Professional Documents
Culture Documents
23
APPLICATION GUIDE
FOR THE CIRCUIT BREAKER
ANALYZER AND MICROOHMMETER
MOD. CBA 2000
Doc. MIE11169 Rev. 1.23 Page 2 of 149
Disclaimer
Every effort has been made to make this material complete, accurate, and up-to-date. In
addition, changes are periodically added to the information herein; these changes will be
incorporated into new editions of the publication. ISA S.R.L reserves the right to make
improvements and/or changes in the product(s) and/or the program(s) described in this
document without notice, and shall not be responsible for any damages, including but not
limited to consequential damages, caused by reliance on the material presented,
including but not limited to typographical errors.
Copies, reprints or other reproductions of the content or of parts of this publication shall
only be permitted with our prior written consent.
All trademarks are the property of their respective holders.
SH O R T FO REW O RD
I often wondered why the users manual is not very much used,
even if it includes valuable information. As me too I am a user of
such manuals, the answer I have given myself is that valuable
information are concealed somewhere in the thick thing, and I do
not have time to waste to find it. So, either the manual is actually
of help, or I ignore it.
Luca Biotti
Q&A Manager
Doc. MIE11169 Rev. 1.23 Page 7 of 149
S AFE TY AT W O R K
IMPORTANT
Dont position the equipment so that is difficult to operate the
disconnecting device, that is the mains supply cord.
The Product includes a rechargeable NI-Cd battery (the battery
must be supplied and substituted only by manufacturer).
The Product deals with voltages and currents that may be lethal
to the unadvertised user. Besides, in order to avoid any danger in
case of fault inside the Product, the device under test should have
the following characteristics:
. Connection sockets must be not accessible;
. Input circuits must have an isolation degree at least equal to the
one of the Product.
. The symbol
! means to read the application guide
before to using the equipment. Its related to dangerous output,
and is located near hazardous sockets.
If you want to clean the test set, dont pour liquids on it: they
could pour in, and damage some card.
sockets!
CB coils not There can be voltage CB coils
disconnected from the from the plant during disconnected.
circuit. tests: this could damage
the test set.
Contact with the plant The DC voltage can be Connections
auxiliary supply. higher than 50 V DC, to the DC
hence very dangerous. supply must
be protected.
Travel transducer does Damage of the travel Verify Open
not have extra stroke transducer or joints and Close
because of extra stroke positions
during tests
CB in service before Danger of short-circuits Disconnect
disconnecting it on the test set. CBA 2000
from CBA 2000. prior to
restoring. In
case, use the
PARK sockets
for coil cables.
Of these points, the first three are very hazardous, both for the
user and the test set. THESE TYPES OF FAULT ARE NOT
COVERED BY THE WARRANTY.
1 I NT RO D U CT IO N
The manual has been revised to the firmware revision 1.21. The
firmware revision is shown at power-on on the main screen.
Doc. MIE11169 Rev. 1.23 Page 12 of 149
1.21
If your test set is not upgraded, you have to connect to the ISA
WEB site:
www.isatest.com,
enter the customer area with the test set number, and download
the firmware RELATED TO : DONT MISTAKE PLEASE. Then use
UPGRADE in the TDMS services to transfer the new file into your
test set.
2 C B T IM I N G T ES T
2.1.1 CB contacts.
If you have more than six chambers per phase, you can split the
test in three, testing one phase at a time: you can connect up to
18 chambers per phase.
Doc. MIE11169 Rev. 1.23 Page 16 of 149
MASTER / SLAVE AUXILIARY CONTACT TRANSDUCER MICRO OHM METER BREAKER COILS
8A - 250VAC MEASURE CAT. III F 63mA 250V MASTER / SLAVE AUXILIARY CONTACT TRANSDUCER MICRO OHM METER BREAKER COILS
8A - 24VDC 8A - 250VAC MEASURE CAT. III F 63mA 250V
MASTER C SLAVE 200A MEASURE
CAT. III V MASTER C SLAVE 8A - 24VDC 200A MEASURE
OUTPUT
5V 10mA x 3
MAX
MAX
MAX
300V 30A
300V 20A OUTPUT CAT. III V MAX 300V 30A
15V MAX
INPUT 5V 10mA x 3 MAX 300V 20A
15V
MAX 10V x 3 ! INPUT
MAX 10V x 3 !
BREAKER CONTACTS BREAKER CONTACTS BREAKER CONTACTS
ANALOG INPUT 2 BREAKER CONTACTS BREAKER CONTACTS BREAKER CONTACTS
ANALOG INPUT 2
O1
O1
MEASURE MEASURE
CAT. III CAT. III PARK MEASURE MEASURE
CAT. III CAT. III PARK
A2 A4 A6
INPUT INPUT
A1 C A3 C A5 C INPUT INPUT
MAX 500V MAX 500V A1 C A2 A3 C A4 A5 C A6
MAX 500V MAX 500V
B1 C B2 B3 C B4 B5 C B6
O2 B2 B4 B6
B1 C B3 C B5 C O2
PARK
PARK
C1 C C2 C3 C C4 C5 C C6
C1 C C2 C3 C C4 C5 C C6
CHARGE CHARGE
BATT. BATT.
MINIMUM TRIP COIL MINIMUM TRIP COIL
O abc def O abc def
! MAX 250V I ! MAX 250V I
+ USB FLASH DRIVE
1 2 3 START
O + USB FLASH DRIVE
1 2 3 START
O
VDC ghi jkl mno VDC ghi jkl mno
IN V 85-264V V 85-264V
4 5 6 C O
50/60Hz - 100W
IN
4 5 6 C O
50/60Hz - 100W
USB V 100-350V USB
pqrs tuv wxyz CO OC pqrs tuv wxyz V 100-350V
CO OC
VDC 100W VDC 100W
OUT 7 8 9 USER OCO
OUT 7 8 9 USER OCO
+ RS232 CBA2000 + CBA2000
.*
RS232
0 DEL
CIRCUIT BREAKER SEL .* 0 DEL
CIRCUIT BREAKER SEL
ANALYZER ANALYZER
The CBA 2000 to the left is the Master: this is the one to which
are connected the coils. The Slave is the one that monitors the
additional chambers: its Slave input is connected to the Master
output of the Master CBA 2000. Before starting the test on the
Master unit, on the Slave unit(s) you have to select the Slave
operation, as follows.
If you dont use the optional connection cables, the cross section
should be enough for the coil drive current: at least 6 sq. mm, in
order to minimize the current drop.
Note that, if you use the three opening coils option, when you
enter the Test options menu and select a sequence involving
Open (Open, Open-Close, Close-Open, OpenCloseOpen), the
following additional selection is displayed.
Doc. MIE11169 Rev. 1.23 Page 20 of 149
Now, select the coil current range. For the best measurement
accuracy, the range should be the smaller one above the coil
current peak. If the coil current is more than the selected range,
the current measurement will be cut at the value of the range.
NOTE: if the test set is supplied by the mains, the LED turns red
as soon as the contact is sensed; if it is supplied by the battery,
contacts are tested only as the test is started.
On the first test, you have to input the tests header, as follows.
You can key in the header by the keyboard, that uses the logic of
portable phones. Once test data are input, you can proceed with
Doc. MIE11169 Rev. 1.23 Page 23 of 149
the test. As you are finished with a CB, dont forget to input the
new CB header.
Selecting Show setups you get the list of the saved setups: you
can choose the desired one. If these values match with what you
want to do, you can immediately start the test; else, you can load
a previously saved setting, or you can enter the menu, modify
the desired parameters and then save the setting, before
proceeding.
Menu > Recording options > Open to close delay > (new
value) RET
Doc. MIE11169 Rev. 1.23 Page 24 of 149
Note that it is possible to time the duration so that the Open coil
command arrives while the Close coil command is still active.
Now, you are ready for the test. As you see, on the front panel
there are two operating buttons in all: one selects the type of
test; the other one starts the test. You dont have to use the
Menu selection to set the desired test. After power-on, the default
selection is OCO; for instance, if you want to perform an O-C test,
Doc. MIE11169 Rev. 1.23 Page 26 of 149
press the SEL button until the OC LED turns on: the test is
selected.
Now, press the START button: the buzzer warns eight times that
the test is about to start, then the test is performed. The LED
above the pushbutton turns on until the test is over.
The dotted lines mark the Open, Close, Open commands. You can
examine the diagram, and verify current peaks and waveform,
main contact timing, pre-insertion resistor timing, by moving the
two cursors. Cursor selection is performed checking t1 or t2: after
selection, turning the wheel the cursor can be moved to the
desired point. As you move it, time measurements are updated;
to the left, current values are the ones selected by the cursor.
Pressing 5 on the keyboard, the movement is slowed down for a
better pointing at the diagram.
If the result is OK, you can save it, and perform other tests;
else, you can repeat it, ignoring the warning message about
saving.
Saving the result is performed using the Result selection:
type in the file name and press the button: the result is saved.
It is possible to continue testing the same CB and saving results
with different tests. It is also possible to recall the former result
of the same CB, so that you can check for deviations in time; this
is easily performed using the software.
The first measurement is the delay between the Open (Close) coil
command and the corresponding main contact Opened (Closed)
position. The following figure shows the recording with an Open
Close command: delays are referred to coil commands.
Doc. MIE11169 Rev. 1.23 Page 29 of 149
In its action, the filter does not show the short bounces; so, there
can be a mismatch between the bounce duration, shown in the
result table, and the display. If it is desired to view these
bounces, it is possible to select Original waveforms in
Preferences: this can be useful also to verify that there is not too
much noise, especially with the BSG option.
Other relevant measurements are the Open coil current and Close
coil current: they are metered moving the cursors on the desired
point.
A
2
Doc. MIE11169 Rev. 1.23 Page 31 of 149
The reported bounce value is the longest one among the contacts
being in series.
CLOSE
CB CONTACT
OPEN
DWELL
The first two (or four) lines of the window above report peak
currents and current flow duration, separately for the different
coils. Next data are the main breaker contacts timing
measurements.
Doc. MIE11169 Rev. 1.23 Page 33 of 149
When you load a result, the Set selection becomes the following
one .
If you want to start a new test, go to Set and press it (or press
the DEL button on the test set): this re-sets test result.
After powering-on CBA 2000, first of all select the type of input
(dry or with voltage); then, check that the contact status
displayed by LEDs on the front matches with the nominal one
(light ON = Closed or With voltage), and correct if something is
wrong. The selection is performed as follows.
The selection defines the type of test. Standard tests are always
available, while tests on the CBs Minimum trip coil and the
minimum voltage test on standard coils depend upon the
availability of the options: Four Coils, MTC (Minimum Trip Coil).
If you are on the main window, you can act on SEL: opening this
window the performed selection is displayed. In the normal
operation way, if you dont need to change the default selections,
after power-on select with SEL the desired test and start it.
Doc. MIE11169 Rev. 1.23 Page 38 of 149
- With the Close selection, the test set issues a Close command.
DC V
JUMPER
MINIMUM
TRIP COIL
0 DC V
10. Fit back the jumper, that keeps the CB's under-voltage coil in
the normal position, then press again the knob: C3 opens, but
the CB's under-voltage coil is supplied via the jumper. The display
will show the following result:
Doc. MIE11169 Rev. 1.23 Page 42 of 149
11. The window shows the O3 coil current, and the CB opening.
In the Results table are reported O3 current and opening timings,
starting from the O3 voltage loss.
12. You can measure the open delay, save the result or repeat
the test: in this instance, you have to send a Close command
before next test.
13. After the last test, while the CB is open, connect again the
CB's under-voltage coil to the auxiliary DC voltage supply: the
test is finished.
OPTION 250 V 70 V
Maximum input voltage 250 V 70 V
Maximum operating voltage 240 V 50 V
Minimum operating voltage 50 V 16 V
Maximum voltage drop 120 V 45 V
Minimum voltage drop 10 V 5V
Voltage adjustment step 2V 0,5 V
Voltage accuracy 2V 0,5 V
Maximum output current 4 A; dV < 10 A; dV <
60 V; 12 V;
2 A; dV > 5 A; dV > 12
60 V V
Maximum test duration 500 ms 500 ms
If the test is stopped, the test set displays the following message.
Doc. MIE11169 Rev. 1.23 Page 45 of 149
The test is selected going to Menu, then test option, and then
checking the Enable minimum voltage test: the following screen
is displayed.
The test result table is the same as usual, with the addition of the
test voltage.
MTC ACTIVATED!
Press the knob to start the test.
Remove the jumper and press the knob: the display says
TEST IN PROGRESS
11. The test set reduces the voltage across the coil with a ramp
having a slope of 20 V/s; the descent is stopped when the voltage
reaches its minimum. During the ramp, the test set monitors the
main contact inputs (actually the test is referred to input A1), in
order to memorize the intervention threshold.
12. Once the minimum is reached, the voltage stays at that value
during the stabilization time, and then increases with a ramp at
the same slope until the nominal voltage is reached. The CB
should not close, as there is no CLOSE command; if it does, the
voltage at which the CB closes is displayed.
13. After this, the following message is displayed.
TEST COMPLETED!
Press the knob to de-activate the MTC.
And the test result is displayed.
14. Set back the bridge: the test is over.
15. Restore the connections to the CB's under-voltage coil before
closing it again.
The test result table shows the voltage threshold at the time of
intervention.
Remove the jumper and press the knob: the display says
TEST IN PROGRESS
12. The test set issues a CLOSE command, to make sure that the
CB is closed; then, about 2 s after, the voltage drops to the
minimum value, stays down during about 0.4 s and goes up
again. The CB should open; the test set records the timing of all
the selected main contact inputs, in order to memorize the
intervention time (actually, the time is measured on input A1).
13. After this, the following message is displayed.
TEST COMPLETED!
Press the knob to de-activate the MTC.
Doc. MIE11169 Rev. 1.23 Page 55 of 149
12. Set back the connections to the CB's under-voltage coil before
closing it again.
The test result table shows breaker opening the time delay, which
is measured on input A1.
Doc. MIE11169 Rev. 1.23 Page 56 of 149
3 C B MO VE ME NT A ND A N A LO G S I G N A LS TE ST
The test set has ten analog inputs plus three digital ones:
. Coil currents; two standard, four optional;
. High voltage (up to 350 V AC): two, with no common point;
. Low voltage (up to 10 V DC): four; three of them with the same
reference, located in the transducer connectors;
. Digital inputs: three, RS422, for digital transducers, located in
the transducer connectors.
3.1.1 Introduction
Travel transducers are used for the purpose of having a better
detail on the mechanical status of the CB. This is accomplished as
follows:
. Comparing the movement trace to CB Open and Close contacts,
it is possible to verify if they act at the desired stroke positions,
including tolerances;
. Computing from the trace the corresponding speed, it is possible
to verify if CB limits are exceeded because of plays caused by
wearing.
. Also auxiliary CB contacts acting positions can be checked.
Vc = V * P / L
From this, the test set computes the stroke of the cursor, and
therefore the CB movement.
For the analog linear transducer there are two ends which cannot
be trespassed during the test.
For the analog angular transducer there is not such a limit: the
transducer can rotate continuously. However, during its rotation,
the angular transducer goes from zero volts to the maximum, and
then, continuing the rotation, the voltage comes back to zero.
Also, the total rotation angle is usually 355; during the
remaining 5, the voltage output is zero.
Doc. MIE11169 Rev. 1.23 Page 60 of 149
Digital transducers are of the rotating type; they differ from the
analog ones also because they can rotate more than one turn
without generating errors: there is no mounting problem.
Transducers differ mainly upon the number of impulses per turn:
the bigger the number, the better the movement reproduction
accuracy. The transducer supplied as an option has 5000
impulses per turn, which ensures an excellent movement
reproduction even on the typical movement of 90.
Doc. MIE11169 Rev. 1.23 Page 61 of 149
TIGHTENING KNOB
Doc. MIE11169 Rev. 1.23 Page 66 of 149
The important thing to know about the adaptive arm is that when
you tighten the knob, all the arm are blocked to their position.
So, the mounting is performed by un-tightening the knob,
locating it to its operating position, and then tightening it.
Test set comes with many connection cables for the two types of
transducers.
For the case that you have your own transducer and cable,
possibly already supplied (but WITH A VOLTAGE LESS THAN 10
V), we provide three cables, 1 m long, terminated with a ten-way
connector on the CBA side, and with three banana connectors,
for the connection to the transducers. The bananas assignment is
the following.
For the case that you have your own transducer and cable, we
supply also three cables, 1 m long, terminated with the 10-way
connector on the CB side, and with banana sockets on the other
side. Sockets are coloured and marked as follows.
You will use cables ended with bananas for the connection to your
transducer.
This selection window is the same for the three transducers of the
three phases, but the calibration will be performed at the
meantime on the three transducers to accommodate for the
mechanical tolerances.
Once the selection is performed, the following window is
displayed.
Doc. MIE11169 Rev. 1.23 Page 72 of 149
You can label the transducer and name the phase, then you can
select the following parameters:
A) For the CB contact:
. Unit of measurement for the circuit breaker (mm, in);
. The nominal breaker contact stroke, that usually is different
from the corresponding transducer stroke.
B) For the transducer:
. Unit of measurement for the transducer (degree, mm, in).
. If the wetting voltage from the transducer comes from the test
set or if it is coming from an external source. In the latter case,
you have to key in the EXACT value of the wetting voltage: an
error here causes a proportional error to all movement
measurements.
. If the actual strokes at the connection point are the same or
they differ on the three phases. In the latter case, you can
program the stroke for phase A, and then, changing the phase,
input different values for phases B and C; else, the programmed
value will be the same for the three phases. Pressing F3, the
selected phase changes. NOTE: the selection is referred to the
stroke of the connection point, not to the nominal transducer
stroke (length): the three transducers are assumed to have the
same stroke.
. Nominal total transducer stroke (i.e., the stoke that the
transducer can perform, before connecting it to the CB).
. Actual transducer stroke, that is the stroke that the transducer
will actually perform after connection to the CB, from the Open to
the Closed CB contact positions. This stroke has to be less than
the total stroke, 80% maximum, in order to accommodate for
extra movements that always occur dynamically, as the CB opens
and closes. The actual transducer stroke can be input in the unit
of measurement selected for the transducer, or in percentage of
the total transducer stroke.
For instance:
. CB contact stroke: 8 in;
. Nominal total transducer stroke: 345 (angular), or 250 mm
(linear);
. Actual transducer stroke: 210 (angular), or 125 mm (linear).
Doc. MIE11169 Rev. 1.23 Page 73 of 149
match the tolerance limits. The following is how does the window
looks like at this moment.
The first problem is that CBA 2000 needs a voltage input. This
problem is easily solved connecting the transducer output to a
resistor of a well-known value. There is an additional problem:
you cannot exceed the maximum resistance rating for the
converter. A reasonable choice is to use a resistor of 1 kOhm, for
the 0 to 5 mA converters, and of 250 Ohm, for the 4 to 20 mA
one. One kOhm resistors are easily found with a tolerance of 1%,
0.5%, or even 0.1%: the accuracy is up to you. For the 250 Ohm
resistor, parallel 4 resistors rated 1 kOhm.
When you connect the resistor to the transducer output, you will
have a voltage drop across it of 0 to 5 V, for the 0-5 mA
converter, or of 1 V at 4 mA, and 5 V at 20 mA, for the second
one. Connect the resistor to our measurement input.
Next, the problem is that the test set asks for the nominal
transducer stroke. The point is that the test set understands that
when you have 0 V you are at stroke 0, and when you have 5 V
Doc. MIE11169 Rev. 1.23 Page 77 of 149
As the variation is 50% of the total, you may say that you have a
stroke of 50 mm, and the transducer stroke is 100 mm long. If
you program the transducer total stroke of 100*5/4 = 125 mm,
the actual stroke will occur between 2 V = 50 mm and 4 V = 100
mm; the delta is 50 mm, as you desire.
The fact is that, when you calibrate, you compute the actual
stroke, between two points which are neither 0 V nor + 5 V, and
you have the correct stroke measurement.
Doc. MIE11169 Rev. 1.23 Page 78 of 149
In conclusion, with this trick you are able to use also these
transducers.
These transducers are of the rotating type; they differ from the
analog ones also because they can rotate more than one turn
without generating errors. Transducers differ mainly upon the
number of impulses per turn: the bigger the number, the better
the movement reproduction accuracy. The transducer supplied as
an option has 5000 impulses per turn, which ensures an eccellent
movement reproduction even on the typical movement of 90.
Doc. MIE11169 Rev. 1.23 Page 79 of 149
This selection window is the same for the three transducers of the
three phases, but the calibration will be performed at the
meantime on the three transducers to accommodate for the
mechanical tolerances.
Once the selection is performed, the following window is
displayed.
Doc. MIE11169 Rev. 1.23 Page 80 of 149
You can label the transducer and name the phase, then you can
select the following parameters:
A) For the CB contact:
. Unit of measurement for the circuit breaker (mm, in);
. The nominal breaker contact stroke, that usually is different
from the corresponding transducer stroke.
B) For the transducer:
. The number of impulses per turn;
. If the actual strokes at the connection point are the same or
they differ on the three phases. In the latter case, you can
program the stroke for phase A, and then, changing the phase,
input different values for phases B and C; else, the programmed
value will be the same for the three phases. Pressing F3, the
selected phase changes. NOTE: the selection is referred to the
stroke of the connection point, not to the nominal transducer
stroke (length): the three transducers are assumed to have the
same number of impulses per turn.
. Actual transducer stroke, that is the stroke that the transducer
will actually perform after connection to the CB, from the Open to
the Closed CB contact positions. The actual transducer stroke can
be input in degrees or in percentage of 360.
NOTE: for all selections, in the test result diagram, the ZERO
value is the CLOSED position. In all diagrams, the CLOSED
position is always designed HIGH.
The test set reports in the results table the key parameters of the
diagram:
Breaker stroke, measured after the oscillations at the
end of the movement;
Wipe: it is the movement from the rest position to
when the contact opens;
Over-travel: it is the stroke beyond the reference Open
or Closed positions;
Rebound: it is the stroke after the over-travel;
Average speed between the two datum points.
The following window shows the performed measurements, which
are found after all other measurements.
verified after the mounting. This operation has been called slow
motion because the mechanism movement is actuated by hand
by the operator. The situation is summarized in the following
drawing.
Where :
. Total stroke is the contact stroke between mechanical ends,
with shock absorbing medium removed ;
. A and B bounds are the stroke limits between mechanical ends,
with shock absorbing medium removed ;
. CB stroke is the contact stroke during normal operation: it is
limited by shock absorbers;
. A and B positions are the stroke limits between the shock
absorbers ;
. Over-travel A and B are the limits dynamically reached by the
contact during normal operation;
. Over-travel stroke is the difference between the contact position
at rest and the position dynamically reached during the
movement;
. A and B margins are the residual shock absorber strokes, before
the mechanism beats against the mechanical ends: these are
what we are looking for.
The fact that the first limit is referred to the Open or Closed
position is not relevant. The calibration will be performed
separately on the three phases; once the first phase has been
selected, press F2: the window becomes the following one.
Doc. MIE11169 Rev. 1.23 Page 89 of 149
The A bound is set to zero, and taken as the reference for further
measurements. As the contact is moved, and taken to B bound,
the B bound figure changes. Of course, this value will be bigger
than the final contact stroke: the difference is the total available
margin. Once the B bound is reached, press F4: the screen
becomes the following one.
Doc. MIE11169 Rev. 1.23 Page 90 of 149
The screen displays the Open and Closed extra strokes: it is now
possible to verify if they are the ones foreseen; otherwise,
pressing DEL it is possible to delete the measurements and to
start over again. If extra strokes are OK, change the phase and
repeat the procedure.
. In the analog inputs window, select type, and then select the
pressure transducer test.
The test execution is the same as for CB contacts test. The result
window displays the analog input waveform during the test.
Measurements are performed moving the cursor on the desired
points, but there is no value in the test result table.
The test set allows monitoring two high (or low) level analog
voltages, plus three low level analog voltages. The sixth analog
input is dedicated to the resistance measurements.
The two high level analog signals inputs are four banana sockets,
black and white. The reference (black) sockets are isolated
between them, and also with respect to the other three analog
inputs; so, any signal can be connected.
Doc. MIE11169 Rev. 1.23 Page 94 of 149
The three low-level analog inputs are located in the three multi-
pins connectors. Pin connection has been explained in paragraph
3.2. The connection is performed using the connectors provided.
These inputs have in common the reference input.
Note that if the voltage input is AC instead of DC, you can select
the AC input checking the sinusoidal waveform icon.
The result window displays the analog input waveform during the
test. Measurements are performed moving the cursor on the
desired points, but there is no value in the test result table.
The three low level analog inputs can be used to measure any
analog input, up to the limit of 10 V DC (or 7 V AC), in
alternative to the transducer measurements. The selection is
performed as follows.
Doc. MIE11169 Rev. 1.23 Page 96 of 149
It is possible to:
Label the current to be measured;
Select if the current is DC or AC; in the latter instance,
check the sinusoidal waveform icon
Select the V/A clamp constant;
Select the maximum test current.
With these parameters, the screen will perform the conversion
from voltage to current, and display the current.
Doc. MIE11169 Rev. 1.23 Page 97 of 149
4 ST AT I C RE S IST A N CE T E ST
The test is performed applying a high current to the test object (if
it is a CB contact it must be Closed), and measuring the voltage
drop across it.
The high current cables are connected to the high current sockets
(19): they should have a cross section of 25 sq. mm, in order to
minimize the voltage drop; besides, they should be twisted
along the maximum possible cable length, in order to
minimize the inductance.
The figure above shows the connection, with one pole grounded.
However, during this test both ends can be grounded: usually,
the resistance of ground connections, that is in parallel to CB
contact resistance, is in the range of 10 mOhm, and it does not
affect very much the measurement, that is in the range of 100
Ohm or less: if the CB contact resistance is greater than 100
Ohm, then the contact is damaged and needs an intervention for
repairing it. In fact, with a nominal current is 1000 A, the contact
resistance power dissipation is 1 W per Ohm: 1 mOhm causes
the power dissipation of 1 kW.
Now you can perform the test: with different test numbers and
labels they will be separately recorded.
Doc. MIE11169 Rev. 1.23 Page 100 of 149
A. TEST OF A RESISTANCE
The time to wait between two tests is about 1 minute, that serves
to charge the capacitor.
This is how the screen would look like after the test.
Doc. MIE11169 Rev. 1.23 Page 103 of 149
5 D Y N AM IC RE SI S T A N CE T E ST
MEASUREMENT CONNECTIONS
AFTER HIGH CURRENT CLAMPS
The figure above shows the connection, with one pole grounded.
However, during this test both ends can be grounded: the
resistance of ground connections, that is in parallel to CB contact
resistance, is in the range of 10 mOhm, and it does not affect
very much the measurement, that is in the range of 100 Ohm or
less: if the CB contact resistance is greater than 100 Ohm, then
the contact is damaged and needs an intervention for repairing it.
In fact, with a nominal current is 1000 A, the contact resistance
power dissipation is 1 W per Ohm: 1 mOhm means a 1 kW
power dissipation.
A) CLOSE TEST
Before test start, the current is zero. Then, current increases (to
190.9 A in our case), the voltage drop is measured (29.88 mV)
and the resistance is computed (0.156 mOhm). After some while,
the capacitor is discharged, and the current goes to zero.
The following is the same as above, but with only the resistance
and the auxiliary contacts.
Doc. MIE11169 Rev. 1.23 Page 111 of 149
Now it can be clearly seen that the resistance has a starting peak,
that goes down in few ms: at the meantime, the current goes to
a stable value (see the former diagram). This part of the
resistance diagram is to be ignored, as it comes from the
connection cables inductance, that cannot be eliminated.
We have cut the recording in four areas, separated by the dotted lines.
Doc. MIE11169 Rev. 1.23 Page 112 of 149
1 2 3 4
In the area 1), the contact is still open: current is zero, resistance
is > 3 mOhm: in this test, we used the 1 mOhm range. The red
cursor, T1, reads > 3 mOhm.
In the area 3), the current is stable: the blue cursor, at T2,
measures the static contact resistance (0.106 mOhm).
You may see that the current was interrupted after about 24 ms,
and the resistance was out of scale. The point here is that this CB
had passed all other tests, and was OK to be put in service!
Luckily, the operator decided to perform also the dynamic
resistance test. They found all nuts being loose. This single test
paid the expenses to purchase CBA!
B) OPEN TEST
Let us also zoom on the dynamic test, and move the cursor on
the resistance measurement: the window is the following.
The total test duration is 27 ms, so the timing is the right one
considering a discharge time of the capacitor of 30 ms.
Before test start, the current is zero. Then, as the contact is
closed, unlike the Close test, the current increases (to 193.1 A in
our case), the voltage drop is measured (36.65 mV) and the
resistance is computed (0.189 mV). As the contact opens, the
current and the voltage drop to zero, while the resistance is
infinite.
The following is the zoom of the transition portion, with only the
resistance movement and the auxiliary contacts.
Doc. MIE11169 Rev. 1.23 Page 117 of 149
We have cut the recording in four areas, separated by the dotted lines.
Doc. MIE11169 Rev. 1.23 Page 118 of 149
1
3 4 5
In the area 1), the contact is still closed, but current is zero,
resistance is > 3 mOhm: in this test, we used the 1 mOhm range.
In the area 3), the current is stable: the red cursor, at T1,
measures the static contact resistance (0.191 mOhm).
In the area 4), the contact starts opening; the resistance grows
to 2.5 mOhm (blue cursor at T2) in about 3.6 ms. This resistance
variation is to be considered, because, given the high currents
involved in short-circuit opening, there is a lot of heat being
generated. The ideal is no resistance change until the opening.
Close 1: Dynamic CLOSE test; only one end grounded. When the
test is started, the current cannot flow, until the contact closes.
At cursor 1 the resistance is infinite; at cursor 2, we measure the
contact resistance (0,629 mOhm).
Doc. MIE11169 Rev. 1.23 Page 120 of 149
Unlike the dynamic resistance Close test, when both ends are
grounded, the current flows as soon as it is started; so, it is
important to program the Close delay, as it is in the Open test.
Doc. MIE11169 Rev. 1.23 Page 121 of 149
Open 1: Dynamic OPEN test; only one end grounded. When the
test is started, the current flows through the closed contact; next
to CB opening, the current is stopped. At cursor 1 the resistance
is the one of the contact (0,629 mOhm); at cursor 2, the
resistance is infinite.
Open 2: Dynamic OPEN test; both ends grounded. When the test
is started, the current flows through the closed contact; next to
CB opening, the current flows through the ground connection. At
cursor 1 the resistance is the one of the contact (0,601 mOhm);
at cursor 2, the resistance is 45 mOhm (the grounding
resistance).
Unfortunately, these resistances are too small for the BSG option
to operate. However, with the dynamic resistance test it is
possible to measure the CB timing, as follows.
The procedure is the same for the CLOSE delay. In our example,
the resistance with the CB open is 70 Ohm; as the CB closes, it
decreases to 24 Ohm. The screen is the following one.
6 BO TH S ID ES GR O U N D E D T EST W IT H BS G 10 0 0
BYPASS RESISTANCE
BYPASS:
ONE PER
CONTACT
Doc. MIE11169 Rev. 1.23 Page 127 of 149
NOTE: if the measuring head is for two poles and the CB has only
one pole, connect the unused clamp to the black clamp: dont
leave them un-connected!
. Prior to connect BSG, perform a resistance test, in order to
measure the resistance in Closed and Open situation. This serves
to set the resistance thresholds during the test. The ideal Open
contact resistance is close to 5 mOhm, that is the maximum
threshold setting. If the Open resistance is more than 8 mOhm,
clamp another shunt in parallel.
. Connect the BSG1000 measuring heads current outputs to CB
main contacts, using the three 10sq. mm cables;
. Connect the BSG1000 measuring heads measurement inputs to
CB main contacts, using the four measurement clamps;
. Connect the three BSG1000 measuring heads to BSG1000 main
unit (three shielded cables, 10 m long).
BSG
DETECTED
Dont
forget to
define the
Open and
Closed
Contact
resistances
in the
BREAKER
CHANNEL
SETTINGS
. Select the page for the setting of breaker channels, that will
modify as follows. This means that CBA 2000 has detected that
the option is present;
. For the Open threshold, input a value equal to 95% of the Open
resistance measurement that you have previously performed (but
not more than 5 mOhm).
. For the Closed threshold, two choices:
Doc. MIE11169 Rev. 1.23 Page 130 of 149
Any resistance value less than the Closed threshold is sensed and
displayed as Closed. Any resistance value more than the Open
threshold is sensed and displayed as Open. Any resistance value
between these thresholds is sensed and displayed as a Resistive
contact: this is the case for graphite nozzles.
. Issue an Open CB command: the CBA diagram should show the
contact OPEN. If not, go to the BSG1000 menu, threshold
selection, and reduce the threshold until contacts are sensed
Open;
. Now, issue a Close CB command: the CBA diagram should show
the contact CLOSED. If not, go to the LOW threshold selection,
and increase it until contacts are sensed Closed.
The possible cases for the threshold settings are summarized in
the following table.
CB DIAGRAM CORRECTION
OPEN CLOSED REDUCE HIGH
THRESHOLD
CLOSED OPEN INCREASE LOW
THRESHOLD
NOTE: LEDS on the front panel dont follow the contact status
unless as the test is performed.
and the similar messages for clamps black and blue: please,
correct.
Sequence of operation
The BSG measuring heads generate the current only when a
test is to be performed.
The current generation is initiated by CBA1000 or CBA 2000,
by means of the Master output. The START signal goes to the
BSG Main unit, that, in turn, commands the Measuring Heads.
When the current is OK the heads inform the CBA via the Main
unit; at this moment, the CBA drives the CB coils, and the time
measurement starts.
The head measures the contact resistance, compares it to the
threshold and communicates in real time the result (CLOSED,
GRAPHITE or OPEN) to the Main unit;
The Main unit converts this information into a voltage level
that is sensed by CBA as CLOSED, RESISTANCE or OPEN. These
Doc. MIE11169 Rev. 1.23 Page 132 of 149
When time tests are executed with the BSG option, it is possible
to observe some time deviation when you compare time results
taken with the option to the same results taken without the
option. The reason is explained in the following sketch, that
shows the resistance change versus time during the Close and
Open transitions.
The sketch shows that, during the Close operation, there is some
time between the moment when the CB contact resistance is 100
mOhm and when it goes down to values such as 10 mOhm and 1
mOhm; the same occurs during the Open operation. The amount
of time difference depends upon the type of CB and of its
isolation mean. Now, the point is that during the normal time
Doc. MIE11169 Rev. 1.23 Page 133 of 149
On the other side, it has to be considered that during our test the
reading head does not measure the CB contact resistance: it
measures the PARALLEL between the CB contact and the ground
connections and the shunt cable. The resistance change with time
becomes the dashed line.
7 F IR ST TR I P T ES T
PRIMARY
CURRENTS
SECONDARY
CURRENTS
CB
COIL
. Next, set the three analog inputs as follows. First, select the
analog inputs 10 V.
Doc. MIE11169 Rev. 1.23 Page 137 of 149
Now the test set is ready to perform the test. Go to the test
diagram and press START: the display will show the message
TEST IN PROGRESS, then TRIGGER ARMED. From this
moment, you have 60 s to issue the CB Open command. The test
set will wait until the selected AUX input is sensed; then, it will
display a screen such as the following one (three secondary
currents and the open coil current).
The timing can be read by zooming the first part of the diagram,
and then View to select the input 1 only: the screen is now the
following one.
Doc. MIE11169 Rev. 1.23 Page 141 of 149
8 SW IT CH S Y N C T EST
PRIMARY
CURRENTS
VT
SECONDARY
CURRENTS
CB
CLOSE
COIL
. Connect the clamps to the three low range analog inputs, and
clamp them to the CTs secondary sides. To this purpose, use the
three connectors which are included in the standard set cables,
plus six 2 m long banana to banana cables.
Doc. MIE11169 Rev. 1.23 Page 145 of 149
In the Test options window select Close (one test only); in the
recording options select: 50 ms as Close coil duration, 20 ms as
pre-trigger duration, 0.15 s as test duration: this improves the
test result resolution.
In the breaker contacts, select a contact per phase (that is not
connected): this improves the display.
Now the test set is ready to perform the test. Go to the test
diagram and press START: the display will show the message
TEST IN PROGRESS, then TRIGGER ARMED. From this
moment, you have 60 s to issue the CB Close command. The test
set will wait until the selected AUX input is sensed; then it will
display a screen such as the following one.
Doc. MIE11169 Rev. 1.23 Page 148 of 149