Professional Documents
Culture Documents
Research Article
GC Method Validation for the Analysis of Menthol in
Suppository Pharmaceutical Dosage Form
Murad N. Abualhasan,1 Abdel Naser Zaid,1 Nidal Jaradat,1 and Ayman Mousa2
1
Department of Pharmacy, Faculty of Medicine & Health Sciences, An Najah National University, Nablus, State of Palestine
2
R&D Department, Avalon Pharma, Riyadh, Saudi Arabia
Received 26 November 2016; Revised 22 January 2017; Accepted 1 February 2017; Published 6 March 2017
Copyright 2017 Murad N. Abualhasan et al. This is an open access article distributed under the Creative Commons Attribution
License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly
cited.
Menthol is widely used as a fragrance and flavor in the food and cosmetic industries. It is also used in the medical and
pharmaceutical fields for its various biological effects. Gas chromatography (GC) is considered to be a sensitive method for the
analysis of menthol. GC chromatographic separation was developed using capillary column (VF-624) and a flame ionization
detector (FID). The method was validated as per ICH guidelines for various parameters such as precision, linearity, accuracy,
solution stability, robustness, limit of detection, and quantification. The tested validation parameters were found to be within
acceptable limits. The method was successfully applied for the quantification of menthol in suppositories formulations. Quality
control departments and official pharmacopeias can use our developed method in the analysis of menthol in pharmaceutical dosage
formulation and raw material.
ratio of the peak area of Menthol and Thymol of the replicate Table 2: System suitability and system precision result.
injections of standard solution should have an RSD not more
than 2.0%. Injection Area of menthol Area of thymol
Ratio
number (mVot) (mVot)
1 76756 124806 0.615
2.5.2. Linearity and Range. In order to evaluate the lin-
2 76491 124418 0.615
earity of assay procedure, a series of standards at different
concentrations of the target concentration for menthol was 3 74717 122117 0.612
prepared in the range of 0.30.9 mg/mL which corresponds 4 76281 125238 0.609
to 50%150% relative to menthol measuring concentration in 5 75446 123684 0.610
standard solution. After chromatographing each preparation 6 74402 121399 0.613
in triplicate, a linear regression analysis was performed on 7 75591 124163 0.609
the average peak ratio versus the concentrations of the levels
8 73692 121011 0.609
studied.
The correlation coefficient was calculated by plotting 9 78280 125238 0.625
component average peak ratio versus component concen- 10 78409 125059 0.627
trations. Linear regression was applied to the plots and the AVG 76007 123713 0.614
correlation coefficients for component data were calculated. RSD% 2.052 1.308 1.071
In order for the test to pass the square of correlation
coefficient should not be less than 0.998.
The limit of detection (LOD) and limit of quantification 9.0 8.29
Thymol
(LOQ) were calculated based on the standard deviation (s.d) 8.0
of the data and the slop of the regression line. The LOD (mVol) 7.0
was calculated according to the following equation: 3.3 s.d 6.0
5.0 1.21
/slope. The LOQ was calculated according to the following 4.0 6.65
equation: 10 s.d/slope. 3.0
Menthol
2.0 4.59
1.0
2.5.3. Accuracy. The accuracy was performed based on three 0
concentrations around the test concentration (80%, 100%, 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0
and 120%); three replicates of each concentration were Time (min)
injected. The percentage of recovery and percentage of RSD
Figure 2: GC chromatogram of menthol and internal standard
were calculated for each of the repeated samples. Thymol.
2.6. Stability of Analytical Solution. The stability of analytical 3.2. Method Validation
solution was determined by analyzing menthol in both the
standard solutions and sample solution over 24-hour period. 3.2.1. Linearity and Range. The correlation coefficient was
The percentage of recovery of menthol was calculated versus calculated by plotting component average peak ratio versus
the fresh injections for both standard solutions and sample component concentrations and was found to be 0.9998.
solutions. In order to prove the stability of the solution the Linear regression was applied to the plots and the slop was
standard deviation of all the test solutions must be within found to 0.9775 with an intercept of 0.0112 (Table 3 &
2.0%. Figure 3).
4 International Journal of Analytical Chemistry