You are on page 1of 11

http://www.jhltonline.

org

STATE OF ART

Axial and centrifugal continuous-flow rotary pumps:


A translation from pump mechanics to clinical practice
Nader Moazami, MD,a,b Kiyotaka Fukamachi, MD, PhD,c Mariko Kobayashi, MD,c
Nicholas G. Smedira, MD,a,b Katherine J. Hoercher, RN,b Alex Massiello, MEBME,c
Sangjin Lee, MD, MS,b,d David J. Horvath, MSME,c and Randall C. Starling, MD, MPHb,d

From the aDepartment of Thoracic and Cardiovascular Surgery; bKaufman Center for Heart Failure; cDepartment of Biomedical
Engineering, Lerner Research Institute; and the dDepartment of Cardiovascular Medicine, Heart and Vascular Institute Cleveland Clinic,
Cleveland, Ohio.

KEYWORDS: The recent success of continuous-flow circulatory support devices has led to the growing acceptance of
LVAD; these devices as a viable therapeutic option for end-stage heart failure patients who are not responsive
continuous flow; to current pharmacologic and electrophysiologic therapies. This article defines and clarifies the major
axial flow; classification of these pumps as axial or centrifugal continuous-flow devices by discussing the difference
centrifugal flow; in their inherent mechanics and describing how these features translate clinically to pump selection and
rotary pumps; patient management issues. Axial vs centrifugal pump and bearing design, theory of operation,
pump mechanics; hydrodynamic performance, and current vs flow relationships are discussed. A review of axial vs
head curves centrifugal physiology, pre-load and after-load sensitivity, flow pulsatility, and issues related to
automatic physiologic control and suction prevention algorithms is offered. Reliability and
biocompatibility of the two types of pumps are reviewed from the perspectives of mechanical wear,
implant life, hemolysis, and pump deposition. Finally, a glimpse into the future of continuous-flow
technologies is presented.
J Heart Lung Transplant 2013;32:1–11
r 2013 International Society for Heart and Lung Transplantation. All rights reserved.

The evolution of mechanical circulatory support from perative for patient care and decisions regarding pump
volume-displacement pulsatile pumps to continuous-flow selection. In this review, we describe the basic engineering
(CF) rotary pumps has ushered in a new era for treatment of differences between the 2 types of CF pumps and provide a
end-stage heart failure. Increasing use of these pumps in the description of how these features translate clinically to
clinical arena is related to multiple positive attributes of this patient management.
class of pumps, including their smaller size, improved
durability, and enhanced survival with less morbidity.1,2 As
acceptance of these classes of pumps continues to grow and
newer-generation pumps are developed, a complete under- Axial vs centrifugal CF pump design and theory
standing of their mechanics and the inherent differences of operation
between centrifugal and axial-flow pumps becomes im-
CF rotary pumps generally consist of blood inlet and outlet
ports and a single rotating element that imparts energy to the
Reprint requests: Nader Moazami, MD, Surgical Director, Kaufman
blood to increase arterial blood flow and pressure. The rigid
Center for Heart Failure, Cardiac Transplantation and Mechanical
Circulatory Support, Cleveland Clinic, 9500 Euclid Ave, Desk J4-1, stationary housing(s) that surrounds and/or lies in the center
Cleveland, OH 44195. Telephone: 216-444-6708. Fax: 216-636-1212. of the rotating element incorporates some combination of
E-mail address: moazamn@ccf.org motor windings, permanent magnets, electromagnets, or
1053-2498/$ - see front matter r 2013 International Society for Heart and Lung Transplantation. All rights reserved.
http://dx.doi.org/10.1016/j.healun.2012.10.001
2 The Journal of Heart and Lung Transplantation, Vol 32, No 1, January 2013

Table 1 Comparison of Axial and Centrifugal Rotary Pump Response to Physiologic Conditions

k, decreased; m, increased; CBC, circulating blood volume; Pump dP, pressure difference across pump inlet and outlet; RHF, right heart failure; SVR,
systemic vascular resistance.
a
Axial pumps in clinical use do not show a linear power vs flow relationships over the entire ventricular assist device operating flow range, making
characterization of flow vs power relationships difficult.

mechanical bearing surfaces that act to drive and support the and illustrated in Figure 1 are several different kinds of
rotating element. bearing elements that are used to accomplish this:
The primary difference between centrifugal-flow and
axial-flow pumps lies in the design of their rotating elements  Mechanical/pivot: Rotor is suspended with mechanical
(Table 1). When one considers the theory of operation of a bearings on spherical surfaces rotating in sockets (eg,
centrifugal CF pump, its rotating element acts as a spinning HeartMate II, Thoratec Corporation, Pleasanton, CA).
disk with blades that can be viewed as a ‘‘thrower,’’  Hydrodynamic: Rotor derives lifting from fluid forces
meaning that the fluid is captured and thrown tangentially in thin, fluid blood films separating rotor and pump
out off the blade tips. In contrast, axial CF pump rotating housing based on the relative motion of surfaces (eg,
elements operate like a propeller in a pipe and can be HeartWare HVAD, HeartWare International Inc,
viewed as a ‘‘pusher.’’ This mechanism can also be viewed Framingham, MA).
as an ‘‘auger’’ trying to screw itself into the inlet fluid,  Electromagnet/position sensor: Suspends rotor using
against the ‘‘resistance force’’ at the outlet, to overcome the electronic position control and electromagnets (eg,
difference between pre-load and after-load. DuraHeart, Terumo Heart Inc, Ann Arbor, MI; and
HeartMate III, Thoratec Corp).
Bearing design  Permanent magnet: Repelling magnets in rotor and
pump housing suspend the rotor. Permanent magnets
A further distinction made between CF pumps is the method alone cannot suspend a rotating element because
used to support the rotating element (rotor). Listed below magnetic forces change continuously with the position
Moazami et al. Comparison of Axial vs Centrifugal Pumps 3

Figure 1 Rotary pump bearing types. (A) Mechanical pivot bearing: a drawing of a generic axial-flow pump illustrates how the rotating
element (rotor) is suspended by pivot bearings on either end as it spins. These bearings, which are usually made from very hard jewel or
ceramic materials, are in mechanical contact while the rotor is spinning. (B) Electromagnetic bearing: Illustration shows electromagnetic
levitation of the INCOR (Berlin Heart GmbH) axial continuous-flow pump rotor. Levitating electromagnets act on opposing permanent
magnets (black) on both ends of the rotating element to suspend the rotor axially, while the permanent follower magnets in the rotating
element and electromagnetic forces induced in the motor stator suspend the rotating element radially. The arrows show the directions of
magnetic forces. (C) Hydrodynamic radial bearing: Diagram shows a cross-section through a generic rotating cylinder (white disc in cross-
section) within a stationary cylinder that is suspended by thin fluid film pressures (fluid is grey in the diagram). This bearing has no
mechanical contact because the hydrodynamic forces keeping the rotating element away from the stationary element increase as the distance
between them decreases. The magnitude and direction of these fluid forces is represented by the height and direction of the arrows.
(D) Hydrodynamic thrust bearing: Diagram shows the cross-section of a generic thrust bearing, consisting of 2 plates, where thin fluid film
pressure acts to separate a rotating surface (2) moving to the right in this figure relative to a stationary surface (3). These forces are created by
the fluid wedge created between geometry of the moving plate relative to the stationary plate and again are represented by the direction and
height of the arrow in the diagram. An example of this bearing type is shown in Panel E. (E) Hydrodynamic thrust and permanent magnet
bearing combination: Assembled component view of the HeartWare HVAD centrifugal continuous flow LVAD on the right and cross-section
on the left. Permanent magnets (1-1) align and separate the rotating disc impeller (2) from the conical stationary center shaft of the pump
housing to form a permanent magnet bearing. This bearing is combined with thin film fluid pressures generated at the tapered hydrodynamic
thrust bearing surface of the rotating disc impeller (2) that creates the fluid wedge needed to create sufficient hydrodynamic forces to lift the
impeller off the stationary housing bearing surface (3). The red line between the rotor (2) and housing (3) is the blood fluid film.
4 The Journal of Heart and Lung Transplantation, Vol 32, No 1, January 2013

Figure 2 (A) Representation of the systemic arterial pressure (AoP) and left ventricular pressure (LVP) relationship is shown during rotary
LVAD support in a failing ventricle. Comparison of (B) centrifugal-flow and (C) axial rotary pump performance characteristics are shown as
pressure differential across pump inlet and outlet in mm Hg and pump flow in liters/min. (D) Representative centrifugal and axial rotary pump flow
waveforms during rotary LVAD support in a failing ventricle are compared. Reprinted with permission from Pagani FD. Continuous-flow rotary
left ventricular assist devices with ‘‘3rd generation’’ design. Semin Thorac Cardiovasc Surg 2008;20:255–263. r 2008 Elsevier, Inc.

of the rotor. Therefore, permanent magnets are generally magnetic bearing elements are frequently backed up with
used in combination with one of the other bearing types. hydrodynamic bearings, which require no control.
In a magnetically suspended pump, stabilization in one or Hydrodynamic bearings similarly produce no life-
more directions of motion is achieved by using limiting contact between the bearing surfaces, but in
hydrodynamic or electromagnetic bearing elements. contrast, are simple and reliable. Once the rotor starts
spinning, it essentially ‘‘water skis’’ on a film of blood,
deriving lift and separation from its own motion. However,
As a practical matter, the primary difference in blood
the load-bearing blood fluid film is prone to higher shear
pump bearings lies in their complexity and reliability.
stress, and theoretically, more hemolysis can occur. Another
Mechanical bearings have tiny and precise ceramic
design issue is that the rotor and housing surfaces are in
components whose advantage is that they position the
contact at startup and shutdown, when there is no relative
rotating assembly in all 6 directions of motion and are stable
motion. The bearing surface material properties need to
at all speeds and operating conditions. Although their small
accommodate this friction to avoid damage at these 2 times.
size limits the surface area exposed to mechanical contact
Surface coatings, if used, require extended endurance testing
and friction, these surfaces are potential sites for thrombus/
to verify multiyear reliability.
fibrin deposition, which is not present in other bearing
designs that completely suspend the rotor. The concentra-
tion of hydraulic loads on the rotor at these small pivot
bearings also makes them theoretically life-limiting for wear Hydrodynamic performance
and fragile with respect to impact.
Electromagnetic rotor positioning generates real-time The pressure across the inlet and outlet of any hydro-
electromagnetic forces on the rotor to actively counteract the dynamic pump is termed by engineers as the ‘‘pump delta
surrounding fluid and magnetic forces that could cause P’’ or ‘‘head pressure.’’ Figure 2A shows the typical
unstable rotor operation. Among the advantages of this pressure difference between the left ventricle (LV; pump
approach are that there is no contact between the bearing inlet) and aorta (pump outlet) of a failing heart supported by
surfaces, so there is no life-limiting wear, and that it a rotary pump. Figure 2B and C compare the typical
provides the lowest shear stress to the blood because of hydrodynamic performance curves (pump head curves) for
relatively large clearances between the rotating element and centrifugal vs axial CF pumps, respectively. A pump head
the housing. However, these systems are complex, requiring curve compares the relationship between pump flow and
position sensors, electromagnets and extra conductors, pressure difference across the pump ports (delta P) at one
connector pins, and electronics to execute the dedicated operating pump speed. Figure 3 shows a full set of pump
position control algorithm. If, for example, there is an head curves over the full range of operating pump speeds
electrical contact failure in a connector pin, or momentary for the axial HeartMate II and the centrifugal HeartWare
instability encountered in the control algorithm, pump HVAD and Terumo DuraHeart left ventricular assist device
failure can occur. To make the system fail-safe, electro- (LVAD).3
Moazami et al. Comparison of Axial vs Centrifugal Pumps 5

Figure 3 Comparison of head curves for axial vs centrifugal continuous flow (CF) pumps is shown for the (A) HeartMate II axial CF, the
(B) HeartWare HVAD centrifugal CF,3 and the (C) Terumo DuraHeart centrifugal CF. Reprinted with permission from Frazier OH et al.
Optimization of axial-pump pressure sensitivity for a continuous flow total artificial heart. J Heart Lung Transplant 2010;29:687–91. r 2010
Elsevier Inc. and Alex Medvedev, Terumo Heart, Inc.The HeartWare HVAD is an investigational device limited by Federal Law to
investigational use.

Most important to note is that most centrifugal pumps demonstrate that the actual degree of flatness of centrifugal
have what is called a flat head curve, where they operate pump head curves can vary with design, with the HeartWare
over a very wide range of flows for a very small change in HVAD being less flat than the Terumo DuraHeart design.
delta P across the pump. The example in Figure 3 shows that
for 1 cardiac cycle in which pump delta P swings from 40 to
80 mm Hg, centrifugal pumps have a very large swing in Inlet cannula suction and control of pump
flow (0 to 10 liters/min), acting almost like a pulsatile pump operating speed
with high peak systolic flows and low, even negative,
diastolic flows. One can think of a centrifugal pump with a The likelihood of high inlet cannula suction in axial vs
flat head curve as an on/off CF pump cycling between high- centrifugal CF pumps is related to the flow vs delta P
flow and low-flow as its output surges with the beating of relationship described above. A CF pump is blind to the
the ventricle. This creates inherent high pump flow pulsatility in absolute value of the pressures at the inlet and outlet ports
response to changing LV pressures (Figure 2D). In contrast, a and responds only to the total differential pressure across the
typical axial-flow pump has a steep head curve where there is a pump. At any given pump speed, a CF pump would have
linearly related increase and decrease in flow with decreasing the same flow at inlet and outlet pressure values of 0/100,
and increasing pump delta P.4 In this example, the 40- to 80- 100/200, and –50/50 mm Hg, because it sees only the delta
mm Hg swing across the pump conduits produces less flow P of 100 mm Hg. This has clinical significance during a low
pulsatility, ranging from 3 to 7 liters/min during a cardiac cycle. LV volume state in which pump flow decreases, such as one
As the text that follows explains, this difference in pump-flow might expect during conditions related to hypovolemia
pulsatility affects the diagnostic and control feedback available associated with right ventricular (RV) failure or bleeding.
to the respective pump controllers. Figure 3B and C As can be seen from the head curves in Figure 2B, a
6 The Journal of Heart and Lung Transplantation, Vol 32, No 1, January 2013

centrifugal LVAD with a flat head curve has a fixed relies on an ultrasonic flow probe incorporated into its outlet
differential pressure as flows decrease from 5 to 0 liters/min, conduit for flow determination. The centrifugal HeartWare
so no significant increase in inlet suction occurs at low flows HVAD and Terumo DuraHeart LVAD flow estimations
or even total occlusion. However, as a result of the steep derived from motor power and speed inputs have been
head curve characteristic of axial pumps, the increasing generally considered accurate and reliable enough to use in
pressure differential generated across them as flow decreases clinical assessment.
translates into much higher suction developed at the inlet Accurate flow estimation for axial and centrifugal rotary
during conditions of low LV volume or inlet cannula pumps also depends on the viscosity of blood, which can be
obstruction.4 This means axial pumps pull the hardest at the estimated from the patient’s hematocrit. Currently, the
lowest flow, creating the potential for a self-latching HeartWare HVAD has a programmable hematocrit setting,
condition wherein the pump continues to increase suction whereas the HeartMate II does not. Implementation of
as flows fall, resulting in the ventricular wall being sucked automatic control algorithms is expected to be more difficult
in around the inlet. Therefore, centrifugal pumps have an in axial pumps without direct flow measurements than in
advantage over axial pumps in the avoidance of ventricular centrifugal pumps.
suck down at low flow.
A complication of intermittent LV suck down is Axial vs centrifugal physiology and control
ventricular arrhythmias. Morshuis et al5 reported a 65%
lower prevalence of ventricular arrhythmia than reported for
axial flow LVADs in their clinical experience with the Pre-load Sensitivity
DuraHeart centrifugal pump. Another pathologic condition
that higher LV inlet suction can create is leftward shift of the Cardiac output in the natural heart is determined by the
interventricular septum. This not only negatively alters the interaction of after-load, myocardial contractility, heart rate,
mechanics of the septal contribution to RV output but can pre-load sensitivity, and LV compliance. Pre-load sensitivity,
also increase tricuspid valve regurgitation secondary to the as it relates to mechanical circulatory assist devices, mimics the
anatomic connection of the septal leaflet of this valve, relationship between LV filling pressures and ventricular
causing it to be ‘‘pulled’’ away.6,7 stroke volume defined by the Frank-Starling curves.12 It is
Recent trends in patient management have been to calculated from the ratio of pump output to pump filling
operate CF LVADs at lower speeds and at less than the pressures at the pump inlet in liters/min/mm Hg. Salmonsen
maximum level of cardiac support to ensure continuous or et al13 reported an average pre-load sensitivity for the
intermittent opening of the aortic valve during operation. DuraHeart, HeartWare HVAD, HeartMate II, and INCOR
This is to prevent reported complications of aortic valve (Berlin Heart AG, Berlin Germany) CF rotary pumps of
fusion and reduce the potential for aortic regurgitation, but it 0.105 ⫾ 0.096 liters/min/mm Hg, which is almost 3 times
also reduces the likelihood of suction events.8–10 lower than the 0.275 liters/min/mm Hg reported for the
human heart.14 The limited pre-load sensitivity of CF devices
theoretically limits their rate of increased output in response to
increasing LV venous return. There is no evidence in the
Flow estimation accuracy based on flow vs
literature to suggest that axial or centrifugal pumps would have
power relationships any significant clinical difference in pre-load sensitivity.
Centrifugal pumps have a linear current-to-flow relationship Reports of greater LV volume unloading by pulsatile vs
across the full range of operating pump flows. In addition, CF LVAD pumps15,16 may be a consequence of the low pre-
the characteristic current and flow pulsatility in response to load sensitivity of the CF LVADs. Interestingly, a review of
the pressures generated across the pump makes the centrifugal all bridge-to-recovery cases at the German Heart Institute
pump motor current an accurate sensorless index of pump from 1992 to 2009 showed that patients with a pulsatile-
flow and a sensitive virtual index of the LV pressures during flow LVAD had an almost 3-fold chance for myocardial
the cardiac cycle. This information allows centrifugal pump recovery compared with those who received CF devices.
controllers to accurately monitor pump flow and the degree of This may be due to greater volume unloading of the LV by
LV unloading by simply monitoring the motor current or pulsatile pumps.17 To date, no published reports have
power.4 compared the exercise tolerance of patients with axial vs
In contrast, the correlation between flow and current in centrifugal CF LVADs.
axial pumps is not nearly linear over the full operating flow
range, is not as well defined as in centrifugal pumps, and After-load sensitivity
offers less accuracy for flow estimation. For example, the
HeartMate II axial flow pump does not display flows below Data from Salamonsen et al13 on after-load sensitivity of CF
3.0 liters/min. In an intraoperative study of estimated flow rotary pumps support the general understanding that these
accuracy in 20 HeartMate II patients, Slaughter et al11 pumps have higher after-load sensitivity (0.09 ⫾ 0.034
concluded that the device’s estimated flow values can be liters/min/mm Hg) than the human heart (0.03 ⫾ 0.01 liters/
used only ‘‘to provide directional information for trend min/mm Hg). This high after-load sensitivity creates the
purposes rather than absolute values of pump flow.’’ The need to control systemic vascular resistance (SVR) in these
MicroMed DeBakey (MicroMed, Houston, TX) axial pump patients to guarantee sustained outputs.
Moazami et al. Comparison of Axial vs Centrifugal Pumps 7

Typically, the targeted mean systemic arterial pressures inversely related to the degree of LV unloading by the
for CF pump patients is 70 to 90 mm Hg, with pressures pump and directly proportional to the strength of LV
exceeding 90 mm Hg to be avoided.18 Although axial and contraction and as such can be used as a measure of the LV
centrifugal CF pumps share this generalized categorization, function under VAD support. Any significant decrease in
analysis of their head curves suggests some relative pump flow pulsatility without a change in pump speed
differences that can have significant clinical implications. should be investigated clinically for causes of decreasing
Because centrifugal pumps operate on a flatter head curve LV pressures during LVAD support. This typically
than axial pumps, they demonstrate larger changes in includes decreasing LV contractility or low LV volume
flow for any given change in pressure across the pump. states caused by right heart failure or dehydration. Reasons
If SVR were to increase (increase in delta P), the pump for the flow PI to increase, without a change in pump
outlet pressure increase would produce an instantaneous speed, include an increase in LV contractility via inotropes,
drop in pump flow to maintain a constant pump outlet myocardial recovery, and exercise or increased pre-load
pressure, which means that centrifugal pumps cannot pump Starling effects.
against a high blood pressure, thereby causing a lower flow It is not surprising that automatic physiologic pump
condition. control algorithms have not been used to any significant
In contrast, the steeper head curve of axial pumps extent in the clinical axial-flow pumps given their
responds to an increase in SVR by increasing the pressure previously stated poor flow estimation accuracy, lower flow
generated across the pump ports, limiting the decrease in pulsatility, and the increased danger of LV suction events.
flow by increasing outlet pressure. At low-flow conditions, They are typically operated only in fixed-speed mode with
this retains the capability to enforce high blood pressures a suction-detection control algorithm. The CorAide cen-
with adequate LV volume. However, it also results in high trifugal LVAD used sensorless flow estimates, high pump
inlet suction that in a low LV volume state can potentiate flow pulsatility response to changes in the status of the LV,
arrhythmias, create suction events, or lead to hemolysis. and much lower LV suction levels during transient periods
of low LV volume to successfully implement a physiologic
automatic pump control algorithm in its European clinical
Pump flow pulsatility and automatic trial.19,21 In this limited trial of 21 patients, 72% of the total
physiologic pump control trial implant duration (16.6 patient-years) was run in
automatic mode (personal unpublished data).
The degree of pulsatility in the pump flow waveform during Although the other centrifugal CF pump systems can
a cardiac cycle for CF rotary pumps is termed the flow theoretically implement similar physiologic automatic con-
pulsatility and is quantified in various ways for different trol schemes, there have been no reports to our knowledge
pump systems as a pulsatility index (PI). It is typically of any other axial or centrifugal CF pump system that
displayed as some form of the ratio of the peak flow during routinely uses a physiologic control algorithm. For the
systole (flowmax) or the total swing in flow during a cardiac centrifugal CF pump systems, this is most likely attributable
cycle (flowmax – flowmin) over the average pump flow to the excellent results to date using fixed-speed mode of
(flowavg).3,18,19, This is normally derived by averaging the operation. Any potential advantages to physiologic auto-
recorded instantaneous peak flows or flow swings over each matic control algorithms in CF LVADs will have to wait for
cardiac cycle for approximately 10 to 15 seconds and then future control system developments in the newer CF pump
calculating the average total pump flow over that same systems.
period. The formula for calculating a PI for the HeartMate II
(PIHM II )20 is {PIHM II ¼ [(flowmax – flowmin)/flowavg]  10. LV suction-detection algorithms
For example, if rotary pump flow values for flowmax,
flowmin,, and flowavg were 8, 4, and 6 liters/min, respectively, Centrifugal and axial CF pump systems use some mechan-
the PIHM II would be 6.7. Multiply the PIHM II by 10, and it ism to detect a suction event; however, the reliability of the
can be interpreted as the percentage of the average flow that centrifugal pump flow calculations at low-flow levels and
the pump flow swings through during a cardiac cycle, which their higher flow pulsatility levels allow these systems to
for this example would be a 4 liters/min swing or 67% of the implement pre-suction avoidance before suction occurs by
average flow of 6 liters/min. The average PIHM II for the axial dropping pump speed when flow PI drops below pre-
HeartMate II bridge-to-transplantation (BTT) trial was programmed lower limits. Because rotary pump suction
5.0 ⫾ 0.9, and the average PI calculated as PIHM II for the events first typically occur during early ventricular diastole,
centrifugal CorAide European clinical trial was 12.1 ⫾ 2.6, during the lowest LV volumes and pressures, the centrifugal
demonstrating more than twice the flow pulsatility.3,18,19 HeartWare HVAD pre-suction detection system looks for
Although flow pulsatility data for the HeartWare HVAD a sudden decrease in the pump’s diastolic minimum flow
clinical trial are not available, it is recommended to maintain a level. The baseline diastolic minimum flow level is
minimum of 2 to 4 liters/min for pump flow operating ranges continuously averaged and recalculated every 2 seconds,
of 3.5 to 7.0 liters/min, which equates to a PIHM II value and a pre-suction detection alarm occurs if the baseline flow
of 5.7. decreases by more than 40% for 10 seconds.3 Only an alarm
The significance of flow pulsatility for rotary blood is activated, because there is no automatic response to
pumps lies in the fact that the degree of flow pulsatility is decreased speed under LV pre-suction detection conditions.
8 The Journal of Heart and Lung Transplantation, Vol 32, No 1, January 2013

Because axial CF pumps typically do not derive accurate their rotors and large hydrodynamic bearing surfaces capable of
pump flow values and do not have the degree of pump flow safely distributing transient touchdown forces due to shock
pulsatility of centrifugal CF pumps, they tend to rely on loads. The HeartMate III centrifugal electromagnetic rotor
detection of a suction event after it occurs based on the levitation system was designed to withstand accelerations of
sensed disturbed flow and motor current feedback when more than 7 times that of gravity.29
intermittent inlet cannula suction takes place.22–23 The As is expected in any initial trials of new technology, the
HeartMate II system does automatically decrease pump later-arriving centrifugal CF rotary pumps identified relia-
speed in response to a suction event. Hertzer et al24 have bility issues not identified in pre-clinical testing. In the
reported that the INCOR axial flow pump has an automatic HeartWare HVAD BTT clinical trial, Strueber et al30
anti-suction algorithm based on flow pulsatility falling below reported no device failure; however, 4% (2 of 50) of the
a pre-determined minimum level, and recent publications devices were electively exchanged because of what was
indicate that Berlin Heart Inc, is developing and testing determined to be manufacturing variability of the pumps’
algorithms for automatic pre-load–sensitive control of rotary hydrodynamic thrust bearings, resulting in an area of
blood pumps.25 reduced flow that may have been attributable to thrombus
formation. After resolving the manufacturing process
problem, they reported no further events occurring in more
Axial vs centrifugal reliability and than 450 additional implants.30
biocompatibility The initial European experience with the DuraHeart
centrifugal CF rotary LVAD in 68 patients reported no
Mechanical wear and implant life pump mechanical failures; however, the pumps in 2 patients
were replaced electively because of temporary flow
With CF LVADs now being used as permanent destination interruption identified by the manufacturer caused by a
therapy for end-stage heart failure patients, high reliability out motor manufacturing process that has since been cor-
to 5 years is generally considered a minimum requirement for rected.31 As experience with longer LVAD support duration
permanent use, and a 10-year device is a targeted goal for accumulates, the answer to the relative reliability of axial vs
newer device designs. Most centrifugal CF LVADs are centrifugal rotary blood pumps will need to play itself out in
defined as third-generation pumps in which the pump rotating the clinical arena despite the theoretic advantages offered by
element within the blood flow path is totally suspended during third-generation pumps.
operation by some combination of hydrodynamic and/or
magnetic (passive or active) forces, giving these pumps no
life-limiting mechanical loads. Hemolysis
Most axial CF pumps use second-generation rotating
assemblies that are pivoted on small bearing surfaces in the Pump hemolysis is directly related not only to pump speed
blood flow path. An exception is the Berlin Heart INCOR and areas of high shear but also to the residence time of
axial CF LVAD, which uses magnetic bearings to actively blood in the high fluid shear areas of the pumping elements
suspend the rotating assembly. Bearing contact points in or bearings. Axial pumps generally have significantly
principle are pre-disposed to mechanical wear and heat smaller clearances at the impeller blade tips and higher
generation, limiting the operating life of the pump. Despite speeds at the outer edge of their blades, creating higher
the stated theoretic limits to pivoted bearing pumps, the shear levels than for the centrifugal pumps; however, the
Jarvik 2000 (Jarvik Heart Inc, New York, NY)26 and the high shear around the blade edges tends to be relatively
HeartMate II axial rotary pumps have both shown excellent short in duration, helping to mitigate this blood damage.
reliability, with no reported life-limiting wear of their blood- Hydrodynamic bearing design is tricky because of the
immersed bearings.27 significant residence time in the high shear fluid film that
However, the German Heart Institute reported in 2010 suspends the rotating assembly. However, hydrodynamic
that of 65 Berlin Heart INCOR and 18 MicroMed DeBakey bearing hemolysis has been mitigated to low levels by
axial rotary pumps implanted between 1991 and 2009, proper bearing design, as demonstrated by the HeartWare
3 stopped because of technical failure and 2 Berlin Heart HVAD. The current literature does not definitively support
INCOR devices had bearing problems.28 Interestingly, the any difference in hemolysis rates between the current axial
HeartMate II axial pivoted-bearing CF pump was designed vs centrifugal CF pumps, nor has hemolysis been a
as a 5-year device, whereas the HeartMate III centrifugal CF significant complication for CF pumps in general; however,
pump, now in pre-clinical animal testing, has a stated design the literature indicates that axial pumps may produce mild
goal of a 10-year operating life.29 subclinical hemolysis whereas the newer centrifugal pumps
Another potential failure mode for the small pivoted do not.
bearings in axial CF pumps is susceptibility to shock loads, A 2011 report on implantation of 2 HeartWare HVAD
such as in a car accident, in which high decelerating forces are pumps for biventricular support by the German Heart
imposed. These forces are concentrated over very small bearing Institute reported no significant increase in free hemoglobin,
surface areas and can lead to catastrophic failure. In contrast, bilirubin, or lactate dehydrogenase (LDH) blood levels
most centrifugal pumps have relatively stiff permanent magnet during the first 6 post-operative months.32 The initial
suspension systems to prevent significant axial displacement of European experience with the DuraHeart centrifugal CF
Moazami et al. Comparison of Axial vs Centrifugal Pumps 9

rotary LVAD showed a decrease in total bilirubin, glutamic- increased experience in both axial and centrifugal rotary
oxaloacetic transaminase, glutamic-pyruvic transaminase, blood pump application, anti-coagulation requirements have
free hemoglobin, and LDH from baseline values for decreased for some of these pumps. Analysis of the monthly
those patients reaching 6 months’ duration, indicating no INR of a 331-patient sub-set of patients successfully
clinically significant hemolysis.31 discharged in the BTT arm of the U.S. HeartMate II pivotal
The reported incidence of hemolysis for the HeartMate II trial recommends an INR of 1.5 to 2.5 for the HeartMate II
axial rotary pump is 0.06, 0.02, and 0.13 events per to retain a low incidence of thromboembolic events while
patient-year in its BTT U.S., destination therapy U.S., and minimizing hemorrhagic events.40 El-Banayosy et al41 used
European multicenter clinical trials, respectively.33 In a a more moderate anti-coagulation protocol to manage all
study comparing hemolysis indices in CF LVAD patients34 rotary blood pumps. They report successfully using a target
supported by the axial CF rotary pump (HeartMate II) and INR of 2.0 to 2.5 and aspirin at 100 mg daily to anti-
the (VentrAssist)35–36 centrifugal CF rotary pump, patients coagulate 6 patients with HeartMate II axial, 10 with
with the centrifugal LVAD displayed normal haptoglobin DuraHeat, and 8 with VentrAssist centrifugal rotary pumps,
levels, low plasma free hemoglobin, and moderately with no device thrombosis, no hemorrhagic or ischemic
increased LDH. In comparison, the HeartMate II patients stroke, and no other thromboembolic complications.41
showed lower haptoglobin and higher free hemoglobin and We must note that the recommendations for maintaining
LDH values. Hemolysis indices reported by Hetzer et al24 INR in the range of 1.5 to 2.5 to balance the dangers of
for the initial experience with the Berlin Heart INCOR axial thromboembolic events with bleeding events have not been
rotary pump showed a statistically significant decrease in confirmed prospectively to date. The lack of mechanical
LDH to below baseline followed out to 6 months. Plasma contact, lower bearing shear levels, and use of more
free hemoglobin did not increase statistically, but they biocompatible pump surfaces in the electromagnetically
reported some persistent elevation from baseline in the suspended centrifugal pumps may allow operation at even
study.24 Westaby et al37 documented no significant change lower anti-coagulation regimens. Use of only anti-platelet
in free hemoglobin or LDH for 4 Jarvik 2000 patients, agents is being proposed for the HeartMate III centrifugal
whereas Siegenthaler et al38 reported mild hemolysis not LVAD now in pre-clinical testing.29
requiring transfusion, stable hemoglobin levels, increased The reported ischemic and hemorrhagic stroke event rates
LDH levels, and low haptoglobin levels in 3 Jarvik 2000 per year for the HeartWare HVAD were 0.11 and 0.05,
patients. Despite these reports of low hemolysis based on respectively, in the BTT clinical trial vs 0.09 and 0.05 for the
pump design, new-onset hemolysis is often the first HeartMate II clinical trial and 0.05 and 0.01 in a post-approval
manifestation of problems in CF LVADs that lead to study.42 The neurologic complication rate for the DuraHeart
turbulent flow such as can be seen with a kinked outflow centrifugal rotary pump improved after the anticoagulation
graft or partial pump thrombosis. regimen was decreased from an INR of 2.5–3.5 to 2.0–2.5 in
the European clinical experience; however, the incidence of
transient ischemic attacks (0.23) was somewhat higher than
Pump deposition for the HeartMate II and HVAD pumps (31).

From a perspective of pump design, mechanical wear and Future directions


heat generation at the bearing contact points in the second-
generation axial pumps increase the chances of fibrin The excellent reliability and biocompatibility of the Heart-
deposition and thrombus forming in the blood flow path. Mate II axial-flow LVAD has been widely published. This
All CF pumps are designed around an operating point being review, however, suggests that centrifugal rotary CF pumps
a single-flow and pump-speed condition; however, centri- may have some theoretic and real advantages over axial CF
fugal pumps can be less sensitive to off design conditions pumps (Table 2). Going forward, the pendulum may swing
because they do not have stator vanes. Axial pumps have to this pump design in the continuum of CF rotary pump
stator (stationary) vanes at the inlet and outlet that guide the design and development. It is not a coincidence that the
flow at the inlet to prevent pre-rotation of the blood as it HeartMate III centrifugal CF rotary pump eliminates many
enters the pump and catch or ‘‘scoop up’’ and straighten the of the limitations of axial pump design, including (1)
flow at the outlet. The geometry of the vanes is just right for bearing-less low shear magnetic levitation of the pump rotor
only 1 value of flow/speed (eg, 5 liters/min/10,000 rpm), to increase reliability and durability, (2) an accurate sensor-
representing the ‘‘design point’’ for the device. At this point, less flow estimator, (3) the ability to withstand high shock
flow separation from the surfaces (meaning areas of poor loads, (4) incorporation of a pulse mode to increase the level
wash) is minimized. At conditions of high flow/low speed of pulsatility and even produce physiologic pulse pressures,
(low SVR) or low flow/high speed (high SVR), these stator and (5) the potential for reduction of anti-coagulation
vanes are off design and may accumulate thrombus in the requirements to only daily anti-platelet medications. This,
zones of flow separation. combined with the capability to implement physiologic
A more aggressive anti-coagulation and anti-platelet control algorithms based on accurate flow estimates and the
therapy was used initially for many of the axial CF rotary inherent advantages of flat centrifugal pump head curves
pumps (international normalized ratio [INR] 2.5 to 3.5) over steep axial pump head curves, allows (1) much lower
compared with the HeartMate I pulsatile devices.39 With inlet suction at low flows; (2) lower outlet pressures at high
10 The Journal of Heart and Lung Transplantation, Vol 32, No 1, January 2013

Table 2 Summary of Key Physiologic Differences in Axial and Centrifugal Continuous Flow Pumps

Pump characteristics C vs A Qualitative pump comparisons


Flow pulsatility C4A Centrifugal pumps have significantly higher flow pulsatility.
Estimated flow accuracy C4A Centrifugal pumps have significantly higher estimated flow accuracy.
Inlet suction A4C Centrifugal pumps have significantly lower inlet suction at low flow conditions.
Ability to scale size down A4C Axial pumps can be more easily scaled down to sizes sufficient to be implanted intravascularly.
Pre-load sensitivity A¼C Axial and centrifugal continuous-flow pumps both have low preload sensitivity relative to the
native ventricle and pulsatile VAD.
After-load sensitivity C4A Axial and centrifugal continuous flow pumps both have high after-load sensitivity relative to
the native ventricle and pulsatile VAD; however, centrifugal pumps, by hydraulic
performance characteristics, have higher after-load sensitivity.
Susceptibility to infection A ¼ C No difference.
Biocompatibility A ¼ C No difference.
Hemolysis A ¼ C Not enough clinical data to suggest one is superior over the other.
Anti-coagulation A ¼ C Not enough clinical data to suggest one is superior over the other.
Ability to wean the pump A ¼ C Not enough clinical data to suggest one is superior over the other.
A, axial; C, centrifugal; VAD, ventricular assist device.

SVRs, preventing the creation of exceedingly high systemic throwing blood tangentially off their blades to create
arterial pressures; (3) higher flow pulsatility that is more centrifugal forces that can be achieved only with sufficient
sensitive to LV pressure swings, providing both a more pump diameters that would preclude them from intra-
sensitive pre-suction detection capability and increased vascular applications. Biocompatible sub-cutaneously im-
diagnostic feedback as to the functional status of the LV; planted or percutaneously delivered intravascular pumps
and (4) safer long-term management of pump speed and capable of providing permanent circulatory support at more
pump output optimization. The lower inlet suction at low than 3.0 liters/min minimum appear to be the new Holy
flows demonstrated by centrifugal rotary blood pumps may Grail of the rapidly expanding universe of cardiac
be one of the first clinical criteria to be used in selecting circulatory support devices.
centrifugal pumps rather than axial pumps for some patients For now, the axial and centrifugal CF rotary pumps have
in diastolic heart failure. VAD candidates presenting with overcome the decades-old challenges of biocompatibility,
hypertrophic ventricles with small end-diastolic volumes reliability, and hemocompatibility to thrust LVAD technol-
may be less likely to experience arrhythmias due to contact ogy into accepted clinical therapy for end-stage heart failure
of the inlet cannula with the LV wall when a centrifugal that is refractory to pharmacologic and electrophysiologic
rotary pump is used. treatment.
HeartWare International is currently in pre-clinical
testing of its next-generation device, the MVAD, which is Disclosure statement
a wide-blade axial CF pump.43 Whereas the Berlin Heart
INCOR rotary LVAD has been designed as a third- The authors thank Tess Parry and Christine Kassuba for editorial
generation axial pump using active electromagnetic suspen- assistance.
sion of the rotor, the MVAD features a magnetic axial rotor None of the authors has a financial relationship with a commercial
and uses a combination of passive permanent magnets and entity that has an interest in the subject of the presented manuscript or
other conflicts of interest to disclose.
hydrodynamic bearing surfaces to totally suspend the rotor
in operation with a displacement volume of only 15 cm3.
This is accomplished by a unique rotor with impeller blades References
that are wider than the blood flow channel between them,
affording sufficient surface area for passive hydrodynamic 1. Miller LW, Pagani FD, Russell SD, et al. Use of a continuous-flow
device in patients awaiting heart transplantation. N Engl J Med
radial rotor suspension. This eliminates the mechanical pivot 2007;357:885-96.
bearings typical of axial pumps and significantly reduces the 2. Slaughter MS, Rogers JG, Milano CA, et al. Advanced heart failure
pump size by using passive rotor suspension rather than the treated with continuous-flow left ventricular assist device. N Engl J
more complex and larger active electromagnetic bearing Med 2009;361:2241-51.
designs. Also not typical of axial pump design is an outflow 3. Larose JA, Tamez D, Ashenuga M, Reyes C. Design concepts and
principle of operation of the HeartWare ventricular assist system.
port angled 901 from the pump housing inlet port, similar to ASAIO J 2010;56:285-9.
centrifugal pump designs. Various versions of the MVAD are 4. Pagani FD. Continuous-flow rotary left ventricular assist devices with ‘‘3rd
being tested within the LV and at the LV apex in animal generation’’ design. Semin Thorac Cardiovasc Surg 2008;20:255-63.
studies. 5. Morshuis M, Schoenbrodt M, Nojiri C, et al. DuraHeart magnetically
levitated centrifugal left ventricular assist system for advanced heart
Not addressed in this report is the growing interest in
failure patients. Expert Rev Med Devices 2010;7:173-83.
miniaturized intraventricular and intravascular CF rotary 6. Santamore WP, Gray LA Jr. Left ventricular contributions to right
pumps, which have been and are expected to be primarily ventricular systolic function during LVAD support. Ann Thorac Surg
axial-flow pumps. Centrifugal pumps are dependent on 1996;61:350-6.
Moazami et al. Comparison of Axial vs Centrifugal Pumps 11

7. John R, Lee S, Eckman P, Liao K. Right ventricular failure—a 26. Siegenthaler MP, Frazier OH, Beyersdorf F, et al. Mechanical reliability
continuing problem in patients with left ventricular assist device of the Jarvik 2000 Heart. Ann Thorac Surg 2006;81:1752-8: discussion
support. J Cardiovasc Transl Res 2010;3:604-11. 1758–9.
8. John R, Mantz K, Eckman P, Rose A, May-Newman K. Aortic valve 27. Reichenbach SH, Masterson KB, Butler KC, Farrar DJ. Negligible
pathophysiology during left ventricular assist device support. J Heart bearing wear in explanted HeartMate II LVADs following clinical
Lung Transplant 2010;29:1321-9. support for up to four years. In Proceedings of the Annual Meeting of
9. Adamson RM, Dembitsky WP, Baradarian S, et al. Aortic valve the International Societyof Rotary Blood Pumps, Berlin, Germany,
closure associated with HeartMate left ventricular device support: November 2010.
technical considerations and long-term results. J Heart Lung Transplant 28. Drews T, Stepanenko A, Dandel M, Buz S, Lehmkuhl HB, Hetzer R.
2011;30:576-82. Mechanical circulatory support in patients of advanced age. Eur J Heart
10. Cowger J, Pagani FD, Haft JW, Romano MA, Aaronson KD, Kolias TJ.
Fail 2010;12:990-4.
The development of aortic insufficiency in left ventricular assist device-
29. Farrar DJ, Bourque K, Dague CP, Cotter CJ, Poirier VL. Design
supported patients. Circ Heart Fail 2010;3:668-74.
features, developmental status, and experimental results with the
11. Slaughter MS, Bartoli CR, Sobieski MA, et al. Intraoperative
Heartmate III centrifugal left ventricular assist system with a
evaluation of the HeartMate II flow estimator. J Heart Lung Transplant
magnetically levitated rotor. ASAIO J 2007;53:310-5.
2009;28:39-43.
30. Strueber M, O’Driscoll G, Jansz P, Khaghani A, Levy WC,
12. Patterson SW, Starling EH. On the mechanical factors which determine
the output of the ventricles. J Physiol 1914;48:357-79. Wieselthaler GM. Multicenter evaluation of an intrapericardial left
13. Salamonsen RF, Mason DG, Ayre PJ. Response of rotary blood pumps ventricular assist system. J Am Coll Cardiol 2011;57:1375-82.
to changes in preload and afterload at a fixed speed setting are 31. Morshuis M, El-Banayosy A, Arusoglu L, et al. European experience
unphysiological when compared with the natural heart. Artif Organs of DuraHeart magnetically levitated centrifugal left ventricular assist
2011;35:E47-53. system. Eur J Cardiothorac Surg 2009;35:1020-7: discussion 1027–8.
14. Levine BD, Lane LD, Buckey JC, Friedman DB, Blomqvist CG. Left 32. Krabatsch T, Potapov E, Stepanenko A, et al. Biventricular circulatory
ventricular pressure-volume and Frank-Starling relations in endurance support with two miniaturized implantable assist devices. Circulation
athletes. Implications for orthostatic tolerance and exercise perfor- 2011;124:S179-186.
mance. Circulation 1991;84:1016-23. 33. Slaughter MS. Hematologic effects of continuous flow left ventricular
15. Klotz S, Deng MC, Stypmann J, et al. Left ventricular pressure and assist devices. J Cardiovasc Transl Res 2010;3:618-24.
volume unloading during pulsatile versus nonpulsatile left ventricular 34. Heilmann C, Geisen U, Benk C, et al. Haemolysis in patients with
assist device support. Ann Thorac Surg 2004;77:143-9: discussion ventricular assist devices: major differences between systems. Eur J
149–50. Cardiothorac Surg 2009;36:580-4.
16. Haft J, Armstrong W, Dyke DB, et al. Hemodynamic and exercise 35. Esmore D, Spratt P, Larbalestier R, et al. VentrAssist left ventricular
performance with pulsatile and continuous-flow left ventricular assist assist device: clinical trial results and Clinical Development Plan
devices. Circulation 2007;116:I8-15. update. Eur J Cardiothorac Surg 2007;32:735-44.
17. Krabatsch T, Schweiger M, Dandel M, et al. Is bridge to recovery more 36. James NL, Wilkinson CM, Lingard NL, van der Meer AL, Woodard
likely with pulsatile left ventricular assist devices than with JC. Evaluation of hemolysis in the VentrAssist implantable rotary
nonpulsatile-flow systems? Ann Thorac Surg 2011;91:1335-40. blood pump. Artif Organs 2003;27:108-13.
18. Slaughter MS, Pagani FD, Rogers JG, et al. Clinical management of 37. Westaby S, Banning AP, Saito S, et al. Circulatory support for long-
continuous-flow left ventricular assist devices in advanced heart failure. term treatment of heart failure: experience with an intraventricular
J Heart Lung Transplant 2010;29:S1-39. continuous flow pump. Circulation 2002;105:2588-91.
19. Doi K, Golding LA, Massiello AL, et al. Preclinical readiness testing of 38. Siegenthaler MP, Martin J, van de Loo A, Doenst T, Bothe W, Beyersdorf
the Arrow International CorAide left ventricular assist system. Ann F. Implantation of the permanent Jarvik-2000 left ventricular assist device:
Thorac Surg 2004;77:2103-10.
a single-center experience. J Am Coll Cardiol 2002;39:1764-72.
20. Griffith BP, Kormos RL, Borovetz HS, et al. HeartMate II left
39. Jahanyar J, Noon GP, Koerner MM, et al. Recurrent device thrombi
ventricular assist system: from concept to first clinical use. Ann Thorac
during mechanical circulatory support with an axial-flow pump is a
Surg 2001;71:S116-120: discussion S124–6.
treatable condition and does not preclude successful long-term support.
21. Gazzoli F, Alloni A, Pagani F, et al. Arrow CorAide left ventricular
J Heart Lung Transplant 2007;26:200-3.
assist system: initial experience of the cardio-thoracic surgery center in
40. Boyle AJ, Russell SD, Teuteberg JJ, et al. Low thromboembolism and
Pavia. Ann Thorac Surg 2007;83:279-82.
22. Voigt O, Benkowski RJ, Morello GF. Suction detection for the MicroMed pump thrombosis with the HeartMate II left ventricular assist device:
DeBakey left ventricular assist device. ASAIO J 2005;51:321-8. analysis of outpatient anti-coagulation. J Heart Lung Transplant
23. Loforte A, Montalto A, Ranocchi F, et al. Heartmate II axial-flow left 2009;28:881-7.
ventricular assist system: management, clinical review and personal 41. El-Banayosy A, Arusoglu L, Morshuis M, Kizner L, Cobaugh D,
experience. J Cardiovasc Med (Hagerstown) 2009;10:765-71. Koerfer R. Safety of managing rotary blood pumps with a mild
24. Hetzer R, Weng Y, Potapov EV, et al. First experiences with a novel anticoagulation regimen. J Heart Lung Transplant 2008:27.
magnetically suspended axial flow left ventricular assist device. Eur J 42. Garbade J, Bittner HB, Barten MJ, Mohr FW. Current trends in implantable
Cardiothorac Surg 2004;25:964-70. left ventricular assist devices. Cardiol Res Pract 2011;2011:290561.
25. Arndt A, Nusser P, Lampe B. Fully autonomous preload-sensitive 43. Slaughter MS, Sobieski MA 2nd, Tamez D, et al. HeartWare miniature
control of implantable rotary blood pumps. Artif Organs 2010;34: axial-flow ventricular assist device: design and initial feasibility test.
726-35. Tex Heart Inst J 2009;36:12-6.

You might also like