FEATURES
1. Compact and lightweight
The adoption of the latest electronics and digital
signal processing technologies has reduced the size
and weight of the flow transmitter by 30% and 30%, Flow transmitter (FSC)
respectively, in comparison with the Fuji conventional
portable flowmeter (Model FSC). (in comparison to
our existing model)
2. Battery operation
The flowmeter is designed for 12 hours of continu- Detector (FSSC)
ous operation via built-in battery which is recharge-
able in 3 hours with the exclusive power adapter.
3. Full variety of detectors
The flowmeter is suitable for various types of detec-
tors applicable for small to large diameter pipe (pipe
inner diameter 13 to 6000mm) and low to high
Detector for high-temperature(FSSH)
temperature (40 to +200C).
4. High accuracy and high-speed response
The flowmeter is designed for high accuracy 10. Graphic printer connection (option)
(1.0%). Easy recording with the Integral type printer.
Response time is within 1 second. 11. Flow velocity profile measurement (option)
5. Improved anti-bubble characteristic Flow profile may be observed in real time.
Anti-bubble characteristic is greatly improved by
digital signal processing. SPECIFICATIONS
6. Excellent performance and easy operation
Large graphic LCD that is outside but easy to read. Measuring objects
Minimum number of function keys are used for page
Measurement fluid:
selection, allowing easy setting.
Uniform liquid in which ultrasonic
While battery is working, the flowmeter is water waves can propagate.
resistant and tolerates exposure to rain. Turbidity of fluid: 10000 mg/L or less
7. Large capacity storage by SD memory card State of fluid: Well-developed turbulent or laminar
Measured data is periodically stored in SD memory flow in a filled pipe.
card. For example, in the case of 256MB (option), it Fluid temperature: 40 to +200C
can be saved about 1 year measurement date(In case Measuring range: 00.3 to 32m/s
of saving period 30 seconds, 14 kinds of saved data).
Available up to 8MB. Piping conditions
8. Serial communication
Applicable piping material:
Use of a USB port allows easy connection to a per-
Select from carbon steel, stainless
sonal computer. Measured date collection panel and
steel, cast iron, PVC, FRP, copper,
Loader software for PC (standard) which is available
aluminum, acrylic or material of known
for display and change of parameter (site setting) are sound velocity.
prepared. Pipe size: Flow rate measurement
9. Heat quantity (calorie) measurement 13 to 6000mm
Heat quantity (calorie) may be measured by tem- Flow velocity profile measurement
perature input, making energy management easy for 40 to 1000mm
cooling and heating.
EDS6-147a
Date Feb. 14, 2013
FSC-2, FSS, FSD
Lining material: Select from no lining, tar epoxy, Analog output signals:
mortar, rubber, Teflon, pyrex glass or 4 to 20mA DC, one point (load resis-
material of known sound velocity. tance, 600 or less)
Note) No gap allowed between the Instantaneous velocity, instantaneous
lining and the pipe. flow rate or heat quantity (calorie) after
Straight pipe length: scaling.
10D or more upstream and 5D or more Analog input signal:
downstream (D: internal pipe diam- 4 to 20mA DC, one point
eter) (input resistance, 200 or
Refer to Japan Electric Measuring In- less) Total
struments Manufactures Associations 4 to 20mA DC, one point (in- 2 points
standard JEMIS-032 for details. put resistance, 200 or less)
or 1 to 5V DC, one point
Performance specifications Used to input temperature for heat
quantity measurement, etc.
Accuracy rating:
SD memory card: Used for data logger function and
Pipe inner Flow velocity Accuracy
diameter range
recording screen data.
Available up to 8GB (Option256MB)
13 to 50mm 2 to 32m/s 1.5% of rate
0 to 2m/s 0.03m/s Compliant media
50 to 300mm 2 to 32m/s 1.0% to 1.5% of rate SD memory card: speed class 2, 4, 6
0 to 2m/s 0.02 to 0.03m/s SDHC memory card: speed class 4, 6
300 to 6000mm 1 to 32m/s 1.0% to 1.5% of rate Format
0 to 1m/s 0.01 to 0.02m/s FAT16: 64MB to 2GB
Note1) Reference conditions are based on JEMIS-032. FAT32: 4GB, 8GB
Note2) Refer to the 4 pages for the accuracy according to kind Otherwise, reading and saving are
of detector. impossible.
File format
Flow transmitter (Type: FSC) Date logger: CSV file
Power supply: Built-in battery or AC power adapter Screen date: Bit map file
Built-in battery: Exclusive lithium button battery Serial communication:
(5000m Ah) USB port (device* compatible):
Continuous operation time, approx. 12 Mini B receptacle
hours (without printer, back light OFF, Connectable number of Mini B recep-
output current not used and at normal tacles:
ambient temperature (20C)) 1 unit
Recharging time, approx. 3 hours Transmission distance: 3m max.
(power adapter used) Transmission speed: 500kbps
Recharging temperature range: 0 to Data:
+40C Instantaneous velocity, instanta-
Power consumption: Min. 3W and neous flow rate, total value, heat
Max. 16W quantity (calorie) value, error infor-
The consumption varies depending on mation, logger data, etc.
the use conditions. * Device: Connected plug from PC
Power adapter: Exclusive power adapter 90V to 264V Printer (option): To be mounted on top of transmitter
AC (50/60Hz), 70VA or less. unit
LCD: Semi-transmissive color graphic display Thermal line dot printing
240 320 (with back light) Note) When the Chinese display is selected,
Measurement value (instantaneous printing is made in kanji characters.
flow rate, integrated flow rate) and Ambient temperature:
various settings are displayed. 10 to +55C (Without printer)
Excellent visibility even outdoors in 10 to +45C (With printer)
direct sunlight. Ambient humidity: 90%RH or less
LED display: Status display when using AC power Type of enclosure: IP64 (Without printer)
adapter. Enclosure case: Plastic case
DC IN (green): Power supply status Outer dimensions: H210 W120 D65mm (Without printer)
CHARGE (red): Battery charging under- H320 W120 D65mm (With printer)
way Weight: 1.0kg (Without printer)
Operation keypad: 1.2kg (With printer)
11 buttons
(ON, OFF, ENT, ESC, MENU, , , , Various functions
, LIGHT, PRINT) Display language: Selectable from Japanese, English,
Power failure backup: German, French, Spanish or Chinese
Measurement value is backed up by (switchable by key operation).
nonvolatile memory. Clock display function:
Clock backup with lithium battery (ef- Time (year, month, day, hour, minute)
fective term, 10 years or more) display (configurable)
Response time: 1 second Monthly error: about 1 minutes at nor-
mal temperature (20C).
2
Instantaneous value display function: Temperature display function:
Instantaneous velocity, instantaneous Fluid temperature be displayed by cur-
flow rate display (The flow in reverse rent input from temperature transmit-
direction is displayed with minus .) ter.
Numeric value: 10 digits (decimal point Metric system
equals 1 digit) Temperature unit: C or K
Unit: Metric/English system selectable English system
Metric system Temperature unit: F or K
Velocity: m/s Site data storage function:
Flow rate: L/s, L/min, L/h, L/d, kL/d, Max. 32 locations (sites) data (pipe
ML/d, m3/s, m3/min, m3/h, size, material, fluid type and etc) can
m3/d, km3/d, Mm3/d, BBL/s, be stored into built-in non-volantile
BBL/min, BBL/h, BBL/d, memory.
kBBL/d, MBBL/d Damping: 0 to 100sec (every 0.1sec) configurable
English system for analog output and velocity/flow
Velocity: ft/s rate display
Flow rate: gal/s, gal/min, gal/h, gal/d, Low flow cut: Equivalent to 0 to 5m/s
kgal/d, Mgal/d, ft3/s, ft3/min, Output setting function:
ft3/h, ft3/d, kft3/d, Mft3/d, Current output scaling, output type,
BBL/s, BBL/min, BBL/h, burnout setting and calibration
BBL/d, kBBL/d, MBBL/d Serial communication function:
Total value display function: Instantaneous velocity, instantaneous
Display of forward or reverse total flow rate, total value, heat flow, er-
(reverse is displayed as minus) ror information, received waveform,
Numeric value: 10 digits (decimal point analog input, velocity profile data,
is corresponding to 1 digit) logger data, etc. may be downloaded
Unit: Metric/English system selectable to personal computer.
Metric system Logger function: Instantaneous velocity, instantaneous
Flow rate total: mL, L, m3, km3, Mm3, flow rate, total value, heat flow, error
mBBL, BBL, kBBL information, received waveform, ana-
English system log input, velocity profile date can be
Flow rate total: gal, kgal, ft 3, kft3, Mft3, saved in a SD memory card.
mBBL, BBL, kBBL, ACRE-ft Waveform display function:
Consumed heat quantity (calorie) display function: Bi-directional received waveforms may
Display of consumed heating medium be displayed.
Metric system Graph display function:
Heat flow: MJ/h, GJ/h Flow rate trend graph may be dis-
Total heat quantity: MJ, GJ played.
English system Printing function (option):
Heat flow: MJ/h, GJ/h, BTU/h, kBTU/h, Hard copy output of a screen
MBTU/h, kWh, MWh Periodic printing (type: text, graph)
Total heat quantity: Logger date (type: text, graph)
MJ, GJ, BTU, kBTU, MBTU, Flow velocity profile measurement (option):
kW, MWh Flow velocity profile may be observed
J : Joule in real time using the exclusive detec-
BTU : British thermal unit tor (option).
W : Watt (Refer to page 5 for details.)
Computation function of consumed heat quantity (calorie):
This function calculates the heat Detector (Type: FSS)
quantity received and sent with liquid
Type of detector:
(water) in cooling and heating.
Classification Type Internal pipe Fluid temperature
diameter (mm)
Consumed heat quantity Middle diameter FSSC 50 to 1200 -40 to 120C
q = K Q (T1T2)
Converter
K : heat quantity (calorie)
Small diameter FSSD 13 to 300 -40 to 100C
conversion factor Large diameter FSSE 200 to 6000 -40 to 80C
(For heating K = 4.123,
Temperature
For cooling = 4.186) High temperature FSSH 50 to 400 -40 to 200C
sensor
Mounting method: Mounting on outside of pipe
T1: Fluid temperature Sensor mounting method:
(inlet)
V or Z method
exchanger
Heat
Heat
FSC)
Converter Connection method:
Detector Transmitter side
Temperature
sensor
(type : FSS) Exclusive connector
Detector side (FSSE)
T2: Fluid temperature (outlet) Q: Flow rate of the fluid Screw terminal
Others: BNC connector
3
FSC-2, FSS, FSD
Ambient temperature:
20 to +60C
MEASURING PRINCIPLE
Ambient humidity: With ultrasonic pulses propagated diagonally between the
FSSE 100%RH or less upstream and downstream sensors, flow rate is measured by
Other 90%RH or less detecting the time difference obtained by the flow of fluid.
Type of enclosure: Detector
Upstream sensor
FSSC IP65
(When waterproof BNC con-
nector is provided) Output
D Flow signal
FSSE IP67 t2 trans-
Flow t1 mitter
Others IP52 velocity V
Water-proof treatment type IP68
Downstream sensor
(Submerged resistant struc-
ture for 5 days)
Material of detector: MOUNTING OF DETECTOR
Classification Type Sensor Rail material PZT oscillator Plastic wedge
case Pipe
Small diameter FSSC Plastic Aluminum alloy + Plastic
Middle diameter FSSD Plastic Aluminum alloy + Plastic Lining
Large diameter FSSE Plastic
High temperature FSSH SUS304 Aluminum alloy
transmitter
DC
Extension cable(option):
Flow
Extended when the length of the detec-
tor signal cable is not sufficient.
Length: 10m, 50m (2) When Z method is used for mounting
Cable
Detector
transmitter
4 to 20mA
DC
Flow
V*2) 1.5
to 2.5 1.0
FSSD
Z *1)
1.0
V 1.5 1.0
FSSC
Z 1.0
V 1.5 1.0
FSSE
Z 1.5 1.0
V 1.0
FSSH
Z*1) 1.0
*1) When FSSD or FSSH is mounted using the Z method, guide rail (option) is required additionally.
*2) For the pipe inner diameter of 13mm, the sensor mounting dimension may be 0.0mm or less depending on pipe material and thickness.
When the sensor mounting dimension is 0.0mm or less,measurement error is about 2 to 5%.
<Description of the table> Required min. pipe thickness (fluid: water) (Unit: mm)
It shows pipe thickness of each material that the sensor mounting Steel pipe 2.15 or more FRP 3.21 or more
size is to be 0.0mm, when fixing a pipe. If the fluid is the one other Stainless pipe 1.87 or more Ductile cast iron 2.15 or more
than water, and if the sound velocity of fluid is faster than the one PVC pipe 3.69 or more PEEK 3.69 or more
of water, the sensor mounting size is to be 0.0mm or more. Copper pipe 3.82 or more PVDF 3.69 or more
Cast-iron pipe 2.98 or more Acrylic pipe 2.90 or more
Aluminum pipe 1.99 or more Polypropylene 3.69 or more
4
FLOW VELOCITY PROFILE DISPLAY FUNCTION Measurement principle
(OPTION) <Pulse Doppler method>
Ultrasonic pulses are transmitted through the fluid flow.
Flow velocity profile can be observed in real time using Entrained bubbles and microscopic particles within the
the dedicated detector from the outside. It is specifiable fluid create frequency phase shifts (Doppler effect.) The
by the code symbol of flow transmitter. resulting doppler shifts are integrated across the inside
pipe diameter cross section. The resulting profile curve
is a real-time dynamic display of the flow profile within
APPLICATION the pipe.
Pulse Doppler method is applicable to observe flow velocity
profile in real time, display the flow status in the pipe, and Sensor 1 (FSD)
decide the appropriate measurement location. Also, it can
be used for diagnosis of flow and laboratory test.
Air Flow
bubbles
SPECIFICATIONS Reflection
Sensor 2
Measuring fluid: Uniform liquid in which ultrasonic (FSD)
waves can propagate.
Flow Flow
Turbidity of fluid: Axisymmetric flow in a filled pipe. velocity velocity
Fluid temperature: Profile 2 Profile 1
Flow velocity
40 to +100C (FSDP2)
40 to +80C (FSDP1,FSDP0)
Air bubble quantity: Piping wall Center Piping wall
0.02 to 15Vol% (Velocity is 1m/s) The above shows an example when using two sensors.
Pipe size: Small type sensor : 40 to 200mm
One detector displays the flow velocity profile for a ra-
Middle type sensor :100 to 400mm
dius.
Large type sensor :200 to 1000mm
Measurement range:
0 to 0.3: Maximum Velocity (de-
pending on the pipe diameter)
Refer to chart, table.1.
Block diagram
Note) This function is to observe flow (1) Using one sensor
velocity profile, and it may be different
Cable
from actual flow rate. Power
supply
Sensor (FSD)
DETECTOR FOR FLOW VELOCITY PROFILE Flow
MEASUREMENT (TYPE: FSDP) Transmitter
Mounting method:
Mounting on outside of existing pipe
Ambient temperature: 20 to +80C
(2) Using two sensors
Ambient humidity: 100% RH or less
Type of enclosure: Power
Cable supply
IP67 (with waterproof BNC connector Sensor (FSD)
provided.)
Material: Sensor housing: PBT
Flow
Guide frame: Aluminum alloy Transmitter
Mounting belt: Plastic cloth belt/stain-
less belt Sensor (FSD)
5
FSC-2, FSS, FSD
CODE SYMBOL
<Flow transmitter> <Detector>
1 2 3 4 5 6 7 8 9 10 11 (for transit time)
F SC 2 - 0 Description 1 2 3 4 5 6 7 8 9 10
6
CODE SYMBOL OPTIONAL ITEMS
Name Specifications Arrange-
<Detector> ment No.
(for flow velocity profile measurement) 1 Battery Special type Li-ion battery ZZP*TK7N6384P1
1 2 3 4 5 6 7 8 (7.4V, 2500mAh) *Order in two pairs.
F SD 0 Y 1 Description 2 AC power Special type power adapter and
<Kind> adapter 90 to 264V AC, 50/60Hz
P2 Small type (40 to 200mm) ZZP*TK7N6380C4
P1 Middle type (100 to 400mm)
P0 Large type (200 to 1000mm)
<Terminal mold> 3 Power code Japan, North America:125V AC 2m ZZP*TK7N6621P1
0 None Europe, Korea: 250V AC 2m ZZP*TK7N6608P1
China: 250V AC 2m ZZP*TK7N6609P1
<Structure>
Y General use 4 Printer To be mounted on top of converter ZZP*TK4J2634C1
1 Modification No. Thermal serial dot system (8 x 384
dot)
7
FSC-2, FSS, FSD
Diagram To connect signal cables for FSSC, FSSD, FSSH, FSDP2, FSDP1, FSDP0
BNC
connector
Printer Detector
(option)
Detector
(the downstream side)
Flow
transmitter Detector input
(the upstream side)
ON OFF
ESC Analog input/output cable
AO analog output 4 to 20mA DC
FEED
ENT
Print
MENU
AICH1 analog input 4 to 20mA DC or 1 to 5 V DC
AICH2 analog input 4 to 20mA DC
CHARGE DC IN
PC
SD Power connector AC90 to 264V
Conversion cord
SD memory Power cord
card AC power adaptor
To connect PC
L 5D
10D or more
L 10D
90 bend
Detector
10D or more
L 10D
L 50D
Tee
10D or more
or more
L 30D L 5D
0.5D
Diffuser
D
1.5D or more
L 10D L 50D
Contraction pipe
L 30D L 10D
Valve
Isolation valve
Check valve
L 50D
Pump
P
8
OUTLINE DIAGRAM (Unit:mm)
Flow transmitter
1201.8 651.8
PORTAFLOW
Sensor (downstream)
DOWNSTREAM
SENSOR
UPSTREAM
2102.5
Sensor (upstream)
AI/AO
DC12V
ON OFF
ESC
FEED
Analog input/output
ENT
Print
MENU DC power input
CHARGE DC IN
Power switch
9
FSC-2, FSS, FSD
1201.8 651.8
Printer
PORTAFLOW
Sensor (downstream)
DOWNSTREAM
SENSOR
UPSTREAM
Sensor (upstream)
AI/AO
Analog input/output
DC12V
ON OFF
FEED
ESC
ENT
Print DC power input
MENU
CHARGE DC IN
Power switch
10
OUTLINE DIAGRAM (Unit:mm)
4802
SPACING: 0 to 300
BNC connector
Scale (inch) Scale (mm)
53
BNC CONN. Lock nut
88max.
Sensor unit
54
Rail end Tag plate Name plate Guide rail
Extended rail
Type.
Mfd.
(880)
L
700max.
400max. 240
Absorber unit
100 100 100 100
Ser.No.
Extended rail
Made in Japan
W1
Weight
Type Diameter (mm) L H W Approx. (kg)
Straight plug BNC Connector
FSDP2 40 to 200 2601.2 70 57 0.8
Straight plug
(12.3)
AO
AI ch1
AI ch2
(5)
+100
1500 -0 1005
Specification
Sensor frequency: 2MHz
Mounting method: V method, Z method (FSSD3)
Fluid temperature: 40 to 100C CODE SYMBOL
Applicable pipe material: PVC, SS, carbon steel pipe, <Detector>
copper pipe, aluminum pipe, etc. 1 2 3 4 5 6 7 8 9 10
[In case lining is removed from the pipe, Measurement F S SD1 1 Description
can not be conducted] <Senser type>(4th digits)
D 13 to 100mm
Rated accuracy of combination with the flow transmitter
<Guide rail>(5th digits)
(Applicable piping: plastic, metal pipe)
1 Provided (L=320mm)100mm
Internal diameter Velocity Accuracy 3 Iong rail (L=540mm)300mm
(mm) <Mounting belt>(6th digits)
Y None
13 to 50 2 to 32m/s 1.5% to 2.5% of rate A Stainless belt (1.5m2)
0 to 2m/s 0.03 to 0.05m/s B Plastic cloth belt (3m1)
50 to 100 2 to 32m/s 1.0% of rate C SS belt fasten with screws (1.0m4)
(300) 0 to 2m/s 0.02m/s <Acoustic coupler> (7th digit)
Y None
A
Mounting belt: according to specified code of symbol. Silicon rubber (KE348)
B Silicone-free grease (HIGH-Z)
Material: PBT, guide rail: aluminum alloy + plastic C Silicone grease (G40M)
Type of enclosure: IP52 <Water-proof treatment>(9th digit)
Acoustic coupler: according to specified code of symbol. Y None
<Tag plate> (10th digit)
Mass: 0.6kg, 0.8kg Y None
A Provided
OPTIONAL ACCESSORIES
Name Drawing No.
Sillicon grease (GM40M) ZZP*45231N5
Sillicon-free grease (HIGH-Z) ZZP*TK7M0981P1
Scope of delivery
D
etector, acoustic coupler and set of the mounting belt
according to specified code of symbol
OUTLINE DIAGRAM (unit: mm)
<Detector>
Saddle Saddle
Scale(inch) Scale(mm)
Connector
Connector
mm
Scale Scale
(inch) (mm)
Distance between two
Name prate
sensors 0 to 133
Spacing: 0 to 348
3202
Lock nut
5402
Element holder
Lock nut
Cursor
inch
Element holder
Cursor
36
90 max.
360.5
90max.
52.50.5
52.50.5
Weight : Approx. 0.6kg Weight : Approx. 0.8kg
Specification
Sensor frequency: 2MHz
Mounting method: V method (50 to 250mm) or Z method
CODE SYMBOL
(150 to 400mm)
<Detector>
Fluid temperature: 40 to 200C 1 2 3 4 5 6 7 8 9 10
A pplicable pipe material: PVC, SS, carbon steel pipe, F S SH1 1 Description
copper pipe, aluminum pipe,etc. <Senser type>(4th digits)
H 50 to 400mm (-40 to 200C)
[In case lining is removed from the pipe, Measurement
<Guide rail>(5th digits)
can not be conducted]
1 Provided
Rated accuracy of combination with the flow transmitter
<Mounting belt>(6th digits)
(Applicable piping: plastic,metal pipe) Y None
A Stainless belt (1.5m2)
Internal diameter Velocity Accuracy C SS belt fasten with screws (1.0m4)
(mm) <Acoustic coupler> (7th digit)
50 to 300 2 to 32m/s 1.0% of rate Y None
0 to 2m/s 0.02m/s D High-temperature grease (KS62M)
300 to 400 0.75 to 32m/s 1.0% of rate <Water-proof treatment>(9th digit)
Y None
0 to 0.75m/s 0.0075m/s
<Tag plate> (10th digit)
Y None
Mounting belt: according to specified code of symbol.
A Provided
Material: sensor housing: SUS304
guide rail: SUS304 + aluminum alloy
Type of enclosure: IP52
Acoustic coupler: according to specified code of symbol. OPTIONAL ACCESSORIES
Mass: 1.6kg Name Drawing No.
Guide rail for high-temperature sensor ZZP*TK4J5917C3
(Z method)
High-temperature grease(KS62M) ZZP*TK7G7983C1
OUTLINE DIAGRAM (unit: mm)
<Detector>
Connector Saddle
Scale(inch)
Scope of delivery
Scale(mm)
D
etector, acoustic coupler and set of the mounting belt
according to specified code of symbol
Name plate
two sensors 0 to 330
Distance between
5302
Lock nut
Element holder
Cursor
26 330.5
90max.
205max.
441
13
FSC-2, FSS, FSD
Specification
Sensor frequency: 0.5MHz
CODE SYMBOL
Mounting method: V or Z method
<Detector>
Fluid temperature: 40 to 80C 1 2 3 4 5 6 7 8 9 10
A pplicable pipe material: PVC, SS, carbon steel pipe, F SSE1 1 Description
copper pipe, aluminum pipe,etc. <Senser type>(4th digits)
E 200 to 6000mm
* In case lining is removed from the pipe, Measurement <Guide rail>(5th digits)
can not be conducted 1 Provided
Also applicable to water-proof type according to specified <Mounting belt>(6th digits)
code of symbol (submerged resistant structure for 5days Y None
D Wire 1500mm
including 10m cable) E Wire 6000mm
Rated accuracy of combination with the flow transmitter <Acoustic coupler> (7th digit)
(Applicable piping: plastic, metal pipe) Y None
A Silicon rubber (KE348)
Internal diameter Velocity Accuracy B Silicone-free grease (HIGH-Z)
C Silicone grease (G40M)
(mm)
200 to 300 2 to 32m/s 1.5% of rate <Water-proof treatment>(9th digit)
Y None
0 to 2m/s 0.03m/s B Provided (with signal cable 10m)
300 to 1200 0.75 to 32m/s 1.5% of rate <Tag plate> (10th digit)
0 to 0.75m/s 0.0113m/s Y None
1000 to 6000 1 to 32m/s 1.0% of rate A Provided
0 to 1m/s 0.02m/s
Name plate
Scope of delivery
D etector, acoustic coupler and set of the mounting belt
according to specified cord of symbol
841
Mounting spring
150
671
19
14
Checked items before purchase
Following conditions may cause failure of the measurement
or to reduce the accuracy by this flow meter.
Please consult and ask Fuji Electric for checking with actual
equipment previously if you have hard to judge the appropri-
ate application.
1)Fluid
I f fluid contains a large amount of bubbles (approx.
12vol% or more at 1m/s flow rate)
If fluid has bad turbidity 10000(mg/L) or more,
If fluid contains slurry or solid materials (about 5wt%)
If flow rate is low Reynolds No.10000 or less,
(reference: flow rate 5m3/h with 100mm)
If it is circulating oil, liquid medicine of low concentration,
waste liquid and hot spring,
2)Pipe
If inside pipe is rusty carbon steel pipe,
If inside pipe having adhering substances and sediment
If outer surface of cast-iron pipe is rough,
If pipe wall is tick such as ruinous pipe, (PP material
15mm or more, PVDF material 9mm or more)
If it is SGPW pipe,
If lining pipe is removed from pipe,
If it is rubber pipe,
3) Length of the straight pipe
For accurate measurement, straight pipes are needed
between up and down stream side of the measuring part.
Please meet the straight pipe conditions according page2.
Caution on use
1) Do not damage the sensor or signal mounted on the pipe.
2) Make sure to fill the fluid inside the pipe to measure .
3) When you use horizontal pipe, it is recommended to install
the sensor horizontally.
4) When you use the grease as acoustic coupler to install
the sensor for outdoor use, it is recommended to install
the waterproof cover to prevent from the degradation.
15
FSC-2, FSS, FSD
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.