Professional Documents
Culture Documents
Richmond B.V.
Lintelerweg 60A
NL-7556 PD Hengelo(Ov)
Warning:
Validation report
Contents
1. Summary ...................................................................................................................................... 3
2. Installation qualification: Steam sterilizer 4 .................................................................................. 5
3. Operational qualification: Steam sterilizer 4 ................................................................................. 8
4. Measurement: Leakage test ....................................................................................................... 10
5. Graph: Leakage test ................................................................................................................... 11
6. Measurement: Bowie & Dick test 3min/134................................................................................ 13
7. Graph: Bowie & Dick test 3min/134............................................................................................ 15
8. Measurement: Unloaded 4min/134 ............................................................................................ 17
9. Graph: Unloaded 4min/134 ........................................................................................................ 19
10. Graph: Unloaded 4min/134 ........................................................................................................ 21
11. Measurement: 50% load Instruments 4 min/134........................................................................ 23
12. Graph: 50% load Instruments 4 min/134 .................................................................................... 25
13. Measurement: 100% load Instruments 4min/134....................................................................... 27
14. Graph: 100% load Instruments 4min/134 ................................................................................... 30
15. Measurement: Mixed load containers/paper 4min/134 .............................................................. 32
16. Graph: Mixed load containers/paper 4min/134 .......................................................................... 34
17. Measurement: Worst case container/textile 4min/134 ............................................................... 36
18. Graph: Worst case container/textile 4min/134 ........................................................................... 39
19. Calibration: Default protocol Steam Sterilizers (120-130-140) ................................................... 41
20. Description of the validation system ........................................................................................... 42
21. Attachments................................................................................................................................ 45
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 3 of 45
Richmond BV UIN: 398
1. Summary
This validation document reports the planning and execution of measurements and tests regarding the
validation the sterilizer mentioned in the assignment description.
The previous contents show the planned and executed validation tasks. The results and other details
are elaborated in the indicated paragraphs of this report. Per measurement they describe the process,
the load under test and the arrangement of the measurement sensors.
The measurement results and their graphical representations are added including the detected
deviations and further measurement details.
This summary holds the details of the validation assignment, the persons and organizations involved
with the validation. The justification of the applied standards are given as well as the used
measurement equipment and the persons responsible fort he interpretation of the result and the final
validation conclusions.
In the conclusion a summary is given of the validation results and eventual detected deviations and
where applicable comments from already held discussions on that and comparing them with previous
validation results or otherwise important circumstances.
Assignment
Assignee: Richmond BV
Assignment contract: our offer dated 02-02-2007
Justification
The validation is set up and executed according to the following applicable European standards:
ISO-EN 285: Sterilizers – Steam sterilizers – Large sterilizers.
ISO-EN 554: Sterilization of medical devices. Validation and routine control of steam (Large
sterilizers).
ISO-EN 285 describes general specifications regarding steam sterilizers and with respect to validation
specific requirements for leakage test and indicating/recording instruments.
ISO-EN 554 describes the actual validation process and the requirements for the measurement
equipment used as well as the specific required measurement values and accuracies with regard to
the course of the sterilization process.
De actual execution of the measurements en the processing of the measurement results was done
with the help of IAVS-M2 validation measurement system with ValidR validation and reporting
software. This measurement system, developed by Richmond BV, Hengelo, the Netherlands,
complies with the requirements set by the standards and it performs most of the criteria verifications
automatically. A detailed description of this equipment and the accompanying calibration certificates is
given further on in this report.
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 4 of 45
Richmond BV UIN: 398
Conclusions
1. The sterilizer complies with the involved standards and in relation to the combinations of process,
product, product packaging and load concerning the following processes:
Ster. Ster.
Nr. temp. time Load
(°C) (min.)
6 - - Leakage test
9 134 4 50% loaded Instruments 4min/134
10 134 4 100% loaded Instruments 4min/134
11 134 4 Mixed load containers/paper 4min/134
[Commentaar]
2. The sterilizer DOES NOT COMPLY with the involved standards and in relation to the combination
of process, product, product packaging and load concerning the following processes:
Ster. Ster.
Nr. temp. time Load
(°C) (min.)
7 134 4 Bowie&Dick test:
- Jacket temperature lower than theoretical temperature
1. Recommende revalidation
According to the involved standards a revalidation is to be performed at latest in March 2008 , or
earlier when any intervention is done that could influence the combination of process, product,
product packaging and load.
2. It is advised to make written work instructions and use those in the daily routine of
loading/unloading of the sterilizer.
Signature
Signature:
Report
The following paragraphs contain the specifications and measurement results for every part of the
validation and a description of the calibration and measurement equipment according to the
arrangement in the contents.
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 5 of 45
Richmond BV UIN: 398
Machine data
Machine number : 4
Location : University Clinic Hengelo
Reason for identification : Revalidation 2007
Identification number : 6648GG6
Date identification : January 7, 2006
Date previous identification : n.a.
Changes after previous identification : n.a.
General Data
If the sterilizers uses several different methods (depending on the cycle), these should all be
described clearly per cycle.
Main data
Inner dimensions :
Sterilizer chamber (H*W*D) : 670 mm * 670 mm * 1330 mm
Volume (H*W*D) : 670 mm * 670 mm * 1300 mm
Orientation of the chamber : Horizontal
Double jacket : Yes
Heating surface as % of the
chamber surface excl. doors : More than 50% %
Way of loading : Carts
Door principle : Sliding
: Vertical
: Single door
Door closing system : Steam
Steam
Filter Type :
Position condensate outlet : central in the floor
Air
Vacuum
Control system
Registration system
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 7 of 45
Richmond BV UIN: 398
Performed by : V. Engineer
Job title : Validation engineer
Date & time : March 15, 2007
Steam pressure
Movable : yes
Fully opened : yes
External leakage : no
Movable : yes
Fully opened : yes
External leakage : no
Movable : yes
Fully opened : yes
External leakage : no
Movable : yes
Fully opened : yes
External leakage : no
Steamvalves
Steamfilter
Doorseal
Clean : yes
Damaged : no
Airfilter
Visible damage : no
Wet (or previously wet) : no
Filter replacing interval : 6 months
Date last replaced : Oct. 2006
Clean : yes
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 9 of 45
Richmond BV UIN: 398
Steam inlet
Attachment ok : yes
Clean : yes
Free of water : yes
Other remarks
Noticable leakages : no
Odd noises : no
Registration
Deepest Vacuum
Jacket pressure
Chamber pressure
Sterilisation temperature
General
Sterilizer name CSD4
Description Steam sterilizer 4
Performed by V. Engineer
Job title Validation Engineer
Date & time 15-3-2007 14:03:45
Conclusion
The validation system concludes the following:
- Compliance with the guidelines was found.
Results
Start time Duration Pressure Difference Recorded
(hh:mm:ss:) (hh:mm:ss:) at Start Pressure
(kPa) (kPa) (kPa)
Evacuate 0:00:00 0:06:06 101.3 -97.3
Stabilizing 0:06:06 0:05:00 4.0 0.1 4.2 ( 0.2)
Leakage test 0:11:06 0:09:57 4.1 0.0 4.2 ( 0.1)
Aeration phase 0:21:03 0:02:00 4.1 97.1 4.6 ( 0.5)
End of measurement 0:23:03 101.2
Legend
── Pressure
End of report
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 13 of 45
Richmond BV UIN: 398
Machine
Name : CSD4
Description : Steam sterilizer 4
Dimensions : 670 x 670 x 1300 (H x W x D, in mm)
Cycle
- Nominal sterilisation temperature 134 °C
- Nominal sterilisation time 0:03:00 uu:mm:ss
- Applicable guideline: EN554: Sterilization of Medical Devices. Validation and routine control of
sterilization by moist heat. (large sterilizers)
Conclusion
The validation system concludes the following:
- Compliance with the guidelines was not found (see further in this report).
Product combination
Product Bowie & Dick test package
Package Sheets of 100% cotton; size 940 x 1180 mm.
Contents 32 sheets of 100% cotton , weigth 7090 grams
Load Size about 240 x 340 x 250 mm.
Thermocouple positioning
Tc. Nr. Type pt100 nr. Level X-Position Y-Position
1 H 1 1 3 4
In the drain
2 J 0 3 1 3
Heat conducting against the jacket
3 R 0 10 3 4
Free in the open space
4 P 0 4 2 3
In package heart
5 P 0 4 2 3
In corner of same sheet as couple 4.
6 P 0 4 4 3
In corner of same sheet as couple 4.
8 P 0 5 3 4
In top of package; under first textile layer
9 P 0 3 3 4
In bottom of package; above first layer of textile
10 R 0 6 3 4
2 cm above package
Thermocouple set
S/N 1080
Description CSD set 1, 1/2 inch feedthrough
Shielded set No
Number of couples 10
Time table
Duration
Start Pre phase 14:13:51 0:00:00 0:18:26
Start Sterilisation phase 0:18:26 0:03:01
Start Post phase 0:21:27 0:08:52
End of measurement 0:30:19
Equilibrationtime: 0:00:08
Sterilisation phase
Pressure (kPa) Start (0:18:27) Middle (0:19:57) End (0:20:58)
ValidR 306.0 (0.0) 312.1 (0.0) 312.0 (0.0)
Control 311.6 (5.6) 312.0 (-0.1) 311.3 (-0.7)
Registration 311.6 (5.6) 312.0 (-0.1) 311.3 (-0.7)
End of report
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 17 of 45
Richmond BV UIN: 398
Machine
Name : CSD4
Description : Steam sterilizer 4
Dimensions : 670 x 670 x 1300 (H x W x D, in mm)
Cycle
- Nominal sterilisation temperature 134 °C
- Nominal sterilisation time 0:04:00 uu:mm:ss
- Applicable guideline: EN554: Sterilization of Medical Devices. Validation and routine control of
sterilization by moist heat. (large sterilizers)
Conclusion
The validation system concludes the following:
- Compliance with the guidelines was not found (see further in this report).
Product combination
Product n.a.
Package n.a.
Contents n.a.
Load n.a..
Thermocouple positioning
Tc. Nr. Type pt100 nr. Level X-Position Y-Position
1 H 1 1 3 4
In the drain
2 J 0 3 1 3
Heat conducting against side wall; on first rib (not heated)
3 J 0 3 1 3
Heat conducting against side wall; on second rib (heated)
4 R 0 2 1 1
5 R 0 2 5 1
6 R 0 10 1 7
On pincher
8 R 0 10 5 7
9 R 0 10 1 1
10 R 0 2 5 7
Thermocouple set
S/N 1080
Description CSD set 1, 1/2 inch feedthrough
Shielded set No
Number of couples 10
Time table
Duration
Start Pre phase 14:17:56 0:00:00 0:17:51
Start Sterilisation phase 0:17:51 0:03:59
Start Post phase 0:21:50 0:36:06
End of measurement 0:57:56
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 18 of 45
Richmond BV UIN: 398
Equilibrationtime: 0:00:26
Sterilisation phase
Pressure (kPa) Start (0:17:53) Middle (0:19:53) End (0:21:23)
ValidR 306.6 (0.0) 311.6 (0.0) 312.2 (0.0)
Control 303.9 (-2.7) 311.3 (-0.3) 311.0 (-1.2)
Registration 308.2 (1.6) 311.3 (-0.3) 311.0 (-1.2)
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 19 of 45
Richmond BV UIN: 398
End of report
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 21 of 45
Richmond BV UIN: 398
End of report
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 23 of 45
Richmond BV UIN: 398
Machine
Name : CSD4
Description : Steam sterilizer 4
Dimensions : 670 x 670 x 1300 (H x W x D, in mm)
Cycle
- Nominal sterilisation temperature 134 °C
- Nominal sterilisation time 0:04:00 uu:mm:ss
- Applicable guideline: EN554: Sterilization of Medical Devices. Validation and routine control of
sterilization by moist heat. (large sterilizers)
Conclusion
The validation system concludes the following:
- Compliance with the guidelines was found.
Product combination
Product Instruments
Package Containers
Contents Instruments
Load Half full
Thermocouple positioning
Tc. Nr. Type pt100 nr. Level X-Position Y-Position
1 H 1 1 3 4
I the drain; through the sive, about 10 cm inside (and back upwards).
2 J 0 3 1 3
Heat conducting against the jacket
3 R 0 11 3 4
Free in the open space
4 P 0 2 2 1
Attched to pincher
5 P 0 2 3 1
At the bottom of the net
6 P 0 2 3 4
On pincher
8 P 0 2 3 4
On sissers
9 P 0 4 5 7
Attached to metal handle
10 R 0 4 4 7
2 cm above container
Thermocouple set
S/N 1080
Description CSD set 1, 1/2 inch feedthrough
Shielded set No
Number of couples 10
Time table
Duration
Start Pre phase 14:21:42 0:00:00 0:22:42
Start Sterilisation phase 0:22:42 0:03:59
Start Post phase 0:26:41 0:35:11
End of measurement 1:01:52
Equilibrationtime: 0:00:15
Sterilisation phase
Pressure (kPa) Start (0:22:44) Middle (0:24:46) End (0:26:16)
ValidR 308.4 (0.0) 311.7 (0.0) 311.8 (0.0)
Control 308.1 (-0.3) 310.9 (-0.8) 311.6 (-0.2)
Registration 308.1 (-0.3) 310.9 (-0.8) 311.6 (-0.2)
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 25 of 45
Richmond BV UIN: 398
End of report
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 27 of 45
Richmond BV UIN: 398
Machine
Name : CSD4
Description : Steam sterilizer 4
Dimensions : 670 x 670 x 1300 (H x W x D, in mm)
Cycle
- Nominal sterilisation temperature 134 °C
- Nominal sterilisation time 0:04:00 uu:mm:ss
- Applicable guideline: EN554: Sterilization of Medical Devices. Validation and routine control of
sterilization by moist heat. (large sterilizers)
Conclusion
The validation system concludes the following:
- Compliance with the guidelines was found.
Product combination
Product Instruments
Package Containers
Contents Instruments
Load Full
Thermocouple positioning
Tc. Nr. Type pt100 nr. Level X-Position Y-Position
1 H 1 1 3 4
I the drain; through the sive, about 10 cm inside (and back upwards).
2 J 0 3 1 3
Heat conducting against the jacket
3 R 0 11 3 4
Free in the open space
4 P 0 2 2 1
Attached to kidney bowle
5 P 0 2 3 1
At the bottom of the net
6 P 0 4 3 3
On stainless steel bowle
8 P 0 4 4 3
At the bottom of the net
9 P 0 6 5 7
Attached to metal handle
10 R 0 6 4 7
2 cm above container
Thermocouple set
S/N 1080
Description CSD set 1, 1/2 inch feedthrough
Shielded set No
Number of couples 10
At the top, from load side to unload side, from left to right:
- Jaw surgery extraction pincer, Jaw surgery extraction pincer
- Jaw 38 extracter; Jaw extraction pincer
- Jaw surgery extraction pincer; Jaw surgery extraction pincer
- Jaw surgery 38, Instruments; Jaw 38, Instruments 38
Time table
Duration
Start Pre phase 14:23:39 0:00:00 0:29:11
Start Sterilisation phase 0:29:11 0:04:00
Start Post phase 0:33:11 0:36:10
End of measurement 1:09:21
Equilibrationtime: 0:00:12
Sterilisation phase
Pressure (kPa) Start (0:29:12) Middle (0:31:14) End (0:32:43)
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 29 of 45
Richmond BV UIN: 398
End of report
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 32 of 45
Richmond BV UIN: 398
Machine
Name : CSD4
Description : Steam sterilizer 4
Dimensions : 670 x 670 x 1300 (H x W x D, in mm)
Cycle
- Nominal sterilisation temperature 134 °C
- Nominal sterilisation time 0:04:00 uu:mm:ss
- Applicable guideline: EN554: Sterilization of Medical Devices. Validation and routine control of
sterilization by moist heat. (large sterilizers)
Conclusion
The validation system concludes the following:
- Compliance with the guidelines was found.
Product combination
Product Instruments
Package Containers
Contents Instruments
Load Half full
Thermocouple positioning
Tc. Nr. Type pt100 nr. Level X-Position Y-Position
1 H 1 1 3 4
I the drain; through the sive, about 10 cm inside (and back upwards).
2 J 0 3 1 3
Heat conducting against the jacket
3 R 0 11 3 4
Free in the open space
4 P 0 6 2 1
In laminate packaging
5 R 0 6 3 1
2 cm aboce packaging witk couple 4.
6 P 0 4 3 4
Folded in jacket
8 P 0 4 4 4
Folded in jacket; under first layer.
9 P 0 2 3 7
Attached to kidney bowle
10 R 0 2 4 7
on the container
Thermocouple set
S/N 1080
Description CSD set 1, 1/2 inch feedthrough
Shielded set No
Number of couples 10
Behind that:
- Basket with textile (coats)
- Two half heigth containers 'Base net'
Behind that:
- Basket with textile (coats) and 1 laminate packaging
- Two half heigth containers 'Base net' and 'Hip set'
Behind that:
- Basket with laminate packaging
- Two half heigth containers 'Base net' and 'Belly net'
Time table
Duration
Start Pre phase 14:26:38 0:00:00 0:28:20
Start Sterilisation phase 0:28:20 0:03:59
Start Post phase 0:32:19 0:39:16
End of measurement 1:11:35
Equilibrationtime: 0:00:04
Sterilisation phase
Pressure (kPa) Start (0:28:21) Middle (0:30:23) End (0:31:51)
ValidR 305.7 (0.0) 312.0 (0.0) 311.7 (0.0)
Control 307.4 (1.7) 311.6 (-0.4) 311.0 (-0.7)
Registration 307.4 (1.7) 311.6 (-0.4) 311.0 (-0.7)
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 34 of 45
Richmond BV UIN: 398
End of report
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 36 of 45
Richmond BV UIN: 398
Machine
Name : CSD4
Description : Steam sterilizer 4
Dimensions : 670 x 670 x 1300 (H x W x D, in mm)
Cycle
- Nominal sterilisation temperature 134 °C
- Nominal sterilisation time 0:04:00 uu:mm:ss
- Applicable guideline: EN554: Sterilization of Medical Devices. Validation and routine control of
sterilization by moist heat. (large sterilizers)
Conclusion
The validation system concludes the following:
- Compliance with the guidelines was not found (see further in this report).
Product combination
Product
Package
Contents
Load
Thermocouple positioning
Tc. Nr. Type pt100 nr. Level X-Position Y-Position
1 H 1 1 3 4
I the drain; through the sive, about 10 cm inside (and back upwards).
2 J 0 3 1 3
Heat conducting against the jacket
3 R 0 11 3 4
Free in the open space
4 P 0 2 3 2
On pincher
5 P 0 2 4 2
On sissers
6 P 0 6 3 3
Folded in the textile
8 P 0 6 4 3
In same sheet as couple 6
9 P 0 6 3 7
Folded in the textile
10 R 0 6 4 7
2 cm aboce packaging witk couple 4.
Thermocouple set
S/N 1080
Description CSD set 1, 1/2 inch feedthrough
Shielded set No
Number of couples 10
Loaded with 4 stacks of containers (2 high) with on top of that two baskets with cotton (sheets and
laminate).
Behind that:
- Textile
- OK Base net
- Ok Base net
Behind that:
- Textile
- OK Base net
- Ok Base net
Behind that:
- Textile
- OK Bone net small
- Ok Base net
Time table
Duration
Start Pre phase 14:28:40 0:00:00 0:33:41
Start Sterilisation phase 0:33:41 0:04:01
Start Post phase 0:37:42 0:37:04
End of measurement 1:14:46
Equilibrationtime: 0:00:12
Sterilisation phase
Pressure (kPa) Start (0:33:42) Middle (0:35:44) End (0:37:14)
ValidR 306.1 (0.0) 312.2 (0.0) 312.1 (0.0)
Control 306.5 (0.4) 311.8 (-0.4) 312.0 (-0.1)
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 38 of 45
Richmond BV UIN: 398
End of report
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 41 of 45
Richmond BV UIN: 398
Protocol
Name : DefSteam
Description : Default protocol Steam Sterilizers (120-130-140)
Results
Thermocouple set : 1080: CSD set 1, 1/2 inch feedthrough
(10 Thermocouples)
Measurement number : 2
Date and time : 20-3-2007 16:35:34
Performed by : V. Engineer
Job title : Validation Engineer
Initial : No
ITCS used : Yes
Approval for each point : Yes
Thermocouples
(The table shows recalculated digital values followed by the deviation from the intial and previous
calibrations).
Calibration temperatures in °C
Nr Code 120.0 134.0 140.0
1 0 35805.941 39976.350 41763.668
-0.2 -0.2 -0.2 -0.2 -0.2 -0.2
2 0 35776.152 39946.561 41704.091
0.0 0.0 -0.1 -0.1 0.0 0.0
3 0 35805.941 39976.350 41763.668
0.0 0.0 -0.1 -0.1 -0.1 -0.1
4 0 35776.152 39946.561 41733.880
0.0 0.0 0.0 0.0 0.0 0.0
5 0 35805.941 39976.350 41763.668
0.1 0.1 0.1 0.1 0.1 0.1
6 0 35686.786 39857.195 41644.514
0.2 0.2 -0.1 -0.1 -0.1 -0.1
7 1 0.000 0.000 0.000
--- --- --- --- --- ---
8 0 35835.730 40035.927 41823.245
-0.3 -0.3 -0.4 -0.4 0.1 0.1
9 0 35776.152 39946.561 41733.880
-0.1 -0.1 -0.1 -0.1 0.1 0.1
10 0 35746.364 39946.561 41733.880
0.0 0.0 -0.1 -0.1 0.1 0.1
The measurements in this report were performed with the Integrated Autoclave Validation System IAVS-
M2, equipped with the ValidR validation- and reporting software. The system and the software are
developed by Richmond BV, the Netherlands.
The system complies to the criteria of all applicable European guidelines regarding measurements
systems for thermometric validation of sterilization and cleaner/disinfectors. The system consists of the
following parts:
2
- An integrated Thermocouple scanner / AD converter / Amplifier TSA measurement system ([S/N
xxxxxxxxxx] ), responsible for central registration and processing of the measurement values for
temperature and pressure and also for controlling the metal block oven for the automatic calibration
of the thermocouples. For that, the measurement system contains amongst others a 16 channel
thermocouple scanner, a pressure transducer amplifier and a pt100 reference amplifier;
- An absolute pressure sensor in a temperature conditioned casing ([S/N xxxxxxxxxx]) with an
measurement tolerance not exceeding 0,7 kPA;
- An unshielded thermocouple set ([S/N xxxxxxxxxx]) with 16 (Type K) Thermocouples, mounted by
an airtight throughput with a total system measurement tolerance of 0,2°C;
- A thermocouple calibration oven (Metal block oven) ([S/N xxxxxxxxxx]) with a built in pt100
reference element with a measurement tolerance of 0,05°C;
- An Independent Temperature Control System ITCS (Type [Xxxxxxxxxxx], [S/N xxxxxxxxxx]) with a
measurement tolerance of 0,1°C;
- A portable Windows computer under Windows XP/ /Office2003 with ValidR Validation and reporting
software version 1.12 installed;
- An HP-DeskJet printer.
The ValidR validation and reporting software package has a user interface to set up detailed validation
protocols and thermocouple calibration protocols for all different combinations of process, product,
packaging and load configurations. Next to these measurement protocols text protocols can be designed
for the Installation and Operational Qualification activities as well as for other free text paragraphs of the
validation end report.
The validation part of ValidR instructs the validation engineer in how to prepare loads, the placements of
the thermocouples and the operation of the sterilizers/desinfector. This software part also controls the
measurement system according to the set up protocols; the measurement system performs the
measurements and checks the results against the applicable standards criteria, which are integrated in
the software.
The report part of ValidR collects all measurement results and the other textual paragraphs and merges
those together to a complete validation end report.
During the yearly maintenance cycle the validation system is extensively checked on correct functionality
and the system is adjusted (when necessary) within the required accuracies and always recalibrated.
The results of maintenance are logged in a logbook which is available with the system.
The pressure sensor is calibrated once per year. The calibration results (correction values) are
programmed into the validation system so measurement results are always automatically corrected.
The attachments to this report hold a copy of the applicable calibration certificate of the pressure sensor.
The ITCS temperature meter is used to verify the temperature of the metal block oven during the
calibration of the thermocouples. The calibration results (correction values) are programmed into the
validation system so measurement results are always automatically corrected.
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 43 of 45
Richmond BV UIN: 398
The attachments to this report hold a copy of the applicable calibration certificate of the ITCS
temperature meter.
Thermocouples have a type denomination in relation to their position either in the load (product) or
elsewhere in the sterilizer/desinfector chamber during validation. Validation standards demand specific
test criteria for the following different positions:
- type C Control couple In a product and with a pT100;
- type P Product couple In a product but not with a pT100;
- type H cHamber couple Not in a product but with a pT00;
- type R Room couple Not in a product and not with a pT100;
- type J Jacket couple On the jacket (heat conducting);
- type I Inactive couple Thermo couple is not used.
- With steam sterilization processes for medical devices and with washer/disinfectors the H-type
thermocouple (at the control pT100) is used for the comparison of the validation system with the
control and measurement system of the sterilizers/desinfector;
- With steam/hot-water sterilization processes for liquids (liquid sterilizers) and with hot-air
sterilizers the C-type thermocouple (at the pT100 in the product) is used for the comparison of the
validation system with the control and measurement system of the sterilizers;
- With H- and C-type thermocouples, so thermocouples placed at a pT100 sensor, the number of
the pT100 sensor concerned is noted with the couple.
The Level-, X- and Y- positions indicate the location of a thermocouple in the sterilizer/desinfector
chamber:
- The sterilizer/desinfector chamber is divided in 11 levels of height, each 5 positions wide and 7
positions deep;
- The level determines the height in the chamber: Level 1 is the bottom of the chamber, level 11 is
the top of the chamber, both actually used for positions outside the load. Levels 2 to 10 are
intermediate heights within the load with level 6 in the middle;
- The X-position determines the width in the chamber: X-position 1 is at the left (seen from the
load side to the unload side); position 3 is in the middle and position 5 is all right;
- The Y-position determines the depth in the chamber: Y-position 1 is at the front (at the load
side), position 4 is in the middle and position 7 is at the back (at the unload side);
- Examples: 1,1,1: On the bottom, left, at the front (load side);
1,5,1: On the bottom, right, at the front (load side);
1,5,7: On the bottom, right, at the back (unload side);
6,3,4: In the geometric heart of the chamber;
10,5,7: In the top layer of the load, right, at the back (unload side);
11,5,7: At the top (ceiling), right, at the back (unload side);
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 44 of 45
Richmond BV UIN: 398
“ITCS” stands for the Independent Temperature Control System. Thermocouples are calibrated at a
reference temperature generated by the metal block oven (MBO); validation standards require an
accurate independent verification of that reference temperature by means of the ITCS.
Validation report Date: 31-5-2010
ValidR Version 1.12b Page 45 of 45
Richmond BV UIN: 398
21. Attachments
This is a table of the supporting documents for this report. The documents are sorted by number of the
individual tasks and identified by the paragraph number in the report. These documents are an
indispensable essential of this report.
Task
Identification Description
Nr.
5 Leakage test Process graph charge 21
6 Bowie & Dick test Bowie & Dick sheet
Process graph charge 24
7 Unloaded 4min/134 Process graph charge 22
8 50% loaded Instruments 4min/134 Process graph charge 26
9 100% loaded Instruments 4min/134 Process graph charge 23
10 Mixed load containers/paper 4min/134 Process graph charge 27
11 Worst case load container/textile 4min/134 Process graph charge 28