Professional Documents
Culture Documents
DOI 10.1007/s10008-015-3105-3
ORIGINAL PAPER
Abstract This work describes the development of a simple, Keywords Prazosin . Pharmaceutical analysis . Boron-doped
fast and low-cost method for determining prazosin (PRA) in diamond . Multiple-pulse amperometry . Flow injection
pharmaceutical samples by flow injection analysis with analysis
multiple-pulse amperometric (FIA-MPA) detection using a
boron-doped diamond film electrode. Electrochemical detec-
tion of PRA was optimized in phosphate buffer pH 4.0 by Introduction
cyclic voltammetry, in which PRA presented two oxidation
processes around at 0.97 and 1.40 V versus Ag/AgCl Prazosin (PRA) is the prototype of a family of agents that
(3.0 mol L−1 KCl). In these conditions, PRA also showed contain a core of piperazinyl quinazoline and acts as a potent
one reduction process at −0.75 V that is dependent on the and selective antagonist of α1 receptors, promoting decrease
oxidation processes. Thus, the determination of PRA by in peripheral vascular resistance and venous return to the
FIA-MPA detection consisted on the application of a two- heart, which consequently leads to its therapeutic effect in
potential waveform, E 1 (generator potential) = 1.6 V/ the treatment of hypertension [1]. Its structural formula is
400 ms and E2 (collector potential) = −1.0 V/30 ms, with shown in Fig. 1.
sample loop of 150 μL and flow rate of 3.0 mL min−1. The PRA is a drug with narrow therapeutic index [2], which
method showed good repeatability (RSD < 3.0 %) and high makes its determination very important to analyze this ac-
analytical frequency (70 injections per h). The working tive ingredient in formulations for the quality control in
linear range was obtained from 2 to 200 μmol L−1 with a pharmaceutical formulations and in biological fluids for
limit of detection of 0.5 μmol L−1. The recovery tests in all pharmacological studies. The official methods for determi-
samples were approximately 100 %, and the results were nation of PRA in pharmaceutical formulations are de-
compared with chromatographic methods. scribed in USP and British Pharmacopoeia [3, 4], which
are based on chromatography (HPLC with UV detection).
It can also be found in the literature various methods for
determination of PRA in formulations and biological fluids
* Wallans T. P. dos Santos
using HPLC with UV detection [5, 6] and other different
wallanst@yahoo.com.br detectors: fluorescence [7, 8], rayleigh light-scattering [9],
and amperometric using glassy carbon electrode [10]. In
1
addition, other methods for determination of PRA, such
Departamento de Química, Universidade Federal dos Vales do
Jequitinhonha e Mucuri, Campus JK, 39100-000 Diamantina, MG,
as spectrophotometry [11], thin layer chromatography
Brazil [12], capillary electrophoresis [13], and electroanalytical
2
Departamento de Farmácia, Universidade Federal dos Vales do
techniques [14–17].
Jequitinhonha e Mucuri, Campus JK, 39100-000 Diamantina, MG, Among all these methods, the electroanalytical techniques
Brazil have shown to be an attractive alternativeness for the
J Solid State Electrochem
The addition-recovery studies of PRA were also carried out in for sensitive and selective detection of PRA. After cathodic
these pharmaceutical samples. and anodic electrochemical treatments applied to BDD elec-
trode, the electrochemical behavior of PRA was re-evaluated
Instrumentation and apparatus in the supporting electrolyte. The cathodic treatment was
chosen due to its better definition of the reduction peak
The electrochemical measurements (amperometric and observed for PRA. Therefore, all studies were performed
voltammetric) were performed using a Potentiostat from after cathodic electrochemical pretreatment of the working
Autolab (Eco Chemie, the Netherlands) model PGSTAT electrode surface. In these conditions, the cyclic voltam-
128 N with GPES 4.9 software. An electrochemical flow cell mograms are presented in Fig. 2.
(Bwall jet^ type) with three-electrode was lab-made. A wire of As can be seen in Fig. 2, there are two oxidation peaks
platinum and an Ag/AgCl with KCl saturated were used as (0.97 and 1.40 V) and one reduction peak (−0.75 V) for
auxiliary and reference electrodes, respectively. The working PRA on BDD electrode. The oxidation and reduction process-
electrode was a thin film (ca.1.2 mm) of BDD (approximately es were described by Arranz and co-authors, which used a
8000 ppm doping level) supported on a polycrystalline silicon carbon paste electrode modified with nafion [16]. These au-
wafer (NeoCoat SA, La Chaux-de-Fonds, Switzerland). One thors observed at first scan just one oxidation process, which
piece of BDD material (1.0 × 1.0 cm) was used in the flow posteriorly generated a quasi-reversible cathodic process.
cell, and the electrode area was delimited by using a rubber O- Although it could not be confirmed by cyclic voltammetry at
ring with 0.4 cm of diameter (electrode area = 0.13 cm2). BDD electrode that the reduction is a quasi-reversible electro-
Background current correction was carried out using the chemical process, FIA-MPA confirmed that this reduction
GPS software from Autolab. The electrochemical pretreat- process on BDD electrode presents similar electrochemical
ment of the BDD electrode for PRA detection was evaluated behavior to carbon paste electrode modified.
by cyclic voltammetry. The BDD electrode was anodically The electrochemical mechanism elucidated in the literature
pretreated in a 0.5mol L−1 H2SO4 solution by applying for PRA on carbon paste electrode proposes that the first ox-
0.001 A during 120 s. The cathodic pretreatment was carried idation current is controlled by adsorption of the analyte at the
out by applying −0.03 A during 360 s using the same solution. electrode surface [16]. According to these authors, this anodic
The BDD electrode was pretreated only once, prior to the process corresponds to the oxidation of the adsorbed species
measurements. of PRA, which can produce dimer compounds through the
The FIA system (single-line) was composed of polyethyl- oxidation of the amine group present in this molecule along
ene tubing with 0.5 mm internal diameter, an acrylic manual with a four-electron and four-proton transfer process. Thus,
injector and flow rate was controlled by the pressure generated the PRA oxidized form might be reduced by an azo group,
by a water column [38]. The flow rate and injection volume in an almost reversible process with two-electron and two-
were investigated in the range of 0.5 to 4.0 mL min−1 and 50 proton transfer [16]. We suggest that the electrochemical
to 350 μL, respectively. oxidation and reduction mechanism of the PRA on BDD
The comparison of the method proposed for PRA determi- electrode is similar to verified on carbon paste electrode.
nation in pharmaceutical samples was performed by the offi- However, the first oxidation process should occur in two
cial method using HPLC-UV [4]. The HPLC system used in steps on BDD electrode. Moreover, the plot of the current
this work was a Shimadzu (Kyoto, Japan). The column used for the oxidation peaks of PRA (0.97 and 1.40 V) as a
was C18 (10, 4.6, 5 mm) from Macherey-Nagel (Germany). function of the root of the scan rate (10 to 500 mV s−1)
The mobile phase was composed of methanol/water (70:30)
with 1 % acetic acid (v/v). All HPLC analysis were carried out
at room temperature (30 °C) with an injection volume of
20 μL and a flow rate of 1.0 mL min−1 (isocratic), while
detection was performed at 254 nm.
determination in pharmaceutical formulations and in human (−1.0 V). However, due to high concentrations of AA and
urine. The stability study of the BDD electrode coupled to the UA in urine samples, the concentration effect of these
FIA system was evaluated by detection of PRA at collector interferents on current signal of analyte also was evaluated
potential pulse (−1.0 V). The repeatability study (Fig. 4) was (Table 1). As can be observed in Table 1, the current signal
performed by 10 successive injections of solutions containing of PRA was relatively constant in the presence of up to 100
10 μmol L−1 PRA, then more 10 injections (with a concentra- times more of AA or UA. For higher concentrations of AA, a
tion 10 times higher than before) and, finally, more 10 injec- decrease in the PRA signal was observed. The UA was not
tions in the same initial concentration of PRA. soluble in the supporting electrolyte at higher concentrations,
As can be noted in Fig. 4, even after the injections of high and so, it did not interfere in the analysis of PRA.
PRA concentrations, the first 10 peaks did not differ signifi- The addition-recovery studies in pharmaceutical samples
cantly from the last 10 measurements. The RSD for PRA were carried out to verify the performance of the proposed
detection were 2.46 % for the first 10 peaks (Fig. 4a), method. The results obtained for recovery of PRA in formu-
2.85 % for 10 injections at high concentrations (Fig. 4b) and lations pharmaceutical (n = 3) and human urine (n = 3) were
2.79 % for the last 10 peaks (Fig. 4a). These results demon- 100.1 % (±1.0) and 98.0 % (±5.0), respectively. The values
strated that the FIA system using BDD electrode can present obtained were close to 100 %, indicating the absence of matrix
high reproducibility for PRA quantification, mainly because effect in these samples.
the phenomenon of working electrode contamination (or pas- The results obtained for the determination of PRA in phar-
sivation) is practically absent due to inert surface of the BDD. maceutical samples using the proposed method versus the
The amperometric responses for triplicate injections of so- official method (HPLC-UV) are presented in Table 2.
lutions containing increasing concentrations of PRA and the Statistical tests (F and t) were applied to results obtained by
respective calibration curve are presented in Fig. 5. The linear both methods with a confidence level of 95 %. The calculated
working range was from 2 to 200 μmol L−1 with linear corre- values from the statistical tests were smaller than the tabulated
lation coefficient greater than 0.99. The linear regression critical values, and thus the results can be considered similar
equation for PRA analysis in these conditions was for both methods.
I(A) = 0.0936 (±0.0324) + 0.1531(±0.0043) c (mol L−1). The
detection limit was estimated at 0.5 μmol L−1. Although the
LOD value obtained for PRA by proposed method is higher
Conclusions
than some values reported by others methods, the FIA-MPA
detection provides enough sensibility for PRA determination
This paper presented a fast, simple and inexpensive method
in pharmaceutical formulations, as well as in urine samples.
for the determination of PRA in pharmaceutical formulations
As advantages in relation to other methods, this work presents
using FIA-MPA detection with BBD electrode. Moreover, this
a simpler and faster method using an unmodified working
work demonstrated that an unmodified working electrode can
electrode, offering a good alternative for routine analysis of
be used for selective determination of PRA in the presence of
PRA in pharmaceutical samples and a potential application in
ascorbic acid and uric acid (electroactive interfering com-
biological fluids.
pounds), offering a good possibility to apply the proposed
As previously shown (Fig. 3), the AA and UA are not
method for the analysis of biological samples that contains
interferents for PRA detection at collector potential pulse
other species with electrochemical behavior similar to PRA.
34. Suffredini HB, Pedrosa VA, Codognoto L, Machado SAS, Rocha- 37. Peckova K, Musilova J, Barek J (2009) Boron-doped diamond film
Filho RC, Avaca LA (2004) Enhanced electrochemical response of electrodes—New tool for voltammetric determination of organic
boron-doped diamond electrodes brought on by a cathodic surface substances. Crit Rev Anal Chem 39(3):148–172. doi:10.1080/
pre-treatment. Electrochim Acta 49(22–23):4021–4026. doi:10. 10408340903011812
1016/j.electacta.2004.01.082 38. dos Santos WTP, Ceolin MP, Albuquerque YDT, Richter EM
35. Chailapakul O, Siangproh W, Tryk DA (2006) Boron-doped dia- (2007) The use of pressure generated by a water column to control
mond-based sensors: a review. Sens Lett 4(2):99–119. doi:10.1166/ flow rates in flow analysis systems. Quim Nova 30(7):1754–1758
sl.2006.008 39. Dey RD, Gupta S, Paira R, Chen SM, Raj CR (2012) Flow
36. Luong JHT, Male KB, Glennon JD (2009) Boron-doped diamond injection amperometric sensing of uric acid and ascorbic acid
electrode: synthesis, characterization, functionalization and analyt- using the self-assembly of heterocyclic thiol on Au elec-
ical applications. Analyst 134(10):1965–1979. doi:10.1039/ trode. J Solid State Electrochem 16(1):173–178. doi:10.
B910206J 1007/s10008-011-1311-1