You are on page 1of 12

Product Data

No. MPDCT0170EAB

Multislice HELICAL CT SCANNER

MULTI
APPLICATION
This system, the Asteion, is a multislice Helical CT scanner capable of scanning the whole-body. The system generates four slices per revolution, using the Selectable Slicethickness Multi-row Detector (SSMD). Since it can also perform high speed reconstruction at a maximum 12 frames per second, every examination can be processed at ultra-high speed. In combination with the Aspire CI technology, it is possible to quickly perform more accurate scanning and examinations.

FEATURES
High-speed scanning This system permits simultaneous 4-slice acquisition per 0.75 second rotation using the SSMD and providing exceptional volumetric data acquisition. The waiting time for cooling of the X-ray tube is reduced, because the exposure time is short. Therefore, patient waiting time and throughput is improved. Wide range of slice thicknesses The SSMD, which provides both high-speed and high-resolution scanning, allows slice thicknesses of 0.5 mm, 1 mm, 2 mm, 3 mm, 4 mm, 5 mm, 8 mm, or 10 mm, to be selected, based on the objectives of the examinations. High quality image The system achieves low-contrast resolution of 2.5 mm at 0.25%, and high-contrast resolution of 0.35 mm. Multislice helical CT can be performed using thin slices in routine examinations, so that high-precision, 3D and MPR images can be generated from the high-resolution voxel data. In CT angiography of the brain, for example, a 40-mm distance can be scanned in 10 seconds at a slice thickness of 1 mm*1. The use of "stack scan" with thin slices minimizes partial volume artifact. High patient throughput In combination with fast scanning, the high-power 60-kW X-ray generator*2 and 6.5-MHU X-ray tube*2, enable the data for routine examinations and the biopsy and 3D generation to be acquired in one scan. The image of any slice thickness can be generated later by reconstruction retry. For example, after the 2-mm slice thickness scanning is performed, the images of several slice thicknesses can be generated from raw data such as a 10-mm thick image for routine examinations, a 5-mm thick image for detail examination, and 2-mm thick image for generation of a 3D volume. Multislice CT fluoroscopy (option) Conventional CT fluoroscopic shows only a single slice, but multislice CT permits realtime image reconstruction to display 3 images obtained by combining 4 slices. Multislice CT fluoroscopy significantly improves operability in biopsy and interventional procedures. *1 Helical pitch 3 *2 Option

Product Data
No. MPDCT0170EAB

Easy operations The system consists of two stand-alone consoles (one for monitoring and the other for image processing). Both consoles are provided with a hybrid keyboard, monitor and mouse, and are capable of simultaneous parallel processing, significantly improving efficiency in performing image processing and diagnosis. The use of a large-size 21-inch color monitor, a sophisticated GUI, and high-performance CPU has improved ease of operation and reduced operation time.

Optimal space utilization The Asteion MULTI consists of only three components - gantry, couch and console. Even though it has a large gantry aperture of 72 cm in diameter and a couchtop width of 47 cm for enhanced patient comfort and safety, the whole system fits into an area as small as 20 meter square.

COMPOSITION (MODEL NUMBER)

MULTI
standard composition (TSX-021A/3)
(1) Asteion system (TSX-021A/1) Gantry ..............................................................1 Patient couch....................................................1 Console ............................................................1 Accessories Set of phantoms ..........................................1 Head rest......................................................1 Set of operation manuals ............................1 (2) Multislice CT system upgrade kit (CGS-25A) SSMD ..............................................................1 Data Acquisition System (DAS) for multislice CT..............................................1 MUDAT for multislice CT ............................1 Gantry modification kit ..................................1 Console for multislice CT................................1 Phantom set......................................................1 System software ................................................1 Set of operation manuals..................................1 This kit is assembled into the main Asteion system at the manufacturing site, prior to delivery. Note: The console desk is not included as standard.

Fly-through software (CFT-02A) Quantitative bone mineral study system** (CBM-13C) Display system for Dental Application (CDP-06C) High speed 3D processing system (VolumeView, CTD-11A) Multislice CT fluoroscopy (TSXF-003C) Pediatric scanning system (CHKS-001A) Patient observation system (CPM-05A)* Video interface (CVF-02B) DICOM Storage SCP (COT-30A) DICOM MWM (COT-32A) DICOM MPPS (COT-33C) DICOM Q/R SCP (COT-34C) DICOM Q/R SCU (COT-35C) PC online interface (COT-37B) Color printer interface (CCP-02A) ECG-gating scan system (CHEG-002A/3) ECG-gating reconstruction system (CHEG-002A/4) Foot switch for patient couch (CAFS-004A) Extend image data disk (EISC-005A) Extend raw data disk (EISC-006A) System transformer (CETF005B) 48-kW X-ray generator (CXGS-009A) 54-kW X-ray generator (CXT-09B) 60-kW X-ray generator (CXGS-008A) 6.5-MHU X-ray tube (CXB-650B) * Not available for sale in European countries. ** Not available for sale in U.S.A.

Optional items
Diagnostic image display unit (color 21 inch) (CMCM-002B) Diagnostic image display unit (Flat monitor) (CMM-003B) Image processing software, ALATOVIEW (TIWS-001B)
2

Multislice HELICAL CT SCANNER

MULTI

PERFORMANCE SPECIFICATIONS
Scan parameters
Scan regions: Scan system: Scan time CT scan: Whole body, including head 360 continuous rotate/rotate

0.5 s (Partial), 0.75, 1, 1.5, 2 and 3 s Scanoscope: Arbitrary up to 14 seconds depending on the length of the scan field along the patient axis. Scan cycle time (for 0.75 s scan) SCAN & VIEW mode: Min. 1.8 s (one image display) SCAN & SCAN mode: Min. 1.8 s (rapid sequence, couch movement 10 mm) Note: Scan cycle time refers to the time between one scan initiation and the next. SCAN & VIEW mode permits immediate viewing of images upon completion of each individual slice. Scan field CT scan: 180 mm (SS) 240 mm (S) 320 mm (M) 400 mm (L) 500 mm (LL) Scanoscope: Lateral direction Axial direction 240 mm (S) Arbitrary from 400 mm (L) 200 mm to 1,390 mm 500 mm (LL) Slice thickness/ Number of slices: 0.5 mm 4 1 mm 4 2 mm 4 3 mm 4 4 mm 4 5 mm 4 8 mm 4 10 mm 2 Number of slices: Max 4 slices/scan Gantry tilting angle: From forward 30 to backward 30 (in increments of 0.5) Remote control from the console is possible. Tube position for scanoscope: 0, 90, 180 and 270 Gantry aperture: 720 mm diameter

Patient couch Vertical movement System: Speed of vertical movement: Stroke: Minimum couchtop height:

Hydraulically driven 16 to 30 mm/s Approx. 570 mm

Approx. 300 mm Remote control from the console is possible. Note: This function allows the user to check the image on the console and adjust the couchtop without leaving the console. Adjustment can be made in 10 mm increments at the console. Couchtop movement System: Motor driven or manual Speed of movement: 100 mm/s (fast) 10 mm/s (slow) Stroke: 1,820 mm Scannable range: 1,430 mm (with head rest) Step feed pitch: 0.5 to 600 mm in 0.5 mm unit Reproducibility: 0.25 mm Repeatable to within 0.25 mm after 600 mm of movement Remote control from the console is possible. Couchtop width: 470 mm Load limitation Max. allowable load: 205 kg (450 lb)

Voice-recorded instruction and scan system (VoiceLink)


Vocal instructions to the patient can be recorded electronically by the operator and replayed during scan sequences automatically as a part of the eXam Plan. Available number of messages: Max. of 32 Available number of seconds: Max. of 128 s for a total of 32 messages. Delay time setting: Time between the end of reproduction and scan start can be set up to 10 s in increments of 1 s.

Helical Scan
X-ray tube rotation speed: 0.75, 1, 1.5 s/360 Maximum Helical Scan time: 100 s
3

Product Data
No. MPDCT0170EAB

Scan start time delay: Scan field in axial direction:

Min. 1 s Setting is possible in steps of 0.1 s. Max. 1,390 mm/scan Up to 10 scan plans are programmable in one eXam Plan. (Multiple and/or Multi-directional Helical) This can be specified arbitrarily within the range of 1.7 mm to 64 mm/s. (in steps of 0.1 mm/s) 2.5/3/3.5/4.5/5/5.5/6 Min. 1.2 s 8 images/s (1 slice/512 512 matrix) 12 images/s (1 slice/256 256 matrix)

Couch speed:

Helical pitch (Note) Image slice Setting slice thickness thickness 3.0 4.5 6.0 0.5 mm 0.5 mm 15.0 mm 22.5 mm 30.0mm 1 mm 1 mm 30.0 mm 45.0 mm 60.0 mm 2 mm 2 mm 60.0 mm 90.0 mm120.0 mm 5 mm 5 mm 150.0 mm225.0 mm300.0 mm 10 mm 8 mm 240.0 mm360.0 mm480.0 mm Note: The couchtop traveling distance per rotation is shown in proportion to the slice thickness.

Helical pitch Image reconstruction time: Real time Helical reconstruction time:

Dynamic scan
Scan time: Programmable time: 0.75, 1, 1.5 s/360 Max. 1 hour This refers to the maximum time within which a series of scans are performed following a predetermined eXam Plan. Max. 10 Maximum time of one continuous scan is 100 s.

SureStart Max continuous scan time: 100 s Region of interest (ROI): Max. 3 ROIs CT number measurement interval: 0.125 s Scan start time delay: Min. 3 s Display function: CT number lapse time Specification of reconstruction position: By entering the couchtop position, or on the scanogram (in minimum units of 0.1 mm) Image slice thickness: Can be set in 0.5 mm increments if it is more than the typical value of the scan slice thickness, and up to five times the scan slice thickness or less. (Max. image slice thickness is 20 mm.) Reconstruction Mode: Full image Half image Detail image Reconstruction method: MUSCOT Reconstruction (MUltiSlice COne-beam Tomography) The relation between the slice thickness and the imaging area scanned in 7.5 s (0.75 s, 10 rotations) in the longitudinal direction are shown for helical pitches 3.0, 4.5, and 6.0.

Number of programmable scans: Scan plan Scan interval:

Min. interval is 1 s. Setting is possible in steps of 0.1 s in scan interval more than 1 s. Note: When a scanning mode with patient couch movement is used, the minimum scan interval is limited by the time for movement. Scan start time delay: Min. 0.5 s Setting is possible in steps of 0.1 s. Maximum scan rate: Max. 133 scans/100 s (0.75 s scan, 133 rotations) Image reconstruction Full image: 1 image/scan Partial image: Reconstruction is possible in steps of 0.1 s. Reconstruction time: Min. 1.2 s Real time reconstruction time: 8 images/s (1 slice/512 512 matrix) 12 images/s (1 slice/256 256 matrix) Rapid sequence scan: Minimum scan cycle time Approx. 2.0 s at 10 mm pitch with 0.75 s scan

Multislice HELICAL CT SCANNER

MULTI

Multi-position dynamic scan:

Maximum number of 8 positions

Reconstruction filter functions:

X-ray generation
Fan shaped, fan angle 49.2 Continuous 80, 100, 120 and 135 kV 10 mA to 300 mA (36 kW)* 10 mA to 400 mA (48 kW) 10 mA to 450 mA (54 kW)* 10 mA to 500 mA (60 kW)* (in increments of 10 mA) X-ray tube heat capacity: 4.0 MHU (6.5 MHU)* X-ray tube cooling rate: Max. 864 kHU/min (924 kHU/min: 6.5 MHU)* Focal spot size IEC standard, nominal:(4 MHU) 0.9 mm 1.3 mm (small) 1.7 mm 1.6 mm (large) (6.5 MHU)* 0.9 mm 0.8 mm (small) 1.6 mm 1.4 mm (large) X-ray beam shape: X-ray exposure: X-ray tube voltage: X-ray tube current:

From FC01 From FC10 From FC20 From FC30 From FC40 From FC50 FC70 From FC80 FC90

Total of 99 available filter spaces Region Abdomen with BHC Abdomen without BHC Brain with BHC Inner ear and bone Brain without BHC Lung Maintenance Super resolution mode for inner ear, bone and lung High resolution mode

X-ray detection
Detection system: Main detector: Data acquisition: Reference detector: Viewing rate: Solid state detectors 896 channels 34 elements 896 channels 4 rows 1 set 1,200 views/s (max.)

Reconstruction matrix: 512 512 Picture element (pixel) size CT image: Scan field Pixel size Scanogram: Enlargement ratio (area) Pixel size Standard 2 4 SS S ** to ** to 0.35 0.47 M ** to 0.63

Data processing

Image reconstruction time CT scan: Min. 0.5 s (for 0.75 s scan) Scanoscope: Reconstructed and displayed simultaneously with scanning (real time reconstruction) Data processor Central processing unit: 64-bit microprocessor Memory size: 1 Gbyte Magnetic disk unit: Raw data, 36 Gbytes Image data, 9 Gbytes Magneto-optical disk unit: 5-inch type, 2.6 Gbytes (double sided)

Data storage
Unit: mm L LL ** to ** to 0.78 0.98 Unit: mm 8 16 Magnetic disk Raw data: Image data: Magneto-optical disk: Max. 4,000 rotations (8000 rotations)* (for 0.75 s scan) Max. 16,000 images (32,000 images)* Approximately 9,600 images or 600 (rotation) raw image data per cassette (2.6 Gbyte) 21-inch color 2 units 1280 1024 1024 1024 (max) From 1,024 to +8,191 Continuously variable (adjustable at variable speed) Continuously variable (adjustable at variable speed)

2.00 (LL) 1.41 1.0 0.71 0.50 1.00 (L) 0.71 0.50 0.50 (S)

Image display
Display monitor: Display matrix: Image matrix: CT number display range: Window width: Window level:

* Option ** Depending on Vari-Area or Zoom factors

Product Data
No. MPDCT0170EAB

Preset window: Window types: Image retrieval Method: Mode: Autoview function: Multi-frame display:

3 kinds of window settings may be preset for each image. Linear, non-linear (6 userprogrammable), and double windows. On-screen menus and keyboard Image, series and patient Software control, function key Reduction/cut-off display, ROI processing

Inset scanogram display Selective related information display Cine display Number of frames: Up to 128 Image display speed: The display speed can be varied up to a maximum of 20 frames/s.

Top/bottom, right/left, black/white reversal of image Image filtering Screen save High speed axial interpolation MultiView (Auto MPR) Raw data processing Multi-zooming (processing of raw data for 4 ROIs and multiple images) Stack reconstruction Protect/Unprotect Play/Reverse reconstruction System management Warm-up Calibration data Picture retouching Scanogram/CT image switching

Clinical application
Dynamic study A mathematical function is fitted to the time curve generated for the designated ROI within an image, and the functional parameters are displayed. ROI shape: Rectangular, polygonal, elliptical, irregular Fitting function: Gamma function (Perform least squares approximation.) Functional image In addition to processing images, Asteion can create functional images or image maps as well. Parameters which can be used to generate a functional image: Peak height Peak time Appearance time Area under curve Mean transit time 2nd moment Transit time

Image processing
Scanogram processing Slice position display (display of planned slice, preset slice, and last scanned slice) Anatomical scale (display of relative position, taking any point selected as standard as 0) Slice position setting Enlargement (2, 4 for L and LL size; 8, 16 for LL size) CT image processing ROI setting and processing ROI shape: Point, rectangular, polygonal, elliptical, irregular ROI processing: Mean value, standard deviation, area, number of pixels ROI display: 3 ROIs can be displayed on an image ROI control: Size, position, rotation Measurement of distance and angle between two points Profile (oblique profile also available) Histogram CT number display Mark display (grid display, scale display) Volume calculation Enlargement (2, 4 and arbitrary size) Addition/subtraction between images Band display (non-linear windowing) Comment and arrow insertion
6

Multislice HELICAL CT SCANNER

MULTI

3D color image processing


High quality 3D images can be obtained very quickly with easy operation. 3D surface rendering Clipping, texture or non-texture 3D volume rendering Maximum intensity projection (Max-IP) Minimum intensity projection (Min-IP) X-ray volume rendering Intensity volume rendering Shaded volume rendering (opacity curve can be set arbitrarily) Display/processing function Zooming, panning, measurement (distance, angle), annotation, cutting, drilling Cine display Segmentation Partial extraction can be performed MPR 3 orthogonal planes/oblique image Curved MPR

Spatial resolution: Scan parameter Tube voltage: Tube current: Scan time: Slice thickness: Scan field: Function type: Phantom: High contrast resolution High resolution mode (FC90): Standard mode (FC30): Scan parameters Tube voltage: Tube current: Scan time: Slice thickness: Scan field: Phantom: Low contrast resolution:

14.5 lp/cm at MFT2% 18 lp/cm at cut off 120 kV 200 mA 1s 2 mm S FC90 16 cm CATPHAN 0.35 mm 0.05 mm 0.55 mm 0.05 mm 120 kV 200 mA 0.75 s 2 mm S (use of zooming) Toshiba standard 2.5 mm at 0.25% or 4 mm 3 mm at 0.3% at 0.2% 120 kV 330 mA* 0.75 s* FC41 10 mm S 16 cm CATPHAN 120 kV 200 mA 0.75 s FC41 10 mm S 16 cm CATPHAN

Image transfer
Toshiba protocol DICOM3 storage SCU

Auto filming
Ethernet network: Toshiba protocol DICOM PRINT Scan parameters Tube voltage: Tube current: Scan time: Function type: Slice thickness: Scan field: Phantom:

IMAGE QUALITY
Noise Standard deviation: Scan parameters Tube voltage: Tube current: Scan time: Function type: Slice thickness: Scan field: Phantom: Less than 0.25% 120 kV 330 mA* 0.75 s* FC41 10 mm S 16 cm CATPHAN Less than 0.5% 120 kV 300 mA 1.5 s FC70 10 mm S 24 cm Water

* 250 mAs in the 36-kW system

* 250 mAs in the 36-kW system

Product Data
No. MPDCT0170EAB

SYSTEM COMPONENTS AND THEIR FUNCTION


Gantry
The gantry and the patient couch constitute the scanner. The scanner uses a fan-shaped continuous X-ray beam to scan the region to be examined. Transmitted X-rays are detected and converted into electric signals by SSMD. The gantry consists of the main body and its support mechanism. The X-ray tube and the SSMD are mounted facing each other on either side of the gantry aperture. The X-ray tube and detectors rotate continuously about the aperture of the gantry. A slip-ring is employed to transmit power between the gantry and the rotating X-ray high-voltage generator assembly. The gantry may be tilted forward and backward in order to perform tilt scanning. Three-dimensional alignment lights are provided for setting slice positions. Gantry and patient couch operating controls are located on both sides of the front of the gantry housing. The patient guide display indicates the scan status to the operator and the patient. X-ray high-voltage generator is built-in the gantry. The system employs a high-frequency inverter for generating and stabilizing the high voltage supplied to the X-ray tube. Included in the generator electronics are circuits for controlling the speed of the rotating anode in the X-ray tube. Use of a high-frequency inverter system results in high power output coupled with excellent stability. In addition, the system is compact and light weight.
G

Scanoscope control Scan control Remote control of couchtop movement Remote control of gantry tilt Functions of the console for scanning and image processing Window level and window width adjustment Other various mouse-operated image processing functions

Magneto-optical disk unit


The magneto-optical disk unit permits repeated recording and retrieval of data, like a floppy disk drive. It is a compact, externally mounted, large-capacity digital storage device that can be used to save a large amount of data. Images can be automatically archived to the magnetooptical disk unit via the eXam Plan.

OPERATING FEATURES
Patient handling and positioning
The couchtop can be lowered to within 300 mm of the floor thus simplifying patient transfer to and from a bed or stretcher. Fast and accurate patient positioning is assisted by alignment lights mounted on the gantry aperture. High-precision couchtop positioning is possible from the integrated console or by manual operation from the control panel and clear digital readouts are located on the gantry. Patient observation system (optional) in the operator's room ensures patient safety during scan procedure. The couchtop can be pulled out manually in case of an emergency.

Patient couch
The patient couch is located in front of the gantry and supports the patient. The entire unit moves vertically and the top moves longitudinally. In case of an emergency, the couchtop may be pulled out manually with very little effort. The couchtop can be lowered to a minimum height of 30 cm from the floor. This facilitates transferring the patient from a low bed or stretcher.

Scanning
Toshiba Scanoscope provides a radiographic-like fullview image of the patient for high-precision advance planning of the slice positions. The axial length of the scanning field for Scanoscope can be adjusted up to 1,390 mm at any time even while monitoring the image being reconstructed in real time. This minimizes patient radiation dosage. The auto index functions allows automatic incremental couchtop movement based on the slice positions determined while viewing the Scanogram.

Console
The two stand-alone consoles (one for scanning, and the other for image processing) are both provided with a hybrid keyboard, monitor and mouse. These consoles can be operated independently, and simultaneously parallel processing can significantly improve the efficiency of image processing and diagnosis. G Functions of the console for scanning Selection of scan parameters
8

Multislice HELICAL CT SCANNER

MULTI

The eXam Plan function allows simple selection of pre-programmed scanning parameters for routine examinations and for efficient patient throughput. The Vari-area function allows the user to pre-select a region of interest for zooming using raw data and immediate post-scan analysis. Zooming using raw data yields higher resolution than enlarging an already reconstructed image. Dynamic and rapid sequence scan modes. Multislice Helical Scan acquires raw data by rotating the X-ray tube continuously while moving the patient continuously through the scanner. A volume of data is acquired from which slices can be reconstructed at any point along the patient axis. This scan mode is best used for rapid patient scanning within one breath hold and for high definition three-dimensional and MPR imaging. Realtime helical reconstruction mode makes it possible to observe the images being scanned in real time at maximum 12 frames per second. This mode shows shift in the slice position in real time and helps the technician to check the scan field on the image, contrast study timing, presence of patient's body motion, etc. Therefore the patient can be released immediately after scanning. SureStart function helps the technician to initiate the helical scanning at peak enhancement timing of contrast study. SureStart starts monitoring the scan from the start of contrast study at a certain slice position, to monitor the change in CT number on the image being displayed in real time. When the contrast reaches the set value, the helical scan automatically starts. This technique provides contrast study at peak enhancement timing regardless of each patient's blood flow speed and also saves contrast medium.

Image display and processing


Reconstructed images are automatically displayed according to window settings preset in the eXam Plan. The window save function allows the user to store an image with window settings different from the ones set by the eXam Plan. Filter parameters can be customized through simple on-screen menu selections. These parameters include the number of filtering passes, matrix size and filter coefficients. Images can be rotated and reversed either right/left, top/bottom, or black/white. The Multi-frame feature allows up to 36 images to be retrieved and simultaneously displayed on the screen. The three-dimensional image display allows the generation of color three-dimensional and real time MPR images from the volumetric scan data acquired using Helical Scan. This results in higher definition and image quality than images reconstructed from conventional single slice scanning. This is because helical scanning achieves better data continuity along the patient axis than conventional scanning.

Image storage and archiving


The system comes with a 45 Gbyte magnetic hard disk as standard equipment allowing for the on-line storage of approximately 16,000 images with max. 4,000 rotations raw data. A 2.6 Gbyte magneto-optical disk is provided as standard equipment. A user-selectable automatic archive function can be enabled to automatically store images on the magneto-optical disk after reconstruction. The image storage capacity of the magnetooptical disk is approx. 9,600 images per disk.

Image filming Data processing


Rapid reconstruction reduces reconstruction time to as short as 0.5 second for 0.75 second scans. (Scan & View mode) A variety of reconstruction algorithms are available according to the anatomy being studied and the clinical objectives. These include algorithms for the abdomen, head, bone, lung, high-resolution, soft tissue, etc. Filming of images can be carried out manually or automatically from the console. Automatic filming sends an entire study to the laser camera. Filming is carried out in background mode so that other scanner and image processing functions can be performed without interruption or delay.

Patient throughput
Patient throughput and cost effectiveness have been given great consideration in designing and building the Aquilion CT scanner. The system incorporates a 4.0-MHU X-ray tube with a very fast cooling rate of max. 864 kHU/min.
9

Product Data
No. MPDCT0170EAB

High speed scans can be conducted in as little time as 0.5 second per scan. Routine scans can be performed as quickly as 0.75 second per scan. In multislice helical scanning, the scanning can be performed at maximum 8 slices per second. Real time reconstruction for scanoscope. CT image reconstruction in 0.5 second for 0.75 second routine scans.

Routine scan cycle times as short as 1.8 seconds. (Scan & View mode) Ease of operation with an optimized use of hybrid keyboard, mouse-driven menus and large color monitor. Convenient patient handling with a couchtop that can be lowered very near the floor.

DIMENSIONS AND MASS

Unit Gantry Patient couch Console (when a recommended dedicated desk is used) Navibox

Dimensions L W H mm (in) 890 (35.0 2,390 (94.1 800 (31.5 420 (16.5 1,970 77.6 620 24.4 1,600 63.0 1010 39.8 1,760 69.3) 390 15.4) 1,225 48.2) 700 27.6)

Mass kg (lb) 1,300 (2,866) 330 (728) 220 (485) 135 (298)

SITING REQUIREMENTS
Power requirements
Phase: Three-phase Voltage: 200 V* Frequency: 50 Hz or 60 Hz 0.5 Hz Capacity: 75 kVA*** Voltage fluctuation due to load change: Less than 5% Power voltage fluctuation: Less than 10%** * Please consult Toshiba for solutions to other voltages and excessive power fluctuations. ** This represents the total voltage fluctuation of load variation and power. *** When installing the optional 60-kW generator, please pay attention to the power capacity. Refer to the site planning guide MOICT0046EA or the site planning manual No. 2A201-384E for details.
10

Grounding
In accordance with local regulations for medically used electrical equipment.

Ambient conditions
Scanner room: Room temperature: Relative humidity: (no condensation) Operator's room: Room temperature: Relative humidity: Gantry Patient couch 18C to 28C 40% to 80% Console 16C to 28C 40% to 80% (no condensation)

Multislice HELICAL CT SCANNER

MULTI

OUTLINE DRAWINGS

Gantry and Patient Couch

990 (39.0) 890 (35.0) 354 (13.9) 536 (21.1)

30 30 (TILT ANGLE)

1,760 (69.3)

1,970 (77.6)

Gantry
Unit: mm (in)

910 (35.8)

1,900 (74.8)

11

Product Data
No. MPDCT0170EAB

Multislice HELICAL CT SCANNER

MULTI

OUTLINE DRAWINGS
470 (18.5) 620 (24.4) 1,820 (71.7) 2,000 (78.7)

Couchtop stroke
1,490 (58.7) 2,390 (94.1) 390 570 (15.4) (22.4) 960 (37.8) 420 (16.5)

870 300 (34.3)

(11.8)

Patient Couch

1,600 (55.1)

800 (31.5)

Navibox

1,010 (39.8)

700 (27.6) 800 (31.5)

1225 (48.2)

Stroke

Console

Unit: mm (in)

Note: Console desk is not included as standard. The photograph of the system on the cover differs form the drawings above.

TOSHIBA CORPORATION MEDICAL SYSTEMS COMPANY

http://www3.toshiba.co.jp/medical
Toshiba Corporation 2000 all rights reserved. Design and specifications subject to change without notice. 2000-12 TME/SB
Toshiba Corporation Medical Systems Company meets internationally recognized standards for Quality Management System ISO 9001, ISO 13485, EN 46001. Toshiba Nasu Operations meets the Environmental Management System standard, ISO 14001.

Printed in Japan