You are on page 1of 2

SPECIFICATIONS FOR PHASED ARRAY ULTRASONIC SYSTEM

1. The system should be compact, light and ideal for field use with splash
proof casing.YES-ALL
2. The system should have fully featured A scan with user-defined colour
scheme with recording facility for full inspection data.
3. B scan facility for corrosion mapping with zoom option for region of
interest. YES-ALL
4. Built-in 2 axis encoder interface. YES-16-32
5. TOFD capability for accurate amplitude-independent defect sizing
complaint with ASME 2235 code. YES-16-32
6. Encoded C- scan storage and imaging capability with user defined
colour scheme. YES-16-32
7. Volume corrected sectorial scan view. YES-16-32
8. Programmable array probe to perform linear scan with electronic
movement of the ultrasonic beam along with the probes. YES-ALL
9. Real time Linear scan, Volume corrected and interpolated display.
10. Multiple display layout capability with split screen. YES-ALL
11. The system should have pulser/receiver with 16 pulser and number
of
elements to be 128 with 256 focal laws. YES-16-32
12. Compatibility with conventional normal, angle and dual probes.
YES-16-32
13. Display resolution 800X600 pixels or better. 8.4 inches diagonal.
YES-16-32
14. Battery operation minimum 6 hours. YES-16-32
15. System should have capability to do minimum 3 linear inspections
simultaneously and display upto 3 linear inspections simultaneously.
YES-16-32
16. Communication port: USB & Ethernet. YES-16-32
17. The phased array system should have the facility for automatic
probe recognition and set-up. YES-16-32
18. System bandwidth: 0.75 to 18 MHz (-3dB). YES-16-32
19. Storage through compact flash card 512 MB. YES-16-32
20. Alarm: Number of alarms- 3. Conditions: any logical combination of
gates.
Analog outputs-2 YES-16-32
21. System should have programmable Time corrected gain. TCG
number of
points as 16 (1 TCG curve per channel for focal laws). YES-16-32
22. Pulser voltage should be 80 V per element. YES-16-32
23. Pulse width should be adjustable from 30ns to 500 ns, resolution
2.5ns.
24. System should have a fall time less than 7 ns. YES-16-32
25. System must have digitizing frequency of 100 MHz (10 bits). YES-
16-32
26. System must be modular in construction. YES-16-32
27. System should have code complaint with the following coded and
code cases :
ASME Section V Article 4 in general & ASME code 2541,Upcoming
ASME codecases 2557 and 2558 for manual raster scans using S scans
and E scan and ASTM 2841-06 standard guide for evaluation of
characteristics of phased array instruments and system along with API
and AWS D1:1 codes. YES-16-32
28. System should have fast PA focal law calculation processing time of
7 sec. YES-16-32
29. System should have wizard type (step by step assistance) for
complete set up creation and sensitivity, wedge delay and velocity
calibration. YES-16-32
30. System should have TCG calibration. An attenuation curve (TCG) is
applied for each beam. YES-16-32
31. System should have sensitivity calibration: A different gain can be
applied over each beam. YES-16-32
32. System should have wedge delay calibration code complaint. YES-
16-32
33. System should have compatibility for Automation in future. YES-
16-32

You might also like