You are on page 1of 20

Dental Radiography Series

SUCCESSFUL
INTRAORAL
RADIOGRAPHY

DENTAL
Content

Successful Intraoral Radiography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Proper Film Positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Film exposure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Recommended Exposure Settings for KODAK Intraoral Dental films . . . . . . . . . . . . . . 11

Processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Manual Processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

KODAK Chemistry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Handling Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Summary of Intraoral Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

3
Successful Intraoral Radiography

Every dentist would like to achieve the goal of a quality correct them. For similar information on panoramic
intraoral radiograph. A quality intraoral radiograph radiography please see the Kodak publication titled
will reveal maximum detail in the image with anatomic “Successful Panoramic Radiography.”
accuracy and optimal density and contrast. This will
give maximum diagnostic yield. This pamphlet will
address some of the common pitfalls and errors seen in
intraoral radiography and address how to prevent and

QUALITY RADIOGRAPH
The goal of all radiography should be to produce a The film must be properly positioned to ensure proper
quality radiograph. Such a radiograph will exhibit geometry and prevent distortion and overlap. Second,
maximum detail to resolve fine objects. It will show the exposure technique factors must be appropriate for
the teeth and anatomic structures accurately with a the patient and the film selected. And last, proper pro-
minimum of distortion or magnification. It will have cessing time, temperature and handling requirements
the optimal density and contrast (visual characteristics) must be followed to produce a quality radiograph.
to maximize its use for the detection of dental disease.
To create such a film, the dental staff must pay atten-
tion to all three steps in the production of the radi-
ograph, positioning, exposure and processing.

THREE STEPS
IN PRODUCING
A QUALITY
RADIOGRAPH

Figure 1

4
Positioning

1
STEP 1: PROPER FILM POSITIONING
Film placement for proper anatomic coverage is beyond
the scope of this pamphlet and can be reviewed in any
quality dental radiography text. This pamphlet will
discuss improper film placement that can lead to errors
such as overlapped contacts and distorted teeth and
roots. This is due to the fact that dental radiography is Figure 2-
a shadow casting technique, in that we cast an image Shadow Casting
of the tooth onto the film. Shadow casting can cause
geometric distortions in the final radiograph such as
elongation, magnification and overlapping contacts.

Geometric distortion can be minimized by using a long


X-ray source to object distance while maintaining the
shortest distance between the film and object. This can
be obtained using a long cone (40 cm) technique. Geo-
metric distortion can also occur if the film is not at
right angles to the beam. For this reason, it is recom-
Figure 3 – mended to always use a film holding and position-indi-
excessive vertical cating device (PID). Many practitioners assume these
angulation, note devices can control all angulation problems. They only
inferior border of hold the film perpendicular to the X-ray beam. They
mandible visible,
elongation of roots.
do not totally prevent errors in the vertical and
Correct by moving horizontal angulation of the film to the tooth itself.
film further back This can lead to commonly seen errors of overlapped
into the mouth contacts and vertical distortion of teeth on the radi-
ograph.
They can be minimized by proper use of the paralleling
technique.

Figure 4 – Figure 5 –
Distortion from Bent film corners
film bending in causing black lines
corner of arch. on film. Correct
Correct by placing with proper film
film further into placement, use soft
center of mouth film packets

5
Positioning

DISTORTION
(VERTICAL ANGULATION)
Although film-holding devices hold the film at right
angles to the X-ray beam, they do not prevent rotation
of the whole device in a vertical axis. This rotation
places the film at an angle to the tooth and can result
in distortion when the angle is large. This is commonly
Figure 6 – Vertical
seen when the film is not placed far enough in the angulation due
center of the mouth and the film must be angulated to to fulcrum rotation
avoid the slope of the palate or the mandibular on lower molar
vestibule. It can be avoided by simply placing the film
deeper into the center of the mouth, so that tipping is
not necessary. Another type of distortion occurs when
the film is allowed to bend on biting down by the
patient. This can also be avoided by placing the film
deep enough into the mouth to avoid contact with the
palate. The bending of film corners for patient comfort
can also cause errors, as the pressure of the bend can
cause partial film development. This shows up as black
lines on the film.

OVERLAPPED CONTACTS
(HORIZONTAL ANGULATION)
Figure 7 – In order to maximize the amount of contact opening,
Improper the beam should be directed at right angles to the con-
horizontal tact area. In the mandible this is fairly straightforward.
angulation, con- In the maxilla, though, the molar contacts are often
tacts overlapped directed mesially due to the triangular shape of the
maxillary molars. This means the beam must also be
directed from the mesial to open these contacts. Often
times the reverse is done. The beam is directed from
the mesial in the bicuspid area and at right angles or
distal in the molar region. This will most often result
in overlapped contacts. Contact areas should always be
Figure 8 – visualized prior to taking bitewing radiographs.
Proper
horizontal
angulation, con-
tacts open Figure 9
Angle for bicuspid

Angle for molar

6
Positioning

Cone cuts
Dental X-ray beams are collimated or restricted to a appear clear. This is known as a cone cut. Proper use of
diameter of 7 cm at the end of the cone or even less position indicating devices (PIDs) will help to prevent
when a rectangular collimator is used. When the exit this problem, which can occur with either round or
pattern of the beam is not aligned with the film, part rectangular cones.
of the film will not be exposed to x-radiation and will

Figure 10

Figure 11 – Cone cut using Figure 12 – Cone cut using


round cone rectangular cone

Reversed Films
Dental X-ray film is marked with an indicator dot to
help indicate the tube side of the film and to help
distinguish the patient’s right or left side. In addition,
the film packet contains a sheet of lead foil which pre-
vents unnecessary radiation from passing through into
the patient and reduces scatter radiation. This sheet of
Figure 13 – Reversed film, Figure 14 – Foil from packet
lead foil is marked with a special pattern. When a film note dot on pattern on left showing pattern of dots
is exposed from the wrong side, the pattern is visible edge of radiograph, light
on the radiograph. Due to the attenuation of the foil, overall density
the radiograph also appears overall light in density.

RADIOGRAPH ERROR FIX


Teeth elongated, cusps don’t Excessive vertical angulation Correct film placement and reduce
overlap, sinus structures or inferior vertical angulation
border of mandible visible

Contacts of teeth overlapped Improper horizontal angulation Visualize contact area and adjust

Dark lines on corner of film Film bent Use of SUPER POLY-SOFT®


packaging and proper placement can
reduce need for bending film

Clear area on one edge of film Cone cut Align cone to position indicating
either in arc or straight the device

Film light in density, unusual Packet was reversed and exposed Follow instructions on packet for
pattern across film (“tire tracks” or through the back side, pattern is orientation
“herringbone”) from foil inside packet
Table 1 – film positioning errors

7
Film exposure

2
Step 2: Film exposure

Selecting a film
Film selection is important to both radiographic suc- helpful exposure guidelines please see page 11. Using
cess and to provide the lowest practical exposure to the these guidelines, the practitioner can verify that their
patient. To achieve consistent quality radiographs you exposure factors are within the suggested normal
must use a consistent quality film. Low cost films may ranges for good radiographic technique.
vary from batch to batch or may be from different
manufacturers. This makes establishing consistent
exposure and development technique factors very
difficult. Dental films are provided in different speed
groups with D-speed films being the slowest and
F speed the fastest. Kodak’s newest film, InSight®, is a D
F speed film that can provide an additional up to 20%
Exposure
reduction in exposure over the E-speed films (up to time
60% over D-speed films) with no loss of image E
contrast or quality. In accordance with the ALARA
principle (keep doses As Low As Reasonably Achiev- F
able), use of F speed film is highly encouraged. For
Film speed

Figure 15 – Exposure time versus film speed

Influence of milliamperes
Most modern dental X-ray machines no longer allow
for the adjustment of mA or milliamperage. Since the
effect of increasing or decreasing mA is the same as
for exposure time, it is common to combine the two
and talk of mAs or milliamp-seconds. In dentistry we
are mainly concerned with exposure time as discussed
below.

Influence of time
Film density (how light or dark overall a film is) is
directly related to exposure time. The longer the expo-
sure time the more X-ray photons reach the film and
expose it. Therefore, the film is darker. The X-ray timer
can be thought of as a faucet. It turns the flow of X-rays
Figure 16 – on or off. If you open the faucet twice as long, you will
exposure time is
like a faucet get twice as many X-rays out of the machine. If you
double the time, the film will be darker (approx twice
as dark).

8
Film exposure

Figure 17 – 0.25 second Figure 18 – 0.5 second Figure 19 – 1.0 second


exposure (underexposed) exposure (proper exposure) exposure (over exposed)

RADIOGRAPH EXPOSURE TIME FIX


Too dark Too long Use shorter time, fewer impulses

Too light Too short Use longer time, longer impulses


Table 2 – expose time errors

Influence of peak kilovoltage


Many modern dental X-ray machines no longer allow the adjustment
of peak kilovoltage (kV). The kilovoltage affects both the quantity of
the X-rays produced and their average energy. The average energy is
sometimes referred to as the “beam quality.” The effect of peak kilovoltage
can be thought of as a nozzle. It controls the force of the emerging stream
of X-rays, whereas the faucet (timer) controls the volume.

Figure 20 – kV is like a spray nozzle

kV has two effects on the quality of the final radiograph.


First, it affects the contrast or gray scale. Low kV is like
opening up the nozzle. The lower energy X-rays have
less penetrating power. This gives a high contrast image
with more of a black and white appearance. High kV
is like closing down the nozzle. The beam is “harder”
with higher energy. High kV gives a low contrast image,
but with more shades of gray to show subtle contrast
changes. Figure 21 – Low contrast long gray scale above,
high contrast short gray scale below
Second, using higher kV produces more X-rays. This
is not a linear relationship. For instance, increasing
or decreasing the kV by 15% should lead to a dose
decrease or increase respectively by a factor 2. As a
result, a good rule of thumb is:

EVERY INCREASE OF 10 kV = EXPOSURE TIME DIVIDED BY 2

9
Film exposure

Figure 22 – 55 kV Figure 23 – 70 kV Figure 24 – 85 kV Figure 25 – Film mistakenly


shot at 90 kV, all other
exposure factors were set as
normal

Although many modern X-ray machines do not allow


changes in kV, modern DC (direct current) machines RADIOGRAPH kV
are actually equivalent to older machines operating at Too dark Too high
higher voltages. For example, a modern 70 kV DC Too light Too low
machine has a beam efficiency similar to an older
80 kV machine. Too much contrast Too low
Too washed out Too high
Table 3 – kV Errors
Other errors that can occur during exposure include
patient movement and double exposures. One way to
decrease patient movement errors is to be sure a head-
rest stabilizes the patient’s head during film placement
and exposure. Double exposures are usually caused by
operator inattention. Using unit dosing and keeping
unexposed films separated from exposed films can help
alleviate this problem. It is important to note that when a
double exposure occurs there is usually a corresponding
blank film in the series. Figure 26 – Patient Figure 27 – Double exposure,
movement, note blurring of not multiple images of teeth
image, soft focus

RADIOGRAPH ERROR FIX


Blurring of structures Patient movement Remind patient to hold still, use shorter
exposure times, tube movement is not as
bad as patient movement

Multiple images on film Double exposure Exposed films should always be separated
from unexposed while taking radiographs

Table 4 – Errors during exposure

10
Recommended Exposure Settings for KODAK Intraoral Dental films

These settings should be used as guidelines and may require adjustments to accomodate local configurations
(equipment, processing etc.)

ULTRA-SPEED Settings kV 50 60 65 65 65 70 70 70 80
D speed mA 7 7 7,5 8 10 7 8 10 10
20 cm Maxillary Incisor 1,37 0,55 0,32 0,30 0,24 0,27 0,24 0,19 0,10
Cuspid 1,37 0,55 0,32 0,30 0,24 0,27 0,24 0,19 0,10
Bicuspid 1,83 0,73 0,43 0,40 0,32 0,37 0,32 0,26 0,13
Molar 2,06 0,82 0,48 0,45 0,36 0,41 0,36 0,29 0,14
Mandibular Incisor 1,14 0,46 0,27 0,25 0,20 0,23 0,20 0,16 0,08
Cuspid 1,14 0,46 0,27 0,25 0,20 0,23 0,20 0,16 0,08
Bicuspid 1,26 0,50 0,29 0,28 0,22 0,25 0,22 0,18 0,09
Molar 1,37 0,55 0,32 0,30 0,24 0,27 0,24 0,19 0,10
Bite Wing Anterior (Incisor) 1,14 0,46 0,27 0,25 0,20 0,23 0,20 0,16 0,08
Posterior (Bicuspid) 1,37 0,55 0,32 0,30 0,24 0,27 0,24 0,19 0,10
Occlusal 2,29 0,91 0,53 0,50 0,40 0,46 0,40 0,32 0,16
40 cm Maxillary Incisor 5,49 2,19 1,28 1,20 0,96 1,10 0,96 0,77 0,38
Cuspid 5,49 2,19 1,28 1,20 0,96 1,10 0,96 0,77 0,38
Bicuspid 7,31 2,93 1,71 1,60 1,28 1,46 1,28 1,02 0,51
Molar 8,23 3,29 1,92 1,80 1,44 1,65 1,44 1,15 0,58
Mandibular Incisor 4,57 1,83 1,07 1,00 0,80 0,91 0,80 0,64 0,32
Cuspid 4,57 1,83 1,07 1,00 0,80 0,91 0,80 0,64 0,32
Bicuspid 5,03 2,01 1,17 1,10 0,88 1,01 0,88 0,70 0,35
Molar 5,49 2,19 1,28 1,20 0,96 1,10 0,96 0,77 0,38
Bite Wing Anterior (Incisor) 4,57 1,83 1,07 1,00 0,80 0,91 0,80 0,64 0,32
Posterior (Bicuspid) 5,49 2,19 1,28 1,20 0,96 1,10 0,96 0,77 0,38
Occlusal 3,66 2,13 2,00 1,60 1,83 1,60 1,28 0,64

INSIGHT Settings kV 50 60 65 65 65 70 70 70 80
E/F speed mA 7 7 7,5 8 10 7 8 10 10
20 cm Maxillary Incisor 0,62 0,25 0,14 0,14 0,11 0,12 0,11 0,09 0,04
Cuspid 0,62 0,25 0,14 0,14 0,11 0,12 0,11 0,09 0,04
Bicuspid 0,82 0,33 0,19 0,18 0,14 0,16 0,14 0,12 0,06
Molar 0,93 0,37 0,22 0,20 0,16 0,19 0,16 0,13 0,06
Mandibular Incisor 0,51 0,21 0,12 0,11 0,09 0,10 0,09 0,07 0,04
Cuspid 0,51 0,21 0,12 0,11 0,09 0,10 0,09 0,07 0,04
Bicuspid 0,57 0,23 0,13 0,12 0,10 0,11 0,10 0,08 0,04
Molar 0,62 0,25 0,14 0,14 0,11 0,12 0,11 0,09 0,04
Bite Wing Anterior (Incisor) 0,51 0,21 0,12 0,11 0,09 0,10 0,09 0,07 0,04
Posterior (Bicuspid) 0,62 0,25 0,14 0,14 0,11 0,12 0,11 0,09 0,04
Occlusal 1,03 0,41 0,24 0,23 0,18 0,21 0,18 0,14 0,07
40 cm Maxillary Incisor 2,47 0,99 0,58 0,54 0,43 0,49 0,43 0,35 0,17
Cuspid 2,47 0,99 0,58 0,54 0,43 0,49 0,43 0,35 0,17
Bicuspid 3,29 1,32 0,77 0,72 0,58 0,66 0,58 0,46 0,23
Molar 3,70 1,48 0,86 0,81 0,65 0,74 0,65 0,52 0,26
Mandibular Incisor 2,06 0,82 0,48 0,45 0,36 0,41 0,36 0,29 0,14
Cuspid 2,06 0,82 0,48 0,45 0,36 0,41 0,36 0,29 0,14
Bicuspid 2,26 0,91 0,53 0,50 0,40 0,45 0,40 0,32 0,16
Molar 2,47 0,99 0,58 0,54 0,43 0,49 0,43 0,35 0,17
Bite Wing Anterior (Incisor) 2,06 0,82 0,48 0,45 0,36 0,41 0,36 0,29 0,14
Posterior (Bicuspid) 2,47 0,99 0,58 0,54 0,43 0,49 0,43 0,35 0,17
Occlusal 4,11 1,65 0,96 0,90 0,72 0,82 0,72 0,58 0,29
Note: for large patients increase time by approximately 25% for children and / or small patients decrease time by approximately 30%

11
Processing

3
Step 3: Processing

Development
RADIOGRAPH DEVELOPIING PROBLEM
Even with the excellent automatic processors available
today, many errors can occur during processing. Many Too light Temperature or time too low
of these errors revolve around improper film handling, Too dark Temperature or time too high
but some can be due to the processor itself. Processing
Too light Contaminated or weak developer
is a chemical reaction therefore: (replace or replenish)
INCREASED TEMPERATURE = Too dark (fogging) Over concentrated developer
INCREASED DEVELOPMENT = DARKER FILM Table 5 – Processing Errors
INCREASED TIME =
INCREASED DEVELOPMENT = DARKER FILM
CONTAMINATED OR DEPLETED CHEMISTRY =
INCOMPLETE DEVELOPMENT = LIGHT FILM
For these reasons, manufacturer’s recommendations for
development time and temperature should be closely
Developer must be replenished following manufacturer’s
followed. Automatic processors should still be checked
recommendations or it will become exhausted. These
for developer temperature as heating elements can fail
recommendations are usually based on the amount of
or overheat. Proper attention to chemical dilution,
radiographs processed. However, developer exhaustion
mixing and loading must be followed. Fixer should
is determined by the surface area of the films processed
always be poured first into the processor as a small spill
not the number of films. If large numbers of panoramic
of fixer into the developer can drastically weaken the
or cephalometric films are processed, more frequent
developer.
replenishment will be needed.

12
Manual processing

Manual processing of intra-oral Kodak Dental X-ray Films

1 STIR SOLUTIONS
Dilute developer and fixer solutions
as directed on the containers. Use separate
2 CHECK TEMPERATURES
OF SOLUTIONS
Check temperatures of solutions with an
3 CHECK DEVELOPMENT TIME
Refer to the table above and check the
development time based on the temperature
paddles for each solution to avoid possible accurate thermometer. Rinse the thermometer of the developer in preparation for step 5.
contamination. Stir the solutions gently. This thoroughly in running water before checking the The times in the table are recommended for
ensures uniformity of solutions and temperatures. other solution. Developer temperature should use with KODAK Dental X-ray Developer.
be within 18-24 °C. Refer to the instruction
leaflet for fixer temperatures.

4 LOAD FILM ON HANGER


Remove films from packets and attach
carefully to a multiple-clip hanger, avoid
5 IMMERSE FILMS IN DEVELOPER
AND START TIMING
Immerse the films smoothly and without pause;
6 AGITATE FILMS
Immediately raise and lower hanger
(agitate it) several times so that air bubbles
finger-marks, scratching or bending. this minimizes streaking. are removed from the film surfaces.
Start timing.

7 RINSE THOROUGHLY
At the end of development time, quickly
remove hanger from the developer and place in
8 FIX ADEQUATELY
Place films in the fixer solution and
agitate the hanger vigorously. Films should
9 WASH THOROUGHLY
Remove hanger from fixer and place in
wash section. Wash for 10 to 30 minutes in
wash section for 30 seconds with clean running remain in fixer for 10 min. running water. Eight volume changes per hour
water at a temperature of 15-24 °C. See specific instructions for fixing times are recommended.
Lift from water and allow to drain over the and temperatures. Extra-oral film should be washed for 5 minutes.
wash section.

10 DRY
Suspend hanger from a drying rack in
dust-free area. Use a fan to accelerate drying.
When dry, remove films from hanger, mount
Note: The times indicated are appropriate with
correct exposure time of the intra-oral films.
and identify.

13
KODAK Chemistry

KODAK Chemistry for Manual Processing


Product Cat. No. Quantity / Unit Processing Characteristics
Dental Monobath 508 7911 6 x 500 ml Developing: Ready to use
4 - 8 min 20°C
(4 min as a minimum,
recommended are 8 min)

Wash:
10 min 20°C
Rapid Access Developer 501 0459 6 x 500 ml Developing: Ready to use
15 sec. 20°C
Rapid Access Fixer 501 0491 6 x 500 ml Fixing:
30 - 60 sec. 20°C
Wash:
1 - 2 min 20°C
(For archiving wash 10 min)

Dental Developer 501 6316 1 x 2,25L Developing: Concentrate – requires dilution


6 min 18°C 1 bottle concentrate + 3 bottles water
506 0686 4 x 1L 5 min 20°C
4min 22°C Please use all bottle contents at once,
3 min 24°C leftover developer or fixer will oxidise

Rinse:
30 sec in
running water
Dental Fixer 501 6308 1 x 2,25L Fixing: Concentrate – requires dilution
10 min 20°C 1 bottle concentrate + 3 bottles water
506 0694 4 x 1L Wash:
10 min in
running water

KODAK Chemistry for Automatic Processing


Product Cat. No. Quantity / Unit Processing Characteristics
Dental Readymatic 524 6970 2 x 5L See processor’s Ready to use /
Developer 524 6996 4 x 2L manual For all dental processors
Dental Readymatic 524 6988 2 x 5L See processor’s Ready to use /
Fixer 524 7002 4 x 2L manual For all dental processors

14
Handling

Handling
Film must be handled carefully under proper safelight can overlap or stick together. Other errors can occur
conditions during processing. Many newer E and F from emulsion tears, fingerprints, static electricity and
speed dental films recommend a red safety filter on chemical spills onto the film. Only clean, dry, powder
daylight loaders instead of the more common amber free gloves should contact the bare film prior to pro-
ones, especially if the processor is in a brightly lit cessing. Unprocessed film should not come in contact
room. Use of an amber filter in such conditions can with wet or contaminated surfaces as this may lead to
result in film fogging. Care must also be taken when film spotting. Kodak ClinAsept® barriers allow film
feeding film into a processor. Opening the lid too soon to be handled with clean hands after removal from the
on a daylight loader can result in room light fogging barrier envelope and can greatly reduce handling arti-
the trailing edge of the film. It can take up to 15-20 facts. After processing, films must not come into con-
seconds for film to completely enter an automatic tact until completely dry as the wet emulsions can stick
processor. Film fed too quickly or too close together together and peel off the films when they are separated.

15
Handling

Figure 28 – Emulsion tear, Figure 29 – Finger Print, Figure 30 – Overlap of


note white area under pontic dirty finger had fixer on it films during processing
where no emulsion is left leaving white mark on film

Figure 31 – Stain from Figure 32 – Light fog on left


incomplete fixation and edge of film from opening lid
washing of film on daylight loader too soon
Handling Errors
RADIOGRAPH HANDLING PROBLEM FIX
Too dark (similar to fogging) Improper safelight Use red safelight with new fast
intraoral films such as Insight®

Random sized dark spots: Stored in humid or hot conditions Store film between 10 to 24°C
v shaped, mottle “noise” in dry conditions

Dark rectangular on film Overlap during processing Feed films slowly or side by side

Dark edge on film Exposed to light before safely in Allow 15-30 seconds from last
processor film to enter processor

Fingerprints Improper handling Clean, dry hands and hold the film
on the edges

Dark spots Developer stains Use fresh developer

White spots Fixer stains Use fresh fixer

Streaks or scratches Emulsion tears Never allow wet films to contact one
another or fingernails, emulsion is
delicate when wet. Keep films away
from extreme sides of processor.

Clean rollers with mild detergent


Dark spots in pattern Roller marks & rinse thoroughly or try Roller
Transport Clean-up film

Add humidifier and or open


Dark spots branched in comet shape Static (most common in packets slowly to minimize static
winter months) discharge

Table 6 – Handling errors

16
Summary of Intraoral Errors

Teeth elongated, cusps don’t overlap, sinus structures Excessive vertical angulation, correct film placement
or inferior border of mandible visible
Contacts of teeth overlapped Improper horizontal angulation, visualize contact
area and adjust
Dark lines on corner of film Film bent, use of Super Polysoft® packaging and proper
placement can reduce need for bending film
Clear area on one edge of the film either Cone cut, use position indicating device
in arc or straight
Film light in density, unusual pattern across film Film was reversed and exposed through the back side,
(“tire tracks” or “herringbone”) pattern is from lead foil
Too dark Exposure too long
kV too high
Processor temperature to high
Development time too long
(fogging) Over concentrated developer
(similar to fogging) Improper safelight
Too light Exposure too short
kV too low
Processor temperature too low
Development time too short
Contaminated or weak developer (replace or replenish)
Too much contrast kV low
Too washed out or gray kV high
Blurring of structures Patient movement
Multiple images on film Double exposure
Mottle “noise” Stored in humid or hot conditions
Dark rectangular area on film Overlap during processing
Dark edge on film Exposed to light before safely in processor
Fingerprints Clean, dry hands and hold the film on the edges
Dark spots Developer stains
White spots Fixer stains
Clear streaks, splotches or scratches Emulsion tears
Dark spots in pattern Dirty Processor Rollers
Dark spots in random or comet pattern Static due to over-dry conditions

Table 7 – Summary of Intraoral Errors

17
KODAK DENTAL
Hedelfinger Strasse 60
70327 Stuttgart
GERMANY
Tel. ++ 49 711 406 3910
Fax ++ 49 711 406 3331
DoldeMedien 0184/03 UK

You might also like