You are on page 1of 6

Common artifacts in computerized

tomography: A review

Issa Al-Shakhrah, PhD and Tariq Al-Obaidi, PhD

A
computed tomographic (CT) • Errors in X-ray attenuation measurements; tients and outpatients, but not all CT
image is a display of the • Alterations in the energy spectrum of studies of uncooperative patients from
anatomy of a thin slice of the the X-ray beam (beam hardening) as intensive and coronary care units were
body developed from multiple X-ray it passes through the patient; repeated. The CT systems employed in
absorption measurements made around • The presence of high-density foreign this study included a Siemens Somotom
the body’s periphery. Unlike conven- materials in the body; Plus 4 spiral CT (Siemens Medical Solu-
tional tomography, in which the image • Partial-volume averaging effect; tions, Erlangen, Germany) and a Philips
of a thin slice is created by blurring out • Motion of the patient; Tomoscan AV spiral CT (Philips Med-
the information from unwanted • Quantum mottle (noise); ical Systems, Best, Netherlands).
regions, the CT image is constructed • Malfunction of the detector arising
mathematically, using data arising only from errors in detector calibrations Results
from the section of interest. Generating and balance, geometric effects, or a Repeat CT studies were performed
such an image is confined to cross sec- machine peculiarity; or due to artifacts in 432 patients, and repeat
tions of the anatomy that are oriented • Inadequate temperature, humidity, or CT studies were obtained in 6% of all CT
essentially perpendicular to the axial the presence of small dust particles studies performed during the review
dimension of the body.1 within the computer that causes an inad- period. Major artifacts were found in the
An artifact is any distortion or error equacy in the reconstruction algorithm. form of streaks, rings, and black and
in an image that is unrelated to the sub- This article will present common arti- white bands. Figures 1 through 8 provide
ject being studied. Artifacts are rela- facts such as streaks, rings, and black examples of artifacts on CT scans in dif-
tively common in CT imaging and may and white bands that appear in CT stud- ferent regions and organs in which arti-
be considered as a source, or type, of ies. This review addresses causes of arti- facts can clearly be seen. Figure 9
noise. Their cause may not always be facts, their effects on the quality of the presents the causes of artifacts as well as
obvious. However, there are a number radiographic image, and procedures that the incidence of each artifact found in the
of different effects that may be respon- can be used to reduce the presence of images studied. Regardless of cause,
sible for artifacts in CT. such artifacts. The reduction of artifacts most CT artifacts found in the study were
Because artifacts in CT arise as a enhances CT interpretive accuracy and manifest as streaks in the CT image.
result of the interaction between the helps to establish a correct diagnosis.
subject and the machine, it is useful to Discussion
classify the artifacts by the nature of Method Joseph2 has reported two reasons for
the error made in the scanning process. During a 1-year period, 7197 CT stud- streak artifacts. First, each individual
In CT, artifacts may be produced by: ies were performed in a large (600-bed) measurement involves the evaluation
teaching hospital. Within this total num- of a single ray or straight-line path
Dr. Al-Shakhrah is an Associate Professor ber of studies, 432 repeat studies were through the slice. A second, more sub-
of Medical Physics, University of Jordan, performed due to image artifacts. These tle cause for streak artifacts arises
Amman, Jordan; and Dr. Al-Obaidi is
a Professor of Medical Physics, Mu’tah repeat studies were collected and classi- when there is an abrupt discrepancy or
University, Al-Karak, Jordan. fied according to the cause of the artifacts inconsistency between views, as might
viewed. The study included both inpa- be seen with patient motion.

August 2003 www.appliedradiology.com APPLIED RADIOLOGY ©


■ 25
COMMON CT ARTIFACTS

The following discussion will


A B
address our results and their causes,
according to each artifact etiology.

Motion
Patient motion has a devastating
effect on image quality. This was the pri-
mary reason for the development of a
body unit (spiral CT) that could com-
plete a scan during a patient’s breath-
hold.
The artifacts found in our study
caused by motion were manifested as
black or white bands, dark spots, loss
of resolution, or distortion of anatomy3
(Figure 1). These artifacts account for
C D
15% of repeated studies. Clinically,
such artifacts are important not only
because they degrade image quality,
but also because they can sometimes
be mistaken for pathologic changes,
such as bronchiectasis.4
Theoretically, motion artifacts can
be reduced by fast scanning,5 gating,6
tube alignment,7 corrective reconstruc-
tion,8 or postprocessing of the scan.9
Knowing the pattern, magnitude,
and frequency of motion in advance
would allow the use of an algorithm to
remove the motion from the projection
data and then reconstruct a motion-free
or, at least, a motion-reduced image.10
A technical problem is image mis-
FIGURE 1. CT artifacts caused by motion.
(A) and (B) Black or white bands and dark registration due to variation in breath-
spots appear on brain CT scans. (C) Loss of holding from one scan to the next.11
resolution appears on sinus CT (coronal view). Misregistration leads to failure to
(D) Distortion of anatomy appears on brain CT. image part of the lung, such that a
lesion might be partially or entirely
excluded on the final scan. This prob-
lem is eliminated by using spiral CT
during breath-holding.12,13

X-ray geometry, geometric oversights


Precise control of fan-beam position
is important for the production of high-
quality CT images; failure to maintain
fan-beam orientation can produce image
artifacts.14 Performance of CT systems
relies mainly on geometric precision and
measurement quality. Inaccurate geome-
FIGURE 2. Black and white
try, inaccurate alignment of the X-ray
bands and blurring artifacts
on lung CT were produced tube with the detectors, or incorrect data
by inaccurate X-ray tube can produce artifacts and blurring that
geometry. limit spatial resolution (Figure 2). In our

26 ■ APPLIED RADIOLOGY ©
www.appliedradiology.com August 2003
COMMON CT ARTIFACTS

FIGURE 3. Ring artifact appears on brain CT caused by inaccurate FIGURE 4. Multiple artifacts due to misalignment of X-ray tube and
matching and inaccurate intercalibration of the detectors. detectors, foreign heavy metal on the film, and beam hardening
(posterior fossa near the petrous bones).

study, artifacts due to geometric causes Beam hardening more pronounced with the iodinated con-
accounted for 5% of repeat studies. A Beam-hardening artifacts result from trast when compared with gadopentetate
precise geometric calibration proce- the preferential absorption of low- dimeglumine (Gd-DTPA).
dure is required, and some corrections energy photons from the beam. The
must be applied to the raw attenuation effect is more pronounced in areas of High-density foreign material
data in order to obtain accurate mea- large attenuation, such as bone. The The presence of objects that have an
surements. An X-ray cone-beam CT artifact is seen as a shadow beneath ribs, exceptionally high or low attenuation
system has been developed by Rizo for example, or increased shadows in can create streaking artifacts by forcing
and Martin.15 The machine was the mediastinum or skull (Figure 4). the detectors to operate in a nonlinear
designed to control small parts limited This type of artifact accounted for 21% response region.1 Figure 5 demonstrates
to a few centimeters, with a high spa- of the repeat scans in our study. This the star pattern caused by high-density
tial resolution close to 30 microns. effect occurs throughout the image but foreign materials. In our study, this type
They introduced the machine setup and usually is not perceived except where of artifact caused 25% of repeat studies,
described the calibration computing there is a great deal of hardening, such one of the most common reasons for
resources involved in the system. They as in the vicinity of bone. This effect can image degradation. A small metal frag-
also discussed the performance on be compensated for by the use of special ment produces a star pattern, and the star
experimental data.15 filters or a special correction algorithm.1 effect is accentuated by any motion. The
Recent experiments using CT and only way to avoid this problem with cur-
Detector artifacts transmission radiography show that rent mathematical reconstructions is to
Artifacts from errors in detector cal- gadolinium (Gd) agents can increase change the angle of the slice to exclude
ibrations and balance are common. A image contrast by up to a factor of 2 the foreign body, but this approach
malfunction of any one detector would when compared with more commonly might also exclude pathology. A similar
incorrectly backproject along the data used iodinated agents on an equi-molar pattern can be produced by gas, for
ring to produce the artifacts. If detec- basis. It has also been suggested that example, in the gastric fundus, but the
tors are not matched or intercalibrated beam-hardening artifacts may be re- effect is less marked.4
accurately, the backprojection for each duced with the use of Gd. This hypothe-
data ring would be slightly different, sis was tested by Ruth and Joseph16 on Partial-volume averaging
causing multiple rings (Figure 3). three different CT scanners using a circu- When tissues of widely different
These artifacts caused by equipment lar water equivalent phantom with a con- absorption occupy the same voxel, the
malfunction can be eliminated by trast-filled tube inserted. It was found beam attenuation is proportional to the
repair or good preventive maintenance. that the artifacts were 1.3 to 1.8 times average value of the attenuation

August 2003 www.appliedradiology.com APPLIED RADIOLOGY ©


■ 27
COMMON CT ARTIFACTS

A B C

D E F

FIGURE 5. CT heavy streaks or star artifacts. Metallic fillings in teeth, (A) axial view and (B) coronal view. (C) Metallic filling in a tooth appears
on a CT of the sinuses (coronal view). (D) Metallic stent in the heart appears on a chest CT. (E) Surgical clips (prosthesis) in the common intra-
hepatic bile duct appear on abdominal CT. (F) Surgical clips around the intracavernous part of the left internal carotid artery appear on brain CT.

A B C

FIGURE 6. (A through C) The skull appears thicker on successively more cephalad scans, because the orientation of the bone becomes more
oblique and less perpendicular to the scan section (partial-volume averaging effect).

coefficient of the voxel. A volume aver- increased slice thickness when com- Quantum mottle (noise) artifacts
age is computed for such voxels, leading pared with conventional CT scanning.18 In a study of 20 examiniations, soft-
to the partial-volume error. The scan of Partial-volume effects on measurements tissue imaging degraded by scattering
the skull-brain shown in Figure 6 of CT numbers may be minimized by artifacts was reported in 14 examina-
demonstrates this effect. Images gener- the use of thin sections and by the selec- tions.20 Quantum mottle is dose-related
ated with helical scanning are degraded tion of a section that lies in the center of image noise that has the appearance
by partial-volume artifacts caused by an the object of interest for measurement.19 of granular steaks arising from high-

28 ■ APPLIED RADIOLOGY ©
www.appliedradiology.com August 2003
COMMON CT ARTIFACTS

A B

FIGURE 7. Effects of quantum mottle. (A) Heavy black and white bands appear on pituitary gland (coronal view). (B) Foggy pelvis CT.

B attenuation structures, such as the shoulder


A
region. A generalized adaptive median fil-
ter (GAMF) was introduced by Hsieh21 for
more robust noise suppression and edge
preservation in CT to combat severe
streaking artifacts resulting from excessive
X-ray quantum noise (Figure 7).

Temperature and humidity


Some computer components are more
sensitive to extremes of temperature and
humidity than is conventional X-ray
equipment. For this reason, manufac-
turer’s recommendations regarding air
conditioner installation should be
FIGURE 8. (A) and (B) Black and white bands and less resolution appear on two different closely followed. Whenever possible, a
slices of brain CT due to malfunction of the CT computer caused by inadequate temperature
regulation in the computer room.
backup air-conditioning system should
be available. The low-temperature limits
below which solid-state devices cannot
operate appropriately will probably
never be encountered in a hospital
installation. Failure from moderately
increased temperature, as well as
humidity, will frequently take the form
of unexplained malfunctions of the com-
puter,19 including increasing numbers of
artifacts and inappropriate responses to
instructions (Figure 8).
One aspect of the computer environ-
ment that must be considered is dust par-
ticle size. The particle size of cigarette
smoke is only slightly larger than the
size of the air gap between the playback
head and disk drive of the computer sys-
tem. In addition, its relatively small size
FIGURE 9. Classification of the repeated studies (artifacts) according to the following causes, makes it particularly difficult to filter
from left to right: motion; X-ray geometry, geometric oversights; detectors; beam hardening; from the atmosphere. For this reason,
high-density foreign materials; partial-volume averaging; quantum mottle (noise); and tempera- smoking in the computer room should
ture and humidity. be strictly forbidden.

August 2003 www.appliedradiology.com APPLIED RADIOLOGY ©


■ 29
COMMON CT ARTIFACTS
10. Cameron JR, Godwin JD, Crawford CR, et al.
Conclusion Acknowledgment Minimum scan speeds for suppression of motion
In our study, CT artifacts were found The authors thank Mr. Jameel Al- artifacts in CT. Radiology. 1992;185:37-42.
to be produced by: the presence of high- Sarayrah for his assistance in collect- 11. Naidich DP, Zerhouni EA, Siegelman SS. Prin-
ciples and techniques of thoracic CT and MR. In:
density foreign materials in the body ing CT studies and in classifying Computed Tomography and Magnetic Resonance
(25%); an error in X-ray attenuation imaging artifacts according to etiology. Imaging of the Thorax. 2nd ed. New York, NY:
measurements, as a result of alterations Raven; 1991:1-34.
12. Kalender WA. Technical foundations of spiral
in the energy spectrum of the X-ray
beam (beam hardening) as it passes
REFERENCES CT. Semin Ultrasound CT MRI. 1994;15:81-89.
1. Haaga JR, Alfidi RJ. Computed Tomography of 13. White CS, Romney BM, Andrew C, et al. Primary
through the patient (21%); partial- the Whole Body. 2nd ed. St. Louis, MO: The C.V. carcinoma of the lung overlooked at CT: Analysis of
Mosby Company; 1988. findings in 14 patients. Radiology. 1996;199:109-115.
volume–averaging effects (16%); 14. Gard, Michael F. Real-time X-ray fan-beam z-
2. Joseph PM. Artifacts in computed tomography.
motion of the patient (15%); quantum In: Newton TH, Potts DG, eds. Radiology of the axis position measurement. IEEE Trans Instr
mottle (noise) (7%); malfunction of the Skull and Brain, Part XVI, General Theory of Com- Meas. 1994;43:295-298.
puted Tomography. Vol. 5. Technical Aspects of 15. Rizo SP, Martin P. X-ray cone-beam CT system
detector arising from errors in detector calibration. Proceedings of SPIE—The International
Computed Tomography. St. Louis, MO: The C.V.
calibrations and balance (6%); geomet- Mosby Company; 1981:3956-3992. Society of Optical Engineering. 1993;229-239.
ric effects or a machine peculiarity (5%); 3. Mayo RJ, Muller NL, Henkelman RM. The dou- 16. Ruth C, Joseph PM. A comparison of beam-
ble-fissure sign: A motion artifact on thin-section hardening artifacts in x-ray computerized tomogra-
and by inadequate temperature, humid- phy with gadolinium and iodine contrast agents.
CT scan. Radiology. 1987;165:580-581.
ity, or the presence of small dust parti- 4. Tarver RD, Conces DJ, Godwin JD. Motion arti- Med Phys. 1995;22:1977-1982.
cles within the computer causing an facts on CT simulate bronchiectasis. AJR Am J 17. Curry TS, Dowdey JE, Murry RC. Chris-
Roentgenol. 1988;151:1117-1119. tensen’s Physics of Diagnostic Radiology. 3rd ed.
inaccuracy in the reconstruction algo- Philadelphia, PA: Lea and Febiger; 1984.
5. Goldberg HI, Gould RG, Feuerstein IM, et al.
rithm (5%). Regardless of the causes of Evaluation of ultrafast CT scanning of the adult 18. Grawford CR, King KF, Toth TL, Hu H. Moving
the artifacts found in this study, most CT abdomen. Invest Radiol. 1989;24:537-543. beam helical CT scanning. IEEE Trans Med Imag-
6. Moore SC, Judy PF, Garmic JD. Prospectively ing. 1996;15:188-196.
artifacts manifested as streaks. Films 19. Morgan CL, Miller MD. Basic Principles of
gated cardiac computed tomography. Med Phys.
that were repeated due to artifacts 1983;10:846-855. Computed Tomography: Theory and Techniques
accounted for 6% of the total 7197 films 7. Grawford CR, Pelc NJ. Method for reducing of Reconstruction. Baltimore, MD: University Park
motion induced image artifacts in projection imag- Press; 1983.
taken during the 1-year period. Patient 20. Dietrich U, Kalff R, Sturmer KM, et al. Comput-
ing. U.S. patent 4,994,965. 1991.
cooperation, use of thin sections, repair 8. Grawford CR, King KF, Ritchie CJ, Godwin JD. erized tomography after internal fixation of the
and/or good preventive maintenance, a Respiratory compensation in projection imaging spine. Neurosurg Rev. 1989;12:211-215.
[abstract]. Radiology. 1990;177:277. 21. Hsieh J. Generalized adaptive median filters
clean computer environment, and suit- and their application in computed tomography.
9. Helenon O, Chanin DS, Laval-Jeantet M, Frija J.
able temperature and humidity can Artifacts on lung CT scan: Removal with Fourier fil- Proceedings of SPIE—The International Society of
reduce artifacts to a minimum. AR tration. Radiology. 1989;171:572-574. Optical Engineering. 1994;662-672.

AIR FORCE HEALTHCARE.


GOOD PAY.
PROFESSIONAL RESPECT.
Experience the best of everything: best facilities, best benefits. Outstanding opportunities

for travel, training, advancement and 30 days of vacation with pay. For an information

packet, call 1-800-423-USAF or visit airforce.com. CROSS INTO THE BLUE

30 ■ APPLIED RADIOLOGY ©
www.appliedradiology.com August 2003

You might also like