You are on page 1of 1

Tolerance classes for thermocouples according to ASTM E230/E230M-12

The International standard IEC 60584:2013 is a parallel standard, but aware that both
tolerances and temperature ranges could be different from those of the ASTM standard.
Note that prescribed tolerances are valid for unused thermocouple material only.

Special tolerances Standard tolerances Standard tolerances


Thermocouple type
(C) (C) (C)

R&S
Temperature range 0 < T < 1480 0 < T < 1480
0.6 1.5 -
The greater of...
0.001T 0.0025T
B
Temperature range 870 < T < 1700 870 < T < 1700
-
Tolerance 0.0025T 0.005T
J
Temperature range 0 < T < 760 0 < T < 760
1.1 2.2 -
The greater of...
0.004T 0.0075T
T
Temperature range 0 < T < 370 0 < T < 370 -200 < T < 0
0,5 1 1
The greater of...
0,004T 0,0075T 0,015T
E1
Temperature range 0 < T < 870 0 < T < 870 -200 < T < 0
1.0 1.7 1.7
The greater of...
0.004T 0.005T 0,01T
K&N (K only)
Temperature range 0 < T < 1260 0 < T < 1260 -200 < T < 0
1.1 2.2 2.2
The greater of...
0.004T 0.0075T 0.01T
C
Temperature range 0 < T < 2315
- -
Tolerance 0.01T

Reference junction 0C
Updated 2014-10-10

Notes:
1) MIMS type of metal sheathed thermocouples E will have greater standard tolerances: 2.2 C or
0.0075T, whichever is greatest from 0 to 870 C and the greater of 2.2 C or 0.02T from -200 to
0 C.

You might also like