You are on page 1of 7

ARTICLE IN PRESS

Medical Dosimetry ■■ (2018) ■■–■■

Medical Dosimetry
j o u r n a l h o m e p a g e : w w w. m e d d o s . o r g

Dosimetry Contribution:

Offline adaptive radiation therapy in the treatment of


prostate cancer: a case study
Evgenia Nigay, M.S., R.T.(T), Heath Bonsall, M.S., C.M.D., R.T.(T),
Beverly Meyer, M.S., R.T.(R)(T), Ashley Hunzeker, M.S., C.M.D., and
Nishele Lenards, M.S., C.M.D., R.T.(R)(T), F.A.A.M.D.
Medical Dosimetry Program at the University of Wisconsin, La Crosse, WI

A R T I C L E I N F O A B S T R A C T

Article history: The purpose of this case study is to develop a method to account for the difference in the
Received 21 October 2017 daily volumes in the bladder, rectum, and targets in prostate radiotherapy and to compare
Received in revised form 18 the predicted dose to the actual dose to these organs. Five patients, both prospectively and
December 2017
retrospectively, were selected from 2 different cancer centers, with a biopsy-confirmed di-
Accepted 18 December 2017
agnosis of prostate cancer. The patients’ planning target volume (PTV) and organs at risk
(OAR) were contoured on the computed tomography (CT) dataset using either Eclipse or
Keywords:
Monaco treatment planning systems (TPSs). Cone-beam computed tomography (CBCT) scans
Adaptive radiation therapy
were collected before each daily treatment and exported to MIM software for analysis. The
prostate cancer
automatically generated reports evaluated the organ volume changes, the actual dose re-
MIM software
ceived during a single fraction, and the projected dose to each organ at the completion of
the treatment course via comparative cumulative dose-volume histograms (DVHs). Volume
changes in the bladder and rectum can cause notable variations in the prescribed dose vs
the actual dose received. MIM software was proven to have utility prospectively by tabu-
lating daily dose and projecting final doses, potentially aiding physicians in decisions about
the boost plans, thus making offline adaptive radiation therapy (ART) clinically manageable.
© 2018 American Association of Medical Dosimetrists.

Introduction response can reach 70 Gy. Due to such high doses, delivery
of a sufficient dose to the prostate or prostate bed while lim-
Among men in the United States, prostate cancer is the iting dose to adjacent radiosensitive structures, such as the
second most common cancer and one of the most common bladder and rectum, proves to be difficult.2 Organs at risk
causes of death. 1 Radiation therapy, both external and (OAR) are highly mobile, morphing shapes that can make
brachytherapy, for prostate cancer is a highly utilized form dose delivery highly speculative, creating questionable dose
of treatment in cancer centers throughout the United States. to the OAR. Even as sophisticated as the plans have become
With external radiotherapy, doses required for tumor for radiation therapy, they cannot account for daily changes
in patient anatomy.3
Adaptive radiation therapy (ART) has been around for
decades.3 However, technical complications can make ART
Reprint requests to Evgenia Nigay, M.S., R.T.(T), Medical Dosimetry
difficult to come to fruition, namely, the timely efforts in re-
Program at the University of Wisconsin, La Crosse, WI. contouring and re-planning each patient who is receiving
E-mail: evgenia.nigay@gmail.com radiation therapy treatment. 3 Cone beam computed

https://doi.org/10.1016/j.meddos.2017.12.005
0958-3947/Copyright © 2018 American Association of Medical Dosimetrists
ARTICLE IN PRESS
2 H. Bonsall et al. / Medical Dosimetry ■■ (2018) ■■–■■

tomography (CBCT) is considered to be the standard check a set of permanent marks in the pelvic region, anteriorly and
utilized for online ART before daily radiation treatments. As on either side, to provide a reference point for treatment
studies have shown, the use of CBCT can decrease the acute planning. Radiopaque markers were placed on these marks
and chronic side effects to the genitourinary and gastroin- to visualize them during the scan. Patients were scanned
testinal systems.4 head first, using 2-mm slices on either Siemens SOMATOM
ART can play an important role in decreasing unwanted Emotion 16 or Siemens SOMATOM Definition AS (Simens,
dose to critical structures and normal tissues. Prostate bed Erlangen, Germany) CT scanner.
and intact prostate radiation treatments are heavily reliant As per the clinic’s protocol, the patients were instructed
upon proper and consistent bladder filling as well as rectal to empty their rectum and fill their bladder before the sim-
emptying. When these 2 criteria are not adequately repro- ulation and each day before treatment. If the rectum was
duced according to the computed tomography (CT) planning too full, patients were asked to have a bowel movement
scan, the dose delivered to the target can be considerably before repeating the scan. Patients were also instructed to
altered.2 Basing the plan on 1 CT scan, daily changes in patient have a low-residue diet for the duration of the treatment to
positioning by even a few millimeters can result in devia- help prevent or reduce diarrhea.
tion from the intended dose.4 The International Commission
on Radiological Units and Measurements recommends Target delineation
regular reports be completed on the received dose to the
planning OAR volume, which is now available through the Once the simulation scan was complete, the target de-
use of daily CBCT.5 lineation and treatment planning were performed using
Daily changes in positions of OAR and their respective either the Elekta Monaco 5.11.01 TPS (Elekta, Stockholm,
volumes can alter dose considerably, warranting investiga- Sweden) or Varian Eclipse 13.7 TPS (Varian, Palo Alto, CA).
tion. Retrospectively analyzing the dose to the target can The clinical target volume (CTV) was contoured by the phy-
reveal a high fluctuation in minimum dose to the prostate.6 sician on the CT dataset. The planning target volume (PTV)
To track the daily doses to the target and OAR, MIM soft- was created by expanding the CTV by 3 mm posteriorly and
ware (Cleveland, OH) has been utilized.7 MIM software is a 5 mm in all other directions for the intact prostate or 8 mm
suite of tools that allows for efficient contouring, rigid and in all directions for the prostate fossa. Once the CT dataset
deformable registration, and plan evaluation. It has been uti- was received by the medical dosimetrist, OAR were con-
lized in an accurate and automated ART workflow in toured. The OAR of interest for this study included the rectum
conjunction with methods that use deformable CBCT con- and bladder volumes.
tours. In this case study, the adaptive radiotherapy assistant Before each treatment, the patients were imaged using
automated the process of fusing the daily CBCT to the plan- CBCT. The scans were matched to the initial planning CT scan
ning CT, deforming the dose and the contours, and generating to ensure the precise target location as well as consistency
reports of the findings. The goal of this research study was in bladder filling and rectal emptying during each treat-
to develop a method to account for the actual difference in ment. Each site aligned the CBCTs based on the priorities
the volumes from day to day and to compare the predicted provided by the respective attending physician. Generally,
dose to the actual dose delivered to the bladder, rectum, and patients were aligned to implanted fiducials if they were
targets. The workflows were designed to edit the trans- present. If nodal volumes or prostate fossa were being
ferred contours and evaluate them as part of the process. treated, the local bony anatomy was used. The CBCT scans
along with any shifts made were sent to the MIM software
Case Description for analysis.

Patient selection Treatment planning

Patients for this study were selected in a retrospective as Specified by the physician, the treatment goals included
well as prospective manner. Each patient was diagnosed with uniform coverage of the PTV while keeping the dose to the
prostate adenocarcinoma confirmed via biopsy. Selected pa- bladder and rectum as low as possible. The prescription dose
tients were to have the external radiotherapy treatments to varied depending on whether the target was the intact pros-
the intact prostate with or without nodal involvement or to tate or the prostate fossa. The treatment plans were created
the prostate bed. using either the volumetric arc therapy or static intensity-
For the simulation, patients were placed in the supine po- modulated radiation therapy. Patient no. 1 was treated to
sition with a pillow under the head and arms high on their the intact prostate using a 2-full arc volumetric arc therapy.
chest, holding a ring. The legs were immobilized in a Vac- Patients 2 through 5 were treated using static intensity-
Lok (CIVCO Radiotherapy, Coralville, IA). Each patient received modulated radiation therapy with a 9-beam arrangement.
ARTICLE IN PRESS
H. Bonsall et al. / Medical Dosimetry ■■ (2018) ■■–■■ 3

Table 1
Prescription details for the initial plan for patient nos. 1-5

Site Patient no. 1 Patient no. 2 Patient no. 3 Patient no. 4 Patient no. 5
Prostate Prostate fossa Prostate Prostate Prostate
Prescription dose 54 Gy/30 fx 50.4 Gy/28 fx 45 Gy/25 fx 50 Gy/20 fx 45 Gy/25 fx
Planning technique VMAT, 2 full arcs IMRT, 9 beams IMRT, 9 beams IMRT, 9 beams IMRT, 9 beams

The prescription for each patient is summarized in Table 1. For each patient, the daily CBCT was completed and
For the purpose of this study, only the initial plans without exported into MIM software, where the assistant looked
the boost were used. for the daily CBCTs as they were exported. Once it recog-
Desired dose to critical structures was not to exceed the nized a new CBCT, MIM automatically ran through a
guidelines set forth by the Quantitative Analyses of Normal workflow that performed a deformable registration with
Tissue Effects in the Clinic (QUANTEC) and the Radiation dose and contour transfer. Another workflow was then used
Therapy Oncology Group (RTOG). Constraints by the RTOG by the reviewing medical dosimetrist to go through each
were utilized for both prostate and prostate fossa plans; ad- of the deformed contours and make edits where neces-
ditionally, QUANTEC constraints were used for the prostate sary. The software then adjusted and generated the report
fossa plans. As per QUANTEC constraints, volume of rectum of the findings. These reports provided information on the
receiving 50 Gy (V50) cannot exceed 50% of the total organ organ volume changes, planning dose to each organ com-
volume (V50 ≤ 50%), volume receiving 60 Gy cannot exceed pared to the actual dose received during a single fraction,
35% (V60 ≤ 35%), and volume receiving 65 Gy cannot exceed and the projected dose to each organ at the completion of
25% (V65 ≤ 25%). For the bladder, volume receiving 65 Gy the treatment course.
cannot exceed 50% of the organ volume (V65 ≤ 50%) and
volume receiving 70 Gy cannot exceed 35% (V70 ≤ 35%). Fol-
lowing the RTOG constraints, volume of the rectum receiving Plan analysis and evaluation
40 Gy cannot exceed 55% (V40 ≤ 55%) and volume of the
bladder receiving 40 Gy cannot exceed 70% (V40 ≤ 70%). The Patient no. 1 was analyzed retrospectively for all 30 treat-
summary of the dose-volume constraints can be found in ment fractions. The comparative cumulative dose-volume
Table 2. histograms (DVHs) showed no significant variations in the
actual delivered dose when compared to the planning dose
Table 2
(Fig. 1). The PTV coverage was slightly less than antici-
Constraints from QUANTEC and RTOG for OAR pated, with 95.85% of the volume being covered by 95% of
the prescription dose vs 98.07% of the volume covered by
QUANTEC RTOG
95% of the prescription dose during the initial planning.
Bladder V65 ≤ 50% V40 ≤ 70%
Overall, the actual dose to the bladder was slightly lower than
V70 ≤ 35%
Rectum V50 ≤ 50% V40 ≤ 55% the planned dose, and the actual dose to the rectum was
V60 ≤ 35% slightly higher than the planned dose but without signifi-
V65 ≤ 25% cant variation. Doses to the CTV were also comparable.

Fig. 1. Patient no. 1—cumulative DVH showing planned dose vs total delivered dose; 30 fractions (fx). (Color version of figure is available online.)
ARTICLE IN PRESS
4 H. Bonsall et al. / Medical Dosimetry ■■ (2018) ■■–■■

Fig. 2. Patient no. 2—cumulative DVH showing planned dose vs total delivered dose; 28 fx. (Color version of figure is available online.)

Patient no. 2 was analyzed while on treatment. The com- received the dose that was similar to the planned dose. The
parative cumulative DVH showed actual dose delivered to dose to the rectal volume was higher than planned, with 15%
the CTV to be comparable to the planning dose (Fig. 2). The of the volume receiving 43.36 Gy vs 37.47%.
PTV received lower dose than planned, with 95% of the dose Patient no. 5 was analyzed while on treatment. The CTV
received by 95.53% of the volume vs the planned 98.29% of and the PTV coverage were comparable between the planned
the volume. The dose received by the rectum was consis- dose and actual received dose (Fig. 5). The cumulative dose
tent during daily treatments but was overall higher than the to the bladder volume also did not deviate from the initial-
initially planned dose. The dose to the bladder was compa- ly planned dose. Rectal volume, however, received a higher-
rable and slightly lower than the planned dose. than-anticipated dose. The dose to 15% of the rectal volume
The comparative cumulative DVH for patient no. 3, who was 43.29 Gy, an increase from 40.55 Gy.
was analyzed while on treatment, showed a comparable cov- The reports were generated for each daily fraction for all
erage of the CTV and the PTV (Fig. 3). The actual dose to 15% the patients. The DVH created showed the variations between
of the rectum was 36.75 Gy, an increase from the planned planning doses and actual single fraction doses. Daily varia-
dose of 32.91 Gy. The rectal volume overall also received a tions of dose received by the OAR were recorded. Overall,
higher dose than planned. The dose received by the bladder the dose received by the OAR on a daily basis fluctuated only
was very comparable. slightly, with more notable changes recorded in the dose re-
Patient no. 4 was analyzed while on treatment using a ceived by the rectal volume. In all 5 patients, the greatest
hypo-fractionated prescription and showed no deviation in variation in dose for the 15% of the rectal volume was 15.72%
CTV coverage on the comparative cumulative DVH (Fig. 4). higher than the planned dose. The dose received by 15% of
The PTV coverage was reduced from 96.66% of the volume the bladder volume was 3% lower than the planned dose.
covered by the prescription dose to 91.25%. The bladder Daily variations in OAR volumes were also recorded with

Fig. 3. Patient no. 3—cumulative DVH showing planned dose vs total delivered dose; 25 fx. (Color version of figure is available online.)
ARTICLE IN PRESS
H. Bonsall et al. / Medical Dosimetry ■■ (2018) ■■–■■ 5

Fig. 4. Patient no. 4—cumulative DVH showing planned dose vs total delivered dose; 20 fx. (Color version of figure is available online.)

significant fluctuations from fraction to fraction. Fig. 6 shows therefore not reviewed by physicians. For further research,
the variations in bladder volume between the initial plan- physicians can be involved to determine whether the results
ning CT and each daily fraction for all 5 patients. Fig. 7 shows are clinically significant, as dose tracking during the initial
the daily variations in rectal volume for all patients. portion of the plan will provide useful information when de-
signing the boost plan. Because the actual delivered dose will
Conclusion be available from the initial plan, it will provide the planner
with a better idea of how much of the dose is still allow-
For each patient, there were notable variations when com- able to the OAR. The target coverage can also be adjusted
paring planning dose with the actual dose delivered to the for the boost plan if it was not adequate during the initial
target and OAR, especially the rectum daily dose. The daily plan. In addition, knowing the actual doses received might
changes in turn add up to a cumulative dose that differs from aid physicians in deciding whether further dose escalation
the initial planning dose. These findings prove that there is is feasible or if the OAR would be overdosed. The next step
utility in using the MIM software to track the changes. When in utilizing the software could be to re-optimize the plan
completed while the patients are on treatment, the entire before the daily treatment if the variations are significant
process from uploading the daily CBCT scan into the soft- enough.
ware to editing the contours and generating the reports takes Because the software monitors daily doses, it takes each
between 5 and 10 minutes. Analyzing all the CBCTs retro- fraction into account and not only compares the daily varia-
spectively was proven to be too time-consuming. This renders tions, but also creates a projected final dose. This function
the software most useful for the patients who are cur- of MIM software makes analyzing the dose delivered on a
rently on treatment. daily basis very useful. It makes possible to detect changes
The data collected were for research purposes and for eval- that are so significant that they might prompt another plan-
uation of the efficiency of the offline ART process, and were ning scan and a new treatment plan. Despite some of the

Fig. 5. Patient no. 5—cumulative DVH showing planned dose vs total delivered dose; 25 fx. (Color version of figure is available online.)
ARTICLE IN PRESS
6 H. Bonsall et al. / Medical Dosimetry ■■ (2018) ■■–■■

Fig. 6. Daily variations in bladder volume for patient nos. 1-5 (fx. 0 = initial planning volume). (Color version of figure is available online.)

software limitations, such as the need to manually edit de- manageable. Based on the study findings, the changes dis-
formable contours, the workflows designed by the MIM played during daily treatments as well as the actual
software engineers make the process of offline adaptive cumulative dose variations justify the extra workload that
therapy as automated as possible, making it clinically goes into analyzing the daily CBCTs with MIM software.

Fig. 7. Daily variations in rectal volume for patient nos. 1-5 (fx. 0 = initial planning volume). (Color version of figure is available online.)
ARTICLE IN PRESS
H. Bonsall et al. / Medical Dosimetry ■■ (2018) ■■–■■ 7

References patients treated with IMRT. Int. J. Radiat. Oncol. Biol. Phys. 87(2,
suppl.):S176; 2013. http://dx.doi.org/10.1016/j.ijrobp.2013.06.454.
1. Prostate Cancer Statistics. Centers for Disease Control and Prevention 5. Prabhakar, R.; Oates, R.; Jones, D.; et al. A study on planning organ
Web site. Updated May 23, 2017. Available at: https://www.cdc.gov/ at risk volume for the rectum using cone beam computed tomography
cancer/prostate/statistics. Accessed June 21, 2017. in the treatment of prostate cancer. Med. Dosim. 39(1):38–43; 2014.
2. Huang, T.C.; Chou, K.T.; Yang, S.N.; et al. Fractionated changes in http://dx.doi.org/10.1016/j.meddos.2013.09.003.
prostate cancer radiotherapy using cone-beam computed tomography. 6. Noel, C.E.; Santanam, L.; Olsen, J.R.; et al. An automated method for
Med. Dosim. 40(3):222–5; 2015. http://dx.doi.org/10.1016/j.meddos adaptive radiation therapy for prostate cancer patients using
.2014.12.003. continuous fiducial-based tracking. Phys. Med. Biol. 55(1):65–82; 2010.
3. Lim-Reinders, S.; Keller, B.M.; Al-Ward, S.; et al. Online adaptive http://dx.doi.org/10.1088/0031-9155/55/1/005.
radiation therapy. Int. J. Radiat. Oncol. Biol. Phys. 99(4):994–1003; 2017. 7. Pirozzi, S.; Piper, J.; Nelson, A.; et al. Evaluation of deformable prostate
http://dx.doi.org/10.1016/j.ijrobp.2017.04.023. cone beam computed tomography (CBCT) contouring methods for
4. Tonlaar, N.Y.; Marina, O.; Brabbins, D.S.; et al. Use of weekly cone beam adaptive radiation therapy. Int. J. Radiat. Oncol. Biol. Phys. 87(2,
CT for adaptive radiation therapy decreases toxicity for prostate cancer suppl.):S719; 2013. http://dx.doi.org/10.1016/j.ijrobp.2013.06.1904.

You might also like