Professional Documents
Culture Documents
pubs.acs.org/ac
Pzer Worldwide Research and Development, Statistics, 558 Eastern Point Road, Groton, Connecticut 06340, United States
Pzer Worldwide Research and Development, Analytical Research and Development Department, 558 Eastern Point Road, Groton,
Connecticut 06340, United States
Pzer Worldwide Research and Development, Analytical Research and Development Department, 674 Ramsgate Road, Sandwich,
Kent CT13 9NJ, United Kingdom
Pzer Consumer Healthcare, Analytical Development, 1121 Sherwood Avenue, Richmond, Virginia 23220, United States
2014 American Chemical Society 11930 dx.doi.org/10.1021/ac503551r | Anal. Chem. 2014, 86, 1193011936
Analytical Chemistry Perspective
testing. A few of these monograph assay methods state to use Table 2. These components comprise the standard error (SE)
not less than 10 units. Other monograph methods make no of the reportable potency assay and may be termed as the
mention of the number of dosage units and only state to place a repeatability of the assay. Additional variance components enter
suitable number of tablets or capsules into a suitable volumetric into the totality of uncertainty of data generated by an assay
ask to yield a specic concentration. There exists no method, such as analyst, instrument, laboratory, etc.7 These
consensus, nor does there appear to be a justied rationale components must also be understood when making a decision
for the number of dosage units selected in these methods. of lot release, for example.
An appropriate number of units in the composite sample and An illustration of these method variance components is
number of sample replicates can be determined as a component displayed through a uniformity of dosage units (UDU)
of the variability minimization strategy of the assay result for a experiment. As shown in eq 1, total variability of a typical
given batch. In cases where the dosage unit has high content uniformity of dosage units experiment consists of at least the
variability, increasing the number of units in the composite following variance components, sample preparation, injection
sample preparation will reduce the variability in the assay result. precision, standard preparation, dosage unit, etc., and may
Testing replicate sample preparations can reduce variability for include other sources such as day and or analyst.
cases where there is high assay variability due to the analytical
method itself. It is important to understand the composition of
UDU 2 = samplepreparation 2 + injection precision 2
the variability in potency assays in order to eectively
determine an appropriate number of dosage units in the + standard preparation 2 + dosage unit 2 + ... (1)
composite sample preparation and an appropriate number of
sample replicates to generate the assay value for a given batch.
To minimize method uncertainty, it is important to focus on
In this work, a sampling strategy was developed for
reducing the largest contributors to this total variability by
preparation of sample replicates and composites of assay test
understanding the method settings (sample agitation rate,
methods for solid oral drug products by chromatographic
analysis. This strategy accounts for variability due to the injection volume, etc.) that minimize these contributors as
analytical method and dosage form and uses the most illuminated through appropriately designed experiments. When
restrictive compendia and regulatory requirements2,46 for these variance components can no longer be reduced from a
potency assay sampling and variability to determine an mechanistic understanding, an additional strategy is to dene a
appropriate sample strategy. sampling scheme or replication strategy that minimizes those
UDU 2 = method 2 + dosage unit 2 (1a) Total Variability: Experimental Model, Uniformity of
Dosage Units. In a uniformity of dosage (UDU) experiment,
Why Are We Concerned with Uniformity of Dosage an individual dosage unit (dui) is prepared into an individual
Units in a Potency Assay? To determine a sampling scheme sample preparation (Si). Scheme 2 illustrates such an
that minimizes, to an appropriate level, the contribution of the experimental run.
method and dosage unit components to the overall potency
assay variability, consider a potency assay method experiment Scheme 2. Illustration of the Samples Prepared to Determine
that produces a nal reportable value (Y). A potency assay Uniformity of Dosage Units
consists of a number of dosage units (duj) prepared into a
sample (Si) of which there may be a number of replicate (r)
sample preparations composited to form the reportable assay
value. Scheme 1 illustrates such an experimental run.
Yi = + i + i i = 1 to r (3)
Equation 3 illustrates the signal (Y) representing the amount
of an analyte with a true mean () and individual signals which
vary about this mean according to the contribution of method
(i) and dosage unit (i) eects. The eects i and i are
Each sample preparation (Si) consists of a preparation of k assumed to be independent of one another and are eects from
dosage units that may be added as whole units or ground into a random sample with population mean of 0 and variances of
powder and equivalently weighed. The analyte amount derived 2 and 2, respectively. The variability of an UDU value is
from the sample matrix calculation can be represented by eq 2.8 then as follows in eq 3a.8
Yij = + i + ij i = 1 to r ; j = 1 to k (2) var(Y ) = ( 2 + 2) (3a)
Equation 2 illustrates the signal representing the amount of Note: 2 can be composed of both potency and weight of
an analyte (Y) with a true mean () value and individual signals dosage unit variability.
which vary about this mean according to the contribution of As in the potency assay, the error attributed to method and
method (i) and dosage unit (ij) eects. The eects i and ij dosage units is inseparable: a single measurement is made on
are assumed to be independent of one another and are eects each sample preparation that consists of a single dosage unit.
from a random sample with population mean of 0 and variances To determine a sampling scheme (solving for r and k in eq 1a),
2 and 2, respectively. The variability of such an assay is then an estimate for each individual component must be obtained. A
illustrated in eq 2a, where (Y) is the reportable value for the general solution (If the UDU and potency assay method are the
potency assay that consists of the average of r sample equivalent method, then a simple algebraic solution exists for
preparations each composed of k dosage units. the solution of method and dosage unit variance estimates. The
1 1 viability of this technique is dependent upon the correctness of
var(Y ) = 2 + 2 the potency assay variability estimate.) that works for any
r rk (2a)
combination of UDU and potency assay method techniques is
The variability of the reportable potency assay value consists dened by obtaining independent variance component
of a contribution from both method (2) and dosage unit estimates.
variability (2). Because the reportable value (Y) is the only In the development of the potency assay method, there are
value observed (i.e., the individual dosage units are not several experiments that provide estimates of the method
observed since they are dissolved in the composite sample variability (2) alone. With an estimate of method variability
solution), the variability attributed to the potency assay cannot (2), an estimate of the dosage unit variability (2) can be
be decomposed into the error attributed to dosage units and solved via subtraction in eq 1a. One possible resource for
method using observed potency assay values alone. That is, providing the estimate of method variability (2) is the
these variability components are inseparable. accuracy assessment experiment during the methods develop-
The variability estimates of the method (2) and dosage ment and validation. This seems to be a logical choice since an
units (2) are required to determine the number of replicate accuracy assessment is usually executed early in the develop-
sample preparations and dosage units per sample preparation in ment of a method, hence providing the means to determine a
a sampling scheme: r and k in eq 2a. How do we obtain these sample composite strategy as early as the rst uniformity of
estimates if they cannot be determined from the observed dosage unit experiment is completed. See Figure 1 for a
potency values alone? The answer is through uniformity of owchart of a possible sampling scheme executed workow. At
dosage unit and other method development experiments such least one other strategy exists to estimate the method and
as an accuracy assessment experiment.9,10 Uniformity of dosage dosage unit variance components; however, it requires two
unit experiments provides an estimate of total reportable things. The rst is that the UDU and potency assay method are
potency assay value variability: method plus dosage unit as equivalent. Second, some minimum number of potency assays
illustrated in eq 1a, above. The section below illustrates a is completed on the same homogeneous material of the UDU
typical uniformity of dosage experiment to see the variance experiment such that the method variability is viable for
contributions in these experiments. calculations. Due to random chance, this might not always be
11932 dx.doi.org/10.1021/ac503551r | Anal. Chem. 2014, 86, 1193011936
Analytical Chemistry Perspective
Yi = + i* + i* i = 1 to r (4)
Equation 4 illustrates the signal representing the amount of
an analyte (Y) with a true mean () and individual signals
which vary about this mean according to the contribution of
method (i*) and measured spiked content (i*) eects. The
eects i* and i* are assumed to be independent of one
another and are eects from a random sample with a
population mean of 0 and variances of *2 and *2,
respectively. The variability of potency assay results from an
experiment described above is dened in eq 4a.
var(Y ) = ( *2 + *2) (4a)
The estimate of *2 in eq 3a represents the variability about
measured concentrations that are exceedingly precise. It is
logical, therefore, to assume that *2 is approximately zero.
Then, the variance of the observed accuracy values (eq 4a)
consist of only variance due to the concentration preparations
(*2). This value is thought to be a good approximation to 2
in eqs 2a and 3a as long as there exists no unrealized eect of
the dosage unit manufacture that is not captured in this spiking
experiment; i.e., *2 in eq 4a is not an exaggerated
underestimate of 2 in eqs 2a and 3a. Other experiments
may be utilized to determine an estimate of the method
variability (2); however, for most methods, a well-dened
accuracy experiment is the most complete and ecient means
for obtaining this estimate.
Illustrating the Sampling Strategy. With estimates of
*2 (from an accuracy experiment as dened above) and
uniformity of dosage unit variability consisting of the sum of
(2 + 2), the eect of increasing the number of dosage units
(k) and/or sample preparation replicates (r) can be determined
through eq 2a. See the case study below for an example of all
the data and calculations necessary to achieve this.
Figure 2 displays the eect of increasing the number of
dosage units (k) and sample preparation replicates (r) in eq 2a
a
Note: The red star pertains to the case study discussed below.
stopping criteria, labeled Target SE in Figure 2, can be Table 4. Uniformity of Dosage Unit Results for Product A
derived using the logic outlined in the Discussion section, Capsules
below.
CASE STUDY
The composites and replicates strategy was applied to a capsule
130 99.6 99.3 99.5
overall
99.5 (0.1)
100.1 (0.5)
formulation of product A, a product under development. As
shown in Table 4, uniformity of dosage unit data was available
for nine manufactured batches. Table 5 shows the results of Table 5. (b) Variance of method = method2 = (0.5 0.5) = 0.25.
accuracy experiments. The data in Tables 4 and 5 are used to (3) Estimate dosage unit variance (dosage unit2): (a) Dosage unit
calculate method and dosage unit variability as shown below. variance is obtained by subtraction using eq 1a.
This information is then used to determine the appropriate UDU 2 = method 2 + dosage unit 2
sampling scheme from Table 3.
Product A Capsules: (1) Calculate variance of UDU (UDU2):
6.3 = 0.25 + dosage unit 2
(a) Overall %RSD for UDU from Table 4 = 2.51; (b) Variance
of UDU = UDU2 = (2.51 2.51) = 6.3. (2) Calculate variance
6.05 = dosage unit 2
of method (method2): (a) Overall Method %RSD is 0.5% from
11934 dx.doi.org/10.1021/ac503551r | Anal. Chem. 2014, 86, 1193011936
Analytical Chemistry Perspective
(4) Estimate method variability and dosage unit variability as a the number of sample replicates and dosage units in the
percentage of total variability: composite sample preparations for a potency assay method was
selected on the basis of regulatory criteria for uniformity of
component 2 (% total) = (component 2/UDU 2) 100 dosage units.2,46
To develop the most restrictive criterion, the standard error
method 2 (% UDU) = (0.25/6.3) 100 = 4% of the drug product assay should be minimized. As dened by
eq 2a, the greater the number of dosage units prepared in a
dosage unit 2 (% UDU) = (6.05/6.3) 100 = 96% sample, the greater is the reduction in the variability of a
potency assay. This is a square root relationship; therefore, a
(5) Determine sampling scheme (number of sample prepara- balance between number of dosage units (and hence impact on
tion replicates and number of dosage units per sample sample preparation eorts due to volumes and subdilutions
preparation): (a) Using the overall %RSD from Table 1 required) and net gain is needed. The greatest number of
(2.5%) and the ratio of % sample preparation variability (4%) dosage units dened in a regulatory document is k = 20 dosage
and % dosage unit variability (96%) from step 4 above, nd the units as recommended in TGO 78.2 A greater number of
number of replicates and units from Table 3 (see star location). dosage units per sample preparation would further reduce the
(b) Sampling scheme is 1 replicate of composite of 5 dosage variability of the potency assay; however, the choice of k = 20 is
units (k = 5, r = 1).
DISCUSSION
A sampling scheme was developed as shown in Table 3 to
USP <905> Acceptance Value (AV) Criterion when n = 30
and Batch Mean is 95% or 105% Label Claim
xbar = 95 98.5 < xbar < 101.5 xbar = 105
determine an appropriate number of sample replicates and
5.75 7.50 5.75
dosage units in the composite sample preparation for a potency
assay. This sampling scheme is based on the sample preparation
variability, dosage unit variability and product UDU (%RSD), criterion of AV < 15. The values in Table 7 are derived by
and a criterion of SE of NMT 1.3. The basis for the criterion of solving for the standard deviation term in the AV equation for n
the SE is discussed below. = 30 dosage units and a specied xbar (s = (AV |M X |)/k).
Criterion. In order to determine a table akin to the one A batch mean of 95% (105%) label claim was chosen as this
illustrated in Table 3 above, a criterion for assessing the is a typical solid oral dosage form drug product release
adequacy of potency assay variability must be established. This specication. Then, the most conservative (smallest value)
criterion should contain the most restrictive elements of current criterion for SE of an assay can be computed as 5.75/sqrt(20) =
regulatory guidance in order to maximize the condence in 1.3, as illustrated in Table 8.
decisions made by such an assay. The above illustrates the basic components germane to
Establishing a Minimal Criterion for the Standard determine a criterion for the standard error of a potency assay.
Error of a Drug Product Assay. The sampling strategy for Criterion for the standard error of a potency assay other than
11935 dx.doi.org/10.1021/ac503551r | Anal. Chem. 2014, 86, 1193011936
Analytical Chemistry Perspective
Table 8. Maximum Standard Error of a Potency Assay Using (8) Kutner, M. H.; Nachtsheim, C. J.; Neter, J.; Li, W. Applied Linear
Standard Deviation of 5.75% and k = 20 Dosage Units Per Statistical Models, 5th ed.; McGraw-Hill/Irwin: Boston, 1996.
Preparation (9) ICH Q2(R1) Validation of Analytical Procedures: Text and
Methodology; ICH: Geneva, 1994.
xbar = 95 98.5 < xbar < 101.5 xbar = 105 (10) ICH Q2B Validation of Analytical Procedures: Methodology; ICH:
1.29 1.68 1.29 Geneva, 1996.
CONCLUSION
A sampling strategy for the number of sample replicates and
dosage units in the composite sample preparation for a potency
assay method was developed. This sampling strategy takes into
account the sample preparation variability, dosage unit
variability, and product UDU. It uses the criterion that the
standard error is not more 1.3. A table of solutions for the
number of sample replicates and the number of dosage units
per sample preparation was derived for any scenario of an
analytical method variability composition from such a criterion.
The sampling scheme is demonstrated on a case study. The
procedure illustrated allows a company to dene a sampling
strategy for drug product potency assay methods that conforms
to an internally dened minimal risk-based standard and allows
uncertainty to be communicated in regards to potency assay
decisions. This work provides the assay method basis of
variability understanding upon which additional components of
method variability (e.g., instrument, analyst, environment, etc.)
can now be determined through additional experimentation
such as during the analytical method transfer exercise.
AUTHOR INFORMATION
Corresponding Authors
*E-mail: brent.harrington@pzer.com.
*E-mail: beverly.nickerson@pzer.com.
Notes
The authors declare no competing nancial interest.
ACKNOWLEDGMENTS
The authors would like to thank Debbie Krause, Kim
Vukovinsky, and Loren Wrisley for their encouragement and
support of this work.
REFERENCES
(1) ICH Q6A Specications: Test Procedures and Acceptance Criteria for
New Drug Substances and New Drug Products: Chemical Substances;
ICH: Geneva, 1999.
(2) Therapeutic Goods Order No. 78 Standard for Tablets and
Capsules; Australian Government Department of Health and Ageing,
Therapeutic Goods Administration: Woden, Australia, 2008.
(3) USP 36-NF31 through Second Supplement; The United States
Pharmocopeial Convention: Rockville, MD, 2013.
(4) <905> Uniformity of Dosage Units in USP 36-NF31 through Second
Supplement; The United States Pharmocopeial Convention: Rockville,
MD, 2013.
(5) EP 2.9.40 Uniformity of Dosage Units in European Pharmacopoeia,
Supplement 8.2 to the 8th ed; European Directorate for the Quality of
Medicines & Healthcare: Strasbourg, France, 2014.
(6) 6.02 Uniformity of Dosage Units, in Japanese Pharmacopoeia
Supplement II, 16th ed.; The Ministry of Health, Labour and Welfare:
Tokyo, 2014.
(7) Ermer, J.; Agut, C. J. Chromatogr., A 2014, 1353, 7177.