Flow measurement
On/Off switching
according to flowrate
Monitoring
cooling
systems
Monitoring
cooling
systems
2
Functions Measuring principle
In accordance with Faraday’s law of Microprocessor-controlled electronics,
electromagnetic induction, a voltage (Ue) utilizing patented “Integrated Autozero
is induced in a conductor that is moved Circuit” guarantees a stable zero-point
through a magnetic field. In the electro- and converts the voltage into an analog
magnetic principle of measurement the 4 to 20 mA scaled output signal.
flowing and electrically conductive fluid
represents the moving conductor. The
induced voltage is proportional to the
flow velocity and is fed to the amplifier by
a pair of electrodes.
Sensor
Magnetic lines
of force
B Exciter coil
Magnetic core
Ue
D
Electrodes
Velocity V
Ue = B x V x D
Removable housing
cover for access to
wiring terminals,
indicators and operating
elements
1/2" NPT
conduit entries
External Electronics housing
ground
terminal
Plastic
Union nut union nut
M30 x 2 (supplied with
Magphant)
SS Sensor
Adapter piece
(supplied with
PVDF tip
≥12"
10"
Magphan for
and SS Electrodes 8" platic pipe onlyt)
6"
4"
3"
1.5/2"
3
Mounting Instructions Orientation in the piping
Preferred Mounting
For best results, the Magphant should be
mounted in vertical piping. A minimum of
5 to 10 straight pipe diameters before
and after the sensor should be main-
tained to ensure an even flow over the
electrodes.
45°
Should horizontal piping be used, the
sensor should be mounted to the side or
downwards (45° below horizontal) to
ensure the electrodes are always 5 to 10 x pipe diameters pipe length
recommended for inlet and outlet
immersed in flowing fluid.
M 30 x 2 M 30 x 2
Plastic
union nut
(supplied with
3.15" Magphant)
(80)
3.15" ≥12"
(80)
Adapter piece
10" (supplied with
8" Magphant)
6"
4"
3"
1.5"/2"
Steel pipe version, 1-1/2” and NOTE: The Magphant for steel piping
larger cannot be used in plastic pipe. The
For 1-1/2” piping and larger, the weld insertion length is different and will not
stub has a graduated scale marked for measure properly.
proper insertion according to pipe
diameter. The opening in the pipe to The Magphant can be installed in
insert the weld stub is 1.18” (30 mm). existing T-fittings (or user-supplied T-
Ensure the weld stub is perpendicular to fittings) where other styles of insertion
the axis of the piping before welding. type instruments are being used.
4
Electrical Connection Warning!
Do not install, wire or dismantle the
instrument when the power supply is
switched on.
27 - 4 to 20 mA output
26 + Max. 750Ω load
25
SPDT Relay output
24
De-energized failsafe
23 condition
2 - 20 to 30 VDC
1 + Power supply
Terminal Block
1 2 23 24 25 26 27
Take a test drive: Free simulation software which fully simulates the operation of the
Magphant is available from Endress+Hauser on diskette, or can be down loaded from
the Endress+Hauser web site at www.us.endress.com.
5
Dimensions 3.27"
(83)
1.32"
(38.5)
Pipe Diameter Dia. L L
in. mm in. mm
1 25 9.23 237
1-1/2 to 2 40/50 9.21 234 1.57"
3 89 9.06 230 (40)
4 100 8.94 227
6 150 8.68 220.5
8 200 8.44 214.5
10 250 8.17 207.5
≥ 12 300 7.93 201.5
3.90"
5.91"
(99)
(150)
0.63" 3.39"
(16) (86)
1.73"
(44) 2.58"
(65.5)
2.95"
(75)
3.66"
(93)
L
9.61" 7.03"
(244) (178.5)
0.87"
(18.5)
0.73"
(22)
Nominal Size
(T-fitting)
ID DN
1/2" 15
Metal union nut 3/4" 20
1" 25
1-1/4" 30
Plastic
1-1/2" 40
union nut
2" 50
(supplied with
Magphant)
Adapter piece
(supplied with
Magphant)
9.29"
T-fitting (236 mm)
(user-supplied)
ID (DN)
6
Power Supply 20 to 30 VDC (24 VDC nominal)
Technical Data
Power Consumption < 2.5 W
Outputs Current: 4 to 20 mA, active. The measuring
system can monitor flow in both directions, i.e.
bidirectional. The current output is always
positive. The relay responds in both flow
directions.
Current (mA)
20.5 mA
20 mA
Reverse Forward
Flow Flow
Flow Rate
Scaled full- (ft/s)
Scaled full-
scale value scale value
7
Ordering Information DTI 200 - 1 2 3 4 5
4 to 20 mA ENDRESS+HAUSER
F1 F2 F3
AC Power
DC input ESC DIAGS.
- + E