Professional Documents
Culture Documents
INTRODUCTION
RELATED INFORMATION
3-15-07
3-15-07
Using the graph paper, plot the meter reading on the X axis
and the percent concentrate reading on the Y axis. This
plotted curve will serve as the known baseline for the
proportioning test. Meters may have different scales and
the plotted resolution should be as large as possible with
the complete range of the calibration solutions known. The
scale you set must be linear.
Collect foam solution sample from the proportioning
system, using care to ensure that the sample is taken at an
adequate distance downstream from the proportioner being
tested. (Usually any place downstream of the first change in
direction is adequate if samples are taken in the riser
room.) Stabilize a water flow through your discharge device
within the known acceptable range for the proportioner and
open the concentrate control valve. Take the sample about
30 to 60 seconds after foam appears from the test connection. Take meter readings of the samples and compare
them to the plotted curve to determine the percentage of
the samples.
Per NFPA 11, Standard for Low-, Medium-, and HighExpansion Foam, acceptable ranges of proportioning
systems are not less than the rated concentration and not
more than 30% above the rated concentration or one
percentage point above the rated concentration, whichever
is less. For example, the acceptable range for a 3%
concentrate is from 3 to 3.9%.
Note: There is some variability dependant on the accuracyy
of the meter, the accuracy to which the calibration samples
were made, and if temperatures between calibration standards and discharge samples have not normalized.
Conductivity is especially susceptible to reading fluctuation,
thus the sample should be continuously stirred until a stabilized reading can be estimated. Care should be taken to
not bump the sides or bottom of the sample container or
contaminate the sample with salts or solids, i.e. dirty fingers
while stirring. The conductivity of water stored in or for
systems can vary throughout a discharge which may affect
results. Conductivity should not be used for sea, salt, or
brackish water supplies. Although conductivity can be the
most accurate means of measuring foam solution in clean
water supplies under ideal conditions, with the stated
reasons A N S U L recommends that the refractive index
method is best for most real world testing.
If discharge samples are sent to ANSULs Lab for testing,
they should be packaged per the instructions with the Test
Kit and sent in for immediate processing. Largely dependent on water quality and the type of concentrate used,
foam solutions may start to biodegrade in a short period of
time which can affect results (possibly as little as three to
five days with nutrient-rich water and detergent-based foam
concentrates).
The following graphs are examples of typical refractive
index values for various concentrate pre-mixes. Results
may vary depending on the concentrate pre-mix type, site
water supply, and actual concentrate lot number. Therefore,
it is important that a new calibration curve be developed on
each date a proportioning test is required.
006876
NOTES:
006912
3-15-07
007596
3-15-07
4. Slowly remove the vertical center tube out through the top
flange opening. This tube must be removed on all horizontal tanks. (If the center tube cannot be completely
removed on the vertical tanks, drop the tube through the
lower flange opening and remove bladder out through the
top flange opening sliding the bladder past the center
tube.)
Remember the vent port(s) must be left open to atmosphere and not plugged, otherwise the valve will not open.
3-15-07
If the in label has been removed, with the valve in the closed
position, put a finger over both ports and manually open the
valve with a wrench. The port(s) that is pulling a suction is the
port(s) to pipe the actuation piping to.
METER READING
FOAM CONCENTRATE
WATER
PRE-MIX #1
PRE-MIX #2
PRE-MIX #3
CALIBRATION CURVE
METER READING
METER READING
715-735-7411
3-15-07
ANSUL INCORPORATED
MARINETTE, WI 54143-2542