EDITION APPLICABLE TO NEW CCD

REPAIR MANUAL ISSUED...............JAN. 2005


REVISED ............

ELECTRONIC DIGITAL LEVEL

DL-101C/102C

TOPCON CORPORATION

Printed in Japan
NOTICE
This document contains proprietary and confidential information of
Topcon Corporation, and its description is protected under copyright
law and international conventions.
Any unauthorized use, duplication, reproduction, modification,
disclosure, distribution or transfer of any part of this document is
strictly prohibited.

Copyright © 2005 Topcon Corporation, All rights reserved.


CONTENTS

Page
1. INTRODUCTION ........................................................................................................................... 1-1
1-1 Introduction ................................................................................................................................. 1-1
1-2 How to Use This Manual ............................................................................................................ 1-2

2. PRODUCT OUTLINE ................................................................................................................... 2-1


2-1 Nomenclature .............................................................................................................................. 2-1
2-2 Operating Keys and Functions .................................................................................................... 2-2

3. PRINCIPLE ..................................................................................................................................... 3-1


3-1 Principle ...................................................................................................................................... 3-1
3-2 Block Diagram ............................................................................................................................ 3-2
3-3 Optical Components Layout ....................................................................................................... 3-3

4. INSPECTION AND ADJUSTMENT ........................................................................................ 4-1


4-1 Inspection and Adjustment Rating .............................................................................................. 4-1
4-2 Operation Check ......................................................................................................................... 4-2
4-3 Error Code ................................................................................................................................... 4-5

5. DISASSEMBLY PROCEDURE ................................................................................................. 5-1

6. TROUBLESHOOTING FLOW CHART .................................................................................. 6-1

7. REPAIR OUTLINE ........................................................................................................................ 7-1


7-1 Digital System Repairs ............................................................................................................... 7-2
7-2 Backup of User Data ................................................................................................................... 7-3
7-3 Upload of Software ..................................................................................................................... 7-8
7-4 Setting of Model ......................................................................................................................... 7-12
7-5 Restoring of the User Data .......................................................................................................... 7-13
7-6 Setting of Parameters .................................................................................................................. 7-16

8. AUTOMATIC LEVELING SYSTEM REPAIRS .................................................................... 8-1


8-1 Compensator Mounting Adjustment ........................................................................................... 8-1
8-2 Compensator Threshold Adjustment .......................................................................................... 8-4

9. CCD READING SYSTEM REPAIRS ....................................................................................... 9-1


9-1 Nomenclature and Connection Diagram ..................................................................................... 9-1
9-2 Positions of PCB Check Pins and VRs ....................................................................................... 9-2
9-3 CCD Adjustment ......................................................................................................................... 9-3
9-4 CCD Focus Adjustment .............................................................................................................. 9-4
9-5 CCD Inclination Adjustment ...................................................................................................... 9-6
9-6 CCD Optical Axis Adjustment ................................................................................................... 9-7
9-7 CCD Offset Adjustment .............................................................................................................. 9-8
9-8 General Adjustment (1) (by Special Collimator) ........................................................................ 9-10
9-9 General Adjustment (2) (by Special Staff) ................................................................................. 9-13
9-10 Final Check ................................................................................................................................. 9-14
10. APPENDICES .............................................................................................................................. 10-1
10-1 Repair Tool List .......................................................................................................................... 10-1
10-2 Adhesive Types and Areas of Application ................................................................................. 10-2
10-3 Oil Types and Areas of Application ........................................................................................... 10-4

11. SPECIFICATIONS ...................................................................................................................... 11-1


1. INTRODUCTION
1-1 Introduction
The features of DL-101C/102C are as follows.

1) The automatic measurement can be done with a high degree of accuracy. There will be little error by the
operator in measurement. (Electronic reading by special pattern staff)

2) The measurement data and others can be recorded.

Note: The production of the CCD unit, which has been used for measuring distance and reading the
relative height so far, is discontinued on account of its maker's affairs. This repair manual is
made as a revised edition in which the parts related to the above-mentioned discontinuation are
changed to new ones. Please make a good use of this manual.
The new parts will be applied on and after the following S/N.
DL-101C: UG0051 ~
DL-102C: UH0051 ~

Some of the optional components with different specifications are not mentioned in this manual.
Please understand it.

1-1 Confidential : duplication and reproduction prohibited


1-2 How to Use This Manual
The disassembly and repair of the DL-101C/102C are executed according to the following items.

Mechanism Description Repair


Automatic leveling system Magnifies the staff’s image and focuses at the Replace the units according
reticle position. to the manual.
CCD reading system Reads the staff’s bar code pattern with the Replace the units or PCBs
CCD and displays the staff’s height. according to the manual.

Confidential : duplication and reproduction prohibited 1-2


2. PRODUCT OUTLINE
2-1 Nomenclature

5 No. Name
1 Objective lens
2 Base
3 Leveling screw
4 Horizontal tangent screw
5 Battery (BT-31Q)
Dry battery box (DB-31)
6 Hand grip
7 Telescope focusing knob
1 4 8 Measuring key
9 Serial signal connector
3 10 Circular level adjusting
screws
2
11 Operating keys
12 Telescope eyepiece
6
13 POWER key
15 14 Display

7 15 Circular level window

14
13
8
12

10 9
11

2-1 Confidential : duplication and reproduction prohibited


2-2 Operating Keys and Functions

REC SET MENU


7 8 9
SRCH IN/SO DIST
4 5 6
MANU REP
1 2 3
0 MEAS key
ESC
POWER
C ENT

Keys Name of key Function


REC Record key Records the measured data or enters the displayed data to the
instrument. The displayed data are recorded by pressing this key.
SET Set key This key initiates the set mode. The set mode is used to set measur-
ing mode, record mode and other parameters.
MENU Menu key This key initiates the menu mode. The menu mode has the following
options; Standard measuring, Formatting Ram/Card, Adjusting and
Utility modes, Adjust mode.
SRCH Search key This key allows for the searching and display of recorded data.
IN/SO Intermediate point / This key is used to initiate intermediate point or set-out mode col-
Set out mode key lection during line leveling.
DIST Distance measuring Distance data is measured and displayed when this key is pressed.
key
MANU Manual entry key When measurement with the [MEAS] key is not possible, the man-
ual key can be used to input the data from the keyboard.
Select key This key is used to page through menu screens or data display
screens.
Digit shift key If the displayed value overflows the screen this key can be used to
shift the display to the right or left in the screen.
REP Repeat measurement This key is used to remeasure the previous backsight or foresight
key point during line leveling.
ESC/C Escape/Clear key This key can be used to escape from the menu mode or any of the
setting modes. This key can be used as a backspace key when
inputting data.
0~9 Numerical key These keys are used to input numeric values.
.( ) Numeric, symbol When in alpha mode this key will change between numeric, alpha,
alpha key or symbol input mode.
-( ) Inverse staff mode This key can be used to measure with an inverse staff. Inverse staff
key mode is necessary to set “USE” within the set mode beforehand.
ENT Enter key Use this key to confirm mode parameters and to enter displayed data
values.
MEAS Start measuring key This key is used to initiate a measurement.
POWER Power key This key is used to cut the instrument ON or OFF.

Confidential : duplication and reproduction prohibited 2-2


3. PRINCIPLE
3-1 Principle
The image of the bar cord pattern of the staff used exclusively for DL in front of the object is formed on the
CCD in the position common to the reticle through the auto. level optical system, and is converted to an elec-
tric signal.
Soft processing is applied to the electric signal to read the staff and measure the distance to the staff.

3-1 Confidential : duplication and reproduction prohibited


3-2 Block Diagram
31800 6300A, 6300B, 6300D, 6300E and 6300F

CPU PCB

Compensator
unit Compensator
detection unit
Display (LCD) unit
Power switch
CPU
Ten key unit
Gate array

MEAS key
CCD
CCD Data RAM
processing
circuit
Backup battery

Card unit

Confidential : duplication and reproduction prohibited 3-2


3-3 Optical Components Layout

1 2 3 4 6 7 8 9

• Telescope

No. Name Function No. Name Function


1 Objective lens  5 Compensation mirror Automatic compensation

2 Objective lens  6 Roof prism Left/right image inversion
 Focusing
3 Internal focusing lens  7 Dichroic prism Optical axis division for CCD

4 Mirror  8 Reticle Adjustment for line of sight
9 Eyepiece lens Adjustment of acuity

3-3 Confidential : duplication and reproduction prohibited


4. INSPECTION AND ADJUSTMENT
4-1 Inspection and Adjustment Rating
Item Inspection Method Rating Adjustment Method
Circular level Center the air bubble using the leveling ±1' Make one half of the
mounting accu- screws, rotate the instrument in 180° on its adjustment using the
racy vertical axis and check the amount of bub- adjusting screws and the
ble movement. other half using the level-
ing screws. (Refer to the
instruction manual.)
Stadia spacing When aligning the stadia wires of the colli- 0.4% max. Adjust by shifting the
mator with one of the level stadia wires (7") objective lens back and
(top or bottom), check the difference of the forth.
other end.
Compensation Level the collimator set horizontally and DL-101C: Adjust by sliding the com-
accuracy check the leveling of the level. Next, check 3" max. pensator back and forth.
the leveling when the instrument is tilted by DL-102C: (Refer to page 8-1 “AUTO-
80% (DL-101C: 9' 36" / 102C: 12') of the 5" max. MATIC LEVELING SYS-
indicated compensation range. The differ- TEM REPAIRS” about the
ence will be the compensation accuracy. details.)
Line of sight Center the air bubbles using the leveling ±2" max. Adjust the eyepiece reticle.
leveling screws.
Sight the horizontal collimator and check
the difference from the reticles of the tele-
scope.
Circular scale Place a level between the two horizontally ±15' max. Replace the circular scale.
eccentricity mounted collimators, rotate the level in
180° in reference to 0° and 90° and check
the scale error.
Compensator Mount the instrument on the inclination DL-101C: Refer to page 8-4 “Com-
division stand and level it. Incline the instrument in ±12' ±1' pensator Threshold Adjust-
accuracy +14' ~ +16' and -14' ~ -16' and press the DL-102C: ment”.
[MEAS] key. Check if “Cmpe Err” is dis- ±15' ±1'
played. Press the [ESC] key to cancel the
error.
CCD focus and Check the focus and optical axis, referring Refer to page 9-14 “Final
optical axis to the final check items. Check”.
check
Measurement Using the collimator exclusively for DL, Height:
accuracy check level the instrument and sight the collima- 1.500m
tor. When the horizontality is within ±2", Distance:
sight the pattern, carry out measurement The value
and check the height. shown in the
collimator
Resolution Adhere the 3" resolution chart 10m before 3" Make adjustment by turn-
check and collimate it. ing the objective lens unit.
Make sure that 4 patterns can be recog-
nized.

4-1 Confidential : duplication and reproduction prohibited


4-2 Operation Check
Procedure Key Display Probable
Operation Problem
1 Turn on the power switch. POWER
TOPCON
DL-102C

Menu
Meas

2 Battery check SET


SET
Set
(1) Press the 8 key and select the 8
title of “Check Battery” with the Mode
2 and keys.
2

Check
Battery

(2) Press the ENT key.


ENT Battery
The voltage is displayed for N sec-
onds and then the title of “Check 7.92V
Battery” is displayed again.
* The display time is changed by the
instrument setting.
(Refer to the instruction manual.) Check
Battery

ESC
(3) Press the C key. ESC
Menu
C
Meas

Confidential : duplication and reproduction prohibited 4-2


Procedure Key Display Probable
Operation Problem
3 Key operation check SET
SET Set
(1) Press the 8 key and select the 8
title of “Check Time” with the Mode
2 and keys.
2

Check
Time

(2) Press the ENT key.


ENT Date
02: 05: 94

ESC
(3) Press the C key. ESC
Date ?
C

(4) Press the ten keys and make sure Any key you like.
that the numerals whose keys have
Date?
been pressed are displayed. 010203

(5) Press the ENT key.


ENT Date
01/02/03

Time
04:05:06

ESC
(6) Press the C key. ESC
Time?
C

4-3 Confidential : duplication and reproduction prohibited


Procedure Key Display Probable
Operation Problem
(7) Press the ten keys and make sure Any key you like.
that the numerals whose keys have
Time?
been pressed are displayed. 012345

ESC
(8) Press the C key several times ESC
Menu
and return the display to “Menu C
Meas”. Meas

4 Adjustment and check of liquid crystal SET Set


brightness 8
(1) Press the
SET
key and enter the Mode
8
“Set Mode”. Select the title of “Set
2
Bright” with the 2 and
keys.
Set
Bright

(2) Make sure with the 2 and


Set
keys that the numerals move 2
up and down and the brightness Bright 3
changes.

ESC
(3) Press the C key several times ESC
Menu
and return the display to “Menu C
Meas”. Meas

Confidential : duplication and reproduction prohibited 4-4


4-3 Error Code
Dispaly Contents Countermeasure Ref.
page
Already Exists Same job or group name already Enter a new job name or group
------
exists. name.
Cannot Cont The job going to be continued Press [ESC] to return to menu.
caused GH error or illegal format ------
data.
Cannot Copy Illegal copy (from card to card) is Confirm the data.
carried out. When card to RAM
------
copy is carried out, there is illegal
data or format error within a card.
Cannot Make Card capacity is full. Use other card. ------
Cannot Read Data The coordinate data cannot be read. Confirm the files in the card. ------
Card over 90% 90 percent of the card memory area Press [ESC]. Delete unnecessary
------
is used by data. job within the card.
Card Full Card memory is full. Use other card. ------
Card is Broken! Management block of card mem- Confirm the card.
------
ory is broken.
Card Changed When the power is ON, Data card is Confirm the card. If you want to
------
changed with other one. continue the work, press [ESC].
Card Not Format Data card is not formatted. Format the card. ------
Card Not Ready Data card is not inserted. Insert data card. ------
Compe Err The inclination of the instrument Press [ESC]. Level the instrument P. 6-3
exceeds the capacity of the compen- properly.
sator.
Coll Error The adjustment value has exceeded Press [ESC]. Start the adjustment
±10mm during the adjustment pro- procedure from the beginning. ------
cess.
Coordi Not Found There is not coordinate data file Use the card including the coordi-
------
appointed. nate data.
Dark Err The level cannot read the rod due to Press [ESC]. The data will have to
lack of light. be entered manually until lighting ------
has improved.
E61 EEPROM writing error Start the adjustment procedure from P. 6-6
the beginning. If the error message
is still displayed, replace the CPU
PCB.

4-5 Confidential : duplication and reproduction prohibited


Dispaly Contents Countermeasure Ref.
page
E70's Measuring error Press [ESC], and remeasure. P. 6-2
• Level is not properly sighted on • Collimate the staff properly.
the staff.
• The distance between the instru- • The distance between the instru-
ment and the staff is either too far ment and the staff should be
or near. within the range of 2m to 100m, 7'
to 328'. (Fiberglass staff)
• The staff is obstructed by 30 per- • Remove the obstructions.
cent or more.
• The cross-hairs are not within the • The staff should be within the
range of the staff. range of the cross-hairs.
E88 Any abnormality occurs within the Turn the power switch off, then on
------
CPU. again.
E90 ~ 96 Communication error Press [ESC]. Check the parameters P. 6-5
of the level and/or external devices.
E98 The battery level of the internal Press [ESC]. Contact your dealer to P. 6-4
backup battery is poor. replace the built-in battery.
Gh Error The calculated elevation exceeds Press [ESC] to return to the mea-
------
the level’s proper range. surement screen.
Group Nothing There is no group appointed in Data Confirm the group or the card, and
------
Card. set again.
Illegal Card Card more than 2Mbytes is used. Capacity of a card must be 2Mbytes
------
or less.
Illegal Format A format of data card is different. Format the card. ------
Input Error Invalid data was input. Reinput the corrected data. ------
Job Not Found In a card, there is no job appointed. Confirm the job, and set again. ------
Job Over 256 jobs exist within RAM or a Press [ESC]. Delete the job in
group. RAM or group after saving the job ------
data.
Light Err The level cannot read the rod due to Press [ESC]. Interrupt the sunlight
an excess of light such as direct or glare. ------
sunlight or glare.
Memory full The data memory area of the instru- Press [ESC]. Delete unnecessary
------
ment is full. job within RAM.
Memory 90 percent of the data memory area Press [ESC]. Delete unnecessary
------
over 90% is used by data. job within RAM.
Must do End The menu option just selected is not Finish the leveling operation and
mode available during leveling in the return to menu mode before
------
instrument. attempting to execute the function
again.
No Job To Cont There is no job to continue. Press [ESC] to return to menu. ------
Opr. Err Method type A and B of adjusting Press [ESC]. Retry the adjusting in
mode: The collimated direction is right direction. ------
opposite.

Confidential : duplication and reproduction prohibited 4-6


Dispaly Contents Countermeasure Ref.
page
Read Error By abnormality of a card, a reading Change the card for other card.
is not completed. An error occurred ------
in cluster read.
Same Group A file is copied into same group. Copy a file into other group. ------
Setting Error Method type A of adjusting mode: Press [ESC]. Set the instrument
The horizontal distance between Aa half way between points a and b
and Ab exceeds 1 meter. within 1 meter.
The horizontal distance Ba is too Set the instrument approximately
near or far from staff (a). 3m from the staff (a). ------
Method type B of adjusting mode: Press [ESC]. Set the instrument at
Either position A or B is not set up correct position divided into 1/3
correctly on 1/3 of the distance between staff (a) and (b).
between staff (a) and (b).
Write Error Data cannot be written into data Check the data card. If necessary,
card. replace the card with a new one.
An error occurred in cluster writing. ------
There is not space cluster in file
generation.
Write Protect! Write protection is applied to data Cancel write protection.
------
card.

• If errors still persist after attempting to clear them, contact your local TOPCON dealer.

4-7 Confidential : duplication and reproduction prohibited


5. DISASSEMBLY PROCEDURE
: Indicates the direction of moving or removing.
: Indicates a part that should be removed for disassembly.

Procedure Illustration

Remove the six setscrews of the display panel.

Remove the lid.


Cover disassembly

Remove the setscrew of the focusing knob.

5-1 Confidential : duplication and reproduction prohibited


Procedure Illustration

Remove the focusing knob.

Remove the eight setscrews from the instrument


cover.
Cover disassembly

Remove the power supply connector.

Confidential : duplication and reproduction prohibited 5-2


Procedure Illustration

Remove the MEAS key connector.

Remove the ten key unit connector of the display


panel.
Cover disassembly

Remove the display unit connector of the display


panel.

5-3 Confidential : duplication and reproduction prohibited


Procedure Illustration

Remove the display panel.


Cover disassembly

Remove the four setscrews of the main PCB.


PCB removal

Remove the two compensator connectors.


(Refer to P. 9-2 for the connector position.)

Confidential : duplication and reproduction prohibited 5-4


Procedure Illustration

Remove the serial signal connector for outside.

Remove the CCD connector.


PCB removal

Remove the three setscrews of the card unit.

5-5 Confidential : duplication and reproduction prohibited


Procedure Illustration

Remove the two mounting screws of the com-


pensator unit.
Compensator unit removal

Remove the objective lens hood.


Objective lens disassembly

Loosen the objective lens retainer ring with the


pin face wrench (ST-DL-03).

Confidential : duplication and reproduction prohibited 5-6


Procedure Illustration

Remove the objective lens retainer ring.

Pull out the objective lens.


Objective lens disassembly

5-7 Confidential : duplication and reproduction prohibited


Procedure Illustration

Loosen the setscrew.


Eyepiece unit removal

Turn and remove the eyepiece unit.

Remove the setscrew on the bottom of the CCD


Ass’y.
CCD unit disassembly

Confidential : duplication and reproduction prohibited 5-8


Procedure Illustration

Turn and remove the CCD inclination adjust-


ment nut.

Remove the CCD Ass’y.


CCD unit disassembly

Turn and remove the optical axis adjusting


screw.

5-9 Confidential : duplication and reproduction prohibited


Procedure Illustration

Remove the two mounting screws for CCD base.

Remove the CCD base.


CCD unit disassembly

Confidential : duplication and reproduction prohibited 5-10


Procedure Illustration

Remove the three leveling screws.


Baseplate disassembly

Turn the shaft retaining nut.

5-11 Confidential : duplication and reproduction prohibited


Procedure Illustration

Remove the shaft retaining nut.

Remove the retaining nut.


Baseplate disassembly

Confidential : duplication and reproduction prohibited 5-12


Procedure Illustration

Remove the lens barrel.


Baseplate disassembly

5-13 Confidential : duplication and reproduction prohibited


6. TROUBLESHOOTING FLOW CHART
The following sample flow chart is for identifying the faulty unit.

Example:

Problem

NO
Diagnosis

YES

NO NO
Diagnosis Diagnosis

YES YES

* Item to be followed.

Identify and replace faulty unit

6-1 Confidential : duplication and reproduction prohibited


(1) Measurement cannot be done. (Err70 and subsequent Nos. of the level)

Measurement is impossible.

NO Align the level and YES


Are the level and Normal OK
focus correct? focus.

NO
YES

NO Connect the YES


Is the CCD connector Normal OK
disconnected? connector correctly.

NO
YES

NO YES
Is the CCD signal Replace the CCD
Normal OK
outputted? unit.

NO
YES

NO YES
Are the CCD focus and Adjust the CCD.
optical axis correct? Normal OK

NO
YES

Replace the PCB.

Confidential : duplication and reproduction prohibited 6-2


(2) Compensator error

Compensator error

NO YES
Is the instrument leveled? Level the instrument. Normal OK

NO
YES

Is the connector NO Connect the YES


Normal OK
disconnected? connector correctly.

NO
YES

NO YES
Is the PCB adjustment OK? Can adjustment
be done?

NO
YES

Replace the PCB.


Replace the PCB.

Can adjustment YES


be done?

NO

Replace the compensator. Adjust the


compensator threshold.

6-3 Confidential : duplication and reproduction prohibited


(3) Power is not turned on.

Power is not turned on.

Is the battery NO
terminal voltage 7.2V? Recharge.
(6.2V or more)

YES

After the display (LCD) NO Is the terminal voltage NO


unit is replaced, is it OK? 7.2V or more?

YES YES

OK Replace the main PCB. OK Repair the internal


power supply.

Confidential : duplication and reproduction prohibited 6-4


(4) Communication error (E90~96)

E90~96

Is the setting of each NO YES


Correct the setting. Normal OK
apparatus correct?

NO
YES

After the YES


cable is replaced, OK
is it OK?

NO

After the serial


YES
cable of the instrument is OK
replaced, is it OK?

NO

Replace the PCB.

6-5 Confidential : duplication and reproduction prohibited


(5) Keys are not effective.

Keys are not effective.

Is the connector NO Connect the YES


connector correctly. Normal OK
disconnected?

YES NO

YES
After the key unit is
OK
replaced, is it OK?

NO

Replace the PCB.

(6) Display is abnormal.

Display is abnormal.

Is the connector NO Connect the YES


disconnected? connector correctly. Normal OK

NO
YES

After the display unit YES


is replaced, is it OK? OK

NO

Replace the PCB.

(7) E61

E61

* When replacing the PCB, carry out the general adjustment and
compensator threshold adjustment.
Replace the PCB.

Confidential : duplication and reproduction prohibited 6-6


7. REPAIR OUTLINE
Repairs for the DL-101C/102C are divided into the following three intems.

• AUTOMATIC LEVELING SYSTEM REPAIRS


• CCD READING SYSTEM REPAIRS
• PCB REPAIRS

Each system is repaired by replacing the units for each item. When replacing a unit, the adjustment method
depends on the unit. The chapter for each system has a flow chart, so adjust according to the flow chart.
After repairs are complete, inspect and adjust according to the flow chart below.

8. AUTOMATIC LEVELING 9. CCD READING 7-1 Digital System Repairs


SYSTEM REPAIRS SYSTEM REPAIRS

9-8 General Adjustment (P. 9-10)

4-2 Operation Check

9-10 Final Check (P. 9-14)

Inspection is completed.

7-1 Confidential : duplication and reproduction prohibited


7-1 Digital System Repairs
Repair the digital system of DL-101C/102C according to the flow chart below.

Angular measurement digital


system repairs

Replacement of display unit Replacement of card Replacement of main


PCB unit PCB unit

Replacement of ten key unit Replacement of power


supply unit

CCD offset adjustment for


main PCB unit (P. 9-8)

To "General Adjustment" and


after in the flow chart (P. 7-1)

Confidential : duplication and reproduction prohibited 7-2


7-2 Backup of User Data
Tools to be used: PC card, personal computer (PC) and F4 cable

Procedure Key Display


Operation
1 Preparation
If the leveling measurement is in process,
end the leveling mode as shown below.
MANU
1. Press the 1 key. MANU
1

2. Input the height and distance directly. Optional numeral keys

MENU
3. Press the 9 key and select "Close MENU
Leveling" to end the leveling mode. 9
MENU
Note: If you press the 9 key and the
"Close Leveling" selection display does
not appear, the leveling measurement is
MANU
in process. The input with the 1
key is necessary.

2 Format of the PC card

1. Insert the PC card into the DL instrument.


MENU
2. Press the 9 key. MENU
Menu
9
Measure

3. Select "Menu Format" with the 2 and


2 Menu
keys.
Format
ENT

4. Input the password with the 7 , 5 , 7 5 3 Set !


3 and ENT keys.
ENT Password

7-3 Confidential : duplication and reproduction prohibited


Procedure Key Display
Operation

5. Select "Format Card" with the 2 and


keys. 2 Format
* Don't select "Format RAM". The data in Card
the DL instrument are erased.
Format
Ram

Format
ENT
Card ?

6. The volume label is displayed. Press the ENT V Label


ENT key.
xxxxxxxxxx
* Unless the volume label is set, it is not dis-
played.

7. The question whether the battery has been ENT Changed


replaced with a new one is displayed.
Battery ?
Press the ENT keys.

8. The next date for replacing the battery ENT Date ?


with a new one is displayed. Press the or
xxxxxxxxxx
ENT key. ESC
* "After 3 years" as default is displayed for C + Optional keys
the battery replacement date. If you set
the replacement date optionally, press the
ESC
C key and input the data.

9. Formatting is performed.
Format
>>

Menu
Format

Confidential : duplication and reproduction prohibited 7-4


Procedure Key Display
Operation
3 Making of the PC card directory

1. After formatting, select "Menu Utility"


2 Menu
and decide it with the ENT key.
Utility
ENT

2. Select "Make Group" and decide it with


2 Make
the ENT key.
Group
ENT

3. Make a directory. After making, decide it Optional keys + ENT Group


with the ENT key.
xxxxxxxxxx

Menu
Utility

4 Copying data into the PC card


MENU
1. Press the 9 key. MENU
Menu
9
Utility

2. Select "Copy Job" with the 2 and


2 Copy
keys and decide it with the ENT
Job
key. ENT

3. Select "S Group RAM" with the 2


2 S Group
and keys and decide it with the
RAM
ENT key. (Selection of copy source) ENT

4. Select "Job Type (All)" (Copy all the files)


2 Job Type
with the 2 and keys and decide
(All)
it with the ENT key. ENT

7-5 Confidential : duplication and reproduction prohibited


Procedure Key Display
Operation
5. Select the name of the directory made as
mentioned above for the PC card and 2 D Group
decide it with the ENT key. (Selection of xxxxxxxxxx
ENT
copy destination)
Copy
Now

6. Copying is completed and "Copy Job" is


Copy
displayed.
OK !

Copy
Job

5 Checking the data in the PC card


SET
1. Press the 8 key. SET
Set
8
Measure

2. Select "Out Module" with the 2 and


2 Out
keys and decide it with the ENT
Module
key. ENT

3. Select "Module Card" with the 2 and


2 Module
keys and decide it with the ENT
Card
key. ENT

4. Select the directory made as mentioned ENT


Group
above for the PC card and decide it with
the ENT key. xxxxx

ESC
5. Press the C key. ESC
Out
C
Module

Menu
Utility

Confidential : duplication and reproduction prohibited 7-6


Procedure Key Display
Operation
SRCH
6. Press the 4 key and then press the SRCH
Srch
ENT key. 4 ENT
Job No

Job Type
Measure

7. The data copied in the PC card are classi-


2 Job Type
fied according to categories. Check the
xxxxxxxxxx
data. ENT
Job Type Measure : Measured data
Adjust : Distance data
Level : Horizontality data
6 Write-protect
Apply write-protect to the PC card taken out WP
of the DL instrument to prevent the data
from being rewritten. Change the switch to "WP"

7-7 Confidential : duplication and reproduction prohibited


7-3 Upload of Software
7-3-1 Upload of software by using the card

Procedure Key Display


Operation
1 Prepare the SRAM card (512K) in which the
DL-101C/102C software is recorded.
2 Turn on the power as pressing the ten keys, 9 6
9 , 6 , 1 and 0 on the dis- TOPCON
1 0
play frame. DL-101C

POWER

3 Press the ENT key. ENT Program


Load ?

4 The instrument requests the password. Press 4 7 2


the keys, 4 , 7 , 2 and ENT in Password
this order.
ENT

Password
Bingo !

5 The device selection display appears. Select


"CARD" with the 2 and keys. 2 Device
Card

6 Press the ENT key. ENT


Flash
Erase

Road
> Write

Road
>>> Write

Confidential : duplication and reproduction prohibited 7-8


Procedure Key Display
Operation

7 Check sum is displayed.


Sum
7A06

8 Upload is finished.
TOPCON
Note: Just after the download of the software,
DL-101C
nothing is displayed for about 5 sec-
onds even if turning on the power.
Don't operate the keys during this
period. If you have operated the keys,
perform "7-5 Initialization".

7-9 Confidential : duplication and reproduction prohibited


7-3-2 Upload of software by using RS-232C
Prepare the DL-101C/102C instrument, the interface cable (F4 cable or F9 cable) and a personal computer
(PC) and connect them as shown below.

DL-101C/102C F4 cable PC

Procedure Key Display


Operation
1 Turn on the power as pressing the ten keys, 9 6
9 , 6 , 1
TOPCON
and 0 on the dis-
1 0 DL-101C
play frame. +
POWER

2 Press the ENT key. ENT Program


Load ?

3 The instrument requests the password. Press 4 7 2 Password


the keys, 4 , 7 , 2 and ENT in
ENT
this order.

Password
Bingo !

4 The device selection display appears. Select


2 Device
"RS-232C" with the 2 and keys.
RS-232C
ENT

Set Board
38400

5 Selection of communication speed

1. In the case of data communication 38400


2 ENT RS-232C
[bps]
38400 ?
The device selection display appears.
Select "38400" with the 2 key and
decide it with the ENT key.

Confidential : duplication and reproduction prohibited 7-10


Procedure Key Display
Operation
2. In the case of data communication 9600
[bps] ENT RS-232C
The device selection display appears. 9600 ?
Select "9600" with the key and
decide it with the ENT key.

6 Press the ENT key. ENT Flash


Erase

Road
> Write

Road
>>> Write

7 Check sum is displayed.


Sum
7A06

8 Upload is finished.
TOPCON
Note: Just after the download of the software,
DL-101C
nothing is displayed for about 5 sec-
onds even if turning on the power.
Don't operate the keys during this
period.

7-11 Confidential : duplication and reproduction prohibited


7-4 Setting of Model

Procedure Key Display


Operation
When setting the model, DL-101C and DL-102C
are changed each other by the procedure below.
1 Turn on the power as pressing the 7 , 7 9 TOPCON
9 and keys.
+ DL-101C
POWER

2 If the display shown at the right appears and 7 9 SET


the power is automatically turned off, initial-
+ DL-102C
ization is finished and the model is changed.
POWER

TOPCON
DL-102C

3 If the model is not changed to a desirable


SET
one, perform the model change again.
DL-101C

Confidential : duplication and reproduction prohibited 7-12


7-5 Restoring of the User Data

Procedure Key Display


Operation
(1) When copying the data of the PC card into
the RAM area
1 Format of the RAM area
MENU
1. Press the 9 key. MENU
Menu
9
Measure

2. Select "Menu Format" with the 2 and


2 Menu
keys.
Format
ENT

3. The instrument requests the password. 7 5 3 Set !


Press the keys, 7 , 5 , 3 and
ENT Password
ENT in this order.

4. Select "Format Ram" with the 2 and


2 Format
keys and decide it by pressing the
Ram
ENT key. ENT

5. Press the ENT key, and formatting is per- ENT Format


formed.
Ram ?

Format
>>

6. The menu display appears again.


Menu
Format

2 Copying the data into the RAM area


MENU
1. Press the 9 key. MENU
Menu
9
Utility

2. Select "Copy Job" with the 2 and


2 Copy
keys and decide it with the ENT
Job
key. ENT

7-13 Confidential : duplication and reproduction prohibited


Procedure Key Display
Operation

3. Select "S Group xxxxx" (directory of PC


2 S Group
card) with the 2 and keys and
xxxxx
decide it with the ENT key. (Selection of ENT
copy source)

4. Select "Job Type (All)" (Copy all the files)


2 Job Type
with the 2 and keys and decide
(All)
it with the ENT key. ENT

5. Select "D Group RAM" with the 2


2 D Group
and keys and decide it with the
RAM
ENT key. ENT

Copy
Now

Copy
OK !

Copy
Job

3 Checking the data of the RAM area


SET
1. Press the 8 key. SET
Set
8
Measure

2. Select "Out Module" with the 2 and


2 Out
keys and decide it with the ENT
Module
key. ENT

3. Select "Module Ram" with the 2 and


2 Module
keys and decide it with the ENT
Ram
key. ENT
ESC
4. Press the C key. ESC
Out
C
Module

SRCH
5. Press the 4 key. SRCH
Copy
4
Job

Confidential : duplication and reproduction prohibited 7-14


Procedure Key Display
Operation

6. Select "Srch Job No" with the 2 and


keys and press the ENT key. 2 Srch
Job No
ENT

Job No ?

7. Input the job name and check it.


Job No ?
xxxxx

8. Check all the file names. Checking the


data is finished.

7-15 Confidential : duplication and reproduction prohibited


7-6 Setting of Parameters
Set the parameters about measurement and display by using the parameter setting mode (SET MODE).
The set parameters when shipped from the factory are shown by thick characters in the "Selecting Item" col-
umn in the table below.

Menu Selecting Item Contents


Check Battery
Set Measure N Time Measurements taken N times
N is between 2 and 99
Single Single measurement mode
Cont Continuous measurement mode
Set Fix Standard DL-101C: 0.1mm / DL-102C: 1mm
Precise DL-101C: 0.01mm / DL-102C: 0.1mm
Set Item Standard Do not display extended data
Extended Display extended data
Display Time 3 Set the duration in seconds, 1~9 secs., that the data screens
will be displayed.
Display Unit m Measuring unit : m (meter)
ft Measuring unit : US. ft (feet) 1m=3.280833333333333ft
Out Module Ram Measured data is stored to internal memory
Card PCMCIA SRam Card
RS-232C Communication with an external data collector is enabled
Off Measured data is not stored
Point Number Incremnt
Decremnt
File Out
Set Comm Comm Standard setting
Std (1200 baud, 7 word length, 1 stop bit, even parity)
Comm Set 300/600/ Baud rate 300/600/1200/2400/4800/9600
Manu Baud 1200/
2400/
4800/
9600
Set Even/ Set Parity
Parity Odd/ Even / Odd / None
None
Set Off/On If term is set to ON all data strings sent to the computer
Term will be terminated with the carriage return/line feeds char-
acters. If term is OFF then CR/LF characters will not be
appended to the data string.
Auto Cutoff 5 min Auto cutoff mode ON
Off Auto cutoff mode OFF
Set Bright 3 Set the brightness to one of nine steps
Set Light On/Off
Check Time
Inverse Mode Not Use Not use the Inverse Mode
Use Use the Inverse Mode
Swing Correct On/Off

Confidential : duplication and reproduction prohibited 7-16


8. AUTOMATIC LEVELING SYSTEM REPAIRS
8-1 Compensator Mounting Adjustment
Procedure Illustration

Inspect the horizontality of the collimator used Illuminate the


exclusively for the electronic level. eyepiece of the
collimator.
1. Illuminate the eyepiece of the collimator.
Fit the optical
2. Put the instrument on the inclination stand. axis height to
each other.
By leveling the inclination stand, fit the opti-
Preparation

cal axis height of the collimator to that of the


DL objective lens.
3. Collimate the reticle lines (cross-hairs) with-
out mounting the compensator. Make adjust-
ment with the reticle so that the collimator's
sighting line may align with the reticle lines
(cross-hairs) of the DL instrument. Rotation
1. Put the compensator into the instrument until
Compensator mounting

it contacts the right side and temporarily


tighten the mounting screws.
2. If the line of sight is off by about 4’, move the Align them
with each other.
reticle up and down to align it with the colli- Reticle lines
mator's sighting line. Collimator's
sighting line

1. Incline the inclination stand in an elevation


angle from the horizontal status.
Adjustment of forward/backward correction residual

The amount of inclination is different


between the models.
DL-101C: 14' ±3"
DL-102C: 16' ±5"

2. Make sure that the distance between the hori-


zontal lines of reticle and collimator does not
exceed the standard while the instrument is
being inclined.
The standard is different between the models.
DL-101C: Within ±3"
DL-102C: Within ±5"
If it is out of the standard, move the compensa-
tor forward and backward to make adjustment.

8-1 Confidential : duplication and reproduction prohibited


Procedure Illustration

3. Incline the inclination stand in a depression


Adjustment of forward/backward

angle in the same way. Make sure that the dis-


tance between the horizontal lines of reticle
and collimator does not exceed the standard.
The amount of inclination and the standard
are the same as Procedure 1 and 2.
correction residual

4. Set the instrument in the Procedure 1 status


again. Make sure that the correction residual
is within the standard.

5. Level the inclination stand.

1. Incline the inclination stand in the right-and-


left direction.
The amount of inclination is different
between the models.
DL-101C: +14' ±3", -14' ±3"
DL-102C: +16' ±5", -16' ±5"
Adjustment of right/left correction residual

2. Make sure that the distance between the hori-


zontal lines of reticle and collimator does not
exceed the standard.
The standard is different between the models.
DL-101C: Within ±3"
DL-102C: Within ±5"
If it is out of the standard, loosen one com-
pensator mounting screw a little and turn the
compensator to make adjustment.

3. Set the instrument in the Procedure 1 status Setscrew Setscrew


again. Make sure that the correction residual Keep it tightened. Loosen it a little.
is within the standard.

4. Tighten the screw with torque of 5kgf·cm and


apply the screw lock (red).

Confidential : duplication and reproduction prohibited 8-2


Procedure Illustration

1. As shown in the illustration, grip the eyepiece


Reticle adjustment
with the tool and adjust the reticle inclinaiton. screw
The mounting screws should be loosened
lightly beforehand.
Standard: Within 2"
Note: If the mounting screws are loosened
excessively, the reticle will move left-
ward/rightward and the CCD opti-
cal axis will be off. Be careful.

Special tool
Reticle inclination adjustment

8-3 Confidential : duplication and reproduction prohibited


8-2 Compensator Threshold Adjustment
Procedure Illustration

1. Preparation Indication of display

(1) Mount the main PCB and others on the TOPCON


instrument. Connect all the cables except the DL-102C
MEAS key (J1).

(2) Connect the minus and plus sides of the digi-


tal multi-meter to TP1 and TP2 respectively.
VerX.XX
Initial
(3) Supply 7.2V from the constant-voltage power
supply.

(4) Level the inclination stand and instrument.


CCD adj
REC
Test CCD
2. Turn on the power while pressing the 7
Compensator threshold adjustment

ESC
and C key.

3. Make sure that the inclination stand and the


bubble level of the instrument are not tilted
(0') and adjust the voltage of TP9 with VR1 to
within the standard.
Standard: 2.0 ±0.1V

4. Adjust the voltage of TP10 with VR2 to


within the standard.
Standard: 2.0 ±0.1V

Confidential : duplication and reproduction prohibited 8-4


9. CCD READING SYSTEM REPAIRS
9-1 Nomenclature and Connection Diagram
Repairing is done by replacing each unit.

Measuring key

Compensator unit

CCD unit

J4
J9 J3

J5 J1 J2

J6
J8 Serial signal
J7 connector for
outside

Power supply
Numerical key unit

Display unit

9-1 Confidential : duplication and reproduction prohibited


9-2 Positions of PCB Check Pins and VRs
PCB number: 31800 6300A/B/D/E/F

J5
VR2

TP10 J9
VR1 J4
TP9

J3

J6 J1

TP1 TP19
J7
TP2
J2

VR3 TP13 J8

Check pin Function Ref. VR Function Ref.


page page
TP1 GND VR1 Compensator threshold voltage 1
TP2 Battery voltage VR2 Compensator threshold voltage 2
TP9 Compensator threshold voltage 1 VR3 CCD offset voltage
TP10 Compensator threshold voltage 2
TP13 Trigger signal
TP19 CCD signal

Confidential : duplication and reproduction prohibited 9-2


9-3 CCD Adjustment
Procedure Illustration

1 Preparation
Fluorescent lamp
When you perform the CCD adjustment, carry out
the adjustment of focus, inclination and optical
axis. If one of these three is adjusted, others may
be incorrect.

(1) Inspect the horizontality of the collimator used


exclusively for the electronic level. (Use a flu-
orescent lamp as the collimator's illuminator.) Collimator used exclusively for the
electronic level

(2) Put the instrument on the inclination stand and


level it.

(3) Align the eyepiece acuity.

(4) Check the horizontality.

2 Connection of power supply and synchroscope Don't tighten


(1) Connect the main PCB to the card unit, and the screws.
connect the compensator connector and CCD
connector to the main PCB.
* After connecting the cable, move the main
PCB toward the objective side for CCD
adjustment.

(2) Connect both of the key connector from the


display panel and the connector of the display
unit to the main PCB, and set them as illus-
trated in right.
* Put cardboard between the instrument and
PCB for insulation.

(3) Connect CH1, CH2 and GND to TP13, TP19


and TP1 respectively.
* CH1 is a trigger signal. CH2 is a CCD signal.

(4) Connect the power cable (ST-GS-05) to the


power connector of the main PCB (J7).

(5) The battery supplies 7.2V.

9-3 Confidential : duplication and reproduction prohibited


9-4 CCD Focus Adjustment
Procedure Illustration

1. Set the synchroscope as shown below.


CH1: Vertical axis 5V
CH2: Vertical axis 1V
Horizontal axis: 40µs
Trigger: 472, 800µs
Display of the display unit

2. Turn on the power while pressing the REC


7 TOPCON
ESC Buzzer sound
and C keys. DL-102C
REC ESC
Note: Keep pressing the 7 and C
keys till the buzzer sounds again after
it sounds once first.
VerX.XX
Buzzer sound
Initial

CCD adj
3. Short-circuit the connector J1 pins 1 and 2 of
CCD focus adjustment

the MEAS key. Test CCD

Chg Tim N
XXX

When the 20m pattern is not collimated


4. Collimate the 20m pattern and turn the fine
adjustment screw of the instrument till the
waveform appears on the synchroscope.

When the 20m pattern is collimated

20m pattern
(Narrower bar code)

Confidential : duplication and reproduction prohibited 9-4


Procedure Illustration

5. Adjustment

(1) Adjust the focus by turning the focusing ring


so that the 20m pattern can be seen clearly.
Change the setting of synchroscope CH2 to Focus is not
“Horizontal axis: 10µs”. adjusted yet. R
pattern is short.
R R
(2) Adjust the focus adjustment screw till the R
pattern is largest while watching the wave-
form on the synchroscope.

Note: Loosen the focus fixation screw lightly


before adjustment.

(3) Tighten the fixation screw and make sure


that the focus of the eyepiece is fit to that of Focus is already
adjusted. R pat-
the CCD. If it is not so, adjust the focus tern is long.
again.
CCD focus adjustment

R R

R R

Focus fixation screw


Focus adjustment screw

9-5 Confidential : duplication and reproduction prohibited


9-5 CCD Inclination Adjustment
Procedure Illustration

1. Move the 20m pattern from the left to the When the 20m pattern is not collimated
right slowly and adjust so that the signals may
be started simultaneously on both of the left/
right sides.
When the 20m pattern edge is collimated

When the 20m pattern edge is collimated


(CCD inclination is not adjusted.)
2. Turn the inclination adjustment nut so that the The signal on
signals may be started simultaneously on both the left side is
of the left/right sides. started first in
this illustra-
tion.
When the signal on the right side is started
CCD inclination adjustment

first:
Tighten the inclination adjustment nut. Loosen the in-
When the signal on the left side is started clination ad-
first: justment nut.
Loosen the inclination adjustment nut.
When the 20m pattern edge is collimated
(CCD inclination adjustment is completed.)
The right and
left of the
waveform are
started simul-
taneously.

Inclination
adjustment nut

Confidential : duplication and reproduction prohibited 9-6


9-6 CCD Optical Axis Adjustment
Procedure Illustration

1. Swing the instrument from the left to the right


while the 20m pattern is the center to find the 20m pattern
signal starting position as shown in the right.

2. Fit the reticle of the instrument to the left end


of the 20m pattern.
Fit the retic lines to the
3. Loosen the two screws of the CCD base left end of 20m pattern.
mounting unit lightly.

4. Turn the optical axis adjustment screw and CCD start


waveform
stop it where the signal is started.
• When the signal is not started at the pat-
tern's left end and, in the right direction, the
waveform as shown in the right is dis-
played:
Tighten the optical axis adjustment screw.
• When the signal is started at the far left side
CCD optical axis adjustment

from the pattern’s left end:


Loosen the optical axis adjustment screw. Optical axis
adjustment screw
5. Tighten the two screws of the CCD base
mounting unit with torque of 5kgf⋅cm. Mounting
screw
6. Move the reticle to the right end of the 20m
pattern and make sure that the CCD output
waveform is started as shown above.
Standard: The difference of start between
right and left should be within ±15"
from the pattern center.
If the difference is not within the standard, adjust
again from Procedure 1.

7. Check the CCD focus adjustment and the


inclination adjustment and make sure that the
focus, the inclination and the optical axis are
within the standard. If any of them is out of
the standard, make adjustment again.

8. Apply the screw lock (red) to the two mount-


ing screws and the adjustment screw.
(Loosen the adjustment screw by one turn
before applying the screw lock.)

9-7 Confidential : duplication and reproduction prohibited


9-7 CCD Offset Adjustment
Procedure Illustration

1. Connect all the connectors to the main PCB


and the card unit.

2. Level the instrument.

3. Connect CH1 and CH2 of the synchroscope


to TP13 and TP19 respectively.
* CH1 is a trigger signal.
* CH2 is a CCD output waveform.

REC
4. Turn on the power as pressing the 7 and
Compe
ENT keys at the same time.
Adjust

ESC
5. Press the C key.
H_Adj
6. Collimate the collimator to adjust focus. Set1.5m?
CCD offset adjustment

7. Press the MEAS key to start measurement.

8. Make sure that automatic measurement is Waveform before adjusting VR3


done ten times by the instrument and the
waveform as shown in the right is displayed
on the synchroscope.
Synchroscope:
0V line

Lower end of CCD output waveform

9. Turn VR3 to fit the lower end of the wave- Waveform after adjusting VR3
form to “0V”.
* Adjustment and checking can be done only
while the instrument is measuring.
Standard: 0.0 ±0.1V

Confidential : duplication and reproduction prohibited 9-8


Procedure Illustration

The following procedures should be done if


adjustment and checking cannot be performed in
Procedure 7 ~ 9. When the adjustment and
checking procedures have already been per-
formed, it is not necessary to carry out Procedure
10 ~ 13.

10. Make sure that the value is displayed as


H_Adj
shown in the right and then press the ENT
key. 1.48534m
Example: 1.48534m

11. Make sure that the message appears on the


display unit as shown in the right. Dist?

12. To adjust distance, input the standard value


of the collimator through the ten keys of the
DL instrument.
CCD offset adjustment

* The standard value is written on the colli-


mator.
Dist
Example: 20.052m 20.052m

13. Press the ENT key and turn off the power.
* If an error or other troubles occur during
ESC
adjustment, press the C key. You can
continue the adjustment.

9-9 Confidential : duplication and reproduction prohibited


9-8 General Adjustment (1) (by Special Collimator)
Procedure Illustration

1. Preparation

(1) Inspect the collimator used exclusively for


the electronic level.

Reticle adjustment
(2) Level the inclination stand. screw

(3) Level the instrument.

2. Adjustment of horizontality

(1) Check the horizontality with the collimator. Special tool


If it is outside the standard, adjust with the
reticle.
Standard: ±2"

(2) Check the reticle inclination. If it is outside


the standard, adjust with a special tool.
General adjustment (1)

Standard: Collimator line width

Key Operation Display


3. Model selection

REC
(1) Turn on the power while pressing the 7 , REC MENU
MENU
and keys. 7 • 9 • Set
9
+ DL-101C
POWER

(2) DL-101C or DL-102C is selected by the


above operation. Set
DL-102C

Confidential : duplication and reproduction prohibited 9-10


Procedure Key Operation Display

4. Adjustment of compensator threshold volt-


age

REC REC
(1) Turn on the power while pressing the 7 7 • ENT
and ENT keys. +
POWER

(2) Level the instrument.


TOPCON
DL-102C

(3) When Fig. 1 appears on the screen of the ENT


instrument, press the ENT key.
Compe
Adjust?
(4) The instruction to tilt by +12 or 15' appears ENT Fig. 1
on the screen of the instrument. Tilt the
General adjustment (1)

instrument in the elevation angle direction


Tilting
and press the ENT key. (Press the key, and
the sign on the screen of the instrument is
+12or15'
changed to minus.)
Standard DL-101C: +12' ±1'
DL-102C: +15' ±1'
The tilt direction is also checked. Be sure to
tilt the instrument in the elevation angle
direction.

(5) When the displayed sign on the screen of the ENT Tilting
instrument is changed to “-” (minus), tilt the
instrument in the depression angle direction
-12or15'
and press the ENT key.
Standard DL-101C: -12' ±1'
DL-102C: -15' ±1'
The tilt direction is also checked. Be sure to
tilt the instrument in the depression angle
direction.

9-11 Confidential : duplication and reproduction prohibited


Procedure Key Operation Display

(6) When the check display “Set OK” appears


ENT
on the screen of the instrument, press the Set OK
ENT key to record the adjusted value. ENT or ESC
The next adjustment display appears on the
screen of the instrument.
Press the ESCC
key, and the adjusted value
is not recorded and the next adjustment dis-
play appears.

5. 1.5m writing
(1) Level the instrument and collimate the 20m
H Adj
pattern of collimator.
Set 1.5m?
(2) Press the MEAS key. Measurement is done
MEAS
ten times automatically.
Note: The displayed value as height includes H Adj
“1.5m” and offset. The value must not 1.5XXXm
be unstable while measurement is done
General adjustment (1)

Ten times
ten times.

(3) Press the ENT key. ENT Set OK


1.5XXX
6. Input of distance offset

(1) Input the distance. Input the numerical


The distance is displayed on the collimator. value with the Dist?
numerical keys of the
(Example)19.943 main unit.

(2) Press the ENT key.


ENT
Measurement is done ten times automatically.
Dist?
(3) Press the ENT key.
19.943
ENT
The offset input is completed and the power
is turned off automatically.
Dist
19.9XX
Ten times

Dist OK
19.943

Confidential : duplication and reproduction prohibited 9-12


9-9 General Adjustment (2) (by Staff Used Exclusively for DL, SG-3M or SA-5M)
Procedure Illustration

1. Set 20.000m as the distance between the Keeping tool


instrument center and the staff surface.
* Fix the staff with a keeping tool. 1.500m
DL
Tripod
Staff

20.000m

2. Carry out the rough adjustment by changing


the tripod’s height to set approx. 1.500m so
that the horizontal scale lines of the isntru-
ment may cross the 1.500m position of the
staff.
General adjustment (2)

1.500m
(center of R pattern)
Horizontal
3. Carry out the fine adjustment by turning the scale line
instrument’s three adjustment screws to the
same degree and set 1.500m correctly.
15

R pattern

Vertical scale line

The 1.500m position is the center of R pattern.

4. Carry out “1.5m writing”.


The operation is the same as Procedure 5 and
6 in “9-8 General Adjustment (1)”. (Refer to
P. 9-12.)

9-13 Confidential : duplication and reproduction prohibited


9-10 Final Check
Procedure Illustration

1. Preparation

(1) Set a staff at a point of the 95~100m dis- Staff


DL
tance.

(2) Set the instrument on a tripod.

95 100m
2. Measurement

(1) Level the center of the staff bar cord pattern


and press the MEAS key to carry out mea-
surement. Make sure that measurement is
possible. If not, move the focus and perform
measurement. Make certain which side the
focus is tilted and then focus the CCD again.

(2) Fit the reticle line to the 1/5 positions on the


left and right of the staff bar cord pattern, and
make sure that measurement is possible.
Final check

If not, check the difference and carry out the


CCD optical axis adjustment again.

Confidential : duplication and reproduction prohibited 9-14


10. APPENDICES
10-1 Repair Tool List

Tool Model Application Remarks

Collimator used exclusively ST-DL-01 CCD adjustment


for DL

Inclination adjusting tool ST-DL-02 Scale inclination adjustment

Pin face wrench ST-DL-03 Installation/removal of objective lens


unit
Power cable ST-GS-05 Power supply to PC board Common use
with GTS-4/5/6.

Compensator mounting ST-LK-30 Compensator mounting surface


surface inspection tool accuracy inspection

10-1 Confidential : duplication and reproduction prohibited


10-2 Adhesive Types and Areas of Application

THREE-BOND 1401C

THREE-BOND 1401C

THREE-BOND 1401C

THREE-BOND 1401

THREE-BOND 1401 (including sand) THREE-BOND 1401C


THREE-BOND 1401 KE45B CEMEDINE 30 min.

KE45B

THREE-BOND 1401 (including sand)

CEMEDINE HIGH-SUPER KE347T

THREE-BOND 1401C

Confidential : duplication and reproduction prohibited 10-2


THREE-BOND 1401C THREE-BOND 1401C

CEMEDINE HIGH-SUPER•KE347T

THREE-BOND 1401C

THREE-BOND 1401C

THREE-BOND 1401C CEMEDINE #007

THREE-BOND 1401C

10-3 Confidential : duplication and reproduction prohibited


Oil No. 7-2 Oil No. 7-2
Oil No. 7-2

Oil K-2

Oil K-2

Confidential : duplication and reproduction prohibited


Oil No. 7-2

Oil No. 7-2


10-3 Oil Types and Areas of Application

10-4
Oil G31 Oil G31 Oil 10000CS

Oil K-2 Oil K-2 Oil SH-33M


11. SPECIFICATIONS
Telescope
Magnification : 32× (DL-101C)
30× (DL-102C)
Objective aperture : 45mm
Field of view : 1° 20'
Resolving power : 3"

Compensator
Working range : ±12' (DL-101C)
±15' (DL-102C)
Setting accuracy : 0.3" (DL-101C)
0.5" (DL-102C)

Height Measurement
Accuracy (Standard deviation for 1km) : Electronic reading 0.4mm Invar staff (DL-101C)
1.0mm Fiber glass staff (DL-102C)
Optical reading 1mm (DL-101C)
1.5mm (DL-102C)
Least count : 0.01mm / 0.1mm (DL-101C)
0.01mm / 0.1mm (DL-102C)

Distance Measurement
Least count : 1cm/1mm (DL-101C)
1cm (DL-102C)
Accuracy : 1cm to 5cm
(Using the [MEAS] key)

Measuring Range : 2m to 100m (7ft to 328ft) : Fiber glass staff


2m to 60m (7ft to 197ft) : Invar staff

Measuring Time : 4sec.

Circular Level Sensitivity : 8' / 2mm (DL-101C)


10' / 2mm (DL-102C)

11-1 Confidential : duplication and reproduction prohibited


Others
Display : 2-line, 8-digit per line, Dot matrix LCD
Data Storage : Internal memory 400KB (approx. 8000 data)
Data transmission : RS-232C port provided
Keyboard : Alpha / Numerical input
Timer : Built-in timer
Horizontal circle : 360° or 400gon
Power supply : On-board rechargeable battery, BT-31Q NiCd 7.2V
On-board dry battery holder, DB-31
(Alkaline manganese dry batteries 6 pcs.)
Operating time : BT-31Q : 10 hours
DB-31 : 10 hours (Alkaline manganese dry batteries)
Ambient temperature range : -20°C to +50°C [-4°F to +122°F]
Dimensions : 237 × 196 × 141mm [9.33 × 7.72 × 5.55 inch]
Weight : 2.8kg [6.16 lbs] (including on-board battery)
On-board rechargeable battery, BT-31Q does not contain mercury.

Data Card
: PC card based on PCMCIA
(SRAM: 64Kbytes to 2Mbytes)

Staffs
Fiber glass staff : Length : 3m [9.84ft] (1.5m [4.92ft] × 2 pcs.)
Graduation : 1cm graduation with 5mm pattern (front surface)
Invar staff : Length : 3.0m [9.84ft]
Aluminum staff : Length : 5.0m [16.4ft]

Confidential : duplication and reproduction prohibited 11-2


TOPCON CORPORATION
75-1 Hasunuma-cho, Itabashi-ku, Tokyo 174-8580, Japan
Phone : 3-3558-2520 Fax : 3-3960-4214 www.topcon.co.jp

TOPCON POSITIONING SYSTEMS, INC. TOPCON SOUTH ASIA PTE. LTD.


7400 National Drive, Livermore, CA 94551, U.S.A. Blk 192 Pandan Loop, #07-01 Pantech Industrial Complex, Singapore 128381
Phone: 925-245-8300 Fax: 925-245-8599 www.topcon.com Phone: 62780222 Fax: 62733540 www.topcon.com.sg
TOPCON CALIFORNIA TOPCON AUSTRALIA PTY. LTD.
3380 Industrial Blvd, Suite 105, West Sacramento, CA 95691, U.S.A. Unit 18, 4 Avenue of Americas Newington NSW 2127, Australia
Phone: 916-374-8575 Fax: 916-374-8329 Phone: 02-8748-8777 Fax: 02-9647-2926 www.topcon.com.au
TOPCON EUROPE B.V. TOPCON INSTRUMENTS (THAILAND) CO., LTD.
Essebaan 11, 2908 LJ Capelle a/d IJssel, The Netherlands. 77/162 Sinn Sathorn Tower, 37th Fl.,
Phone: 010-4585077 Fax: 010-4585045 www.topconeurope.com Krungdhonburi Rd., Klongtonsai, Klongsarn, Bangkok 10600 Thailand.
IRELAND OFFICE Phone: 02-440-1152~7 Fax: 02-440-1158
Unit 69 Western Parkway Business Center TOPCON INSTRUMENTS (MALAYSIA) SDN. BHD.
Lower Ballymount Road, Dublin 12, Lreland
Phone: 01460-0021 Fax: 01460-0129 Excella Business Park Block C, Ground & 1st Floor, Jalan Ampang Putra,
Taman Ampang Hilir, 55100 Kuala Lumpur, MALAYSIA
TOPCON DEUTSCHLAND G.m.b.H. Phone: 03-42701068 Fax: 03-42704508
Giesserallee 31, 47877 Willich, GERMANY
Phone: 02154-885-100 Fax: 02154-885-111 info@topcon.de TOPCON KOREA CORPORATION
www.topcon.de 2F Yooseoung Bldg., 1595-3, Seocho-Dong, Seocho-gu, Seoul, 137-876, Korea.
Phone: 82-2-2055-0321 Fax: 82-2-2055-0319 www.topcon.co.kr
TOPCON S.A.R.L.
89, Rue de Paris, 92585 Clichy, Cedex, France.
TOPCON OPTICAL (H.K.) LIMITED
Phone: 33-1-41069490 Fax: 33-1-47390251 topcon@topcon.fr 2/F., Meeco Industrial Bldg., No. 53-55 Au Pui Wan Street, Fo Tan Road,
Shatin, N.T., Hong Kong
TOPCON ESPAÑA S.A. Phone: 2690-1328 Fax: 2690-2221 www.topcon.com.hk
Frederic Mompou 5, ED. Euro 3, 08960, Sant Just Desvern, Barcelona, Spain. TOPCON CORPORATION BEIJING OFFICE
Phone: 93-473-4057 Fax: 93-473-3932 www.topconesp.com Building A No.9, Kangding Street
TOPCON SCANDINAVIA A. B. Beijing Economic Technological Development Area, Beijing, China 100176
Neongatan 2 S-43151 Mölndal, SWEDEN Phone: 10-6780-2799 Fax: 10-6780-2790
Phone: 031-7109200 Fax: 031-7109249 TOPCON CORPORATION BEIRUT OFFICE
TOPCON (GREAT BRITAIN) LTD. P. O. BOX 70-1002 Antelias, BEIRUT-LEBANON.
Phone: 961-4-523525/961-4-523526 Fax: 961-4-521119
Topcon House Kennet Side, Bone Lane, Newbury, Berkshire RG14 5PX U.K.
Phone: 44-1635-551120 Fax: 44-1635-551170 TOPCON CORPORATION DUBAI OFFICE
survey.sales@topcon.co.uk laser.sales@topcon.co.uk C/O Atlas Medical FZCO., P. O. Box 54304, C-25, Dubai Airport Free Zone,UAE
Phone: 971-4-2995900 Fax: 971-4-2995901

Printed in Japan 0501-25TO 1