You are on page 1of 7

J Solid State Electrochem

DOI 10.1007/s10008-015-3105-3

ORIGINAL PAPER

Determination of prazosin in pharmaceutical samples by flow


injection analysis with multiple-pulse amperometric detection
using boron-doped diamond electrode
Tiago de Jesus Guedes 1 & Morgana F. Alecrim 2 & Fernando M. Oliveira 1 &
Amanda B. Lima 1 & Sandro L. Barbosa 2 & Wallans T. P. dos Santos 2

Received: 18 October 2015 / Revised: 7 December 2015 / Accepted: 14 December 2015


# Springer-Verlag Berlin Heidelberg 2015

Abstract This work describes the development of a simple, Keywords Prazosin . Pharmaceutical analysis . Boron-doped
fast and low-cost method for determining prazosin (PRA) in diamond . Multiple-pulse amperometry . Flow injection
pharmaceutical samples by flow injection analysis with analysis
multiple-pulse amperometric (FIA-MPA) detection using a
boron-doped diamond film electrode. Electrochemical detec-
tion of PRA was optimized in phosphate buffer pH 4.0 by Introduction
cyclic voltammetry, in which PRA presented two oxidation
processes around at 0.97 and 1.40 V versus Ag/AgCl Prazosin (PRA) is the prototype of a family of agents that
(3.0 mol L−1 KCl). In these conditions, PRA also showed contain a core of piperazinyl quinazoline and acts as a potent
one reduction process at −0.75 V that is dependent on the and selective antagonist of α1 receptors, promoting decrease
oxidation processes. Thus, the determination of PRA by in peripheral vascular resistance and venous return to the
FIA-MPA detection consisted on the application of a two- heart, which consequently leads to its therapeutic effect in
potential waveform, E 1 (generator potential) = 1.6 V/ the treatment of hypertension [1]. Its structural formula is
400 ms and E2 (collector potential) = −1.0 V/30 ms, with shown in Fig. 1.
sample loop of 150 μL and flow rate of 3.0 mL min−1. The PRA is a drug with narrow therapeutic index [2], which
method showed good repeatability (RSD < 3.0 %) and high makes its determination very important to analyze this ac-
analytical frequency (70 injections per h). The working tive ingredient in formulations for the quality control in
linear range was obtained from 2 to 200 μmol L−1 with a pharmaceutical formulations and in biological fluids for
limit of detection of 0.5 μmol L−1. The recovery tests in all pharmacological studies. The official methods for determi-
samples were approximately 100 %, and the results were nation of PRA in pharmaceutical formulations are de-
compared with chromatographic methods. scribed in USP and British Pharmacopoeia [3, 4], which
are based on chromatography (HPLC with UV detection).
It can also be found in the literature various methods for
determination of PRA in formulations and biological fluids
* Wallans T. P. dos Santos
using HPLC with UV detection [5, 6] and other different
wallanst@yahoo.com.br detectors: fluorescence [7, 8], rayleigh light-scattering [9],
and amperometric using glassy carbon electrode [10]. In
1
addition, other methods for determination of PRA, such
Departamento de Química, Universidade Federal dos Vales do
Jequitinhonha e Mucuri, Campus JK, 39100-000 Diamantina, MG,
as spectrophotometry [11], thin layer chromatography
Brazil [12], capillary electrophoresis [13], and electroanalytical
2
Departamento de Farmácia, Universidade Federal dos Vales do
techniques [14–17].
Jequitinhonha e Mucuri, Campus JK, 39100-000 Diamantina, MG, Among all these methods, the electroanalytical techniques
Brazil have shown to be an attractive alternativeness for the
J Solid State Electrochem

presents a simple, fast, and reproducible analytical method


for PRA determination in pharmaceutical formulations and
of potential application in biological fluids by FIA-MPA using
BDD electrode.

Fig. 1 Structural formula of PRA


Experimental
determination of drugs in pharmaceutical formulations and in
biologic fluids. This is because the electrochemical detec- Reagents and solutions
tion often presents high selectivity and sensitivity for the
determination of electroactive compounds. Furthermore, The used reagents were of analytical grade, and all solutions
the electroanalytical methods provide low-cost and fast were prepared in deionized water with a resistivity of no less
analysis with less waste generation [18]. However, in the than 18.2 MΩ cm obtained from the ELGA purification sys-
literature, only four methods were published for PRA de- tem, Di-MKZ model. Initially, a stock solution of
termination based on electrochemical detection without 1.0 × 10−3 mol L−1 PRA (Sigma-Aldrich, São Paulo, Brazil)
chromatographic separation. Two electrochemical works was prepared in water. This stock solution of PRA was diluted
were carried out by differential pulse polarography [14, in the following supporting electrolytes for studies of electro-
15] and, in the others, a modified carbon paste electrode chemical detection: sulfuric acid from Vetec (Duque de
was used [16, 17]. Due to similar structure of the PRA and Caxias, Brazil); acetate buffer, acetic acid/sodium acetate from
doxazosin, Arranz and co-authors also evaluated the ability Merck (Rio de Janeiro, Brazil); phosphate buffer, sodium
to pre-concentrate PRA on the Tenax-modified carbon dihydrogen phosphate/sodium hydrogen phosphate from
paste electrode [19]. However, the results showed that PRA Vetec (Duque de Caxias, Brazil); borate buffer, boric acid/
did not accumulate with time in this working electrode be- borate sodium from Merck (Rio de Janeiro, Brazil); all
cause there is not a benzodioxan moiety in your molecule, supporting electrolytes were prepared at concentration of
which is involved in the adsorption between doxazosin and 0.1 mol L−1. The electrochemical response to analyte was
Tenax. These electroanalytical studies can be still improved in studied in a large range of pH using Britton-Robinson buffer
order to obtain a practical and efficient method for routine and the above electrolytes by cyclic voltammetry. A Britton-
analysis in pharmaceutical laboratories, eliminating the need Robinson buffer solution was composed of a mixture of
for modification, or use of electrodes with toxic potential (e.g., 0.1 mol L−1 acetic acid, boric acid, and phosphoric acid, and
mercury). Furthermore, these methods can also improving the its different pH values were adjusted with sodium hydroxide.
analytical parameters by increasing reproducibility, simplicity, The best conditions for determination of PRA by FIA-MPA
and analytical frequency. using BDD electrode were obtained in 0.1 mol L−1 phosphate
In this context, this work presents a new method for PRA buffer (pH 4.0).
determination by flow injection analysis with multiple-pulse The interference studies in urine samples were carried out
amperometric detection (FIA-MPA), which may yield a with presence of the ascorbic acid and uric acid, both stan-
higher analytical frequency and greater electrochemical re- dards from Sigma-Aldrich (São Paulo, Brazil). Standards of
sponse reproducibility when compared with stationary tech- the interfering species were diluted in buffer phosphate solu-
niques. It is worth noting that the MPA detection enables up to tion at concentration of 1 to 1000 times higher than the analyte
10 potential pulses as a function of time and the acquisition of concentration for evaluating interference of the PRA detection
up to 10 amperograms, independent, simultaneous which by FIA-MPA. The human urine samples were collected from
brings a number of advantages over other methods such as one healthy patient and were diluted in phosphate buffer (pH
use of internal standard in flow systems [20], and for increas- 4.0) solution without any treatment for FIA-MPA detection.
ing the selectivity of the electrochemical method for the de- The addition-recovery studies of PRA were carried out in
tection of the products of oxidation/reduction, even in the these urine samples.
presence of interfering species [21, 22] and also may enable The capsules of PRA, manufactured by Pfizer (Guarulhos,
simultaneous analysis of electroactive compounds [23–26]. Brazil) and marketed under trade name Minipress® SR, were
The use of boron-doped diamond (BDD) film electrode has obtained containing 1.0 mg of active ingredient. The capsules
been widely used for analysis of other drugs in pharmaceutical (20) of PRA were powdered in a mortar, and a weight corre-
formulations [27–30] and is an excellent alternative in im- sponding to one capsule was dissolved in electrolyte using an
proving the electrochemical response of PRA due to its high ultrasonic bath for 30 min for the MPA-FIA detection or in
stability and reproducibility. The BDD electrode also offers mobile phase for the HPLC-DAD detection. The sample and
several advantages including a broad potential window and standard solutions were filtered through a membrane filter
low background [31–37]. In these perspectives, this work (pore size of 0.45 mm) before injection in the HPLC system.
J Solid State Electrochem

The addition-recovery studies of PRA were also carried out in for sensitive and selective detection of PRA. After cathodic
these pharmaceutical samples. and anodic electrochemical treatments applied to BDD elec-
trode, the electrochemical behavior of PRA was re-evaluated
Instrumentation and apparatus in the supporting electrolyte. The cathodic treatment was
chosen due to its better definition of the reduction peak
The electrochemical measurements (amperometric and observed for PRA. Therefore, all studies were performed
voltammetric) were performed using a Potentiostat from after cathodic electrochemical pretreatment of the working
Autolab (Eco Chemie, the Netherlands) model PGSTAT electrode surface. In these conditions, the cyclic voltam-
128 N with GPES 4.9 software. An electrochemical flow cell mograms are presented in Fig. 2.
(Bwall jet^ type) with three-electrode was lab-made. A wire of As can be seen in Fig. 2, there are two oxidation peaks
platinum and an Ag/AgCl with KCl saturated were used as (0.97 and 1.40 V) and one reduction peak (−0.75 V) for
auxiliary and reference electrodes, respectively. The working PRA on BDD electrode. The oxidation and reduction process-
electrode was a thin film (ca.1.2 mm) of BDD (approximately es were described by Arranz and co-authors, which used a
8000 ppm doping level) supported on a polycrystalline silicon carbon paste electrode modified with nafion [16]. These au-
wafer (NeoCoat SA, La Chaux-de-Fonds, Switzerland). One thors observed at first scan just one oxidation process, which
piece of BDD material (1.0 × 1.0 cm) was used in the flow posteriorly generated a quasi-reversible cathodic process.
cell, and the electrode area was delimited by using a rubber O- Although it could not be confirmed by cyclic voltammetry at
ring with 0.4 cm of diameter (electrode area = 0.13 cm2). BDD electrode that the reduction is a quasi-reversible electro-
Background current correction was carried out using the chemical process, FIA-MPA confirmed that this reduction
GPS software from Autolab. The electrochemical pretreat- process on BDD electrode presents similar electrochemical
ment of the BDD electrode for PRA detection was evaluated behavior to carbon paste electrode modified.
by cyclic voltammetry. The BDD electrode was anodically The electrochemical mechanism elucidated in the literature
pretreated in a 0.5mol L−1 H2SO4 solution by applying for PRA on carbon paste electrode proposes that the first ox-
0.001 A during 120 s. The cathodic pretreatment was carried idation current is controlled by adsorption of the analyte at the
out by applying −0.03 A during 360 s using the same solution. electrode surface [16]. According to these authors, this anodic
The BDD electrode was pretreated only once, prior to the process corresponds to the oxidation of the adsorbed species
measurements. of PRA, which can produce dimer compounds through the
The FIA system (single-line) was composed of polyethyl- oxidation of the amine group present in this molecule along
ene tubing with 0.5 mm internal diameter, an acrylic manual with a four-electron and four-proton transfer process. Thus,
injector and flow rate was controlled by the pressure generated the PRA oxidized form might be reduced by an azo group,
by a water column [38]. The flow rate and injection volume in an almost reversible process with two-electron and two-
were investigated in the range of 0.5 to 4.0 mL min−1 and 50 proton transfer [16]. We suggest that the electrochemical
to 350 μL, respectively. oxidation and reduction mechanism of the PRA on BDD
The comparison of the method proposed for PRA determi- electrode is similar to verified on carbon paste electrode.
nation in pharmaceutical samples was performed by the offi- However, the first oxidation process should occur in two
cial method using HPLC-UV [4]. The HPLC system used in steps on BDD electrode. Moreover, the plot of the current
this work was a Shimadzu (Kyoto, Japan). The column used for the oxidation peaks of PRA (0.97 and 1.40 V) as a
was C18 (10, 4.6, 5 mm) from Macherey-Nagel (Germany). function of the root of the scan rate (10 to 500 mV s−1)
The mobile phase was composed of methanol/water (70:30)
with 1 % acetic acid (v/v). All HPLC analysis were carried out
at room temperature (30 °C) with an injection volume of
20 μL and a flow rate of 1.0 mL min−1 (isocratic), while
detection was performed at 254 nm.

Results and discussion

Electrochemical behavior of prazosin

The electrochemical behavior of PRA was studied by cyclic


Fig. 2 Cyclic voltammograms on BDD electrode in 0.1 mol L−1
voltammetry at BDD electrode in various electrolytes. The phosphate buffer pH 4.0 without (black line) and with addition of PRA
0.1 mol L −1 phosphate buffer (pH 4.0) was chosen as 1.0 × 10−4 mol L −1 (red line). The scan was started at −1.0 V in the
supporting electrolyte because it presented the best results positive direction. Scan rate 50 mV s −1
J Solid State Electrochem

was linear, thus supporting that the oxidation processes are


controlled by diffusion.
As one of the purposes of this work is to determine the
PRA in urine, ascorbic acid (AA) and uric acid (UA) are
common interferents in this sample type due to overlapping
of their oxidation processes with of the analyte [39]. So, PRA
determination is not possible by application of anodic poten-
tial pulses in the MPA detection. However, as the AA and UA
do not exhibit reduction processes on the BDD electrode, the
PRA determination is possible by application of cathodic po-
tential pulse in the MPA detection without interference of Fig. 4 Repeatability data obtained from successive injections of a
these compounds. Thus, the PRA firstly must be oxidized solution containing 10 μmol L−1 (a) or 100 μmol L−1 (b) of PRA. Flow
rate 3.0 mL min−1; injection volume 150 μL. Supporting electrolyte
in the generator potential pulse and quantified by the ap- 0.1 mol L −1 phosphate buffer pH 4.0. Potential pulses on BDD
plication of a second potential pulse (collector) through the electrode: 1.6 V/400 ms (not shown) and −1.0 V/30 ms
reduction of the product generated at BDD electrode. Thus,
all optimization studies by FIA-MPA detection of this It is worth mentioning that was necessary the application of
work are given in function of the collector potential pulse. more potential pulses between the potentials selected to de-
crease the capacitive current in the MPA detection due to the
high potential jump (2.6 V) over BDD electrode. To solve this
Optimization parameters of FIA-MPA detection problem, the potential pulses of 0.8 V/100 ms and −0.2 V/
100 ms were selected, but the current signals obtained at these
Optimization parameters were performed for PRA determina- potential pulses were not acquired, and the respective
tion in urine and pharmaceutical formulations by FIA-MPA. amperograms are not shown in Fig. 3, as well as in other
The potential pulses as well as their application times were figures in this work. The flow rate and sample loop volume
optimized in function of the selectivity and sensitivity for in the FIA system are also important to better efficiency of the
analyte quantification. These pulses were continuously ap- method [21]. So, these parameters were optimized based on
plied in function of time (cyclical form) and the following better analytical frequency, analytical response, and repeat-
potential pulses were selected (amperograms shown in Fig. 3): ability for PRA determination. The best conditions were ob-
I) +1.6 V for 400 ms: PRA and common interferents as AA tained using 3.0 mL min−1 and 150 μL as flow rate and the
e UA are oxidized, but only the electrochemical generated injection volume, respectively. In this condition, the analytical
product from PRA is electroactive in the cathodic region on frequency was estimated in 70 injections per h, which pro-
BDD electrode. vides faster analysis than the other applications reported for
II) −1.0 V for 30 ms: the electrochemical generated product the determination of PRA.
from PRA in the oxidation step is reduced and quantified,
without the interference of AA and UA. Analytical parameters

The analytical parameters of the FIA-MPA method were eval-


uated (after optimization of the conditions) for PRA

Fig. 3 Amperograms obtained by FIA-MPA detection in supporting


electrolyte with duplicate injections of solutions containing
10 μmol L−1 of only PRA, only UA, only AA and all together (PRA + Fig. 5 Amperogram obtained by FIA-MPA detection after triplicate in-
UA + AA) in the same concentration. Flow rate 3.0 mL min−1; injection jections of standard solutions containing PRA at concentrations from 2 to
volume 150 μL. Potential pulses on BDD electrode 1.6 V/400 ms and 200 μmol L−1 (a–i) and the respective calibration curve is showed in the
−1.0 V/30 ms inset. Other conditions, see Fig. 4
J Solid State Electrochem

Table 1 FIA-MPA responses


obtained after triplicate injections Concentration ratio of the Current signal of PRA (%) Current signal of PRA (%)
of solutions containing interfering substance for relation AA/PRA for relation UA/PRA
2.0 μmol L−1 PRA and increasing
concentrations of AA and UA 1 100.0 100.0
50 99.5 103.2
100 104.1 102.9

determination in pharmaceutical formulations and in human (−1.0 V). However, due to high concentrations of AA and
urine. The stability study of the BDD electrode coupled to the UA in urine samples, the concentration effect of these
FIA system was evaluated by detection of PRA at collector interferents on current signal of analyte also was evaluated
potential pulse (−1.0 V). The repeatability study (Fig. 4) was (Table 1). As can be observed in Table 1, the current signal
performed by 10 successive injections of solutions containing of PRA was relatively constant in the presence of up to 100
10 μmol L−1 PRA, then more 10 injections (with a concentra- times more of AA or UA. For higher concentrations of AA, a
tion 10 times higher than before) and, finally, more 10 injec- decrease in the PRA signal was observed. The UA was not
tions in the same initial concentration of PRA. soluble in the supporting electrolyte at higher concentrations,
As can be noted in Fig. 4, even after the injections of high and so, it did not interfere in the analysis of PRA.
PRA concentrations, the first 10 peaks did not differ signifi- The addition-recovery studies in pharmaceutical samples
cantly from the last 10 measurements. The RSD for PRA were carried out to verify the performance of the proposed
detection were 2.46 % for the first 10 peaks (Fig. 4a), method. The results obtained for recovery of PRA in formu-
2.85 % for 10 injections at high concentrations (Fig. 4b) and lations pharmaceutical (n = 3) and human urine (n = 3) were
2.79 % for the last 10 peaks (Fig. 4a). These results demon- 100.1 % (±1.0) and 98.0 % (±5.0), respectively. The values
strated that the FIA system using BDD electrode can present obtained were close to 100 %, indicating the absence of matrix
high reproducibility for PRA quantification, mainly because effect in these samples.
the phenomenon of working electrode contamination (or pas- The results obtained for the determination of PRA in phar-
sivation) is practically absent due to inert surface of the BDD. maceutical samples using the proposed method versus the
The amperometric responses for triplicate injections of so- official method (HPLC-UV) are presented in Table 2.
lutions containing increasing concentrations of PRA and the Statistical tests (F and t) were applied to results obtained by
respective calibration curve are presented in Fig. 5. The linear both methods with a confidence level of 95 %. The calculated
working range was from 2 to 200 μmol L−1 with linear corre- values from the statistical tests were smaller than the tabulated
lation coefficient greater than 0.99. The linear regression critical values, and thus the results can be considered similar
equation for PRA analysis in these conditions was for both methods.
I(A) = 0.0936 (±0.0324) + 0.1531(±0.0043) c (mol L−1). The
detection limit was estimated at 0.5 μmol L−1. Although the
LOD value obtained for PRA by proposed method is higher
Conclusions
than some values reported by others methods, the FIA-MPA
detection provides enough sensibility for PRA determination
This paper presented a fast, simple and inexpensive method
in pharmaceutical formulations, as well as in urine samples.
for the determination of PRA in pharmaceutical formulations
As advantages in relation to other methods, this work presents
using FIA-MPA detection with BBD electrode. Moreover, this
a simpler and faster method using an unmodified working
work demonstrated that an unmodified working electrode can
electrode, offering a good alternative for routine analysis of
be used for selective determination of PRA in the presence of
PRA in pharmaceutical samples and a potential application in
ascorbic acid and uric acid (electroactive interfering com-
biological fluids.
pounds), offering a good possibility to apply the proposed
As previously shown (Fig. 3), the AA and UA are not
method for the analysis of biological samples that contains
interferents for PRA detection at collector potential pulse
other species with electrochemical behavior similar to PRA.

Table 2 Quantity-control of the PRA in pharmaceutical formulation


obtained by the proposed method (FIA-MPA) and by official method
(HPLC-UV) and their respective values of standard deviation (n = 3) Acknowledgments The authors are grateful to Conselho Nacional de
Desenvolvimento Científico e Tecnológico (CNPq – Project: 481683/
Sample Ingredient Labeled (mg) FIA-MPA (mg) HPLC-UV (mg) 2013-5) and Fundação de Amparo à Pesquisa do Estado de Minas Gerais
(FAPEMIG) for financial support. This work is a collaboration research
Capsules PRA 1.00 0.92 ± 0.06 0.96 ± 0.02 project of members of the Rede Mineira de Química (RQ-MG) supported
by FAPEMIG (Project: CEX - RED-00010-14).
J Solid State Electrochem

References urine and pharmaceutical tablets using. Carbon paste electrodes.


Analyst 122(8):849–854
20. Gimenes DT, Santos WTP, Muñoz RAA, Richter EM (2010)
1. Hardman JG, Limbird LE (2006) Goodman & Gilman’s the phar- Internal standard in flow injection analysis with amperometric de-
macological basis therapeutics. McGraw Hill, New York tection. Electrochem Commun 12(2):216–218
2. L.X. Yu and B.V. Li (eds.), (2014) FDA Bioequivalence Standards, 21. dos Santos WTP, Gimenes DT, Richter EM, Angnes L (2011) Flow
AAPS Advances in the Pharmaceutical Sciences. Series 13, doi:10. injection analysis with multiple pulse amperometric detection: po-
1007/978-1-4939-1252-0. Springer, New York tentialities and applications. Quim Nova 34(10):1753–1761
3. British Pharmacopoeia (1988) Vols. I and II, by Great Britain, 22. Gimenes DF, dos Santos WTP, Tormin TF, Muñoz RAA, Ricther
British Pharmacopoeia Commission Publisher: London HMSO EM (2010) Flow-injection amperometric method for indirect deter-
1988, Description: xli, 1575p mination of dopamine in the presence of a large excess of ascorbic
4. Healthwise Knowledgebase (1998) US Pharmacopeia, Rockville. acid. Electroanalysis 22(1):74–78
http://www.healthwise.org . Accessed 21 Sept 2015 23. Pereira PF, Marra MC, Lima AB, dos Santos WTP, Muñoz RAA,
5. Batchman WJ (1986) High Performance liquid chromatographic Richter EM (2013) Fast and simultaneous determination of
determination of diuretic-antihypertensive combination products. nimesulide and paracetamol by batch injection analysis with am-
I. Prazosin and polythiazide. J Liq Chromatogr 9(5):1033–1049 perometric detection on bare boron-doped diamond electrode.
6. Sultana N, Saeed Arayne M, Shah SN (2014) Development and Diam Relat Mater 39:41–46. doi:10.1016/j.diamond.2013.07.010
validation for the simultaneous quantification of prazosin, 24. Medeiros RA, Lourencao BC, Rocha-Filho RC, Fatibello-Filho O
amlodipine, diltiazem and verapamil in API, dosage formulation (2010) Simple flow injection analysis system for simultaneous de-
and human serum by RP-HPLC: application to in vitro interaction termination of phenolic antioxidants with multiple pulse ampero-
studies. Med Chem 4(12):770–777 metric detection at a boron-doped diamond electrode. Anal Chem
7. Yee YG, Rubin PC, Meffin P (1979) Prazosin determination by 82(20):8658–8663. doi:10.1021/ac101921f
high-pressure liquid chromatography using flourescence detection. 25. Chaves SC, Aguiar PNC, Torres LMFC, Gil ES, Luz RCS, Damos
J Chromatogr A 172(1):313–318, 21 April 1979 FS, Munoz RAA, Richter EM, dos Santos WTP (2015)
8. Fletcher AJ, Addison RS, Mortimer RH, Cannell GR (1995) Rapid Simultaneous determination of caffeine, ibuprofen, and paraceta-
determination of prazosin in perfusion media by HPLC with solid mol by flow-injection analysis with multiple-pulse amperometric
phase extraction. J Liq Chromatogr 18(14):2911–2923. doi:10. detection on boron-doped diamond electrode. Electroanalysis
1080/10826079508009334 27(12):2785–2791. doi:10.1002/elan.201500306
9. Li PA, Peng H, Peng DJ, Zhou MQ, Zhang J (2015) Rayleigh light 26. dos Santos WTP, Almeida EGN, Ferreira HEA, Gimenes DF,
scattering detection of three α1-adrenoceptor antagonists coupled Richter EM (2008) Simultaneous flow injection analysis of para-
with high performance liquid chromatograph. Spectrochim Acta A cetamol and ascorbic acid with multiple pulse amperometric detec-
Mol Biomol Spectrosc 147:178–184, 5 August 2015 tion. Electroanalysis 20(17):1878–1883. doi:10.1002/elan.
10. Rathinavelu A, Malave A (1995) High-performance liquid chroma- 200804262
tography using electrochemical detection for the determination of 27. Faria EO, Lopes-Junior ACV, Souto DEP, Leite FRF, Damos FS,
prazosin in biological samples. J Chromatogr B Biomed Sci Appl Luz RCS, Santos AS, Franco DL, dos Santos WTP (2012)
670(1):177–182, 4 August 1995 Simultaneous determination of caffeine and acetylsalicylic acid in
11. Sreedhar K, Sastry CSP, Reddy MN, Sankar DG (1996) pharmaceutical formulations using a boron-doped diamond film
Spectrophotometric methods for the determination of prazosin hy- electrode by differential pulse voltammetry. Electroanalysis 24(5):
drochloride in tablets. Talanta 43(11):1847–1855 1141–1146. doi:10.1002/elan.201500306
12. Daldrup T, Susanto F, Michalke P, Fresenius Z (1981) Kombination 28. Sartori ER, Medeiros RA, Rocha-Filho RC, Fatibello-Filho O
yon DC, GC (OV 1 und OV 17) und HPLC (RP 18) zur schnellen (2010) Square-wave voltammetric determination of propranolol
Erkennung von Arzneimitteln, Rauschmitteln und verwandten and atenolol in pharmaceuticals using a borondoped diamond elec-
Verbindungen. Fresenius Z Anal Chem 308:413–427 trode. Talanta 81(4–5):1418–1424. doi:10.1016/j.talanta.2010.02.
13. Soini H, Riekkola ML, Novatny MV (1994) Mixed polymer net- 046
works in the direct analysis of pharmaceuticals in urine by capillary 29. Santos MCG, Tarley CRT, Dall’Antonia LH, Sartori ER (2013)
electrophoresis. J Chromatogr A 680(2):62–634, 7 October 1994 Evaluation of boron-doped diamond electrode for simultaneous
14. Altiokka G, Tuncel M (1997) Differential pulse polarographic de- voltammetric determination of hydrochlorothiazide and losartan in
termination of prazosin hydrochloride in tablets. Die Pharmazie pharmaceutical formulations. Sensors Actuators B Chem 188:263–
[0031–7144] Altiokka, G 52(5):401–402 270. doi:10.1016/j.snb.2013.07.025
15. Özgür MÜ, Aycan S, Islimyeli S (1995) Differential pulse polaro- 30. Lima AB, Torres LMFC, Guimaraes CFRC, Verly RM, da Silva
graphic determination of prazosin hydrochloride in tablets. Die LM, Carvalho-Junior AD, dos Santos WTP (2014) Simultaneous
Pharm 50(6):435–436 determination of paracetamol and ibuprofen in pharmaceutical sam-
16. Arranz A, de Betoño SF, Echevarria C, Moreda JM, Cid A, Valentín ples by differential pulse voltammetry using a boron-doped dia-
JFA (1999) Voltammetric and spectrophotometric techniques for mond electrode. J Braz Chem Soc 25(3):478–483. doi:10.5935/
the determination of the antihypertensive drug Prazosin in urine 0103-5053.20140005
and formulations. J Pharm Biomed Anal 21:797–807 31. Macpherson JV (2015) A practical guide to using boron doped
17. Arranz A, Moreda JM, Arranz JF (2000) Anodic stripping diamond in electrochemical research. Phys Chem Chem Phys
voltammetric assay for the determination of prazosin at carbon 17(5):2935–2949. doi:10.1039/C4CP04022H
paste electrodes. Química Analítica [0212–0569] Arranz, A 32. Einaga Y, Foord JS, Swain GM (2014) Diamond electrodes: diver-
19(1):31–37 sity and maturity. Mater Res Bull 39(6):525–532. doi:10.1557/mrs.
18. Uslu B, Ozkan AS (2011) Electroanalytical methods for the deter- 2014.94
mination of pharmaceuticals: a review of recent trends and devel- 33. Pleskov YV (2000) Synthetic diamond, a New electrode material
opments. Anal Lett 44(16):2644–2702 for electroanalysis. J Anal Chem 55(11):1045–1050, Translated
19. Arranz A, de Betoño SF, Moreda JM, Cid A, Arranz JF (1997) from Zhurnal Analificheskoi Khimii, Vol. 55, No. I1, 2000, pp.
Cathodic stripping voltammetric determination of doxazosin in 1165–1171
J Solid State Electrochem

34. Suffredini HB, Pedrosa VA, Codognoto L, Machado SAS, Rocha- 37. Peckova K, Musilova J, Barek J (2009) Boron-doped diamond film
Filho RC, Avaca LA (2004) Enhanced electrochemical response of electrodes—New tool for voltammetric determination of organic
boron-doped diamond electrodes brought on by a cathodic surface substances. Crit Rev Anal Chem 39(3):148–172. doi:10.1080/
pre-treatment. Electrochim Acta 49(22–23):4021–4026. doi:10. 10408340903011812
1016/j.electacta.2004.01.082 38. dos Santos WTP, Ceolin MP, Albuquerque YDT, Richter EM
35. Chailapakul O, Siangproh W, Tryk DA (2006) Boron-doped dia- (2007) The use of pressure generated by a water column to control
mond-based sensors: a review. Sens Lett 4(2):99–119. doi:10.1166/ flow rates in flow analysis systems. Quim Nova 30(7):1754–1758
sl.2006.008 39. Dey RD, Gupta S, Paira R, Chen SM, Raj CR (2012) Flow
36. Luong JHT, Male KB, Glennon JD (2009) Boron-doped diamond injection amperometric sensing of uric acid and ascorbic acid
electrode: synthesis, characterization, functionalization and analyt- using the self-assembly of heterocyclic thiol on Au elec-
ical applications. Analyst 134(10):1965–1979. doi:10.1039/ trode. J Solid State Electrochem 16(1):173–178. doi:10.
B910206J 1007/s10008-011-1311-1

You might also like