You are on page 1of 32

3300 XL 8mm Proximity Transducer System

Bently Nevada* Asset Condition Monitoring


Description
The 3300 XL 8 mm Proximity Transducer System consists of:
a 3300 XL 8 mm probe,
a 3300 XL extension cable1, and
a 3300 XL Proximitor* Sensor2.
The system provides an output voltage that is directly proportional to the
distance between the probe tip and the observed conductive surface and can
measure both static (position) and dynamic (vibration) values. The systems
primary applications are vibration and position measurements on fluid-film
bearing machines, as well as Keyphasor* reference and speed measurements3.
The 3300 XL 8 mm system delivers the most advanced performance in our eddy
current proximity transducer systems. The standard 3300 XL 8 mm 5-metre
system also fully complies with the American Petroleum Institutes (API) 670
Standard (4th Edition) for mechanical configuration, linear range, accuracy, and
temperature stability. All 3300 XL 8 mm proximity transducer systems provide
this level of performance and support complete interchangeability of probes,
extension cables, and Proximitor sensors, eliminating the need to match or
bench calibrate individual components
Each 3300 XL 8 mm Transducer System component is backward-compatible and
interchangeable4 with other non-XL 3300 series 5 mm and 8 mm transducer
system components5. This compatibility includes the 3300 5 mm probe, for
applications in which an 8 mm probe is too large for the available mounting
space6,7.
Proximitor Sensor
The 3300 XL Proximitor Sensor incorporates numerous improvements over
previous designs. Its physical packaging allows you to use it in high-density DIN-
rail installations. You can also mount the sensor in a traditional panel mount
configuration, where it shares an identical 4-hole mounting footprint with older
Proximitor sensor designs. The mounting base for either option provides
electrical isolation and eliminates the need for separate isolator plates. The 3300
XL Proximitor Sensor is highly immune to radio frequency interference, allowing
you to install it in fiberglass housings without adverse effects from nearby radio
frequency signals. The 3300 XL Proximitor Sensors improved RFI/EMI immunity
satisfies European CE mark approvals without requiring special shielded conduit
or metallic housings, resulting in lower installation costs and complexity.
The 3300 XLs SpringLoc terminal strips require no special installation tools and
facilitate faster, more robust field wiring connections by eliminating screw-type
clamping mechanisms that can loosen.

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 1 of 32
Notes:
Proximity Probe and Extension Cable
1. 1-metre systems do not use an extension cable.
The 3300 XL probe and extension cable also reflect
improvements over previous designs. A patented 2. Proximitor sensors are supplied by default from the
TipLoc* molding method provides a more robust factory calibrated to AISI 4140 steel. Calibration to
bond between the probe tip and the probe body. other target materials is available upon request.
The probes cable incorporates a patented CableLoc* 3. Consult Bently Nevada* Applications Note,
design that provides 330 N (75 lbf) pull strength to Considerations when using Eddy Current Proximity
more securely attach the probe cable and probe tip. Probes for Overspeed Protection Applications, when
considering this transducer system for tachometer or
You can also order 3300 XL 8 mm probes and overspeed measurements.
extension cables with an optional FluidLoc* cable
4. 3300 XL 8 mm components are both electrically and
option. This option prevents oil and other liquids
physically interchangeable with non-XL 3300 5 mm
from leaking out of the machine through the cables and 8 mm components. Although the packaging of
interior. the 3300 XL Proximitor Sensor differs from its
Connectors predecessor, its design fits in the same 4-hole
mounting pattern when used with the 4-hole
The 3300 XL probe, extension cable, and Proximitor mounting base, and will fit within the same mounting
sensor have corrosion-resistant, gold-plated space specifications (when minimum permissible
ClickLoc* connectors. These connectors require only cable bend radius is observed).
finger-tight torque (the connectors will "click" when 5. Mixing XL and non-XL 3300-series 5 mm and 8 mm
tight), and the specially-engineered locking system components limits system performance to the
mechanism prevents the connectors from loosening. specifications for the non-XL 3300 5 mm and 8 mm
These connectors require no special tools for Transducer System.
installation or removal. 6. The 3300-series 5 mm probe (refer to Specifications
You can order the 3300 XL 8 mm probes and and Ordering Information p/n 141605-01) uses
smaller physical packaging, but does not reduce the
extension cables with connector protectors already side view clearances or tip-to-tip spacing
installed. We can also supply connector protectors requirements as compared to an 8 mm probe. It is
separately for field installations (such as when an used when physical (not electrical) constraints
application must run the cable through restrictive preclude the use of an 8 mm probe. When your
conduit). We recommend connector protectors for application requires narrow side view probes, use the
all installations to provide increased environmental 3300 NSv* Proximity Transducer System (refer to
protection8. Specifications and Ordering Information p/n 147385-
01).
Extended Temperature Range Applications
7. 8 mm probes provide a thicker encapsulation of the
An extended temperature range (ETR) probe and ETR probe coil in the molded PPS plastic probe tip. This
extension cable are available for applications in results in a more rugged probe. The larger diameter
which either the probe lead or extension cable may of the probe body also provides a stronger, more
exceed the 177 C (350 F) temperature robust case. We recommend that you use 8 mm
probes when possible to provide optimal robustness
specification. The ETR probe has an extended
against physical abuse.
temperature rating for up to 260 C (500 F) for the
probe lead and connector, although the probe tip 8. Each 3300 XL extension cable includes silicone tape
must remain below 177 C (350 F). The ETR that you can use instead of connector protectors. We
do not recommend silicone tape for applications that
extension cable rating is also up to 260 C (500 F). will expose the probe-to-extension cable connection
Both the ETR probe and cable are compatible with to turbine oil.
standard temperature probes and cables., so that,
for example, you can utilize an ETR probe with the
330130 extension cable. The ETR system uses the
standard 3300 XL Proximitor Sensor. Note that when
you use any ETR component as part of your system,
the ETR component limits the system accuracy to
the accuracy of the ETR system.

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 2 of 32
Specifications Probe Length RPROBE ()

Unless otherwise noted, the following specifications 2.0 7.88 0.50


are for a 3300 XL 8 mm Proximitor Sensor, extension
5.0 8.73 0.70
cable and 8 mm probe between +18 C and +27 C
(+64 F to +80 F), with a -24 Vdc power supply, a 10 9.0 9.87 0.90
k load, an AISI 4140 steel target, and a probe
gapped at 1.27 mm (50 mils). Performance
characteristics apply to systems that consist solely Nominal
of 3300 XL 8 mm components. The system accuracy Extension Cable
and interchangeability specifications do not apply to DC Resistance
transducer systems that are calibrated to any target
See Table 2 and Table 3.
other than our AISI 4140 steel target.
Table 2: Resistance from Center Conductor to
Electrical Center Conductor (RCORE)
Proximitor
Sensor Input Length of Extension
RCORE ()
Cable (m)
Accepts one noncontacting 3300-
series 5 mm, 3300 8 mm or 3300 3.0 0.66 0.10
XL 8 mm Proximity Probe and 3.5 0.77 0.12
Extension Cable.
4.0 0.88 0.13
Power
4.5 0.99 0.15
Requires -17.5 Vdc to -26 Vdc
without barriers at 12 mA 7.0 1.54 0.23
maximum consumption, -23 Vdc
7.5 1.65 0.25
to -26 Vdc with barriers.
Operation at a more positive 8.0 1.76 0.26
voltage than -23.5 Vdc can result
in reduced linear range. 8.5 1.87 0.28

Supply
Sensitivity Table 3: Resistance from Outer Conductor to Outer
Conductor (RJACKET)
Less than 2 mV change in output
voltage per volt change in input Length of Extension
voltage. RJACKET ()
Cable (m)
Output 3.0 0.20 0.04
Resistance
3.5 0.23 0.05
50
4.0 0.26 0.05
Nominal Probe
DC Resistance 4.5 0.30 0.06
See Table 1. 7.0 0.46 0.09
Table 1: Resistance from Center Conductor to 7.5 0.49 0.10
Outer Conductor (RPROBE)
8.0 0.53 0.11
Probe Length RPROBE () 8.5 0.56 0.11
0.5 7.45 0.50
1.0 7.59 0.50
1.5 7.73 0.50
Specifications and Ordering Information
Part Number 141194-01
Rev. L (07/09)

Page 3 of 32
Extension Cable
Capacitance
69.9 pF/m (21.3 pF/ft) typical
Extended
Field Wiring Temperature
Range (ETR) for
0.2 to 1.5 mm2 (16 to 24 AWG) .
5- and 9-Metre
Recommend using 3-conductor
Systems
shielded triax cable and tinned
field wiring. Maximum length of 7.87 V/mm (200 mV/mil) 6.5%
305 metres (1,000 feet) between including interchangeability error
the 3300 XL Proximitor Sensor when measured in increments of
and the monitor. See the 0.25 mm (10 mils) over the 80 mil
frequency response graphs in linear range from 0 C to +45 C
through Figure 13 (pages 25 and (+32 F to +113 F).
26) for signal rolloff at high
Deviation from best fit straight line (DSL)
frequencies when using longer
field wiring lengths. Standard 5- or
1-metre System
Linear Range
Less than 0.025 mm (1 mil) with
2 mm (80 mils). Linear range
components at 0 C to +45 C
begins at approximately 0.25 mm
(+32 F to +113 F).
(10 mils) from target and is from
0.25 to 2.3 mm (10 to 90 mils) Standard
(approximately 1 to 17 Vdc). 9-metre System
Recommended Less than 0.038 mm (1.5 mil)
Gap Setting for with components at 0 C to +45
Radial Vibration C (+32 F to +113 F).
-9Vdc [approximately 1.27 mm Extended
(50 mils)] Temperature
Range 5 and
Incremental
9-metre
Scale Factor
Systems
(ISF)
Less than 0.038 mm (1.5 mil)
with components at 0 C to +45
Standard 5- or C (+32 F to +113 F).
1-metre System
7.87 V/mm (200 mV/mil) 5%
Standard 5- or
including interchangeability error
1-metre System
when measured in increments of
Performance
0.25 mm (10 mils) over the 80 mil
Over Extended
linear range from 0 C to +45 C
Temperatures
(+32 F to +113 F).
Over a probe temperature range
Standard
of 35 C to +120 C (-31 F to
9-metre System
+248 F) with the Proximitor
7.87 V/mm (200 mV/mil) 6.5% sensor and extension cable
including interchangeability error between 0 C to +45C (+32 F to
when measured in increments of +113 F), the ISF remains within
0.25 mm (10 mils) over the 80 mil 10% of 7.87 V/mm (200 mV/mil)
linear range from 0 C to +45 C and the DSL remains within
(+32 F to +113 F). 0.076 mm (3 mils).
Specifications and Ordering Information
Part Number 141194-01
Rev. L (07/09)

Page 4 of 32
Over a Proximitor sensor and Frequency
extension cable temperature Response
range of 35 C to +65 C (-31 F
(0 to 10 kHz), +0, -3 dB, with up to
to +149 F) with the probe
305 metres (1000 feet) of field
between 0 C to +45 C (+32 F to
wiring.
+113 F), the ISF remains within
10% of 7.87 V/mm (200 mV/mil) Minimum
and the DSL remains within Target Size
0.076 mm (3 mils).
15.2 mm (0.6 in) diameter (flat
Standard target)
9-metre System
Shaft Diameter
Performance
Over Extended Minimum
Temperatures
50.8 mm (2 in)
Over a probe temperature range
Recommended
of 35 C to +120 C (-31 F to
Minimum
+248 F) with the Proximitor
sensor and extension cable 76.2 mm (3 in)
between 0 C to +45C (+32 F to
When gapped at the center of the
+113 F), the ISF remains within
linear range, the interaction
18% of 7.87 V/mm (200 mV/mil)
between two separate transducer
and the DSL remains within
systems (cross-talk) will be less
0.152 mm (6 mils).
than 50 mV on shaft diameters of
Over a Proximitor sensor and at least 50 mm (2 in) or greater.
extension cable temperature You should take care to maintain
range of 35 C to +65 C (-31 F minimum separation of
to +149 F) with the probe transducer tips, generally at least
between 0 C to +45 C (+32 F to 40 mm (1.6 in) for axial position
+113 F), the ISF remains within measurements or 38 mm (1.5 in)
18% of 7.87 V/mm (200 mV/mil) for radial vibration measurements
and the DSL remains within to limit cross-talk to 50 mV or less.
0.152 mm (6 mils). Radial vibration or position
measurements on shaft
Extended
diameters smaller than 76.2 mm
Temperature
(3 in) will generally change the
Range System
scale factor.
Performance
Over Extended
Temperatures
Over a probe and extension cable
temperature range of 35 C to
+260 C (-31 F to +500 F) with
the Proximitor sensor between 0
C to +45 C (+32 F to +113 F),
the ISF remains within 18% of
7.87 V/mm (200 mV/mil) and the
DSL remains within 0.152 mm
(6 mils).

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 5 of 32
Effects of 60 Hz North America
Magnetic Fields
3300 XL
up to 300 Gauss
Proximitor
See Table 4. Sensor and
probe, ia:
Table 4: Output Voltage in Mil pp/Gauss
Ex ia IIC T4/T5; Class I Zone 0 or
5- or Class 1 Division 1; Groups A, B, C,
9-metre
1-metre Ext. and D, when installed with
Gap Proximitor Probe
Proximitor Cable intrinsically safe zener barriers
Sensor
Sensor per drawing 141092 or when
10 mil 0.0119 0.0247 0.0004 0.0004 installed with galvanic isolators.
Certificate number 1109248(LR
50 mil 0.0131 0.0323 0.0014 0.0014 26744-222)
90 mil 0.0133 0.0348 0.0045 0.0045
3300 XL
Proximitor
Sensor and
Electrical Classification probe, nA:
Complies with the European CE Ex nA IIC T4/T5 Class I Zone 2 or
mark. Class I, Division 2, Groups A, B, C,
Hazardous Area Approvals and D when installed without
barriers per drawing 140979.
Note: Multiple approvals for hazardous areas certified by
T5 @ Ta = -35 C to +85 C.
Canadian Standards Association (C/US) in North America
T4 @ Ta= -51 C to +100 C.
and by Baseefa for Europe and IEC Ex.
Certificate number 1109248(LR
Table 5: Field Wiring Limitations: 26744-222)
Europe
3300 XL
L/R
Type Gas Capacitance Inductance Proximitor
Ratio
Approval: Group (F) (mH)* Sensor, ia:
(H/)
II 1 G EEx ia IIC T4/T5. EC
ATEX certificate number
and IEC IIC 0.078 0.99 29.2 BAS99ATEX1101, when installed
Zone 0/1 per drawing 141092.
IIB 0.645 7.41 117.0 3300 XL
IIA 2.144 15.6 234.0 Proximitor
CSA A Sensor, nA:
Div 1 and 0.070 1.0 29.2
B II 3 G Ex nA II T4/T5. EC
certificate number Baseefa
C 0.600 5.0 117.0
07ATEX0189X, when installed per
D 2.09 11.0 234.0 drawing 140979.
CSA T5 @ Ta= -35 C to +85 C
All 0.460 100.0 N/A
Div 2 T4 @ Ta= -51 C to +100 C.

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 6 of 32
3300 XL 8mm 3300 XL 8mm
probe, ia: and 3300 5mm
Eddy Current
II 1 G EEx ia IIC, Temperature
Probes, nA
Classification per table 4-5, EC
certificate number IECEx BAS04.0058X
BAS99ATEX1099 when installed
Ex nA II for Zone 2 Temperature
per drawing 142491.
Classification per table 5.
3300 XL 8mm
probe, nA:
Table 6: Probe Ex ia and Ex nA Temperature
II 3 G EEx nA II, Temperature
Classification
Classification per table 4-5, EC
certificate number Temperature Ambient Temperature
BAS99ATEX3098 when installed Classification (Probe Only)
per drawing 142491.
T1 -51C to +232C
IEC Ex
T2 -51C to +177C
3300 XL
Proximitor T3 -51C to +120C
Sensor, ia
T4 -51C to +80C
IECEx BAS04.0055X
T5 -51C to +40C
Ex ia IIC T4 (-51C Ta +100C) /
T5 (-35C Ta +85C)
Hazardous Area
Terminal Block Conditions of
Connections Safe Use:
Ui= -28V Ci = 0
Ii= 140mA Li =10H ATEX: Follow the conditions of
safe use included on the
Pi= 0.84W Declaration of Conformance sent
3300 XL with each product.
Proximitor Canadian Standards Association
Sensor, nA: (CSA):
IECEx BAS04.0057X Division 1 (Intrinsically
Ex nA II T4 (-51C Ta +100C) / safe): Install per Bently Nevada
T5 (-35C Ta +85C) drawing 141092.
Ui = -28V Division 2 (non-Incendive):
Install per Bently Nevada drawing
3300 XL 8mm 140979.
and 3300 5mm
Eddy Current IECEx
Probes, ia: Zone 0 (Intrinsically safe): The
IECEx BAS04.0056 Proximitor Sensor must be
installed to minimize the risk of
Ex ia IIC Temperature impact or friction with other
Classification per table 5. metallic surfaces.
Ui = -28V Ci = 1.5 nF
Li = 140 mA Li = 200 H
Pi 0.84 W
Specifications and Ordering Information
Part Number 141194-01
Rev. L (07/09)

Page 7 of 32
Zone 2 (non-Incendive): The Extension Cable
probe must be supplied from a
Material
voltage-limited source.
75 triaxial, fluoroethylene
propylene (FEP) insulated.
Compliance and Certifications Extended
EMC: Temperature
European Community Directives: Range (ETR)
Extension Cable
EMC Directive 2004/108/EC Material
Standards: 75 triaxial, perfluoroalkoxy (PFA)
EN61000-6-2 insulated.

EN61000-6-4 Proximitor
Sensor Material
American
Bureau of A308 aluminum
Shipping (ABS) System Length
Type Approval
5 or 9 metres including extension
Certification Number: cable, or 1 metre
09-HS446965B-PDA

Mechanical Standard Probe


Probe Tip and Extension
Material Cable Armor
(optional)
Polyphenylene sulfide (PPS).
Flexible AISI 302 or 304 SST with
Probe Case FEP outer jacket.
Material
Extended
AISI 303 or 304 stainless steel Temperature
(SST). Range Probe
Probe Cable and Extension
Specifications Cable Armor
(optional)
Standard cable
Flexible AISI 302 or 304 SST with
75 triaxial, fluoroethylene PFA outer jacket.
propylene (FEP) insulated probe
cable in the following total probe Tensile Strength
lengths: 0.5, 1, 1.5, 2, 5, or 9 (Maximum
metres. Rated)

Extended 330 N (75 lbf) probe case to probe


Temperature lead. 270 N (60 lbf) at probe lead
Range cable to extension cable connectors.

75 triaxial, perfluoroalkoxy (PFA) Connector


insulated probe cable in the Material
following total probe lengths: 0.5, Gold-plated brass or gold-plated
1, 1.5, 2, 5, or 9 metres. beryllium copper.

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 8 of 32
Probe Case Total System
Torque Mass (Typical)
See Table 7. 0.7 kg (1.5 lbm)
Table 7: Probe Case Torque Probe

Maximum 323 g (11.4 oz)


Probe Type Recommended
Rated Extension Cable
Standard 33.9 Nm 11.2 Nm 34 g/m (0.4 oz/ft)
forward-mounted
probes (300 inlbf) (100 inlbf) Armored
Extension Cable
Standard
22.6 Nm 7.5 Nm 103 g/m (1.5 oz/ft)
forward-mount
probes - first (200 inlbf) (66 inlbf) Proximitor
three threads Sensor
Reverse-mount 22.6 Nm 7.5 Nm 246 g (8.67 oz)
probes (200 inlbf) (66 inlbf)
Environmental Limits
Probe Temperature Range
Operating and
Storage
Temperature
Standard Probe
Connector-to-connector torque
-51 C to +177 C (-60 F to +350
Recommended F)
Torque
Extended
See Table 8. Temperature
Table 8: Recommended Connector-to-Connector Range Probe
Torque -51 C to +218 C (-60 F to
+425F) for the probe tip; -51 C to
Connector Type Tightening Instructions
+260 C (-60 F to +500 F) for the
Two 3300 XL gold probe cable and connector.
Finger tight
"click" type connectors Note: Exposing the probe to temperatures below 34 C (-30 F)
may cause premature failure of the pressure seal.
One non-XL stainless
Finger tight plus 1/8 turn
steel connector and Extension Cable Temperature Range
using pliers
one 3300 XL connector
Operating and
Storage
Maximum Temperature:
Connector Standard Cable
Torque
-51 C to +177 C (-60 F to +350
0.565 Nm (5 inlbf) F)
Minimum Cable Extended
Bend Radius Temperature
25.4 mm (1.0 in) Range Cable
-51 C to +260 C (-60 F to +500
F)

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 9 of 32
Proximitor Ordering Information
Sensor
Temperature 3300 XL 8 mm Proximity Probes:
Range 330101 3300 XL 8 mm Probe, 3/8-24 UNF thread,
Operating without armor3
Temperature 330102 3300 XL 8 mm Probe, 3/8-24 UNF thread, with
-51 C to +100 C (-60 F to +212 armor3
F) Part Number-AXX-BXX-CXX-DXX-EXX
Storage A: Unthreaded Length Option
Temperature Note: Unthreaded length must be at least 0.8 inches less than
the case length.
-51 C to +105 C (-60 F to +221
F) Order in increments of 0.1 in
Length configurations:
Relative Maximum unthreaded length: 8.8 in
Humidity Minimum unthreaded length: 0.0 in
Less than a 3% change in Example: 0 4 = 0.4 in
Average Scale Factor (ASF) when B: Overall Case Length Option
tested in 93% humidity in Order in increments of 0.1 in
accordance with IEC standard Threaded length configurations:
68-2-3 for up to 56 days. Maximum case length: 9.6 in
Minimum case length: 0.8 in
Probe Pressure Example: 2 4 = 2.4 in
C: Total Length Option
3300 XL 8 mm probes are
05 0.5 metre (1.6 feet)
designed to seal differential
10 1.0 metre (3.3 feet)
pressure between the probe tip
15 1.5 metre (4.9 feet)
and case. The probe sealing
20 2.0 metres (6.6 feet)
material consists of a Viton 50 5.0 metres (16.4 feet)1
O-ring. Probes are not pressure 90 9.0 metres (29.5 feet)
tested prior to shipment. Contact
our custom design department if D: Connector and Cable-Type Option
you require a test of the pressure 01 Miniature coaxial ClickLoc
seal for your application. connector with connector
Note: It is the responsibility of the customer or user to ensure protector, standard cable
that all liquids and gases are contained and safely 02 Miniature coaxial ClickLoc
controlled should leakage occur from a proximity probe. connector, standard cable
In addition, solutions with high or low pH values may
11 Miniature coaxial ClickLoc
erode the tip assembly of the probe causing media
connector with connector
leakage into surrounding areas. Bently Nevada, LLC will
protector, FluidLoc cable
not be held responsible for any damages resulting from
12 Miniature coaxial ClickLoc
connector, FluidLoc cable
leaking 3300 XL 8 mm proximity probes. In addition, 3300
E: Agency Approval Option
XL 8 mm proximity probes will not be replaced under the
00 Not required
service plan due to probe leakage.
05 Multiple Approvals
Patents:
3300 XL 8 mm Proximity Probes, Metric:
Components or procedures 330103 3300 XL 8 mm Probe, M10 x 1 thread, without
described in one or more of the armor3
following patents apply to this
product: 5,016,343; 5,126,664; 330104 3300 XL 8 mm Probe, M10 x 1 thread, with
5,351,588; and 5,685,884. armor3
Part Number-AXX-BXX-CXX-DXX-EXX

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 10 of 32
A: Unthreaded Length Option Note: The FluidLoc cable option 12 is not necessary on the vast
Note: Unthreaded length must be at least 20 mm less than the majority of 330105 and 330106 installations due to the
case length. presence of the probe sleeve. Consider carefully the
application before ordering the FluidLoc cable option for
Order in increments of 10 mm. these probes.
Length configuration:
Maximum unthreaded length: 230 . E: Agency Approval Option
mm 00 Not required
Minimum unthreaded length: 0 mm 05 Multiple Approvals
Example: 0 6 = 60 mm
B: Overall Case Length Option 3300 XL 8 mm Proximity Probes, Smooth Case:
Order in increments of 10 mm. 330140 3300 XL 8 mm Probe without armor2
Metric thread configurations:
Maximum length: 250 mm 330141 3300 XL 8 mm Probe with armor2
Minimum length: 20 mm Part Number-AXX-BXX-CXX-DXX
Example: 0 6 = 60 mm
C: Total Length Option Option Descriptions
05 0.5 metre (1.6 feet) A: Overall Case Length Option
10 1.0 metre (3.3 feet) Order in increments of 0.1 in
15 1.5 metres (4.9 feet) Length configurations:
20 2.0 metres (6.6 feet) Maximum length: 9.6 in
50 5.0 metres (16.4 feet) 1 Minimum length: 0.8 in
90 9.0 metres (29.5 feet) Example: 2 4 = 2.4 in
D: Connector and Cable-Type Option B: Total Length Option
01 Miniature coaxial ClickLoc 05 0.5 metre (1.6 feet)
connector with connector 10 1.0 metre (3.3 feet)
protector, standard cable 15 1.5 metres (4.9 feet)
02 Miniature coaxial ClickLoc 20 2.0 metres (6.6 feet)
connector, standard cable 50 5.0 metres (16.4 feet) 1
11 Miniature coaxial ClickLoc 90 9.0 metres (29.5 feet)
connector with connector C: Connector and Cable-Type Option
protector, FluidLoc cable 01 Miniature coaxial ClickLoc
12 Miniature coaxial ClickLoc connector with connector
connector, FluidLoc cable protector, standard cable
E: Agency Approval Option 02 Miniature coaxial ClickLoc
00 Not required connector, standard cable
05 Multiple Approvals 11 Miniature coaxial ClickLoc
3300 XL 8 mm Reverse Mount Probes connector with connector
protector, FluidLoc cable
330105-02-12-CXX-DXX-EXX, 3/8-24 UNF threads3
12 Miniature coaxial ClickLoc
330106-05-30-CXX-DXX-EXX, M10 x 1 threads3 connector, FluidLoc cable
D: Agency Approval Option
Option Descriptions 00 Not required
C: Total Length Option 05 Multiple Approvals
05 0.5 metre (1.6 feet)
10 1.0 metre (3.3 feet) 3300 XL 8 mm Extended Temperature Range
15 1.5 metre (4.9 feet) (ETR) Proximity Probes:
20 2.0 metres (6.6 feet) 330191 3300 XL 8 mm ETR Probe, 3/8-24 UNF thread,
50 5.0 metres (16.4 feet) 1 without armor
90 9.0 metres (29.5 feet) 330192 3300 XL 8 mm ETR Probe, 3/8-24 UNF thread,
D: Connector Option with armor
02 Miniature ClickLoc coaxial
connector Part Number-AXX-BXX-CXX-DXX
12 Miniature ClickLoc coaxial A: Unthreaded Length Option
connector, FluidLoc cable

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 11 of 32
Note: Unthreaded length must be at least 0.8 inches less than D: Agency Approval Option
the case length. 00 Not required
Order in increments of 0.1 in 05 Multiple Approvals
Length configurations: 3300 XL 8 mm Extended Temperature Range
Maximum unthreaded length: 8.8 in (ETR) Reverse Mount Probes
Minimum unthreaded length: 0.0 in
Example: 1 5 = 1.5 in 330195-02-12-CXX-DXX, 3/8-24 UNF threads
B: Overall Case Length Option 330196-05-30-CXX-DXX, M10 x 1 threads
Order in increments of 0.5 in
Threaded length configurations: C: Total Length Option
Maximum case length: 9.6 in 05 0.5 metre (1.6 feet)
Minimum case length: 0.8 in 10 1.0 metre (3.3 feet)
Example: 2 5 = 2.5 in 15 1.5 metre (4.9 feet)
C: Total Length Option 20 2.0 metres (6.6 feet)
05 0.5 metre (1.6 feet) 50 5.0 metres (16.4 feet) 1
10 1.0 metre (3.3 feet) 90 9.0 metres (29.5 feet)
15 1.5 metre (4.9 feet) D: Agency Approval Option
20 2.0 metres (6.6 feet) 00 Not required
50 5.0 metres (16.4 feet)1 05 Multiple Approvals
90 9.0 metres (29.5 feet) 3300 XL 8 mm Extended Temperature Range
D: Agency Approval Option (ETR) Proximity Probes, Smooth Case:
00 Not required
05 Multiple Approvals 330197 3300 XL 8 mm Probe without armor2
330198 3300 XL 8 mm Probe with armor2
3300 XL 8 mm Extended Temperature Range
(ETR) Proximity Probes, Metric: Part Number-AXX-BXX-CXX
330193 3300 XL 8 mm Probe, M10 x 1 thread, without A: Overall Case Length Option
armor Order in increments of 0.5 in
330194 3300 XL 8 mm Probe, M10 x 1 thread, with armor Length configurations:
Maximum length: 9.5 in
Part Number-AXX-BXX-CXX-DXX Minimum length: 1.0 in
A: Unthreaded Length Option Example: 3 5 = 3.5 in
Note: Unthreaded length must be at least 20 mm less than the B: Total Length Option
case length. 05 0.5 metre (1.6 feet)
10 1.0 metre (3.3 feet)
Order in increments of 10 mm. 15 1.5 metres (4.9 feet)
Length configuration: 20 2.0 metres (6.6 feet)
Maximum unthreaded length: 230 50 5.0 metres (16.4 feet) 1
mm 90 9.0 metres (29.5 feet)
Minimum unthreaded length: 0 mm C: Agency Approval Option
Example: 0 6 = 60 mm 00 Not required
B: Overall Case Length Option 05 Multiple Approvals
Order in increments of 10 mm.
Metric thread configurations:
3300 XL Proximitor Sensor
Maximum length: 250 mm
Minimum length: 20 mm 330180-AXX-BXX
Example: 0 6 = 60 mm A: Total Length and Mounting Option
C: Total Length Option 10 1.0 metre (3.3 feet) system
05 0.5 metre (1.6 feet) length, panel mount
10 1.0 metre (3.3 feet) 11 1.0 metre (3.3 feet) system
15 1.5 metres (4.9 feet) length, DIN mount
20 2.0 metres (6.6 feet) 12 1.0 metre (3.3 feet) system
50 5.0 metres (16.4 feet) 1 length, no mounting hardware
90 9.0 metres (29.5 feet) 50 5.0 metre (16.4 feet) system
length, panel mount
Specifications and Ordering Information
Part Number 141194-01
Rev. L (07/09)

Page 12 of 32
51 5.0 metre (16.4 feet) system A: Cable Length Option
length, DIN mount 030 3.0 metres (9.8 feet)
52 5.0 metre (16.4 feet) system 035 3.5 metres (11.5 feet)
length, no mounting hardware 040 4.0 metres (13.1 feet)
90 9.0 metres (29.5 feet) system 045 4.5 metres (14.8 feet)
length, panel mount 070 7.0 metres (22.9 feet)
91 9.0 metres (29.5 feet) system 075 7.5 metres (24.6 feet)
length, DIN mount 080 8.0 metres (26.2 feet)
92 9.0 metres (29.5 feet) system 085 8.5 metres (27.9 feet)
length, no mounting hardware
B: Agency Approval Option B: Cable Option
00 Not required 00 Standard cable
05 Multiple approvals 01 Armored cable
C: Agency Approval Option
3300 XL Extension Cable 00 Not required
330130-AXXX-BXX-CXX 05 Multiple Approvals
Note: Make sure that the extension cable length and the probe
length, when added together, equal the Proximitor Sensor
Accessories
total length. 141078-01

A: Cable Length Option Manual.


030 3.0 metres (9.8 feet) 175751
035 3.5 metres (11.5 feet)
040 4.0 metres (13.1 feet) 3300 XL Multi-Purpose Stainless
045 4.5 metres (14.8 feet) Steel Housing. 12x12x6. Can
070 7.0 metres (22.9 feet) hold up to 8 Proximitor Sensors in
075 7.5 metres (24.6 feet) a DIN-mount configuration or 6
080 8.0 metres (26.2 feet) Proximitor Sensors in a panel-
085 8.5 metres (27.9 feet) mount configuration. (Available
with ATEX Zone 0 and Zone 1
certifications.)
B: Connector Protector and Cable Option
00 Standard cable
01 Armored cable 176467
02 Standard cable with connector
3300 XL Multi-Purpose Stainless
protectors
03 Armored cable with connector Steel Housing. 12x8x6. Can
protectors hold up to 4 3300XL Proximitor
10 FluidLoc cable Sensors in both DIN-mount and
11 Armored FluidLoc cable panel-mount configurations.
12 FluidLoc cable with connector (Available with ATEX Zone 0 and
protectors Zone 1 certifications.)
13 Armored FluidLoc cable with 330181
connector protectors
C: Agency Approval Option 3300 XL Multi-Purpose Stainless
00 Not required Steel Housing. 13x9.5x7. Can
05 Multiple Approvals hold up to 8 Proximitor sensors in
a DIN-mount configuration or 6
3300 XL Extended Temperature Range (ETR) Proximitor Sensors in a panel-
Extension Cable mount configuration. Primarily
330190-AXXX-BXX-CXX used by customers requiring
Note: Make sure that the extension cable length and the probe hazardous area approvals fro
length, when added together, equal the Proximitor Sensor their installations. Available with
total length. ATEX Zone 0 and Zone 1 and

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 13 of 32
North American Division 1 and 138493-01
Division 2 certifications.
Replacement DIN-mount
02120015 mounting pad.
Bulk field wire. 1.0 mm2 (18 148722-01
AWG), 3 conductor, twisted,
3300 XL test plug. The 3300 XL
shielded cable with drain wire.
Test Plug contains 3 small test
Specify length in feet.
pins attached to 3 color-coded 1-
metre wires, each terminated in a
Aluminum probe clamp bracket2 banana plug. The 3-pin adapter
137491-AXX plugs into the test pin holes on
A: Mounting screw option 3300 XL-style Proximitor sensors.
01 10-24 UNC-2A mounting You can use this test plug to
screws check the performance of the
02 M5 x 0.8-6g mounting screws Proximitor sensor from the test
The aluminum clamp bracket is an unthreaded pin holes in the terminal strip
mounting bracket designed for use with the smooth without removing the field wiring.
case probes (330140, 330141, 330197 and 330198). 04310310
After gapping the probe, tighten the clamp bracket
by tightening the screws. The mounting screws have 3300 XL Proximitor Sensor
pre-drilled holes for safety wire. panel-mount screw. One 6-32
UNC thread forming mounting
Aluminum probe threaded mounting bracket screw. Four screws are required
137492-AXX for each Proximitor sensor.
(Screws supplied standard with
A: Thread size Proximitor housings [3300 XL
01 3/8-24 panel-mount option]).
04 M10 x 1
The aluminum probe threaded mounting bracket is 03200006
the standard mounting bracket for most 3300 and Silicone self-fusing tape. A
3300 XL probe installations. The -01 option includes 9.1-metre (10-yard) roll of silicone
two 10-24 UNC-2A mounting screws. The -04 option tape to protect connectors. This
includes two M5 x 0.8-6g mounting screws. The tape is easy to install and
mounting screws have pre-drilled holes for safety provides excellent electrical
wire. isolation and protection from the
environment. We do not
Phenolic threaded probe mounting bracket recommend using this tape inside
27474-AXX the casing of the machine.
A: Thread size 40113-02
01 3/8-24
04 M10 x 1 Connector Protector Kit.
We recommend the phenolic threaded mounting Connector Protector Kit for 3300
bracket if your application requires additional XL 8 mm probes and extension
electric isolation from the mounting location (as in cables, including connector
some generator and electrical motor bearing protectors and installation tools.
locations). The -01 option includes two 10-24 UNC- 136536-01
2A mounting screws. The -04 option includes two M5
x 0.8-6g mounting screws. The mounting screws Connector protector adapter.
have pre-drilled holes for safety wire. This allows you to use connector
protector installation tools
138492-01 manufactured prior to 1998 with
Replacement panel-mount 75 ClickLoc connectors.
mounting pad.
Specifications and Ordering Information
Part Number 141194-01
Rev. L (07/09)

Page 14 of 32
40180-02 sleeves, 2 pieces of slit FEP tubing,
and 1 strip of silicone tape.
Connector protectors. Package
contains 10 pairs of connector 330153-09
protectors for 3300 XL 8 mm
3300 XL ETR Connector Kit. Used
probes and extension cables.
on ETR 3300 XL 8mm probes and
03839410 3300 ETR XL extension cables.
Contains one pair of male and
75 triaxial male connector
female ClickLoc connectors, two
protector. Male connector
color-coded sleeves, two pieces
protectors install onto the
of high temperature slit FEP
extension cable and attach to the
tubing, and one strip of silicon
female connector protector on
tape.
the probe, providing
environmental protection of 163356
connectors.
Connector Crimp Tool Kit.
03839420 Includes 1 set of multiconnector
inserts and connector installation
75 triaxial female connector
instructions. Compatible only
protector. Female connector
with 330153 connector kits or
protectors install onto the probe
with probes shipped in 2003 or
lead and attach to the male
later with ClickLoc connectors
connector protector on the
uninstalled. Supplied with
extension cable, providing
carrying case.
environmental protection of
connectors. You can also place Notes:
the connector protector onto the
1. 5-metre probes are designed for use with
extension cable to slide over the
the 5-metre Proximitor sensor only.
connection to the Proximitor 2. Mounting clamps must be ordered
sensor to protect that connection separately for 330140, 330141, 330197, and
from the environment. 330198.
04301007 3. For a shorter delivery time, order commonly
stocked probes. Currently, stocked probes
3/8-24 probe lock nut with consist of the following part numbers:
safety wire holes. Single probe
lock nut with 2 holes drilled
through the nut in order to secure 330101-00-08-05-02-00, 330101-00-08-05-02-05,
the lock nut in place with safety 330101-00-08-10-02-00, 330101-00-08-10-02-05,
wire. 330101-00-12-10-02-00, 330101-00-12-10-02-05,
330101-00-16-10-02-00, 330101-00-16-10-02-05,
04301008 330101-00-20-05-02-00, 330101-00-20-10-02-00,
M10 x 1 probe lock nut with 330101-00-20-10-02-05, 330101-00-30-10-02-00,
safety wire holes. Single probe 330101-00-30-10-02-05, 330101-00-40-05-02-00,
lock nut with 2 holes drilled 330101-00-40-10-02-00, 330101-00-40-10-02-05,
through the nut in order to secure 330101-00-60-10-02-00, 330101-00-60-10-02-05,
the lock nut in place with safety 330102-00-20-10-02-00, 330103-00-02-10-02-05,
wire. 330103-00-04-10-02-00, 330103-00-05-10-02-00,
330104-00-06-10-02-00, 330104-01-05-50-02-00,
330153-01 330105-02-12-05-02-00, 330105-02-12-05-02-05,
3300 XL connector kit. Used on 330105-02-12-10-02-00, 330105-02-12-10-02-05,
3300 XL 8 mm probes and 330106-05-30-05-02-00, 330106-05-30-05-02-05,
extension cables. Contains 1 pair 330106-05-30-10-02-00 and 330106-05-30-10-02-
each of male and female ClickLoc 05.
connectors, 2 color-coded
Specifications and Ordering Information
Part Number 141194-01
Rev. L (07/09)

Page 15 of 32
Graphs and Figures
Note: All dimensions shown in millimetres (inches) except as noted.

Gap (m m )0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
4 0.10
DSL Error (mils)

DSL Error (mm)


7.87 V/mm (200
Referenced to

2 0.05
mV/mil)

0 0.00

-2 -0.05

-4 -0.10

10
Referenced to 7.87
V/mm (200 mV/mil)
ISF Error (%)

-5

-10

-24
-22
-20
-18
Output (Volts)

-16
-14
-12
-10
-8
-6
-4
-2
0
0 10 20 30 40 50 60 70 80 90 100
Gap (m ils)

5m @ 25 C (77 F) 5m @ 45 C (113 F) 5m @ 0 C (32 F)

Figure 1: Typical 3300 XL 8 mm 5m or 1m System over API 670 Testing Range

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 16 of 32
Gap (m m ) 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
6 0.15

DSL Error (mils) 4 0.10

DSL Error (mm)


7.87 V/mm (200
Referenced to 2 0.05

mV/mil)
0 0.00

-2 -0.05

-4 -0.10

-6 -0.15

10
Referenced to 7.87
V/mm (200 mV/mil)
ISF Error (%)

-5

-10

-24
-22
-20
-18
Output (Volts)

-16
-14
-12
-10
-8
-6
-4
-2
0
0 10 20 30 40 50 60 70 80 90 100
Gap (m ils)

9m @ 25 C (77 F) 9m @ 45 C (113 F) 9m @ 0 C (32 F)

Figure 2: Typical 3300 XL 8 mm 9m System over API 670 Testing Range

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 17 of 32
Gap (m m ) 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
6 0.15

4 0.10

DSL Error (mils)

DSL Error (mm)


7.87 V/mm (200
Referenced to
2 0.05

mV/mil)
0 0.00

-2 -0.05

-4 -0.10

-6 -0.15

10
Referenced to 7.87
V/mm (200 mV/mil)
ISF Error (%)

-5

-10

-24
-22
-20
-18
Output (Volts)

-16
-14
-12
-10
-8
-6
-4
-2
0
0 10 20 30 40 50 60 70 80 90 100

Gap (m ils)
1m P ro be @ 25 C (77 F) 1m P ro be @ -35 C (-31F)
1m P ro be @ 120 C (248 F)

Figure 3: Typical 3300 XL 8mm Probe over API 670 Operating Range

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 18 of 32
Gap (m m ) 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
6 0.15

4 0.10
DSL Error (mils)

DSL Error (mm)


7.87 V/mm (200
Referenced to 2 0.05

mV/mil)
0 0.00

-2 -0.05

-4 -0.10

-6 -0.15

10
Referenced to 7.87
V/mm (200 mV/mil)
ISF Error (%)

-5

-10

-24
-22
-20
-18
-16
Output (Volts)

-14
-12
-10
-8
-6
-4
-2
0
0 10 20 30 40 50 60 70 80 90 100
Gap (m ils)

TC=25 C (77 F) Tc=-34 C (-30 F) Tc=-51C (-60 F)

Figure 4: Typical 3300 XL 8 mm 5m Proximitor Sensor with 4m Extension Cable @ Tc (Probe is at 25 C)

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 19 of 32
Gap (m m ) 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
8 0.20
6 0.15
DSL Error (mils)

DSL Error (mm)


Referenced to
4 0.10

200(mV/mil)
2 0.05
0 0.00
-2 -0.05
-4 -0.10
-6 -0.15
-8 -0.20

15

10
Referenced to
ISF Error (%)

200(mV/mil)

-5

-10

-15

-20

-18

-16

-14
Output (Volts)

-12

-10

-8

-6

-4

-2

0
Gap (m ils) 0 10 20 30 40 50 60 70 80 90 100

25 C (77 F) 65 C (149 F) 85 C (185 F) 100 C (212 F)

Figure 5: Typical 3300 XL 8 mm 5m Proximitor Sensor with 4m Extension Cable @ Th (Probe is at 25C)

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 20 of 32
Gap (m m )
0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
8 0.20
6 0.15

DSL Error (mils)

DSL Error (mm)


4 0.10
2 0.05
0 0.00
-2 -0.05
-4 -0.10
-6 -0.15
-8 -0.20

15

10
ISF Error (%)

-5

-10

-15

-22
-20
-18
-16
Output (Volts)

-14
-12
-10
-8
-6
-4
-2
0
0 10 20 30 40 50 60 70 80 90 100

Gap (m ils)

Tc=+25 C (+77 F) Tc=-34 C (-30 F) Tc=-51C (-60 F)

Figure 6: Typical 3300 XL 8mm 9 m Proximitor Sensor with 8m of Extension Cable @ Tc (Probe is at 25 C)

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 21 of 32
Gap (m m )
0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
8 0.20
6 0.15

DSL Error (mils)

DSL Error (mm)


4 0.10
2 0.05
0 0.00
-2 -0.05
-4 -0.10
-6 -0.15
-8 -0.20

15

10
ISF Error (%)

-5

-10

-15

-22
-20
-18
-16
Output (Volts)

-14
-12
-10
-8
-6
-4
-2
0
0 10 20 30 40 50 60 70 80 90 100

Gap (m ils)
Th=+25 C (+77 F) Th=+65 C (+149 F)
Th=+85 C (+185 F) Th=+100 C (+212 F)

Figure 7: Typical 3300 XL 8mm 9m Proximitor Sensor with 8m Extension Cable @ Th (Probe is at 25 C)

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 22 of 32
Gap (m m ) 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
10 0.25
8 0.20
6 0.15

DSL Error (mils)

DSL Error (mm)


4 0.10
2 0.05
0 0.00
-2 -0.05
-4 -0.10
-6 -0.15
-8 -0.20
-10 -0.25

15

10
ISF Error (%)

-5

-10

-15

-24
-22
-20
-18
Output (Volts)

-16
-14
-12
-10
-8
-6
-4
-2
0
Gap (m ils) 0 10 20 30 40 50 60 70 80 90 100

Th=+25 C (+77 F) Th =+260 C (+500 F)

Figure 8: Typical 3300 XL Extended Temperature Range Probe and 4m Extended Temperature Range Extension
Cable @ Th (Proximitor Sensor and Probe Tip with 1-foot Cable are at +25 C)

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 23 of 32
Gap (m m ) 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50
10 0.25
8 0.20
6 0.15

DSL Error (mils)

DSL Error (mm)


4 0.10
2 0.05
0 0.00
-2 -0.05
-4 -0.10
-6 -0.15
-8 -0.20
-10 -0.25

15

10
ISF Error (%)

-5

-10

-15

-24
-22
-20
-18
Output (Volts)

-16
-14
-12
-10
-8
-6
-4
-2
0
Gap (m ils) 0 10 20 30 40 50 60 70 80 90 100

Th=+25 C (+77 F) Th=+260 C (+500 F)

Figure 9: Typical 3300 XL Extended Temperature Range Probe and 8m Extended Temperature Range Extension
Cable @ Th (Proximitor Sensor and Probe Tip with 1-foot Cable are at +25 C)

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 24 of 32
Frequency Response to Different Field Wiring Lengths
without Barriers (5 m System)
1

0
Magnitude (dB)

-1

-2

-3

-4

-5
100 1,000 10,000 100,000
Frequency (Hz)

No field wiring 1000' wiring 2000' wiring


5000' wiring 12,000' wiring

Figure 10: Frequency Response, Typical 3300 XL 8mm 5m or 1m System with Varying Lengths of Field Wiring
Attached, No Barriers

Phase Response with Different Field Wiring Lengths, No


Barriers (5 m System)
0
-10
Phase Angle (degrees)

-20
-30
-40
-50
-60
-70
-80
-90
-100
100 1,000 10,000 100,000
Frequency (Hz)
No field wiring 1000' wiring 2000' wiring
5000' wiring 12,000' wiring

Figure 11: Phase Response, Typical 3300 XL 8mm 5m or 1m System with Varying Lengths of Field Wiring Attached,
No Barriers

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 25 of 32
Frequency Response to Different Field Wiring Lengths without
Barriers (9 m System)
1

Magnitude (dB) -1

-2

-3

-4

-5
100 1,000 10,000 100,000
Frequency (Hz)

No field wiring 1000' field wiring 2000' field wiring


5000' field wiring 12,000' field wiring

Figure 12: Frequency Response, Typical 3300 XL 8mm 9m System with Varying Lengths of Field Wiring Attached,
No Barriers

Phase Response with Different Field Wiring Lengths, No


Barriers (9 m System)
0.00
-10.00
Phase Angle (degrees)

-20.00
-30.00
-40.00
-50.00
-60.00
-70.00
-80.00
-90.00
-100.00
100 1,000 10,000 100,000
Frequency (Hz)

No field wiring 1000' field wiring 2000' field wiring


5000' field wiring 12,000' field wiring

Figure 13: Phase Response, Typical 3300 XL 8mm 9m System with Varying Lengths of Field Wiring Attached, No
Barriers

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 26 of 32
4
2

1 3
5

2.5 (0.10) 6
7

8
9
10

1. Probe tip, 8.0 mm (0.31 in) diameter


2. 9/16 in for 3/8-24 threads, M17 for M10 threads (see Note 2)
3. Case thread
4. 5/16 in wrench flats for 3/8-24 threads; 8mm wrench flats for M10 threads.
5. 75 cable, 3.68 mm (0.145 in) maximum outside diameter, 3.94 mm (0.155 in) maximum outside diameter for FluidLoc cable,
7.67 mm (0.302 in) outside diameter of armor, 9.5 mm (0.38 in) maximum diameter of armor ferrule
6. Miniature male coaxial connector, 7.24 mm (0.285 in) maximum outside diameter D
7. Unthreaded length A
8. Case length B
9. 6.0 mm (0.235 in) maximum
10. Total length C, +30%, -0% (see Note 3)
Figure 14: 3300 XL 8mm Proximity Probes, Standard Mount
330101 and 330191, 3/8-24 UNF-2A, without armor 7

330102 and 330192, 3/8-24 UNF-2A, with armor 6

330103 and 330193, M10X1 thread, without armor 7


330104 and 330194, M10X1 thread, with armor 6

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 27 of 32
1 36.3 3
Max. 51.1 (2.01) Max.
(1.43)

1. 12.4 mm (0.49 in) maximum diameter


2. Connector protector (fluorosilicone material)
3. 12.4 mm (0.49 in) maximum diameter
Figure 15: Installed Connector Protectors

2
3
1
4

2 (0.08)
5 (0.2) 5
6
7
8
9
1. Probe tip, 8.0 mm (0.31 in) diameter
2. 7/16 in or M10 hexagonal
3. Case thread
4. 75 cable, 3.68 mm (0.145 in) outside diameter
5. Miniature male coaxial connector, 7.24 mm (0.285 in) maximum outside diameter D
6. Unthreaded length A, 5.0 mm (0.20 in)
7. Case length B, 30 mm (0.12 in)
8. 6.0 mm (0.235 in) maximum
9. Total length C, +30%, -0% (see Note 3)

Figure 16: 3300 XL 8mm Proximity Probes, Reverse Mount 4, 7


330105 and 330195, 3/8-24 UNF-2A threads
330106 and 330196, M10X1 threads

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 28 of 32
2 4
1 3
5

2.54 (0.100)
6 7
8
9

1. Probe tip, 8 mm (0.31 in) diameter


2. 9.66 mm (0.38 in) maximum diameter
3. 5/16 in wrench flats, 4 each
4. 75 cable, 3.68 mm (0.145 in) maximum diameter, 3.94 mm (0.155 in) maximum diameter for FluidLoc cable, 7.67 mm (0.302
in) outside diameter with armor, 10.2 mm (0.4 in) maximum diameter for armor ferrule
5. Miniature male coaxial connector, 7.24 mm (0.285 in) maximum outside diameter D
6. Case length A,
7. 349.3 (13.75) max. distance
8. 6.0 mm (0.235 in) maximum
9. Total length C, +30%, -0% (see Note 3)
Figure 17: 3300 XL 8mm Proximity Probes, Smooth Case
330140 and 330197, without armor 7
330141 and 330198, with armor 6

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 29 of 32
2
4
83.8 83.8
1 (3.30) (3.30)
5
3

83.8 83.8
(3.30) (3.30) 7
6 8
9
1. 7.24 mm (0.285 in) maximum diameter
2. Miniature male coaxial connector
3. FEP or PFA coated armor, armor length 300 mm (11.8 in) less than cable length (see Note 6)
4. 75 cable, 3.7 mm (0.15 in) maximum outside diameter, 3.94 mm (0.155 in) maximum diameter for FluidLoc cable, 7.67 mm
(0.302 in) maximum outside diameter of armor, 10.2 mm (0.40 in) maximum diameter of armor ferrule
5. 7.24 mm (0.285 in) maximum diameter
6. Stainless steel ferrules, 10.2 mm (0.40 in) max diameter
7. FEP or PFA insulated triaxial cable
8. Miniature female coaxial connector
9. Cable length, +20%, -0%

Figure 18: 330130, 3300 XL Extension Cable (FEP Armor and Insulation)
30190, 3300 XL ETR Extension Cable (PFA Armor and Insulation)

1. Mounting option A, Options 50 or -90


Figure 19: Panel Mount 3300 XL Proximitor Sensor

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 30 of 32
1. Mounting option A, Options 51 or 91
2. 35mm DIN rail (not included)
3. 89.4 mm (3.52 in). Additional 3.05 mm (0.120 in) clearance required to remove DIN rail.

Figure 20: DIN Mount 3300 XL Proximitor Sensor

1. Mounting option A, Options 50 or -90

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 31 of 32
Figure 21: Physical Mounting Characteristics Showing Interchangeability of 3300 and 3300 XL Proximitor Sensors
when 4-hole Mounting Option Is Used8
Notes:
1. All dimensions on figures are in millimetres (inches) unless otherwise noted.
2. Standard mount 8 mm probes supplied with M17 or 9/16 inch lock nut.
3. Probes ordered with 5 or 9 metre integral cables have a length tolerance of +20%, -0%.
4. Reverse mount probes not available with armor or connector protector options.

5. Letters inside quotation marks on figures refer to probe ordering options.


6. Stainless steel armor is supplied with FEP outer jacket for standard probes, PFA outer jacket for ETR probes.
7. FEP jacket is standard non-armored portion of the cable for standard probes, PFA jacket on non-armored
portion for ETR probes.
8. Use M3.5 or #6 screws for panel-mount Proximitor Sensors (screws provided when purchasing Bently Nevada
housings).

* denotes trademarks of Bently Nevada, LLC, a wholly owned subsidiary of General Electric Company.

Viton is a trademark of E. I. Du Pont de Nemours and Company.

1999-2009. Bently Nevada LLC. All rights reserved.

1631 Bently Parkway South, Minden, Nevada USA 89423


Phone: 775.782.3611 Fax: 775.215.2873
www.ge-energy.com/bently

Specifications and Ordering Information


Part Number 141194-01
Rev. L (07/09)

Page 32 of 32