You are on page 1of 55

Analytical Method Validation

Analytical Method Validation

Why Method Validation is Important?

1. The purpose of analytical measurement is to get consistent,


reliable and accurate data.
Incorrect measurement results can lead to tremendous costs.
2. Equal importance for those working in a regulated and in an
accredited environment.
U.S. FDA, EMEA, EPA, AOAC, ISO

Analytical Method Validation

Background
NDA and ANDA must include the analytical procedures necessary
to ensure:
Identity, Strength, Quality, Purity, and Potency of the Drug Substances
and Drug product [21CFR 314.50(d)(l) and 314.94(a)(9)(i)]
Data to establish and reliability [21CFR 211.169(e) and 211.194(a)
(2)]

Analytical Method Validation

Validation is an Old Concept


But There are Many Problems
Lack of documented procedures and documented validation results
Sampling or Sample preparation step contribute to overall error.
Accessories and materials used for equipment qualification are not
qualified.
Limits for Operational Qualification
Lack of software validation and computer system validation
Qualification and validation are done at just one particular point in time.
Adaptation of acceptance criteria for qualification of new system
Analytical Method Validation

Validation Activity Including


the Complete Analytical Procedure
Sampling

Sample Preparation

Analysis

Data Evaluation Reporting


Analytical Method Validation

Optimization of Validation

Additional value and Cost VS. Completeness of validation


Analytical Method Validation

Considerations Prior to
Method Validation
Suitability of Instrument
Status of Qualification and Calibration
Suitability of Materials
Status of Reference Standards, Reagents, Placebo Lots
Suitability of Analyst
Status of Training and Qualification Records
Suitability of Documentation
Written analytical procedure and proper approved protocol with
pre-established acceptance criteria
Analytical Method Validation

Validation Step

Define the application, purpose and scope of the method.


Analytes? Concentration? Sample matrices?
Develop a analytical method.
Develop a validation protocol.
Qualification of instrument.
Qualify/train operator

Analytical Method Validation

Qualification of material.
Perform pre-validation experiments.
Adjust method parameters and/or
acceptance criteria if necessary.
Perform full validation experiments.
Develop SOP for executing the method in routine analysis.
Document validation experiments and results in the
validation report.
Analytical Method Validation

System Suitability
Validation
Calibration
Pump

Injector

Detector

Data System

Analyst

Method
Sample
Analytical Method Validation

10

Verification vs. Validation

Compendial vs. Non-compendial Methods


Compendial methods-Verification
Regulatory analytical procedure in USP/NF
Non-compendial methods-Validation
Alternative analytical procedure proposed by the applicant for
use instead of the regulatory analytical procedure

Analytical Method Validation

11

Regulations and Guidelines of


Validation
US FDA 21 CFR (Code of Federal Regulations) Part 210 and 211
Part 210: cGMP in Manufacturing, Processing, Packaging, or Holding of
Drugs; General
Part 211: cGMP for Manufacturing Practice for Finished Pharmaceuticals

ICH Guidelines
Q2A, Text on Validation of Analytical procedures
(March 1995)
Q2B, Validation of Analytical Procedures: Methodology (May 1997)

USP Chapter <1225>


Validation of Compendial Methods
Analytical Method Validation

12

The accuracy, sensitivity, specificity, and reproducibility of test methods


employed by the firm shall be established and documented. Such validation
and documentation may be accomplished in accordance with 211.194(a)
(2).

21 CFR PART 211 - CURRENT GOOD MANUFACTURING


PRACTICE FOR FINISHED PHARMACEUTICALS
Subpart I-Laboratory Controls
211.165 Testing and release for distribution (e)

Methods validation means establishing, through documented evidence, a


high degree of assurance that an analytical method will consistently yield
results that accurately reflect the quality characteristics of the product
tested.
21 CFR PART 210 - CURRENT GOOD MANUFACTURING
PRACTICE IN MANUFACTURING, PROCESSING, PACKING,
OR HOLDING OF DRUGS
210.3 Definitions (b) (25)

Analytical Method Validation

13

The objective of validation of an analytical procedure is to


demonstrate that it is suitable for its intended purpose

ICH Guideline for Industry Q2A

In practice, it is usually possible to design the experimental work such


that the appropriate validation characteristics can be considered
simultaneously to provide a sound, overall knowledge of the capabilities of
the analytical procedure, for instance: Specificity, Linearity, Range,
Accuracy, and Precision.

ICH Guideline for Industry Q2B


Analytical Method Validation

14

ICH/USP Validation Requirements

Precision
Specificity
Linearity
Range
Accuracy

Repeatability
Intermediate Precision
Limit of Detection
Limit of Quantitation
Robustness

Analytical Method Validation

15

USP Data Elements Required For Assay Validation


Analytical
Performance
Parameter

Assay Category 2
Assay Category 1

Quantitative

Limit Tests

Assay
Category 3

Accuracy

Yes

Yes

Precision

Yes

Yes

No

Yes

Specificity

Yes

Yes

Yes

LOD

No

No

Yes

LOQ

No

Yes

No

Linearity

Yes

Yes

No

Range

Yes

Yes

Ruggedness

Yes

Yes

Yes

Yes

* May be required, depending on the nature of the specific test.


Category 1: Quantitation of major components or active ingredients
Category 2: Determination of impurities or degradation products
Category 3: Determination of performance characteristics
Analytical Method Validation

16

ICH Validation Characteristics vs.


Type of Analytical Procedure

Type of Analytical
Procedure

Impurity testing
Identification

Assay

Quantitative

Limit Tests

No

Yes

No

Yes

Repeatability

No

Yes

No

Yes

Interm. Prec.

No

Yes

No

Yes

Specificity

Yes

Yes

Yes

Yes

LOD

No

No

Yes

No

LOQ

No

Yes

No

No

Linearity

No

Yes

No

Yes

Range

No

Yes

No

Yes

Accuracy
Precision

Analytical Method Validation

17

Specificity
Ability of an
analytical method to
measure the analyte
free from
interference due to
other components.

Selectivity
Bias

Analytical Method Validation

18

Specificity: ICH/USP
An investigation of specificity should be conducted
during the validation of an identification test, an
impurities assay, and a potency assay.
Procedures used will depend on the intended
objective of the analytical procedure.
If a method can not completely discriminate, two of
more procedures are recommended.

Analytical Method Validation

19

Specificity: Identification
Should be able to discriminate between compounds
closely related in structure.
Confirmed by obtaining negative results for samples
with spiked related compounds and positive results
for samples with analyte.
Choice of potential interfering substances should be
based on sensible scientific judgment considering
substances that could likely occur.
Analytical Method Validation

20

Specificity: Impurities/Assay
Chromatographic Methods
Demonstrate Resolution

Impurities/Degradants Available
Spike with impurities/degradants
Show resolution and a lack of interference

Impurities/Degradants Not Available


Stress Samples
For assay, Stressed and Unstressed Samples should be compared.
For impurity test, impurity profiles should be compared.

Analytical Method Validation

21

Pure and Impure HPLC peaks

Peak purity tests can also be evaluated with


The spectra of Photodiode array detectors
Mass spectrometry

Analytical Method Validation

22

Specificity: Potential Interference

Placebo
Drug Substance Degradants
Drug Product Degradants
Related Substances
Packaging Extractables

Analytical Method Validation

23

Forced Degradation Studies


Heat

High Temperature (50 to 60 oC)

Humidity

Humidity (70 to 80%)

Acid Hydrolysis

Acid Hydrolysis (0.1 N)

Base Hydrolysis

Base Hydrolysis (0.1 N)

Oxidation

Peroxide Oxidation (3 to 30%)

Light

Intense UV/Visible Light

Intent is to create 10 to 30 % Degradation


Analytical Method Validation

24

Specificity Study

Condition

DEG
#1

DEG
#2

DEG
#3

DEG
#4

Active
ingredients

2 N HCl

17.23

4.71

95.17%

0.1 N
NaOH

100%

30%
H2O2

1.12

1.41

1.65

1.2

99.98%

UV/Vis

4.68

94.67%

Analytical Method Validation

25

Linearity
Ability of an assay
to elicit a direct and
proportional
response to
changes in analyte
concentration.
Analytical Method Validation

26

Linearity Should be Evaluated


By Visual Inspection of plot of signals vs. analyte
concentration
By Appropriate statistical methods
Linear Regression (y = mx + b)
Correlation Coefficient, y-intercept (b), slope (m), residual
sum of squares

Requires a minimum of 5 concentration levels


Analytical Method Validation

27

Linearity Example

R square = 0.999
Slope = 0.97
y-intercept = 0.233
Line Eq.: Y = 0.97X + 0.233
Std. Error = 1.319
Std. Deviation of Slopes = 0.0079
Analytical Method Validation

28

Range
The interval between the
upper and lower
concentrations of analyte
in the sample that have
been demonstrate to have
a suitable level of
precision, accuracy, and
linearity.
Analytical Method Validation

29

Range

Normally derived from Linearity studies.


Established by confirming that the method provides
acceptable degree of linearity, accuracy, and
precision.
Specific range dependent upon intended application
of the procedure.

Analytical Method Validation

30

Minimum Specified Range:


For Drug Substance & Drug product Assay
80 to 120% of test Concentration

For Content Uniformity Assay


70 to 130% of test Concentration

For Dissolution Test Method


+/- 20% over entire Specification Range

For Impurity Assays


From Reporting Level to 120% of Impurity Specification for Impurity
Assays
From Reporting Level to 120% of Assay Specification for
Impurity/Assay Methods
Analytical Method Validation

31

Accuracy

Closeness of the test


results obtained by the
method to the true
value.

Analytical Method Validation

32

Accuracy
Should be established across specified range of analytical
procedure.
Should be assessed using a minimum of 3 concentration levels,
each in triplicate (total of 9 determinations)
Should be reported as:
Percent recovery of known amount added (reference material) or
The difference between the mean assay result and the accepted
value
Analytical Method Validation

33

Accuracy Data Set (1 of 3)

Amount
Added (mg)

Amount
Found
(mg)

Percent
Recovery

0.0

0.0

---

50.2

50.4

100.5

79.6

80.1

100.6

99.9

100.7

100.8

% Recovery

% Recovery

99.2

98.9

99.3

99.3

99.4

99.7

99.3

99.3

Std.dev.

0.1

0.4

95%C.I

99.30.25

99.30.99

Mean

Analytical Method Validation

34

Analyte recovery at different


concentration
Analyte Ingred. (%)

Analyte ratio

Unit

Mean recovery (%)

100

100 %

98-102

10

10-1

10 %

98-102

10-2

1%

97-103

0.1

10-3

0.1%

95-105

0.01

10-4

100 ppm

90-107

0.001

10-5

10 ppm

80-110

0.0001

10-6

1 ppm

80-110

0.00001

10-7

100 ppb

80-110

0.000001

10-8

10 ppb

60-115

0.0000001

10-9

1 ppb

40-120

AOAC manual for the Peer-Verified Methods program


Analytical Method Validation

35

Precision
The closeness of agreement

Ball Ball
Ball
Ball Ball Ball
Ball Ball

Strik Strike
Strike
e Strike
Strike
Strike

(degree of scatter) between a


series of measurements obtained
from multiple samplings of the
same homogeneous sample.
Should be investigated using
homogeneous, authentic samples.

Analytical Method Validation

36

Precision Considered at 3 Levels

Repeatability
Intermediate Precision
Reproducibility

Analytical Method Validation

37

Repeatability
Express the precision

Should be assessed

under the same

using minimum of 9

operating conditions

determinations

over a short interval

(3 concentrations/ 3

of time.

replicates) or

Also referred to as

Minimum of 6

Intra-assay precision

determinations at the
100% level.
Analytical Method Validation

38

Intermediate Precision
Express within-laboratory variations.

Depends on the

Expressed in terms of standard

circumstances under which

deviation, relative standard

the procedure is intended

deviation (coefficient of variation)

to be used.

and confidence interval.

Studies should include

Known as part of Ruggedness in

varying days, analysts,

USP

equipment, etc.

Analytical Method Validation

39

Repeatability & Intermediate Precision


Day 1

Day 2

100.6

99.5

100.8

99.9

100.1

98.9

100.3

99.2

100.5

99.7

100.4
Mean = 100.5
RSD = 0.24%
CI = 100.5 0.24

99.6
Mean = 99.5
RSD = 0.36%
CI = 99.5 0.36

Grand
Mean = 100.0
RSD = 0.59%

Analytical Method Validation

40

Reproducibility

Definition: Ability reproduce data within the


predefined precision
Determination: SD, RSD and confidence interval
Repeatability test at two different labs.

Note: Data not required for BLA/NDA

Analytical Method Validation

41

Reproducibility Study
Lab 1

Lab 2

Lab 3

Day 1

Day 2

Day 1

Day 2

Day 1

Day 2

Analyst
1

Analyst
2

Analyst
1

Analyst
2

Analyst
1

Analyst
2

3 Preps

3 Preps

3 Preps

3 Preps

3 Preps

3 Preps

Analytical Method Validation

42

Analyte concentration versus precision


Analyte %

Analyte ratio

Unit

RSD (%)

100

100 %

1.3

10

10-1

10 %

2.7

10-2

1%

2.8

0.1

10-3

0.1%

3.7

0.01

10-4

100 ppm

5.3

0.001

10-5

10 ppm

7.3

0.0001

10-6

1 ppm

11

0.00001

10-7

100 ppb

15

0.000001

10-8

10 ppb

21

0.0000001

10-9

1 ppb

30

AOAC manual for the Peer-Verified Methods program


Analytical Method Validation

43

Detection Limit (DL)

Quantitation Limit (QL)

Lowest amount of analyte in a

Lowest amount of analyte in a

sample that can be detected

sample that can be quantified

but not necessarily quantitated.

with suitable accuracy and

Estimated by Signal to Noise


Ratio of 3:1.

precision.
Estimated by Signal to Noise
Ratio of 10:1.

Analytical Method Validation

44

Detection Limit (DL) and Quantitation


Limit (QL) Estimated by
1. Based in Visual Evaluations
- Used for non-instrumental methods

2. Based on Signal-to Noise-Ratio


- 3:1 for Detection Limit
- 10:1 for Quantitation Limit

3. Based on Standard Deviation of the Response


and the Slope
Analytical Method Validation

45

Based on Signal-to Noise-Ratio

Analytical Method Validation

46

Detection Limit (DL) and


Quantitation Limit (QL)

DL =

3.3s
S

QL =

10s
S

S = slope of calibration curve


s = standard deviation of blank readings or
standard deviation of regression line
Validated by assaying samples at DL or QL
Analytical Method Validation

47

Robustness
Definition: Capacity to remain unaffected by small
but deliberate variations in method parameters
Determination: Comparison results under differing
conditions with precision under normal conditions
Variations may include: stability of analytical solution,
variation of pH in a mobile phase, different column
(lot/supplier), temperature, flow rate.

Analytical Method Validation

48

Confuse of Precision Terms

Repeatability

Reproducibility

Ruggedness
Intermediate
Precision

Robustness

Analytical Method Validation

49

Precision Terms

Instrument Precision

- 10 Std. Injections

Repeatability

- One Analysis (6 preps)

Intermediate Precision

- Two Analyses

Reproducibility

- Two Lab.

Ruggedness

- Many Variables

Robustness

- Intentional Changes
Analytical Method Validation

50

Robustness Variations

All Assays

-Sample Prep Manipulation


-Extraction Time

HPLC Assays

-Mobile Phase Composition


-Different Columns
-Temperature
-Flow Rate

GC Assays

-Different Columns
-Temperature
-Flow Rate
Analytical Method Validation

51

Robustness-Mobile Phase Change

MeOH/
Water

Retention
Time 1

Retention
Time 2

Resolution

75:25

11.94

16.41

7.39

80:20

8.47

11.17

6.17

85:15

7.81

10.18

5.93

Analytical Method Validation

52

ICH/USP System Suitability


ICH

USP 23 <621>
Definition: evaluation of
equipment, electronic,
analytical operations and
samples as a whole
Determination: repeatability,
tailing factor (T), capacity

System Suitability Requirements


Parameters

Recommendations

In general k 2.0

R > 2, between the peak of


interest and the closest
potential interferent
(degradant, internal STD,
impurity, excipient, etc..)

factor (k), resolution (R), and


theoretical Plates (N)

T2

In general N > 2000

Repeatability

RSD 2.0% (n 5)

Analytical Method Validation

53

Guidance on Re-Validation
When sponsors make changes in the analytical
procedure, drug substance, drug product, the
changes, may necessitate revalidation of the
analytical procedures.
The degree of revalidation depends on the nature of
the change.
FDA intends to provide guidance in the future on
post-approval changes in analytical procedures.
Analytical Method Validation

54

References
Analytical Methods Validation for FDA Compliance The Center for
Professional Advancement 2003. 3.12-14.
Guideline for submitting samples and analytical data for kethods validation (Feb.
1987)
ICH Q2A
ICH Q2B
21 Code of Federal Registrations Part 210 and 211
Michael E. Swatrz and Ira S. Krull, Analytical method development and validation.
Mrcel Dekker, Inc. New York, 1997.
USP 23 <1225>
http://www.waters.com
Ludwig Huber, Validation and Qualification in Analytical Laboratories, Interpharm
Press Inc. Buffalo Grove, Illinois, 1999.
Analytical Method Validation

55

You might also like