Rev. 6-12
Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Total Protein
Chemistry
Standard
0
37 C
6.2 8.5
1.0 15.0, Use Reflex Test, Dilution Ratio = 2
g/dl
1
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Reagent [Total Protein Reagent], Volume [200.0 ul]
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High
Incubate 00:01:00 [0h 1min 0sec]
Read with Primary Filter [545], Differential Filter [700]
Add Sample, Volume [4.0 ul]
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low
Incubate 00:05:00 [0h 5min 0sec]
Read with Primary Filter [545], Differential Filter [700]
Document number: APP-T7528-CW2910 Rev. 6-12
Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com
Test: Triglycerides
Catalog # : T7532
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this
instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control
validation program.
Assay Name:
Assay Type:
Assay Mode:
Version:
Temperature:
Normal:
Linear:
Units:
# of decimals:
Assay Description:
Triglycerides
Chemistry
Standard
0
37 C
44 148
<= 1000, Use Reflex Test, Dilution Ratio = 2
mg/dl
0
********************************************************************************************************************
Blank used:
Valid interval: 7 days 0 hours
Controls:
Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours
Standards:
Chemistry Calibrator, Curve valid interval: 7 days 0 hours
********************************************************************************************************************
ASSAY STEPS
Add Reagent [Triglycerides Reagent], Volume [200.0 ul]
Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High
Incubate 00:02:00 [0h 2min 0sec]
Read with Primary Filter [505], Differential Filter [630]
Add Sample, Volume [2.5 ul]
Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High
Incubate 00:10:00 [0h 10min 0sec]
Read with Primary Filter [505], Differential Filter [630]
Document number: APP-T7532-CW2910 Rev. 6-12
Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com