SONOLINE G20

US

Software
System
Standard Software Upgrade

08647831
08647815

© Siemens AG 2004
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
Siemens Medical
Ultrasound
ejt DivisionSolutions
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: US06-101.816.01.02.02 English


Replaces: US06-101.816.01.01.02 Doc. Gen. Date: 01.05
n.a.

Part No.:
sd
2 Revision / Disclaimer
1Revision / Disclaimer

Copyright
Copyright © 2004 by Siemens Corporation. All rights reserved. No part of this publication
may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into
any language or computer language, in any form or by any means, electronic, mechanical,
magnetic, optical, chemical, manual, or otherwise, without the prior written permission of
Siemens Corporation. Siemens Corporation reserves the right to change its products and
services at any time. In addition, this manual is subject to change without notice. Siemens
Corporation welcomes customer input on corrections and suggestions for improvements
to this manual.Although Siemens Corporation has attempted to ensure accuracy through-
out this manual, Siemens Corporation assumes no liability for any errors or omissions, nor
for any damages resulting from the application or use of this information.
Trademarks
ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE,
Adara, Antares, G20, G40, G50, G60S, Omnia and KinetDx are trademarks of Siemens
Corporation registered in the U.S. Patent and Trademark Office. CWS3000, DS3000,
WS3000, and WebPro are trademarks of Siemens Corporation.
Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet
Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are
registered trademarks of Netscape Communications Corporation. Java and the JavaScript
name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered
trademark of Adobe Systems, Inc.
Other products and brand names are trademarks of their respective owners.
Disclaimer
The service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie-
mens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
License Agreement
All computer programs copyright 1990-2004 by Siemens Corporation or its suppliers. Pro-
grams are licensed under the following agreement:
Siemens or its suppliers retain(s) ownership of and title to any computer program supplied
with the equipment and to the trade secrets embodied in such computer programs. Sub-
ject to the Buyer’s acceptance and fulfillment of the obligations in this paragraph, Siemens
grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any
computer program supplied with the Equipment that is necessary to operate the Equip-
ment solely on the medium in which such program is delivered for the purpose of operat-
ing the equipment in accordance with the instructions set forth in the operator’s manuals
supplied with the Equipment and for no other purpose whatsoever. Buyer may not
reverse-assemble, reverse-compile or otherwise reverse-engineer such computer pro-
grams nor may Buyer make a copy of such program or apply any techniques to derive the
trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms

SONOLINE G20 US06-101.816.01.02.02 Page 2 of 8 Siemens


01.05 sd Medical Solutions USA, Inc.
Revision / Disclaimer 3

of this license, the license granted by this paragraph shall terminate. Further, because
unauthorized use of such computer programs will leave Siemens without an adequate
remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to
restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s
suppliers of software is a direct and intended beneficiary of this end-user sublicense and
may enforce it directly against Buyer with respect to software supplied by such supplier,
and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indi-
rect, consequential, incidental or other damages arising out of the sublicense of the com-
puter programs supplied with the equipment.

Siemens US06-101.816.01.02.02 Page 3 of 8 SONOLINE G20


Medical Solutions USA, Inc. 01.05 sd
4 Table of Contents
1- 0Table of Contents

1 _______ Standard Software Upgrade _______________________________________ 5

Loading System Software Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5


Updating and Reloading System Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
OS-System Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Software-Based Option Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

SONOLINE G20 US06-101.816.01.02.02 Page 4 of 8 Siemens


01.05 sd Medical Solutions USA, Inc.
Standard Software Upgrade 5
2-
Loading System Software Software
1Standard Software Upgrade

The system software resides in the system memory, so loading software using disks and
the system"s compact disk (CD) drive is not normally required. It may be necessary to
load software if it is received as part of a system upgrade.

NOTE While loading the new software version, selecting Retain Current
Preset will retain all enabled features and the user-defined settings
for system presets, QuickSets, and OB tables and formulas. Se-
lecting Initialize Preset resets all settings to factory default set-
tings and deletes all enabled features and the user-defined
settings for the system presets, QuickSets, and OB tables and for-
mulas. The selection of Initialize Presets may require the confirma-
tion of language selections that are installed with the system
software.

Updating and Reloading System Software 0

1. With the system powered on, insert the System Disk into the CD drive located below the
control panel on the front of the ultrasound system.
2. Press the F6 key to access the system presets.
The system displays the Preset Main Menu screen.
3. Select the down arrow at the bottom of the left pane of the Preset Main Menu to access
the Serviceability menu.
4. Select System/Language from the Serviceability menu.
The system displays language and system selections on the right side of the screen.
5. Select Retain Current Preset.

NOTE Selecting Initialize Preset deletes all the user-defined settings for
the system presets, QuickSets, and OB tables and formulas; this
selection may set the language to English. When selecting Initial-
ize Preset, Siemens recommends to first make backups of system
presets, QuickSets, and OB tables and formulas; enabled features
must be reloaded using the original source media.

6. Select the Upgrade System button.


If Retain Current Preset was selected, wait approximately two minutes for the system
to display the following message: Install Finished.
If Initialize Preset was selected, the system displays a caution message.

Siemens US06-101.816.01.02.02 Page 5 of 8 SONOLINE G20


Medical Solutions USA, Inc. 01.05 sd
6 Standard Software Upgrade

- Selecting Yes displays a message that instructs to restart the system. All the us-
er-defined settings for the system presets, QuickSets, and OB tables and formulas
are then deleted and the language is usually set to English.
- Select No displays the System/Language screen again; manually back up the sys-
tem presets, QuickSets, and OB tables and formulas. Perform the backup procedure
and then start again at step 1 of this software update/reload procedure.
7. Select the Shut Down button.
The system status changes to standby.
8. Wait 20 seconds and then press the Standby button to power on the system.
The system displays a progress bar at the bottom of the screen as system software
loads. When the process is completed, the system displays the message “Restart the
system”.
9. Press the CD drive Eject button and remove the System Disk.

NOTE Selecting Initialize Preset during this procedure may change the
system language. Use the system presets to select another lan-
guage. To change the language, press the F6 key, select the down
arrow at the bottom of the left pane of the Preset Main Menu, select
System/Language from the Serviceability Menu, select another
language, and then select the Change Language button. Wait for
the system to power off and then press the Standby button to pow-
er on the system.

10. Before using the system for the first time after a software update/reload, press the
Standby button to power off the system again, wait 20 seconds, and then press the
Standby button to power on the system.

OS-System Disk 0

CAUTION Use extreme caution when loading OS-System Disk software.


¹ The OS-System Disk overwrites the system"s hard drive.

The OS-System disk is an auto bootable CD. If inserted in the CD drive, the disk will over-
write the local hard drive while cycling power to the ultrasound system.

1. With the system powered on, insert the OS-System Disk into the CD drive.
2. Press the Standby button to power off the system.
3. Wait 20 seconds and then press the Standby button to power on the system.
The system displays an “R” prompt on the image screen in approximately five minutes.
4. Remove the OS-System Disk from the CD drive.

SONOLINE G20 US06-101.816.01.02.02 Page 6 of 8 Siemens


01.05 sd Medical Solutions USA, Inc.
Standard Software Upgrade 7

5. Press the Standby button to power off the system, wait approximately 20 seconds, and
then press the Standby button to power on the system.
The system boot sequence takes approximately three minutes. If an “Extended Post”
message displays, then select OK to dismiss the dialog box.
6. Press the Standby button again to power off the system, wait approximately 20 sec-
onds, and then press the Standby button to power on the system.

Software-Based Option Upgrade 0

Perform an upgrade or unload a software-based option (feature) on the ultrasound system


using a key disk (USB memory stick). One key disk is required to upgrade one option per
system. A key disk cannot be in used to concurrently upgrade an option on multiple sys-
tems. If subsequently unloaded, the option key disk can be loaded on a different SONO-
LINE G20 ultrasound system.
Observe the following precautions when using a key disk:

1. Insert only one key disk at a time.


2. Do not remove the key disk during the option upgrade or unload.

Upgrading a Software-based option

1. Insert the media containing the option into one of the USB ports on the I/O panel.
2. Press the F6 key.
The system displays the Preset Main Menu.
3. Select Installation from Key Disk from the Options menu.
4. Select the Install button.
The system displays a confirmation dialog box.
5. Select the OK button.
The system displays a message box as the key disk flashes while the option is loaded.
When the process is complete, the system displays a new message box.
6. Select the OK button.
The option is listed on the Installed Option screen.
7. Remove the key disk.

Uninstalling a Software-based Option

1. Insert the source media for the option into one of the USB ports on the I/O panel.
2. Press the F6 key.
The system displays the Preset Main Menu.

Siemens US06-101.816.01.02.02 Page 7 of 8 SONOLINE G20


Medical Solutions USA, Inc. 01.05 sd
8 Standard Software Upgrade

3. Select Installation from Key Disk.


4. Select the Uninstall button.
5. Select the OK button
The system displays a message box as the key disk flashes while the option is
uninstalled. When the process in complete, the system displays a new message box.
6. Select the OK button.
The option is no longer listed on the Installed Option screen.
7. Remove the key disk.

SONOLINE G20 US06-101.816.01.02.02 Page 8 of 8 Siemens