Green Belt:
-Sandip Mohite
Team Members:
-Tejas Bengali
SCOPE:
Selection from alternatives(No
design changes)
DELIVERABLES:
Improve product quality.
Patient
Monitor
Mother
Board
Navigation Spo2-
Key Pad Modules
Keys Module
Spo2
Sensor
Technical
Comfort of use Performance
Reliability
Biocompatibility
Standard
Cost
Patient Safety
Technical Performance
DOE
Experiment was designed to remove all the
biases viz.
1. Previous knowledge
2. Inclination
3. Effect of day
10 Persons and 6 probes selected for experiment.
1. Variety in terms of gender, age.
Rating from 1-10
-1- Least Comfortable
-10- Most Comfortable
Probes of
different makes
labeled
Black box
To avoid bias
90 1
Residual
Percent
50 0
10 -1
1
-2
0.1
-2 -1 0 1 2 0.0 2.5 5.0 7.5 10.0
Residual Fitted Value
Residual
30
0
20
-1
10
-2
0
-2.25 -1.50 -0.75 0.00 0.75 1.50 2.25 1 10 20 30 40 50 60 70 80 90 100 110 120
Residual Observation Order
7
Mean of A
4
1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6
9
6 10
2
1 2 3 4 5 6
PROBE
Interactions suggest people rate same probe differently. People have their
own perception of comfort level.
LARSEN & TOUBRO 18/23
Medical Equipment & Systems
To Select Spo2 probe DMADV Six-Sigma
SN Ratio Analysis
Main Effects Plot (data means) for SN ratios
Human Probe
17
SN ratios
16
Mean of S
15
14
13
1 1 2 3 4 5 6
Signal-to-noise: Larger is better
Mean Ratings
Comfort
Probe Name Probe No. Level
D 4 6.4
C 1 6.55
E 3 6.7
A 5 6.7
B 6 7.2
Datum 2 7.5
Results
Cost Avoidance of Rs. 2500 per unit.(Cumulative
till July-2010 is Rs 23L)
Qualitative and quantitative Feedback from four
sales regions indicate good customer acceptance
and 20% reduction in initial failures(1-3 months)