You are on page 1of 10

IP 585/10

Determination of fatty acid methyl esters (FAME), derived


from bio-diesel fuel, in aviation turbine fuel — GC-MS with
selective ion monitoring/scan detection method

1 Scope 2 Normative references

This document specifies a method for the The following referenced documents are
identification and quantification in the range 4,5 indispensable for the application of this document.
mg/kg to 150 mg/kg of selected fatty acid methyl For dated references, only the edition cited applies.
ester (FAME) species, see Table 1, in aviation For undated references, the latest edition of the
turbine fuel (AVTUR). referenced document (including any amendments)
applies.
NOTE 1 - AVTUR (and AVTUR contaminated with diesel/
biodiesel) can contain some high molecular weight
IP 365, Determination of density— Oscillating
naphtha components which may mask the FAME at U-tube method (EN ISO 12185)
very low levels. Low carbon number methyl esters, such IP 367, Determination and application of precision
as those derived from coconut oil, may be subject to
data in relation to methods of test
interference from AVTUR components and may not be
quantified at low levels. IP 475, Petroleum Liquids—Manual
sampling (EN ISO 3170)
NOTE 2 - The FAME species listed in Table 1 have been
specifically chosen as they typically make up greater than 3 Principle
95 % of the composition of the major biofuel feedstocks
currently used in biodiesel blends with conventional
The test portion of AVTUR is run neat together
mineral diesel. In addition methyl heptadecanoate is
included as it may be present in animal fats.
with an internal standard on the GC-MS with a long
polar column to separate the polar FAME species
NOTE 3 - In the round-robin study, two samples were relative to the non-polar hydrocarbon matrix of
blended containing no FAME. No positive biases were the jet fuel and any mineral diesel contamination.
found in either the Merox treated or hydrotreated fuels Due to natural background of high end naphtha
used in the round robin. components present in some samples, complete
WARNING – The use of this standard may involve separation is not possible and a multiple selective
hazardous materials, operations and equipment. ion monitoring (SIM) is performed. The ions used in
This standard does not purport to address all of the SIM method are indicative of the FAME species
the safety problems associated with its use. It is and not the background hydrocarbon matrix of the
the responsibility of the user of this standard to fuel. The SIM method is used to measure the levels
establish appropriate safety and health practices of the specified FAME species in the sample by
and determine the applicability of regulatory reference to an external standard calibration of the
limitations prior to use. specified FAME species in dodecane.

Table 1: Fatty acid methyl esters determined

Fatty acid methyl ester Molecular formula Symbol used

methyl hexadecanoate (methyl palmitate) C17H34O2 C16:0


methyl heptadecanoate (methyl margarate) C18H36O2 C17:0
methyl octadecanoate (methyl stearate) C19H38O2 C18:0
methyl octadecenoate (methyl oleate) C19H36O2 C18:1
methyl octadecadienoate (linoleate) C19H34O2 C18:2
methyl octadecatrienoate (linolenate) C19H32O2 C18:3

585.1

Issued under licence for use by Intertek employees only.


IMPORTANT: This file is subject to a licence agreement issued by the Energy Institute, London, UK. All rights reserved. It may only be used in accordance with
the licence terms and conditions. It must not be forwarded to, or stored or accessed by, any unauthorised user. Enquiries: e: pubs@energyinst.org.uk t: +44 (0)207 467 7100
FAME IN AVTUR – GC-MS, IP 585

Variations in density between the AVTUR NOTE 3 - The SIM ion data is used for the quantitative
sample and dodecane are taken into account measurements and the scan mode total ion chromatogram
in the final result calculation. The mixed FAME (TIC) data to aid identification of species.
calibration is run at the same time as the test
portion and a quality check standard is run 5.2 GC polar capillary column HP INNOWAX
every five samples to check for calibration drift. (200 µm x 0,4 µm film) open tubular capillary
column of 50 to 60 m length, with a helium
NOTE - The role of the internal standard is to minimise carrier gas flow of approximately 0,6 ml/min.
any effects due to instrumental drift.
5.3 Analytical balance, capable of weighing to
4 Reagents and materials ± 0,0001 g.

4.1 Fatty acid methyl esters, as detailed 5.4 Conical glass flask fitted with a ground
in Table 1, with a purity of 99 % or greater glass closure, nominal capacity 250 ml.
for preparation of FAME calibration solutions.
5.5 10 ml volumetric flask, Grade A.
NOTE - These compounds can be obtained commercially
from a range of chemical suppliers. It is recommended 5.6 Dropping pipette.
that individual FAME compounds are used.
5.7 Calibrated adjustable autopipettes, capable
4.2 Helium carrier gas. of delivering 10 to 100µl and 100 to 1 000 µl

4.3 Dodecane, purity > 90 %. 5.8 GC Autosampler vials (1,5-2 ml) with crimp
caps.
4.4 Cyclohexane, purity > 99 %.
NOTE – Screw-cap vials with septa may also be used.
4.5 Methyl heptadecanoate-d33 (methyl
margarate-d33) – internal standard, > 98 atom % 5.9 1 ml volumetric pipette, Grade A.
atom deuterated.
5.10 Syringe, glass, 10 µL capacity.
5 Apparatus
5.11 100 ml volumetric flask, glass, Grade A.
NOTE – It is strongly recommended that dedicated
glassware is used for this method to avoid cross- 6 Sampling
contamination.

6.1 Unless otherwise specified, take


5.1 GC-MS, capable of operating in an
a sample of at least 60 ml in accordance
electron impact ionisation mode, ideally with
with IP 475 and/or in accordance with the
simultaneous SIM/SCAN data collection, and
requirements of national standards or regulations
equipped with a split/splitless or cold on-column
for the sampling of petroleum products.
injector. If the instrument is capable of running
simultaneous SIM and full scan modes then the 6.2 Use dedicated aviation turbine fuel sampling
full scan total ion chromatogram (TIC) is used in equipment.
the identification of the individual FAME species.
6.2.1 Ensure that gloves are clean prior to
NOTE 1 - It is strongly recommended to use an injection sampling and that cross-contamination from the
packed liner. gloves can not occur.
NOTE 2 - If the instrument cannot run simultaneous SIM/ 6.2.2 Sample lines shall be flushed prior to taking
SCAN then it is recommended that the samples are run samples to ensure that there is no cross contamination
twice, once in SIM mode for quantification and then again
of the sample from material previously sampled.
in full scan mode if positive identification of detected
FAME species by spectral matching is required.
6.3 Use amber glass or epoxy-lined metal
The instrument used must be capable of detecting containers with inert closures.
0,5 mg/kg of each of the specified FAME species in NOTE – ASTM D4306 provides guidance in Section
dodecane with a signal to noise ratio of at least 10:1 6 on a procedure that may be used for testing the
and shall be equipped with data collection and analysis suitability of sample containers for tests affected by
software to allow accurate peak area measurement. trace contamination.

585.2

Issued under licence for use by Intertek employees only.


IMPORTANT: This file is subject to a licence agreement issued by the Energy Institute, London, UK. All rights reserved. It may only be used in accordance with
the licence terms and conditions. It must not be forwarded to, or stored or accessed by, any unauthorised user. Enquiries: e: pubs@energyinst.org.uk t: +44 (0)207 467 7100
FAME IN AVTUR – GC-MS, IP 585

6.3.1 Used containers are permitted provided it 7.2.1 Place the conical flask (5.4) onto the
can be confirmed they have only been used for balance (5.3) and tare. Using a dropping pipette
aviation turbine fuel containing <5 mg/kg FAME. (5.6) place 0,1000 g ± 0,0010 g of each FAME
species into the flask and record the masses added.
NOTE - New sample containers are strongly recommended Add dodecane (4.3) to give a total mass of 100 g
due to concerns over the difficulty in removing all traces ± 0,05 g. Stopper the flask and mix the contents.
of FAME retained from previous samples.
7.2.2 Transfer the BCS to a tightly closed glass
6.3.2 Rinse all sample containers and their closures container and store in a refrigerator held at 4 ºC
with the product to be sampled at least three times. +/- 2 ºC when not in use and use within three months.
Each rinse shall use product with a volume of Before use, examine for any phase separation or
10 % to 20 % of the container volume. Each rinse discoloration, shake vigorously, and let stand to allow
shall include closing and shaking the container for a for removal of air bubbles. Discard the standard if it
minimum of 5 seconds and then draining the product. shows sediment, phase separation or discoloration.

7.2.3 Calculate the stock solution concentration


7 Internal standard and calibration for each FAME component (Y) using equation [1].
standards preparation
Conc (mg/kg) of FAME species Y in BCS =
7.1 Internal standard solution (ISS) preparation
Prepare a stock internal standard solution containing mass of FAME species Y (g) x 106 [1]
nominally 1 000 mg/L of methyl heptadecanoate- total mass of BCS (g)
d33 internal standard as follows.
7.3 Working standard solution (WSS) and
7.1.1 Place a 100 ml volumetric flask (5.11) working calibration standards
onto the balance (5.3) and tare. Using a dropping
pipette (5.6), add 100 mg +/- 0,5 mg of methyl 7.3.1 Prepare a nominal 100 mg/kg working
heptadecanoate-d33 (4.5) and make up to 100 standard solution (WSS) by pipetting 1 000 µl of
ml with dodecane (4.3) to give a 1 000 mg/L BCS into the 10 ml volumetric flask and dilute
solution. to 10 ml with dodecane. Calculate the working
standard solution (WSS) concentration for each
7.1.2 Store 1 000 mg/L standard in a tightly closed FAME component (Y) using equation [2].
glass container in a refrigerator held at 4 ºC +/- 2 ºC
when not in use and use within three months. Before
Conc (mg/kg) of FAME species Y in WSS =
use, examine for any phase separation or discoloration,
shake vigorously, and let stand to allow for removal of
air bubbles. Discard the standard if it shows sediment, conc of FAME species Y in BCS [2]
phase separation or discoloration. 10
7.2 Bulk calibration solution (BCS) preparation 7.3.2 Prepare a set of working calibration
Prepare a bulk calibration solution (BCS) containing standards containing nominally 2, 4, 6, 8, 10,
nominally 1 000 mg/kg each of all of the FAME 20, 40, 60, 80 and 100 mg/kg of each reference
reference compounds listed in Table 1 in dodecane. compound by dilution of the WSS (7.3.1)

Table 2: Volumetric Dilutions for preparing standards in the range 0 – 100 mg/kg from the 100 mg/kg WCS
Nominal std concentration (mg/kg) Volume WSS (µl) Volume dodecane Volume internal standard
(µl) (µl)
100 1 000 0 10
80 800 200 10
60 600 400 10
40 400 600 10
20 200 800 10
10 100 900 10
8 80 920 10
6 60 940 10
4 40 960 10
2 20 980 10
0 0 1 000 10

585.3

Issued under licence for use by Intertek employees only.


IMPORTANT: This file is subject to a licence agreement issued by the Energy Institute, London, UK. All rights reserved. It may only be used in accordance with
the licence terms and conditions. It must not be forwarded to, or stored or accessed by, any unauthorised user. Enquiries: e: pubs@energyinst.org.uk t: +44 (0)207 467 7100
FAME IN AVTUR – GC-MS, IP 585

in dodecane. This can be done directly into Sample injection


autosampler vials (5.8) using volumetric dilution of Splitless injector, temperature 260 ºC.
the 100 mg/kg working standard as listed in Table Sample volume 1 µl.
2. In addition add 10 µL of the internal standard Injector purge flow 100 ml/min 2 min after
solution (7.1) to each vial using the glass syringe injection.
(5.10) and then immediately seal each autosampler Autosampler syringe wash solvent - cyclohexane.
vial with a crimp cap (5.8) and mix thoroughly.
The sample injection parameters such as sample
If the accuracy of the autopipettes are not within the pumps and syringe solvent washes pre-and post-
manufacturers specified tolerance then all volumes injection are instrument dependent and shall be
shall be weighed and masses used instead of volume. chosen to eliminate sample carryover between
sequential injections of the highest standard and a
7.3.3 Calculate the actual concentration of each blank dodecane sample (see Section 5.1 note 1).
of the FAME species (Y) in the final standard
solutions using equation [3]. Data collection solvent delay
In order to prolong filament life employ a solvent
Conc of FAME species Y (mg/kg) = delay to allow the bulk of the sample matrix
to elute before the detector is switched on.
(mg/kg of FAME species Y in WSS) x (volume of WSS (µl))
1 000 NOTE 2 - A delay of 20 min after sample injection was
[3] typical using the conditions specified above in this
method.
NOTE - The calculations used in the preparation of
standards are suitable for a spreadsheet application and 8.2 SIM ion window setup The elution order and
generation of a suitable spreadsheet application for this
approximate elution timings of the standard FAME
is recommended.
species used in this method are listed in Table 3.
Typical chromatograms for calibration standards
8 Apparatus preparation and fuels are shown in Annex A.

8.1 GC-MS setup Follow the manufacturer’s NOTE - The values listed in Table 3 are those for
instructions for setting up the apparatus. Instrument elution order and approximate retention times for FAME
manufacturers differ in how their instruments are standards using the conditions and Innowax column
set up and therefore it is not possible to publish detailed in this method.
a specification for all manufacturers at this time.
In some FAME feeds two peaks have been seen
NOTE 1 - For some instruments the following settings for C18:1 which elute within half a minute
have given satisfactory results. However, settings may of each other. These have been attributed
have to be optimised for specific instruments. to cis and trans isomers and the peak areas
shall be added together and counted as one.
GC oven temperature programme Determine the retention time windows for each
Initial temp 150 ºC and hold for 5 min. FAME component in Table 3 by injecting the
Ramp at 12 ºC per min until 200 ºC and hold for 100 mg/kg standard in full scan mode and record
17 min.
Ramp at 3 ºC per min until 252 ºC and hold for
6,5 min.
Total run time 50 min.
Table 3: Elution order and approximate elution timing of FAME species
Fatty acid methyl ester Approximate retention time Symbol
(min)
methyl hexadecanoate (methyl palmitate) 24,9 C16:0
methyl heptadecanoate-d33 (methyl margarate-d33) 28,4 C17:0 (-d33)

methyl heptadecanoate (methyl margarate) 30,1 C17:0


methyl octadecanoate (methyl stearate) 34,7 C18:0
methyl octadecenoate (methyl oleate) 35,5 -36,1 (see below) C18:1
methyl octadecadienoate (linoleate) 37,7 C18:2
methyl octadecatrienoate (linolenate) 40,3 C18:3

585.4

Issued under licence for use by Intertek employees only.


IMPORTANT: This file is subject to a licence agreement issued by the Energy Institute, London, UK. All rights reserved. It may only be used in accordance with
the licence terms and conditions. It must not be forwarded to, or stored or accessed by, any unauthorised user. Enquiries: e: pubs@energyinst.org.uk t: +44 (0)207 467 7100
FAME IN AVTUR – GC-MS, IP 585

the retention time of each. The scan range for TIC 10 Calibration data analysis
full scan identification shall cover the range m/z
33 to 320 amu. Set up the SIM data collection 10.1 Calibration standard peak identification
windows for each component in the instrumental Identify the peaks resulting from the working
method using the designated ions in Table 4. calibration solutions by reference to their relative
retention times using the data obtained from the
The SIM ions selected to give maximum SIM ion window setup injection (8.2). Confirm
selectivity and sensitivity for the specified identification by comparing the SCAN TIC spectra
species are detailed in Table 4. For each FAME for each peak to the pure component spectra,
species the summed responses for all SIM either produced in-house, or with a recognised
ions specified for that particular FAME are mass spectral database such as produced by NIST.
employed in the calibration and quantification.
10.2 Calibration graph preparation
Table 4: SIM ions to be employed for quantification Measure the area of each individual FAME
of specified FAME species determined component from the combined SIM ion trace using
the data package peak integration facility. Calculate
Species to be SIM ions to be used for the FAME to internal standard peak area ratio (R) for
detected quantification each individual FAME species using equation [4].
C16:0 227, 239, 270, 271
C17:0 (d33)Int std 317 R=measured peak area for the detected FAME components
C17:0 241, 253, 284 measured peak area for the internal standard
C18:0 255, 267, 298 [4]
C18:1 264, 265, 296
NOTE 1 - Many instrument software packages include
C18:2 262, 263, 264, 294, 295
an option to use internal standard for calibration and
C18:3 236, 263, 292, 293 calculation of concentrations. This may be used to
automatically calculate calibration curves and results in
9 Procedure place of equation 4.

9.1 Add 1,00 ml of each test sample  for Two calibration plots for each FAME component
analysis to a glass autosampler GC  vial (5.8) covering the working ranges 0 mg/kg to 10 mg/kg
using a volumetric pipette (5.9) or calibrated (using the 0, 2, 4, 6, 8, 10 mg/kg standards) and
autopipette (5.7). To each vial, add 10 µl of internal 0 mg/kg to 100 mg/kg (using the 0, 20, 40, 60,
standard stock solution (7.1) using a syringe (5.10), 80, 100 mg/kg standards) are used in this method.
add cap, crimp  and shake well prior to analysis.
Produce calibration plots for both working ranges
9.2 Place the required number of calibration by plotting peak area ratio R (from equation 4)
standards and test samples and a dodecane versus individual FAME component concentrations
blank into the autosampler vials so that the test and force through zero (verified by the fact that
portions are run immediately after the standards. no peaks were detected in the 0 mg/kg blank
If more than five test portions are to be analysed dodecane). Ensure that the plots are linear for each
in a batch, the 2 mg/kg standard shall be run after component and calculate a calibration equation.
every five samples as a quality control standard
to check for variation in detector response. The correlation coefficients for all the fitted
calibration shall be greater than 0,985. If
this is not so, the calibrations shall be re-run.
9.3 Re-run the full set of calibration standards
after 25 test portions have been analysed or if NOTE 2 - In practice correlation coefficients greater than
the quality control standard peak area ratio for 0,995 should be achievable.
C18:0 (10.2) deviates by more than 5 % from the
initial calibration. If a quality control standard falls The fitted calibration line for each FAME species
outside the 5 % tolerance then the results from the will have the equation [5]:
previous five sample solutions shall be discarded
and the samples re-run using the new calibration. R = slope x concentration of FAME species in mg/kg

585.5

Issued under licence for use by Intertek employees only.


IMPORTANT: This file is subject to a licence agreement issued by the Energy Institute, London, UK. All rights reserved. It may only be used in accordance with
the licence terms and conditions. It must not be forwarded to, or stored or accessed by, any unauthorised user. Enquiries: e: pubs@energyinst.org.uk t: +44 (0)207 467 7100
FAME IN AVTUR – GC-MS, IP 585

11 Sample data analysis If the density of the FAME species and dodecane are
not known, determine the densities in accordance
Identify the individual FAME species in each with IP 365.
sample using the relative SIM ion intensities and
retention time for each species relative to the 12 Expression of results
standards. In addition to the ion patterns of the
measured ion species other qualifying ions are Report the total FAME content in mg/kg
used to positively identify specific FAME species. to the nearest 0,1 mg/kg. Individual FAME
component concentrations can also be reported
NOTE - The quantitative SIM ions are used to create the if required for FAME source identification.
calibration data however other qualifying ions may be
used to help verify the presence of a particular FAME
species in the jet fuel background that may co-elute with 13 Precision
the FAME. These ions can be selected from the spectra
obtained from the pure component standards in full scan 13.1 General
mode. The precision values given in 13.2 and 13.3 were
derived from statistical analysis, by IP 367, of data
After the sample peaks have been identified and obtained during a round robin carried out in 2009
qualified using a combination of retention time, using 12 instruments/operators and a sample set
SIM ion pattern and other qualifying ions, measure comprising 16 jet fuel samples in duplicate.
the peak areas of all the FAME components
which have been detected in each sample. NOTE – Data from three samples with nominal values of 0
and 1 mg/kg were removed during the statistical analysis
Calculate the concentration of each individual and together with additional outliers, this resulted in the
FAME species from the relevant calibration loss of >20 % of data.
graph depending on the level of FAME detected
in accordance with equation [6]. The lower 13.2 Repeatability, r
working range shall be employed for all FAME The difference between two test results
species present in any sample at less than obtained by the same operator with the same
10 mg/kg. The 0 mg/kg to 100 mg/kg calibration apparatus under constant operating conditions
range shall be employed for all FAME species on identical test material, would in the long run,
present in any sample at greater than 10 mg/kg. in the normal and correct operation of the test
method, exceed the value below in only one
case in 20. Table 5 shows calculated values.
Concentration FAME component = R / slope of calibration [6]

r = 0,1632 (X+3)
The total FAME concentration shall be obtained
by summing the concentrations obtained for the where X is the average of the two results being
six specified FAME species using the specified compared, in mg/kg.
calibration ranges.
13.3 Reproducibility, R
As the GC injector injects a fixed volume of sample, The difference between two single and
the density of the samples will vary and differ independent test results, obtained by different
from the dodecane used to prepare the standards, operators working in different laboratories on
a density correction factor shall be applied to identical test material, would in the long run, in
the concentrations calculated from the standard the normal and correct operation of the test
calibration in accordance with equation [7]: method, exceed the value below in only one
case in 20. Table 5 shows calculated values.
Total FAME, mg/kg =
R = 0,2579 (X+3)
(measured FAME, mg/kg) x (density of dodecane)
(sample density) where X is the average of the two results being
[7] compared, in mg/kg.

585.6

Issued under licence for use by Intertek employees only.


IMPORTANT: This file is subject to a licence agreement issued by the Energy Institute, London, UK. All rights reserved. It may only be used in accordance with
the licence terms and conditions. It must not be forwarded to, or stored or accessed by, any unauthorised user. Enquiries: e: pubs@energyinst.org.uk t: +44 (0)207 467 7100
FAME IN AVTUR – GC-MS, IP 585

Table 5: Tabulated repeatability (r) and reproducibility 14 Test report


(R)
The test report shall include at least the following
Level of result r (mg/kg) R (mg/kg) information:
(mgkg)
5,0 1,3 2,1 a) Reference to this standard.
10,0 2,1 3,4 b) Type and identification of the product under
20,0 3,8 5,9 test.
30,0 5,4 8,5 c) Result of the test (see clause 12).
40,0 7,0 11,1 d) Any deviation from the procedure described.
50,0 8,7 13,7 e) Date of the test.
60,0 10,3 16,3
75,0 12,7 20,1
100 16,8 26,6
140 23,3 36,9

NOTE - In the round-robin study, three Merox based


samples were included containing nominally 0, 30 and
100 mg/kg FAME. No positive biases were found in
either the Merox treated or hydrotreated fuels.

585.7

Issued under licence for use by Intertek employees only.


IMPORTANT: This file is subject to a licence agreement issued by the Energy Institute, London, UK. All rights reserved. It may only be used in accordance with
the licence terms and conditions. It must not be forwarded to, or stored or accessed by, any unauthorised user. Enquiries: e: pubs@energyinst.org.uk t: +44 (0)207 467 7100
FAME IN AVTUR – GC-MS, IP 585

Annex A
(informative)

Examples of chromatograms
A.1 Example of 2 mg/kg and 0 mg/kg FAME standards overlaid
(Innowax J&W 60 m X 0,25 mm 0,5 µm film)

Abundance
C18:0
D33 IS
12000
C17:0 C17:0
11000
C16:0
10000
C18:1
9000

8000
C18:2
7000
C18:3
6000

5000

4000

3000

2000

1000

0
24.00 26.00 28.00 30.00 32.00 34.00 36.00 38.00 40.00 42.00 44.00 46.00 48.00
Time-->

A.2 Example of 5 mg/kg RME in HDT


A.2 Example of 5 mg/kg RME in HDT

Abundance D33 IS
C17:0
13000

12000

11000

10000

9000

8000

7000

6000

5000

4000

3000

24.00 26.00 28.00 30.00 32.00 34.00 36.00 38.00 40.00 42.00 44.00 46.00 48.00
Time-->

585.8

Issued under licence for use by Intertek employees only.


IMPORTANT: This file is subject to a licence agreement issued by the Energy Institute, London, UK. All rights reserved. It may only be used in accordance with
the licence terms and conditions. It must not be forwarded to, or stored or accessed by, any unauthorised user. Enquiries: e: pubs@energyinst.org.uk t: +44 (0)207 467 7100
FAME IN AVTUR – GC-MS, IP 585

A.3 Example of 30 mg/kg RME in MEROX


A.3 Example of 30 mg/kg RME in MEROX
Abundance

30000

25000

20000

15000
D33 IS
C17:0
10000

5000

24.00 26.00 28.00 30.00 32.00 34.00 36.00 38.00 40.00 42.00 44.00 46.00 48.00
Time-->

15

585.9

Issued under licence for use by Intertek employees only.


IMPORTANT: This file is subject to a licence agreement issued by the Energy Institute, London, UK. All rights reserved. It may only be used in accordance with
the licence terms and conditions. It must not be forwarded to, or stored or accessed by, any unauthorised user. Enquiries: e: pubs@energyinst.org.uk t: +44 (0)207 467 7100
FAME IN AVTUR – GC-MS, IP 585

Bibliography
[1] Ministry of Defence Standard 91-91, [2] ASTM D4306 Standard practice for aviation
Turbine fuel, aviation kerosine type, jet fuel sample containers for tests affected by
A1, NATO code: F-35, joint services trace contamination
designation: AVTUR

585.10

Issued under licence for use by Intertek employees only.


IMPORTANT: This file is subject to a licence agreement issued by the Energy Institute, London, UK. All rights reserved. It may only be used in accordance with
the licence terms and conditions. It must not be forwarded to, or stored or accessed by, any unauthorised user. Enquiries: e: pubs@energyinst.org.uk t: +44 (0)207 467 7100

You might also like