Quality Control Area

1. Chemical Testing Room : 20 Square Meter


2. Instrument Room

: 25 Square Meter

3. Microbiology Room

: 25 Square Meter

Total Area

: 70 Square Meter

Chemical Testing Room Provision :


Laboratories designed , constructed and maintained prevent
Entry of insects and rodents besides cross contamination
Interior surface (walls, floor, and ceilings) smooth free from cracks,
Easy cleaning and disinfection Joints Coved wall to wall ceiling to wall floor to
wall
utilities like Regular supply DM water for Washing Laboratory Accessories
Glassware Etc.
Sampling Area of water Testing Purpose
Hand washing area with drier
ROOM AIR with AIR CONDITION
Temperature 20- 22 C
RH 45-60
Lighting & ventilation
Drainage System Back flow
Construction Materials resistant from acid, alkali, solvent
Separate storage facilities Test samples, retained samples ,reagents and laboratory
accessories , reference substances and reference materials, Records
,Etc..securely locked.

Secure Waste Disposal Area

First aid kit and Fire fighting equipments located at the right places
Fire fighting equipment fire extinguishers, fire blankets and gas masks,
Instrument Room Provision

Laboratories designed , constructed and maintained prevent


Entry of insects and rodents besides cross contamination
Interior surface (walls, floor, and ceilings) smooth free from cracks,
Easy cleaning and disinfection Joints Coved wall to wall ceiling to wall floor to
wall
Lighting & ventilation
Construction Materials resistant from acid, alkali, solvent
Separate storage facilities samples, retained samples ,reagents and laboratory
accessories , reference substances and reference materials Records
,Etc..securely locked.
Hand washing area with drier

AIR CONDITION PROVIDED


Temperature 20- 22 C
RH 45-60
First aid kit and Fire fighting equipments located at the right places
Fire fighting equipment fire extinguishers, fire blankets and gas masks,

Microbiology ROOM Provision :


Laboratories designed , constructed and maintained prevent
Entry of insects and rodents besides cross contamination

Interior surface (walls, floor, and ceilings) smooth free from cracks,
Easy cleaning and disinfection Joints Coved all to wall ceiling to wall floor to
wall
to avoid mix ups, contamination and cross-contamination. samples, reference
organisms, media testing and records. separate storage locations
AIR LOCK
Separate AHU,
LAMINAR AIR FLOW WORK STATION
Temperature 20- 22 C
RH 45-60
Air pressure 15 pa

Secure Waste Disposal Area

Hand washing area with drier

General Condition :
5 People Working area
Electrical Points provided for Equipments
Esd chair or SS Stool
Computer Printers
Each room log book providing facility