You are on page 1of 5

Int. J. Radiation Oncology Biol. Phys., Vol. 71, No. 1, Supplement, pp.

S113S117, 2008
Copyright 2008 Elsevier Inc.
Printed in the USA. All rights reserved
0360-3016/08/$see front matter

doi:10.1016/j.ijrobp.2007.10.001

QA FOR RT SUPPLEMENT

HELICAL TOMOTHERAPY QUALITY ASSURANCE

JOHN BALOG, PH.D.,* AND EMILIE SOISSON, M.S.y


* Department of Radiation Oncology, Mohawk Valley Medical Physics, Rome, NY; and y Department of Human Oncology,
University of Wisconsin, Madison WI

Helical tomotherapy uses a dynamic delivery in which the gantry, treatment couch, and multileaf collimator leaves
are all in motion during treatment. This results in highly conformal radiotherapy, but the complexity of the deliv-
ery is partially hidden from the end-user because of the extensive integration and automation of the tomotherapy
control systems. This presents a challenge to the medical physicist who is expected to be both a system user and an
expert, capable of verifying relevant aspects of treatment delivery. A related issue is that a clinical tomotherapy
planning system arrives at a customers site already commissioned by the manufacturer, not by the clinical phys-
icist. The clinical physicist and the manufacturers representative verify the commissioning at the customer site
before acceptance. Theoretically, treatment could begin immediately after acceptance. However, the clinical phys-
icist is responsible for the safe and proper use of the machine. In addition, the therapists and radiation oncologists
need to understand the important machine characteristics before treatment can proceed. Typically, treatment
begins about 2 weeks after acceptance. This report presents an overview of the tomotherapy system. Helical
tomotherapy has unique dosimetry characteristics, and some of those features are emphasized. The integrated
treatment planning, delivery, and patient-plan quality assurance process is described. A quality assurance protocol
is proposed, with an emphasis on what a clinical medical physicist could and should check. Additionally, aspects of
a tomotherapy quality assurance program that could be checked automatically and remotely because of its
inherent imaging system and integrated database are discussed. 2008 Elsevier Inc.

INTRODUCTION structed to provide an image that can be registered with the


planning reference image (4, 5).
The helical TomoTherapy Hi-ART system was designed for
The integrated, dynamic nature of helical tomotherapy
dedicated integrated image-guided intensity-modulated
requires quality assurance (QA) that is different, than that
radiotherapy (IMRT). Helical tomotherapy is characterized
for conventional linear accelerators. Although the dynamic
by radiation delivery in which the patient is constantly trans-
lated through a continuously rotating radiation beam components and synchrony requirements add new challenges
mounted on a slip-ring gantry (13). A short in-line 6-MV to QA (68), other design simplifications make the process
linear accelerator and its associated radiofrequency system easier. For example, the collimator does not rotate, no
are mounted on a computed tomography (CT)-like ring gan- wedges or other accessories are used, and the couch does
try. A primary collimator shapes the radiation beam to be 40 not rotate. At present, only three longitudinal field widths,
cm in the transverse direction and 5 cm in the longitudinal 5, 2.5, and 1.0 cm, have been commissioned into the planning
direction (into the gantry bore), creating a fan beam. A set system. All field widths are collimated by the single move-
of moveable jaws further collimates the fan beam down to able set of jaws. Only one photon energy is provided for treat-
as little as 1 cm wide in the longitudinal direction, and ment (nominally 6 MV), and no electron mode is available.
a binary multileaf collimator (MLC) continuously provides Furthermore, without a flattening filter and associated beam
intensity modulation during the helical radiation delivery. steering, it is less likely that related beam profile asymmetries
The MLC has 64 leaves across the transverse beam direction. will arise.
Image guidance is accomplished with an on-board megavolt- Specific QA protocols vary among users but most should
age CT (MVCT) detector mounted on the gantry opposite the follow American Association of Physicists in Medicine
linear accelerator. The linear accelerator is detuned to operate Task Group report 40 guidelines (9), applicable. Conven-
at approximately 3 MV when used as an imaging source. tional external beam QA is component based such that the
Data collected in the detector during the scan are recon- clinical physicist starts measuring basic machine parameters

Reprint requests to: John Balog, Ph.D., Department of Radiation TomoTherapy Inc; and E. Soisson receives financial support from
Oncology, Mohawk Valley Medical Physics, 127 Dixon Dr., Rome, TomoTherapy Inc.
NY 13440. Tel: (315) 271-9614; (315) 671-1803; E-mail: JPB@ Received Feb 17, 2007, and in revised form Sept 13, 2007.
mvmp.net Accepted for publication Oct 2, 2007.
Conflict of interest: J. Balog owns common stock in
S113
S114 I. J. Radiation Oncology d Biology d Physics Volume 71, Number 1, Supplement, 2008

(i.e., profiles, percentage depth dose [PDD], output factors) TomoImage verification also occurs during commission-
and then progresses to planning system commissioning and ing. A scan of the TomoTherapy-provided cylindrical phan-
plan verification. Tomotherapy QA is process orientated. tom, cheese phantom, with a set of density and resolution
The focus is on verification of dynamic treatment delivery. plugs inserted should be scanned to ensure that all plugs are
At present, it is important for physicists to take responsibility visible on a normal scan with no ring, streak, or button
to thoroughly educate themselves about the system and pro- artifacts present in the image. The scan dose should be mea-
cess to determine first what can go wrong and then what tests sured at this time and be <2.0 cGy. The largest three rows of
to perform, and the associated specifications, to prevent any holes on the resolution plugs should be able to be distin-
adverse consequences. guished from their neighbors.
As is consistent with other treatment units, commissioning
includes taking a full set of profiles in the transverse and lon-
TOMOTHERAPY TREATMENT PLANNING
gitudinal directions at several depths using a water tank, in
COMMISSIONING AND ACCEPTANCE TESTING
addition to a PDD measurement for each field width. The pro-
In contrast to conventional systems, TomoTherapy sets the files collected with a water tank system should match the
output of the linear accelerator and the configuration of the standard model. If the profiles do not match, the beam-line
collimation only to a fixed standard. This defines a specific parameters should be adjusted until the shapes agree to
energy fluence model for its convolution-superposition- a gamma value of 1 that corresponds to a 2% dose difference
based planning system that depends on the slice width (1, and 2 mm distance-to-agreement value. The 25% width of the
2.5, or 5 cm width defined at the isocenter). A full set of transverse profiles at 15 cm should match to within 1% of the
beam data is collected at the factory, alignment tests are reference values. Likewise, the 50% field width should be
performed, and the machine is adjusted to meet in-house within 1% of the standard. The PDD profiles should be within
specifications for agreement with the standard model. Clini- 2% of the reference values from 10 to 200 mm.
cal physicists should receive copies of those data for eventual Beam stability should be checked to ensure a rotational
comparison with the annual QA data. On-site acceptance test- variation of less than 2 mm. The output should be measured
ing of the machine consists of a set of alignment and dosimet- in centigrays per minute at a 1.5 cm depth and 85 source-
ric verification tests. The site acceptance testing procedure is to-axis distance, and the monitor chambers should be cali-
performed by the clinical physicist in cooperation with an on- brated so that 1 monitor unit is approximately equal to 1
site TomoTherapy representative. cGy on this display, although this is not necessary to ensure
The linear accelerator and MLC mechanical alignment is correct output. Finally, the beam ramp-up time should be
tested first. The MLC offset must be less than 1.5 mm. It determined to be <10 s.
is also necessary to ensure that the linear accelerator is cen- After the static beam measurements have been verified,
tered with respect to the primary collimator in the longitudi- IMRT delivery verification can be performed using the
nal direction to within 0.5 mm. Finally, it is important to standard factory IMRT plans. This set of six plans is set
ensure that the beam does not diverge in the longitudinal up the same at every site. Two plans are run for each field
direction more than 0.5 mm. Also, the jaw and MLC twist width, one with an on-axis target and the other with an off-
should be less than 0.5 . axis target. Six measurements are taken across a plane
The treatment machine has a stationary green laser system across the central axis, with two measurements in the
that is centered on the virtual isocenter, which is 70 cm out of high-dose region and four outside the target volume in gra-
the bore from the actual isocenter. This facilitates patient dient regions. The measured dose should be within 3% of
setup, because it is done unencumbered by the bore. All green the calculated in-treatment region and within 3 mm in the
lasers must be accurate to within 1 mm. The home positions gradient regions.
of the moveable red lasers are matched to this location to Before commissioning is complete, baseline values should
within 1 mm. be acquired that can be used for future testing. Once the
It must be verified that the couch moves perpendicular machine has met all the aforementioned specifications,
to the gantry plane with a divergence in the International a Rotational Variation procedure is run in which data
Electrotechnical Commission-X direction of <2 mm. Diver- are collected from the detector during a 5-min procedure in
gence must also be <2 mm with the couch moving to its which the gantry is rotating constantly for 5 min while the
full extent in the International Electrotechnical Commis- field is opened to the maximal jaw setting (42 mm). The
sion-Z direction. The cobra motion of the couch is also tested energy fluence is collected by the on-board detectors and
at this time. The couch must be within 2 mm of its expected compared with the standard data set regularly by the Tomo-
horizontal position throughout the useful vertical range. In ad- Therapy field service engineer.
dition, the couch must be level in its fully retracted position.
The MLC characteristics are tested as a part of the machine
TOMOTHERAPY QA
commissioning. The MLC response is quantified during
commissioning by way of detector analysis (10). The leaves Daily machine QA
do not move in and out of the fan beam instantaneously but Daily QA can be performed by qualified radiation thera-
have latencies resulting from their transit time of about 20 ms. pists but must be under the direction of a qualified medical
Helical TomoTherapy QA d J. BALOG AND E. SOISSON S115

physicist. Daily QA should verify the MVCT image quality, purposes only. Using density and image resolution plugs
red and green laser positions, visible and audible indica- in the cheese phantom, high- and low-contrast resolution,
tors, couch translation accuracy, energy, and integral dose noise, and uniformity can be quantified, if desired.
output. A measure of beam energy by sampling the output ratios at
After running an Airscan procedure to calibrate the de- two depths should be obtained. The monthly energy ratio
tector array, MVCT image quality can be confirmed on a daily should be within 2% of the baseline. TomoTherapy recom-
basis by noting the presence of any new imaging artifacts. mends taking measurements at 10- and 20-cm depths in solid
Red and green lasers are generally checked by noting their water and comparing both to a measurement at 1.5 cm for
position relative to some baseline position. Sites that choose completeness. In addition, the static integral dose output at
to treat any patients without MVCT image guidance should 1.5 cm should be within 2% of the expected value.
be more diligent about daily laser testing. Intensity-modulated RT plan verification should be per-
Daily output measurements should be integrations of the formed using a standard plan. An absolute dose measurement
dose as the chamber and phantom pass through the beam. should be taken in the high-dose region and should be within
This IMRT delivery verification should be in lieu of (or in 3% of the calculated dose. It is important to test the plans
addition to) the static output measurements typically made run for all commissioned field widths. Verifying a plan run
at the beam central axis, because the dose delivered to any with one field width does not ensure that plans run with all
point is the integration of dose from the start to the end of field widths are within the specifications. All field widths
patient motion. Integral dose measurements made with the should be verified in this manner on a schedule deemed rea-
phantom setup at the virtual isocenter test the couch and laser sonable by the user.
accuracy, the beam output, and the longitudinal field width. The moving couch presents many QA challenges. The pa-
Daily output checks can be done for one treatment field width tient would be overdosed 1% if the couch traveled 1% too
with a tolerance of 3% for physicist notification and 5% for slow, because all treated areas would be within the primary
halting treatment. beam 1% longer. Additionally, the patient areas would even-
In addition to a rotational output check, daily energy ver- tually be mistreated as the couch-speed error continued. A
ification is also recommended, as well as a check of the 5% speed error for a total treatment length of 20 cm would
displayed monitor units per minute. Energy is generally mea- result in a 1-cm positional error at the end of treatment.
sured by taking a ratio of the measurements at two different The most dangerous error possible with tomotherapy would
depths and ensuring that the ratio of these values is within be for the couch to stop during treatment. Therapists should
2% of the baseline. Any major shifts in the monitor units be instructed to verify that the couch position continuously
per minute (more than 2%) should also be reported to changes during treatment by observing the graphical user
a physicist. interface readout. However, that relies on the manufacturers
system properly displaying such information and is not inde-
Monthly machine QA pendent. Visual verification is difficult because of the slow
A qualified clinical medical physicist should perform QA speed (<1 mm/s). The manufacturer has extensive verifica-
tests on a monthly basis. The monthly QA should include tion in the couch control, but clinical testing is important.
more comprehensive tests of those parameters included in The clinical physicist should independently verify couch
the daily tests, as well as testing couch motion integrity and motion during treatment at least once a month.
machine synchrony. Monthly integral dose output and longitudinal beam profile
Again, all visible and audible indicators should be checked shapes should be checked at a single depth for each commis-
monthly. The bunker door is interlocked and should be in- sioned field width, because they are so important for the inte-
cluded in the monthly QA program. Likewise, the numerous gral dose delivered. A 1-mm profile change for a 10-mm field
emergency-off buttons should be cycled through the monthly width would result in an approximate 10% dose change. The
QA program. profile shape should be analyzed for each commissioned field
More comprehensive laser testing should be done at the
width. This should include an analysis of the full width half
monthly QA testing. An easy test is to center the cheese phan-
maximum of each profile to submillimeter accuracy. Gener-
tom on the virtual isocenter and then acquire TomoImages.
ally, field widths are checked using one of two methods. To-
Several slices should be selected in front of the virtual isocen-
ter and the same number behind it. The resultant images pographic procedures using an ion chamber passing through
should confirm that the phantom image is centered. The radi- the beam and resultant charge profiles can be measured. In ad-
opaque markers present on the phantom center should be dition, film can be used by irradiating it with a static field for
viewable on the center slice. all three field widths. With the film at 85 cm source-to-axis
The extent of imaging QA should depend on how the sys- distance, the full width half maximum of the Y-axis film pro-
tem is being used. If MVCT images are regularly used for file should match the nominal field to within 1%.
treatment planning, it is important to ensure that the CT num- Some measure of the transverse profile shape should be
bers remain constant and that the images are void of any quantified on a monthly basis to ensure that no indications
significant artifacts. It is necessary only to perform a qualita- of target failure are present. Off-axis ratios should be within
tive assessment of overall image quality for image fusion 2% of expected values. One method to record lateral profile
S116 I. J. Radiation Oncology d Biology d Physics Volume 71, Number 1, Supplement, 2008

constancy is to use film. Devices are also available from var- verse field shape, and MLC response could all change
ious manufacturers for this purpose. within tolerances, but the combined effect on a treatment
The couch and gantry synchrony can be tested using film might not be clear. The clinical physicist should anticipate
as described by Fenwick et al. (6). Alternatively, they can this possibility and design a few IMRT plans to benchmark
be tested using the TomoTherapy-supplied multichannel treatment delivery.
electrometer that can sample as fast as every 100 ms, which
allows output vs. gantry position and output vs. couch posi-
tion to be analyzed (7). Patient-specific QA
Patient-specific QA starts by transferring the delivered
Annual machine QA dose distribution onto a phantom using an integrated tab
Annual QA measurements should be performed by aquali- within the planning station. The user selects a previously
fied clinical physicist. Annual testing should mimic the tomo- scanned phantom, positions the phantom with respect to
therapy commissioning procedures and specifications but be the machine geometry so that is at a convenient location for
adapted to use the equipment and QA tools available to the dose measurement, and then recalculates the patient dose
physicist performing the tests. The physicist should refer to on that phantom. Most users choose the cheese phantom
the machine commissioning section for specific procedures for QA, in which films can be irradiated in the coronal or sag-
and specifications. Some details of the annual QA testing ittal orientation to acquire a planar representation of the dose
are described in the following paragraphs. distribution. The absolute dose is measured with an A1SL ion
First, static beam data should be collected using a water chamber that can be positioned anywhere across the central
tank. Longitudinal profile shapes at the standard commis- axis of the phantom. Separate QA procedures are created
sioning depths (1.5, 5.0, 10.0, 15.0, and 20.0 cm) should be from the delivery quality assurance plan on the tomotherapy
measured and compared with initial commissioning data. planning station that can then be delivered at the operators
The transverse field profiles, cone profiles, at commission- station. The QA phantom can be imaged first for setup veri-
ing depths should also be measured. A Task Group report 51- fication with the planned phantom position using image fu-
type calibration can be performed for a static beam and has sion. After the film is exposed and processed, the film
been described previously (12), or another check of machine analysis features included in the planning system can be
output should be performed. used with a film image and associated film calibration file.
Annual measurements should include a check of the The analysis tools include profile and isodose comparisons
beam geometry (11). All tests described in the machine and a gamma calculation. The absolute measured dose with
commissioning should be included. The linear accelerator an ion chamber located in a homogenous high-dose region
and MLC alignment can be checked as described in To- should be 5% of the calculated dose.
moTherapy Treatment Planning Commissioning and Ac-
ceptance Testing but using film instead of the detector if
CONCLUSION
no software to analyze the detector data is available. Most
of the necessary .xml files for the tests performed in The tomotherapy machine and treatment process differs
commissioning should be located in the hard drive of the from conventional treatment accelerators in a number of
operators station computer, if they are not already listed ways, making QA for this unit different from that for con-
in the database, and can be used in the annual testing. ventional accelerators. Clinical physicists need to under-
The beam geometry should be consistent with the commis- stand the tomotherapy differences and similarities to fulfill
sioning specifications. their role in ensuring correct technical treatment delivery.
Image parameters such as resolution, geometric accuracy, Helical tomotherapy involves many moving parts. QA pro-
density, and dose should also be verified. Again, the extent of grams should test them collectively. We presented a QA
this testing is up to the discretion of the user. program for daily, monthly, and annual QA testing that in-
Demanding, but clinically realistic, IMRT plans should cluded procedures, specifications, and schedules, but each
be delivered either annually or whenever beam changes clinical physicist is responsible for the actual QA program
are suspected. The PDD, longitudinal field shapes, trans- in their clinic.

REFERENCES
1. Mackie TR, Holmes T, Swerdloff S, et al. Tomotherapy: A new 4. Ruchala K, Olivera GO, Schloesser EA, et al. Megavoltage CT
concept for the delivery of conformal radiotherapy. Med. Phys on a tomotherapy system. Phys Med Biol 1999;44:25972621.
1993;20:17091719. 5. Ruchala K, Olivera GO, Kapatoes J, et al. VCT image recon-
2. Mackie TR, Balog J, Ruchala K, et al. Tomotherapy. Semin struction during tomotherapy. Phys Med Biol 2000;45:
Radiat Oncol 1999;9:108117. 35453562.
3. Yang JN, Mackie TR, Reckwerdt P, et al. An investigation 6. Fenwick JD, Tome WA, Jaradat HA, et al. Quality assurance for
of tomotherapy beam delivery. Med Phys 1997;24: a helical tomotherapy machine. Phys Biol Med 2004;49:
425436. 29332955.
Helical TomoTherapy QA d J. BALOG AND E. SOISSON S117

7. Balog JP, Holmes T, Vaden R. Helical tomotherapy dynamic 10. Balog JP, Olivera GO, Kapatoes J. Clinical helical tomotherapy
quality assurance. Med Phys 2006;33:39393950. commissioning dosimetry techniques. Med Phys 2003;30:
8. Parham A, Susanta H, Patrick H, et al. The use of a commercial 30973106.
QA device for daily output check of a helical tomotherapy unit. 11. Balog JP, Mackie TR, Pearson D, et al. Benchmarking beam
Med Phys 2006;33:36803682. alignment for a clinical helical tomotherapy device. Med Phys
9. Kutcher GJ, Coia L, Gillin M, et al. Comprehensive QA for 2003;30:11181127.
radiation oncology: Report of AAPM Radiation Therapy Com- 12. Jeraj R, Mackie TR, Balog J, et al. Radiation characteristics of
mittee Task Group 40. Med Phys 1994;21:581618. helical tomotherapy. Med Phys 2004;31:396404.

You might also like