Introduction
The 16 Calibration Information instruction sheet provides the
information necessary to calibrate and verify the Fluke
Model 16 Multimeter with Temperature (hereafter known as
the Meter). This instruction sheet provides the following
information:
Product specifications
Replacement parts
See the instruction sheet for operating instructions.
Service
The Meter is warranted to be free from defects in material
and workmanship for one year. The complete warranty
statement is in the Meters instruction sheet.
Specifications
Specifications are in Table 1. Accuracy is specified for a
period of one year after calibration, at 18C to 28C (64F to
82F) with relative humidity to 90%. AC conversions are accoupled, average responding, and calibrated to the rms
value of a sine wave input.
Table 1. Specifications
Maximum Voltage Between any Terminal and Earth Ground (excludes 10% tolerance) 600 V ac or dc
Display
Operating Temperature
-10C to 50C
Storage Temperature
Temperature Coefficient
Relative Humidity
Altitude Operation/Non-operation
2000 m
Battery Type
Battery Life
Shock, Vibration
Size (HxWxL)
Weight
10 oz (286g)
EMI Regulations
Safety
Designed to Protection Class II requirement of UL1244, ANSI/ISA-S82.01 - 1988, CSA C22.2 No 231,
and VDE 0411, and IEC 1010-1 overvoltage category III (CAT III), 600 V
Certification
P$
16
Multimeter with Temperature
Accuracy specifications are given as follows: ([% of reading] + [number of least significant digits]).
Table 1. Specifications (cont)
Function
Range
1
Resolution
Accuracy
4000 mV
1 mV
(1.9%+3)
4.000 V
0.001 V
(1.9%+3)
40.00 V
0.01 V
(1.9%+3)
300.0 V
0.1 V
NA
400.0 V
0.1 V
(1.9%+3)
600.0 V
1V
(1.9%+3)
4000 mV
1 mV
(0.9%+2)
4.000 V
0.001 V
(0.9%+2)
40.00 V
0.01 V
(0.9%+1)
300.0 V
0.1 V
NA
400.0 V
0.1 V
(0.9%+1)
600 V
1V
(0.9%+1)
400.0
0.1
(0.9%+2)
4.000 k
0.001 k
(0.9%+1)
40.00 k
0.01 k
(0.9%+1)
400.0 k
0.1 k
(0.9%+1)
4.000 M
0.001 M
(0.9%+1)
40.00 M
0.01 M
(1.5%+3)
1.000 F
0.001 F
(1.9%+2)
10.00 F
0.01 F
(1.9%+2)
100.0 F
0.1 F
(1.9%+2)
1000 F
1 F
(1.9%+2)
10000 F
10 F
(10%+90)
typical
0 to 200 A
0.1 A
(1% + 2 counts)
0 to 200 A
0.1 A
2% + 3 counts)
-10C to 400C
14F to 75F
(1% + 0.8C)
(1% + 1.5F)
-40C to -10C
-40F to 14F
0.1C
0.2F
0.1C
0.2F
2.000 V
0.001 V
Temperature
(Type K Thermocouple)
C2
(5% + 1.5C)
Typical
(5% + 3.3F) Typical
(0.9%+2)
1.
The 4000 mV range can be entered only in the manual range mode. Use the 4000 mV range with accessories.
2.
The beeper is guaranteed to come on at <25 and turn off at >250. The Meter detects opens or shorts of 250 s or
longer.
3.
Specifications
Table 1. Specifications (cont)
Function
Overload
1
Protection
J
L2
1.
2.
Input Impedance
(Nominal)
2
Common Mode
Rejection Ratio 2
(1 k Unbalanced)
600 V dc
>5 M <100 pF
Automatic Selection
and LoZ = >2 k
<200 pF
3
(ac coupled)
600 V dc
>10 M <100 pF
Automatic Selection
and LoZ = >2 k
3
<200 pF
>100 dB at dc, 50 Hz or
60 Hz
Normal Mode
2
Rejection
>60 dB at dc 50 Hz or
60 Hz
<450 mV dc
>50 dB at 50 Hz or 60
Hz
Short Circuit
Current
<500 A
600 V dc
<1.5 V dc
<1.5 V dc
600 V dc
2.4-3.0 V dc
2.400 V dc
0.95 mA (typical)
Burden Voltage
Normal Mode
Rejection
600 V rms or dc
<5 mV/A
>100 dB @ dc 50 Hz or 60
Hz
600 V rms or dc
<5 mV/A
>60 dB at dc, 50 Hz or
60 Hz
>50 dB @ 50 Hz or
60 Hz
.
3 x 106 V Hz maximum.
~2 k input impedance up to 50 V. Impedance increases with input voltage to >300 k at 600 V.
16
Multimeter with Temperature
Remove the test leads and set the rotary knob to OFF.
2.
3.
4.
5.
Cleaning
W Warning
To avoid electrical shock, remove test leads
and any input signals before cleaning.
To clean the case, wipe it with a cloth lightly dampened with
water and a mild detergent. Do not use abrasives, solvents,
or alcohol.
7.
To reinsert the pca: Place the pca over the four screw
posts in the case top, then press gently on the center of
the pca while using the small flat-edge screwdriver to
shoehorn the pca under the snap on a side of the case
top. Repeat on the other side and the top.
8.
9.
9.
Recommended
Model
AC/DC Calibrator
DC Voltage: 0 to
600 V
Accuracy: 0.25%
AC Voltage: 0 to
600 V
Accuracy: 0.5%
Frequency: 50 to
400 Hz
5700A, or 5500A
Decade Resistor
Resistance: 1.0 to
40 M
Accuracy: 0.25%
Fluke 5500A
Decade
Capacitor
Capacitance: 0 to
1.000 F
Accuracy: 0.5%
Fluke 5500A
Note
Before reinserting the pca, make sure that the
rotary knob is in the OFF position.
Minimum
Specification
Thermocouple
Wire K-Type
---
Fluke 80 PK-1
Temperature
Probe
Accuracy: Certified
to 0.2C ambient
Fluke 80T-150U
Mercury
Thermometer
0.02C resolution
0.05F resolution
---
Dewar Flask
Decal
Case, Top
LCD
Conn.
Elastomeric
Conn.
Elastomeric
LCD to PCA
Switch Assembly
Actuator, Switch
Support, Switch Assembly
Battery
Shock Absorber
Main PCA
Case, Bottom
Shield, Bottom
Foot
ig3f.eps
16
Multimeter with Temperature
Snap
Snaps
Snap
ig2f.eps
Performance Test
Performance Test
7.
8.
W Warning
To avoid electric shock, do not execute the
performance tests procedures unless the
Meter is fully assembled.
To ensure that the Meter is working properly and performs
to specifications, use the following procedures. If the Meter
fails this test, it needs calibration adjustment or repair.
Multimeter
Note
The following performance test is for all functions
except for temperature. Temperature performance
test follows this procedure.
1.
2.
3.
Calibration Adjustments
Perform the following calibration adjustment procedures if
the Meter fails the performance test.
Initial Steps
1.
2.
Note
To avoid stretching or bending the battery
contacts, connect leads to the base of the
contacts as shown in Figure 4.
3.
Temperature
Note
Multimeter
Note
1.
2.
1.
3.
Set the calibrator for 0 V dc. Put the Meter in the 4.000
V dc range.
2.
4.
5.
3.
6.
5.
16
Multimeter with Temperature
Table 3. Performance Tests
Meter Range
Test No.
1.
2.
3.
4.
Input To Meter
Switch
Position
Display
Minimum
Display
Maximum
0.003
0.000
Center
0V
4.000 V ac
1
0.040
0.000
Center
0V
4.000 V ac, MIN MAX
2
0.013
0.007
Center
10 mV, 50 Hz
4.000 V ac
3
0.013
0.007
Center
10 mV, 400 Hz
4.000 V ac
4
3.570
3.430
Right
3.5 V, 50 Hz
4.000 V ac
5
3.570
3.430
Center
3.5 V, 400 Hz
4.000 V ac
6
35.70
34.30
Center
35 V, 50 Hz
40.00 V ac
7
35.70
34.30
Right
35 V, 400 Hz
40.00 V ac
8
357.0
343.0
Right
350 V, 80 Hz
400.0 V ac
91
614
586
Center
600 V, 400 Hz
600 V ac
10
-0.012
0.012
11
4.000 V dc, MIN MAX
0V
Center
4.000 V dc
0V
Center
-0.002
0.002
12
13
40.00 V dc
+20 mV dc
Center
00.01
00.03
Center
-000.3
-000.1
14
400.0 V dc
-200 mV dc
15
4.000 V dc
+3.5 V
Center
+3.466
+3.534
-35.33
-34.67
16
40.00 V dc
-35 V dc
Center
17
400.0 V dc
+100 V dc
Center
+99.0
+101.0
-606
-594
18
600 V dc
-600 V dc
Center
0.2
Right
-0.2
200 A dc
0.0 A, dc
19
9.7
10.3
10.0 A, dc
Right
20
200 A dc
-10.3
-9.7
-10.0 A, dc
Right
21
200 A dc
202.2
Right
197.8
200 A dc
200.0 A, dc
22
-197.8
Right
-202.2
200 A dc
-200.0 A, dc
23
41.1
Right
38.9
200 A ac
40.0 A, 50 Hz
24
41.1
38.9
200 A ac
40.0 A, 400 Hz
Right
25
195.7
204.3
200.0 A, 50 Hz
Right
26
200 A ac
195.7
204.3
200.0 A, 400 Hz
Right
27
200 A ac
1.980
+2 V dc2
Right
2.020
28
Continuity
3.3973
Right
3.4633
292
Continuity, Automatic Selection
+3.43 V dc2
Right
Continuity, Automatic Selection
302
-0.5 V dc2
-0.5073
-0.4933
Right
400.0
31
0.0
0.0
0.2
Right
400.0
32
1.0
0.8
1.2
Right
400.0
33
350.0
346.6
353.4
Right
4.000 k
34
1.0 k
0.990
1.010
Right
40.00 k
35
35 k
34.67
35.33
Right
400.0 k
36
100 k
99.0
101.0
Right
4.000 M
37
2.9 M
2.873
2.927
Right
40.00 M
38
35 M
34.44
35.56
Right
, Automatic Selection
39
1.96 V dc
1.9403
1.9803
Right
1.000 F
404
0.0 F
-0.001 F
0.001 F
Right
1.000 F
0.95 F
0.930 F
0.970 F
414
In the Automatic Selection mode, the Meter uses a low-impedance thermistor (~2.5k for circuit protection and load testing
(referred to as low-Z input circuitry). When using the 5100B, 5500A, or 5700A to drive the Meter with high voltages, avoid an
overload/current limit condition by gradually stepping the voltage up (waiting two seconds between each step) from 90.0 V,
120.0 V, 180.0 V, and 350.0 V at 80 Hz each step.
Calibrator 50 divider override.
The dc volts annunciator must be on.
Conducting performance tests of the 400 , 4 k, 40 k, and 1 F ranges (tests no. 22, 23, 24, 25, 26, 31, and 32) verifies
that the discrete and integrated circuitry needed to support the other capacitance ranges are working within specifications.
Therefore, the tests indirectly verify that the Meter will meet specification in the 10 F, 100 F, 1000 F, and 10,000 F
ranges.
Display Readings
C
0.0 0.8
23.0 1.0
100.0 1.8
F
32.0 1.5
73.4 2.2
212.0 3.6
Calibration Adjustments
Temperature
AK-80
5700A
Red
Yellow
Copper
+
Ice Bath
Banana
Plug
ig6f.eps
+
+
Input Receptacles
9V
BATTERY
ig1f.eps
16
Multimeter with Temperature
Temperature
4.
5.
Note
Adjust DC voltage (R4) before adjusting
temperature (R38).
The following describes three temperature calibration
procedures:
2.
3.
4.
5.
2.
3.
4.
5.
2.
3.
10
Replacement Parts
Replacement parts are listed in Table 4 and shown in
Figure 1.
Table 4. Replacement Parts
Parts
Part No.
Window decal
Case, top
LCD
Connection elastomeric
Connection Elastomeric LCD to PCA
Switch assembly
Actuator, Switch
Support, Switch Assembly
Battery
Case, Bottom assembly
Screw, THD Form, PH.P.STL, 5014 x .750
Foot
617974
617966
643541
867242
867247
618022
203445
879031
696534
618097
832246
885884
Replacement Parts
45
V
1000V
750V
10A
mV
FUSED
MAX
CAL
ENABLE
100
mA
COM
500mA
F 250V
ALL INPUTS
1kV MAX
REL
FREQ
AUTO
COMP
HI
LO
dB
2ND
REF#
REF
HOLD
MN MX
RATE
THRESH
ADDR
BAUD
LOCAL
POWER
80T-150U
+
+
+
9V
BATTERY
R38
ig4f.eps
Thermometer
+
80AK +
R38
9V
BATTERY
Thermocouple
ig5f.eps
11
16
Multimeter with Temperature