Renishaw plc warrants its equipment provided that it is installed exactly
as defined in associated Renishaw documentation.
PATENTS
Features of Renishaw’s PH50 and similar products are the subjects of
the following patents and patent applications:
JP 1735296
JP 2,084,393
US 4168576
US 4313263
US 4571847
GB 2298488
DE 19607680.3
Renishaw plc
Metrology Division
PH50
Motorised Probe Head
System
with
TP50 Touch Trigger Probe
User’s Guide
SAFETY
Pinch hazards exist between moving parts and between moving and static parts.
Do not hold the probe head during movements, or during manual probe changes.
Beware of unexpected movement. The user should remain outside of the full
working envelope of probe head/extension/probe combinations.
In all applications involving the use of machine tools or CMMs, eye protection is
recommended.
There are no user serviceable parts inside Renishaw mains powered units. Return
defective units to an authorised Renishaw Customer Service Centre.
Replace blown fuses with new components of the same type. Refer to the
ELECTRICAL REQUIREMENTS section of the relevant product documentation.
For instructions regarding the safe cleaning of Renishaw products, refer to the
MAINTENANCE section of the relevant product documentation.
It is the machine supplier’s responsibility to ensure that the user is made aware
of any hazards involved in operation, including those mentioned in Renishaw
product documentation, and to ensure that adequate guards and safety interlocks
are provided.
Under certain circumstances the probe signal may falsely indicate a probe
seated condition. Do not rely on probe signals to stop machine movement.
SECURITE
Aucune pièce des machines Renishaw alimentées sur secteur ne peut être réparée
par l’utilisateur. Renvoyer toute machine défectueuse à un Centre Après Vente
Renishaw agréé.
Remplacer les fusibles grillés par des composants neufs du même type. Consulter
la section SPECIFICATIONS ELECTRIQUES de votre documentation.
Les conseils de nettoyage en toute sécurité des produits Renishaw figurent dans
la section MAINTENANCE de votre documentation.
Dans certains cas, il est possible que le signal du palpeur indique à tort l’état que
le palpeur est au repos. Ne pas se fier aux signaux du palpeur qui ne garantissent
pas toujours l’arrêt de la machine.
SICHERHEITSANWEISUNGEN
Auf unerwartete Bewegungen achten. Der Anwender soll sich immer außerhalb
des Meßtasterkopf-Arm-Meßtaster-Bereichs aufhalten.
Es obliegt dem Maschinenlieferanten, den Anwender über alle Gefahren, die sich
aus dem Betrieb der Ausrüstung, einschließlich der, die in der Renishaw
Produktdokumentation erwähnt sind, zu unterrichten und zu versichern, daß
ausreichende Sicherheitsvorrichtungen und Verriegelungen eingebaut sind.
SICUREZZA
Tra le parti in moto o tra le parti in moto e quelle ferme esiste effettivamente il
pericolo di farsi del male pizzicandorsi. Evitare di afferrare la testina della sonda
quando è in moto, oppure quando si effettuano spostamenti a mano.
I fusibili bruciati dovranno essere sostituiti con quelli dello stesso tipo. Consultare
la sezione SPECIFICHE ELETTRICHE della documentazione del prodotto.
Per le istruzioni relative alla pulizia dei prodotti Renishaw, fare riferimento alla
sezione MANUTENZIONE (MAINTENANCE) della documentazione del prodotto.
SEGURIDAD
Existe el peligro de atraparse los dedos entre las distintas partes móviles y entre
partes móviles e inmóviles. No sujetar la cabeza de la sonda mientras se mueve,
ni durante los cambios manuales de la sonda.
Tener cuidado con los movimientos inesperados. El usuario debe quedarse fuera
del grupo operativo completo compuesto por la cabeza de sonda/extensión/
sonda o cualquier combinación de las mismas.
Se recomienda usar protección para los ojos en todas las aplicaciones que
implican el uso de máquinas herramientas y máquinas de medición de
coordenadas.
Dentro de las unidades Renishaw que se enchufan a la red, no existen piezas que
puedan ser mantenidas por el usuario. Las unidades defectuosas deben ser
devueltas a un Centro de Servicio al Cliente Renishaw.
Sustituir los fusibles fundidos con componentes nuevos del mismo tipo. Remitirse
a la sección titulada ELECTRICAL REQUIREMENTS en la documentación sobre
el producto.
SEGURANÇA
Figo de constrição entre peças móveis e entre peças móveis e estáticas. Não
segurar a cabeça da sonda durante o movimento ou durante mudanças manuais
de sonda.
Não há peças que possam ser consertadas pelo utilizador dentro das unidades
Renishaw alimentadas pela rede. Devolver unidades avariadas a um Centro de
Atendimento a Clientes Renishaw.
SIKKERHED
Der er risiko for at blive klemt mellem bevægelige dele og mellem bevægelige og
statiske dele. Hold ikke sondehovedet under bevægelse eller under manuelle
sondeskift.
Pas på uventede bevægelser. Brugeren bør holde sig uden for hele sondehovedets/
forlængerens/sondens arbejdsområde.
Der er ingen dele inde i Renishaw-enhederne, som sluttes til lysnettet, der kan
efterses eller repareres af brugeren. Send alle defekte enheder til Renishaws
kundeservicecenter.
Se maskinleverandørens brugervejledning.
VEILIGHEID
De onderdelen van Renishaw units die op het net worden aangesloten kunnen
niet door de gebruiker onderhouden of gerepareerd worden. U kunt defecte units
naar een erkend Renishaw Klantenservice Centrum brengen of toezenden.
Voor het veilig reinigen van Renishaw produkten wordt verwezen naar het hoofdstuk
ONDERHOUD (MAINTENANCE) in de produktendocumentatie.
SÄKERHET
Risk för klämning existerar mellan rörliga delar och mellan rörliga och stillastående
delar. Håll ej i sondens huvud under rörelse eller under manuella sondbyten.
Se upp för plötsliga rörelser. Anvåndaren bör befinna sig utanför arbetsområdet
för sondhuvudet/förlängningen/sondkombinationerna.
Det finns inga delar som användaren kan utföra underhåll på inuti Renishaws
nätströmsdrivna enheter. Returnera defekta delar till ett auktoriserat Renishaw
kundcentra.
Se maskintillverkarens bruksanvisning.
Under vissa omständigheter kan sondens signal falskt ange att en sond är
monterad. Lita ej på sondsignaler för att stoppa maskinens rörelse.
TURVALLISUUTTA
Fuse replacement
There are two 2 Amp (T) slow-blow fuses which are used for all voltages (one is a
spare). Fuses are replaced as follows:
WARNING: Make sure that only fuses of the specified type are used for
replacement.
ENVIRONMENTAL REQUIREMENTS
The following environmental conditions comply with (or exceed) BS EN 61010-
1:1993
Indoor use IP30 (no protection against water)
AltitudeUp to 2000m
Operating temperature: PH50 +10° to +44°C
PHC50 0°C to +50°C
Storage temperature -10°C to +70°C
Relative humidity 80% maximum for temperatures up to
+31°C, linear decrease to 50% at
Transient overvoltages +50°C
Pollution degree Installation category II
1
Remplacement du fusible
Deux fusibles retardés de 2A sont fournis dont un de remplacement. Pour procéder
à l'échange, veuillez vous conformer à la procédure suivante:
Cet équipement doit impérativement être relié à une terre de protection par
un câble secteur à 3 conducteurs. La prise équipant ce câble doit être
connectée à une prise secteur effectivement équipée d'une borne reliée à la
terre. La liaison entre la terre de protection et le contrôleur PHC50 ne doit
en aucun cas être interrompue par l'utilisation d'un prologateur dépourvu
de conducteur reliant l'équipement à cette terre de protection.
Dieses System muß mit einem geerdeten Schutzleiter über ein 3-adriges Hauptakel
verbunden sein. Der Hauptstecker sollte nur in eine geerdete Steckdose gesteckt
werden. Der geerdete Kontakt sollte nicht unwirksam gemacht werden, indem
man ein Verlängerungskabel ohne Schutzleiter benutzt.
UMGEBUNGSBEDINGUNGEN
Die Forderungen der Richtlinie BS EN 61010-1:1993 sind erfüllt
Note on Documentation 19
3.0 INSTALLATION 39
5.0 MAINTENANCE 48
5.1 General Maintenance 48
5.2 Probe Replacement 49
5.3 Fuse Replacement 52
5.4 Cleaning 52
6.0 FAULT-FINDING 53
6.1 Poor Measurement Performance 55
6.2 Unwanted Triggers 56
6.3 Probe Fails to Re-arm 57
6.4 No Probe Signal 58
6.5 No Head Movement 59
6.6 Obstruct Error 60
6.7 Overload Error 60
6.8 Obstruct, Overload and Stop LEDs 61
For more detailed information on the PH50 system refer to the PH50
Motorised Probe Head System Installation Guide (Part No H-1000-5096)
and the Guide to the Renishaw Extended Command Set (Part No
H-1000-6001).
Associated documentation:
Part No Title
H-1000-5096 PH50 Installation Guide
H-1000-6001 Guide to the Renishaw Extended Command Set
H-1000-5016 HCU1 Hand Control Unit User’s Guide
H-1000-5021 Touch Trigger Probe System User’s Guide
H-1000-3120 CMM Stylus Guide
H-1000-5080 Machine Checking Gauge User’s Guide
H-1000-5050 CMM Probing Systems Technical Specifications
NOTE
Please fill out the User Registration Form at the end of this guide and
return it to Renishaw plc to obtain further information on the PH50 system
and accessories and to enable Renishaw to keep you informed of the
latest product information, special offers, upgrade news and new product
developments.
You may find it useful to make a note of information about your system
here:
CMM Make
CMM Model
Figure 1
• Compact Z axis
• Very small swept volume
• Fast indexing movement
• 0.5µm (2σ) indexing repeatability*
• 7.5° indexing steps on both axes
• Lightweight design
* See section 1.2 Specifications for test conditions
The TP50 precision touch trigger probe (see Figure 2) is based on proven
Renishaw kinematic principles and is designed for use with the PH50
motorised probe head.
Figure 2
Each PH50 system includes a tool kit containing the following items:
• S7 stylus tools for stylus fitment (see section 4.5) - two supplied
The PHC50 (see Figure 3) manages all head and probe system functions
and interfaces with the CMM computer.
Figure 3
The HCU1 (see Figure 4) is a remote hand control unit that enables the
probe head to be used in a manual mode or during a teach cycle. An
LCD dot matrix display provides system status information and error
analysis.
Figure 4
Figure 5
PHC50 Probe Head Controller
1.2 Specifications
Repeatability of position
(2σ) with 21mm stylus 0.5µm (0.00002in)
Uni-directional repeatability
Max (2σ) at stylus tip 0.5µm (0.000014in)
Figure 6
M3 (4 off)
60mm
(2.36in)
87mm
(3.4in) 33mm
(1.3in)
Figure 7 60mm
(2.36in)
25mm
(0.98in)
54mm
(2.13in)
87mm
(3.4in)
33mm
(1.3in)
A Axis
0° to 105°
7mm
(0.28in) A Axis Centre of Rotation
The shank is fixed to the head using 3 off M3 * 6mm socket head cap
screws.
PH50 User’s Guide 31
SYSTEM OVERVIEW
1.3 Accessories
1.3.1 Styli
The M3 mounting thread range can be used directly with the TP50
probe and offers a variety of ball sizes in industrial ruby from 0.5mm
(0.02in) to 5mm (0.19in) diameter. These are available with steel,
tungsten carbide and ceramic stems to cover a wide range of
performance and price options.
For further information on choosing and using styli with the TP50, see
section 4.5.
Figure 8
For further information on the AM1, please refer to the AM1 User’s Guide
(Part No H-1000-4010). If your PH50 has been fitted with an AM1, this
guide should be supplied. If it is not present, please contact your
Renishaw representative who will be pleased to send a copy free of
charge.
Figure 9
CAUTION
The Renishaw MCG (see Figure 10) enables rapid and effective interim
checking of CMM performance as recommended in many existing
standards for CMM volumetric performance assessment.
It is fully compatible with the TP50 and enables fast, automatic machine
evaluation (go/no go checks) on a regular basis.
Figure 10
The Renishaw universal datum sphere (see Figure 11) features quick
and easy adjustment of the ball stem over a wide range of angles,
allowing clearance for probe qualification above, centrally and below
the ball.
Figure 11
DO requalify the probe after an adjustment has been made to the trigger
force
DO take care to avoid a collision between the probe head and the CMM
bed, workpiece etc
DO ensure that the probe is securely mounted in the head (see section
5.2)
DO ensure that the probe head is moved clear of any obstructions before
requesting initialisation (see section 4.4)
DO NOT apply force to any of the moving parts of the probe head
DO NOT remove power from the PH50 system while the probe head is
unlocked if it can be avoided - on power-up head operation will be
suspended and initialisation will be required (see section 4.4)
DO NOT move the axes of the probe head by hand while it is locked
DO NOT move the axes of the probe head by hand while it is unlocked
3.0 INSTALLATION
For information on how to install the PH50 system, please refer to the
PH50 Installation Guide (Part No H-1000-5096).
In manual mode, the optional HCU1 hand control unit (see section 1.1.5)
may be used to control the PH50 probe head manually.
Please refer to the HCU1 User’s Guide (Part No H-1000-5016) for further
details of operation in manual mode.
If the HCU1 is not connected to the PHC50 when power is applied, the
system will enter automatic mode. The system must then be switched
to manual mode by the CMM computer before the HCU1 can be used
to control head and probe functions.
Figure 12
The HOME button is used to initialise the head. It is only active when
the HOME LED is lit and initialisation is required.
Table 1
LED NAME COLOUR FUNCTION
To indicate this state, the PHC50 front panel HOME LED will be lit. If an
HCU1 hand control unit is fitted, the message INITIALISE HEAD will be
shown on the dot matrix display.
The PH50 can be initialised by pressing either the PHC50 front panel
HOME button or the HCU1 HOME key (if the HCU1 is fitted).
WARNING
NOTE
Always fit and tighten styli securely to the probe using the S7 stylus tool
included in the tool kit.
CAUTION
Ensure that all stylus joints are clean and free from dirt or debris, and
that the stylus ball is clean and not damaged.
Ensure that the probing trigger force is set at an appropriate level (see
section 4.6). Check the measurement performance of the system after
any adjustment has been made to the trigger force. Requalify the probe
in every position used after a stylus has been adjusted or changed.
CAUTION
With the A axis positioned at 90°, the trigger force adjuster is accessed
via a hole underneath the head (see Figure 13).
Figure 13
The TP50 probe has a trigger force range of 11g to 30g (0.39oz to
1.05oz).
For further information on the Renishaw gram gauge and setting trigger
force, please refer to section 1.3.3 in this guide or to the Renishaw
Gram Gauge Data Sheet Part No H-1000-2033.
If your CMM has been fitted with probe DAMPing capability, this feature
will be controlled by the CMM computer. Please refer to your CMM
supplier for details of the specific use and control of probe DAMPing on
your CMM.
The PHC50 front panel PROBE DAMPED LED indicates the status of
probe DAMPing (see section 4.3).
CAUTION
5.0 MAINTENANCE
There are no user serviceable parts inside any PH50 system units.
WARNING
The TP50 can be removed and replaced by the user should it become
damaged or require servicing.
CAUTION
WARNING
CAUTION
Figure 14
2. Gently push the new probe into the probe head (pushing against
the probe body, not the stylus holder), aligning the location
key on the probe with the slot in the A axis, until it is fully engaged
4. Tighten the probe locking screw by hand while pressing the probe
gently into the head until the probe is fixed securely in place. Do
not overtighten.
CAUTION
WARNING
5.4 Cleaning
PH50 system units may be wiped with a clean, dry, lint-free cloth. The
units are not sealed against water.
6.0 FAULT-FINDING
Use Table 2 to identify the problem you are experiencing with your PH50
system and refer to the relevant section in this guide.
Table 2
The optional HCU1 hand control unit is also able to diagnose certain
error conditions. Please refer to the HCU1 User’s Guide (Part No
H-1000-5016) for further details.
NOTE
Table 3
POSSIBLE CAUSE CHECKS/REMEDIES
Loose mounting of Ensure all mounting screws are tight and mounting
probe head to the CMM is secure
Probe not installed Remove probe and re-fit. See section 5.2 for
correctly details of probe replacement
Poor stylus assembly Ensure stylus joints are kept to a minimum and all
joints are clean and secure
Table 4
Table 5
Table 6
Table 7
OBSTRUCT ERROR and STOP LEDs are lit on PHC50 front panel.
The head has been obstructed while moving to the requested position
and is unable to reach that position or unable to lock into it.
Table 8
OVERLOAD ERROR and STOP LEDs are lit on PHC50 front panel.
Table 9
There has been an internal system failure within the PH50 probe head.
Table 10
CAUSE CHECKS/REMEDIES
Email: genenq@renishaw.co.uk
NOTE
Name ..................................................................................
Position ...............................................................................
Company Name ..................................................................
Telephone ...........................................................................
CMM make and model .........................................................
Precision Metrology