TECHNICAL SPECIFICATIONS OF 500MA X-RAY MACHINE WITH MOTORIZED

TABLE

High frequency X-Ray machine suitable for general radiography and Fluoroscopy

 High frequency X-Ray generator having frequency of 40KHz should be provided


 Power output of generator should be 40KW
 Radiography and Fluoroscopy KV range 40 to 125 in steps
 mA Range (Rad.): Max 500 mA
 mA Range (Flo) : 0-3mA
 Exposure Time (Rad.): 1ms to 10 sec with maximum numbers of steps
Exposure Time (Flo) : 5 minute cumulative

dual action hand switch retractable cord provided

CONTROL

Control should be very compact, soft touch panel having following functions & indications. The panel can
be supplied in floor or wall mount with spill proof design, following features are available on the control
panel

 Machine ON/OFF switch


 Digital Display of KV & mAs
 KV & mAs increase and decrease switches
 Tube focal spot selection switch
 Ready and X-Ray on switch with indicators
 Bucky selection switch
 Self diagnostic Programme with indicators for earth fault error. KV error, filament error & Tube’s
thermal overload

X-RAY TUBE

One no manual collimator for Over Couch and One Number Motorized Collimator for Under
Couch adjustment of exposure area

 Two No of Dual Focus Rotating Anode TOSHIBA X-Ray tube thermally protected
having focal spot 1.0mm small focus. 2.0mm large focus.
 Anode heat storage capacity of tube more than 100KHU.
 Two pair of 6 meter H.V Cable.