Professional Documents
Culture Documents
UCAM-500A
High-Speed
Data Logger
Number
Functions:
To control scanners
To collect and store data sent from scanners
To transfer data to the personal computer
Data update rate: See table below.
Measuring modes: Real-time Interval Trigger Monitor
Measuring functions: Initial value Original value True value
Range of initial value available for storage: Same as the measuring
range of scanner
Conversion to values in engineering units: Values in engineering units
are obtained by multiplying measured values (strain, voltage,
potentiometer value) by defined coefficients.
Memory: 64M-bytes (backed for approximately 4 years at 20C). It
stores data except for calculation channels.
Interface:
Dedicated interface between UCAM-500A and USB-500A, total
extended cable length up to 185m
GP-IB interface between UCAM-500A and personal computer;
prepare the GP-IB cable which is commercially available.
Real-time clock: Year, month, day, hour, minute, second (for interval
measurement, etc.; backed for approximately 8 years at 20C)
Operating temperature range: 0 to 40C
Operating humidity range: 20 to 80%R.H. (no dew condensing)
Power requirements: 85 to 264Vac (no switchover required), 50/60Hz;
approx. 50VA
Dimensions: 426 x 88 x 300 mm (excluding protrusions)
Mass: Approx. 6kg
sensors
(1) Strain unit USS-51A
Strain gages/strain-gage transducers
1-gage method/1-gage method (true dummy method):
120, 350
2-gage method (active-dummy & active-active methods):
120 to 1000
4-gage method (active on opposite sides & full bridge methods):
120 to 1000
Potentiometer sensors: 1k to 10k
(2) Voltage unit USV-51A
20.0V
(3) Thermocouple unit UST-51A
Thermocouples: K (CA), T (CC), E (CRC), J (IC), R
Number of measuring channels: 50 max. (with 5 measuring units
mounted)
Frequency response: DC to 7.8Hz (3dB 0.5dB)
Max. Data Update Rate, Number of Measuring Times and Data Acquisition Time vs. Number of Measuring Channels
Max. data update rate
Max. number of
measuring times, approx.
50 times/sec (50Hz)
10 times/sec (10Hz)
2 times/sec (2Hz)
1 time/sec (1Hz)
60,000
30,000
12,000
6,000
20 min.
50 min.
2.1 hr.
1.7 hr.
Number of
measuring channels
100
200
500
1000
Number of scanners
required
2*
4*
10
20
*Synchronous cable US-50A should be used for connection between scanner to scanner.
1 time/sec
33.3 hr.
16.7 hr.
8.3 hr.
5.6 hr.
4.2 hr.
3.3 hr.
1.7 hr..
2.8 hr.
2.1 hr.
1.7 hr.
Number of
measuring times
Number of
measuring channels
120,000
60,000
30,000
20,000
15,000
12,000
6,000
50
100
200
300
400
500
1000
11
UCAM-20PC
Data
Logger
of measuring channels:
30 max. with the dedicated scanner integrated
200 max. with external scanners connected
230 max. with the dedicated scanner integrated and external
scanners connected
Scanning speed:
Comprehensively set for all channels
12.5 channels/sec (0.08 sec/channel)
1 channel/sec (1 sec/channel)
0.2 channel/sec (5 sec/channel)
Interface RS-232-C
Transfer rate 1200, 2400, 4800 or 9600 bps
Operating temperature/humidity range: 0 to 40C, 20 to 85%RH
(no dew condensing)
Power requirements: 90-132/180-264Vac, 50/60Hz, approx. 50VA
or 9-16Vdc, approx. 30W (as specified)
Dimensions & mass:
320 x 40 x 250mm (excluding protrusions), approx. 1.9kg
With a dedicated scanner integrated: 323.6 x 86.7 x 250mm
(excluding protrusions), approx. 6.2kg max.
120
240
350
120
Strain-gage
1-gage method
Transducers
(true dummy method) 350
Active-dummy method
2-gage method
60 to 1000
Active-active method
Common dummy method
Active on opposite sides method
4-gage method
60 to 1000
Full bridge method
Constant-current excitation
Civil Engineering 4-gage method
Transducers
350
Transducer with temp. measuring function
DC voltage-output sensor
Voltage
Thermocouple
K (CA)
Temperature
T (CC)
E (CRC)
J (IC)
R
Pt100 (new JIS)
Resistance-based
temp. measurement
JPt100 (old JIS)
Thunder-resistant terminal
Scanner Interface
12
Civil engineering
General purpose
USB USB USB USB USB USB USB USB
-21A -70A -51A -51AT -20A -23A -70A -20D
-10/20
-50A
-30 -50D
**
USI
-27A
USI-25A
** USI USI
-27A -25A
**Not required
UCAM-70A
Data
Logger
Connectable
to a variety of sensors
scanner provides 10 measuring channels as standard
and is extensible to a maximum of 30 channels.
External scanners provide 50 measuring channels each and
allow a maximum of 1000 measuring channels.
High-speed measurement, 20 channels/sec
Civil engineering model available
Graphic display available
Integral high-speed printer (14 lines/sec)
Onscreen image can be hardcopied.
Variety of useful functions
Data can be saved on the floppy disk or 1Mbyte PC memory card.
GP-IB and EIA-232-D (RS-232-C) interface as standard provisions
Internal
Specifications UCAM-70A
Models:
Processing:
Other functions:
Power requirement:
Dimensions and mass:
Scanner USB-70A
Measuring channels:
Subject variables
Strain
Gage
Transducer
DC Voltage/
Current
Remarks
Transducer
Connector
Thunder
Arrester
UCAM-70A-10
UCAM-70A-20
UCAM-10B Data
Logger
SMD-10A/20A
High-speed
measurement, 20 channels/sec
to a variety of sensors
GP-IB and self-test function as standard provisions
Subject variables: Strain, DC voltage/current, temperature (thermocouple)
Measuring channels: 500 max.; internal scanner for 10 channels;
10 external scanners connectable, each providing 50 measuring
channels
External scanners: USB-51A, USB-51AT, USB-20A, USB-50A
Measurement can be triggered by a relative value (a certain changing value); 30 types of trigger table available
A constant can be individually set for all measuring points in a
range of 0.0001 x 107 to 99999 x 107 or 0.001 x 107 to 9999 x
107
Interval measurement possible
Power requirements
100V version: 85-132Vac, 50/60Hz; approx. 35VA
220V version: 170-264Vac
DC version: 11-30Vdc, approx. 20W
Dimensions and mass: 425 x 88 x 421.5mm, approx. 8kg
Connectable
Handy Digital
Static Strain Indicators
(RS-232-C)
UCAM-1A
Built-in
Handy
Static Strain Meter
channels each
requirement: R6P dry cell x 8 for USB-11A;
USB-12A is supplied from USB-11A.
Dimensions and mass:
USB-11A: 300 x 55 x 140mm,
approx. 1.4kg
USB-12A: 300 x 55 x 70mm,
approx. 1kg
Power
13