Optimus 50/65/80

Specifications

The solution
for any application ...
The X-ray generator for high end hospital and clinical applications.
The design incorporates high versatility and excellent adaptation to all
requirements of radiographic and fluoroscopic applications.
The Optimus is your ideal partner for radiography and fluoroscopy systems.

Optimus
in the DigitalDiagnost

X-ray generator Optimus


The Optimus X-ray generator for radiography
and fluoroscopy is easy to adapt to user modern
requirements, making procedures fast and
highly customized. This is ensured by up to
1,000 APRF programs (Anatomically
Programmed Radiography and Fluoroscopy),
which can either be defaults or can be
programmed by you. That also applies to all
routine applications with old and new X-ray
systems in radiology.

X-ray generator

Optimus 50 / 65 / 80

Applications
Routine radiography

Tomography

Digital radiography PCR

Therapy simulation

(Philips Computed Radiography)


Techniques
Image intensifier TV technique

Fluorography

DSI (Digital Spot Imaging)


Auxiliaries
Units

up to eight

Interface possibility

System CAN-signal bus and / or decade adapter

Note:

This specification describes Version 3 of the software.The technical data conform to IEC 60601-2-7.

Technical data
Power supply

Optimus 50

Mains voltage

Optimus 65

Optimus 80

400 V (10%); 50 Hz or 60 Hz, 3-phase


If the generators are connected up to mains which is not 400 V (10%),
a mains adaptation transformer is required.

Max. mains resistance at 400 V

0,3

0,2

0,2

Max. mains current at 400 V

145 A

190 A

230 A

Fuses

35 A (slow)

50 A (slow)

50 A (slow)

X-ray generator
High-voltage generator

The converter generator generates high voltage equivalent to DC voltage

Nominal power (IEC)

50 kW

65 kW

80 kW

Max. tube voltage

150 kV

150 kV

150 kV

Max. tube current (at 70 kV)

650 mA

900 mA

1100 mA

Max. frame rates


Compatible X-ray tubes

6 fps for camera technique and DSI


Up to 3 double-focus tubes with a total of 6 different focal spots can
be used. SRO- and RO tube assemblies are compatible.

Tube load protection

The tube is thermally monitored by the tube load protection system.


If the load is high, tube output is reduced; exposured being made are
not aborted.The tube load display on the generator control desk
indicates the tube power available.

Nominal tube output


Ratings

650 mA at 70 kV

900 mA at 70 kV

1100 mA at 70 kV

625 mA at 80 kV

812 mA at 80 kV

1000 mA at 80 kV

500 mA at 100 kV

650 mA at 100 kV

800 mA at 100 kV

400 mA at 125 kV

520 mA at 125 kV

640 mA at 125 kV

333 mA at 150 kV

433 mA at 150 kV

533 mA at 150 kV

Ambient conditions
Temperature

+10 C to +40 C

Rel. humidity

15% to 90%, without condensation

Rel. atmospheric pressure

50 kPa to 110 kPa

Sound emission

55 dBa max., averaged over 1 h, at a distance of 1 m

Ventilation demand

2.5 l/s max.

Radiography and Fluoroscopy


Radiography

Optimus 50

Optimus 65

Optimus 80

Exposure techniques:
One-factor technique

1)

kV, continuously falling load, AMPLIMAT

1)

1)

Two-factor technique

kV, mA, constant load, AMPLIMAT

Two-factor technique

kV, mAs, constant load

Three-factor technique

kV, mA, s, constant load

Three-factor technique

kV, mAs, s, constant load

AMPLIMAT-Automatic exposure control

up to 5 ionization measuring chambers or


4 ionization measuring chambers and 1 photomultiplier (secondary electron multiplier) or
3 ionization measuring chambers, 1 photomultiplier and 1 photodiode

Setting ranges
Tube voltage

40 kV to 150 kV programmable in: 27 large steps or


111 small steps of 1 kV each

Radiography without AMPLIMAT:


Tube current

1 mA to 650 mA

2)

1 mA to 900 mA

mAs product
Steps

2)

1 mA to 1100 mA

2)

0.5 mAs to 850 mAs


For tube current and mAs product, programmable as required in:
6.25%, 12.5% or 25% steps

Exposure times

3)

kV-mA-s-technique: you select in steps from 1 ms to 6 s (16 s )


kV-mAs-s-technique: from 1 ms to 6 s the generator switches automatically

Radiography with AMPLIMAT:


mAs product

2)

0.5 mAs to 800 mAs , the maximum can be limited to 600 mAs

Switch-off dose
Exposure

0.5 Gy to 100 Gy
1 ms to 4 s or to 6 s with TDC (Tomographic Density Control)
The display panel on the control desk shows the switched mAs- and
ms values after exposure.

Screen-film combinations

From a library of approx. 100 film and screen data, 5 screen-film


combinations can be read in.The values of new combinations can be
integrated by Customer Service at any time.

2)

The ultimate value depends on the particular tube nomogram.

3)

For special application techniques.

Fluoroscopy
Image intensifier TV system
Tube voltage
Tube current

1
40 kV to 110 kV (programmable up to 125 kV on request)
0.1 mA to 6 mA (maximum selectable between 3 mA and 6 mA)

Quick fluoroscopy
Tube voltage
Tube current

40 kV to 110 kV (programmable up to 125 kV on request)


0.1 mA to 6 mA (maximum selectable between 3 mA and 6 mA)

Adjustment times:
Standard fluoroscopy

0.5 s to 1,0 s (dependent on absorption)

Quick fluoroscopy

0.2 s to 0.4 s (dependent on absorption)


4

Options
Tomography

Optimus 50 / 65 / 80

Automatic tomographic time input:

for entering a tomographic figure and angle patient-side at the auxiliary

Free tomography

kV and mA can be selected at the control desk direct

APRF-technique

kV, mA, tomographic figure incl. time are set when the APRF programme
is selected (Anatomiically Programmed Radiography / Fluoroscopy)

TDC
(Tomographic Density Control)

The automatic TDC system regulates dose rate for tomographic exposures with time limits to ensure optimum average image brightness over
the entire tomographic time, even if object cross-sections are different.

Area dose product display

(for BuckyDiagnost only)

Calculation, can be calibrated

Automatic detection of kV, mAs product, film size and, where applicable,
added filters inserted in the collimator.

Display

0.1 cGycm2 to 999999 cGycm2 (max. resolution 0.1 cGycm2)


The display appears on the generator control desk and / or on the
optional PDO user interface.

Number of auxiliaries

Separate detection of area dose product for up to three tubes.

PDO (Patient Data Organizer)


Features of the PDO software

Digital dokumentation of patient and examination data


Data import and export between PDO and RIS
Input of patient data via barcode reader, floppy disk or keyboard
Display and dokumentation of the area dose product for up to
1)

three tubes (calculated and / or measured area dose product )


Printout of patient and examination data with user-defined print layout
Data output for invoicing or statistical purposes
Hardware and software requirements Standard PC, 486 or better and Windows 3.1 or better
Memory (RAM) 8 Mbyte or better
Adequate number of serial / parallel interfaces, cable
1)

To measure area dose product a Diamentor M4 measuring chamber incl. separate control module is also required.

Variofocus
Variofocus technique

By superimposing the small and large focal spots it is possible to


generate several focuses of variable size.The new focus sizes are
between the small and large focus.

Features

Uniform loading of the focuses lengthens tube life


Focus size is adapted to the object features
Application-oriented focuses are assigned to the APRF programmes
Maximum image definition due to optimum geometric imaging
conditions.

Dimensions and weights


Flat-panel control desk

Control cabinet

Free space required for air circulation

20
60

550
430
200

335

3)

590

260

20
202

1)

1955
2160

430

1290

480

The flat-panel control desk can also be attached


to the wall using a simple wall bracket.

2)

Weight
4 kg
Heat dissipation max. 40 kW
690

Stand for flat-panel control desk

65

520

1)
1) Wall junction support
2) Wall-cabinet spacing angle
3) Fixing points

60

202

60

825

Weight
220 kg
Heat dissipation max. 1.2 kW

All dimensions in mm

Weight
24 kg
(incl. flat-panel control desk)

Philips Medical
Systems is part
of Royal Philips
Electronics
www.medical.philips.com
medical@philips.com
fax: +31 40 27 64 887

Philips Medical Systems


Global Information Center
I.B.R.S. / C.C.R.I.
Numro 11088
5600 VC Eindhoven
Pays-Bas /
The Netherlands
(no stamp required)

Koninklijke Philips
Electronics N.V. 2003
All rights are reserved.

Printed in Germany.
4512 158 10931/742*2003-10

Reproduction in whole or in
part is prohibited without the
prior written consent of the
copyright holder.

Printed on Reviva Mega paper


which is made from 50% pulp
(bleached without the use of
chlorine)

Philips Medical Systems DMC


GmbH reserves the right to
make changes in specifications or
to discontinue any product at any
time without notice or obligation
and will not be liable for any
consequences resulting from the
use of this publication.