Professional Documents
Culture Documents
1 Equipment
Four primary components of the EDS setup are
1. the excitation source (electron beam or x-ray beam)
2. the X-ray detector
3. the pulse processor
4. the analyzer.
EDS spectrum of the mineral crust of the vent shrimp Rimicaris
exoculata[1] It shows peaks caused by X-rays given o as electrons return to the K electron shell and one peak resulting from
the L shell of iron.
5 SEE ALSO
EDS is often contrasted with its spectroscopic counter1. Minimizing SEM beam current allowing for optipart, WDS (wavelength dispersive X-ray spectroscopy).
mization of imaging under analytical conditions,
WDS diers from EDS in that it uses the diraction of X2. Reduced sample damage and
rays on special crystals to separate its raw data into spectral components (wavelengths). WDS has a much ner
3. Smaller beam interaction and improved spatial resspectral resolution than EDS. WDS also avoids the probolution for high speed maps.
lems associated with artifacts in EDS (false peaks, noise
from the ampliers, and microphonics). In WDS, only
one element can be analyzed at a time, while EDS gath- Where the X-ray energies of interest are in excess of ~
ers a spectrum of all elements, within limits, of a sample. 30 keV, traditional silicon-based technologies suer from
poor quantum eciency due to a reduction in the detector stopping power. Detectors produced from high density semiconductors such as cadmium telluride (CdTe)
3 Accuracy of EDS
and cadmium zinc telluride (CdZnTe) have improved
eciency at higher X-ray energies and are capable of
Accuracy of EDS spectrum can be aected by various room temperature operation. Single element systems,
factors. Many elements will have overlapping peaks (e.g., and more recently pixelated imaging detectors such as
Ti K and V K, Mn K and Fe K). The accuracy of the the HEXITEC system, are capable of achieving energy
spectrum can also be aected by the nature of the sample. resolutions of the order of 1% at 100 keV.
X-rays can be generated by any atom in the sample that
is suciently excited by the incoming beam. These X- In recent years, a dierent type of EDS detector, based
rays are emitted in any direction, and so they may not all upon a superconducting microcalorimeter, has also beescape the sample. The likelihood of an X-ray escaping come commercially available. This new technology comthe specimen, and thus being available to detect and mea- bines the simultaneous detection capabilities of EDS
sure, depends on the energy of the X-ray and the amount with the high spectral resolution of WDS. The EDS
and density of material it has to pass through. This can microcalorimeter consists of two components: an abresult in reduced accuracy in inhomogeneous and rough sorber, and a superconducting transition-edge sensor
(TES) thermometer. The former absorbs X-rays emitted
samples.
from the sample and converts this energy into heat; the
latter measures the subsequent change in temperature due
to the inux of heat. The EDS microcalorimeter has his4 Emerging technology
torically suered from a number of drawbacks, including
low count rates and small detector areas. The count rate
There is a trend towards a newer EDS detector, called is hampered by its reliance on the time constant of the
the silicon drift detector (SDD). The SDD consists of a calorimeters electrical circuit. The detector area must be
high-resistivity silicon chip where electrons are driven to small in order to keep the heat capacity small and maxia small collecting anode. The advantage lies in the ex- mize thermal sensitivity (resolution). However, the count
tremely low capacitance of this anode, thereby utilizing rate and detector area have been improved by the impleshorter processing times and allowing very high through- mentation of arrays of hundreds of superconducting EDS
put. Benets of the SDD include:
microcalorimeters, and the importance of this technology
is growing.
1. High count rates and processing,
2. Better resolution than traditional Si(Li) detectors at
high count rates,
3. Lower dead time (time spent on processing X-ray
event),
4. Faster analytical capabilities and more precise X-ray
maps or particle data collected in seconds,
5 See also
Elemental mapping
Transmission electron microscopy
Wavelength dispersive X-ray spectroscopy
3
X-ray uorescence
X-ray microtomography
Scanning electron microscopy
References
[1] Corbari, L; et al. (2008). Iron oxide deposits associated with the ectosymbiotic bacteria in the hydrothermal
vent shrimp Rimicaris exoculata (PDF). Biogeosciences
5: 12951310. doi:10.5194/bg-5-1295-2008.
[2] Joseph Goldstein (2003). Scanning Electron Microscopy
and X-Ray Microanalysis. Springer. ISBN 978-0-30647292-3. Retrieved 26 May 2012.
[3] Joseph Goldstein (2003). Scanning Electron Microscopy
and X-Ray Microanalysis. Springer. ISBN 978-0-30647292-3. Retrieved 26 May 2012.
External links
MICROANALYST.NET Information portal with
X-ray microanalysis and EDX contents
-EDS on the SEM: Primer discussing principles, capabilities and limitations of EDS with the SEM
8.1
Text
8.2
Images
8.3
Content license